SlideShare a Scribd company logo
1 of 1
Download to read offline
RESEARCH POSTER PRESENTATION DESIGN © 2015
www.PosterPresentations.com
Turbidity is the cloudiness or haziness of a liquid sample. For this project a
task was given to develop a method to accurately measure turbidity
without the use of the standard equipment, a nephelometer. In order to
accomplish this it was first necessary to understand what causes turbidity
and under what circumstances this measurement would be necessary.
Turbidity is caused by tiny particles suspended in the liquid sample. These
particles scatter incoming light, causing the appearance of cloudiness. The
particles that cause turbidity in water samples are varied and include
organic and inorganic chemical compounds, disinfectant byproducts from
water treatment and microorganisms. Sources for the particles include but
are not limited to agricultural run-off, construction run-off, industrial
discharge both intentional and accidental and municipal water treatment.
With this information it was determined that turbidity would be measured
in two types of circumstances, in a laboratory and in the field. The
methodology developed addresses both of these applications.
Abstract
Introduction
Method I: Turbidity Tube
Large tube constructed from plastic fluorescent
lightbulb cover, rubber cap filled with sand and
handmade plastic secchi disk .
Calibrated using McFarland
standard solutions from 0.5 to 6.0.
Tabletop version used a 100 mL glass
graduated cylinder and paper secchi
disk, pushed up to exterior bottom.
Method II: Spectrophotometer
Jasco V-630 set at 660nm.
Recorded absorbance of samples of McFarland standard solutions
ranging from 0.5 to 6.0
Both devices were calibrated using the same set of McFarland standard
solutions on the same day as they were prepared to reduce errors caused by
particles settling out from the solutions.
Comparison of Methods
Turbidity Tube Spectrophotometer
Inexpensive Expensive
Portable Stable location
Man-powered Requires power source
Large sample needed Smaller sample size
Accuracy and precision Possible interference
dependent on observer’s to results if sample
eyesight and contains large
interpretation particle
Methods
A calibration graph was created from each method. Determination of the
turbidity of the unknown sample was then calculated.
Results Conclusion
The resulting data from both methods are similar. While the
Spectrophotometer was able to detect McFarland units down to 0.5 units,
its upper detection limit was reached at 5 McFarland units. The turbidity
tube functioned up to 6 McFarland units, but it was unable do determine
turbidity below 1 McFarland unit. Both methods can function with a fair
amount of accuracy under the conditions that they would usually be used
for. The turbidity tube is inexpensive to build and the large plastic tube can
be transported into the field with little concern for breakage. It does,
however, require a fairly large sample size. The accuracy of the data from
this method would improve over time as the operator gains experience.
The spectrophotometer is a more expensive option that yields good results,
but application of this method would be limited in the field by proximity of
water source to areas where a mobile laboratory with a power source could
be set up.
References
The Turbidity Tube: Simple and Accurate Measurement of
Turbidity in the Field; Elizabeth Myre & Ryan Shaw, April 2006
http://cas.umn.edu/assets/pdf/Turbidity%20Tube.pdf
Chromaticity and Turbidity Quantitative Measurement using
UV/Vis Spectrophotometer
http://jasco-europe.businesscatalyst.com/assets/uv-0014-e.pdf
https://water.usgs.gov/owq/FieldManual/Chapter6/Section6.7_v2.1
.pdf
https://en.wikipedia.org/wiki/McFarland_standards
https://en.wikipedia.org/wiki/Secchi_disk
http://bcrc.bio.umass.edu/intro/manual/index.php/Gene_Regulation
_II
Turbidity in water is caused by suspended particulate matter. Standard
laboratory methods for measurement include using a Nephelometer, the
Jackson candle method or the Formazin turbidity method. Each method
has its own units of measure, expressed as NTU (Nephelometric turbidity
units) JTU (Jackson turbidity units) or FTU (Formazin turbidity units).
For this project, it was necessary to create a set of standard solutions for
device calibration. This was done by mixing a series of McFarland
standard solutions. McFarland standard solutions are created by mixing
specific ratios of barium chloride and sulfuric acid together. Mixing these
two compounds together results in a turbid solution. ‘Results from this
method are expressed in McFarland units.
In a lab, a spectrophotometer can be used to determine turbidity. In this
device, light from a source is separated into its different wavelengths. This
light then passes through a wavelength specific slit and continues through
the sample liquid. The amount of light that passes through the sample is
measured in either % absorbance or % transmittance. Manufacturer
literature for the spectrophotometer available (Jasco V-630) recommended
a setting of 660nm wavelength for this type of application. This
wavelength would give optimal results based on particle size and the way
light scatters. It is important to note that this method is not acceptable for
Federal reporting requirements. This is due to the limitations inherent to
the design of the spectrophotometer. Its angle of observation is a direct
light path allowing it to measure transmittance or absorbance. For
measuring turbidity accurately, a nephelometer has an offset light sensor to
measure light scatter.
In a field application, a low cost method commonly used is a turbidity
tube. Instead of a computerized device, it relies on human eyesight to
determine the turbidity level of the solution. This device works by
allowing light to pass through a clear tube containing the sample. The
bottom of the tube contains an optical pattern. Turbid solution is added
until the observer can no longer see this pattern through the solution.
Gwinnett Technical College BSCI 1230 Fall 2015
Jalpa Shah and K. Kala Drake
Turbidity Tube vs Spectrophotometer
0
10
20
30
40
50
60
0 1 2 3 4 5 6 7
mLofsample
McFarland Units
Turbidity tube
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0 1 2 3 4 5 6 7
Absorbance
McFarland units
Spectrophotometer Calibration
Sample
type
Turbidity
tube
Spectrophotometer
unknown 4.4 4.1
Jalpa’s
Pond
Water
< 0.5 < 0.5
Kala’s
Pond
Water
< 0.5 < 0.5

