2. Fortunately regardless the software used there
are fixed tools to be used to manipulate
patients’ complaint
3. It is the threshold of electrical stimulation and refers to the
least amount of stimulation a recipient can detect when
electrical signals are delivered to individual electrodes.
4. It is the most comfortable, the maximum comfortable level or
sounds considered loud but comfortable “ according the
manufacturer”.
6. This parameter determines the range of acoustic signals
that are mapped into the recipient's electric dynamic
range.
The lower edge of the dynamic range determines the
acoustic input level that is mapped near the threshold of
electrical stimulation while the upper limit of the IDR
determines the acoustic signals that are mapped near
the maximum electrical stimulation.
7.
8.
9. Low frequency components of the signal is usually
delivered to electrode contacts that are higher than that
of the input signal. As a result, the pitch of the human
voice is much higher than what the recipient was
accustomed.
Furthermore, despite the sophisticated nature of the
speech coding strategies used with CI still the signal is
markedly different from the normal peripheral auditory
system in response to acoustic stimulation.
The auditory system has not received stimulation for
certain sounds (high frequencies) so, reducing the ability
to processing these sounds again.
Less stimulation sites (electrode contacts ) in CI relative
to 3500 IHCs in normal cochlea.
Different signal type than
that of a hearing aid
10. A period of acclimatization
is necessary for the
auditory system to adjust
to the new signal.
Young children who didn’t
experience sounds before
require extensive exposure
to speech and
environmental sounds
(auditory verbal therapy).
11.
12. Before we judge a complaint of patient a
period of acclimatization should be
considered first after the initial activation.
Sound quality and speech recognition
always improve over the first few months
after activation, so, re-programing during
the early stages of implant use may not
result in better sound quality.
13. Global reduction of the upper stimulation level but without
resulting in soft sounds.
Loudness balances across electrode array.
Reduction of the stimulation rate.
Or reduction of gain for the LF or HF channels (for CIS
strategies).
Sweep to identify the electrode with the ringing character.
Disable the most basal electrodes.
Increasing IDR.
Increase microphone sensitivity.
Increase threshold levels.
Increased loudness growth function (maplaw).
Sweep at USL to ensure pitch percepts in expected
tonotopic pattern
different stimulation rate, or another speech coding
strategy
• Echo or boomy
sound quality.
• Ringing
pinging or
tinny sound
quality.
• Sound is
muffled fuzzy
or not clear.
• Overall poor
sound quality
performance.
14. Ensure loudness balance across USL in electrode array
Global decrease of USL if speech of the others become
too soft, increase sensitivity, T levels or loudness growth
for low level inputs.
Too loud own voice but speech from others is comfortable
15. After considering the acclimatization
issue, reduce microphone sensitivity
and/or IDR.
Adjust T levels to balance loudness at
50% of DR
Global decrease of T levels.
Ensure loudness balance across
electrode array at USL.
Global decrease of USL
Decrease the USL in the LF channels.
Reduce stimulation levels at LF
channels and increase Stimulation
levels at MF and HF channels.
• Buzzing, frying or
humming sounds.
• Roaring.
• Hollow sound.
16. Too loud
Reduce USL
Reduce volume
control setting
Too soft
Increase USL
Increase volume control
setting
Re-measure T levels
Increase Sensitivity if the
complaint was to the low
level speech
Reduce noise reduction
used (e.g. clear voice)
17. Reassess volume control setting
Increase T levels and/or increase IDR, increase T
levels, or increase maplaw setting.
18.
19. Discomfort,
pain, non-
auditory
percept such
as tinnitus,
tactile
sensation
Measure ESRT to find out is USL are too high
Sweep USL to find suspected channel,
increase PW if not resolved switch off the
electrode.
Disable 2-3 electrodes at a time during live
stimulation to find the problem (increase PW
or disable)
Increase PW for all electrodes if a lot of
electrodes are offended.
Reduce USL
Decrease stimulation rate (high rate causes
tinnitus/vibration
Change electrode coupling mode
(monopolar/bipolar).
Change speech coding strategy.
Reduce volume control setting.
20. Facial nerve
stimulation
Single channel stimulation
Increase PW on channel or globally
if PW increase adversely affected
loudness growth disable the
electrode.
Stimulation in live speech mode:
Sweep USL to find a problem
channels.
May increase USL to elicit the facial
stimulation.
Disable 2-3 electrodes at a time
during live stimulation to find the
problem (increase PW or disable).
If stimulation still exists with 25%
electrodes disabled, change electrode
coupling mode.
21. In contrast to hard failure in which a cochlear implant shows more
than a sudden loss of function, such as no communication between
the internal and external devices.
A “soft failure” is a term used by the cochlear implant industry to
describe a situation where a patient’s hearing performance has
noticeably decreased but the implant integrity testing shows that
the implant is functioning as normal. In other words, the implant
testing does not show a problem with the internal device, but the
patient is not hearing as well as they used to.
22. The Consensus Statement from 10th Symposium
on Cochlear Implantation in Children assessed
that soft failure:
• Is an uncommon occurrence
• Is a working diagnosis
• Only confirmed by removal, examination and identification
of failure mode.
• Diagnosis begins with awareness of common
presentations
• Declining performance, aversive symptoms or
intermittent function
• Diagnosis can only be made with complete otological
examination, CT scan, expert re-programming,
exchange of all external hardware, and integrity testing
by the manufacturer.