ANALIZA CALITATIVA SI CANTITATIVA
ANALIZA CALITATIVA SI CANTITATIVA
ANALIZA CALITATIVA SI CANTITATIVA
ANALIZA CALITATIVA SI CANTITATIVA
ANALIZA CALITATIVA SI CANTITATIVA
1. Metoda curbei de calibrare
Concentratia, mL/mL Area
0,200 1,43
0,400 2,86
0,600 4,29
0,800 5,73
1,000 7,16
Proba 4,10
0,573
y = 7,165x - 0,005
R² = 1
0,00
1,00
2,00
3,00
4,00
5,00
6,00
7,00
8,00
0,000 0,200 0,400 0,600 0,800 1,000 1,200
Area
Concentratia, mL/mL
Area
ANALIZA CALITATIVA SI CANTITATIVA
2. Metoda normării ariilor
cu A = 

n
1
j
j
A
Compus Area suma A i Concentratia (%)
A 14171 70499 20.10
B 21573 30.60
C 34755 49.30
Compus Area Factor raspuns Area corectata suma arii corectate Concentratia (%)
A 14171 1 14171 91136.54 15.55
B 21573 1.28 27613.44 30.30
C 34755 1.42 49352.1 54.15
ANALIZA CALITATIVA SI CANTITATIVA
3. Metoda adaosului
100 %
i
i St
St
A
c c
A
 
Solutie standard - 100 mL
Set 1 Concentratia, g/L Area
A - 250 mg 2,5 38801
B - 250 mg 2,5 14236
C - 250 mg 2,5 26301
D - 250 mg 2,5 39594
1000 mg proba solida
Proba - 100 mL
Set 2 A 68994
B 19050
C 53480
D + St (100 mg) 84922
O proba solida de 1000 mg a fost analizata pentru a afla concentratia de
A, B, C si D. Proba a fost supusa unei proceduri de extractie cu
solvent. In extrasul rezultat s-au adaugat 100 mg compus D ca standard
intern (SI) intr-un volum final de 100 mL.
De asemenea a fost analizata o solutie standard care ii contine pe cei 4
compusi (250 mg), in raport procentual de 1:1:1:1, intr-un volum de
100 mL. Sa se determine compozitia probei folosind datele din tabel.
ANALIZA CALITATIVA SI CANTITATIVA
4. Metoda standardului intern
100 %
i
i St
St
A
c c
A
 
O proba de sol a fost analizata prin GC‐FID pentru a determina
concentratia compusului X. Rezultatele masuratorilor sunt:
tr (minute) area peak
Cromatograma 1 21,1 ppm Toluen SI 10,05 38774
concentratia X 14,77 39115
Care este concentratia lui X in proba?
ANALIZA CALITATIVA SI CANTITATIVA
4. Metoda standardului intern
100 %
i
i SI
SI
A
c c
A
 