More Related Content

What's hot

Evaluation of near infrared (NIR) spectroscopy as a tool for determination of...
Evaluation of near infrared (NIR) spectroscopy as a tool for determination of...Evaluation of near infrared (NIR) spectroscopy as a tool for determination of...
Evaluation of near infrared (NIR) spectroscopy as a tool for determination of...SLOPE Project
 
Microfluidics Lecture
Microfluidics LectureMicrofluidics Lecture
Microfluidics LecturePeter Zhou
 
Near infrared spectroscopy-Food analysis
Near infrared spectroscopy-Food analysisNear infrared spectroscopy-Food analysis
Near infrared spectroscopy-Food analysisFoodtech mbg
 
Chemical analysis via NIR spectroscopy
Chemical analysis via NIR spectroscopyChemical analysis via NIR spectroscopy
Chemical analysis via NIR spectroscopyILRI
 
Application Of NIRS In Feed Industry
Application Of NIRS In Feed IndustryApplication Of NIRS In Feed Industry
Application Of NIRS In Feed Industryguest06ad101
 
Smart sensors for BTEX
Smart sensors for BTEXSmart sensors for BTEX
Smart sensors for BTEXreganf
 
Patterning, integration and characterisation of polymer optical oxygen sensor...
Patterning, integration and characterisation of polymer optical oxygen sensor...Patterning, integration and characterisation of polymer optical oxygen sensor...
Patterning, integration and characterisation of polymer optical oxygen sensor...Sherry Huang
 
Proposal presentation lab on chip
Proposal presentation lab on chipProposal presentation lab on chip
Proposal presentation lab on chipBrent_Gifford
 
Microfluidics and Nanofluidics
Microfluidics and Nanofluidics Microfluidics and Nanofluidics
Microfluidics and Nanofluidics Manjesh Saakre
 
Process Solutions from Wet Chemistry to Near-Infrared Spectroscopy
Process Solutions from Wet Chemistry to Near-Infrared SpectroscopyProcess Solutions from Wet Chemistry to Near-Infrared Spectroscopy
Process Solutions from Wet Chemistry to Near-Infrared SpectroscopyMetrohm USA
 
iwa_guidance_notes_on_leak_detection_and_repair_2007
iwa_guidance_notes_on_leak_detection_and_repair_2007iwa_guidance_notes_on_leak_detection_and_repair_2007
iwa_guidance_notes_on_leak_detection_and_repair_2007Stuart Stapely
 
PSC-Best Practices for Using Near Infrared Instrumentation for the Frac Sand ...
PSC-Best Practices for Using Near Infrared Instrumentation for the Frac Sand ...PSC-Best Practices for Using Near Infrared Instrumentation for the Frac Sand ...
PSC-Best Practices for Using Near Infrared Instrumentation for the Frac Sand ...Bob Schumann
 
