Polyaniline-coated electrodes for detection of dissolved oxygen
levels in solution
Thien Dien TO1
, Tung Son Vinh NGUYEN1
, Tien Van TRAN1
, An Duy
NGUYEN1
, Ngan Nguyen LE1
, Tin Chanh Duc DOAN1,
*, Chien Mau DANG1
1
Laboratory for Nanotechnology (LNT), Vietnam National University - Ho Chi Minh City, Community 6, Linh
Trung Ward, Thu Duc District, Ho Chi Minh City, Vietnam;
Abstract: Emeraldine Base Polyaniline (EB-PANI), Emeraldine Salt Polyaniline (ES-PANI) and Leuco
Emeraldine Base Polyaniline (LEB-PANI) were studied for detection of different dissolved oxygen levels in
solutions. Optical properties of the polyaniline forms were characterized by Ultraviolet-Visible (UV-Vis)
spectroscopy. To identify oxygen sensitivity of the PANI films, cyclic voltametry (CV) measurements were
performed with the PANI films drop-casted on working platinum electrodes. Electrolyte solution in the CV
measurements was based on phosphate buffer with salinity of 20‰ and pH 7.3. Pure nitrogen and oxygen gases
were used to change oxygen levels that were monitored by a commercial dissolved oxygen probe. CV
measurements showed the variations of current in corresponding to different oxygen concentrations. The
correlation of the current and dissolved oxygen levels was also determined. The results exhibited the possibility
of using LEB-PANI as sensing electrodes for detection of low concentrations of dissolved oxygen in
aquaculture farming.
Keywords: cyclic voltametry, dissolved oxygen detection, leucoemeraldine base polyaniline, oxidation doping,
polyaniline
I. INTRODUCTION
Dissolved oxygen (DO) concentration in water is one of the most important parameters of water for
aquaculture farming, hence it is needed to be continuously monitored and controlled. DO measurement is used
to identify amount of oxygen dissolved in a solution. Adequate DO concentration in water indicates good
quality of water for aquaculture farming. When DO concentration is lower than 5 mg/l, aquacultural life is in
danger. Large amount of fish might be fatal if DO concentration reaches 1-2 mg/l for several hours.
Commercial DO probes have been in market with various types from different manufacturers. Most DO
sensors employ sensing technologies including optical based method which is commonly referred to as
luminescent and the Clark electrochemical or membrane-covered electrode [1]. Hemera, Sacomam, Thethys
Instruments, YSI supply DO sensors working based on optical method. Amount of oxygen is proportional to
luminescence intensity of chemical dyes. Horiba, Greenspan Analytical, In-situ, Eutech-Thermo Scientific use
polarographic electrodes and electrochemical advanced technologies. However, high cost of these sensors has
hampered the application of these sensor systems in small and medium shrimp/fish farms. Therefore, the need of
compact DO sensor development with low cost, simple operation has been increasing and has attracted the
interest of many scientists in this field.
The optical method was used to measure DO levels in micro-scale water flows with an oxygen
quenchable luminophore platinum (II) octaethyporphrin bound with thin polystyrene film [2]. Another work also
used the optical method to measure DO concentrations within cell culture media by analyzing fluorescence
intensity of dextran and ruthenium complex immobilized on poly(ethylene glycol) diacrylate microarray [3].
However, the optical method with a complicated setup and result analysis is only suitable for measurement in
laboratories and it is difficult to apply for practical applications in aquaculture farming. Besides, electrochemical
technology has been widely used for DO sensor development [4-6]. Chou et al. [5] investigated gold/Nafion
electrodes with H2SO4 acid as a supporting electrolyte for DO detection in water. The sensitivity of DO sensors
was 38.4 µA/ppm and the lowest limit of DO was 3.8 ppm. Peng Wang and his co-workers [6] used a silicon
microprobe including Nafion coating on three platinum electrodes to measure DO levels at 0% atm, 10.5% atm
and 21% atm. The difference between two curves in linear sweep voltammetry in air-equilibrated DI water and
oxygen depleted DI water was attributed to oxygen reduction. However, the aim of this research is to measure
DO concentration in blood for biological and biomedical use; hence, DO concentrations were not high.
Moreover, repeatability was not satisfactory and the calibration curves were not established.
Recently conductive polymers have been investigated for different types of sensors [7-15] but only few
of them have been used for DO measurements. Tieman et al. [16] used ionically conductive polymer
poly(dimethyldiallylammonium chloride) coated on graphite electrodes as working electrodes to measure DO.
Polyaniline-coated electrodes for detection of dissolved oxygen levels in solution
International organization of Scientific Research 15 | P a g e
The authors used Pt auxilary electrode and Ag/AgCl reference electrode. DO concentration was measured by
reduction at potential of -0.40 V versus Ag/AgCl. The linearity of concentration was in the range of 1.4-9.3
ppm. In addition, polyaniline (PANI) was once investigated for DO measurement [17]. A platinum electrode
coated with PANI was examined as a potentiometric sensor. This PANI-coated electrode gave approximately 97
mV/decade of the oxygen concentration and a voltage signal of about 400 mV. A detection limit of DO was
estimated of about 5 x 10-7
mol/l. The proposed mechanism based on the oxidation of reduced form
(leucoemeraldine base PANI) leading to conversion into conducting salt PANI in the presence of oxygen
bubbling in H2SO4 solution. In this research, pH of the medium was kept lower than 7 by a mixture of H2SO4
and NaOH. The oxidative doping due to existing H2SO4 1 M solution could have caused conversion into
conducting salt PANI. Therefore, the proposed mechanism of oxidation by oxygen bubbling is still not clear.
In this report three forms of PANI including Emeraldine Base Polyaniline (EB-PANI), Emeraldine Salt
Polyaniline (ES-PANI) and Leuco Emeraldine Base Polyaniline (LEB-PANI) (Fig. 1) were investigated as
sensing materials for detection of DO levels in solution. UV-Vis spectroscopy and cyclic voltametry (CV) were
used to evaluate the DO sensitivity of these polymers. The increases in current of the CV measurements in
corresponding to different DO concentrations were observed. The results showed possibility of using PANI-
coated electrodes for DO sensors.
Figure 1. (a) Structure of EB-PANI. (b) ES-PANI. (c) LEB-PANI.
II. EXPERIMENTAL
2.1. Materials
Emeraldine Base Polyaniline (EB-PANI) (99.7%, Mw 20,000), dimethyl sulfoxide (DMSO) (99.9%),
phenylhydrazine (PHZ) (97%) were bought from Sigma Aldrich and hydrochloric acid (HCl) (37%) was bought
from Merck. Disodium hydrogen phosphates dodecahydrate (Na2HPO4.12H2O), sodium dihydrogen phosphate
dihydrate (NaH2PO4.2H2O) and sodium chloride (NaCl) were used to prepare testing solutions. All the
chemicals were used as purchased without further purification. Oxygen (O2) gas (99.999%) was supplied by
Singapore Oxygen Air Liquide Pte. Ltd.
2.2. Preparation of polymer solutions
2.2.1. EB-PANI solution
Mix 30 mg of EB-PANI with 2.7 ml DMSO, stirring for 24 hours with a magnetic stirrer at room
temperature to obtain EB-PANI solution. The solution was then filtered through a filter paper. The obtained
polymer solution had dark blue color.
2.2.2. ES-PANI solution
EB-PANI powder was protonated by HCl 0.1 M solution to obtain ES-PANI (green color). The salt was
filtered and rinsed by DI water to remove residual acid. Then ES-PANI was dried under vacuum at 60C in a
vacuum oven. 30 mg ES-PANI was dispersed in 2.7 ml DMSO to obtain ES-PANI solution. This solution
underwent ultrasonification and centrifugal mixing.
2.2.3. LEB-PANI solution
ES-PANI was mixed with PHZ 97% to obtain LEB-PANI. Then the powder was rinsed in DI water,
acetone and methanol to remove residual PHZ and dried at 60C in the vacuum oven. After that LEB-PANI was
dissolved in DMSO and this solution underwent ultrasonification and centrifugal mixing.
2.3. Fabrication of electrochemical chips
Electrochemical chips included one working electrode and one counter electrode (dimension of 3 mm x
9 mm) put separately with a distance of 3 mm. Both electrodes were made of platinum (Pt). Pt layers were
sputtered with Univex 350 (Leybold, Germany) and patterned on silicon (Si) wafers coated with 700 nm thick
SiO2 by photolithography and lift-off techniques. The Pt electrodes had a thickness of 190 nm with a titanium
adhesion layer.
4 µl of the PANI solutions was drop-casted on the working electrode of the electrochemical chips and
these chips were heated in vacuum oven at 60°C for 24 hours to remove completely DMSO.
Polyaniline-coated electrodes for detection of dissolved oxygen levels in solution
International organization of Scientific Research 16 | P a g e
2.4. Characterization
UV-Vis absorption spectroscopy of the PANI samples was obtained from a double beam
spectrophotometer Cary 100 (Varian, USA) in the wavelength range from 200 to 800 nm. To check oxygen
sensitivity of the PANI forms, the quartz cuvettes containing the PANI solutions were bubbled with 0.5 psi of
oxygen gas in 1-5 minutes. After each time of oxygen bubbling, UV-Vis spectra were recorded and compared.
The variation of the UV-Vis spectra in the presence of oxygen (if any) can be applied for optical DO sensors
using PANI layers as a color indicator.
The thickness of the Pt electrodes and the polymer films after fabrication was measured by a stylus
profiler Dektak 6M (Veeco, USA).
2.5. Cyclic voltametry (CV) measurements
Testing electrolyte solutions were prepared from phosphate buffer which was made from 6 g
Na2HPO4.12H2O, 0.5 g NaH2PO4.2H2O and DI water to obtain 200 ml testing solution with pH 7.3. Then 4 g
NaCl was added to make the solution have salinity of 20‰. The testing solution was prepared to imitate the
conditions of water in shrimp ponds.
The testing solutions were bubbled with nitrogen (N2) gas for 15 minutes to minimize DO
concentration in the solution, approximately 1.5 ppm. Then oxygen (O2) gas was bubbled through the solution
with a pressure of 1 psi for 15-90 seconds to create various DO concentrations in the solutions. DO
concentrations were measured by a commercial DO probe DO6400TC/T (Sensorex, US) with a measurement
range of 0-20 mg/l, 0.01 mg/l resolution and ±1.5% accuracy.
The polymer coated chips were dipped in the testing solutions in which the polymer part was immersed
completely. CV measurements were performed with a potentiostat/galvanostat PGSTAT 302N (Metrohm,
Autolab, the Netherlands) and a silver/silver chloride (Ag/AgCl) reference electrode RE-6 (Basi Inc., US). The
polymer chips were mounted on two crocodile clips (Fig. 2). The current was set from 10 µA to 10 mA, the
scanning rate was 50 mV/sec, the temperature was about 28 C ± 1 C.
Figure 2. The polymer chip mounted on two crocodile clips.
III. RESULTS AND DISCUSSION
3.1. UV-Vis spectroscopy of EB-PANI, ES-PANI and LEB-PANI
UV-Vis spectroscopy of EB-PANI, ES-PANI and LEB-PANI was carried out with Cary 100 (Varian,
USA) in the wavelength range from 200 to 800 nm. UV-Vis absorption spectra of EB-PANI, ES-PANI and
LEB-PANI in DMSO are shown in Fig. 3.
Figure 3. UV-Vis absorption spectra of EB-PANI (solid line), ES-PANI (dash line) and LEB-PANI (dash dot
line) in DMSO.
Polyaniline-coated electrodes for detection of dissolved oxygen levels in solution
International organization of Scientific Research 17 | P a g e
UV-Vis spectrum of EB-PANI in DMSO solvent shows two peaks at wavelength of 324 nm and 625
nm. This result is in good agreement with literature values [18-20]. The absorption peak at 324 nm is assigned to
the π - π* transition of the benzenoid rings. The peak at 625 nm originates from the exciton transition of the
quinoid rings [19]. The absorption spectrum of HCl-doped PANI in DMSO solvent showes peaks at 289 nm,
456 nm and 663 nm. The two bands have been assigned to the optical absorption of the polaron band [18, 20].
The absorption spectrum of LEB-PANI has only one peak at 292 nm, which is in good agreement with the
published value [21].
The UV-Vis spectra of PANI solutions with and without pure oxygen bubbling were recorded. The
quartz cuvettes containing the polymer solutions were bubbled with pure oxygen for 1-5 minutes. After each
duration of oxygen bubbling, UV-Vis spectra were recorded and compared (see Supplementary Materials for the
UV-Vis spectra of the PANI forms with oxygen bubbling). The UV-Vis spectra of EB-PANI, ES-PANI and LEB-
PANI solutions almost did not change as the oxygen gas was bubbled into the solutions. However, the UV-Vis
spectra of LEB-PANI solutions which included residual PHZ showed an increasing peak at 450 nm with an
increase in DO concentration (Figure S4 of Supplementary Materials). The color of LEB-PANI changed from
transparent to green when oxygen was bubbled into the solution. It was found that the color change was due to
oxidation of the residual PHZ in the LEB-PANI preparation (Figure S5 of Supplementary Materials).
3.2. Platinum electrodes
Series of the electrochemical chips with Pt electrodes were fabricated by photolithography, sputtering
and lift-off processes. Fig. 4 shows the fabricated Pt electrodes and a diced chip coated with the PANI film on
the working electrode. The thickness of the PANI films varied from 500 nm to several micrometers.
Figure 4. The platinum electrodes on a silicon wafer before dicing into seperate chips (left) and the diced chip
coated with the PANI film on the working electrode (right).
3.3. CV measurements of the PANI-coated chips
CV measurements were performed with the PANI-coated platinum electrode as the working electrode,
another Pt counter electrode and the Ag/AgCl reference electrode. For each CV measurement, at least two chips
of the same batch were measured to check reproducibility. The potential was scanned with various ranges and it
was found that the stable potential range of the fabricated electrodes was from -0.6 V to +1.1 V. Various DO
concentrations in the testing solutions were created by bubbling N2 for 15 minutes and O2 for 15-90 seconds. In
this study, high DO concentrations were used to check the DO sensitivity of the PANI forms. Basing on the
measurement of the commercial DO probe, 1 minute of oxygen bubbling could create 16.05 mg/l of DO
concentration in the testing solutions. Due to the limited measurement range of the commercial DO probe (i.e.
