SlideShare a Scribd company logo
1 of 47
Hepatic Doppler US [2]
Dr. Kamal Sayed MBBS MSc US UAA
Color AND spectral parameters [contd..]
[3] Wall Filters
•
Wall filters @ selectively filter out all acquired information
below an operator-defined frequency threshold.
•
@ Filters eliminate the typically low-frequency–high-intensity
noise that may arise from vessel wall motion.
@ Filters operate at variable frequencies to eliminate signal from
low-velocity blood flow.
@ Filter settings are usually preset by the manufacturer, and a
high, medium, or low filter setting may be applied separately
to spectral, color, and power imaging.
•
“High” refers to the higher range of frequency shifts that are
filtered out and thus are not depicted on the color image or
spectral waveform.
•
In patients with very slow portal venous flow, a high filter
setting may cause this flow to be inadvertently
[unintentionally] obscured or missed.
•
To avoid the loss of signal that characterizes slow flow, filter
settings should be kept at the lowest possible setting
(typically in the 50–100-Hz range).
•
.
•
The cutoff frequencies vary with the velocity scale; the
lowest filter cutoff frequencies cannot be used with the
highest velocity ranges and vice versa.
•
On the color bar, the filter setting is indicated by black areas
on both sides of the baseline.
•
•
# Expansion of the filter shows up as a widening of the black
band (,,,Fig 10) slide [7]
# On the spectral waveform, a high wall filter setting will
•
result in loss of depicted spectral information immediately
above the baseline. Reducing the wall filter setting results in
filling in of spectral data toward the baseline.
•
•
Figure 9. Changing the wall filter.
•
(a) Color duplex US image [slide [7] obtained with a high wall
filter setting shows @ loss of the low-velocity-flow
component of the spectral waveform immediately above the
baseline. @ Higher-velocity flow is well depicted, and
@accurate flow quantification can still occur.
•
(b) Color duplex US image slide [8] demonstrates how the
spectral waveform progressively fills in toward the baseline
as the wall filter is sequentially reduced from high (left arrow)
to medium (middle arrow) to low (right arrow).
Figure 9 [a] Changing the spectral wall filter [high filter]
Figure 9 [b] Changing the spectral wall filter [reduced filter from high to med
to low = arrows]
•
Figure 10. Changing the color Doppler wall filter.
•
(a) CFD image slide [10] obtained with the highest possible
wall filter setting shows how color signal arising from low-
velocity flow may be filtered out.
•
## The change in the filter setting appears as a change in the
width of the horizontal black line in the center of the color
bar. Slide [10]
•
(b) CFD image slide [11] obtained with a low filter setting
demonstrates filling in of flow in the hepatic veins (blue),
which indicates minimal filtering of color signal.
•
Figure 10 [b] (high filter = filters out signals from low flow velocities)
The change in the filter setting appears as a change in the width of the
horizontal black line in the center of the color bar.
Figure 10 [b] filling in of flow in hepatic veins [blue] Changing the color
Doppler wall filter from high to low.
•
•
[4] Inversion of Flow figure 11
•
Inversion refers to the ability to electronically invert the
direction of flow as depicted on both the color flow and
spectral waveforms.
•
a) As such, color inversion will result in a blue-red reversal
and may lead to misinterpretation of the direction of flow in
the vessel being evaluated (Fig 11 a/slide 13).
•
b) On a color Doppler flow US image obtained with reversal of
this inversion, appropriate directional flow is noted.
•
[Figure 11 b/slide 14]
Figure 11a. Inversion of color flow. Image [color Doppler active mode] on
inversion of the color bar, portal venous flow appears blue, which falsely
suggests reversal of flow (ie, away from the transducer).
Figure 11b. Inversion of color flow. On a color Doppler flow US image
obtained with reversal of this inversion, appropriate directional flow is
noted.
•
Similarly, inversion of the spectral waveform will switch the
depicted flow curve from above to below (or below to above)
the baseline (,,,Fig 12).
•
Adjustment of the inversion button will alter either the color
or the spectral scale, depending on which is active at the time
of adjustment.
•
Flow toward the TXR typically appears red (at CFD) OR above
the baseline (positive flow) on the spectral waveform and can
