CS 3600
Summary from “Trueness and precision of 5 intraoral scanners in
the impressions of single and multiple implants: a comparative in
vitro study”
Overview
Article Name
“Trueness and precision of 5 intraoral scanners in the
impressions of single and multiple implants: a comparative in
vitro study”
Authors
• Francesco Guido Mangano, Uli Hauschild, Giovanni Veronesi,
Mario Imburgia, Carlo Mangano, and Oleg Admakin
Date Published
6 June 2019
Background
• Very few peer reviewed studies have evaluated the clinical
efficacy of intraoral scanners in implantology.
• Even fewer studies compare the accuracy of the various
intraoral scanners.
• This study compares the trueness and precision of five
intraoral scanners, and determines which have both accurate
and consistent results.
2
What are Trueness and Precision?
In this study:
Trueness is quantified by distances between generated model and reference model
(smaller value means better trueness).
Precision is quantified by distances between repeatedly generated models (smaller
value means better precision).
Accuracy is a combination of trueness and precision. An intraoral scanner must
perform well in both trueness and precision tests to be considered “highly accurate.”
3
Study Material and Methods
10 scans were performed by a single digital dentistry expert per indication per intraoral
scanner device with a zig-zag technique in the same environmental conditions.
2 reference maxillary plaster models were used as a reference:
• A - Partially edentulous maxillary arch with:
• Implant analog on tooth #23 (implant single support crown simulation)
• Implant analog on teeth #14 & 16 (partial implant supported case simulation)
• B - Totally edentulous maxillary arch:
• Implant analog on teeth #11, 14, 16, 21, 24 & 26 (fixed full arch implant supported denture simulation)
Scanbodies used : Megagen® PEEK scanbodies to enhance scans (no light reflection).
Reference : 3 scans made per indication with Freedom UHD ® desktop scanner and
superimposed in Geomagic Studio 2012® to validate the reference models.
4
How do these intraoral scanners compare in accuracy?
Implantology Intraoral Scanner Comparative Study
5
TRIOS 3
EMERALD
DWIO
CS 3600
OMNICAM
Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners
in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
Implantology Intraoral Scanner Comparative Study
Scanner Producer Technology of
Acquisition
Powder Color System
CS 3600 Carestream
Dental,
Atlanta, Georgia, USA
Structured light-Active
Speed 3D Video™
No Yes Proprietary files (CSZ), but
also open formats (PLY, STL)
immediately available
Trios 3 3-Shape,
Copenhagen,
Denmark
Structured light –Confocal
microscopy and Ultrafast
Optical Scanning™
No Yes Proprietary files (DCM)
available, but possibility to
export STL files via the
new Trios on Dental Desktop®
Omnicam Dentsply-Sirona,
York, Pennsylvania,
USA
Structured light -Optical
triangulation and confocal
microscopy
No Yes Proprietary files (CS3, SDT,
CDT, IDT) are available,
possible to export STL files via
the Cerec Connect®
DWIO Dentalwings,
Montreal, Quebec,
Canada
Blue laser-Multiscan
Imaging™ technology
No No Proprietary files (XORDER),
but also open formats (STL)
immediately available
Emerald Planmeca,
Helsinki, Finland
Red, green and blue
lasers-Projected Pattern
Triangulation™
No Yes Open formats (PLY, STL)
immediately available
6
Summary of 5 Intraoral Scanners Used in the Study
Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners
in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
Single Crown, Best Result in Trueness (μM)
7
Deviation: 22 μm (20)
Triangles: 179,496
Deviation: 14 μm (13)
Triangles: 34,980
Deviation: 31 μm (26)
Triangles: 81,712
Deviation: 23 μm (21)
Triangles: 68,281
Deviation: 21 μm (28)