20 mg/l), the DO concentrations created by durations of oxygen bubbling larger than 1 minute were estimated
correspondingly.
3.3.1. EB-PANI coated chips
Fig. 5 shows cyclic voltammograms of the EB-PANI coated chips immersed in the testing solutions
which were bubbled with N2 and O2 for 15, 30, 60 and 90 seconds (DO concentrations ranged from 1.50 ppm to
26.00 ppm).
Polyaniline-coated electrodes for detection of dissolved oxygen levels in solution
International organization of Scientific Research 18 | P a g e
Figure 5. CV plots of the EB-PANI chips with N2 purging (bold solid line), O2 bubbling for 15 seconds (dot
line), 30 seconds (dash line), 60 seconds (solid line), 90 seconds (dash dot dot line).
When DO concentration in the solution increases, the magnitude of current increases with DO
concentrations from 0.015 mA to 0.085 mA at the cathode peak of -0.3 V. There is large difference in current
when DO concentration increases from 1.5 ppm (corresponding to N2 purging) to 26 ppm (90 seconds of O2
bubbling). The increase in current exhibits the DO sensitivity of the EB-PANI films. The change in current
corresponding to estimated DO concentration is 2.86 µA per 1 ppm of DO concentration. This sensitivity is
however lower than the published sensitivity of 38.4 µA/ppm [5] and 12.5 µA.cm-2
/ppm [16]. This difference in
sensitivity is due to different electrode materials and different sizes of electrodes.
The relationship between the cathode current at -0.3 V and DO concentrations was shown in Fig. 6.
The increase in current of the EB-PANI thin films follows an exponential Decay function. It was observed that
the current increased steeply from 1.5 ppm to 10 ppm DO. Beyond 16 ppm DO, the current did not change much
and it became stable. It could be attributed to the saturation of DO in the testing solution with 90 seconds of
oxygen bubbling. Therefore after this measurement, the duration of oxygen bubbling was limited not more than
90 seconds.
Figure 6. The relationship between the cathode current and DO concentrations at -0.3 V of the EB-PANI coated
chip.
3.3.2. ES-PANI coated chips
Cyclic voltammograms of the ES-PANI coated chips which were bubbled with N2 and O2 for 15, 30, 60
and 90 seconds (DO concentrations ranged from 1.50 ppm to 26.00 ppm) are shown in Fig. 7.
Polyaniline-coated electrodes for detection of dissolved oxygen levels in solution
International organization of Scientific Research 19 | P a g e
Figure 7. CV plots of the ES-PANI chips with N2 purging (bold solid line), O2 bubbling for 15 seconds (dot
line), 30 seconds (dash line), 60 seconds (solid line), 90 seconds (dash dot dot line).
When DO concentration in the solution increases from 1.5 ppm to 9.77 ppm, the magnitude of current
increases with DO concentrations from 0.026 mA to 0.125 mA at the cathode peak of -0.05 V. In comparison
with the EB-PANI chips, the cathode peak is found to shift to -0.05 V. The change in current of the ES-PANI
chips is much larger (11.97 µA/ppm of DO concentration) in the range of 1.5-9.77 ppm DO. However, the
current decreases when DO concentration increases to 16.05 ppm (corresponding to 60 seconds of oxygen
bubbling). Two ES-PANI chips of the same measurement gave the same results. The instable increase of the
current can be attributed to the scratched ES-PANI film after the CV measurements (Figure S6 of
Supplementary Materials). It should be noted that the ES-PANI layer was not durable. The layer was stripped
off gradually after the CV scans from 1.5 ppm to 26 ppm DO and was removed completely after two times of
CV scans. The stripping of the ES-PANI layer can be due to dispersion of the conductive ES-PANI particles
which were charged under an electrical field in the electrolyte solution.
3.3.3. LEB-PANI coated chips
Fig. 8 shows the cyclic voltammograms of the LEB-PANI coated chips exposed to oxygen bubbling for
15, 30, 60 and 90 seconds. Similar to the ES-PANI chips, the cathode peak of the LEB-PANI is -0.05 V. This
peak is similar to the reduction peak of -0.03 V of LEB-PANI in CV measurements at pH 1 of Edward Song et
al. [22].
This potential indicated the reduction of LEB-PANI film at the cathode. Large increase in magnitude of
the current is also observed when DO concentrations increased from 1.5 ppm (corresponding to N2 purging) to
26 ppm (1.5 minutes of oxygen bubbling). The current increases with DO concentrations from 0.018 mA to 0.1
mA at -0.05 V. Fig. 9 shows the relationship between the cathode current at -0.05 V and DO concentrations. The
increase in current of the LEB-PANI thin films also follows an exponential Decay function similar to the EB-
PANI films. The current of the LEB-PANI films increases steeply from 1.5 ppm to 10 ppm DO and does not
change much after 15 ppm DO. The change in current corresponding to estimated DO concentration is 3.35 µA
per 1 ppm of DO concentration. The results showed that the DO sensitivity of the LEB-PANI coated chips is
higher than the EB-PANI form.
Figure 8. CV plots of the LEB-PANI chips with N2 purging (bold solid line), O2 bubbling for 15 seconds (dot
line), 30 seconds (dash line), 60 seconds (solid line), 90 seconds (dash dot dot line).
Polyaniline-coated electrodes for detection of dissolved oxygen levels in solution
International organization of Scientific Research 20 | P a g e
Figure 9. The relationship between the cathode current and DO concentrations at -0.05 V of the LEB-PANI
coated chip.
3.3.4. Stability of the PANI electrodes
To check reversibility and consistency of the current increase between nitrogen and oxygen bubbling,
the CV measurements of the PANI chips were performed when the chipswere exposed to bubbling nitrogen (15
minutes) and oxygen (30 seconds) for two times. CV plots after two times of oxygen bubbling (9.77 ppm DO)
are almost the same (Figure S7 of Supplementary Materials). Therefore, the PANI electrode chips have good
reversibility.
To check stability of the PANI electrodes, the CV measurements of the PANI electrodes were
performed as long as the CV plots could be obtained. DO concentration varied from 1.50 ppm to 26.00 ppm for
the CV measurements of each chip. The EB-PANI electrodes could be measured for 5 weeks and the LEB-PANI
layers could be measured for 7 weeks whereas the ES-PANI electrodes could only be measured for 1-2 weeks.
The increase of the cathode current and DO concentrations of the EB-PANI and LEB-PANI electrodes in 5 and
7 weeks plotted from the CV scans also follows an exponential Decay function (Figure S8 and Figure S9 of
Supplementary Materials). The variations in the current of the EB-PANI and LEB-PANI chips were observed.
The current signal of the EB-PANI chips was stable for 2 weeks and decreased in the following weeks. For the
LEB-PANI chips, although there was large fluctuation at high DO concentrations, the current was somehow
stable at below 10 ppm DO after 7 weeks.
3.3.5. Possible sensing mechanism
The cathode reaction relating to the redox reaction of oxygen in solution with the PANI films coated on
the Pt electrode is shown in the following reactions (1) and (2). For simplicity, a general sign is used in the
reaction (1), in which PANI represents different forms of PANI and [PANI]n-
represents the different
electrochemical redox states of PANI which have been reported in many studies [23-25].
PANI + n e → [PANI]n-
(1)
O2 + 4 e → 2 O2-
(2)
Depending on the form of PANI (EB-PANI, ES-PANI or LEB-PANI) the difference in the consumed
number of electrons causes the different cathodic potentials (as observed in the CV measurements). The
cathodic potential of EB-PANI is more polarized than that of LEB-PANI. This leads to the fact that the oxygen
reduction with the LEB-PANI electrode is easier than with the EB-PANI and the higher DO sensitivity of the
LEB-PANI coated chips.
In the reaction (1), the PANI film receives electrons to transform into a charge state [PANI]n-
and a part
of charge is transferred to oxygen to cause the reaction (2). The CV plots of the PANI forms show only one
cathode peak. It could be implied that the reduction reaction is followed by one-step reaction. The reduced
oxygen O2-
may be combined with [PANI]n-
to produce a new compound like [PANI-O2] (Fig. 10). The same
structure of [PANI-O2] was also proposed by D. V. Zhuzhel’skii [26].
Polyaniline-coated electrodes for detection of dissolved oxygen levels in solution
International organization of Scientific Research 21 | P a g e
Figure 10. Mechanism of reduced oxygen at the PANI-coated Pt electrode.
Figure 11. Resonance in [PANI-O2] structure to form oxoniumcations.
In the presence of DO in solution, two oxygen atoms approach the PANI films and attach to the
benzene rings of the PANI. The resonance in the structure then occurs and oxonium ions are generated (Fig. 11).
When there are more oxygen molecules, more oxonium ions are formed and the conductivity of the PANI films
increases. The increase in oxygen concentration leads to the increase in current intensity. However, oxygen
present in the solution can react at some extent. When all the reaction sites for oxygen are occupied, the current
does not increase much as DO concentration increases. This leads to the observed saturation of the increase in
the current (Fig. 6 and Fig. 9).
IV. CONCLUSIONS
In this study we have investigated the DO sensitivity of the EB-PANI, ES-PANI and LEB-PANI films.
The UV-Vis spectra of the PANI solutions did not show any change in the presence of dissolved oxygen.
However, the CV measurements of the EB-PANI, ES-PANI and LEB-PANI films coated on the Pt electrodes
showed the increase in current as DO concentration in the solutions increased. The steep current increase of the
PANI films in the DO range of 1.5-10 ppm is suitable for monitoring of water quality in aquaculture farming.
The correlation of the current and DO concentrations was studied for the EB-PANI and LEB-PANI electrodes.
The higher DO sensitivity (3.35 µA/ppm) and higher stability of the LEB-PANI films enables the feasibility of
using LEB-PANI films as sensing materials in DO sensors.
V. ACKNOWLEDGMENTS
The authors highly appreciate the financial support of Vietnam National University - Ho Chi Minh City for this
study through the grant number B2013-76-04.
REFERENCES
[1]. YSI. The dissolved oxygen handbook: A practical guide to dissolved oxygen measurements.
www.ysi.com/weknowdo, 2009.
[2]. Song, D.H.; Kim, H.D.; Kim, K.C. Dissolved oxygen concentration field measurement in micro-scale
water flows using ptoep/ps film sensor. Optics and Lasers in Engineering, 50, 2012, 74-81.
[3]. Lee, S.; Ibey, B.L.; Coté, G.L.; Pishko, M.V. Measurement of ph and dissolved oxygen within cell
culture media using a hydrogel microarray sensor. Sensors and Actuators B: Chemical, 128, 2008, 388-
398.
[4]. Eminaga, Y.; Brischwein, M.; Wiest, J.; Clauss, J.; Becker, S.; Wolf, B. Self calibration of a planar
dissolved oxygen sensor. Sensors and Actuators B: Chemical, 177, 2013, 785-791.
[5]. Chou, T.C.; Ng, K.M.; Wang, S.H. Gold-solid polymer electrolyte sensor for detecting dissolved oxygen
in water. Sensors and Actuators B: Chemical, 66, 2000, 184-186.
[6]. Wang, P.; Liu, Y.; Abruña, H.D.; Spector, J.A.; Olbricht, W.L. Micromachined dissolved oxygen sensor
based on solid polymer electrolyte. Sensors and Actuators B: Chemical, 153, 2011, 145-151.
Polyaniline-coated electrodes for detection of dissolved oxygen levels in solution
International organization of Scientific Research 22 | P a g e
[7]. Bartlett, P.N.; Birkin, P.R. The application of conducting polymers in biosensors. Synthetic Metals, 61,
1993, 15-21.
[8]. Partridge, A.C.; Jansen, M.L.; Arnold, W.M. Conducting polymer-based sensors. Materials Science and
Engineering: C, 12, 2000, 37-42.
[9]. Nicolas-Debarnot, D.; Poncin-Epaillard, F. Polyaniline as a new sensitive layer for gas sensors. Analytica
Chimica Acta, 475, 2003, 1-15.
[10]. James, D.; Scott, S.M.; Ali, Z.; O’Hare, W.T. Chemical sensors for electronic nose systems.
Microchimica Acta, 149, 2005, 1-17.
[11]. Bai, H.; Shi, G.Q. Gas sensors based on conducting polymers. Sensors, 7, 2007, 267-307.
[12]. Lange, U.; Roznyatovskaya, N.V.; Mirsky, V.M. Conducting polymers in chemical sensors and arrays.
Analytica Chimica Acta, 614, 2008, 1-26.
[13]. Liu, J.; Agarwal, M.; Varahramyan, K. Glucose sensor based on organic thin film transistor using glucose
oxidase and conducting polymer. Sensors and Actuators B: Chemical, 135, 2008, 195-199.
[14]. Tolani, S.; Craig, M.; DeLong, R.; Ghosh, K.; Wanekaya, A. Towards biosensors based on conducting
polymer nanowires. Analytical and Bioanalytical Chemistry, 393, 2009, 1225-1231.
[15]. Doan, T.C.D.; Ramaneti, R.; Baggerman, J.; van der Bent, J.F.; Marcelis, A.T.M.; Tong, H.D.; van Rijn,
C.J.M. Carbon dioxide sensing with sulfonated polyaniline. Sensors and Actuators B: Chemical, 168,
2012, 123-130.
[16]. Tieman, R.S.; Heineman, W.R.; Johnson, J.; Seguin, R. Oxygen sensors based on the ionically conductive
polymer poly(dimethyldiallylammonium chloride). Sensors and Actuators B: Chemical, 8, 1992, 199-
204.
[17]. Shim, Y.-B.; Stilwell, D.E.; Park, S.-M. Electrochemistry of conductive polymer x: Polyaniline-based
potentiometric sensor for dissolved oxygen. Electroanalysis, 3, 1991, 31-36.
[18]. Huang, W.S.; MacDiarmid, A.G. Optical properties of polyaniline. Polymer, 34, 1993, 1833-1845.
[19]. Hua, M.-Y.; Hwang, G.-W.; Chuang, Y.-H.; Chen, S.-A.; Tsai, R.-Y. Soluble n-doped polyaniline: 
Synthesis and characterization. Macromolecules, 33, 2000, 6235-6238.
[20]. Shimano, J.Y.; MacDiarmid, A.G. Polyaniline, a dynamic block copolymer: Key to attaining its intrinsic
conductivity? Synthetic Metals, 123, 2001, 251-262.
[21]. McCall, R.P.; Ginder, J.M.; Leng, J.M.; Ye, H.J.; Manohar, S.K.; Masters, J.G.; Asturias, G.E.;
MacDiarmid, A.G.; Epstein, A.J. Spectroscopy and defect states in polyaniline. Phys. Rev. B, 41, 1990,
5202-5213.
[22]. Edward Song; Choi, J.-W. Conducting polyaniline nanowire and its applications in chemiresistive
sensing. Nanomaterials, 3, 2013, 498-523.
[23]. Gospodinova, N.; Terlemezyan, L. Conducting polymers prepared by oxidative polymerization:
Polyaniline. Progress in Polymer Science, 23, 1998, 1443-1484.
[24]. Genies, E.M.; Tsintavis, C. Redox mechanism and electrochemical behaviour or polyaniline deposits.
Journal of Electroanalytical Chemistry and Interfacial Electrochemistry, 195, 1985, 109-128.
[25]. Zhuzhel’skii, D.V.; Krylova, V.A.; Ivanov, V.D.; Malev, V.V. Mechanism of electrochemical reactions
of polyaniline films formed under the conditions of cathodic oxygen reduction. Russ J Electrochem, 45,
2009, 145-151.