be changed simply by adjusting the inversion button.
•
Figure 12. Inversion of spectral and color flow falsely suggesting
reversal of portal venous flow.
•
(a) On a color duplex image slide [17] obtained with spectral
Doppler as the active mode, @ the spectral waveform is
below the baseline, @ with appropriate color flow.
•
(b) Color duplex image slide [18] obtained after the inversion
button was reversed demonstrates @ appropriate directional
flow, with the spectral @ waveform now appearing above the
baseline.
•
Note that the color bar does not change when the Doppler
spectrum is inverted.
Figure 12 a. Inversion of spectral AND color flow : spectral [active mode], the spectral waveform
is below the baseline, with appropriate color flow.
Figure 12 b. Color duplex obtained after the inversion button was reversed
demonstrates appropriate directional flow, with the spectral waveform now appearing
above the baseline.
•
(B) Spectral-Specific Parameters
•
[angle correction/ spectral gain/ gate size/ gate position]
•
The spectral waveform contains a host of hemodynamic
information describing the velocity and character of the
blood flow in the specific vessel being insonated.
•
The waveform depicts the spectrum of frequency shifts of all
blood traversing the sampled volume during the period of
data acquisition . With use of the computer to apply an angle
correction, a range of time-dependent velocities are
displayed on the vertical axis (spectral display).
•
[1] angle correction
•
Angle correction refers to adjustment of the Doppler angle
and is used to calibrate the velocity scale for the angle
between the US beam and the blood flow being measured.
•
Spectral Doppler US provides both qualitative and
quantitative data about flow velocity.
•
Flow velocity is calculated from the Doppler frequency shift
according to the Doppler equation :
•
V = ΔfC/2focosθ
•
Flow velocity is calculated from the Doppler frequency shift
according to the Doppler equation :
•
V = ΔfC/2focosθ
•
where
•
V = velocity.
•
Δf = Doppler frequency shift.
•
C = speed of sound in soft tissue.
•
fo = original transmitted frequency.
•
θ = the angle between the transducer and blood flow.
•
Ideally, the direction of flow should be at an approximately
45°–60° angle relative to the transducer.
•
Within this range, a linear relation exists b/w velocity and the
Doppler shifts.
•
Outside this range, the unreliable signal produces inaccurate
estimates of flow (,,,Figs 13, ,,,14).
•
The calculated velocity is inversely related to : the cosine of
the angle b/w the TXR and flowing blood.
•
Because the cosine of 90° is 0, the calculated velocity
approaches 0 as the angle approaches 90°
•
Figure 13. Angle correction.
•
(a) Color duplex US image [slide 24] obtained with no angle
correction [angle 0^] shows how no meaningful velocity data
can be obtained from the portal venous waveform because
the computer automatically assigns an angle of 0° (cos 0° = 1).
•
(b) Color duplex US image slide [25] obtained with correct
definition of the angle between the TXR and the direction of
portal venous flow [angle 52^] demonstrates a flow velocity
of 29.3 cm/sec.
Figure 13 a : Without angle correction [anle 0^], the measured flow velocity is 18.0
cm/sec.
no meaningful velocity data can be obtained [computer automatically assigned
anangle of 0^ [cosine 0 = 1]
Figure 13 b : after angle correction [52^], the measured flow velocity is
29.3 cm/sec.
•
Figure 14. Angle correction.
•
(a) Color duplex image slide [27] obtained with a 30°
corrected angle, which is too low, demonstrates a flow
velocity of 21.3 cm/sec [underestimation of true portal vein
flow velocity].
•
(b) Color duplex image slide [28] obtained with a 70°
corrected angle, which is too high, demonstrates a flow
velocity of 52.8 cm/sec in the portal vein, which represents an
overestimation of flow velocity.
•
Note that the measured flow velocity increases as the
corrected angle increases.
Fig 14 a : with 30^ corrected angle [too low] = duplex image demonsrated
overestimated flow velocity of 21.3 cm/s
Fig 14 b : with 70^ corrected angle [toohigh] = duplex image demonsrated
low velocity of 52.8 cm/s
•
The differences between angle correction and angle of
insonation are important to understand.
•
Angle correction specifies the true Doppler angle [b/w beam
& blood flow] by placing the cursor parallel to the direction of
blood flow (,,,Fig 13).
•
Manually applying this correction allows the computer to
solve the Doppler equation.
•
The angle of insonation is the angle b/w theTXR and the
