Measuring pH Using the Orion
3 Star Plus pH Meter with Printer




        Measuring pH Using the Orion 3 Star Plus pH Meter with Printer   1
Orion 3 Star Plus pH Meter




       Front of Meter & Keypad



     Measuring pH Using the Orion 3 Star Plus pH Meter with Printer   2
Start Up
• Turn Printer ON
• Press POWER Key on
  Meter
    Display will read:
        Star PLUS
        Software Version and
         current method
    Header information will
     print:
        Meter Serial Number
        Meter Software Version
        Method currently set
• Check print quality
                  Measuring pH Using the Orion 3 Star Plus pH Meter with Printer   3
Calibrate Method 1


      Open
     Fill Hole

                                                  Rinse electrode and
                                Select Method (#
                                                  blot carefully
                                decimals) to be used
                                for samples




                                   Press MEASURE
                                   to set method


 Measuring pH Using the Orion 3 Star Plus pH Meter with Printer         4
Calibrate Method 1
                                                            CAL 1 Displayed..




Press CALIBRATE




                        • pH will stop flashing when
                          reading stable
                        • Buffer value will be displayed
                                  Immerse electrode
                          with leading zeroes
                                  in first buffer

                   Measuring pH Using the Orion 3 Star Plus pH Meter with Printer   5
Calibrate Method 1
                                After pressing CALIBRATE,
                                • CAL 2 displayed
                                • pH flashes while reading in
                                  progress




                                                           • Remove electrode from
                                                             first buffer, rinse & blot
                                                           • Immerse electrode in
                                                             second buffer

            • pH will stop flashing when
              reading stable
            • Buffer value will be displayed
              with leading zeroes


 Measuring pH Using the Orion 3 Star Plus pH Meter with Printer                       6
Calibrate Method 1




                       After pressing CALIBRATE,• Remove electrode from
                       • CAL 3 displayed
                         pH will stop flashing when
                                                  second buffer, rinse & blot
                       • pH flashes while reading in
                         reading stable         • Immerse electrode in third
                       • progress
                         Buffer value will be displayed
                                                  buffer
                         with leading zeroes




 Measuring pH Using the Orion 3 Star Plus pH Meter with Printer            7
Completing Calibration
                               Press MEASURE to print
                               & save calibration results




                                                             • Check date and time
                                                               (not visible in photo)
                                                             • Check mV for each
                                                               buffer
                                                             • Check slope 1 and slope
                                                               2
                                                             • Initial & date printout




   Measuring pH Using the Orion 3 Star Plus pH Meter with Printer                   8
Verifying Instrument Performance
                                                                • Remove electrode from
                                                                  buffer, rinse & blot
                                                                • Immerse electrode in QC
Press MEASURE                                                     (6.86) buffer


                                                     pH will stop flashing when
                                                      reading stable


                                 Press MEASURE
                                 to print result


                                 • Initial & date
                                 • Identify sample as
                                   QC Buffer




                Measuring pH Using the Orion 3 Star Plus pH Meter with Printer              9
Documenting Calibration

 Calibration & Initial
 QC Buffer on form

• Check Dates & Times
• Check Method
• Check Calibration #
• Check Limits:
  • 4.01 +177 mV ± 30 mV
  • 7.00 0 mV ± 30 mV
  • 10.01 -177 mV ± 30 mV
  • Slope 1 and 2:
         95% - 102%
  • QC Buffer: 6.86 ± 0.05
      Method 1: 6.8 – 6.9
      Method 2: 6.81 – 6.91




                              Measuring pH Using the Orion 3 Star Plus pH Meter with Printer   10
Measuring Sample pH
                                                                           • Remove electrode from
                                                                             buffer or storage solution,
                                                                             rinse & blot
                                                                           • Immerse electrode in
Press MEASURE                                                                sample


                                                      pH will stop flashing when
                                                       reading stable



                          Press MEASURE
                          again to print result




                                • Initial & date
                                • Identify sample
                Measuring pH Using the Orion 3 Star Plus pH Meter with Printer                     11
Reviewing Results
• Check Dates & Times
  • If > 2 hours since last QC Buffer run,
    instrument performance must be
    verified by running the QC Buffer as a
    Batch control

    • Check Method
      • 1 decimal for any product QC and most
        reagents/solution
      • 2 decimals if reagent procedure
        requires more sensitivity
      • NO ROUNDING

         • Check Calibration #
            • Match to the most recent calibration for
              the method selected

             • Attach product results to Product QC
               Forms
             • Attach reagent/solution pH
               measurements to Reagent pH Log
               [19.4.frm024]
                           Measuring pH Using the Orion 3 Star Plus pH Meter with Printer   12

More Related Content

Viewers also liked

Calibrationofglassware2016
Calibrationofglassware2016Calibrationofglassware2016
Calibrationofglassware2016Abhilasha Pandey
 