More Related Content

What's hot

GRDS International ECG proceedings,October 2016,Hongkong
GRDS International ECG proceedings,October 2016,HongkongGRDS International ECG proceedings,October 2016,Hongkong
GRDS International ECG proceedings,October 2016,HongkongGlobal R & D Services
 
Photochemistry Mediated Synthesis and Characterization of Thyroxine Capped Si...
Photochemistry Mediated Synthesis and Characterization of Thyroxine Capped Si...Photochemistry Mediated Synthesis and Characterization of Thyroxine Capped Si...
Photochemistry Mediated Synthesis and Characterization of Thyroxine Capped Si...priyanka raviraj
 
Using Tunisian Phosphate Rock and Her Converted Hydroxyapatite for Lead Remov...
Using Tunisian Phosphate Rock and Her Converted Hydroxyapatite for Lead Remov...Using Tunisian Phosphate Rock and Her Converted Hydroxyapatite for Lead Remov...
Using Tunisian Phosphate Rock and Her Converted Hydroxyapatite for Lead Remov...IJERA Editor
 
Novel pH sensor based on fiber optic coated bromophenol blue and cresol red
Novel pH sensor based on fiber optic coated bromophenol blue and cresol redNovel pH sensor based on fiber optic coated bromophenol blue and cresol red
Novel pH sensor based on fiber optic coated bromophenol blue and cresol redTELKOMNIKA JOURNAL
 
M.TECH final viva voice
M.TECH final viva voiceM.TECH final viva voice
M.TECH final viva voiceGOKUL D
 
Equilibrium and Kinetics Adsorption of Cadmium and Lead Ions from Aqueous Sol...
Equilibrium and Kinetics Adsorption of Cadmium and Lead Ions from Aqueous Sol...Equilibrium and Kinetics Adsorption of Cadmium and Lead Ions from Aqueous Sol...
Equilibrium and Kinetics Adsorption of Cadmium and Lead Ions from Aqueous Sol...theijes
 
Preparation, characterization and application of sonochemically doped fe3+ in...
Preparation, characterization and application of sonochemically doped fe3+ in...Preparation, characterization and application of sonochemically doped fe3+ in...
Preparation, characterization and application of sonochemically doped fe3+ in...eSAT Journals
 
Estimation of Chlorine in Water Samples-ELECTROANALYSIS
Estimation of Chlorine in Water Samples-ELECTROANALYSISEstimation of Chlorine in Water Samples-ELECTROANALYSIS
Estimation of Chlorine in Water Samples-ELECTROANALYSISIRJET Journal
 
Ionic liquids in selective extraction of co from ni
Ionic liquids in selective extraction of co from niIonic liquids in selective extraction of co from ni
Ionic liquids in selective extraction of co from niKishan Mahilange
 
Jetircp06003 (2)
Jetircp06003 (2)Jetircp06003 (2)
Jetircp06003 (2)aissmsblogs
 
Microwave Assist Green Synthesis of Silver Nano Particles Using Rhynchostylis...
Microwave Assist Green Synthesis of Silver Nano Particles Using Rhynchostylis...Microwave Assist Green Synthesis of Silver Nano Particles Using Rhynchostylis...
Microwave Assist Green Synthesis of Silver Nano Particles Using Rhynchostylis...Dr. Amarjeet Singh
 
adsorption of methylene blue onto xanthogenated modified chitosan microbeads
adsorption of methylene blue onto xanthogenated modified chitosan microbeadsadsorption of methylene blue onto xanthogenated modified chitosan microbeads
adsorption of methylene blue onto xanthogenated modified chitosan microbeadsSiti Nadzifah Ghazali
 
Synthesis and characterization of Silver Nanoparticles
Synthesis and characterization of Silver NanoparticlesSynthesis and characterization of Silver Nanoparticles
Synthesis and characterization of Silver NanoparticlesIJMERJOURNAL
 

What's hot (18)

GRDS International ECG proceedings,October 2016,Hongkong
GRDS International ECG proceedings,October 2016,HongkongGRDS International ECG proceedings,October 2016,Hongkong
GRDS International ECG proceedings,October 2016,Hongkong
 
Photochemistry Mediated Synthesis and Characterization of Thyroxine Capped Si...
Photochemistry Mediated Synthesis and Characterization of Thyroxine Capped Si...Photochemistry Mediated Synthesis and Characterization of Thyroxine Capped Si...
Photochemistry Mediated Synthesis and Characterization of Thyroxine Capped Si...
 
