SlideShare a Scribd company logo
1 of 8
Download to read offline
The Use and Characteristics of Multileaf Collimators in
Radiation Therapy
Introduction and Purpose
According to Takahashi's article1
"Conformation radiotherapy: rotation
techniques as applied to radiography and
radiotherapy of cancer", conventional
collimator where used until the inception
of the multileaf collimator in 1965. These
conventional jaws could only shape
square and rectangular fields.
Another beam-shaping option, at the
time was Cerrobend blocks. They offer
more conformal shapes than conventional
collimator jaws. Jeraj Matjaz and Robar Vlado state that2
"typical MLSs have 40 to
120 leaves, arranged in pairs. By moving and controlling a large number of
narrow, closely abutting individual leaves, one can generate almost any desired
field shape. The advantages of MLCs are simple and less time consuming
preparation, use without needing to enter the treatment room, and simple
change or correction of field shape. The therapy expenses are lower because
individual shielding blocks are not needed, thus eliminating the need to handle
the Wood’s alloy, which is toxic. With MLC, we shorten the therapy time, and thus
also the period during which patient must remain in still position. Other
advantages are constant control and continuous adjusting of the field shape
during irradiation in advanced conformal radiotherapy. MLC has also some
disadvantages, which include a stepping edge effect, radiation leakage
between leaves, wider penumbra, and problems with generating some complex
field shapes."
Figure 1:MLCs from www.bertiehiggins.com
MLC Characteristics
There are three main types of MLC configurations: upper jaw replacement, lower
jaw replacement, or tertiary collimation. Upper jaw replacement is commonly
used by Elekta. Here the leaves are placed close to the source of radiation. This
can be an advantage because there is less motion of the leaves required to move
across the collimated field. This means that you can have a shorter leaf length and
thus a smaller treatment head. The disadvantage of this setup is that the leaf
width must be smaller and tolerances must be higher. Siemens uses the lower
jaw replacement setup. With this type of MLC configuration, both leaf ends and
sides match the beam divergence. The leaves can extend 10cm across the field
centerline. With third level configurations, the MLCs are just below the level of
the conventional upper jaws. Varian uses this configuration. It has the advantage
of limiting downtime in case there is an MLC malfunction becaue you can move
the leaves manually out of the field if failure occurs.2
According to Matjaz et al, "Multileaf collimators that are double focused
(Siemens design) have flat leaf ends that follow the beam divergence. The leaf
ends of Elektra and Varian MLC design are rounded. The material of choice for
leaf construction is tungsten alloy because it has one of the highest width is larger
than the penumbra generated by a focused or divergent edge. Second, the
penumbra width might change as a function of the distance of the leaf end from
the field midline. The leaf position must be detected in real-time to achieve a safe
and reliable position control. Linear encoders and video optical systems are most
commonly used for detection." Linear encoders are used for detection of leaf
positions in MLC systems where "high precision potentiometers are commonly
used. These potentiometers can detect positions of any individual leaf in the
system. For safer work two potentiometers with correlated readings are used in
this system."2
Acceptance Testing
According to the AAPM Task Group 50 report3
, "MLCs should function according
to manufacturer specifications. Acceptance testing provides the opportunity for
the user to become familiar with the MLC and to confirm that it does in fact meet
the stated criteria for acceptance. These tests do not guarantee long-term
accuracy and reliability. As with other equipment, frequent QA testing should be
performed initially, and as confidence builds, the frequency may be relaxed to
balance effort with anticipated need." One acceptance test to perform, is
checking the mechanical axes alignment. Axes that should be included are the
following: gantry, collimator, couch rotation, and jaw and leaf symmetry with the
collimator axis. Optical axes alignment should also be checked by doing light and
radiation coincidence tests. This compares fields with opposing collimator angles.
" This test will also detect flat collimator faces that are out of focus with the
source. Collimator and gantry spoke shots are also useful and should be
registered to the mechanical isocenter. Any misalignment is generally more
serious for collimators which are closer to the source due to geometric
magnification. Therefore, focused MLCs that replace the conventional jaws
require the most careful alignments, while MLCs with rounded leaf faces which
are located below the jaws are usually within tolerances met by the jaws.
Accordingly, these parameters should be tested for the following situations: (1)
jaws or backup diaphragms alone and (2) selected leaf ends and sides from
selected locations within the leaf banks, across the full range of motion, at 0°, 90°,
180°, and 270° gantry angles."3
During acceptance testing MLC performance will also need to be checked. Boyer
et al states in the TG 50 report that "The width of the x-ray attenuation of a leaf at
isocenter is sensitive to the source-to-MLC distance, and it should be verified
during acceptance. The errors in leaf position can be compensated for using
software corrections at the time the apertures are configured. However, for the
sake of uniformity among machines, this condition should be corrected during
installation of the MLC." Leaf position should also be calibrated according to the
vendor technique. Varian uses an optical beam projected over the leaf ends and
then extends them one at a time until the beam is completely blocked. Shaft
encoders help to determine the leaf positions.3
Figure 2: Shows Alignment of 5cm Strips Formed By MLCs (Image from TG-50 Report)
Acceptance tests for MLCs also include: leaf travel, leaf speed, transmission, and
leakage between leaf faces in the closed position. The leaf speed test is to verify
the maximum speed of the leaves as well as a smooth motion. The leaf travel test
checks to see if the leaves can reach their maximum range. The leaf transmission
test verifies the interleaf transmission and transmission beneath the leaves and
jaws combined. Leakage between the leaf faces in the closed position also needs
to be checked, especially when rounded leaves are used. The last check in
acceptance testing of MLCs involves using field-shaping software to make an
irregularly shaped field to test be for using the MLCs clinically.3