Report on Lab on a chip
Report on Lab on a chip Report on Lab on a chip
Report on Lab on a chip Sanjib Pashi
 
Sk microfluidics and lab on-a-chip-ch2
Sk microfluidics and lab on-a-chip-ch2Sk microfluidics and lab on-a-chip-ch2
Sk microfluidics and lab on-a-chip-ch2stanislas547
 
Leakage detection in water pipe networks using Ground Penetrating Radar (GPR)...
Leakage detection in water pipe networks using Ground Penetrating Radar (GPR)...Leakage detection in water pipe networks using Ground Penetrating Radar (GPR)...
Leakage detection in water pipe networks using Ground Penetrating Radar (GPR)...Dai Shi
 
Tutorial: Circuits and Systems for Lab-on-Chip Integration
Tutorial: Circuits and Systems for Lab-on-Chip IntegrationTutorial: Circuits and Systems for Lab-on-Chip Integration
Tutorial: Circuits and Systems for Lab-on-Chip IntegrationWael Badawy
 

What's hot (20)

Evaluation of near infrared (NIR) spectroscopy as a tool for determination of...
Evaluation of near infrared (NIR) spectroscopy as a tool for determination of...Evaluation of near infrared (NIR) spectroscopy as a tool for determination of...
Evaluation of near infrared (NIR) spectroscopy as a tool for determination of...
 
Lab on a chip. Sachin KUMAR KASHYAP
Lab on a chip. Sachin KUMAR KASHYAPLab on a chip. Sachin KUMAR KASHYAP
Lab on a chip. Sachin KUMAR KASHYAP
 
Microfluidics Lecture
Microfluidics LectureMicrofluidics Lecture
Microfluidics Lecture
 
Near infrared spectroscopy-Food analysis
Near infrared spectroscopy-Food analysisNear infrared spectroscopy-Food analysis
Near infrared spectroscopy-Food analysis
 
Chemical analysis via NIR spectroscopy
Chemical analysis via NIR spectroscopyChemical analysis via NIR spectroscopy
Chemical analysis via NIR spectroscopy
 
Application Of NIRS In Feed Industry
Application Of NIRS In Feed IndustryApplication Of NIRS In Feed Industry
Application Of NIRS In Feed Industry
 
Smart sensors for BTEX
Smart sensors for BTEXSmart sensors for BTEX
Smart sensors for BTEX
 
Patterning, integration and characterisation of polymer optical oxygen sensor...
Patterning, integration and characterisation of polymer optical oxygen sensor...Patterning, integration and characterisation of polymer optical oxygen sensor...
Patterning, integration and characterisation of polymer optical oxygen sensor...
 
Proposal presentation lab on chip
Proposal presentation lab on chipProposal presentation lab on chip
Proposal presentation lab on chip
 
Microfluidics and Nanofluidics
Microfluidics and Nanofluidics Microfluidics and Nanofluidics
Microfluidics and Nanofluidics
 
Process Solutions from Wet Chemistry to Near-Infrared Spectroscopy
Process Solutions from Wet Chemistry to Near-Infrared SpectroscopyProcess Solutions from Wet Chemistry to Near-Infrared Spectroscopy
Process Solutions from Wet Chemistry to Near-Infrared Spectroscopy
 
iwa_guidance_notes_on_leak_detection_and_repair_2007
iwa_guidance_notes_on_leak_detection_and_repair_2007iwa_guidance_notes_on_leak_detection_and_repair_2007
iwa_guidance_notes_on_leak_detection_and_repair_2007
 
PSC-Best Practices for Using Near Infrared Instrumentation for the Frac Sand ...
PSC-Best Practices for Using Near Infrared Instrumentation for the Frac Sand ...PSC-Best Practices for Using Near Infrared Instrumentation for the Frac Sand ...
PSC-Best Practices for Using Near Infrared Instrumentation for the Frac Sand ...
 