Purified water
Purified water Purified water
Purified water
 
Using Tunisian Phosphate Rock and Her Converted Hydroxyapatite for Lead Remov...
Using Tunisian Phosphate Rock and Her Converted Hydroxyapatite for Lead Remov...Using Tunisian Phosphate Rock and Her Converted Hydroxyapatite for Lead Remov...
Using Tunisian Phosphate Rock and Her Converted Hydroxyapatite for Lead Remov...
 
MQP Poster
MQP PosterMQP Poster
MQP Poster
 
Novel pH sensor based on fiber optic coated bromophenol blue and cresol red
Novel pH sensor based on fiber optic coated bromophenol blue and cresol redNovel pH sensor based on fiber optic coated bromophenol blue and cresol red
Novel pH sensor based on fiber optic coated bromophenol blue and cresol red
 
M.TECH final viva voice
M.TECH final viva voiceM.TECH final viva voice
M.TECH final viva voice
 
Equilibrium and Kinetics Adsorption of Cadmium and Lead Ions from Aqueous Sol...
Equilibrium and Kinetics Adsorption of Cadmium and Lead Ions from Aqueous Sol...Equilibrium and Kinetics Adsorption of Cadmium and Lead Ions from Aqueous Sol...
Equilibrium and Kinetics Adsorption of Cadmium and Lead Ions from Aqueous Sol...
 
Water Analysis & Test Method
Water Analysis & Test MethodWater Analysis & Test Method
Water Analysis & Test Method
 
Preparation, characterization and application of sonochemically doped fe3+ in...
Preparation, characterization and application of sonochemically doped fe3+ in...Preparation, characterization and application of sonochemically doped fe3+ in...
Preparation, characterization and application of sonochemically doped fe3+ in...
 
tahir20143_2_001
tahir20143_2_001tahir20143_2_001
tahir20143_2_001
 
Estimation of Chlorine in Water Samples-ELECTROANALYSIS
Estimation of Chlorine in Water Samples-ELECTROANALYSISEstimation of Chlorine in Water Samples-ELECTROANALYSIS
Estimation of Chlorine in Water Samples-ELECTROANALYSIS
 
Ionic liquids in selective extraction of co from ni
Ionic liquids in selective extraction of co from niIonic liquids in selective extraction of co from ni
Ionic liquids in selective extraction of co from ni
 
Jetircp06003 (2)
Jetircp06003 (2)Jetircp06003 (2)
Jetircp06003 (2)
 
408madunavalkaj
408madunavalkaj408madunavalkaj
408madunavalkaj
 
Microwave Assist Green Synthesis of Silver Nano Particles Using Rhynchostylis...
Microwave Assist Green Synthesis of Silver Nano Particles Using Rhynchostylis...Microwave Assist Green Synthesis of Silver Nano Particles Using Rhynchostylis...
Microwave Assist Green Synthesis of Silver Nano Particles Using Rhynchostylis...
 
adsorption of methylene blue onto xanthogenated modified chitosan microbeads
adsorption of methylene blue onto xanthogenated modified chitosan microbeadsadsorption of methylene blue onto xanthogenated modified chitosan microbeads
adsorption of methylene blue onto xanthogenated modified chitosan microbeads
 
Synthesis and characterization of Silver Nanoparticles
Synthesis and characterization of Silver NanoparticlesSynthesis and characterization of Silver Nanoparticles
Synthesis and characterization of Silver Nanoparticles
 

Viewers also liked

Glosario de términos - Sistema circulatorio
Glosario de términos - Sistema circulatorioGlosario de términos - Sistema circulatorio
Glosario de términos - Sistema circulatorioNacho Vargas TBc
 
Class 7 bangladesh & global studies capter 9 class 1
Class 7 bangladesh & global studies capter 9 class 1Class 7 bangladesh & global studies capter 9 class 1
Class 7 bangladesh & global studies capter 9 class 1Cambriannews
 
Misterios humanos
Misterios humanosMisterios humanos
Misterios humanosmonsanman
 
Design Thinking - In Business Analytics & Big Data
Design Thinking - In Business Analytics & Big DataDesign Thinking - In Business Analytics & Big Data
Design Thinking - In Business Analytics & Big DataUmakant Jani
 
Armando josemarquezrivas
Armando josemarquezrivasArmando josemarquezrivas
Armando josemarquezrivasARMANDO23493014
 
各國網路販售酒品規定
各國網路販售酒品規定各國網路販售酒品規定
各國網路販售酒品規定vTaiwan.tw
 

Viewers also liked (12)

Glosario de términos - Sistema circulatorio
Glosario de términos - Sistema circulatorioGlosario de términos - Sistema circulatorio
Glosario de términos - Sistema circulatorio
 
A05520108
A05520108A05520108
A05520108
 
B05520913
B05520913B05520913
B05520913
 
D05522329
D05522329D05522329
D05522329
 
F05523541
F05523541F05523541
F05523541
 
Class 7 bangladesh & global studies capter 9 class 1
Class 7 bangladesh & global studies capter 9 class 1Class 7 bangladesh & global studies capter 9 class 1
Class 7 bangladesh & global studies capter 9 class 1
 
E05523034
E05523034E05523034
E05523034
 
Misterios humanos
Misterios humanosMisterios humanos
Misterios humanos
 
Design Thinking - In Business Analytics & Big Data
Design Thinking - In Business Analytics & Big DataDesign Thinking - In Business Analytics & Big Data
Design Thinking - In Business Analytics & Big Data
 
Armando josemarquezrivas
Armando josemarquezrivasArmando josemarquezrivas
Armando josemarquezrivas
 
Cuadro comparativo
Cuadro comparativoCuadro comparativo
Cuadro comparativo
 
各國網路販售酒品規定
各國網路販售酒品規定各國網路販售酒品規定
各國網路販售酒品規定
 

Similar to C05521422

Hollow-Polyaniline-Sphere-Coated Sensor For Measuring Gas-Phase OH Radicals A...
Hollow-Polyaniline-Sphere-Coated Sensor For Measuring Gas-Phase OH Radicals A...Hollow-Polyaniline-Sphere-Coated Sensor For Measuring Gas-Phase OH Radicals A...
Hollow-Polyaniline-Sphere-Coated Sensor For Measuring Gas-Phase OH Radicals A...IJERA Editor
 
International Journal of Engineering and Science Invention (IJESI)
International Journal of Engineering and Science Invention (IJESI)International Journal of Engineering and Science Invention (IJESI)
International Journal of Engineering and Science Invention (IJESI)inventionjournals
 
Distillery Wastewater Decontamination by the Fenton Advanced Oxidation Method
Distillery Wastewater Decontamination by the Fenton Advanced Oxidation MethodDistillery Wastewater Decontamination by the Fenton Advanced Oxidation Method
Distillery Wastewater Decontamination by the Fenton Advanced Oxidation MethodIJRES Journal
 
Synthesis of 2-aminocyclopent-1-ene-1-carbodithioic acid (ACA) Capped Silver ...
Synthesis of 2-aminocyclopent-1-ene-1-carbodithioic acid (ACA) Capped Silver ...Synthesis of 2-aminocyclopent-1-ene-1-carbodithioic acid (ACA) Capped Silver ...
Synthesis of 2-aminocyclopent-1-ene-1-carbodithioic acid (ACA) Capped Silver ...IJERA Editor
 
Highly active pd and pd–au nanoparticles supported on functionalized graphene...
Highly active pd and pd–au nanoparticles supported on functionalized graphene...Highly active pd and pd–au nanoparticles supported on functionalized graphene...
Highly active pd and pd–au nanoparticles supported on functionalized graphene...Science Padayatchi
 
Final Photocatalysis Lab Report (1) (1)
Final Photocatalysis Lab Report (1) (1)Final Photocatalysis Lab Report (1) (1)
Final Photocatalysis Lab Report (1) (1)Henry Hsieh
 
Potential of hydroxamic acid in determination of phenol in industrial waste w...
Potential of hydroxamic acid in determination of phenol in industrial waste w...Potential of hydroxamic acid in determination of phenol in industrial waste w...
Potential of hydroxamic acid in determination of phenol in industrial waste w...Alexander Decker
 
Preparation, characterization and application of
Preparation, characterization and application ofPreparation, characterization and application of
Preparation, characterization and application ofeSAT Publishing House
 
Synthesis & Characterization of Fluorescent Silver Nanoparticles stabilized b...
Synthesis & Characterization of Fluorescent Silver Nanoparticles stabilized b...Synthesis & Characterization of Fluorescent Silver Nanoparticles stabilized b...
Synthesis & Characterization of Fluorescent Silver Nanoparticles stabilized b...IJERA Editor
 
F041033947
F041033947F041033947
F041033947IOSR-JEN
 
Synthesis, Characterization and Electrical Properties of Polyaniline Doped wi...
Synthesis, Characterization and Electrical Properties of Polyaniline Doped wi...Synthesis, Characterization and Electrical Properties of Polyaniline Doped wi...
Synthesis, Characterization and Electrical Properties of Polyaniline Doped wi...IJERA Editor
 
Study-of-Zinc-Oxide-Nanofluids-for-Heat-Transfer-Application (2)
Study-of-Zinc-Oxide-Nanofluids-for-Heat-Transfer-Application (2)Study-of-Zinc-Oxide-Nanofluids-for-Heat-Transfer-Application (2)
Study-of-Zinc-Oxide-Nanofluids-for-Heat-Transfer-Application (2)ujjwal kumar
 
The Effect of Size of the CuO Nanoleaves on the Sunlight Driven Photocatalyti...
The Effect of Size of the CuO Nanoleaves on the Sunlight Driven Photocatalyti...The Effect of Size of the CuO Nanoleaves on the Sunlight Driven Photocatalyti...
The Effect of Size of the CuO Nanoleaves on the Sunlight Driven Photocatalyti...IRJET Journal
 
Electrochemical Behavior of L-Tyrosine at Poly (Dicyclomine Hydrochloride) Fi...
Electrochemical Behavior of L-Tyrosine at Poly (Dicyclomine Hydrochloride) Fi...Electrochemical Behavior of L-Tyrosine at Poly (Dicyclomine Hydrochloride) Fi...
Electrochemical Behavior of L-Tyrosine at Poly (Dicyclomine Hydrochloride) Fi...paperpublications3
 
Novel Conductive Nano-composite Ink Based on Poly Aniline, Silver Nanoparticl...
Novel Conductive Nano-composite Ink Based on Poly Aniline, Silver Nanoparticl...Novel Conductive Nano-composite Ink Based on Poly Aniline, Silver Nanoparticl...
Novel Conductive Nano-composite Ink Based on Poly Aniline, Silver Nanoparticl...Khaled Abdel-Razeq
 
Advanced oxidation of phenolic compounds
Advanced oxidation of phenolic compoundsAdvanced oxidation of phenolic compounds
Advanced oxidation of phenolic compoundsRaafat Alnaizy, PhD
 

Similar to C05521422 (20)

E0262027035
E0262027035E0262027035
E0262027035
 
E0262027035
E0262027035E0262027035
E0262027035
 
Hollow-Polyaniline-Sphere-Coated Sensor For Measuring Gas-Phase OH Radicals A...
Hollow-Polyaniline-Sphere-Coated Sensor For Measuring Gas-Phase OH Radicals A...Hollow-Polyaniline-Sphere-Coated Sensor For Measuring Gas-Phase OH Radicals A...
Hollow-Polyaniline-Sphere-Coated Sensor For Measuring Gas-Phase OH Radicals A...
 