Commissioning of MLCs
During commissioning, interleaf transmissions should be less than 2%. Although,
average transmissions can be used instead, if needed. Central axis profiles should
also be obtained as TPRs or TMRs and PDDs. Penumbra data should be added to
the treatment planning system (TPS) if it has that capability. Flat and rounded
leaf edges have different penumbra to account for. Profiles for symmetric and
asymmetric fields should be acquired to check off-axis ratio.
MLC Quality Assurance
According to Klein et al and the task group 142 report4
, a matched segment test
should be performed weekly. It is sometimes called the picket fence test. Some of
the leaf parameters that affect dose delivery for IMRT include leaf positional
accuracy and transmission values. Simple tests, such as the picket fence test "can
assess positional accuracy qualitatively by the matching of sequential segments
and leaf transmission, particularly interleaf. We recommend the picket fence test
be performed weekly with a careful examination of the image acquired by static
film or on-line portal image."4
Monthly test include: the travel speed test, backup diaphragm settings, setting
vs radiation field for 2 patterns, and leaf position accuracy. The travel speed test
is to account for gantry rotation which may affect leaf motion due to gravitational
effects imposed on the leaf carriage system. Loss of travel speed can result in
increased beam holds or gap width errors. MLC travel speed is evaluated with
vendor software or by MLC log file evaluation.4
The backup diaphragm setting test
is only done for Elekta machines and must be within 2mm. The setting versus
radiation field test is done on non-IMRT machines and has a tolerance of 2mm.
Leaf position accuracy is a test done on IMRT machines and has a tolerance of
1mm for leaf positions of an IMRT field for four cardinal gantry angles.4
Figure 3: From TG 142
Annual tests include the following: MLC transmission test, leaf position
repeatability test, MLC spoke shot, coincidence of light field and x-ray field test,
segmental IMRT test, and moving window IMRT test. The tolerance for these tests
are listed in figure 3, from the table in the task group 142 report.
Component Replacement
After an MLC is replaced it is important to ensure accurate leaf positioning by
the calibration of leaf positions. According to Klein et al,4
"through the calibration,
the measured signals, such as voltages from the potentiometers or pixel numbers
from a solid-state camera, and the actual leaf positions establish a one-to-one
relationship. Periodic checking and recalibration are also needed to ensure the
integrity of the controlling system. The Varian MLC calibrates the leaf positions
using narrow infrared beams built into the collimator assembly that transect the
paths of the leaves.
The calibration procedure is carried out automatically each time the MLC
operating system is initialized. Each leaf is driven through its range of travel. As a
given leaf intersects the infrared beam, the values returned by its position
encoders are acquired. These values are used along with equation to calibrate the
leaf position. The calibration values are saved in a table for use by the control
system. In the Philips MLC system, which uses a video optical controlling
mechanism, four reference reflectors are fixed on the head structure. The
positions of the four reflectors establish a fixed frame of reference, which
requires film exposures of regular fields with different field sizes set by a set of
default calibration values. The actual field sizes measured from the films set the
final calibration. During operation, the positions of the four reference reflectors
are acquired and checked every 0.1 sec. The installation of an MLC on existing
equipment should be accompanied by measurements sufficient to realign the
original equipment, if necessary."
Radiation Safety Concerns
Boyer et al3
states that the " assessment of safety with accelerators and
associated devices is tested only minimally in a manufacturer’s acceptance
procedures. Additional safety tests are warranted because of the increased
complexity of an MLC. The use of multiple, conformed MLC fields in either static
or dynamic modes will render the conventional use of visual inspection as a daily
verification of field shapes impractical or impossible. Active interlock checks
should be carried out for leaf and jaw positional tolerances. These measurements
should include assessment of software interlocks, hardware interlocks, and other
possible independent systems. Non-active interlocks designed to prevent
unauthorized motions should be tested. These would include procedures such as
dynamic imaging of field shape, motion enable power line interrupt, etc.
Communication link interlocks are provided to ensure that the heavy data traffic
that flows between the control computers and the accelerator hardware is not
corrupted. Means of intentionally corrupting the data should be carefully
discussed with the manufacturer. Tests should be devised to demonstrate that
the interlocks are functioning to detect true positive data errors. Interlock checks
to ensure the software will not allow a trailing edge of a leaf to be unshielded by
the jaws must be performed."
When it come to radiation safety, it is important to remember events like the
one that took place in March of 2005, at St. Vincent hospital in New York, in which
the MLCs were not even in place during the treatment of Scott Jerome-Parks. This
overtreatment left him, according to Walt Bogdanich of the New York Times,5
"deaf, struggling to see, unable to swallow, burned, with his teeth falling out, with
ulcers in his mouth and throat, nauseated, in severe pain and finally unable to
breathe".
Conclusion
Multileaf collimators are a great leap forward in radiation oncology from the
limitations of jaws and Cerrobend blocks. With MLCs, we have to ability to further
increase the dose to cancer while limiting the dose to normal structures through
IMRT. In the future, I see greater numbers of smaller MLC leaves. This will help
shape the dose further and offer more conformity to the tumor.
References
1. Takahashi, S., "Conformation radiotherapy: rotation techniques as applied
to radiography and radiotherapy of cancer", Acta Radiol. Suppl. 242 (1965)
1–142
2. Matjaz J, Vlado R. Multileaf Collimator in Radiotherapy. Radiol Oncol 2004;
38(3): 235-40.
3. Boyer A, Biggs P, Galvin J, et al. Basic Applications of Multileaf Collimators:
Report of Task Group No. 50 Radiation Therapy Committee. American
Association of Physicists in Medicine 2001.
4. Klein E, Hanley J, Bayouth J, et al. Quality Assurance of Medical
Accelerators: Task Group 142 Report. Med. Phys. 36(9), 2009.
5. Bogdanich W, Radiation Offers New Cures, and Ways to Do Harm. New York
Times. January 23, 2010.