Triangles: 123,417
TRIOS 3 CS 3600 EMERALD DWIO OMNICAM
Best Result
[Carestream Dental Note: Triangle number is not a measure of acquisition resolution or
mesh quality. In fact, in having the best trueness with the smallest number of triangles, CS
3600 demonstrates the most efficient/optimal mesh generation (by placing more triangles in
areas with more details and vice versa), with minimized mesh file size while maintaining
accuracy.]
Single crown: best result in trueness (standard deviation) in μm, for the 5 examined
scanners, and the number of triangles composing each mesh.
Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners
in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
Partial Prosthesis, Best Result in Trueness (μM)
8
Deviation: 28 μm (24)
Triangles: 207,159
Deviation: 21 μm (25)
Triangles: 55,968
Deviation: 42 μm (32)
Triangles: 113,162
Deviation: 39 μm (39)
Triangles: 135,028
Deviation: 29 μm (32)
Triangles: 192,967
8
TRIOS 3 CS 3600 EMERALD DWIO OMNICAM
Best Result
Partial prosthesis: best result in trueness (standard deviation) in μm, for the 5 examined
scanners, and the number of triangles composing each mesh.
Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners
in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
[Carestream Dental Note: Triangle number is not a measure of acquisition resolution or
mesh quality. In fact, in having the best trueness with the smallest number of triangles, CS
3600 demonstrates the most efficient/optimal mesh generation (by placing more triangles in
areas with more details and vice versa), with minimized mesh file size while maintaining
accuracy.]
Full Arch, Best Result in Trueness (μM)
9
Deviation: 37 μm (44)
Triangles: 326,696
Deviation: 32 μm (49)
Triangles: 83,735
Deviation: 57 μm (58)
Triangles: 178,507
Deviation: 73 μm (64)
Triangles: 227, 491
Deviation: 47 μm (55)
Triangles: 322,929
9
TRIOS 3 CS 3600 EMERALD DWIO OMNICAM
Best Result
Full arch: best result in trueness (standard deviation) in μm, for the 5 examined
scanners, and the number of triangles composing each mesh.
Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners
in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
[Carestream Dental Note: Triangle number is not a measure of acquisition resolution or
mesh quality. In fact, in having the best trueness with the smallest number of triangles, CS
3600 demonstrates the most efficient/optimal mesh generation (by placing more triangles in
areas with more details and vice versa), with minimized mesh file size while maintaining
accuracy.]
Mean Trueness/Standard Deviation
Scanner Single Crown
Mean ± SD
Partial Prosthesis
Mean ± SD
Full Arch
Mean ± SD
p-value1
Trios 3
22.3 ± 0.5† 28.5 ± 0.5†,‡,• 46.3 ± 4.9†,‡,• <.0001
CS 3600
15.2 ± 0.8‡,#,§ 23.0 ± 1.1^,§,# 44.9 ± 8.9^,§,# <.0001
Emerald
43.1 ± 11.5†,‡,•,^ 49.3 ± 5.5†,^,° 66.3 ± 5.6†,^,° <.0001
DWIO
27.8±3.2#,• 49.8±5.0‡,§,* 92.1±24.1‡,§,°,* <.0001
Omnicam
28.4±4.5§,^ 38.1±8.8•,#,°,* 70.4±11.9•,#,* <.0001
10
The same symbol after SD indicates differences in trueness between scanner pairs (Tukey-adjustment for multiple comparison). Minimum
significant difference across scanners: 7.3μm, 6.6μm, 16.8μm for single crown (SC), partial prosthesis (PP) and full arch (FA), respectively. 1p-
value testing the interaction between scanner and context (SC vs. PP vs. FA) from non-parametric, Kruskall-Wallis test. A p-value >0.05
indicates no difference in scanner trueness according to the context.
Best Result
Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners
in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
Colorimetric Trueness Map Comparison
11
TRIOS 3 CS 3600 EMERALD DWIO OMNICAM
+0.100 mm
+0.088 mm
+0.077 mm
+0.065 mm
+0.053 mm
+0.042 mm
+0.030 mm
-0.030 mm
-0.042 mm
-0.053 mm