International Journal of Engineering and Science Invention (IJESI)
International Journal of Engineering and Science Invention (IJESI)International Journal of Engineering and Science Invention (IJESI)
International Journal of Engineering and Science Invention (IJESI)
 
Distillery Wastewater Decontamination by the Fenton Advanced Oxidation Method
Distillery Wastewater Decontamination by the Fenton Advanced Oxidation MethodDistillery Wastewater Decontamination by the Fenton Advanced Oxidation Method
Distillery Wastewater Decontamination by the Fenton Advanced Oxidation Method
 
Synthesis of 2-aminocyclopent-1-ene-1-carbodithioic acid (ACA) Capped Silver ...
Synthesis of 2-aminocyclopent-1-ene-1-carbodithioic acid (ACA) Capped Silver ...Synthesis of 2-aminocyclopent-1-ene-1-carbodithioic acid (ACA) Capped Silver ...
Synthesis of 2-aminocyclopent-1-ene-1-carbodithioic acid (ACA) Capped Silver ...
 
Highly active pd and pd–au nanoparticles supported on functionalized graphene...
Highly active pd and pd–au nanoparticles supported on functionalized graphene...Highly active pd and pd–au nanoparticles supported on functionalized graphene...
Highly active pd and pd–au nanoparticles supported on functionalized graphene...
 
Mercury_JECE
Mercury_JECEMercury_JECE
Mercury_JECE
 
Final Photocatalysis Lab Report (1) (1)
Final Photocatalysis Lab Report (1) (1)Final Photocatalysis Lab Report (1) (1)
Final Photocatalysis Lab Report (1) (1)
 
Potential of hydroxamic acid in determination of phenol in industrial waste w...
Potential of hydroxamic acid in determination of phenol in industrial waste w...Potential of hydroxamic acid in determination of phenol in industrial waste w...
Potential of hydroxamic acid in determination of phenol in industrial waste w...
 
Preparation, characterization and application of
Preparation, characterization and application ofPreparation, characterization and application of
Preparation, characterization and application of
 
Synthesis & Characterization of Fluorescent Silver Nanoparticles stabilized b...
Synthesis & Characterization of Fluorescent Silver Nanoparticles stabilized b...Synthesis & Characterization of Fluorescent Silver Nanoparticles stabilized b...
Synthesis & Characterization of Fluorescent Silver Nanoparticles stabilized b...
 
F041033947
F041033947F041033947
F041033947
 
Db25611616
Db25611616Db25611616
Db25611616
 
Synthesis, Characterization and Electrical Properties of Polyaniline Doped wi...
Synthesis, Characterization and Electrical Properties of Polyaniline Doped wi...Synthesis, Characterization and Electrical Properties of Polyaniline Doped wi...
Synthesis, Characterization and Electrical Properties of Polyaniline Doped wi...
 
Study-of-Zinc-Oxide-Nanofluids-for-Heat-Transfer-Application (2)
Study-of-Zinc-Oxide-Nanofluids-for-Heat-Transfer-Application (2)Study-of-Zinc-Oxide-Nanofluids-for-Heat-Transfer-Application (2)
Study-of-Zinc-Oxide-Nanofluids-for-Heat-Transfer-Application (2)
 
The Effect of Size of the CuO Nanoleaves on the Sunlight Driven Photocatalyti...
The Effect of Size of the CuO Nanoleaves on the Sunlight Driven Photocatalyti...The Effect of Size of the CuO Nanoleaves on the Sunlight Driven Photocatalyti...
The Effect of Size of the CuO Nanoleaves on the Sunlight Driven Photocatalyti...
 
Electrochemical Behavior of L-Tyrosine at Poly (Dicyclomine Hydrochloride) Fi...
Electrochemical Behavior of L-Tyrosine at Poly (Dicyclomine Hydrochloride) Fi...Electrochemical Behavior of L-Tyrosine at Poly (Dicyclomine Hydrochloride) Fi...
Electrochemical Behavior of L-Tyrosine at Poly (Dicyclomine Hydrochloride) Fi...
 
Novel Conductive Nano-composite Ink Based on Poly Aniline, Silver Nanoparticl...
Novel Conductive Nano-composite Ink Based on Poly Aniline, Silver Nanoparticl...Novel Conductive Nano-composite Ink Based on Poly Aniline, Silver Nanoparticl...
Novel Conductive Nano-composite Ink Based on Poly Aniline, Silver Nanoparticl...
 
Advanced oxidation of phenolic compounds
Advanced oxidation of phenolic compoundsAdvanced oxidation of phenolic compounds
Advanced oxidation of phenolic compounds
 

More from IOSR-JEN

More from IOSR-JEN (20)

C05921721
C05921721C05921721
C05921721
 
B05921016
B05921016B05921016
B05921016
 
A05920109
A05920109A05920109
A05920109
 
J05915457
J05915457J05915457
J05915457
 
I05914153
I05914153I05914153
I05914153
 
H05913540
H05913540H05913540
H05913540
 
G05913234
G05913234G05913234
G05913234
 
F05912731
F05912731F05912731
F05912731
 
E05912226
E05912226E05912226
E05912226
 
D05911621
D05911621D05911621
D05911621
 
C05911315
C05911315C05911315
C05911315
 
B05910712
B05910712B05910712
B05910712
 
A05910106
A05910106A05910106
A05910106
 
B05840510
B05840510B05840510
B05840510
 
I05844759
I05844759I05844759
I05844759
 
H05844346
H05844346H05844346
H05844346
 
G05843942
G05843942G05843942
G05843942
 
F05843238
F05843238F05843238
F05843238
 
E05842831
E05842831E05842831
E05842831
 
D05842227
D05842227D05842227
D05842227
 

Recently uploaded

SQL Database Design For Developers at php[tek] 2024
SQL Database Design For Developers at php[tek] 2024SQL Database Design For Developers at php[tek] 2024
SQL Database Design For Developers at php[tek] 2024Scott Keck-Warren
 
Commit 2024 - Secret Management made easy
Commit 2024 - Secret Management made easyCommit 2024 - Secret Management made easy
Commit 2024 - Secret Management made easyAlfredo García Lavilla
 
Are Multi-Cloud and Serverless Good or Bad?
Are Multi-Cloud and Serverless Good or Bad?Are Multi-Cloud and Serverless Good or Bad?
Are Multi-Cloud and Serverless Good or Bad?Mattias Andersson
 
Integration and Automation in Practice: CI/CD in Mule Integration and Automat...
Integration and Automation in Practice: CI/CD in Mule Integration and Automat...Integration and Automation in Practice: CI/CD in Mule Integration and Automat...
Integration and Automation in Practice: CI/CD in Mule Integration and Automat...Patryk Bandurski
 
Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024BookNet Canada
 
Scanning the Internet for External Cloud Exposures via SSL Certs
Scanning the Internet for External Cloud Exposures via SSL CertsScanning the Internet for External Cloud Exposures via SSL Certs
Scanning the Internet for External Cloud Exposures via SSL CertsRizwan Syed
 
Beyond Boundaries: Leveraging No-Code Solutions for Industry Innovation
Beyond Boundaries: Leveraging No-Code Solutions for Industry InnovationBeyond Boundaries: Leveraging No-Code Solutions for Industry Innovation
Beyond Boundaries: Leveraging No-Code Solutions for Industry InnovationSafe Software
 
AI as an Interface for Commercial Buildings
AI as an Interface for Commercial BuildingsAI as an Interface for Commercial Buildings
AI as an Interface for Commercial BuildingsMemoori
 
"Subclassing and Composition – A Pythonic Tour of Trade-Offs", Hynek Schlawack
"Subclassing and Composition – A Pythonic Tour of Trade-Offs", Hynek Schlawack"Subclassing and Composition – A Pythonic Tour of Trade-Offs", Hynek Schlawack
"Subclassing and Composition – A Pythonic Tour of Trade-Offs", Hynek SchlawackFwdays
 
New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024BookNet Canada
 
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 3652toLead Limited
 
Understanding the Laravel MVC Architecture
Understanding the Laravel MVC ArchitectureUnderstanding the Laravel MVC Architecture
Understanding the Laravel MVC ArchitecturePixlogix Infotech
 
"ML in Production",Oleksandr Bagan
"ML in Production",Oleksandr Bagan"ML in Production",Oleksandr Bagan
"ML in Production",Oleksandr BaganFwdays
 
My Hashitalk Indonesia April 2024 Presentation
My Hashitalk Indonesia April 2024 PresentationMy Hashitalk Indonesia April 2024 Presentation
My Hashitalk Indonesia April 2024 PresentationRidwan Fadjar
 
Ensuring Technical Readiness For Copilot in Microsoft 365
Ensuring Technical Readiness For Copilot in Microsoft 365Ensuring Technical Readiness For Copilot in Microsoft 365
Ensuring Technical Readiness For Copilot in Microsoft 3652toLead Limited
 
Designing IA for AI - Information Architecture Conference 2024
Designing IA for AI - Information Architecture Conference 2024Designing IA for AI - Information Architecture Conference 2024
Designing IA for AI - Information Architecture Conference 2024Enterprise Knowledge
 
Developer Data Modeling Mistakes: From Postgres to NoSQL
Developer Data Modeling Mistakes: From Postgres to NoSQLDeveloper Data Modeling Mistakes: From Postgres to NoSQL
Developer Data Modeling Mistakes: From Postgres to NoSQLScyllaDB
 
Powerpoint exploring the locations used in television show Time Clash
Powerpoint exploring the locations used in television show Time ClashPowerpoint exploring the locations used in television show Time Clash
Powerpoint exploring the locations used in television show Time Clashcharlottematthew16
 
My INSURER PTE LTD - Insurtech Innovation Award 2024
My INSURER PTE LTD - Insurtech Innovation Award 2024My INSURER PTE LTD - Insurtech Innovation Award 2024
My INSURER PTE LTD - Insurtech Innovation Award 2024The Digital Insurer
 
Unleash Your Potential - Namagunga Girls Coding Club
Unleash Your Potential - Namagunga Girls Coding ClubUnleash Your Potential - Namagunga Girls Coding Club
Unleash Your Potential - Namagunga Girls Coding ClubKalema Edgar
 

Recently uploaded (20)

SQL Database Design For Developers at php[tek] 2024
SQL Database Design For Developers at php[tek] 2024SQL Database Design For Developers at php[tek] 2024
SQL Database Design For Developers at php[tek] 2024
 
Commit 2024 - Secret Management made easy
Commit 2024 - Secret Management made easyCommit 2024 - Secret Management made easy
Commit 2024 - Secret Management made easy
 
Are Multi-Cloud and Serverless Good or Bad?
Are Multi-Cloud and Serverless Good or Bad?Are Multi-Cloud and Serverless Good or Bad?
Are Multi-Cloud and Serverless Good or Bad?
 
