More Related Content Similar to Pc 400 Application _Ashraful (2).pptx (20) Pc 400 Application _Ashraful (2).pptx2. ABX Pentra C400
© 2018 HORIBA, Ltd. All rights reserved. 2
2018 PC400 PRESENTATION V1.11
7. 1 Physical specifications
2 Technical specifications
3 Computer specifications
Product Overview
© 2018 HORIBA, Ltd. All rights reserved. 7
8. ISE Module
Sample Tray
Reagent Tray
Sampling System
Windows XP
Cuvette Changer
➢ Bench-top analyzer
➢ Dimensions : 100 x 65 x 57 cm
➢ Weight : 120 kg
➢ Screen: 12 inches
© 2018 HORIBA, Ltd. All rights reserved. 8
Main features : Benefits
Physical Specifications
Product Overview
9. • Analysis method types:
- Spectrophotometry: Colorimetry and Turbidimetry,
- Potentiometry: Direct (Serum or Plasma) and Undirect (Urine)
• On-board Test Capacity: 55 parameters on board
- 52 cassettes positions or 39 tests positions with Reagent racks
- 3 ISE's (Na+, K+, Cl-)
• Sample Capacity (60 samples on board)
- 6 racks with 10 sample positions for primary tubes or cups
- Continuous loading
- Positive Sample Identification (integrated Barcode Reader and tube presence detector)
• Sample Needle : level detection, shock detection and clot detection
• 52 Reagents sectors : 44 refrigerated (2-10°C)
• Reagent Needle: Capacitive Level detection & Preheating
Main features : Benefits
© 2018 HORIBA, Ltd. All rights reserved. 9
Product Overview
10. • Hardware characteristics
Processor : Atom™ 1.46 GHz
Hard drive : 250 GB
RAM : 4GB
RS232C, RJ145 LAN, USB 2.0 (7 ports), Audio
• Operating System
Windows 7 ™
• Software
FULL QC
Validation Station
User’s Channels
Logs
Interactive Help
Main features : Benefits
© 2018 HORIBA, Ltd. All rights reserved. 10
Product Overview
12. Software
COMPLETE SOFTWARE
Flexibility:
➢ Id. or pos.: if LIS or not (dedicated, Vet, Industry)
barcode tube or cup position
Easy to use:
➢ Backup: automatic reagent management
automatic recalibration if needed
➢ Create additional automatic filter: for Cal
➢ Automatic maintenance
Quality/traceability :
➢ Automatic data export
© 2018 HORIBA, Ltd. All rights reserved. 12
Product Overview
13. Software
COMPLETE SOFTWARE
Flexibility:
➢ Language: English, French, German, Italian, Spanish, Portuguese, Greek & Polis
Quality/traceability :
➢ Automatic database backup (8 working days max)
➢ Manual database backup (to help for support)
(for sales to follow activity)
Product Overview
© 2018 HORIBA, Ltd. All rights reserved. 13
14. Software
COMPLETE SOFTWARE
Easy to use :
➢ Mono or Bi directional : ASTM format
➢ Query and Requery possible
Support :
➢ Integrated technical support/tool
Product Overview
© 2018 HORIBA, Ltd. All rights reserved. 14
15. Flexibility:
➢ Printer: Customized printout (Logo, header,…)
➢ Printer installation: Only windows Driver required to get manual or
automatic printout
➢ QC Printout: with precision and/or accuracy curve
Easy to use and Secured :
➢ On board a complete validation station
➢ Westgard rules possible
➢ According LIS or not Set what to Review (Quality flag, Delta check,…)
➢ Automatic rerun out of Critical panic range
➢ Automatic post Dilution or post Concentration
Software
COMPLETE SOFTWARE
© 2018 HORIBA, Ltd. All rights reserved. 15
Product Overview
16. Software
COMPLETE SOFTWARE
Easy to use and Flexibility:
➢ Customized Audible alert to be well informed on time
➢ Audible message set up by: type, volume intensity,
language,..