Caracterisitici SI :
- să nu fie conţinută de amestecul de analizat,
- să fie complet separată de restul componenţilor, dar cu un timp de retenţie
apropiat de cel al componentului de determinat,
- să nu difere prea mult în concentraţie faţă de componentul de determinat.
• Q1. Care sunt principalele diferențe între cromatografia de lichide de înaltă
performanță și cea de gaze?
• Q2. Ce tip de detector de GC este utilizat cel mai frecvent? Explicați principiul
său de lucru și care sunt limitele sale?
• Q3. Care sunt gazele purtătoare utilizate în mod obișnuit în analiza GC atunci
când se utilizează detectorul FID?
• Q4. Care sunt caracteristicile dorite ale unui detector GC?
• Q5. Ce înțelegeți prin specificitatea unui detector?
• Q6. Care sunt tipurile de coloane capilare cel mai frecvent utilizate?
• Q7. Ce înțelegeți prin eficiența coloanei și cum se exprimă?
• Q8. Ce înțelegeți prin programarea temperaturii în analiza GC?
• Q9. Când este util modul de lucru izoterm?
• Q10. Ce măsuri ati adopta pentru a prelungi durata de viață utilă a unei coloane?
Test 4
Ans1.
• În HPLC faza mobilă este un lichid; în cromatografie în fază gazoasă faza mobilă sau
purtătorul este un gaz.
• Cromatografia de gaze se aplică pentru analiza compușilor volatili; compușii nevolatili
pot fi ușor analizati prin HPLC.
• Cromatografie de gaze nu poate fi utilizata pentru analiza moleculelor cu masă
moleculară mare; HPLC are aplicații pentru separarea și identificarea compușilor cu
greutate moleculară foarte mare.
• HPLC necesită presiuni de lucru mai mari decât GC, deoarece lichidele necesită
presiuni mai mari decât gazele pentru transport prin intermediul sistemului.
• Coloane HPLC sunt scurte și late, în comparație cu coloanele GC.
Ans2.
Detectorul cel mai frecvent utilizat este detectorul cu ionizare in
flacără. Compusul separat care a ajuns la detector este ars cu ajutorul
gazului combustibil și produce ioni și electroni care pot conduce
electricitate prin flacără. Prin aplicarea unei diferențe mari de potențial
la vârful arzătorului și la electrodul colector localizat deasupra flăcării
si se măsoara curentul dintre electrozi. Detectorul este dependent de
debitul masic și răspunsul nu este afectat de schimbările debitului de
gaz purtător. Cu toate acestea, detectorul nu detecteaza gaze
anorganice, cum ar fi CO, O2, NH3, N2, CS2, CO2, etc.
Ans3.
Atunci când se utilizează detectorul FID gazele inerte frecvent utilizate
în analiza sunt azot si heliu. Azotul este mai frecvent utilizat, deoarece
este mai puțin costisitor decât heliu. Puritatea gazului purtător ar trebui
să fie mai mult de 99,995%, de aceea ar trebui folosite capcane on-line
pentru a preveni umezeala reziduală sau alte impurități de la intrarea în
sistem.
Ans4.
Alegerea detectorului pentru analize are la baza criteriile:
• Reproductibilitate la modificări ale concentrației compușilor de eluat.
• Domeniu dinamic liniar larg.
• Sensibilitate ridicată.
• Ar trebui să aibă volum intern mic pentru a da peak-uri înguste și, de
asemenea, sa facilitează spălarea urmelor din proba anterioara.
• Ar trebui să fie, de preferință, non-distructiv.
Ans5.
Detectoarele se împart în trei categorii, în funcție de răspunsul la
compușii eluati:
• neselectivi – dau răspuns pentru toti compusii din faza mobila cu
excepția gazului purtător.
• selectivi - răspund la o anumită clasă de compuși cu proprietăți fizice
sau chimice comune.
• specifici - răspunsul la un singur compus specific.
Ans6.
Coloanele capilare sunt în general tuburi lungi de 10 - 100 m, cu un
diametru interior cuprins între 0,1 - 0,5mm, realizat din material
flexibil, cum ar fi silice topită. Tipuri de coloane capilare:
• Wall coated open tubular (WCOT) - peretele intern este acoperit cu o
peliculă foarte fină de lichid adsorbant
• Surface coated open tubular (SGOT) - peretele interior este căptușit
cu un strat de suport solid pe care este absorbită faza lichidă.
Coloanele sunt flexibile și înfășurate în mai multe bobine; fixate în
interiorul cuptorului.
Ans7.
Prin utilizare continuă, o coloană își pierde treptat puterea de rezoluție originală.
Eficiența coloanei este exprimată pe baza conceptului teoriei talerelor. In timpul
separarii, fiecare component de separat petrece un timp finit în fiecare taler teoretic.
Eficiența coloanei este cu atat mai buna cu cat înălțimea talerului este mai mică, ceea
ce inseamna un număr de talere (N) cat mai mare.
unde N = număr de talere teoretice; tr = timpul de retenție al unui peak
w = lățimea peak-ului la bază;
N poate fi de asemenea exprimat ca:
unde w1/2 este lățimea peak-ului la jumătatea înălțimii
Eficiența coloanei este exprimată în termeni de înălțime echivalentă a talerelor
teoretice (HETP): HETP = L/N, unde L = lungimea coloanei în de mm; o valoarea
mică a HETP semnifică o eficiență mai mare coloanei.
Ans8.
Programarea temperaturii înseamnă schimbarea temperaturii coloanei
cu o viteză predeterminată în timpul ciclului de analiză. Aceasta are
aceeași influență asupra timpului de eluție a componentelor separate ca
programarea gradientului în analiza HPLC. Programul de temperatura
ajută la reducerea timpului de analiză permițând eluarea timpurie a
componentelor mai puțin volatile.
Ans9.
Modul de lucru izoterm este util atunci când este necesară o rezoluție
ridicată pentru separarea compușilor cu interval de fierbere îngust.
Temperatura este setată la mijlocul intervalului de puncte de de fierbere
ale constituenților. Acest lucru permite obtinerea unei rezolutii bune la
separarea unor componenti cu puncte de fierbere mici, dar ar putea
conduce la peak-uri cu largite in cazul componentilor cu puncte de
fierbere mai mari, datorită retenției lor în coloană timp mai indelungat.
Ans10.
• Conditionarea coloanei înainte de prima utilizare sau după depozitare
timp îndelungat.
• Aveți grijă să nu depășească limita superioară de temperatură
specificată de producător
• A se evita injectarea soluțiilor care sunt puternic acide sau bazice în
natură
• Spalarea coloanei prin injectare cu solvenți cum ar fi metanol, clorură
de metilen sau hexan pentru a îndepărta contaminarea coloanei după
utilizare excesivă.

