Instruction Guide for Gas Phase Chromatograph
Preparation:
      Put a sheet of paper in the printer (always make sure the paper is not jamming in the
      printer).
      Position both 3-way-valve between the bottle table and GC chromographer- make sure
      the nozzle is perpendicular. (For reference see diagram.)

Vorbereitung
      Blatt am Drucker einlegen.
      Stellung der beiden 3-Weg-Ventile zwischen Flashenbank und GC überprüfen (Soll-
      zustand siehe Abbildung)




                                      Step 1. Calibration
                                       Schritt 1. Eichen

Main menu [Hauptmenu (1)] select Data set (Messreihe angeben) (Put the numbers in one at
a time with Enter).

   (5) Loading / laden (5)
       1. Calibration 1 (low gas concentration/ niedrige Gaskonzentrationen, dry location /
           trockene Standort; E0, E1, E2). Important: set the control panel of the
           Chromotograph to:

                 “DET”       “1”     “Enter”     “Range”      ”0”     “Enter”

       2. Calibration 2 (high gas concentration/ hohe Gaskonzentrationen, wet location/
          feuchte Standorte; E1, E2, E3). Important: set the control panel of the
          Chromotograph to:

                  “DET” “1” “Enter” “Range” ”1” “Enter”
   Type 0 to return to main menu
   (3) Performing analysis / Analyse durchführen Enter „j“ (Data name) Enter
   (6) Gas Processor / Gasprozessor
   (6) Graphics / Graphiken
        1. (1) FID (Flame ionisation detector, sensitive for CH4) and ECD (electron capture
detector, sensitive for N20, CO2). Important: while E0 is being measured, correct the base
curve of the measurement.
               -Position the base curve of the FID (gray) over the upper small wheel on the
        Chromotographer, above the top of the red mark on the monitor.
               - Level the white base curve of the ECD above the lower small wheel on
        Chromatographer with the blue marking on display screen.
During or after the calibration, plug the glass bottles into the corresponding plots according to
the numbers listed in the protocol but DON’T open.

More Related Content

Similar to Instruction guide for gas phase chromatograph

Instruction guide for gas phase chromatograph
Instruction guide for gas phase chromatographInstruction guide for gas phase chromatograph
Instruction guide for gas phase chromatographRodrigo Rudge
 
Instruction guide for gas phase chromatograph
Instruction guide for gas phase chromatographInstruction guide for gas phase chromatograph
Instruction guide for gas phase chromatographRodrigo Rudge
 
Cn abi7900 ht_setup_201200808e
Cn abi7900 ht_setup_201200808eCn abi7900 ht_setup_201200808e
Cn abi7900 ht_setup_201200808eElsa von Licy
 
Wiring a pH and Conductivity Probe to a Zeno3200
Wiring a pH and Conductivity Probe to a Zeno3200Wiring a pH and Conductivity Probe to a Zeno3200
Wiring a pH and Conductivity Probe to a Zeno3200TAMUK
 
As practical-exam-notes
As practical-exam-notesAs practical-exam-notes
As practical-exam-noteszohaib201
 
As practical-exam-notes
As practical-exam-notesAs practical-exam-notes
As practical-exam-noteszohaib201
 
Lecture_(4)_R.E.M and Area Calculation
Lecture_(4)_R.E.M and Area CalculationLecture_(4)_R.E.M and Area Calculation
Lecture_(4)_R.E.M and Area CalculationBahzad5
 
Medidor digital de presion 680 UserManual.pdf
Medidor digital de presion 680 UserManual.pdfMedidor digital de presion 680 UserManual.pdf
Medidor digital de presion 680 UserManual.pdfJuanRios958368
 
Cn mx3000 p_20120808e
Cn mx3000 p_20120808eCn mx3000 p_20120808e
Cn mx3000 p_20120808eElsa von Licy
 
Termometro IR con guia laser 320 manual
Termometro IR con guia laser 320 manualTermometro IR con guia laser 320 manual
Termometro IR con guia laser 320 manualAlberto Angel Villa
 

Similar to Instruction guide for gas phase chromatograph (20)

Instruction guide for gas phase chromatograph
Instruction guide for gas phase chromatographInstruction guide for gas phase chromatograph
Instruction guide for gas phase chromatograph
 
Instruction guide for gas phase chromatograph
Instruction guide for gas phase chromatographInstruction guide for gas phase chromatograph
Instruction guide for gas phase chromatograph
 
Phoenics tutorial
Phoenics tutorialPhoenics tutorial
Phoenics tutorial
 
Gc step 2 (1)
Gc step 2 (1)Gc step 2 (1)
Gc step 2 (1)
 
Gc step 2 (1)
Gc step 2 (1)Gc step 2 (1)
Gc step 2 (1)
 
Gc step 2 (1)
Gc step 2 (1)Gc step 2 (1)
Gc step 2 (1)
 
