Technical Note

QIAGEN Rotor Gene Q: Software V2.0
Instrument Setup Instructions for RT2 Profiler™ PCR
Arrays
Preparation
Before the Experiment (Presetting the Machine will save time for your run):
•
Please make note of the Rotor-Gene Q Series software version on your instrument.
•
For more information, please click the help button or refer to the Rotor-Gene Q Manual available at
http://www.qiagen.com.
•

Set up a PCR protocol template file on the Rotor-Gene Q Series software as follows:
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.

12.
13.
14.
15.
16.
17.
18.
19.

Open the Rotor-Gene Q Series Software 2.0 on the desktop of the computer that is connected to the
Rotor-Gene Q.
Select File ► New. The New Run dialog box will appear (Note: the New Run dialog box may open
automatically).
Under the Advanced tab, select Two Step.
Click New.
Under the Welcome to the Advanced Run Wizard! tab, select Rotor-Disc 100.
Ensure locking ring has been attached to the Rotor-Disc 100 and check Locking Ring Attached
box.
Click Next.
Under the Miscellaneous Options tab, set Reaction Volume (µl) to 20.
Click Next.
Click the Edit Profile tab.
Under the Edit Profile tab (Figure 1), adjust parameters to reflect the following:
Hold
•
Hold Temperature: 95°
C
•
Hold Time: 10 mins 0 secs
Cycling
•
This cycle repeats 40 time(s)
•
95° , 15 seconds, Not Acquiring
C
•
60° , 30 seconds, Acquiring to Cycling A on Green
C
Click Insert After ► New Melt. Ensure Optimize gain before melt on all tubes is checked.
Click Ok.
Click Gain Optimisation.
Under the Auto-Gain Optimisation Setup tab, click Optimise Acquiring.
Click Ok.
Ensure Perform Optimisation Before 1st Acquistion is checked.
Click Close.
Click Next.
Click Save Template and enter RT2_Rotor_Gene_Q as the template name.
Click Save.

Page 1 of 5
Technical Note

Figure 1
Edit Profile Tab

Performing Real-Time PCR Detection
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.

If the Rotor-Gene Q is off, switch on the instrument, and ensure the standby light is lit.
Open the Rotor-Gene Q Series Software 2.0.
Under the New Run dialog box, click on the Quick Start tab, and select Open a Template In Another
Folder.
Click New.
Locate RT2_Rotor_Gene_Q Template file.
Click Open.
Under the 1. Rotor Selection tab, select Rotor-Disc 100.
Ensure locking ring has been attached to the Rotor-Disc 100 and check Locking Ring Attached
box.
Click Next.
Under 2. Confirm Profile tab, verify desired profile.
Click Start Run.
Enter name for run and click Save.
Rotor-Gene Q run will now commence.

After the PCR Run
1.
2.
3.
4.
5.
6.

Once the PCR run is complete, observe the Sample Bank (right side of screen, Figure 2).
Click Bank On.
Click All On.
Select Analysis in program bar.
Under Quantitation tab, select Cycling A. Green.
Click Show.
Page 2 of 5
Technical Note

7.

Calculate the threshold cycle (Ct) for each well using the instrument’s software.
a. To define the Baseline (Figure 1):
i.
Observe amplification plots in Linear View.
ii.
Select Dynamic Tube (default analysis setting) to ensure the average background of
each well is determined just before amplification commences.
iii.
(Optional) Select Ignore First. Fluorescent signal from the initial cycles may not be
representative of the remainder of the run. Thus, better results may be achieved if the
initial cycles are ignored. Up to 5 cycles can be ignored.
iv.
(Optional) Select Noise Slope Correction. Selection of this option can improve data
whose baseline (initial cycles) is noticeably sloped. Noise Slope Correction improves
the data when raw data backgrounds are observed to slope upward or downward
before the takeoff point (Ct).
*IMPORTANT: Ensure that all selections remain consistent across all PCR Array runs in the same analysis.
Figure 2
Setting the Baseline

b.