vessel being studied (,,,,Fig 15). The angle of insonation
should also be between 45° and 60°.
•
Figure 15. Angle of insonation.
•
(a, b) Color duplex image [slide 31] of the anterior branch of
the right portal vein obtained with the TXR positioned in an
intercostal (a) and subcostal (b) location depict flow as
moving toward [a] and away [b] from the TXR, respectively.
•
(c) On a color duplex image obtained with theTXR positioned
perpendicular to flow (arrow), no color is assigned, yielding a
false finding of absent flow.
•
The angle of insonation of the vein depends entirely on the
position of the transducer.
Figure 15 a :
Figure 15 b :
Figure 15 c :
•
UNdercorrection applied to the Doppler angle, will result in a
falsely low flow estimate.
•
This result comes directly from the Doppler equation, in
which the smaller the angle between flow direction and
transducer position, the greater the denominator of the
equation.
•
Flow may appear to be reversed when the beam-flow angle
changes about 90°.
•
Complete loss of flow may be evident when the beam-flow
angle is 90°.
•
[2] Spectral Gain
•
The spectral gain setting enhances the intensity of depicted
flow in the spectral display.
•
Gain should be adjusted to outline the contour of the
depicted waveform (,,,,,Fig 16/).
•
Too low a setting falsely suggests absent flow.
•
Too high a setting artificially fills in the spectral waveform,
resulting in falsely increased flow with little meaningful
quantitative flow data.
•
gain settings function independently of other parameters,
•
so that changing the percentage of gain applied to an
acquired signal will not alter any other parameter.
•
A change in the color gain does not alter the spectral gain and
vice versa, with the PRF remaining unchanged.
•
Figure 16. Optimization of gain settings.
•
(a) Duplex image [slide [38] obtained with spectral Doppler as
the active mode and too low a gain setting (0%) falsely
suggests absent flow.
•
(b-d) Duplex images slides [39/40/41] obtained with a gain
setting of 38% (b), 77% (c), and 100% (d) demonstrate gradual
artificial filling in of the spectral waveform, yielding a false
finding of increased flow with little meaningful quantitative
flow data. changing the gain setting will not alter any other
parameter.
•
Changing the color gain does not alter the spectral gain (and
vice versa), and the PRF remains unchanged. Whether the
color or spectral component is active, the gain setting should
be adjusted to outline the contour of the depicted waveform
or color flow depiction.
•
Figure 16 a : spectral Doppler, active mode, too low a gain setting (0%) =
falsely suggests absent flow
Figure 16 b : gain setting 38%
Figure 16 c : gain setting 77%
Figure 16 d : gain setting 100%
•
[3] Gate Size
•
The operator-adjusted gate defines the size and location of
the area from which Doppler information is obtained. The
gate is delineated as a pair of cross-hairs within the 2D image
and should be as small as possible to exclude erroneous signal
arising from adjacent vessels or marginal flow.
•
Too large a gate may @ admit erroneous signal from adjacent
vessels (,Fig 17) @ or may lead to acquisition of data from
extraneous parenchyma.
•
Too small a gate may @ give the false impression of reduced
•
or even absent flow. @ A smaller gate also reduces
computation time @ and increases the frame rate, thereby
@allowing more accurate depiction of flow.
•
Figure 17. Optimizing gate size and position.
•
Color duplex US image slide [45] obtained with @ a wide gate
placed in @ a suboptimal location : shows sampling of flow in
both the portal (above the baseline) and hepatic (below the
baseline) veins.
•
Too large a gate size may result in sampling from too large an
anatomic region.
•
By reducing the gate size and improving the position for
sampling, a normal spectral waveform is obtained.
Figure 17. Optimizing gate size and position
•
[4] Gate Position
•
To maximize depiction of flow, the gate should be positioned
over the central part of the vessel being studied. In central
portions of the liver, where portal and hepatic veins course in
proximity to one another, a small gate should be carefully
placed on the desired vessel to avoid obtaining flow patterns
from adjacent vessels (,Fig 17/slide 45). When helical flow
occurs in the portal vein, as is often seen following liver
transplantation, gate placement over the central part of the
vein will demonstrate expected flow above and below the
spectral baseline.
Hepatic doppler us [2]