Raaj Global Pharma Regulatory Affairs Consultants Thane-mumbai profile-updat...
Raaj Global Pharma Regulatory Affairs  Consultants Thane-mumbai profile-updat...Raaj Global Pharma Regulatory Affairs  Consultants Thane-mumbai profile-updat...
Raaj Global Pharma Regulatory Affairs Consultants Thane-mumbai profile-updat...Rajashri Survase Ojha
 
Importance of instumental analysis
Importance of instumental analysisImportance of instumental analysis
Importance of instumental analysisDr. Nilesh Badgujar
 
Storage and handling of chemicals
Storage and handling of chemicalsStorage and handling of chemicals
Storage and handling of chemicalsSuresh Selvaraj
 
Analytical chemistry_Instrumentation_Introduction
Analytical chemistry_Instrumentation_IntroductionAnalytical chemistry_Instrumentation_Introduction
Analytical chemistry_Instrumentation_IntroductionBivek Timalsina
 
Reference standards in Pharmaceutical Industries
Reference standards in Pharmaceutical IndustriesReference standards in Pharmaceutical Industries
Reference standards in Pharmaceutical Industriesbhavanavedantam
 
Instrument Calibration
Instrument Calibration Instrument Calibration
Instrument Calibration Ramesh Dham
 
Analytical Method Validation
Analytical Method ValidationAnalytical Method Validation
Analytical Method ValidationStefan Holt
 
Introduction to analytical chemistry
Introduction to analytical chemistryIntroduction to analytical chemistry
Introduction to analytical chemistryUMAR ALI
 
Smartest Resume Guide for Students and Freshers
Smartest Resume Guide for Students and FreshersSmartest Resume Guide for Students and Freshers
Smartest Resume Guide for Students and FreshersTwenty19.com
 
Pharma regulatory affairs
Pharma regulatory affairsPharma regulatory affairs
Pharma regulatory affairsGirish Swami
 
HPLC Principle,Instrumentation and Application
HPLC Principle,Instrumentation and ApplicationHPLC Principle,Instrumentation and Application
HPLC Principle,Instrumentation and ApplicationAlakesh Pradhan
 
LinkedIn SlideShare: Knowledge, Well-Presented
LinkedIn SlideShare: Knowledge, Well-PresentedLinkedIn SlideShare: Knowledge, Well-Presented
LinkedIn SlideShare: Knowledge, Well-PresentedSlideShare
 

Viewers also liked (18)

Calibrationofglassware2016
Calibrationofglassware2016Calibrationofglassware2016
Calibrationofglassware2016
 
Manish ppt
Manish pptManish ppt
Manish ppt
 
Raaj Global Pharma Regulatory Affairs Consultants Thane-mumbai profile-updat...
Raaj Global Pharma Regulatory Affairs  Consultants Thane-mumbai profile-updat...Raaj Global Pharma Regulatory Affairs  Consultants Thane-mumbai profile-updat...
Raaj Global Pharma Regulatory Affairs Consultants Thane-mumbai profile-updat...
 
Importance of instumental analysis
Importance of instumental analysisImportance of instumental analysis
Importance of instumental analysis
 
Storage and handling of chemicals
Storage and handling of chemicalsStorage and handling of chemicals
Storage and handling of chemicals
 
Uv,vis,nmr,mass,ir
Uv,vis,nmr,mass,irUv,vis,nmr,mass,ir
Uv,vis,nmr,mass,ir
 
Analytical chemistry_Instrumentation_Introduction
Analytical chemistry_Instrumentation_IntroductionAnalytical chemistry_Instrumentation_Introduction
Analytical chemistry_Instrumentation_Introduction
 
Reference standards in Pharmaceutical Industries
Reference standards in Pharmaceutical IndustriesReference standards in Pharmaceutical Industries
Reference standards in Pharmaceutical Industries
 
P h and buffer
P h and bufferP h and buffer
P h and buffer
 
Instrument Calibration
Instrument Calibration Instrument Calibration
Instrument Calibration
 
Analytical Method Validation
Analytical Method ValidationAnalytical Method Validation
Analytical Method Validation
 
Introduction to analytical chemistry
Introduction to analytical chemistryIntroduction to analytical chemistry
Introduction to analytical chemistry
 
HPLC
HPLCHPLC
HPLC
 
Uv seminar ppt
Uv seminar pptUv seminar ppt
Uv seminar ppt
 
Smartest Resume Guide for Students and Freshers
Smartest Resume Guide for Students and FreshersSmartest Resume Guide for Students and Freshers
Smartest Resume Guide for Students and Freshers
 
Pharma regulatory affairs
Pharma regulatory affairsPharma regulatory affairs
Pharma regulatory affairs
 
HPLC Principle,Instrumentation and Application
HPLC Principle,Instrumentation and ApplicationHPLC Principle,Instrumentation and Application
HPLC Principle,Instrumentation and Application
 
