1. Measuring pH Using the Orion
3 Star Plus pH Meter with Printer
Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 1
2. Orion 3 Star Plus pH Meter
Front of Meter & Keypad
Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 2
3. Start Up
• Turn Printer ON
• Press POWER Key on
Meter
Display will read:
Star PLUS
Software Version and
current method
Header information will
print:
Meter Serial Number
Meter Software Version
Method currently set
• Check print quality
Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 3
4. Calibrate Method 1
Open
Fill Hole
Rinse electrode and
Select Method (#
blot carefully
decimals) to be used
for samples
Press MEASURE
to set method
Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 4
5. Calibrate Method 1
CAL 1 Displayed..
Press CALIBRATE
• pH will stop flashing when
reading stable
• Buffer value will be displayed
Immerse electrode
with leading zeroes
in first buffer
Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 5
6. Calibrate Method 1
After pressing CALIBRATE,
• CAL 2 displayed
• pH flashes while reading in
progress
• Remove electrode from
first buffer, rinse & blot
• Immerse electrode in
second buffer
• pH will stop flashing when
reading stable
• Buffer value will be displayed
with leading zeroes
Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 6
7. Calibrate Method 1
After pressing CALIBRATE,• Remove electrode from
• CAL 3 displayed
pH will stop flashing when
second buffer, rinse & blot
• pH flashes while reading in
reading stable • Immerse electrode in third
• progress
Buffer value will be displayed
buffer
with leading zeroes
Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 7
8. Completing Calibration
Press MEASURE to print
& save calibration results
• Check date and time
(not visible in photo)
• Check mV for each
buffer
• Check slope 1 and slope
2
• Initial & date printout
Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 8
9. Verifying Instrument Performance
• Remove electrode from
buffer, rinse & blot
• Immerse electrode in QC
Press MEASURE (6.86) buffer
pH will stop flashing when
reading stable
Press MEASURE
to print result
• Initial & date
• Identify sample as
QC Buffer
Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 9
10. Documenting Calibration
Calibration & Initial
QC Buffer on form
• Check Dates & Times
• Check Method
• Check Calibration #
• Check Limits:
• 4.01 +177 mV ± 30 mV
• 7.00 0 mV ± 30 mV
• 10.01 -177 mV ± 30 mV
• Slope 1 and 2:
95% - 102%
• QC Buffer: 6.86 ± 0.05
Method 1: 6.8 – 6.9
Method 2: 6.81 – 6.91
Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 10
11. Measuring Sample pH
• Remove electrode from
buffer or storage solution,
rinse & blot
• Immerse electrode in
Press MEASURE sample
pH will stop flashing when
reading stable
Press MEASURE
again to print result
• Initial & date
• Identify sample
Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 11
12. Reviewing Results
• Check Dates & Times
• If > 2 hours since last QC Buffer run,
instrument performance must be
verified by running the QC Buffer as a
Batch control
• Check Method
• 1 decimal for any product QC and most
reagents/solution
• 2 decimals if reagent procedure
requires more sensitivity
• NO ROUNDING
• Check Calibration #
• Match to the most recent calibration for
the method selected
• Attach product results to Product QC
Forms
• Attach reagent/solution pH
measurements to Reagent pH Log
[19.4.frm024]
Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 12