More Related Content

Similar to Hepatic doppler us [2]

Electromagnetic flowmeters
Electromagnetic flowmetersElectromagnetic flowmeters
Electromagnetic flowmetersAJAL A J
 
Cathode ray oscilloscope
Cathode ray oscilloscopeCathode ray oscilloscope
Cathode ray oscilloscopeesmaeel21
 
CH2 Hydraulics and hydrology of HP.pptx
CH2 Hydraulics and hydrology of HP.pptxCH2 Hydraulics and hydrology of HP.pptx
CH2 Hydraulics and hydrology of HP.pptxDawit Girma
 
cathode ray oscilloscope &function generator
cathode ray oscilloscope &function generatorcathode ray oscilloscope &function generator
cathode ray oscilloscope &function generatormegha agrawal
 
Flow around a bent duct theory
Flow around a bent duct theoryFlow around a bent duct theory
Flow around a bent duct theoryGaurav Vaibhav
 
Introduction to Video Signals
Introduction to Video SignalsIntroduction to Video Signals
Introduction to Video SignalsDevashish Raval
 
Singular value decomposition filtering in high-frame-rate cardiac vector flow...
Singular value decomposition filtering in high-frame-rate cardiac vector flow...Singular value decomposition filtering in high-frame-rate cardiac vector flow...
Singular value decomposition filtering in high-frame-rate cardiac vector flow...journalBEEI
 
Echocardiography assessment of Aortic Regurgitation severity
Echocardiography assessment of Aortic Regurgitation severityEchocardiography assessment of Aortic Regurgitation severity
Echocardiography assessment of Aortic Regurgitation severityPRAVEEN GUPTA
 
Doppler principles [1]
Doppler principles [1]Doppler principles [1]
Doppler principles [1]KamalEldirawi
 
5. An analog filer has system fnction Ha(s)--a (a) (10 pts,) Comvert .pdf
5. An analog filer has system fnction Ha(s)--a (a) (10 pts,) Comvert .pdf5. An analog filer has system fnction Ha(s)--a (a) (10 pts,) Comvert .pdf
5. An analog filer has system fnction Ha(s)--a (a) (10 pts,) Comvert .pdfinfo324235
 

Similar to Hepatic doppler us [2] (20)

Hepatic dopp us [1]
Hepatic dopp  us [1]Hepatic dopp  us [1]
Hepatic dopp us [1]
 
Electromagnetic flowmeters
Electromagnetic flowmetersElectromagnetic flowmeters
Electromagnetic flowmeters
 
Hepatic dopp us [1]
Hepatic dopp  us [1]Hepatic dopp  us [1]
Hepatic dopp us [1]
 
Cathode ray oscilloscope
Cathode ray oscilloscopeCathode ray oscilloscope
Cathode ray oscilloscope
 
CH2 Hydraulics and hydrology of HP.pptx
CH2 Hydraulics and hydrology of HP.pptxCH2 Hydraulics and hydrology of HP.pptx
CH2 Hydraulics and hydrology of HP.pptx
 
DUAL TRACE OSCILLOSCOPE
DUAL TRACE OSCILLOSCOPEDUAL TRACE OSCILLOSCOPE
DUAL TRACE OSCILLOSCOPE
 
780069
780069780069
780069
 
780069
780069780069
780069
 
780069
780069780069
780069
 
cathode ray oscilloscope &function generator
cathode ray oscilloscope &function generatorcathode ray oscilloscope &function generator
cathode ray oscilloscope &function generator
 
Flow around a bent duct theory
Flow around a bent duct theoryFlow around a bent duct theory
Flow around a bent duct theory
 
dopplerphysics2.pptx
dopplerphysics2.pptxdopplerphysics2.pptx
dopplerphysics2.pptx
 
Doppler phys (2)
Doppler phys (2)Doppler phys (2)
Doppler phys (2)
 
Dop phys (1)
Dop phys (1)Dop phys (1)
Dop phys (1)
 
Introduction to Video Signals
Introduction to Video SignalsIntroduction to Video Signals
Introduction to Video Signals
 
Singular value decomposition filtering in high-frame-rate cardiac vector flow...
Singular value decomposition filtering in high-frame-rate cardiac vector flow...Singular value decomposition filtering in high-frame-rate cardiac vector flow...
Singular value decomposition filtering in high-frame-rate cardiac vector flow...
 