LinkedIn SlideShare: Knowledge, Well-Presented
LinkedIn SlideShare: Knowledge, Well-PresentedLinkedIn SlideShare: Knowledge, Well-Presented
LinkedIn SlideShare: Knowledge, Well-Presented
 

More from Rick Presley

Standardized MPF Cover Sheet
Standardized MPF Cover SheetStandardized MPF Cover Sheet
Standardized MPF Cover SheetRick Presley
 
Problem Solving Flow Chart
Problem Solving Flow ChartProblem Solving Flow Chart
Problem Solving Flow ChartRick Presley
 
Standardized mpf cover sheet stand alone draft 04
Standardized mpf cover sheet stand alone   draft 04Standardized mpf cover sheet stand alone   draft 04
Standardized mpf cover sheet stand alone draft 04Rick Presley
 
Manual Problem Form Training
Manual Problem Form TrainingManual Problem Form Training
Manual Problem Form TrainingRick Presley
 

More from Rick Presley (6)

Mobile Inventory
Mobile InventoryMobile Inventory
Mobile Inventory
 
Standardized MPF Cover Sheet
Standardized MPF Cover SheetStandardized MPF Cover Sheet
Standardized MPF Cover Sheet
 
Problem Solving Flow Chart
Problem Solving Flow ChartProblem Solving Flow Chart
Problem Solving Flow Chart
 
Standardized mpf cover sheet stand alone draft 04
Standardized mpf cover sheet stand alone   draft 04Standardized mpf cover sheet stand alone   draft 04
Standardized mpf cover sheet stand alone draft 04
 
Blood types
Blood typesBlood types
Blood types
 
Manual Problem Form Training
Manual Problem Form TrainingManual Problem Form Training
Manual Problem Form Training
 

Measuring pH with orion 3star with printer workshop

  • 1. Measuring pH Using the Orion 3 Star Plus pH Meter with Printer Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 1
  • 2. Orion 3 Star Plus pH Meter Front of Meter & Keypad Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 2
  • 3. Start Up • Turn Printer ON • Press POWER Key on Meter  Display will read:  Star PLUS  Software Version and current method  Header information will print:  Meter Serial Number  Meter Software Version  Method currently set • Check print quality Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 3
  • 4. Calibrate Method 1 Open Fill Hole Rinse electrode and Select Method (# blot carefully decimals) to be used for samples Press MEASURE to set method Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 4
  • 5. Calibrate Method 1 CAL 1 Displayed.. Press CALIBRATE • pH will stop flashing when reading stable • Buffer value will be displayed Immerse electrode with leading zeroes in first buffer Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 5
  • 6. Calibrate Method 1 After pressing CALIBRATE, • CAL 2 displayed • pH flashes while reading in progress • Remove electrode from first buffer, rinse & blot • Immerse electrode in second buffer • pH will stop flashing when reading stable • Buffer value will be displayed with leading zeroes Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 6
  • 7. Calibrate Method 1 After pressing CALIBRATE,• Remove electrode from • CAL 3 displayed pH will stop flashing when second buffer, rinse & blot • pH flashes while reading in reading stable • Immerse electrode in third • progress Buffer value will be displayed buffer with leading zeroes Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 7
  • 8. Completing Calibration Press MEASURE to print & save calibration results • Check date and time (not visible in photo) • Check mV for each buffer • Check slope 1 and slope 2 • Initial & date printout Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 8
  • 9. Verifying Instrument Performance • Remove electrode from buffer, rinse & blot • Immerse electrode in QC Press MEASURE (6.86) buffer pH will stop flashing when reading stable Press MEASURE to print result • Initial & date • Identify sample as QC Buffer Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 9
  • 10. Documenting Calibration Calibration & Initial QC Buffer on form • Check Dates & Times • Check Method • Check Calibration # • Check Limits: • 4.01 +177 mV ± 30 mV • 7.00 0 mV ± 30 mV • 10.01 -177 mV ± 30 mV • Slope 1 and 2: 95% - 102% • QC Buffer: 6.86 ± 0.05 Method 1: 6.8 – 6.9 Method 2: 6.81 – 6.91 Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 10
  • 11. Measuring Sample pH • Remove electrode from buffer or storage solution, rinse & blot • Immerse electrode in Press MEASURE sample pH will stop flashing when reading stable Press MEASURE again to print result • Initial & date • Identify sample Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 11
  • 12. Reviewing Results • Check Dates & Times • If > 2 hours since last QC Buffer run, instrument performance must be verified by running the QC Buffer as a Batch control • Check Method • 1 decimal for any product QC and most reagents/solution • 2 decimals if reagent procedure requires more sensitivity • NO ROUNDING • Check Calibration # • Match to the most recent calibration for the method selected • Attach product results to Product QC Forms • Attach reagent/solution pH measurements to Reagent pH Log [19.4.frm024] Measuring pH Using the Orion 3 Star Plus pH Meter with Printer 12

Editor's Notes

  1. Front of Meter: Note the different selection buttons Power button press a second time to turn on backlighting