More Related Content

Featured

How Race, Age and Gender Shape Attitudes Towards Mental Health
How Race, Age and Gender Shape Attitudes Towards Mental HealthHow Race, Age and Gender Shape Attitudes Towards Mental Health
How Race, Age and Gender Shape Attitudes Towards Mental Health
ThinkNow
 
Social Media Marketing Trends 2024 // The Global Indie Insights
Social Media Marketing Trends 2024 // The Global Indie InsightsSocial Media Marketing Trends 2024 // The Global Indie Insights
Social Media Marketing Trends 2024 // The Global Indie Insights
Kurio // The Social Media Age(ncy)
 

Featured (20)

Everything You Need To Know About ChatGPT
Everything You Need To Know About ChatGPTEverything You Need To Know About ChatGPT
Everything You Need To Know About ChatGPT
 
Product Design Trends in 2024 | Teenage Engineerings
Product Design Trends in 2024 | Teenage EngineeringsProduct Design Trends in 2024 | Teenage Engineerings
Product Design Trends in 2024 | Teenage Engineerings
 
How Race, Age and Gender Shape Attitudes Towards Mental Health
How Race, Age and Gender Shape Attitudes Towards Mental HealthHow Race, Age and Gender Shape Attitudes Towards Mental Health
How Race, Age and Gender Shape Attitudes Towards Mental Health
 
AI Trends in Creative Operations 2024 by Artwork Flow.pdf
AI Trends in Creative Operations 2024 by Artwork Flow.pdfAI Trends in Creative Operations 2024 by Artwork Flow.pdf
AI Trends in Creative Operations 2024 by Artwork Flow.pdf
 
Skeleton Culture Code
Skeleton Culture CodeSkeleton Culture Code
Skeleton Culture Code
 
PEPSICO Presentation to CAGNY Conference Feb 2024
PEPSICO Presentation to CAGNY Conference Feb 2024PEPSICO Presentation to CAGNY Conference Feb 2024
PEPSICO Presentation to CAGNY Conference Feb 2024
 
Content Methodology: A Best Practices Report (Webinar)
Content Methodology: A Best Practices Report (Webinar)Content Methodology: A Best Practices Report (Webinar)
Content Methodology: A Best Practices Report (Webinar)
 
How to Prepare For a Successful Job Search for 2024
How to Prepare For a Successful Job Search for 2024How to Prepare For a Successful Job Search for 2024
How to Prepare For a Successful Job Search for 2024
 