Manually define the Threshold Value (Figure 3):
i.
Observe the Log View of the amplification plots.
ii.
In the Ct Calculation box (under Sample Bank) click the button beside the Threshold
box.
iii.
Move mouse to Amplification plot and click mouse to place threshold above the
background signal but within the lower one-third to lower one-half of the linear phase of
the amplification plot.
iv.
Right click on Quant. Results window.
v.
Click Export to Excel.
Page 3 of 5
Technical Note

c.

vi.
This file format can be opened in the Microsoft Excel Program.
Figure 4 illustrates the layout of the Rotor-Gene Q analysis window.

Figure 3
Setting the Threshold

Page 4 of 5
Technical Note

Figure 4
Rotor-Gene Q Analysis Window

Page 5 of 5

More Related Content

Similar to Instrument setuprotorgeneq

Mi script mirna_rgq_technical_note_setup-e
Mi script mirna_rgq_technical_note_setup-eMi script mirna_rgq_technical_note_setup-e
Mi script mirna_rgq_technical_note_setup-e
Elsa von Licy
 
Roche lc4801.2.9.11e
Roche lc4801.2.9.11eRoche lc4801.2.9.11e
Roche lc4801.2.9.11e
Elsa von Licy
 
Mi script mirna_icycler_technical_note_setup-e
Mi script mirna_icycler_technical_note_setup-eMi script mirna_icycler_technical_note_setup-e
Mi script mirna_icycler_technical_note_setup-e
Elsa von Licy
 
Cn lc480 setup_20120808e
Cn lc480 setup_20120808eCn lc480 setup_20120808e
Cn lc480 setup_20120808e
Elsa von Licy
 
Cn bio radinstrumentsetup_20120808e
Cn bio radinstrumentsetup_20120808eCn bio radinstrumentsetup_20120808e
Cn bio radinstrumentsetup_20120808e
Elsa von Licy
 
Mi script mirna_abi7900_technical_note_setup-e
Mi script mirna_abi7900_technical_note_setup-eMi script mirna_abi7900_technical_note_setup-e
Mi script mirna_abi7900_technical_note_setup-e
Elsa von Licy
 
Cn abi vii_a7setup_20120808e
Cn abi vii_a7setup_20120808eCn abi vii_a7setup_20120808e
Cn abi vii_a7setup_20120808e
Elsa von Licy
 
Mi script mirna_lc480_technical_note_setup-e
Mi script mirna_lc480_technical_note_setup-eMi script mirna_lc480_technical_note_setup-e
Mi script mirna_lc480_technical_note_setup-e
Elsa von Licy
 
Mi script mirna_viia7_technical_note_setup-e
Mi script mirna_viia7_technical_note_setup-eMi script mirna_viia7_technical_note_setup-e
Mi script mirna_viia7_technical_note_setup-e
Elsa von Licy
 
Mi script mirna_abi7500_technical_note_setup-e
Mi script mirna_abi7500_technical_note_setup-eMi script mirna_abi7500_technical_note_setup-e
Mi script mirna_abi7500_technical_note_setup-e
Elsa von Licy
 
Cn mx3000 p_20120808e
Cn mx3000 p_20120808eCn mx3000 p_20120808e
Cn mx3000 p_20120808e
Elsa von Licy
 
Cn abi7500 setup_20120808e
Cn abi7500 setup_20120808eCn abi7500 setup_20120808e
Cn abi7500 setup_20120808e
Elsa von Licy
 

Similar to Instrument setuprotorgeneq (20)

Abi7900 ht2.3e
Abi7900 ht2.3eAbi7900 ht2.3e
Abi7900 ht2.3e
 
Mi script mirna_rgq_technical_note_setup-e
Mi script mirna_rgq_technical_note_setup-eMi script mirna_rgq_technical_note_setup-e
Mi script mirna_rgq_technical_note_setup-e
 
Roche lc4801.2.9.11e
Roche lc4801.2.9.11eRoche lc4801.2.9.11e
Roche lc4801.2.9.11e
 
Som lc480 setup_e
Som lc480 setup_eSom lc480 setup_e
Som lc480 setup_e
 
Mi script mirna_icycler_technical_note_setup-e
Mi script mirna_icycler_technical_note_setup-eMi script mirna_icycler_technical_note_setup-e
Mi script mirna_icycler_technical_note_setup-e
 