Report on Lab on a chip
Report on Lab on a chip Report on Lab on a chip
Report on Lab on a chip
 
Near infrared spectroscopy 2
Near infrared spectroscopy 2Near infrared spectroscopy 2
Near infrared spectroscopy 2
 
Nat Sci10
Nat  Sci10Nat  Sci10
Nat Sci10
 
Sk microfluidics and lab on-a-chip-ch2
Sk microfluidics and lab on-a-chip-ch2Sk microfluidics and lab on-a-chip-ch2
Sk microfluidics and lab on-a-chip-ch2
 
Leakage detection in water pipe networks using Ground Penetrating Radar (GPR)...
Leakage detection in water pipe networks using Ground Penetrating Radar (GPR)...Leakage detection in water pipe networks using Ground Penetrating Radar (GPR)...
Leakage detection in water pipe networks using Ground Penetrating Radar (GPR)...
 
Tutorial: Circuits and Systems for Lab-on-Chip Integration
Tutorial: Circuits and Systems for Lab-on-Chip IntegrationTutorial: Circuits and Systems for Lab-on-Chip Integration
Tutorial: Circuits and Systems for Lab-on-Chip Integration
 
water technology guide
water technology guidewater technology guide
water technology guide
 

Viewers also liked

Research Poster Presentation
Research Poster PresentationResearch Poster Presentation
Research Poster Presentationeem5120
 
Shreya poster on drug delivery
Shreya poster  on drug deliveryShreya poster  on drug delivery
Shreya poster on drug deliveryShreya Modi
 
Template For Poster Presentation (Tysa)
Template For Poster Presentation (Tysa)Template For Poster Presentation (Tysa)
Template For Poster Presentation (Tysa)drtanoybose
 
Poster of research with homeless families
Poster of research with homeless familiesPoster of research with homeless families
Poster of research with homeless familiesariemens
 
Research Poster Design
Research Poster DesignResearch Poster Design
Research Poster Designsamharlow
 
The Link Between Alcohol and Breast Cancer
The Link Between Alcohol and Breast CancerThe Link Between Alcohol and Breast Cancer
The Link Between Alcohol and Breast CancerDr. Omer Hameed
 

Viewers also liked (6)

Research Poster Presentation
Research Poster PresentationResearch Poster Presentation
Research Poster Presentation
 
Shreya poster on drug delivery
Shreya poster  on drug deliveryShreya poster  on drug delivery
Shreya poster on drug delivery
 
Template For Poster Presentation (Tysa)
Template For Poster Presentation (Tysa)Template For Poster Presentation (Tysa)
Template For Poster Presentation (Tysa)
 
Poster of research with homeless families
Poster of research with homeless familiesPoster of research with homeless families
Poster of research with homeless families
 
Research Poster Design
Research Poster DesignResearch Poster Design
Research Poster Design
 
The Link Between Alcohol and Breast Cancer
The Link Between Alcohol and Breast CancerThe Link Between Alcohol and Breast Cancer
The Link Between Alcohol and Breast Cancer
 

Similar to Turbidity poster presentation final edit

WATER LEVEL DETECTOR WITH TURBIDITY SENSOR
WATER LEVEL DETECTOR WITH TURBIDITY SENSORWATER LEVEL DETECTOR WITH TURBIDITY SENSOR
WATER LEVEL DETECTOR WITH TURBIDITY SENSORdeepa bhatia
 
TURBIDITY MEASUREMENT USING NIR DIGITAL CAMERAS AND GOOGLE EARTH ENGINE: A re...
TURBIDITY MEASUREMENT USING NIR DIGITAL CAMERAS AND GOOGLE EARTH ENGINE: A re...TURBIDITY MEASUREMENT USING NIR DIGITAL CAMERAS AND GOOGLE EARTH ENGINE: A re...
TURBIDITY MEASUREMENT USING NIR DIGITAL CAMERAS AND GOOGLE EARTH ENGINE: A re...IRJET Journal
 
Turbidity measurement
Turbidity measurementTurbidity measurement
Turbidity measurementgulfamraza
 
Project Report Print
Project Report PrintProject Report Print
Project Report PrintDavid Sharp
 
Paper based analytical devises
Paper based analytical devisesPaper based analytical devises
Paper based analytical devisesShuaib Mubarak
 
Formatted Invited SPIE paper (2)
Formatted Invited SPIE paper (2)Formatted Invited SPIE paper (2)
Formatted Invited SPIE paper (2)Alpha Mansaray
 