Social Media Marketing Trends 2024 // The Global Indie Insights
Social Media Marketing Trends 2024 // The Global Indie InsightsSocial Media Marketing Trends 2024 // The Global Indie Insights
Social Media Marketing Trends 2024 // The Global Indie Insights
 
Trends In Paid Search: Navigating The Digital Landscape In 2024
Trends In Paid Search: Navigating The Digital Landscape In 2024Trends In Paid Search: Navigating The Digital Landscape In 2024
Trends In Paid Search: Navigating The Digital Landscape In 2024
 
5 Public speaking tips from TED - Visualized summary
5 Public speaking tips from TED - Visualized summary5 Public speaking tips from TED - Visualized summary
5 Public speaking tips from TED - Visualized summary
 
ChatGPT and the Future of Work - Clark Boyd
ChatGPT and the Future of Work - Clark Boyd ChatGPT and the Future of Work - Clark Boyd
ChatGPT and the Future of Work - Clark Boyd
 
Getting into the tech field. what next
Getting into the tech field. what next Getting into the tech field. what next
Getting into the tech field. what next
 
Google's Just Not That Into You: Understanding Core Updates & Search Intent
Google's Just Not That Into You: Understanding Core Updates & Search IntentGoogle's Just Not That Into You: Understanding Core Updates & Search Intent
Google's Just Not That Into You: Understanding Core Updates & Search Intent
 
How to have difficult conversations
How to have difficult conversations How to have difficult conversations
How to have difficult conversations
 
Introduction to Data Science
Introduction to Data ScienceIntroduction to Data Science
Introduction to Data Science
 
Time Management & Productivity - Best Practices
Time Management & Productivity -  Best PracticesTime Management & Productivity -  Best Practices
Time Management & Productivity - Best Practices
 
The six step guide to practical project management
The six step guide to practical project managementThe six step guide to practical project management
The six step guide to practical project management
 
Beginners Guide to TikTok for Search - Rachel Pearson - We are Tilt __ Bright...
Beginners Guide to TikTok for Search - Rachel Pearson - We are Tilt __ Bright...Beginners Guide to TikTok for Search - Rachel Pearson - We are Tilt __ Bright...
Beginners Guide to TikTok for Search - Rachel Pearson - We are Tilt __ Bright...
 
Unlocking the Power of ChatGPT and AI in Testing - A Real-World Look, present...
Unlocking the Power of ChatGPT and AI in Testing - A Real-World Look, present...Unlocking the Power of ChatGPT and AI in Testing - A Real-World Look, present...
Unlocking the Power of ChatGPT and AI in Testing - A Real-World Look, present...
 