More Related Content

Viewers also liked

Viewers also liked (9)

Thesis presentation
Thesis presentationThesis presentation
Thesis presentation
 
Contents page
Contents pageContents page
Contents page
 
Каталог Орифлэйм Украина №15 2015
Каталог Орифлэйм Украина №15 2015Каталог Орифлэйм Украина №15 2015
Каталог Орифлэйм Украина №15 2015
 
Contents page
Contents pageContents page
Contents page
 
Презентация каталога 2_2016
Презентация каталога 2_2016Презентация каталога 2_2016
Презентация каталога 2_2016
 
Diapositivas Monica
Diapositivas  MonicaDiapositivas  Monica
Diapositivas Monica
 
Clipping Heraldo de Soria 30/09/11 @iedbarcelona
Clipping Heraldo de Soria 30/09/11 @iedbarcelonaClipping Heraldo de Soria 30/09/11 @iedbarcelona
Clipping Heraldo de Soria 30/09/11 @iedbarcelona
 
Perkins Camping Canoe
Perkins Camping CanoePerkins Camping Canoe
Perkins Camping Canoe
 
Research paper presentation
Research paper presentationResearch paper presentation
Research paper presentation
 

Similar to CALLOS-JOHN-MP506-PROJECTPAPER

Multileaf collimator basic concepts.pptx
Multileaf collimator basic concepts.pptxMultileaf collimator basic concepts.pptx
Multileaf collimator basic concepts.pptxTaushifulHoque
 