-0.065 mm
-0.077 mm
-0.088 mm
-0.100 mm
Best Result
Trueness in the single crown (SC), partial prosthesis (PP) and full-arch (FA) with the 5 examined intraoral scanners (IOSs): colorimetric maps.
The color maps indicated inward (blue) or outward (red) displacement between overlaid structures, whereas a minimal change was indicated by
a green color. For all three models (SC, PP, FA): the color scale ranged from a maximum deviation of +100μm and−100μm, with the best result
given by the deviations comprised between +30μm and−30μm (green color).
Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners
in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
Changes in Mean Trueness Comparison
Changes in mean trueness (standard deviation), in μm, for the 5 examined scanners, in
the different applications (single crown, SC vs. partial prosthesis, PP vs. full-arch, FA).
12
Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners
in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
CS 3600
Mean Precision/Standard Deviation
Scanner Single Crown
Mean ± SD
Partial Prosthesis
Mean ± SD
Full Arch
Mean ± SD
p-value1
Trios 3
15.2 ± 0.8†,‡,• 21.0 ± 1.9‡,• 35.6 ± 3.4†,‡,• <.0001
CS 3600
11.3 ± 1.1^,§,# 17.0 ± 2.3^,§,# 35.7 ± 4.3^,§,# <.0001
Emerald 32.8 ± 10.7†,^ 29.9 ± 8.9^,° 61.5 ± 18.1†,^,°,* 0.0007
DWIO 27.1 ± 10.7‡,§ 34.8 ± 10.8‡,§ 111.0 ± 24.8‡,§,°,ç <.0001
Omnicam
30.6 ± 3.3•,# 43.2 ± 9.4•,#,° 89.3 ± 14.0•,#,*,ç <.0001
13
The same symbol after SD indicates differences in precision between scanner pairs (Tukey-adjustment for multiple comparison). Minimum
significant difference across scanners: 8.8 μm, 9.8 μm, 19.4 μm for single crown (SC), partial prosthesis (PP) and full arch (FA), respectively.
1p-value testing the interaction between scanner and context (SC vs. PP vs. FA) from non-parametric, Kruskall-Wallis test. A p-value > 0.05
indicates no difference in scanner precision according to the context.
Best Result Very closely follows
best result
Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners
in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
Colorimetric Precision Map Comparison
14
Very Close to
Best Results
TRIOS 3 CS 3600 EMERALD DWIO OMNICAM
Best Result
Trueness in the single crown (SC), partial prosthesis (PP) and full-arch (FA) with the 5 examined intraoral scanners (IOSs): colorimetric maps.
The color maps indicated inward (blue) or outward (red) displacement between overlaid structures, whereas a minimal change was indicated by
a green color. For all three models (SC, PP, FA): the color scale ranged from a maximum deviation of +100μm and−100μm, with the best result
given by the deviations comprised between +30μm and−30μm (green color).
+0.100 mm
+0.088 mm
+0.077 mm
+0.065 mm
+0.053 mm
+0.042 mm
+0.030 mm
-0.030 mm
-0.042 mm
-0.053 mm
-0.065 mm
-0.077 mm
-0.088 mm
-0.100 mm
Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners
in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
Changes in Mean Precision Comparison
Changes in mean precision (standard deviation), in μm, for the 5 examined scanners, in
the different applications (single crown vs. partial prosthesis vs. full-arch)
15
CS 3600
Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners
in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
CS 3600
Results Summary
Single Crown:
CS 3600 had the best trueness (15.2 ± 0.8 μm) and the best
precision (11.3 ± 1.1 μm)
Partial Prosthesis
CS 3600 had the best trueness (23 ± 1.1 μm) and the best
precision(21 ± 1.9 μm)
Full Arch
CS 3600 had the best trueness (44.9 ± 8.9 μm)
Trios3 had the best precision (35.6 ± 3.4 μm), followed very
closely by CS 3600 (35.7 ± 4.3 μm)
16
Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners
in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7