MICRO VOLUME UV/VIS SPECTROSCOPY ,U5nano spectrophotometer
MICRO VOLUME UV/VIS SPECTROSCOPY ,U5nano spectrophotometerMICRO VOLUME UV/VIS SPECTROSCOPY ,U5nano spectrophotometer
MICRO VOLUME UV/VIS SPECTROSCOPY ,U5nano spectrophotometersuriyapriya kamaraj
 
relative humidity detection using fiber optics technologies
relative humidity detection using fiber optics technologiesrelative humidity detection using fiber optics technologies
relative humidity detection using fiber optics technologiesAbhas Dash
 
Relative Humidity sensors using optical fiber technologies
Relative Humidity sensors using optical fiber technologiesRelative Humidity sensors using optical fiber technologies
Relative Humidity sensors using optical fiber technologiesAbhas Dash
 
10.1.1.415.6239
10.1.1.415.623910.1.1.415.6239
10.1.1.415.6239Jim Cosman
 
Online Monitoring of Wastewater in Common Effluent Treatment Plant using UV-V...
Online Monitoring of Wastewater in Common Effluent Treatment Plant using UV-V...Online Monitoring of Wastewater in Common Effluent Treatment Plant using UV-V...
Online Monitoring of Wastewater in Common Effluent Treatment Plant using UV-V...Mudasir Gulzar
 
Integrated Optofluidic device to study Interaction of particle or solution wi...
Integrated Optofluidic device to study Interaction of particle or solution wi...Integrated Optofluidic device to study Interaction of particle or solution wi...
Integrated Optofluidic device to study Interaction of particle or solution wi...prathul Nath
 
Report on Diffuse 2D-LOSA
Report on Diffuse 2D-LOSAReport on Diffuse 2D-LOSA
Report on Diffuse 2D-LOSASyed Mukarram
 
Optofluidic Devices
Optofluidic DevicesOptofluidic Devices
Optofluidic Devicesprathul Nath
 
DEVELOPMENT OF A NITRATE CONCENTRATON PREDICTION MODEL USING UV
DEVELOPMENT OF A NITRATE CONCENTRATON PREDICTION MODEL USING UVDEVELOPMENT OF A NITRATE CONCENTRATON PREDICTION MODEL USING UV
DEVELOPMENT OF A NITRATE CONCENTRATON PREDICTION MODEL USING UVEvan Forney
 
IEEE Sensor journal 2016 Title and Abstract
IEEE Sensor journal 2016 Title and AbstractIEEE Sensor journal 2016 Title and Abstract
IEEE Sensor journal 2016 Title and Abstracttsysglobalsolutions
 

Similar to Turbidity poster presentation final edit (20)

WATER LEVEL DETECTOR WITH TURBIDITY SENSOR
WATER LEVEL DETECTOR WITH TURBIDITY SENSORWATER LEVEL DETECTOR WITH TURBIDITY SENSOR
WATER LEVEL DETECTOR WITH TURBIDITY SENSOR
 
TURBIDITY MEASUREMENT USING NIR DIGITAL CAMERAS AND GOOGLE EARTH ENGINE: A re...
TURBIDITY MEASUREMENT USING NIR DIGITAL CAMERAS AND GOOGLE EARTH ENGINE: A re...TURBIDITY MEASUREMENT USING NIR DIGITAL CAMERAS AND GOOGLE EARTH ENGINE: A re...
TURBIDITY MEASUREMENT USING NIR DIGITAL CAMERAS AND GOOGLE EARTH ENGINE: A re...
 
Turbidity measurement
Turbidity measurementTurbidity measurement
Turbidity measurement
 
Project Report Print
Project Report PrintProject Report Print
Project Report Print
 
Paper based analytical devises
Paper based analytical devisesPaper based analytical devises
Paper based analytical devises
 
executivesummary
executivesummaryexecutivesummary
executivesummary
 
COSGC Symposium Paper
COSGC Symposium PaperCOSGC Symposium Paper
COSGC Symposium Paper
 
Formatted Invited SPIE paper (2)
Formatted Invited SPIE paper (2)Formatted Invited SPIE paper (2)
Formatted Invited SPIE paper (2)
 
