Please read the user's manual carefully before use.
AF-50AF-50
Breath Alcohol Tester
Professional Fuel cell sensor
Personal & professional use alcohol tester
Instruction
Manual
50 sets of data memory
Professional Fuel Cell Sensor
Law enforcement grade accuracy
4-digit LCD display with a backlit
Fast analyzing and result
Automatic power off
User adjustable settings
(Blow time, Unit, Beep alarm,
Blow detecting)
Before using the alcohol tester, read all instruction manual carefully and
practice the test.
Maintenance and calibration should be performed by experts.
Test should be performed after 20 minutes after eating, drinking or smoking,
if failure to wait 20 minutes may result in invalid results or incorrect results.
Residual alcohol in the mouth can distort the measurement.
Aromatic drinks or mouth sprays containing alcohol can interfere with
measurements. In this case, wait for 20 minutes or rinse the mouth with
pure water.
Do not permit to blow cigarette smoke into the tester which may damage
the sensor.
Do not drink and drive, even small amount of alcohol can result in driving
impairment.
Recommended to use calibration service every 12 months or after 500
times tests.
Do not use this device as a tool to determine "drink and drive".
The manufacturer and the sellers will not take any legal responsibility for
the test results.
When not used for a long time, keep the device after taking out the batteries.
In case of testing after long-term storage, there might be high tolerance of
test results for the first 3 tests.
Replace the batteries when the battery indicator icon reaches empty for the
accuracy reason.
In case the test result is higher than 0.1% BAC (1.0 g/l) and you try again
without interval, then warm up time might be delayed.
Perform a test within the operating temperature, which is specified in the
instruction manual.
Thank you for purchasing our breathalyzer, ALCOFIND AF-50 which is a
personal & professional use alcohol tester with extremely high-reliability.
It can be used for in the hospital, school and many other industrial fields.
It measures estimated B.A.C(Blood Alcohol Content) in human breath by
means of a fuel cell sensor which is high durable sensors and providing
accurate breath alcohol results.
Important safety information
Introduction
Preparation
Performing the measurement
<Front> <Back>
Install two AA batteries in the battery compartment by paying attention to
the battery polarity.
Insert a mouthpiece into the port on the side of the alcohol tester.
Ensure whether it is securely fit or not.
A new mouthpieces is to be used for every person tested.
Two AA batteries (1.5V Alkaline)
Press the power button
The total number of test performed will be displayed
briefly and start countdown.
When the countdown reaches zero with a double beep sound,
three dashes ( ) will be displayed and "START" will flash.
Take a deep breath and blow the mouthpiece for five seconds
until there is a double beep sound.
Digital Breathalyzer
ALCOFIND AF-50
Two AAsize Batteries
6 (six) Mouthpieces
Carrying Case
User's Manual
Contents
1. Mouthpiece
2. Mouthpiece slot
3. Display
Name of Each Part
4. Power & Start button
5. Mode button
6. Battery cover
Press and pull up the cover to open
Wait at least 20 minutes after eating, drinking or smoking before measurement.
After a few seconds(depending on the alcohol level), the
estimated BAC(Blood Alcohol Concentration) will be displayed.
The value will flash for several seconds.
If you want to test again, pressthe start button to restart the
countdown cycle.
The tester automatically will be turned offafter 15~20 seconds.
When the sampling was successful, LCD outer lines will be
turning along the edge of LCD as you can see the picture.
User Settings & Memory
User may adjust UNIT, BEEP ALARM, BLOW TIME, FLOW DETECTION through Menu mode.
Turn on the tester which may begin a countdown. During the countdown, press and hold
the mode button for more than 5 seconds, the tester may enter Menu mode
& Memory mode
Adjust measurement units
When "Unit" appears on the display, press the start button to adjust the measurement unit.
Adjust beep alarm setting
When "bEEp" appears on the display, press the start button to set the beep alarm.
If user set the beep alarm at 0.05%BAC, the tester will sound an audible warning alarm
when the test result is over 0.05%BAC. If user set "none" the tester will not sound at all.
Recommended to set the value at your country's BAC Limit.
Adjust blow time
When "bL-t" appears on the display, press the start button to adjust the blow time.
A longer blow time may provide a deep lung air which may provide more accurate reading.
Adjust flow detection setting
When "FL-d" appears on the display, press the start button to adjust the flow detection.
Use this option to adjust how hard a user is required to blow.
LO Low level detection, this is default setting.
HI High level flow detection. This level requires a stronger breath. Recommend if previous
test attempts have been unsuccessful.
USER User activated test mode. When testing in this mode, the user must press the start
button during blowing to obtain breath sample and perform a test successfully.
Enter the Memory Mode
When "DAT" appears on the display, press the start button, the most recent test result
will appear. By pressing start button, next test result will appear.
The tester will recall up to 50 test results. If the memory is full, the oldest test result is
deleted automatically.
Replace new batteries.
Replace new batteries
immediately.
Press button again to
repeat the measurement.
Press button to restart
the countdown cycle and retest.
Repeat the test after the unit
has reached a temperature
within the operation range.
Display message & Problem-solving guide
MESSAGE WHAT ITMEANS ACTION
Full battery
Empty battery : Warns when
battery is low. The unit can be
operated but recommended to change
new batteries for correct result!
If the LCD displays "BAt",
the unit cannot operate.
If the sampling volume is too
low or not sufficiently blow.
If there is no breath sample blown
within 30 seconds, the LCD will
display "OUT".
If the unit is powered on beyond the
acceptable temperature range(
) the unit will display it.
5~40 C
or 41~104 F
Full battery
Technical specifications
Indication of B.A.C.
Accuracy
Warm up time
Sensor
Power supply
Battery Life
Dimensions(WxHxD)
Weight
Calibration
Operating temperature
0.000 ~ 0.500%BAC
0.00 ~ 5.00
0.00 ~ 2.50mg/l
+/- 0.005%BAC at 0.1%BAC at 25 C
+/- 0.05 at 1.0 at 25 C
+/- 0.025mg/l at 0.50mg/l at 25 C
Within 15 seconds at 0.50g/l( )
warm up time may vary depending on the BAC measured.
Professional Fuel cell sensor
Two AAsize alkaline batteries
Approx. 1500 tests
55mm x 125mm x 20mm
123g including batteries
every 12 months or after 500 tests
5 ~ 40 C
The specifications are subject to change without prior notice for functional improvement.
Warranty
Customer service
Distributor :
For any questions, after-sales service, product calibration or to purchase
additional mouthpieces, please contact nearest distributor.
The manufacturer warrants the product to be free from defects in workmanship
or material (excluding calibration) under normal use either for 1(one) year from
the date of purchase or 500 breath tests whichever comes earlier.
The manufacturer's obligation under this warranty is limited to replacing, adjusting
or repairing the unit at manufacturer's option, if returned along with proof of
purchase.
For any defects due to natural disaster or customer's negligence, misuse,
reconstruction and repair by unauthorized person, reasonable cost shall be
charged for repair or part replacement.