Echocardiography assessment of Aortic Regurgitation severity
Echocardiography assessment of Aortic Regurgitation severityEchocardiography assessment of Aortic Regurgitation severity
Echocardiography assessment of Aortic Regurgitation severity
 
Doppler principles [1]
Doppler principles [1]Doppler principles [1]
Doppler principles [1]
 
5. An analog filer has system fnction Ha(s)--a (a) (10 pts,) Comvert .pdf
5. An analog filer has system fnction Ha(s)--a (a) (10 pts,) Comvert .pdf5. An analog filer has system fnction Ha(s)--a (a) (10 pts,) Comvert .pdf
5. An analog filer has system fnction Ha(s)--a (a) (10 pts,) Comvert .pdf
 
Paper_Horsley
Paper_HorsleyPaper_Horsley
Paper_Horsley
 

More from KamalEldirawi

More from KamalEldirawi (18)

Us physics (11)
Us physics (11)Us physics (11)
Us physics (11)
 
Introductory us
Introductory usIntroductory us
Introductory us
 
Illustrated carotid doppler 3
Illustrated carotid doppler 3Illustrated carotid doppler 3
Illustrated carotid doppler 3
 
Illustrated carotid doppler 2
Illustrated carotid doppler 2Illustrated carotid doppler 2
Illustrated carotid doppler 2
 
Illustrated carotid doppler 1
Illustrated carotid doppler 1Illustrated carotid doppler 1
Illustrated carotid doppler 1
 
Umbilical artery doppler [1]
Umbilical artery doppler [1]Umbilical artery doppler [1]
Umbilical artery doppler [1]
 
Doppler us (3)
Doppler us (3)Doppler us (3)
Doppler us (3)
 
Doppler ultrasounds (1)
Doppler ultrasounds (1)Doppler ultrasounds (1)
Doppler ultrasounds (1)
 
Doppler Physics (3)
Doppler Physics (3)Doppler Physics (3)
Doppler Physics (3)
 
Us physics (9)
Us physics (9)Us physics (9)
Us physics (9)
 
Us physic (s)
Us physic (s)Us physic (s)
Us physic (s)
 
Mc qs us (1)
Mc qs us (1)Mc qs us (1)
Mc qs us (1)
 
Us physics (4)
Us physics (4)Us physics (4)
Us physics (4)
 
Us physics 3
Us physics 3Us physics 3
Us physics 3
 
Us physics (8)
Us physics (8)Us physics (8)
Us physics (8)
 
Us physics 2
Us physics 2Us physics 2
Us physics 2
 
Us machine
Us machineUs machine
Us machine
 
Us hand book (1)
Us hand book (1)Us hand book (1)
Us hand book (1)
 

Recently uploaded

VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escortsaditipandeya
 
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...Dipal Arora
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Dipal Arora
 
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Tirupati Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...chandars293
 
Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...narwatsonia7
 
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service AvailableCall Girls Gwalior Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service AvailableDipal Arora
 
Top Rated Bangalore Call Girls Mg Road ⟟ 9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Mg Road ⟟   9332606886 ⟟ Call Me For Genuine S...Top Rated Bangalore Call Girls Mg Road ⟟   9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Mg Road ⟟ 9332606886 ⟟ Call Me For Genuine S...narwatsonia7
 
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomLucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomdiscovermytutordmt
 
Call Girls Gwalior Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Gwalior Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Bangalore Call Girls Nelamangala Number 9332606886 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 9332606886  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 9332606886  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 9332606886 Meetin With Bangalore Esc...narwatsonia7
 
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...indiancallgirl4rent
 
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...Dipal Arora
 
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...Taniya Sharma
 
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...tanya dube
 
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋TANUJA PANDEY
 
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Call Girls in Nagpur High Profile
 

Recently uploaded (20)

VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
 
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
 
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Tirupati Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
 
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
 
Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...
 
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service AvailableCall Girls Gwalior Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service Available
 
Top Rated Bangalore Call Girls Mg Road ⟟ 9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Mg Road ⟟   9332606886 ⟟ Call Me For Genuine S...Top Rated Bangalore Call Girls Mg Road ⟟   9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Mg Road ⟟ 9332606886 ⟟ Call Me For Genuine S...
 