Multileaf Collimator
Multileaf CollimatorMultileaf Collimator
Multileaf CollimatorVinay Desai
 
beam modifying devises
beam modifying devisesbeam modifying devises
beam modifying devisessaikishore15
 
Measurements and measuring devices search
Measurements and measuring devices searchMeasurements and measuring devices search
Measurements and measuring devices searchmohamed alamin yahia
 
Improving reel-building with multivariable CD control.pdf
Improving reel-building with multivariable CD control.pdfImproving reel-building with multivariable CD control.pdf
Improving reel-building with multivariable CD control.pdfmaheshbalan206
 
Chapter 4 comparators
Chapter 4 comparatorsChapter 4 comparators
Chapter 4 comparatorsVISHALM580
 
LINAC COMMSSN.ppt
LINAC COMMSSN.pptLINAC COMMSSN.ppt
LINAC COMMSSN.pptBBAdhikari
 
Power Swing Phenomena and Comparative Study of Its Detection on Transmission ...
Power Swing Phenomena and Comparative Study of Its Detection on Transmission ...Power Swing Phenomena and Comparative Study of Its Detection on Transmission ...
Power Swing Phenomena and Comparative Study of Its Detection on Transmission ...ijsrd.com
 
Beam modification ppt.pptx
Beam modification ppt.pptxBeam modification ppt.pptx
Beam modification ppt.pptxBBAdhikari
 
© 2023, IRJET | Impact Factor value: 8.226 | ISO 9001:2008 Certified Journal ...
© 2023, IRJET | Impact Factor value: 8.226 | ISO 9001:2008 Certified Journal ...© 2023, IRJET | Impact Factor value: 8.226 | ISO 9001:2008 Certified Journal ...
© 2023, IRJET | Impact Factor value: 8.226 | ISO 9001:2008 Certified Journal ...IRJET Journal
 
Differential Protection of Power Transformer in Substation
Differential Protection of Power Transformer in SubstationDifferential Protection of Power Transformer in Substation
Differential Protection of Power Transformer in Substationijtsrd
 
Fault identification in transformer winding
Fault identification in transformer windingFault identification in transformer winding
Fault identification in transformer windingeSAT Publishing House
 
Fault identification in transformer winding
Fault identification in transformer windingFault identification in transformer winding
Fault identification in transformer windingeSAT Journals
 
Analysing and Comparing the Behaviours of Cylindrical Fatigue Specimens Made ...
Analysing and Comparing the Behaviours of Cylindrical Fatigue Specimens Made ...Analysing and Comparing the Behaviours of Cylindrical Fatigue Specimens Made ...
Analysing and Comparing the Behaviours of Cylindrical Fatigue Specimens Made ...IRJET Journal
 

Similar to CALLOS-JOHN-MP506-PROJECTPAPER (20)

Session 9 radiation oncology
Session 9 radiation oncologySession 9 radiation oncology
Session 9 radiation oncology
 
Y04506134137
Y04506134137Y04506134137
Y04506134137
 
Multileaf collimator basic concepts.pptx
Multileaf collimator basic concepts.pptxMultileaf collimator basic concepts.pptx
Multileaf collimator basic concepts.pptx
 
Multileaf Collimator
Multileaf CollimatorMultileaf Collimator
Multileaf Collimator
 
beam modifying devises
beam modifying devisesbeam modifying devises
beam modifying devises
 
Measurements and measuring devices search
Measurements and measuring devices searchMeasurements and measuring devices search
Measurements and measuring devices search
 
Asmc2014 5.8
Asmc2014 5.8Asmc2014 5.8
Asmc2014 5.8
 
Mlc
MlcMlc
Mlc
 
Improving reel-building with multivariable CD control.pdf
Improving reel-building with multivariable CD control.pdfImproving reel-building with multivariable CD control.pdf
Improving reel-building with multivariable CD control.pdf
 
Chapter 4 comparators
Chapter 4 comparatorsChapter 4 comparators
Chapter 4 comparators
 
LINAC COMMSSN.ppt
LINAC COMMSSN.pptLINAC COMMSSN.ppt
LINAC COMMSSN.ppt
 
Power Swing Phenomena and Comparative Study of Its Detection on Transmission ...
Power Swing Phenomena and Comparative Study of Its Detection on Transmission ...Power Swing Phenomena and Comparative Study of Its Detection on Transmission ...
Power Swing Phenomena and Comparative Study of Its Detection on Transmission ...
 