Cap 4.ppt

  • 1. ANALIZA CALITATIVA SI CANTITATIVA
  • 2. ANALIZA CALITATIVA SI CANTITATIVA
  • 3. ANALIZA CALITATIVA SI CANTITATIVA
  • 4. ANALIZA CALITATIVA SI CANTITATIVA
  • 5. ANALIZA CALITATIVA SI CANTITATIVA 1. Metoda curbei de calibrare Concentratia, mL/mL Area 0,200 1,43 0,400 2,86 0,600 4,29 0,800 5,73 1,000 7,16 Proba 4,10 0,573 y = 7,165x - 0,005 R² = 1 0,00 1,00 2,00 3,00 4,00 5,00 6,00 7,00 8,00 0,000 0,200 0,400 0,600 0,800 1,000 1,200 Area Concentratia, mL/mL Area
  • 6. ANALIZA CALITATIVA SI CANTITATIVA 2. Metoda normării ariilor cu A =   n 1 j j A Compus Area suma A i Concentratia (%) A 14171 70499 20.10 B 21573 30.60 C 34755 49.30 Compus Area Factor raspuns Area corectata suma arii corectate Concentratia (%) A 14171 1 14171 91136.54 15.55 B 21573 1.28 27613.44 30.30 C 34755 1.42 49352.1 54.15
  • 7. ANALIZA CALITATIVA SI CANTITATIVA 3. Metoda adaosului 100 % i i St St A c c A   Solutie standard - 100 mL Set 1 Concentratia, g/L Area A - 250 mg 2,5 38801 B - 250 mg 2,5 14236 C - 250 mg 2,5 26301 D - 250 mg 2,5 39594 1000 mg proba solida Proba - 100 mL Set 2 A 68994 B 19050 C 53480 D + St (100 mg) 84922 O proba solida de 1000 mg a fost analizata pentru a afla concentratia de A, B, C si D. Proba a fost supusa unei proceduri de extractie cu solvent. In extrasul rezultat s-au adaugat 100 mg compus D ca standard intern (SI) intr-un volum final de 100 mL. De asemenea a fost analizata o solutie standard care ii contine pe cei 4 compusi (250 mg), in raport procentual de 1:1:1:1, intr-un volum de 100 mL. Sa se determine compozitia probei folosind datele din tabel.
  • 8. ANALIZA CALITATIVA SI CANTITATIVA 4. Metoda standardului intern 100 % i i St St A c c A   O proba de sol a fost analizata prin GC‐FID pentru a determina concentratia compusului X. Rezultatele masuratorilor sunt: tr (minute) area peak Cromatograma 1 21,1 ppm Toluen SI 10,05 38774 concentratia X 14,77 39115 Care este concentratia lui X in proba?
  • 9. ANALIZA CALITATIVA SI CANTITATIVA 4. Metoda standardului intern 100 % i i SI SI A c c A   Caracterisitici SI : - să nu fie conţinută de amestecul de analizat, - să fie complet separată de restul componenţilor, dar cu un timp de retenţie apropiat de cel al componentului de determinat, - să nu difere prea mult în concentraţie faţă de componentul de determinat.
  • 10. • Q1. Care sunt principalele diferențe între cromatografia de lichide de înaltă performanță și cea de gaze? • Q2. Ce tip de detector de GC este utilizat cel mai frecvent? Explicați principiul său de lucru și care sunt limitele sale? • Q3. Care sunt gazele purtătoare utilizate în mod obișnuit în analiza GC atunci când se utilizează detectorul FID? • Q4. Care sunt caracteristicile dorite ale unui detector GC? • Q5. Ce înțelegeți prin specificitatea unui detector? • Q6. Care sunt tipurile de coloane capilare cel mai frecvent utilizate? • Q7. Ce înțelegeți prin eficiența coloanei și cum se exprimă? • Q8. Ce înțelegeți prin programarea temperaturii în analiza GC? • Q9. Când este util modul de lucru izoterm? • Q10. Ce măsuri ati adopta pentru a prelungi durata de viață utilă a unei coloane? Test 4
  • 11. Ans1. • În HPLC faza mobilă este un lichid; în cromatografie în fază gazoasă faza mobilă sau purtătorul este un gaz. • Cromatografia de gaze se aplică pentru analiza compușilor volatili; compușii nevolatili pot fi ușor analizati prin HPLC. • Cromatografie de gaze nu poate fi utilizata pentru analiza moleculelor cu masă moleculară mare; HPLC are aplicații pentru separarea și identificarea compușilor cu greutate moleculară foarte mare. • HPLC necesită presiuni de lucru mai mari decât GC, deoarece lichidele necesită presiuni mai mari decât gazele pentru transport prin intermediul sistemului. • Coloane HPLC sunt scurte și late, în comparație cu coloanele GC.
  • 12. Ans2. Detectorul cel mai frecvent utilizat este detectorul cu ionizare in flacără. Compusul separat care a ajuns la detector este ars cu ajutorul gazului combustibil și produce ioni și electroni care pot conduce electricitate prin flacără. Prin aplicarea unei diferențe mari de potențial la vârful arzătorului și la electrodul colector localizat deasupra flăcării si se măsoara curentul dintre electrozi. Detectorul este dependent de debitul masic și răspunsul nu este afectat de schimbările debitului de gaz purtător. Cu toate acestea, detectorul nu detecteaza gaze anorganice, cum ar fi CO, O2, NH3, N2, CS2, CO2, etc.