Easy to use and secured:
➢ Maintenance Alert can be set according use
➢ Most of the maintenance in automatic : daily & weekly
© 2018 HORIBA, Ltd. All rights reserved. 16
Product Overview
17. Quality & Traceability:
➢ Customized by setting the user name and access
➢ Protected instrument use : Automatic logoff after time can be set.
Flexibility:
© 2018 HORIBA, Ltd. All rights reserved. 17
➢ Up to 15 User’s channel confiration
(can be extended to 40 according to hardware modifications)
➢ Sales Tool : Test Counter
➢ Perishable management: use of expired (reagents, calibrators,…),…
Software
Product Overview
19. To create empty new
User’s Channel
(for Admin Level)
To see details or modify
unit, ranges,
Control required,…
To Copy a previous
methodology in a new
User’s Channel
(for Admin Level)
To Update/download new
Horiba’s Channels
© 2018 HORIBA, Ltd. All rights reserved. 19
To Change test order
(printout & Worklist
selection)
Product Overview
Application
21. Application
Details 1/4
Flexibility:
➢ Pretreatment (predilution) ½ up to 1/22 500
➢ Post Dilution / Post Concentration
➢ Unit selection protected (automatic calculation)
➢ Manual validation Possible (for special test independant
than validation station)
➢ Test name according User’s preference
➢ And activate only required test
Product Overview
© 2018 HORIBA, Ltd. All rights reserved. 21
22. Security and rules:
➢ Delta Check or anteriority check
➢ Normal or reference range
➢ Critical rerun range (automatic rerun)
Standardization:
Instrument correlation possible
Product Overview
© 2018 HORIBA, Ltd. All rights reserved. 22
Application
Details 1/4
24. Cleaner before or after R1 R1+S R2
Application
Details 2/4
Sequence & wavelength
15 Wavelengths: mono or bi : 340nm …700nm
© 2018 HORIBA, Ltd. All rights reserved. 24
Product Overview
25. Time
In second
Reagent
Needle
dispensing
Cycle time 12 second:
■ Volume flexibility: 2µL up to 95µL for Sample Needle
15 up to 600µL for Reagent Needle
➢ Dispensing flexibility: in any cycle new solution (reagent or sample or diluent)
can be dispensed by Needles
➢ Mixing Speed Flexibility: can be set from 0 up to 100%
Sample Needle dispensing
0
Absorbance
measurement
In Cycle
0
(R1 alone
reading)
12
1
(R1 + S
reading)
24 36 48
2 3 4
Product Overview
© 2018 HORIBA, Ltd. All rights reserved. 25
Application
Details 2/4
26. Reagent 1
Reagent 2
Sample Reagent 3
Reagent 4
0 12 24 36 48
Time
In second
Product Overview
© 2018 HORIBA, Ltd. All rights reserved. 26
Application
Details 2/4
27. Type: ENDPOINT / KINETIC / KINSEARCH
Reading point : 0 to 99 cycles (12s)
Product Overview
© 2018 HORIBA, Ltd. All rights reserved. 27
Application
Details 3/4
28. Sample Quality:
➢ Sample limit (interferences)
➢ Reaction limit (substrat depletion)
➢ Antigen excess (hook effect or Prozone)
Reaction Quality:
➢ OD deviation check (div_abs)
➢ Reaction direction (Sign)
Product Overview
© 2018 HORIBA, Ltd. All rights reserved. 28
Application
Details 3/4
29. Flexibility: All standard of Biochemistry from Enzyme to DAT &TDM
➢ 3 Calibrator predilutionType
(Factor, Main Direct & Main indirect)
➢ Calibration Mode:
Linear: - Factor
- Slope avg
- Linear regression
Non linear: - Linear inter
- Logit/log4
- Logit/log 5
- Exponen 5
Product Overview
© 2018 HORIBA, Ltd. All rights reserved. 29
Application
Details 4/4
30. ➢ Deviation
➢ Factor Checks:
Calibration & calibrator Quality:
Reagent Quality:
➢ Reagent range & Blank range checks:
© 2018 HORIBA, Ltd. All rights reserved. 30
Product Overview
Application
Details 4/4
31. Application
PC400 Ratio
Ratio: set calculation between tests
➢ Flexibility : Used when the PC400 is not connected to a LIS
© 2018 HORIBA, Ltd. All rights reserved. 31
Product Overview
32. Application
PC400 Profiles
PROFILES: list of tests pre-programmed in the worklist
➢ Flexibility : Used when the PC400 is not connected to a LIS
© 2018 HORIBA, Ltd. All rights reserved. 32
Product Overview
34. Application
PC400 Reagent
To change reagent use in Vial or
Cassette
To create new Reagent or Diluent for
User’s channel
Reagent: Select test in cassette or Vial + create new reagent
➢ Flexibility : Reagent format use (small or big consumption)
User’s Channel configuration (Open Channel)
© 2018 HORIBA, Ltd. All rights reserved. 34
Product Overview
36. PC400 Test Menu
Advantages coming from
HORIBA Medical expertise in CC
More than 10 years with the Pentra family
• A complete range ready to use
• Reagent menu allowing the flexibility with 75
applications including:
✓ Enzymes & substrats
✓ Ions & specific proteins as Ferritin, Transf.,
Hapto,…
✓ CRP & CRP-HS
✓ HbA1C
✓ …
Performances
© 2018 HORIBA, Ltd. All rights reserved. 36
37. Panels example:
• Hepatic: Total Bilirubin, Direct Bilirubin, Transaminase
ASAT & ALAT, ALP & GGT
• Glucidic: Glucose, HbA1c, Fructosamin
• Lipidic panel: Cholesterol, HDL Chol, LDL Chol, Trigly,
Apo A1 & Apo B
• Bone: Calcium, Phosphorus, ALP
• Cardiac: CK, CK-MB ASAT, LDH, Myoglobin, CRP HS
• Renal: Urea (BUN), Creatinine, Uric Acid, Micro Albumin
• …
© 2018 HORIBA, Ltd. All rights reserved. 37
Performances
42. Test Equation
© 2018 HORIBA, Ltd. All rights reserved. 42
r2 correlation coefficient
0.9964
ALT Y = 1.00 X + 5.00 (U/L)
CHOL Y = 0.95 X + 0.04 (mmol/L) 0.993
GLUP Y = 0.98 X + 0.04 (mmol/L) 0.9974
CRP Y = 1.03 X - 0.18 (mg/L) 0.997
Correlation
Passing-Bablock regression
Performances
43. Reagent Stability and Sample Volume
© 2018 HORIBA, Ltd. All rights reserved. 43
Test Open Stability Sample Volume
ALT 42 Days 20 μL/test
CHOL 48 Days 3 μL/test
GLUP 83 Days 4 μL/test
CRP 64 Days 4 μL/test
Performances
44. Fast Instrument
• Albumin (BCG)
• Glucose (PAP)
• CRP (Latex turbidimetry)
• Complete BMP Panel
Performances
© 2018 HORIBA, Ltd. All rights reserved. 44
45. Reagent range extensions
• DAT & TDM are running on Pentra C400
14 DATs validated: Amphetamine, Barbiturate, Benzodiazepine,
Buprenorphine, Cannabinoid, Cocaine, Ecstasy, Ethyl Alcohol,
Methadone , Methadone Metabolite, Opiate, Oxycodone,
Phencyclidine & Propoxyphene
7 TDMs validated: Acetaminophene, Amikacine, Digoxine,
Gentamycine, Tobramycine ,Tricyclique (Antidepressant) &
Vancomycine
• Up to 15 additional User’s channels
(can be extended to 40 according to hardware modifications)
With Open cassettes 30/10 (ref A11A01922) and Vials can be
used
© 2018 HORIBA, Ltd. All rights reserved. 45
Performances
47. • On board Reagent Help
• Website
• International Technical Support
& Etools (For Distributors and subsidiaries only)
Documentation:
Support
© 2018 HORIBA, Ltd. All rights reserved. 47
48. • Communication catalogue
• Excel summary file (in Technical Notes)
• Etools Catalogue
(For Distributors and subsidiaries only)
Catalogue information:
Support
© 2018 HORIBA, Ltd. All rights reserved. 48