Beam modification ppt.pptx
Beam modification ppt.pptxBeam modification ppt.pptx
Beam modification ppt.pptx
 
© 2023, IRJET | Impact Factor value: 8.226 | ISO 9001:2008 Certified Journal ...
© 2023, IRJET | Impact Factor value: 8.226 | ISO 9001:2008 Certified Journal ...© 2023, IRJET | Impact Factor value: 8.226 | ISO 9001:2008 Certified Journal ...
© 2023, IRJET | Impact Factor value: 8.226 | ISO 9001:2008 Certified Journal ...
 
Differential Protection of Power Transformer in Substation
Differential Protection of Power Transformer in SubstationDifferential Protection of Power Transformer in Substation
Differential Protection of Power Transformer in Substation
 
Fault identification in transformer winding
Fault identification in transformer windingFault identification in transformer winding
Fault identification in transformer winding
 
Fault identification in transformer winding
Fault identification in transformer windingFault identification in transformer winding
Fault identification in transformer winding
 
Instron machine
Instron machineInstron machine
Instron machine
 
Master Report
Master ReportMaster Report
Master Report
 
Analysing and Comparing the Behaviours of Cylindrical Fatigue Specimens Made ...
Analysing and Comparing the Behaviours of Cylindrical Fatigue Specimens Made ...Analysing and Comparing the Behaviours of Cylindrical Fatigue Specimens Made ...
Analysing and Comparing the Behaviours of Cylindrical Fatigue Specimens Made ...
 