Som i cycler_setupe
Som i cycler_setupeSom i cycler_setupe
Som i cycler_setupe
 
Abi75001.3.0 e
Abi75001.3.0 eAbi75001.3.0 e
Abi75001.3.0 e
 
Cn lc480 setup_20120808e
Cn lc480 setup_20120808eCn lc480 setup_20120808e
Cn lc480 setup_20120808e
 
Cn bio radinstrumentsetup_20120808e
Cn bio radinstrumentsetup_20120808eCn bio radinstrumentsetup_20120808e
Cn bio radinstrumentsetup_20120808e
 
Mi script mirna_abi7900_technical_note_setup-e
Mi script mirna_abi7900_technical_note_setup-eMi script mirna_abi7900_technical_note_setup-e
Mi script mirna_abi7900_technical_note_setup-e
 
Cn abi vii_a7setup_20120808e
Cn abi vii_a7setup_20120808eCn abi vii_a7setup_20120808e
Cn abi vii_a7setup_20120808e
 
Mi script mirna_lc480_technical_note_setup-e
Mi script mirna_lc480_technical_note_setup-eMi script mirna_lc480_technical_note_setup-e
Mi script mirna_lc480_technical_note_setup-e
 
Stratagene mx3000e
Stratagene mx3000eStratagene mx3000e
Stratagene mx3000e
 
Drv10 lab3 pf755_pcam
Drv10 lab3 pf755_pcamDrv10 lab3 pf755_pcam
Drv10 lab3 pf755_pcam
 
Mi script mirna_viia7_technical_note_setup-e
Mi script mirna_viia7_technical_note_setup-eMi script mirna_viia7_technical_note_setup-e
Mi script mirna_viia7_technical_note_setup-e
 
Mi script mirna_abi7500_technical_note_setup-e
Mi script mirna_abi7500_technical_note_setup-eMi script mirna_abi7500_technical_note_setup-e
Mi script mirna_abi7500_technical_note_setup-e
 
Cn mx3000 p_20120808e
Cn mx3000 p_20120808eCn mx3000 p_20120808e
Cn mx3000 p_20120808e
 
Som abi7500 setup_e
Som abi7500 setup_eSom abi7500 setup_e
Som abi7500 setup_e
 
Cn abi7500 setup_20120808e
Cn abi7500 setup_20120808eCn abi7500 setup_20120808e
Cn abi7500 setup_20120808e
 
User Guide - PID
User Guide - PIDUser Guide - PID
User Guide - PID
 

More from Elsa von Licy

Styles of Scientific Reasoning, Scientific Practices and Argument in Science ...
Styles of Scientific Reasoning, Scientific Practices and Argument in Science ...Styles of Scientific Reasoning, Scientific Practices and Argument in Science ...
Styles of Scientific Reasoning, Scientific Practices and Argument in Science ...
Elsa von Licy
 
L agressivite en psychanalyse (21 pages 184 ko)
L agressivite en psychanalyse (21 pages   184 ko)L agressivite en psychanalyse (21 pages   184 ko)
L agressivite en psychanalyse (21 pages 184 ko)
Elsa von Licy
 
Vuillez jean philippe_p01
Vuillez jean philippe_p01Vuillez jean philippe_p01
Vuillez jean philippe_p01
Elsa von Licy
 
Spr ue3.1 poly cours et exercices
Spr ue3.1   poly cours et exercicesSpr ue3.1   poly cours et exercices
Spr ue3.1 poly cours et exercices
Elsa von Licy
 
Plan de cours all l1 l2l3m1m2 p
Plan de cours all l1 l2l3m1m2 pPlan de cours all l1 l2l3m1m2 p
Plan de cours all l1 l2l3m1m2 p
Elsa von Licy
 
Bioph pharm 1an-viscosit-des_liquides_et_des_solutions
Bioph pharm 1an-viscosit-des_liquides_et_des_solutionsBioph pharm 1an-viscosit-des_liquides_et_des_solutions
Bioph pharm 1an-viscosit-des_liquides_et_des_solutions
Elsa von Licy
 