Aiwa tv-f2400
Aiwa tv-f2400Aiwa tv-f2400
Aiwa tv-f2400
 
Gc step 2
Gc step 2Gc step 2
Gc step 2
 
Gc step 2
Gc step 2Gc step 2
Gc step 2
 
Gc step 2
Gc step 2Gc step 2
Gc step 2
 
Cn abi7900 ht_setup_201200808e
Cn abi7900 ht_setup_201200808eCn abi7900 ht_setup_201200808e
Cn abi7900 ht_setup_201200808e
 
Sokkia set20 k
Sokkia set20 kSokkia set20 k
Sokkia set20 k
 
Wiring a pH and Conductivity Probe to a Zeno3200
Wiring a pH and Conductivity Probe to a Zeno3200Wiring a pH and Conductivity Probe to a Zeno3200
Wiring a pH and Conductivity Probe to a Zeno3200
 
As practical-exam-notes
As practical-exam-notesAs practical-exam-notes
As practical-exam-notes
 
As practical-exam-notes
As practical-exam-notesAs practical-exam-notes
As practical-exam-notes
 
Lecture_(4)_R.E.M and Area Calculation
Lecture_(4)_R.E.M and Area CalculationLecture_(4)_R.E.M and Area Calculation
Lecture_(4)_R.E.M and Area Calculation
 
Data logging (1)
Data logging (1)Data logging (1)
Data logging (1)
 
Medidor digital de presion 680 UserManual.pdf
Medidor digital de presion 680 UserManual.pdfMedidor digital de presion 680 UserManual.pdf
Medidor digital de presion 680 UserManual.pdf
 
Cn mx3000 p_20120808e
Cn mx3000 p_20120808eCn mx3000 p_20120808e
Cn mx3000 p_20120808e
 
Termometro IR con guia laser 320 manual
Termometro IR con guia laser 320 manualTermometro IR con guia laser 320 manual
Termometro IR con guia laser 320 manual
 

More from Rodrigo Rudge

2. site characteristics
2. site characteristics2. site characteristics
2. site characteristicsRodrigo Rudge
 
Instruction guide for gas phase chromatograph
Instruction guide for gas phase chromatographInstruction guide for gas phase chromatograph
Instruction guide for gas phase chromatographRodrigo Rudge
 
Instruction guide for gas phase chromatograph
Instruction guide for gas phase chromatographInstruction guide for gas phase chromatograph
Instruction guide for gas phase chromatographRodrigo Rudge
 

More from Rodrigo Rudge (7)

2. site characteristics
2. site characteristics2. site characteristics
2. site characteristics
 
Gc read out
Gc read outGc read out
Gc read out
 
Instruction guide for gas phase chromatograph
Instruction guide for gas phase chromatographInstruction guide for gas phase chromatograph
Instruction guide for gas phase chromatograph
 
Gc read out
Gc read outGc read out
Gc read out
 
Instruction guide for gas phase chromatograph
Instruction guide for gas phase chromatographInstruction guide for gas phase chromatograph
Instruction guide for gas phase chromatograph
 
Gc read out
Gc read outGc read out
Gc read out
 
3. co2 measurement
3. co2 measurement3. co2 measurement
3. co2 measurement
 

Instruction guide for gas phase chromatograph

  • 1. Instruction Guide for Gas Phase Chromatograph Preparation: Put a sheet of paper in the printer (always make sure the paper is not jamming in the printer). Position both 3-way-valve between the bottle table and GC chromographer- make sure the nozzle is perpendicular. (For reference see diagram.) Vorbereitung Blatt am Drucker einlegen. Stellung der beiden 3-Weg-Ventile zwischen Flashenbank und GC überprüfen (Soll- zustand siehe Abbildung) Step 1. Calibration Schritt 1. Eichen Main menu [Hauptmenu (1)] select Data set (Messreihe angeben) (Put the numbers in one at a time with Enter). (5) Loading / laden (5) 1. Calibration 1 (low gas concentration/ niedrige Gaskonzentrationen, dry location / trockene Standort; E0, E1, E2). Important: set the control panel of the Chromotograph to: “DET” “1” “Enter” “Range” ”0” “Enter” 2. Calibration 2 (high gas concentration/ hohe Gaskonzentrationen, wet location/ feuchte Standorte; E1, E2, E3). Important: set the control panel of the Chromotograph to: “DET” “1” “Enter” “Range” ”1” “Enter” Type 0 to return to main menu (3) Performing analysis / Analyse durchführen Enter „j“ (Data name) Enter (6) Gas Processor / Gasprozessor (6) Graphics / Graphiken 1. (1) FID (Flame ionisation detector, sensitive for CH4) and ECD (electron capture detector, sensitive for N20, CO2). Important: while E0 is being measured, correct the base curve of the measurement. -Position the base curve of the FID (gray) over the upper small wheel on the Chromotographer, above the top of the red mark on the monitor. - Level the white base curve of the ECD above the lower small wheel on Chromatographer with the blue marking on display screen. During or after the calibration, plug the glass bottles into the corresponding plots according to the numbers listed in the protocol but DON’T open.