Integration and Automation in Practice: CI/CD in Mule Integration and Automat...
Integration and Automation in Practice: CI/CD in Mule Integration and Automat...Integration and Automation in Practice: CI/CD in Mule Integration and Automat...
Integration and Automation in Practice: CI/CD in Mule Integration and Automat...
 
Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
Transcript: New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
 
Scanning the Internet for External Cloud Exposures via SSL Certs
Scanning the Internet for External Cloud Exposures via SSL CertsScanning the Internet for External Cloud Exposures via SSL Certs
Scanning the Internet for External Cloud Exposures via SSL Certs
 
Beyond Boundaries: Leveraging No-Code Solutions for Industry Innovation
Beyond Boundaries: Leveraging No-Code Solutions for Industry InnovationBeyond Boundaries: Leveraging No-Code Solutions for Industry Innovation
Beyond Boundaries: Leveraging No-Code Solutions for Industry Innovation
 
AI as an Interface for Commercial Buildings
AI as an Interface for Commercial BuildingsAI as an Interface for Commercial Buildings
AI as an Interface for Commercial Buildings
 
"Subclassing and Composition – A Pythonic Tour of Trade-Offs", Hynek Schlawack
"Subclassing and Composition – A Pythonic Tour of Trade-Offs", Hynek Schlawack"Subclassing and Composition – A Pythonic Tour of Trade-Offs", Hynek Schlawack
"Subclassing and Composition – A Pythonic Tour of Trade-Offs", Hynek Schlawack
 
New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
New from BookNet Canada for 2024: BNC CataList - Tech Forum 2024
 
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365
Tech-Forward - Achieving Business Readiness For Copilot in Microsoft 365
 
Understanding the Laravel MVC Architecture
Understanding the Laravel MVC ArchitectureUnderstanding the Laravel MVC Architecture
Understanding the Laravel MVC Architecture
 
"ML in Production",Oleksandr Bagan
"ML in Production",Oleksandr Bagan"ML in Production",Oleksandr Bagan
"ML in Production",Oleksandr Bagan
 
My Hashitalk Indonesia April 2024 Presentation
My Hashitalk Indonesia April 2024 PresentationMy Hashitalk Indonesia April 2024 Presentation
My Hashitalk Indonesia April 2024 Presentation
 
Ensuring Technical Readiness For Copilot in Microsoft 365
Ensuring Technical Readiness For Copilot in Microsoft 365Ensuring Technical Readiness For Copilot in Microsoft 365
Ensuring Technical Readiness For Copilot in Microsoft 365
 
Designing IA for AI - Information Architecture Conference 2024
Designing IA for AI - Information Architecture Conference 2024Designing IA for AI - Information Architecture Conference 2024
Designing IA for AI - Information Architecture Conference 2024
 
Developer Data Modeling Mistakes: From Postgres to NoSQL
Developer Data Modeling Mistakes: From Postgres to NoSQLDeveloper Data Modeling Mistakes: From Postgres to NoSQL
Developer Data Modeling Mistakes: From Postgres to NoSQL
 
Powerpoint exploring the locations used in television show Time Clash
Powerpoint exploring the locations used in television show Time ClashPowerpoint exploring the locations used in television show Time Clash
Powerpoint exploring the locations used in television show Time Clash
 
My INSURER PTE LTD - Insurtech Innovation Award 2024
My INSURER PTE LTD - Insurtech Innovation Award 2024My INSURER PTE LTD - Insurtech Innovation Award 2024
My INSURER PTE LTD - Insurtech Innovation Award 2024
 
Unleash Your Potential - Namagunga Girls Coding Club
Unleash Your Potential - Namagunga Girls Coding ClubUnleash Your Potential - Namagunga Girls Coding Club
Unleash Your Potential - Namagunga Girls Coding Club
 