MICRO VOLUME UV/VIS SPECTROSCOPY ,U5nano spectrophotometer
MICRO VOLUME UV/VIS SPECTROSCOPY ,U5nano spectrophotometerMICRO VOLUME UV/VIS SPECTROSCOPY ,U5nano spectrophotometer
MICRO VOLUME UV/VIS SPECTROSCOPY ,U5nano spectrophotometer
 
relative humidity detection using fiber optics technologies
relative humidity detection using fiber optics technologiesrelative humidity detection using fiber optics technologies
relative humidity detection using fiber optics technologies
 
Relative Humidity sensors using optical fiber technologies
Relative Humidity sensors using optical fiber technologiesRelative Humidity sensors using optical fiber technologies
Relative Humidity sensors using optical fiber technologies
 
10.1.1.415.6239
10.1.1.415.623910.1.1.415.6239
10.1.1.415.6239
 
Online Monitoring of Wastewater in Common Effluent Treatment Plant using UV-V...
Online Monitoring of Wastewater in Common Effluent Treatment Plant using UV-V...Online Monitoring of Wastewater in Common Effluent Treatment Plant using UV-V...
Online Monitoring of Wastewater in Common Effluent Treatment Plant using UV-V...
 
Integrated Optofluidic device to study Interaction of particle or solution wi...
Integrated Optofluidic device to study Interaction of particle or solution wi...Integrated Optofluidic device to study Interaction of particle or solution wi...
Integrated Optofluidic device to study Interaction of particle or solution wi...
 
Report on Diffuse 2D-LOSA
Report on Diffuse 2D-LOSAReport on Diffuse 2D-LOSA
Report on Diffuse 2D-LOSA
 
Evaluation of Microleakage Caused by Different LED Light Sources in Composite...
Evaluation of Microleakage Caused by Different LED Light Sources in Composite...Evaluation of Microleakage Caused by Different LED Light Sources in Composite...
Evaluation of Microleakage Caused by Different LED Light Sources in Composite...
 
Exploring Sources of Uncertainties in Solar Resource Measurements
Exploring Sources of Uncertainties in Solar Resource MeasurementsExploring Sources of Uncertainties in Solar Resource Measurements
Exploring Sources of Uncertainties in Solar Resource Measurements
 
Optofluidic Devices
Optofluidic DevicesOptofluidic Devices
Optofluidic Devices
 
DEVELOPMENT OF A NITRATE CONCENTRATON PREDICTION MODEL USING UV
DEVELOPMENT OF A NITRATE CONCENTRATON PREDICTION MODEL USING UVDEVELOPMENT OF A NITRATE CONCENTRATON PREDICTION MODEL USING UV
DEVELOPMENT OF A NITRATE CONCENTRATON PREDICTION MODEL USING UV
 
IEEE Sensor journal 2016 Title and Abstract
IEEE Sensor journal 2016 Title and AbstractIEEE Sensor journal 2016 Title and Abstract
IEEE Sensor journal 2016 Title and Abstract
 