More Related Content

Similar to CS 3600 True Precision Implantology Study

IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...
IJERD Editor
 
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...
IJERD Editor
 
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...
IJERD Editor
 

Similar to CS 3600 True Precision Implantology Study (20)

Micro CT settings for caries detection: how to optimize
Micro CT settings for caries detection: how to optimizeMicro CT settings for caries detection: how to optimize
Micro CT settings for caries detection: how to optimize
 
Diagnosis of Vertical Root Fracture Using Digital Radiography, Helical Comput...
Diagnosis of Vertical Root Fracture Using Digital Radiography, Helical Comput...Diagnosis of Vertical Root Fracture Using Digital Radiography, Helical Comput...
Diagnosis of Vertical Root Fracture Using Digital Radiography, Helical Comput...
 
1111.pdf
1111.pdf1111.pdf
1111.pdf
 
1111_2.pdf
1111_2.pdf1111_2.pdf
1111_2.pdf
 
190330 AI & cloud based medical/dental SW (KIST 김영준)
190330 AI & cloud based medical/dental SW (KIST 김영준)190330 AI & cloud based medical/dental SW (KIST 김영준)
190330 AI & cloud based medical/dental SW (KIST 김영준)
 
Jc14-12/2016-subharti dental collegeand hospital,Meerut
Jc14-12/2016-subharti dental collegeand hospital,MeerutJc14-12/2016-subharti dental collegeand hospital,Meerut
Jc14-12/2016-subharti dental collegeand hospital,Meerut
 
Cone beam
Cone beamCone beam
Cone beam
 
Korte-PachymetryACT13
Korte-PachymetryACT13Korte-PachymetryACT13
Korte-PachymetryACT13
 
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...
 
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...
 
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...
IJERD (www.ijerd.com) International Journal of Engineering Research and Devel...
 
Implications of cbct in orthodontics
Implications of cbct in orthodonticsImplications of cbct in orthodontics
Implications of cbct in orthodontics
 
3D tumor spheroid models for in vitro therapeutic screening: a systematic app...
3D tumor spheroid models for in vitro therapeutic screening: a systematic app...3D tumor spheroid models for in vitro therapeutic screening: a systematic app...
3D tumor spheroid models for in vitro therapeutic screening: a systematic app...
 
BIOSTATS
BIOSTATSBIOSTATS
BIOSTATS
 
Jung2019
Jung2019Jung2019
Jung2019
 
Clinical accuracy outcome of open and closed trayimpressions
Clinical accuracy outcome of open and closed trayimpressionsClinical accuracy outcome of open and closed trayimpressions
Clinical accuracy outcome of open and closed trayimpressions
 
Journal club on Connective tissue graft associated or not with low laser ther...
Journal club on Connective tissue graft associated or not with low laser ther...Journal club on Connective tissue graft associated or not with low laser ther...
Journal club on Connective tissue graft associated or not with low laser ther...
 
Muller-JonesHPF_4185022
Muller-JonesHPF_4185022Muller-JonesHPF_4185022
Muller-JonesHPF_4185022
 
20221026 implant LL - Association of prosthetic features and peri-implantitis...
20221026 implant LL - Association of prosthetic features and peri-implantitis...20221026 implant LL - Association of prosthetic features and peri-implantitis...
20221026 implant LL - Association of prosthetic features and peri-implantitis...
 
International Journal of Engineering Research and Development
International Journal of Engineering Research and DevelopmentInternational Journal of Engineering Research and Development
International Journal of Engineering Research and Development
 

Recently uploaded

👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
rajnisinghkjn
 
Pune Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Pune No💰Adva...
Pune Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Pune No💰Adva...Pune Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Pune No💰Adva...
Pune Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Pune No💰Adva...
Sheetaleventcompany
 
Jual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan Cytotec
Jual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan CytotecJual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan Cytotec
Jual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan Cytotec
jualobat34
 
Call Girl In Indore 📞9235973566📞 Just📲 Call Inaaya Indore Call Girls Service ...
Call Girl In Indore 📞9235973566📞 Just📲 Call Inaaya Indore Call Girls Service ...Call Girl In Indore 📞9235973566📞 Just📲 Call Inaaya Indore Call Girls Service ...
Call Girl In Indore 📞9235973566📞 Just📲 Call Inaaya Indore Call Girls Service ...
Sheetaleventcompany
 