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomLucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
 
Call Girls Gwalior Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Gwalior Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 9907093804 Top Class Call Girl Service Available
 
Bangalore Call Girls Nelamangala Number 9332606886 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 9332606886  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 9332606886  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 9332606886 Meetin With Bangalore Esc...
 
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
 
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
 
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
 
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
 
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
 
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
 
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
 
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
 

Hepatic doppler us [2]

  • 1. Hepatic Doppler US [2] Dr. Kamal Sayed MBBS MSc US UAA
  • 2. Color AND spectral parameters [contd..] [3] Wall Filters • Wall filters @ selectively filter out all acquired information below an operator-defined frequency threshold. • @ Filters eliminate the typically low-frequency–high-intensity noise that may arise from vessel wall motion. @ Filters operate at variable frequencies to eliminate signal from low-velocity blood flow. @ Filter settings are usually preset by the manufacturer, and a high, medium, or low filter setting may be applied separately to spectral, color, and power imaging.
  • 3. • “High” refers to the higher range of frequency shifts that are filtered out and thus are not depicted on the color image or spectral waveform. • In patients with very slow portal venous flow, a high filter setting may cause this flow to be inadvertently [unintentionally] obscured or missed. • To avoid the loss of signal that characterizes slow flow, filter settings should be kept at the lowest possible setting (typically in the 50–100-Hz range). • .
  • 4. • The cutoff frequencies vary with the velocity scale; the lowest filter cutoff frequencies cannot be used with the highest velocity ranges and vice versa. • On the color bar, the filter setting is indicated by black areas on both sides of the baseline. • •
  • 5. # Expansion of the filter shows up as a widening of the black band (,,,Fig 10) slide [7] # On the spectral waveform, a high wall filter setting will • result in loss of depicted spectral information immediately above the baseline. Reducing the wall filter setting results in filling in of spectral data toward the baseline. •
  • 6. • Figure 9. Changing the wall filter. • (a) Color duplex US image [slide [7] obtained with a high wall filter setting shows @ loss of the low-velocity-flow component of the spectral waveform immediately above the baseline. @ Higher-velocity flow is well depicted, and @accurate flow quantification can still occur. • (b) Color duplex US image slide [8] demonstrates how the spectral waveform progressively fills in toward the baseline as the wall filter is sequentially reduced from high (left arrow) to medium (middle arrow) to low (right arrow).
  • 7. Figure 9 [a] Changing the spectral wall filter [high filter]
  • 8. Figure 9 [b] Changing the spectral wall filter [reduced filter from high to med to low = arrows]
  • 9. • Figure 10. Changing the color Doppler wall filter. • (a) CFD image slide [10] obtained with the highest possible wall filter setting shows how color signal arising from low- velocity flow may be filtered out. • ## The change in the filter setting appears as a change in the width of the horizontal black line in the center of the color bar. Slide [10] • (b) CFD image slide [11] obtained with a low filter setting demonstrates filling in of flow in the hepatic veins (blue), which indicates minimal filtering of color signal. •
  • 10. Figure 10 [b] (high filter = filters out signals from low flow velocities) The change in the filter setting appears as a change in the width of the horizontal black line in the center of the color bar.
  • 11. Figure 10 [b] filling in of flow in hepatic veins [blue] Changing the color Doppler wall filter from high to low. •
  • 12. • [4] Inversion of Flow figure 11 • Inversion refers to the ability to electronically invert the direction of flow as depicted on both the color flow and spectral waveforms. • a) As such, color inversion will result in a blue-red reversal and may lead to misinterpretation of the direction of flow in the vessel being evaluated (Fig 11 a/slide 13). • b) On a color Doppler flow US image obtained with reversal of this inversion, appropriate directional flow is noted. • [Figure 11 b/slide 14]
  • 13. Figure 11a. Inversion of color flow. Image [color Doppler active mode] on inversion of the color bar, portal venous flow appears blue, which falsely suggests reversal of flow (ie, away from the transducer).