  • 13. Ans3. Atunci când se utilizează detectorul FID gazele inerte frecvent utilizate în analiza sunt azot si heliu. Azotul este mai frecvent utilizat, deoarece este mai puțin costisitor decât heliu. Puritatea gazului purtător ar trebui să fie mai mult de 99,995%, de aceea ar trebui folosite capcane on-line pentru a preveni umezeala reziduală sau alte impurități de la intrarea în sistem.
  • 14. Ans4. Alegerea detectorului pentru analize are la baza criteriile: • Reproductibilitate la modificări ale concentrației compușilor de eluat. • Domeniu dinamic liniar larg. • Sensibilitate ridicată. • Ar trebui să aibă volum intern mic pentru a da peak-uri înguste și, de asemenea, sa facilitează spălarea urmelor din proba anterioara. • Ar trebui să fie, de preferință, non-distructiv.
  • 15. Ans5. Detectoarele se împart în trei categorii, în funcție de răspunsul la compușii eluati: • neselectivi – dau răspuns pentru toti compusii din faza mobila cu excepția gazului purtător. • selectivi - răspund la o anumită clasă de compuși cu proprietăți fizice sau chimice comune. • specifici - răspunsul la un singur compus specific.
  • 16. Ans6. Coloanele capilare sunt în general tuburi lungi de 10 - 100 m, cu un diametru interior cuprins între 0,1 - 0,5mm, realizat din material flexibil, cum ar fi silice topită. Tipuri de coloane capilare: • Wall coated open tubular (WCOT) - peretele intern este acoperit cu o peliculă foarte fină de lichid adsorbant • Surface coated open tubular (SGOT) - peretele interior este căptușit cu un strat de suport solid pe care este absorbită faza lichidă. Coloanele sunt flexibile și înfășurate în mai multe bobine; fixate în interiorul cuptorului.
  • 17. Ans7. Prin utilizare continuă, o coloană își pierde treptat puterea de rezoluție originală. Eficiența coloanei este exprimată pe baza conceptului teoriei talerelor. In timpul separarii, fiecare component de separat petrece un timp finit în fiecare taler teoretic. Eficiența coloanei este cu atat mai buna cu cat înălțimea talerului este mai mică, ceea ce inseamna un număr de talere (N) cat mai mare. unde N = număr de talere teoretice; tr = timpul de retenție al unui peak w = lățimea peak-ului la bază; N poate fi de asemenea exprimat ca: unde w1/2 este lățimea peak-ului la jumătatea înălțimii Eficiența coloanei este exprimată în termeni de înălțime echivalentă a talerelor teoretice (HETP): HETP = L/N, unde L = lungimea coloanei în de mm; o valoarea mică a HETP semnifică o eficiență mai mare coloanei.
  • 18. Ans8. Programarea temperaturii înseamnă schimbarea temperaturii coloanei cu o viteză predeterminată în timpul ciclului de analiză. Aceasta are aceeași influență asupra timpului de eluție a componentelor separate ca programarea gradientului în analiza HPLC. Programul de temperatura ajută la reducerea timpului de analiză permițând eluarea timpurie a componentelor mai puțin volatile.
  • 19. Ans9. Modul de lucru izoterm este util atunci când este necesară o rezoluție ridicată pentru separarea compușilor cu interval de fierbere îngust. Temperatura este setată la mijlocul intervalului de puncte de de fierbere ale constituenților. Acest lucru permite obtinerea unei rezolutii bune la separarea unor componenti cu puncte de fierbere mici, dar ar putea conduce la peak-uri cu largite in cazul componentilor cu puncte de fierbere mai mari, datorită retenției lor în coloană timp mai indelungat.
  • 20. Ans10. • Conditionarea coloanei înainte de prima utilizare sau după depozitare timp îndelungat. • Aveți grijă să nu depășească limita superioară de temperatură specificată de producător • A se evita injectarea soluțiilor care sunt puternic acide sau bazice în natură • Spalarea coloanei prin injectare cu solvenți cum ar fi metanol, clorură de metilen sau hexan pentru a îndepărta contaminarea coloanei după utilizare excesivă.