Gorgeous Call Girls Dehradun {8854095900} ❤️VVIP ROCKY Call Girls in Dehradun...
Gorgeous Call Girls Dehradun {8854095900} ❤️VVIP ROCKY Call Girls in Dehradun...Gorgeous Call Girls Dehradun {8854095900} ❤️VVIP ROCKY Call Girls in Dehradun...
Gorgeous Call Girls Dehradun {8854095900} ❤️VVIP ROCKY Call Girls in Dehradun...
Sheetaleventcompany
 
Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...
Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...
Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...
Sheetaleventcompany
 
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...
Sheetaleventcompany
 

Recently uploaded (20)

👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
 
Pune Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Pune No💰Adva...
Pune Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Pune No💰Adva...Pune Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Pune No💰Adva...
Pune Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Pune No💰Adva...
 
Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...
Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...
Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...
 
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptxANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
 
Call Girls Shahdol Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Shahdol Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Shahdol Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Shahdol Just Call 8250077686 Top Class Call Girl Service Available
 
Jual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan Cytotec
Jual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan CytotecJual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan Cytotec
Jual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan Cytotec
 
Call Girl In Indore 📞9235973566📞 Just📲 Call Inaaya Indore Call Girls Service ...
Call Girl In Indore 📞9235973566📞 Just📲 Call Inaaya Indore Call Girls Service ...Call Girl In Indore 📞9235973566📞 Just📲 Call Inaaya Indore Call Girls Service ...
Call Girl In Indore 📞9235973566📞 Just📲 Call Inaaya Indore Call Girls Service ...
 
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
 
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
 
Kolkata Call Girls Naktala 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Kolkata Call Girls Naktala  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Girl Se...Kolkata Call Girls Naktala  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Girl Se...
Kolkata Call Girls Naktala 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
 
Circulatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanismsCirculatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanisms
 
Intramuscular & Intravenous Injection.pptx
Intramuscular & Intravenous Injection.pptxIntramuscular & Intravenous Injection.pptx
Intramuscular & Intravenous Injection.pptx
 
🚺LEELA JOSHI WhatsApp Number +91-9930245274 ✔ Unsatisfied Bhabhi Call Girls T...
🚺LEELA JOSHI WhatsApp Number +91-9930245274 ✔ Unsatisfied Bhabhi Call Girls T...🚺LEELA JOSHI WhatsApp Number +91-9930245274 ✔ Unsatisfied Bhabhi Call Girls T...
🚺LEELA JOSHI WhatsApp Number +91-9930245274 ✔ Unsatisfied Bhabhi Call Girls T...
 
Gorgeous Call Girls Dehradun {8854095900} ❤️VVIP ROCKY Call Girls in Dehradun...
Gorgeous Call Girls Dehradun {8854095900} ❤️VVIP ROCKY Call Girls in Dehradun...Gorgeous Call Girls Dehradun {8854095900} ❤️VVIP ROCKY Call Girls in Dehradun...
Gorgeous Call Girls Dehradun {8854095900} ❤️VVIP ROCKY Call Girls in Dehradun...
 
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room DeliveryCall 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
 
Kolkata Call Girls Shobhabazar 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Gir...
Kolkata Call Girls Shobhabazar  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Gir...Kolkata Call Girls Shobhabazar  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Gir...
Kolkata Call Girls Shobhabazar 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Gir...
 
Gastric Cancer: Сlinical Implementation of Artificial Intelligence, Synergeti...
Gastric Cancer: Сlinical Implementation of Artificial Intelligence, Synergeti...Gastric Cancer: Сlinical Implementation of Artificial Intelligence, Synergeti...
Gastric Cancer: Сlinical Implementation of Artificial Intelligence, Synergeti...
 