Turbidity poster presentation final edit

  • 1. RESEARCH POSTER PRESENTATION DESIGN © 2015 www.PosterPresentations.com Turbidity is the cloudiness or haziness of a liquid sample. For this project a task was given to develop a method to accurately measure turbidity without the use of the standard equipment, a nephelometer. In order to accomplish this it was first necessary to understand what causes turbidity and under what circumstances this measurement would be necessary. Turbidity is caused by tiny particles suspended in the liquid sample. These particles scatter incoming light, causing the appearance of cloudiness. The particles that cause turbidity in water samples are varied and include organic and inorganic chemical compounds, disinfectant byproducts from water treatment and microorganisms. Sources for the particles include but are not limited to agricultural run-off, construction run-off, industrial discharge both intentional and accidental and municipal water treatment. With this information it was determined that turbidity would be measured in two types of circumstances, in a laboratory and in the field. The methodology developed addresses both of these applications. Abstract Introduction Method I: Turbidity Tube Large tube constructed from plastic fluorescent lightbulb cover, rubber cap filled with sand and handmade plastic secchi disk . Calibrated using McFarland standard solutions from 0.5 to 6.0. Tabletop version used a 100 mL glass graduated cylinder and paper secchi disk, pushed up to exterior bottom. Method II: Spectrophotometer Jasco V-630 set at 660nm. Recorded absorbance of samples of McFarland standard solutions ranging from 0.5 to 6.0 Both devices were calibrated using the same set of McFarland standard solutions on the same day as they were prepared to reduce errors caused by particles settling out from the solutions. Comparison of Methods Turbidity Tube Spectrophotometer Inexpensive Expensive Portable Stable location Man-powered Requires power source Large sample needed Smaller sample size Accuracy and precision Possible interference dependent on observer’s to results if sample eyesight and contains large interpretation particle Methods A calibration graph was created from each method. Determination of the turbidity of the unknown sample was then calculated. Results Conclusion The resulting data from both methods are similar. While the Spectrophotometer was able to detect McFarland units down to 0.5 units, its upper detection limit was reached at 5 McFarland units. The turbidity tube functioned up to 6 McFarland units, but it was unable do determine turbidity below 1 McFarland unit. Both methods can function with a fair amount of accuracy under the conditions that they would usually be used for. The turbidity tube is inexpensive to build and the large plastic tube can be transported into the field with little concern for breakage. It does, however, require a fairly large sample size. The accuracy of the data from this method would improve over time as the operator gains experience. The spectrophotometer is a more expensive option that yields good results, but application of this method would be limited in the field by proximity of water source to areas where a mobile laboratory with a power source could be set up. References The Turbidity Tube: Simple and Accurate Measurement of Turbidity in the Field; Elizabeth Myre & Ryan Shaw, April 2006 http://cas.umn.edu/assets/pdf/Turbidity%20Tube.pdf Chromaticity and Turbidity Quantitative Measurement using UV/Vis Spectrophotometer http://jasco-europe.businesscatalyst.com/assets/uv-0014-e.pdf https://water.usgs.gov/owq/FieldManual/Chapter6/Section6.7_v2.1 .pdf https://en.wikipedia.org/wiki/McFarland_standards https://en.wikipedia.org/wiki/Secchi_disk http://bcrc.bio.umass.edu/intro/manual/index.php/Gene_Regulation _II Turbidity in water is caused by suspended particulate matter. Standard laboratory methods for measurement include using a Nephelometer, the Jackson candle method or the Formazin turbidity method. Each method has its own units of measure, expressed as NTU (Nephelometric turbidity units) JTU (Jackson turbidity units) or FTU (Formazin turbidity units). For this project, it was necessary to create a set of standard solutions for device calibration. This was done by mixing a series of McFarland standard solutions. McFarland standard solutions are created by mixing specific ratios of barium chloride and sulfuric acid together. Mixing these two compounds together results in a turbid solution. ‘Results from this method are expressed in McFarland units. In a lab, a spectrophotometer can be used to determine turbidity. In this device, light from a source is separated into its different wavelengths. This light then passes through a wavelength specific slit and continues through the sample liquid. The amount of light that passes through the sample is measured in either % absorbance or % transmittance. Manufacturer literature for the spectrophotometer available (Jasco V-630) recommended a setting of 660nm wavelength for this type of application. This wavelength would give optimal results based on particle size and the way light scatters. It is important to note that this method is not acceptable for Federal reporting requirements. This is due to the limitations inherent to the design of the spectrophotometer. Its angle of observation is a direct light path allowing it to measure transmittance or absorbance. For measuring turbidity accurately, a nephelometer has an offset light sensor to measure light scatter. In a field application, a low cost method commonly used is a turbidity tube. Instead of a computerized device, it relies on human eyesight to determine the turbidity level of the solution. This device works by allowing light to pass through a clear tube containing the sample. The bottom of the tube contains an optical pattern. Turbid solution is added until the observer can no longer see this pattern through the solution. Gwinnett Technical College BSCI 1230 Fall 2015 Jalpa Shah and K. Kala Drake Turbidity Tube vs Spectrophotometer 0 10 20 30 40 50 60 0 1 2 3 4 5 6 7 mLofsample McFarland Units Turbidity tube 0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0 1 2 3 4 5 6 7 Absorbance McFarland units Spectrophotometer Calibration Sample type Turbidity tube Spectrophotometer unknown 4.4 4.1 Jalpa’s Pond Water < 0.5 < 0.5 Kala’s Pond Water < 0.5 < 0.5