C05521422

  • 1. IOSR Journal of Engineering (IOSRJEN) www.iosrjen.org ISSN (e): 2250-3021, ISSN (p): 2278-8719 Vol. 05, Issue 05 (May. 2015), ||V2|| PP 14-22 International organization of Scientific Research 14 | P a g e Polyaniline-coated electrodes for detection of dissolved oxygen levels in solution Thien Dien TO1 , Tung Son Vinh NGUYEN1 , Tien Van TRAN1 , An Duy NGUYEN1 , Ngan Nguyen LE1 , Tin Chanh Duc DOAN1, *, Chien Mau DANG1 1 Laboratory for Nanotechnology (LNT), Vietnam National University - Ho Chi Minh City, Community 6, Linh Trung Ward, Thu Duc District, Ho Chi Minh City, Vietnam; Abstract: Emeraldine Base Polyaniline (EB-PANI), Emeraldine Salt Polyaniline (ES-PANI) and Leuco Emeraldine Base Polyaniline (LEB-PANI) were studied for detection of different dissolved oxygen levels in solutions. Optical properties of the polyaniline forms were characterized by Ultraviolet-Visible (UV-Vis) spectroscopy. To identify oxygen sensitivity of the PANI films, cyclic voltametry (CV) measurements were performed with the PANI films drop-casted on working platinum electrodes. Electrolyte solution in the CV measurements was based on phosphate buffer with salinity of 20‰ and pH 7.3. Pure nitrogen and oxygen gases were used to change oxygen levels that were monitored by a commercial dissolved oxygen probe. CV measurements showed the variations of current in corresponding to different oxygen concentrations. The correlation of the current and dissolved oxygen levels was also determined. The results exhibited the possibility of using LEB-PANI as sensing electrodes for detection of low concentrations of dissolved oxygen in aquaculture farming. Keywords: cyclic voltametry, dissolved oxygen detection, leucoemeraldine base polyaniline, oxidation doping, polyaniline I. INTRODUCTION Dissolved oxygen (DO) concentration in water is one of the most important parameters of water for aquaculture farming, hence it is needed to be continuously monitored and controlled. DO measurement is used to identify amount of oxygen dissolved in a solution. Adequate DO concentration in water indicates good quality of water for aquaculture farming. When DO concentration is lower than 5 mg/l, aquacultural life is in danger. Large amount of fish might be fatal if DO concentration reaches 1-2 mg/l for several hours. Commercial DO probes have been in market with various types from different manufacturers. Most DO sensors employ sensing technologies including optical based method which is commonly referred to as luminescent and the Clark electrochemical or membrane-covered electrode [1]. Hemera, Sacomam, Thethys Instruments, YSI supply DO sensors working based on optical method. Amount of oxygen is proportional to luminescence intensity of chemical dyes. Horiba, Greenspan Analytical, In-situ, Eutech-Thermo Scientific use polarographic electrodes and electrochemical advanced technologies. However, high cost of these sensors has hampered the application of these sensor systems in small and medium shrimp/fish farms. Therefore, the need of compact DO sensor development with low cost, simple operation has been increasing and has attracted the interest of many scientists in this field. The optical method was used to measure DO levels in micro-scale water flows with an oxygen quenchable luminophore platinum (II) octaethyporphrin bound with thin polystyrene film [2]. Another work also used the optical method to measure DO concentrations within cell culture media by analyzing fluorescence intensity of dextran and ruthenium complex immobilized on poly(ethylene glycol) diacrylate microarray [3]. However, the optical method with a complicated setup and result analysis is only suitable for measurement in laboratories and it is difficult to apply for practical applications in aquaculture farming. Besides, electrochemical technology has been widely used for DO sensor development [4-6]. Chou et al. [5] investigated gold/Nafion electrodes with H2SO4 acid as a supporting electrolyte for DO detection in water. The sensitivity of DO sensors was 38.4 µA/ppm and the lowest limit of DO was 3.8 ppm. Peng Wang and his co-workers [6] used a silicon microprobe including Nafion coating on three platinum electrodes to measure DO levels at 0% atm, 10.5% atm and 21% atm. The difference between two curves in linear sweep voltammetry in air-equilibrated DI water and oxygen depleted DI water was attributed to oxygen reduction. However, the aim of this research is to measure DO concentration in blood for biological and biomedical use; hence, DO concentrations were not high. Moreover, repeatability was not satisfactory and the calibration curves were not established. Recently conductive polymers have been investigated for different types of sensors [7-15] but only few of them have been used for DO measurements. Tieman et al. [16] used ionically conductive polymer poly(dimethyldiallylammonium chloride) coated on graphite electrodes as working electrodes to measure DO.
  • 2. Polyaniline-coated electrodes for detection of dissolved oxygen levels in solution International organization of Scientific Research 15 | P a g e The authors used Pt auxilary electrode and Ag/AgCl reference electrode. DO concentration was measured by reduction at potential of -0.40 V versus Ag/AgCl. The linearity of concentration was in the range of 1.4-9.3 ppm. In addition, polyaniline (PANI) was once investigated for DO measurement [17]. A platinum electrode coated with PANI was examined as a potentiometric sensor. This PANI-coated electrode gave approximately 97 mV/decade of the oxygen concentration and a voltage signal of about 400 mV. A detection limit of DO was estimated of about 5 x 10-7 mol/l. The proposed mechanism based on the oxidation of reduced form (leucoemeraldine base PANI) leading to conversion into conducting salt PANI in the presence of oxygen bubbling in H2SO4 solution. In this research, pH of the medium was kept lower than 7 by a mixture of H2SO4 and NaOH. The oxidative doping due to existing H2SO4 1 M solution could have caused conversion into conducting salt PANI. Therefore, the proposed mechanism of oxidation by oxygen bubbling is still not clear. In this report three forms of PANI including Emeraldine Base Polyaniline (EB-PANI), Emeraldine Salt Polyaniline (ES-PANI) and Leuco Emeraldine Base Polyaniline (LEB-PANI) (Fig. 1) were investigated as sensing materials for detection of DO levels in solution. UV-Vis spectroscopy and cyclic voltametry (CV) were used to evaluate the DO sensitivity of these polymers. The increases in current of the CV measurements in corresponding to different DO concentrations were observed. The results showed possibility of using PANI- coated electrodes for DO sensors. Figure 1. (a) Structure of EB-PANI. (b) ES-PANI. (c) LEB-PANI. II. EXPERIMENTAL 2.1. Materials Emeraldine Base Polyaniline (EB-PANI) (99.7%, Mw 20,000), dimethyl sulfoxide (DMSO) (99.9%), phenylhydrazine (PHZ) (97%) were bought from Sigma Aldrich and hydrochloric acid (HCl) (37%) was bought from Merck. Disodium hydrogen phosphates dodecahydrate (Na2HPO4.12H2O), sodium dihydrogen phosphate dihydrate (NaH2PO4.2H2O) and sodium chloride (NaCl) were used to prepare testing solutions. All the chemicals were used as purchased without further purification. Oxygen (O2) gas (99.999%) was supplied by Singapore Oxygen Air Liquide Pte. Ltd. 2.2. Preparation of polymer solutions 2.2.1. EB-PANI solution Mix 30 mg of EB-PANI with 2.7 ml DMSO, stirring for 24 hours with a magnetic stirrer at room temperature to obtain EB-PANI solution. The solution was then filtered through a filter paper. The obtained polymer solution had dark blue color. 2.2.2. ES-PANI solution EB-PANI powder was protonated by HCl 0.1 M solution to obtain ES-PANI (green color). The salt was filtered and rinsed by DI water to remove residual acid. Then ES-PANI was dried under vacuum at 60C in a vacuum oven. 30 mg ES-PANI was dispersed in 2.7 ml DMSO to obtain ES-PANI solution. This solution underwent ultrasonification and centrifugal mixing. 2.2.3. LEB-PANI solution ES-PANI was mixed with PHZ 97% to obtain LEB-PANI. Then the powder was rinsed in DI water, acetone and methanol to remove residual PHZ and dried at 60C in the vacuum oven. After that LEB-PANI was dissolved in DMSO and this solution underwent ultrasonification and centrifugal mixing. 2.3. Fabrication of electrochemical chips Electrochemical chips included one working electrode and one counter electrode (dimension of 3 mm x 9 mm) put separately with a distance of 3 mm. Both electrodes were made of platinum (Pt). Pt layers were sputtered with Univex 350 (Leybold, Germany) and patterned on silicon (Si) wafers coated with 700 nm thick SiO2 by photolithography and lift-off techniques. The Pt electrodes had a thickness of 190 nm with a titanium adhesion layer. 4 µl of the PANI solutions was drop-casted on the working electrode of the electrochemical chips and these chips were heated in vacuum oven at 60°C for 24 hours to remove completely DMSO.
  • 3. Polyaniline-coated electrodes for detection of dissolved oxygen levels in solution International organization of Scientific Research 16 | P a g e 2.4. Characterization UV-Vis absorption spectroscopy of the PANI samples was obtained from a double beam spectrophotometer Cary 100 (Varian, USA) in the wavelength range from 200 to 800 nm. To check oxygen sensitivity of the PANI forms, the quartz cuvettes containing the PANI solutions were bubbled with 0.5 psi of oxygen gas in 1-5 minutes. After each time of oxygen bubbling, UV-Vis spectra were recorded and compared. The variation of the UV-Vis spectra in the presence of oxygen (if any) can be applied for optical DO sensors using PANI layers as a color indicator. The thickness of the Pt electrodes and the polymer films after fabrication was measured by a stylus profiler Dektak 6M (Veeco, USA). 2.5. Cyclic voltametry (CV) measurements Testing electrolyte solutions were prepared from phosphate buffer which was made from 6 g Na2HPO4.12H2O, 0.5 g NaH2PO4.2H2O and DI water to obtain 200 ml testing solution with pH 7.3. Then 4 g NaCl was added to make the solution have salinity of 20‰. The testing solution was prepared to imitate the conditions of water in shrimp ponds. The testing solutions were bubbled with nitrogen (N2) gas for 15 minutes to minimize DO concentration in the solution, approximately 1.5 ppm. Then oxygen (O2) gas was bubbled through the solution with a pressure of 1 psi for 15-90 seconds to create various DO concentrations in the solutions. DO concentrations were measured by a commercial DO probe DO6400TC/T (Sensorex, US) with a measurement range of 0-20 mg/l, 0.01 mg/l resolution and ±1.5% accuracy. The polymer coated chips were dipped in the testing solutions in which the polymer part was immersed completely. CV measurements were performed with a potentiostat/galvanostat PGSTAT 302N (Metrohm, Autolab, the Netherlands) and a silver/silver chloride (Ag/AgCl) reference electrode RE-6 (Basi Inc., US). The polymer chips were mounted on two crocodile clips (Fig. 2). The current was set from 10 µA to 10 mA, the scanning rate was 50 mV/sec, the temperature was about 28 C ± 1 C. Figure 2. The polymer chip mounted on two crocodile clips. III. RESULTS AND DISCUSSION 3.1. UV-Vis spectroscopy of EB-PANI, ES-PANI and LEB-PANI UV-Vis spectroscopy of EB-PANI, ES-PANI and LEB-PANI was carried out with Cary 100 (Varian, USA) in the wavelength range from 200 to 800 nm. UV-Vis absorption spectra of EB-PANI, ES-PANI and LEB-PANI in DMSO are shown in Fig. 3. Figure 3. UV-Vis absorption spectra of EB-PANI (solid line), ES-PANI (dash line) and LEB-PANI (dash dot line) in DMSO.
  • 4. Polyaniline-coated electrodes for detection of dissolved oxygen levels in solution International organization of Scientific Research 17 | P a g e UV-Vis spectrum of EB-PANI in DMSO solvent shows two peaks at wavelength of 324 nm and 625 nm. This result is in good agreement with literature values [18-20]. The absorption peak at 324 nm is assigned to the π - π* transition of the benzenoid rings. The peak at 625 nm originates from the exciton transition of the quinoid rings [19]. The absorption spectrum of HCl-doped PANI in DMSO solvent showes peaks at 289 nm, 456 nm and 663 nm. The two bands have been assigned to the optical absorption of the polaron band [18, 20]. The absorption spectrum of LEB-PANI has only one peak at 292 nm, which is in good agreement with the published value [21]. The UV-Vis spectra of PANI solutions with and without pure oxygen bubbling were recorded. The quartz cuvettes containing the polymer solutions were bubbled with pure oxygen for 1-5 minutes. After each duration of oxygen bubbling, UV-Vis spectra were recorded and compared (see Supplementary Materials for the UV-Vis spectra of the PANI forms with oxygen bubbling). The UV-Vis spectra of EB-PANI, ES-PANI and LEB- PANI solutions almost did not change as the oxygen gas was bubbled into the solutions. However, the UV-Vis spectra of LEB-PANI solutions which included residual PHZ showed an increasing peak at 450 nm with an increase in DO concentration (Figure S4 of Supplementary Materials). The color of LEB-PANI changed from transparent to green when oxygen was bubbled into the solution. It was found that the color change was due to oxidation of the residual PHZ in the LEB-PANI preparation (Figure S5 of Supplementary Materials). 3.2. Platinum electrodes Series of the electrochemical chips with Pt electrodes were fabricated by photolithography, sputtering and lift-off processes. Fig. 4 shows the fabricated Pt electrodes and a diced chip coated with the PANI film on the working electrode. The thickness of the PANI films varied from 500 nm to several micrometers. Figure 4. The platinum electrodes on a silicon wafer before dicing into seperate chips (left) and the diced chip coated with the PANI film on the working electrode (right). 3.3. CV measurements of the PANI-coated chips CV measurements were performed with the PANI-coated platinum electrode as the working electrode, another Pt counter electrode and the Ag/AgCl reference electrode. For each CV measurement, at least two chips of the same batch were measured to check reproducibility. The potential was scanned with various ranges and it was found that the stable potential range of the fabricated electrodes was from -0.6 V to +1.1 V. Various DO concentrations in the testing solutions were created by bubbling N2 for 15 minutes and O2 for 15-90 seconds. In this study, high DO concentrations were used to check the DO sensitivity of the PANI forms. Basing on the measurement of the commercial DO probe, 1 minute of oxygen bubbling could create 16.05 mg/l of DO concentration in the testing solutions. Due to the limited measurement range of the commercial DO probe (i.e. 20 mg/l), the DO concentrations created by durations of oxygen bubbling larger than 1 minute were estimated correspondingly. 3.3.1. EB-PANI coated chips Fig. 5 shows cyclic voltammograms of the EB-PANI coated chips immersed in the testing solutions which were bubbled with N2 and O2 for 15, 30, 60 and 90 seconds (DO concentrations ranged from 1.50 ppm to 26.00 ppm).