Poly histologie-et-embryologie-medicales
Poly histologie-et-embryologie-medicalesPoly histologie-et-embryologie-medicales
Poly histologie-et-embryologie-medicales
Elsa von Licy
 
Methodes travail etudiants
Methodes travail etudiantsMethodes travail etudiants
Methodes travail etudiants
Elsa von Licy
 
There is no_such_thing_as_a_social_science_intro
There is no_such_thing_as_a_social_science_introThere is no_such_thing_as_a_social_science_intro
There is no_such_thing_as_a_social_science_intro
Elsa von Licy
 

More from Elsa von Licy (20)

Styles of Scientific Reasoning, Scientific Practices and Argument in Science ...
Styles of Scientific Reasoning, Scientific Practices and Argument in Science ...Styles of Scientific Reasoning, Scientific Practices and Argument in Science ...
Styles of Scientific Reasoning, Scientific Practices and Argument in Science ...
 
Strategie Decisions Incertitude Actes conference fnege xerfi
Strategie Decisions Incertitude Actes conference fnege xerfiStrategie Decisions Incertitude Actes conference fnege xerfi
Strategie Decisions Incertitude Actes conference fnege xerfi
 
Rainville pierre
Rainville pierreRainville pierre
Rainville pierre
 
Neuropsychophysiologie
NeuropsychophysiologieNeuropsychophysiologie
Neuropsychophysiologie
 
L agressivite en psychanalyse (21 pages 184 ko)
L agressivite en psychanalyse (21 pages   184 ko)L agressivite en psychanalyse (21 pages   184 ko)
L agressivite en psychanalyse (21 pages 184 ko)
 
C1 clef pour_la_neuro
C1 clef pour_la_neuroC1 clef pour_la_neuro
C1 clef pour_la_neuro
 
Hemostase polycop
Hemostase polycopHemostase polycop
Hemostase polycop
 
Antiphilos
AntiphilosAntiphilos
Antiphilos
 
Vuillez jean philippe_p01
Vuillez jean philippe_p01Vuillez jean philippe_p01
Vuillez jean philippe_p01
 
Spr ue3.1 poly cours et exercices
Spr ue3.1   poly cours et exercicesSpr ue3.1   poly cours et exercices
Spr ue3.1 poly cours et exercices
 
Plan de cours all l1 l2l3m1m2 p
Plan de cours all l1 l2l3m1m2 pPlan de cours all l1 l2l3m1m2 p
Plan de cours all l1 l2l3m1m2 p
 
M2 bmc2007 cours01
M2 bmc2007 cours01M2 bmc2007 cours01
M2 bmc2007 cours01
 
Feuilletage
FeuilletageFeuilletage
Feuilletage
 
Chapitre 1
Chapitre 1Chapitre 1
Chapitre 1
 
Biophy
BiophyBiophy
Biophy
 
Bioph pharm 1an-viscosit-des_liquides_et_des_solutions
Bioph pharm 1an-viscosit-des_liquides_et_des_solutionsBioph pharm 1an-viscosit-des_liquides_et_des_solutions
Bioph pharm 1an-viscosit-des_liquides_et_des_solutions
 
Poly histologie-et-embryologie-medicales
Poly histologie-et-embryologie-medicalesPoly histologie-et-embryologie-medicales
Poly histologie-et-embryologie-medicales
 
Methodes travail etudiants
Methodes travail etudiantsMethodes travail etudiants
Methodes travail etudiants
 
Atelier.etude.efficace
Atelier.etude.efficaceAtelier.etude.efficace
Atelier.etude.efficace
 
There is no_such_thing_as_a_social_science_intro
There is no_such_thing_as_a_social_science_introThere is no_such_thing_as_a_social_science_intro
There is no_such_thing_as_a_social_science_intro
 