CALLOS-JOHN-MP506-PROJECTPAPER

  • 1. The Use and Characteristics of Multileaf Collimators in Radiation Therapy Introduction and Purpose According to Takahashi's article1 "Conformation radiotherapy: rotation techniques as applied to radiography and radiotherapy of cancer", conventional collimator where used until the inception of the multileaf collimator in 1965. These conventional jaws could only shape square and rectangular fields. Another beam-shaping option, at the time was Cerrobend blocks. They offer more conformal shapes than conventional collimator jaws. Jeraj Matjaz and Robar Vlado state that2 "typical MLSs have 40 to 120 leaves, arranged in pairs. By moving and controlling a large number of narrow, closely abutting individual leaves, one can generate almost any desired field shape. The advantages of MLCs are simple and less time consuming preparation, use without needing to enter the treatment room, and simple change or correction of field shape. The therapy expenses are lower because individual shielding blocks are not needed, thus eliminating the need to handle the Wood’s alloy, which is toxic. With MLC, we shorten the therapy time, and thus also the period during which patient must remain in still position. Other advantages are constant control and continuous adjusting of the field shape during irradiation in advanced conformal radiotherapy. MLC has also some disadvantages, which include a stepping edge effect, radiation leakage between leaves, wider penumbra, and problems with generating some complex field shapes." Figure 1:MLCs from www.bertiehiggins.com
  • 2. MLC Characteristics There are three main types of MLC configurations: upper jaw replacement, lower jaw replacement, or tertiary collimation. Upper jaw replacement is commonly used by Elekta. Here the leaves are placed close to the source of radiation. This can be an advantage because there is less motion of the leaves required to move across the collimated field. This means that you can have a shorter leaf length and thus a smaller treatment head. The disadvantage of this setup is that the leaf width must be smaller and tolerances must be higher. Siemens uses the lower jaw replacement setup. With this type of MLC configuration, both leaf ends and sides match the beam divergence. The leaves can extend 10cm across the field centerline. With third level configurations, the MLCs are just below the level of the conventional upper jaws. Varian uses this configuration. It has the advantage of limiting downtime in case there is an MLC malfunction becaue you can move the leaves manually out of the field if failure occurs.2 According to Matjaz et al, "Multileaf collimators that are double focused (Siemens design) have flat leaf ends that follow the beam divergence. The leaf ends of Elektra and Varian MLC design are rounded. The material of choice for leaf construction is tungsten alloy because it has one of the highest width is larger than the penumbra generated by a focused or divergent edge. Second, the penumbra width might change as a function of the distance of the leaf end from the field midline. The leaf position must be detected in real-time to achieve a safe and reliable position control. Linear encoders and video optical systems are most commonly used for detection." Linear encoders are used for detection of leaf positions in MLC systems where "high precision potentiometers are commonly used. These potentiometers can detect positions of any individual leaf in the system. For safer work two potentiometers with correlated readings are used in this system."2
  • 3. Acceptance Testing According to the AAPM Task Group 50 report3 , "MLCs should function according to manufacturer specifications. Acceptance testing provides the opportunity for the user to become familiar with the MLC and to confirm that it does in fact meet the stated criteria for acceptance. These tests do not guarantee long-term accuracy and reliability. As with other equipment, frequent QA testing should be performed initially, and as confidence builds, the frequency may be relaxed to balance effort with anticipated need." One acceptance test to perform, is checking the mechanical axes alignment. Axes that should be included are the following: gantry, collimator, couch rotation, and jaw and leaf symmetry with the collimator axis. Optical axes alignment should also be checked by doing light and radiation coincidence tests. This compares fields with opposing collimator angles. " This test will also detect flat collimator faces that are out of focus with the source. Collimator and gantry spoke shots are also useful and should be registered to the mechanical isocenter. Any misalignment is generally more serious for collimators which are closer to the source due to geometric magnification. Therefore, focused MLCs that replace the conventional jaws require the most careful alignments, while MLCs with rounded leaf faces which are located below the jaws are usually within tolerances met by the jaws. Accordingly, these parameters should be tested for the following situations: (1) jaws or backup diaphragms alone and (2) selected leaf ends and sides from selected locations within the leaf banks, across the full range of motion, at 0°, 90°, 180°, and 270° gantry angles."3 During acceptance testing MLC performance will also need to be checked. Boyer et al states in the TG 50 report that "The width of the x-ray attenuation of a leaf at isocenter is sensitive to the source-to-MLC distance, and it should be verified during acceptance. The errors in leaf position can be compensated for using software corrections at the time the apertures are configured. However, for the sake of uniformity among machines, this condition should be corrected during installation of the MLC." Leaf position should also be calibrated according to the vendor technique. Varian uses an optical beam projected over the leaf ends and then extends them one at a time until the beam is completely blocked. Shaft encoders help to determine the leaf positions.3