  • 14. Figure 11b. Inversion of color flow. On a color Doppler flow US image obtained with reversal of this inversion, appropriate directional flow is noted.
  • 15. • Similarly, inversion of the spectral waveform will switch the depicted flow curve from above to below (or below to above) the baseline (,,,Fig 12). • Adjustment of the inversion button will alter either the color or the spectral scale, depending on which is active at the time of adjustment. • Flow toward the TXR typically appears red (at CFD) OR above the baseline (positive flow) on the spectral waveform and can be changed simply by adjusting the inversion button.
  • 16. • Figure 12. Inversion of spectral and color flow falsely suggesting reversal of portal venous flow. • (a) On a color duplex image slide [17] obtained with spectral Doppler as the active mode, @ the spectral waveform is below the baseline, @ with appropriate color flow. • (b) Color duplex image slide [18] obtained after the inversion button was reversed demonstrates @ appropriate directional flow, with the spectral @ waveform now appearing above the baseline. • Note that the color bar does not change when the Doppler spectrum is inverted.
  • 17. Figure 12 a. Inversion of spectral AND color flow : spectral [active mode], the spectral waveform is below the baseline, with appropriate color flow.
  • 18. Figure 12 b. Color duplex obtained after the inversion button was reversed demonstrates appropriate directional flow, with the spectral waveform now appearing above the baseline.
  • 19. • (B) Spectral-Specific Parameters • [angle correction/ spectral gain/ gate size/ gate position] • The spectral waveform contains a host of hemodynamic information describing the velocity and character of the blood flow in the specific vessel being insonated. • The waveform depicts the spectrum of frequency shifts of all blood traversing the sampled volume during the period of data acquisition . With use of the computer to apply an angle correction, a range of time-dependent velocities are displayed on the vertical axis (spectral display).
  • 20. • [1] angle correction • Angle correction refers to adjustment of the Doppler angle and is used to calibrate the velocity scale for the angle between the US beam and the blood flow being measured. • Spectral Doppler US provides both qualitative and quantitative data about flow velocity. • Flow velocity is calculated from the Doppler frequency shift according to the Doppler equation : • V = ΔfC/2focosθ
  • 21. • Flow velocity is calculated from the Doppler frequency shift according to the Doppler equation : • V = ΔfC/2focosθ • where • V = velocity. • Δf = Doppler frequency shift. • C = speed of sound in soft tissue. • fo = original transmitted frequency. • θ = the angle between the transducer and blood flow.
  • 22. • Ideally, the direction of flow should be at an approximately 45°–60° angle relative to the transducer. • Within this range, a linear relation exists b/w velocity and the Doppler shifts. • Outside this range, the unreliable signal produces inaccurate estimates of flow (,,,Figs 13, ,,,14). • The calculated velocity is inversely related to : the cosine of the angle b/w the TXR and flowing blood. • Because the cosine of 90° is 0, the calculated velocity approaches 0 as the angle approaches 90°
  • 23. • Figure 13. Angle correction. • (a) Color duplex US image [slide 24] obtained with no angle correction [angle 0^] shows how no meaningful velocity data can be obtained from the portal venous waveform because the computer automatically assigns an angle of 0° (cos 0° = 1). • (b) Color duplex US image slide [25] obtained with correct definition of the angle between the TXR and the direction of portal venous flow [angle 52^] demonstrates a flow velocity of 29.3 cm/sec.
  • 24. Figure 13 a : Without angle correction [anle 0^], the measured flow velocity is 18.0 cm/sec. no meaningful velocity data can be obtained [computer automatically assigned anangle of 0^ [cosine 0 = 1]
  • 25. Figure 13 b : after angle correction [52^], the measured flow velocity is 29.3 cm/sec.
  • 26. • Figure 14. Angle correction. • (a) Color duplex image slide [27] obtained with a 30° corrected angle, which is too low, demonstrates a flow velocity of 21.3 cm/sec [underestimation of true portal vein flow velocity]. • (b) Color duplex image slide [28] obtained with a 70° corrected angle, which is too high, demonstrates a flow velocity of 52.8 cm/sec in the portal vein, which represents an overestimation of flow velocity. • Note that the measured flow velocity increases as the corrected angle increases.
  • 27. Fig 14 a : with 30^ corrected angle [too low] = duplex image demonsrated overestimated flow velocity of 21.3 cm/s
  • 28. Fig 14 b : with 70^ corrected angle [toohigh] = duplex image demonsrated low velocity of 52.8 cm/s