❤️Call Girl Service In Chandigarh☎️9814379184☎️ Call Girl in Chandigarh☎️ Cha...
❤️Call Girl Service In Chandigarh☎️9814379184☎️ Call Girl in Chandigarh☎️ Cha...❤️Call Girl Service In Chandigarh☎️9814379184☎️ Call Girl in Chandigarh☎️ Cha...
❤️Call Girl Service In Chandigarh☎️9814379184☎️ Call Girl in Chandigarh☎️ Cha...
 
Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...
Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...
Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...
 
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...
 

CS 3600 True Precision Implantology Study

  • 1. CS 3600 Summary from “Trueness and precision of 5 intraoral scanners in the impressions of single and multiple implants: a comparative in vitro study”
  • 2. Overview Article Name “Trueness and precision of 5 intraoral scanners in the impressions of single and multiple implants: a comparative in vitro study” Authors • Francesco Guido Mangano, Uli Hauschild, Giovanni Veronesi, Mario Imburgia, Carlo Mangano, and Oleg Admakin Date Published 6 June 2019 Background • Very few peer reviewed studies have evaluated the clinical efficacy of intraoral scanners in implantology. • Even fewer studies compare the accuracy of the various intraoral scanners. • This study compares the trueness and precision of five intraoral scanners, and determines which have both accurate and consistent results. 2
  • 3. What are Trueness and Precision? In this study: Trueness is quantified by distances between generated model and reference model (smaller value means better trueness). Precision is quantified by distances between repeatedly generated models (smaller value means better precision). Accuracy is a combination of trueness and precision. An intraoral scanner must perform well in both trueness and precision tests to be considered “highly accurate.” 3
  • 4. Study Material and Methods 10 scans were performed by a single digital dentistry expert per indication per intraoral scanner device with a zig-zag technique in the same environmental conditions. 2 reference maxillary plaster models were used as a reference: • A - Partially edentulous maxillary arch with: • Implant analog on tooth #23 (implant single support crown simulation) • Implant analog on teeth #14 & 16 (partial implant supported case simulation) • B - Totally edentulous maxillary arch: • Implant analog on teeth #11, 14, 16, 21, 24 & 26 (fixed full arch implant supported denture simulation) Scanbodies used : Megagen® PEEK scanbodies to enhance scans (no light reflection). Reference : 3 scans made per indication with Freedom UHD ® desktop scanner and superimposed in Geomagic Studio 2012® to validate the reference models. 4
  • 5. How do these intraoral scanners compare in accuracy? Implantology Intraoral Scanner Comparative Study 5 TRIOS 3 EMERALD DWIO CS 3600 OMNICAM Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
  • 6. Implantology Intraoral Scanner Comparative Study Scanner Producer Technology of Acquisition Powder Color System CS 3600 Carestream Dental, Atlanta, Georgia, USA Structured light-Active Speed 3D Video™ No Yes Proprietary files (CSZ), but also open formats (PLY, STL) immediately available Trios 3 3-Shape, Copenhagen, Denmark Structured light –Confocal microscopy and Ultrafast Optical Scanning™ No Yes Proprietary files (DCM) available, but possibility to export STL files via the new Trios on Dental Desktop® Omnicam Dentsply-Sirona, York, Pennsylvania, USA Structured light -Optical triangulation and confocal microscopy No Yes Proprietary files (CS3, SDT, CDT, IDT) are available, possible to export STL files via the Cerec Connect® DWIO Dentalwings, Montreal, Quebec, Canada Blue laser-Multiscan Imaging™ technology No No Proprietary files (XORDER), but also open formats (STL) immediately available Emerald Planmeca, Helsinki, Finland Red, green and blue lasers-Projected Pattern Triangulation™ No Yes Open formats (PLY, STL) immediately available 6 Summary of 5 Intraoral Scanners Used in the Study Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