Instrument setuprotorgeneq

  • 1. Technical Note QIAGEN Rotor Gene Q: Software V2.0 Instrument Setup Instructions for RT2 Profiler™ PCR Arrays Preparation Before the Experiment (Presetting the Machine will save time for your run): • Please make note of the Rotor-Gene Q Series software version on your instrument. • For more information, please click the help button or refer to the Rotor-Gene Q Manual available at http://www.qiagen.com. • Set up a PCR protocol template file on the Rotor-Gene Q Series software as follows: 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. 15. 16. 17. 18. 19. Open the Rotor-Gene Q Series Software 2.0 on the desktop of the computer that is connected to the Rotor-Gene Q. Select File ► New. The New Run dialog box will appear (Note: the New Run dialog box may open automatically). Under the Advanced tab, select Two Step. Click New. Under the Welcome to the Advanced Run Wizard! tab, select Rotor-Disc 100. Ensure locking ring has been attached to the Rotor-Disc 100 and check Locking Ring Attached box. Click Next. Under the Miscellaneous Options tab, set Reaction Volume (µl) to 20. Click Next. Click the Edit Profile tab. Under the Edit Profile tab (Figure 1), adjust parameters to reflect the following: Hold • Hold Temperature: 95° C • Hold Time: 10 mins 0 secs Cycling • This cycle repeats 40 time(s) • 95° , 15 seconds, Not Acquiring C • 60° , 30 seconds, Acquiring to Cycling A on Green C Click Insert After ► New Melt. Ensure Optimize gain before melt on all tubes is checked. Click Ok. Click Gain Optimisation. Under the Auto-Gain Optimisation Setup tab, click Optimise Acquiring. Click Ok. Ensure Perform Optimisation Before 1st Acquistion is checked. Click Close. Click Next. Click Save Template and enter RT2_Rotor_Gene_Q as the template name. Click Save. Page 1 of 5
  • 2. Technical Note Figure 1 Edit Profile Tab Performing Real-Time PCR Detection 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. If the Rotor-Gene Q is off, switch on the instrument, and ensure the standby light is lit. Open the Rotor-Gene Q Series Software 2.0. Under the New Run dialog box, click on the Quick Start tab, and select Open a Template In Another Folder. Click New. Locate RT2_Rotor_Gene_Q Template file. Click Open. Under the 1. Rotor Selection tab, select Rotor-Disc 100. Ensure locking ring has been attached to the Rotor-Disc 100 and check Locking Ring Attached box. Click Next. Under 2. Confirm Profile tab, verify desired profile. Click Start Run. Enter name for run and click Save. Rotor-Gene Q run will now commence. After the PCR Run 1. 2. 3. 4. 5. 6. Once the PCR run is complete, observe the Sample Bank (right side of screen, Figure 2). Click Bank On. Click All On. Select Analysis in program bar. Under Quantitation tab, select Cycling A. Green. Click Show. Page 2 of 5
  • 3. Technical Note 7. Calculate the threshold cycle (Ct) for each well using the instrument’s software. a. To define the Baseline (Figure 1): i. Observe amplification plots in Linear View. ii. Select Dynamic Tube (default analysis setting) to ensure the average background of each well is determined just before amplification commences. iii. (Optional) Select Ignore First. Fluorescent signal from the initial cycles may not be representative of the remainder of the run. Thus, better results may be achieved if the initial cycles are ignored. Up to 5 cycles can be ignored. iv. (Optional) Select Noise Slope Correction. Selection of this option can improve data whose baseline (initial cycles) is noticeably sloped. Noise Slope Correction improves the data when raw data backgrounds are observed to slope upward or downward before the takeoff point (Ct). *IMPORTANT: Ensure that all selections remain consistent across all PCR Array runs in the same analysis. Figure 2 Setting the Baseline b. Manually define the Threshold Value (Figure 3): i. Observe the Log View of the amplification plots. ii. In the Ct Calculation box (under Sample Bank) click the button beside the Threshold box. iii. Move mouse to Amplification plot and click mouse to place threshold above the background signal but within the lower one-third to lower one-half of the linear phase of the amplification plot. iv. Right click on Quant. Results window. v. Click Export to Excel. Page 3 of 5
  • 4. Technical Note c. vi. This file format can be opened in the Microsoft Excel Program. Figure 4 illustrates the layout of the Rotor-Gene Q analysis window. Figure 3 Setting the Threshold Page 4 of 5
  • 5. Technical Note Figure 4 Rotor-Gene Q Analysis Window Page 5 of 5