  • 29. • The differences between angle correction and angle of insonation are important to understand. • Angle correction specifies the true Doppler angle [b/w beam & blood flow] by placing the cursor parallel to the direction of blood flow (,,,Fig 13). • Manually applying this correction allows the computer to solve the Doppler equation. • The angle of insonation is the angle b/w theTXR and the vessel being studied (,,,,Fig 15). The angle of insonation should also be between 45° and 60°.
  • 30. • Figure 15. Angle of insonation. • (a, b) Color duplex image [slide 31] of the anterior branch of the right portal vein obtained with the TXR positioned in an intercostal (a) and subcostal (b) location depict flow as moving toward [a] and away [b] from the TXR, respectively. • (c) On a color duplex image obtained with theTXR positioned perpendicular to flow (arrow), no color is assigned, yielding a false finding of absent flow. • The angle of insonation of the vein depends entirely on the position of the transducer.
  • 34. • UNdercorrection applied to the Doppler angle, will result in a falsely low flow estimate. • This result comes directly from the Doppler equation, in which the smaller the angle between flow direction and transducer position, the greater the denominator of the equation. • Flow may appear to be reversed when the beam-flow angle changes about 90°. • Complete loss of flow may be evident when the beam-flow angle is 90°.
  • 35. • [2] Spectral Gain • The spectral gain setting enhances the intensity of depicted flow in the spectral display. • Gain should be adjusted to outline the contour of the depicted waveform (,,,,,Fig 16/). • Too low a setting falsely suggests absent flow. • Too high a setting artificially fills in the spectral waveform, resulting in falsely increased flow with little meaningful quantitative flow data.
  • 36. • gain settings function independently of other parameters, • so that changing the percentage of gain applied to an acquired signal will not alter any other parameter. • A change in the color gain does not alter the spectral gain and vice versa, with the PRF remaining unchanged. • Figure 16. Optimization of gain settings. • (a) Duplex image [slide [38] obtained with spectral Doppler as the active mode and too low a gain setting (0%) falsely suggests absent flow.
  • 37. • (b-d) Duplex images slides [39/40/41] obtained with a gain setting of 38% (b), 77% (c), and 100% (d) demonstrate gradual artificial filling in of the spectral waveform, yielding a false finding of increased flow with little meaningful quantitative flow data. changing the gain setting will not alter any other parameter. • Changing the color gain does not alter the spectral gain (and vice versa), and the PRF remains unchanged. Whether the color or spectral component is active, the gain setting should be adjusted to outline the contour of the depicted waveform or color flow depiction. •
  • 38. Figure 16 a : spectral Doppler, active mode, too low a gain setting (0%) = falsely suggests absent flow
  • 39. Figure 16 b : gain setting 38%
  • 40. Figure 16 c : gain setting 77%
  • 41. Figure 16 d : gain setting 100%
  • 42. • [3] Gate Size • The operator-adjusted gate defines the size and location of the area from which Doppler information is obtained. The gate is delineated as a pair of cross-hairs within the 2D image and should be as small as possible to exclude erroneous signal arising from adjacent vessels or marginal flow. • Too large a gate may @ admit erroneous signal from adjacent vessels (,Fig 17) @ or may lead to acquisition of data from extraneous parenchyma.
  • 43. • Too small a gate may @ give the false impression of reduced • or even absent flow. @ A smaller gate also reduces computation time @ and increases the frame rate, thereby @allowing more accurate depiction of flow.
  • 44. • Figure 17. Optimizing gate size and position. • Color duplex US image slide [45] obtained with @ a wide gate placed in @ a suboptimal location : shows sampling of flow in both the portal (above the baseline) and hepatic (below the baseline) veins. • Too large a gate size may result in sampling from too large an anatomic region. • By reducing the gate size and improving the position for sampling, a normal spectral waveform is obtained.
  • 45. Figure 17. Optimizing gate size and position
  • 46. • [4] Gate Position • To maximize depiction of flow, the gate should be positioned over the central part of the vessel being studied. In central portions of the liver, where portal and hepatic veins course in proximity to one another, a small gate should be carefully placed on the desired vessel to avoid obtaining flow patterns from adjacent vessels (,Fig 17/slide 45). When helical flow occurs in the portal vein, as is often seen following liver transplantation, gate placement over the central part of the vein will demonstrate expected flow above and below the spectral baseline.