  • 7. Single Crown, Best Result in Trueness (μM) 7 Deviation: 22 μm (20) Triangles: 179,496 Deviation: 14 μm (13) Triangles: 34,980 Deviation: 31 μm (26) Triangles: 81,712 Deviation: 23 μm (21) Triangles: 68,281 Deviation: 21 μm (28) Triangles: 123,417 TRIOS 3 CS 3600 EMERALD DWIO OMNICAM Best Result [Carestream Dental Note: Triangle number is not a measure of acquisition resolution or mesh quality. In fact, in having the best trueness with the smallest number of triangles, CS 3600 demonstrates the most efficient/optimal mesh generation (by placing more triangles in areas with more details and vice versa), with minimized mesh file size while maintaining accuracy.] Single crown: best result in trueness (standard deviation) in μm, for the 5 examined scanners, and the number of triangles composing each mesh. Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
  • 8. Partial Prosthesis, Best Result in Trueness (μM) 8 Deviation: 28 μm (24) Triangles: 207,159 Deviation: 21 μm (25) Triangles: 55,968 Deviation: 42 μm (32) Triangles: 113,162 Deviation: 39 μm (39) Triangles: 135,028 Deviation: 29 μm (32) Triangles: 192,967 8 TRIOS 3 CS 3600 EMERALD DWIO OMNICAM Best Result Partial prosthesis: best result in trueness (standard deviation) in μm, for the 5 examined scanners, and the number of triangles composing each mesh. Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7 [Carestream Dental Note: Triangle number is not a measure of acquisition resolution or mesh quality. In fact, in having the best trueness with the smallest number of triangles, CS 3600 demonstrates the most efficient/optimal mesh generation (by placing more triangles in areas with more details and vice versa), with minimized mesh file size while maintaining accuracy.]
  • 9. Full Arch, Best Result in Trueness (μM) 9 Deviation: 37 μm (44) Triangles: 326,696 Deviation: 32 μm (49) Triangles: 83,735 Deviation: 57 μm (58) Triangles: 178,507 Deviation: 73 μm (64) Triangles: 227, 491 Deviation: 47 μm (55) Triangles: 322,929 9 TRIOS 3 CS 3600 EMERALD DWIO OMNICAM Best Result Full arch: best result in trueness (standard deviation) in μm, for the 5 examined scanners, and the number of triangles composing each mesh. Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7 [Carestream Dental Note: Triangle number is not a measure of acquisition resolution or mesh quality. In fact, in having the best trueness with the smallest number of triangles, CS 3600 demonstrates the most efficient/optimal mesh generation (by placing more triangles in areas with more details and vice versa), with minimized mesh file size while maintaining accuracy.]
  • 10. Mean Trueness/Standard Deviation Scanner Single Crown Mean ± SD Partial Prosthesis Mean ± SD Full Arch Mean ± SD p-value1 Trios 3 22.3 ± 0.5† 28.5 ± 0.5†,‡,• 46.3 ± 4.9†,‡,• <.0001 CS 3600 15.2 ± 0.8‡,#,§ 23.0 ± 1.1^,§,# 44.9 ± 8.9^,§,# <.0001 Emerald 43.1 ± 11.5†,‡,•,^ 49.3 ± 5.5†,^,° 66.3 ± 5.6†,^,° <.0001 DWIO 27.8±3.2#,• 49.8±5.0‡,§,* 92.1±24.1‡,§,°,* <.0001 Omnicam 28.4±4.5§,^ 38.1±8.8•,#,°,* 70.4±11.9•,#,* <.0001 10 The same symbol after SD indicates differences in trueness between scanner pairs (Tukey-adjustment for multiple comparison). Minimum significant difference across scanners: 7.3μm, 6.6μm, 16.8μm for single crown (SC), partial prosthesis (PP) and full arch (FA), respectively. 1p- value testing the interaction between scanner and context (SC vs. PP vs. FA) from non-parametric, Kruskall-Wallis test. A p-value >0.05 indicates no difference in scanner trueness according to the context. Best Result Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