  • 5. Polyaniline-coated electrodes for detection of dissolved oxygen levels in solution International organization of Scientific Research 18 | P a g e Figure 5. CV plots of the EB-PANI chips with N2 purging (bold solid line), O2 bubbling for 15 seconds (dot line), 30 seconds (dash line), 60 seconds (solid line), 90 seconds (dash dot dot line). When DO concentration in the solution increases, the magnitude of current increases with DO concentrations from 0.015 mA to 0.085 mA at the cathode peak of -0.3 V. There is large difference in current when DO concentration increases from 1.5 ppm (corresponding to N2 purging) to 26 ppm (90 seconds of O2 bubbling). The increase in current exhibits the DO sensitivity of the EB-PANI films. The change in current corresponding to estimated DO concentration is 2.86 µA per 1 ppm of DO concentration. This sensitivity is however lower than the published sensitivity of 38.4 µA/ppm [5] and 12.5 µA.cm-2 /ppm [16]. This difference in sensitivity is due to different electrode materials and different sizes of electrodes. The relationship between the cathode current at -0.3 V and DO concentrations was shown in Fig. 6. The increase in current of the EB-PANI thin films follows an exponential Decay function. It was observed that the current increased steeply from 1.5 ppm to 10 ppm DO. Beyond 16 ppm DO, the current did not change much and it became stable. It could be attributed to the saturation of DO in the testing solution with 90 seconds of oxygen bubbling. Therefore after this measurement, the duration of oxygen bubbling was limited not more than 90 seconds. Figure 6. The relationship between the cathode current and DO concentrations at -0.3 V of the EB-PANI coated chip. 3.3.2. ES-PANI coated chips Cyclic voltammograms of the ES-PANI coated chips which were bubbled with N2 and O2 for 15, 30, 60 and 90 seconds (DO concentrations ranged from 1.50 ppm to 26.00 ppm) are shown in Fig. 7.
  • 6. Polyaniline-coated electrodes for detection of dissolved oxygen levels in solution International organization of Scientific Research 19 | P a g e Figure 7. CV plots of the ES-PANI chips with N2 purging (bold solid line), O2 bubbling for 15 seconds (dot line), 30 seconds (dash line), 60 seconds (solid line), 90 seconds (dash dot dot line). When DO concentration in the solution increases from 1.5 ppm to 9.77 ppm, the magnitude of current increases with DO concentrations from 0.026 mA to 0.125 mA at the cathode peak of -0.05 V. In comparison with the EB-PANI chips, the cathode peak is found to shift to -0.05 V. The change in current of the ES-PANI chips is much larger (11.97 µA/ppm of DO concentration) in the range of 1.5-9.77 ppm DO. However, the current decreases when DO concentration increases to 16.05 ppm (corresponding to 60 seconds of oxygen bubbling). Two ES-PANI chips of the same measurement gave the same results. The instable increase of the current can be attributed to the scratched ES-PANI film after the CV measurements (Figure S6 of Supplementary Materials). It should be noted that the ES-PANI layer was not durable. The layer was stripped off gradually after the CV scans from 1.5 ppm to 26 ppm DO and was removed completely after two times of CV scans. The stripping of the ES-PANI layer can be due to dispersion of the conductive ES-PANI particles which were charged under an electrical field in the electrolyte solution. 3.3.3. LEB-PANI coated chips Fig. 8 shows the cyclic voltammograms of the LEB-PANI coated chips exposed to oxygen bubbling for 15, 30, 60 and 90 seconds. Similar to the ES-PANI chips, the cathode peak of the LEB-PANI is -0.05 V. This peak is similar to the reduction peak of -0.03 V of LEB-PANI in CV measurements at pH 1 of Edward Song et al. [22]. This potential indicated the reduction of LEB-PANI film at the cathode. Large increase in magnitude of the current is also observed when DO concentrations increased from 1.5 ppm (corresponding to N2 purging) to 26 ppm (1.5 minutes of oxygen bubbling). The current increases with DO concentrations from 0.018 mA to 0.1 mA at -0.05 V. Fig. 9 shows the relationship between the cathode current at -0.05 V and DO concentrations. The increase in current of the LEB-PANI thin films also follows an exponential Decay function similar to the EB- PANI films. The current of the LEB-PANI films increases steeply from 1.5 ppm to 10 ppm DO and does not change much after 15 ppm DO. The change in current corresponding to estimated DO concentration is 3.35 µA per 1 ppm of DO concentration. The results showed that the DO sensitivity of the LEB-PANI coated chips is higher than the EB-PANI form. Figure 8. CV plots of the LEB-PANI chips with N2 purging (bold solid line), O2 bubbling for 15 seconds (dot line), 30 seconds (dash line), 60 seconds (solid line), 90 seconds (dash dot dot line).
  • 7. Polyaniline-coated electrodes for detection of dissolved oxygen levels in solution International organization of Scientific Research 20 | P a g e Figure 9. The relationship between the cathode current and DO concentrations at -0.05 V of the LEB-PANI coated chip. 3.3.4. Stability of the PANI electrodes To check reversibility and consistency of the current increase between nitrogen and oxygen bubbling, the CV measurements of the PANI chips were performed when the chipswere exposed to bubbling nitrogen (15 minutes) and oxygen (30 seconds) for two times. CV plots after two times of oxygen bubbling (9.77 ppm DO) are almost the same (Figure S7 of Supplementary Materials). Therefore, the PANI electrode chips have good reversibility. To check stability of the PANI electrodes, the CV measurements of the PANI electrodes were performed as long as the CV plots could be obtained. DO concentration varied from 1.50 ppm to 26.00 ppm for the CV measurements of each chip. The EB-PANI electrodes could be measured for 5 weeks and the LEB-PANI layers could be measured for 7 weeks whereas the ES-PANI electrodes could only be measured for 1-2 weeks. The increase of the cathode current and DO concentrations of the EB-PANI and LEB-PANI electrodes in 5 and 7 weeks plotted from the CV scans also follows an exponential Decay function (Figure S8 and Figure S9 of Supplementary Materials). The variations in the current of the EB-PANI and LEB-PANI chips were observed. The current signal of the EB-PANI chips was stable for 2 weeks and decreased in the following weeks. For the LEB-PANI chips, although there was large fluctuation at high DO concentrations, the current was somehow stable at below 10 ppm DO after 7 weeks. 3.3.5. Possible sensing mechanism The cathode reaction relating to the redox reaction of oxygen in solution with the PANI films coated on the Pt electrode is shown in the following reactions (1) and (2). For simplicity, a general sign is used in the reaction (1), in which PANI represents different forms of PANI and [PANI]n- represents the different electrochemical redox states of PANI which have been reported in many studies [23-25]. PANI + n e → [PANI]n- (1) O2 + 4 e → 2 O2- (2) Depending on the form of PANI (EB-PANI, ES-PANI or LEB-PANI) the difference in the consumed number of electrons causes the different cathodic potentials (as observed in the CV measurements). The cathodic potential of EB-PANI is more polarized than that of LEB-PANI. This leads to the fact that the oxygen reduction with the LEB-PANI electrode is easier than with the EB-PANI and the higher DO sensitivity of the LEB-PANI coated chips. In the reaction (1), the PANI film receives electrons to transform into a charge state [PANI]n- and a part of charge is transferred to oxygen to cause the reaction (2). The CV plots of the PANI forms show only one cathode peak. It could be implied that the reduction reaction is followed by one-step reaction. The reduced oxygen O2- may be combined with [PANI]n- to produce a new compound like [PANI-O2] (Fig. 10). The same structure of [PANI-O2] was also proposed by D. V. Zhuzhel’skii [26].
  • 8. Polyaniline-coated electrodes for detection of dissolved oxygen levels in solution International organization of Scientific Research 21 | P a g e Figure 10. Mechanism of reduced oxygen at the PANI-coated Pt electrode. Figure 11. Resonance in [PANI-O2] structure to form oxoniumcations. In the presence of DO in solution, two oxygen atoms approach the PANI films and attach to the benzene rings of the PANI. The resonance in the structure then occurs and oxonium ions are generated (Fig. 11). When there are more oxygen molecules, more oxonium ions are formed and the conductivity of the PANI films increases. The increase in oxygen concentration leads to the increase in current intensity. However, oxygen present in the solution can react at some extent. When all the reaction sites for oxygen are occupied, the current does not increase much as DO concentration increases. This leads to the observed saturation of the increase in the current (Fig. 6 and Fig. 9). IV. CONCLUSIONS In this study we have investigated the DO sensitivity of the EB-PANI, ES-PANI and LEB-PANI films. The UV-Vis spectra of the PANI solutions did not show any change in the presence of dissolved oxygen. However, the CV measurements of the EB-PANI, ES-PANI and LEB-PANI films coated on the Pt electrodes showed the increase in current as DO concentration in the solutions increased. The steep current increase of the PANI films in the DO range of 1.5-10 ppm is suitable for monitoring of water quality in aquaculture farming. The correlation of the current and DO concentrations was studied for the EB-PANI and LEB-PANI electrodes. The higher DO sensitivity (3.35 µA/ppm) and higher stability of the LEB-PANI films enables the feasibility of using LEB-PANI films as sensing materials in DO sensors. V. ACKNOWLEDGMENTS The authors highly appreciate the financial support of Vietnam National University - Ho Chi Minh City for this study through the grant number B2013-76-04. REFERENCES [1]. YSI. The dissolved oxygen handbook: A practical guide to dissolved oxygen measurements. www.ysi.com/weknowdo, 2009. [2]. Song, D.H.; Kim, H.D.; Kim, K.C. Dissolved oxygen concentration field measurement in micro-scale water flows using ptoep/ps film sensor. Optics and Lasers in Engineering, 50, 2012, 74-81. [3]. Lee, S.; Ibey, B.L.; Coté, G.L.; Pishko, M.V. Measurement of ph and dissolved oxygen within cell culture media using a hydrogel microarray sensor. Sensors and Actuators B: Chemical, 128, 2008, 388- 398. [4]. Eminaga, Y.; Brischwein, M.; Wiest, J.; Clauss, J.; Becker, S.; Wolf, B. Self calibration of a planar dissolved oxygen sensor. Sensors and Actuators B: Chemical, 177, 2013, 785-791. [5]. Chou, T.C.; Ng, K.M.; Wang, S.H. Gold-solid polymer electrolyte sensor for detecting dissolved oxygen in water. Sensors and Actuators B: Chemical, 66, 2000, 184-186. [6]. Wang, P.; Liu, Y.; Abruña, H.D.; Spector, J.A.; Olbricht, W.L. Micromachined dissolved oxygen sensor based on solid polymer electrolyte. Sensors and Actuators B: Chemical, 153, 2011, 145-151.
  • 9. Polyaniline-coated electrodes for detection of dissolved oxygen levels in solution International organization of Scientific Research 22 | P a g e [7]. Bartlett, P.N.; Birkin, P.R. The application of conducting polymers in biosensors. Synthetic Metals, 61, 1993, 15-21. [8]. Partridge, A.C.; Jansen, M.L.; Arnold, W.M. Conducting polymer-based sensors. Materials Science and Engineering: C, 12, 2000, 37-42. [9]. Nicolas-Debarnot, D.; Poncin-Epaillard, F. Polyaniline as a new sensitive layer for gas sensors. Analytica Chimica Acta, 475, 2003, 1-15. [10]. James, D.; Scott, S.M.; Ali, Z.; O’Hare, W.T. Chemical sensors for electronic nose systems. Microchimica Acta, 149, 2005, 1-17. [11]. Bai, H.; Shi, G.Q. Gas sensors based on conducting polymers. Sensors, 7, 2007, 267-307. [12]. Lange, U.; Roznyatovskaya, N.V.; Mirsky, V.M. Conducting polymers in chemical sensors and arrays. Analytica Chimica Acta, 614, 2008, 1-26. [13]. Liu, J.; Agarwal, M.; Varahramyan, K. Glucose sensor based on organic thin film transistor using glucose oxidase and conducting polymer. Sensors and Actuators B: Chemical, 135, 2008, 195-199. [14]. Tolani, S.; Craig, M.; DeLong, R.; Ghosh, K.; Wanekaya, A. Towards biosensors based on conducting polymer nanowires. Analytical and Bioanalytical Chemistry, 393, 2009, 1225-1231. [15]. Doan, T.C.D.; Ramaneti, R.; Baggerman, J.; van der Bent, J.F.; Marcelis, A.T.M.; Tong, H.D.; van Rijn, C.J.M. Carbon dioxide sensing with sulfonated polyaniline. Sensors and Actuators B: Chemical, 168, 2012, 123-130. [16]. Tieman, R.S.; Heineman, W.R.; Johnson, J.; Seguin, R. Oxygen sensors based on the ionically conductive polymer poly(dimethyldiallylammonium chloride). Sensors and Actuators B: Chemical, 8, 1992, 199- 204. [17]. Shim, Y.-B.; Stilwell, D.E.; Park, S.-M. Electrochemistry of conductive polymer x: Polyaniline-based potentiometric sensor for dissolved oxygen. Electroanalysis, 3, 1991, 31-36. [18]. Huang, W.S.; MacDiarmid, A.G. Optical properties of polyaniline. Polymer, 34, 1993, 1833-1845. [19]. Hua, M.-Y.; Hwang, G.-W.; Chuang, Y.-H.; Chen, S.-A.; Tsai, R.-Y. Soluble n-doped polyaniline:  Synthesis and characterization. Macromolecules, 33, 2000, 6235-6238. [20]. Shimano, J.Y.; MacDiarmid, A.G. Polyaniline, a dynamic block copolymer: Key to attaining its intrinsic conductivity? Synthetic Metals, 123, 2001, 251-262. [21]. McCall, R.P.; Ginder, J.M.; Leng, J.M.; Ye, H.J.; Manohar, S.K.; Masters, J.G.; Asturias, G.E.; MacDiarmid, A.G.; Epstein, A.J. Spectroscopy and defect states in polyaniline. Phys. Rev. B, 41, 1990, 5202-5213. [22]. Edward Song; Choi, J.-W. Conducting polyaniline nanowire and its applications in chemiresistive sensing. Nanomaterials, 3, 2013, 498-523. [23]. Gospodinova, N.; Terlemezyan, L. Conducting polymers prepared by oxidative polymerization: Polyaniline. Progress in Polymer Science, 23, 1998, 1443-1484. [24]. Genies, E.M.; Tsintavis, C. Redox mechanism and electrochemical behaviour or polyaniline deposits. Journal of Electroanalytical Chemistry and Interfacial Electrochemistry, 195, 1985, 109-128. [25]. Zhuzhel’skii, D.V.; Krylova, V.A.; Ivanov, V.D.; Malev, V.V. Mechanism of electrochemical reactions of polyaniline films formed under the conditions of cathodic oxygen reduction. Russ J Electrochem, 45, 2009, 145-151.