  • 4. Figure 2: Shows Alignment of 5cm Strips Formed By MLCs (Image from TG-50 Report) Acceptance tests for MLCs also include: leaf travel, leaf speed, transmission, and leakage between leaf faces in the closed position. The leaf speed test is to verify the maximum speed of the leaves as well as a smooth motion. The leaf travel test checks to see if the leaves can reach their maximum range. The leaf transmission test verifies the interleaf transmission and transmission beneath the leaves and jaws combined. Leakage between the leaf faces in the closed position also needs to be checked, especially when rounded leaves are used. The last check in acceptance testing of MLCs involves using field-shaping software to make an irregularly shaped field to test be for using the MLCs clinically.3 Commissioning of MLCs During commissioning, interleaf transmissions should be less than 2%. Although, average transmissions can be used instead, if needed. Central axis profiles should also be obtained as TPRs or TMRs and PDDs. Penumbra data should be added to the treatment planning system (TPS) if it has that capability. Flat and rounded leaf edges have different penumbra to account for. Profiles for symmetric and asymmetric fields should be acquired to check off-axis ratio. MLC Quality Assurance According to Klein et al and the task group 142 report4 , a matched segment test should be performed weekly. It is sometimes called the picket fence test. Some of the leaf parameters that affect dose delivery for IMRT include leaf positional accuracy and transmission values. Simple tests, such as the picket fence test "can assess positional accuracy qualitatively by the matching of sequential segments and leaf transmission, particularly interleaf. We recommend the picket fence test be performed weekly with a careful examination of the image acquired by static film or on-line portal image."4
  • 5. Monthly test include: the travel speed test, backup diaphragm settings, setting vs radiation field for 2 patterns, and leaf position accuracy. The travel speed test is to account for gantry rotation which may affect leaf motion due to gravitational effects imposed on the leaf carriage system. Loss of travel speed can result in increased beam holds or gap width errors. MLC travel speed is evaluated with vendor software or by MLC log file evaluation.4 The backup diaphragm setting test is only done for Elekta machines and must be within 2mm. The setting versus radiation field test is done on non-IMRT machines and has a tolerance of 2mm. Leaf position accuracy is a test done on IMRT machines and has a tolerance of 1mm for leaf positions of an IMRT field for four cardinal gantry angles.4 Figure 3: From TG 142 Annual tests include the following: MLC transmission test, leaf position repeatability test, MLC spoke shot, coincidence of light field and x-ray field test, segmental IMRT test, and moving window IMRT test. The tolerance for these tests are listed in figure 3, from the table in the task group 142 report.
  • 6. Component Replacement After an MLC is replaced it is important to ensure accurate leaf positioning by the calibration of leaf positions. According to Klein et al,4 "through the calibration, the measured signals, such as voltages from the potentiometers or pixel numbers from a solid-state camera, and the actual leaf positions establish a one-to-one relationship. Periodic checking and recalibration are also needed to ensure the integrity of the controlling system. The Varian MLC calibrates the leaf positions using narrow infrared beams built into the collimator assembly that transect the paths of the leaves. The calibration procedure is carried out automatically each time the MLC operating system is initialized. Each leaf is driven through its range of travel. As a given leaf intersects the infrared beam, the values returned by its position encoders are acquired. These values are used along with equation to calibrate the leaf position. The calibration values are saved in a table for use by the control system. In the Philips MLC system, which uses a video optical controlling mechanism, four reference reflectors are fixed on the head structure. The positions of the four reflectors establish a fixed frame of reference, which requires film exposures of regular fields with different field sizes set by a set of default calibration values. The actual field sizes measured from the films set the final calibration. During operation, the positions of the four reference reflectors are acquired and checked every 0.1 sec. The installation of an MLC on existing equipment should be accompanied by measurements sufficient to realign the original equipment, if necessary." Radiation Safety Concerns Boyer et al3 states that the " assessment of safety with accelerators and associated devices is tested only minimally in a manufacturer’s acceptance procedures. Additional safety tests are warranted because of the increased complexity of an MLC. The use of multiple, conformed MLC fields in either static or dynamic modes will render the conventional use of visual inspection as a daily verification of field shapes impractical or impossible. Active interlock checks should be carried out for leaf and jaw positional tolerances. These measurements should include assessment of software interlocks, hardware interlocks, and other possible independent systems. Non-active interlocks designed to prevent
  • 7. unauthorized motions should be tested. These would include procedures such as dynamic imaging of field shape, motion enable power line interrupt, etc. Communication link interlocks are provided to ensure that the heavy data traffic that flows between the control computers and the accelerator hardware is not corrupted. Means of intentionally corrupting the data should be carefully discussed with the manufacturer. Tests should be devised to demonstrate that the interlocks are functioning to detect true positive data errors. Interlock checks to ensure the software will not allow a trailing edge of a leaf to be unshielded by the jaws must be performed." When it come to radiation safety, it is important to remember events like the one that took place in March of 2005, at St. Vincent hospital in New York, in which the MLCs were not even in place during the treatment of Scott Jerome-Parks. This overtreatment left him, according to Walt Bogdanich of the New York Times,5 "deaf, struggling to see, unable to swallow, burned, with his teeth falling out, with ulcers in his mouth and throat, nauseated, in severe pain and finally unable to breathe". Conclusion Multileaf collimators are a great leap forward in radiation oncology from the limitations of jaws and Cerrobend blocks. With MLCs, we have to ability to further increase the dose to cancer while limiting the dose to normal structures through IMRT. In the future, I see greater numbers of smaller MLC leaves. This will help shape the dose further and offer more conformity to the tumor.
  • 8. References 1. Takahashi, S., "Conformation radiotherapy: rotation techniques as applied to radiography and radiotherapy of cancer", Acta Radiol. Suppl. 242 (1965) 1–142 2. Matjaz J, Vlado R. Multileaf Collimator in Radiotherapy. Radiol Oncol 2004; 38(3): 235-40. 3. Boyer A, Biggs P, Galvin J, et al. Basic Applications of Multileaf Collimators: Report of Task Group No. 50 Radiation Therapy Committee. American Association of Physicists in Medicine 2001. 4. Klein E, Hanley J, Bayouth J, et al. Quality Assurance of Medical Accelerators: Task Group 142 Report. Med. Phys. 36(9), 2009. 5. Bogdanich W, Radiation Offers New Cures, and Ways to Do Harm. New York Times. January 23, 2010.