  • 11. Colorimetric Trueness Map Comparison 11 TRIOS 3 CS 3600 EMERALD DWIO OMNICAM +0.100 mm +0.088 mm +0.077 mm +0.065 mm +0.053 mm +0.042 mm +0.030 mm -0.030 mm -0.042 mm -0.053 mm -0.065 mm -0.077 mm -0.088 mm -0.100 mm Best Result Trueness in the single crown (SC), partial prosthesis (PP) and full-arch (FA) with the 5 examined intraoral scanners (IOSs): colorimetric maps. The color maps indicated inward (blue) or outward (red) displacement between overlaid structures, whereas a minimal change was indicated by a green color. For all three models (SC, PP, FA): the color scale ranged from a maximum deviation of +100μm and−100μm, with the best result given by the deviations comprised between +30μm and−30μm (green color). Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
  • 12. Changes in Mean Trueness Comparison Changes in mean trueness (standard deviation), in μm, for the 5 examined scanners, in the different applications (single crown, SC vs. partial prosthesis, PP vs. full-arch, FA). 12 Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7 CS 3600
  • 13. Mean Precision/Standard Deviation Scanner Single Crown Mean ± SD Partial Prosthesis Mean ± SD Full Arch Mean ± SD p-value1 Trios 3 15.2 ± 0.8†,‡,• 21.0 ± 1.9‡,• 35.6 ± 3.4†,‡,• <.0001 CS 3600 11.3 ± 1.1^,§,# 17.0 ± 2.3^,§,# 35.7 ± 4.3^,§,# <.0001 Emerald 32.8 ± 10.7†,^ 29.9 ± 8.9^,° 61.5 ± 18.1†,^,°,* 0.0007 DWIO 27.1 ± 10.7‡,§ 34.8 ± 10.8‡,§ 111.0 ± 24.8‡,§,°,ç <.0001 Omnicam 30.6 ± 3.3•,# 43.2 ± 9.4•,#,° 89.3 ± 14.0•,#,*,ç <.0001 13 The same symbol after SD indicates differences in precision between scanner pairs (Tukey-adjustment for multiple comparison). Minimum significant difference across scanners: 8.8 μm, 9.8 μm, 19.4 μm for single crown (SC), partial prosthesis (PP) and full arch (FA), respectively. 1p-value testing the interaction between scanner and context (SC vs. PP vs. FA) from non-parametric, Kruskall-Wallis test. A p-value > 0.05 indicates no difference in scanner precision according to the context. Best Result Very closely follows best result Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
  • 14. Colorimetric Precision Map Comparison 14 Very Close to Best Results TRIOS 3 CS 3600 EMERALD DWIO OMNICAM Best Result Trueness in the single crown (SC), partial prosthesis (PP) and full-arch (FA) with the 5 examined intraoral scanners (IOSs): colorimetric maps. The color maps indicated inward (blue) or outward (red) displacement between overlaid structures, whereas a minimal change was indicated by a green color. For all three models (SC, PP, FA): the color scale ranged from a maximum deviation of +100μm and−100μm, with the best result given by the deviations comprised between +30μm and−30μm (green color). +0.100 mm +0.088 mm +0.077 mm +0.065 mm +0.053 mm +0.042 mm +0.030 mm -0.030 mm -0.042 mm -0.053 mm -0.065 mm -0.077 mm -0.088 mm -0.100 mm Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7
  • 15. Changes in Mean Precision Comparison Changes in mean precision (standard deviation), in μm, for the 5 examined scanners, in the different applications (single crown vs. partial prosthesis vs. full-arch) 15 CS 3600 Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7 CS 3600
  • 16. Results Summary Single Crown: CS 3600 had the best trueness (15.2 ± 0.8 μm) and the best precision (11.3 ± 1.1 μm) Partial Prosthesis CS 3600 had the best trueness (23 ± 1.1 μm) and the best precision(21 ± 1.9 μm) Full Arch CS 3600 had the best trueness (44.9 ± 8.9 μm) Trios3 had the best precision (35.6 ± 3.4 μm), followed very closely by CS 3600 (35.7 ± 4.3 μm) 16 Mangano F, Logozzo S, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. June 6, 2019. Trueness and precision of 5 intraoral scanners in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 19(101): DOI 10.1186/s12903-019-0792-7