More Related Content

Viewers also liked

Receiving the gift of salvation
Receiving the gift of salvationReceiving the gift of salvation
Receiving the gift of salvationapologeticspress
 
Kyvyt e-portfolio
Kyvyt e-portfolioKyvyt e-portfolio
Kyvyt e-portfolioIT Oamk
 
TDM Transport over mpls v 1.1
TDM Transport over mpls v 1.1TDM Transport over mpls v 1.1
TDM Transport over mpls v 1.1A Achyar Nur
 
‫Job description-Human Resource Management
‫Job description-Human Resource Management‫Job description-Human Resource Management
‫Job description-Human Resource ManagementAbdalrhman Alawdy
 
BUILDING THE DX LAB FROM SCRATCH: WHAT IS THIS AND WHY DO IT?
BUILDING THE DX LAB FROM SCRATCH: WHAT IS THIS AND WHY DO IT?BUILDING THE DX LAB FROM SCRATCH: WHAT IS THIS AND WHY DO IT?
BUILDING THE DX LAB FROM SCRATCH: WHAT IS THIS AND WHY DO IT?DX Lab
 
Refluxo laringofaríngeo
Refluxo laringofaríngeoRefluxo laringofaríngeo
Refluxo laringofaríngeoElyabe Rosário
 
ENSAYO RESPONSABILIDAD DE LA ADMINISTRACIÓN
ENSAYO  RESPONSABILIDAD DE LA ADMINISTRACIÓNENSAYO  RESPONSABILIDAD DE LA ADMINISTRACIÓN
ENSAYO RESPONSABILIDAD DE LA ADMINISTRACIÓNABDELBONILLA2015
 
logotipos - isotipos -isologotipos
logotipos - isotipos -isologotiposlogotipos - isotipos -isologotipos
logotipos - isotipos -isologotiposIsvan Cardenas
 

Viewers also liked (10)

Receiving the gift of salvation
Receiving the gift of salvationReceiving the gift of salvation
Receiving the gift of salvation
 
Tema #01
Tema #01Tema #01
Tema #01
 
Creative integrated solutions profile V1.0
Creative integrated solutions profile V1.0Creative integrated solutions profile V1.0
Creative integrated solutions profile V1.0
 
Kyvyt e-portfolio
Kyvyt e-portfolioKyvyt e-portfolio
Kyvyt e-portfolio
 
TDM Transport over mpls v 1.1
TDM Transport over mpls v 1.1TDM Transport over mpls v 1.1
TDM Transport over mpls v 1.1
 
‫Job description-Human Resource Management
‫Job description-Human Resource Management‫Job description-Human Resource Management
‫Job description-Human Resource Management
 
BUILDING THE DX LAB FROM SCRATCH: WHAT IS THIS AND WHY DO IT?
BUILDING THE DX LAB FROM SCRATCH: WHAT IS THIS AND WHY DO IT?BUILDING THE DX LAB FROM SCRATCH: WHAT IS THIS AND WHY DO IT?
BUILDING THE DX LAB FROM SCRATCH: WHAT IS THIS AND WHY DO IT?
 
Refluxo laringofaríngeo
Refluxo laringofaríngeoRefluxo laringofaríngeo
Refluxo laringofaríngeo
 
ENSAYO RESPONSABILIDAD DE LA ADMINISTRACIÓN
ENSAYO  RESPONSABILIDAD DE LA ADMINISTRACIÓNENSAYO  RESPONSABILIDAD DE LA ADMINISTRACIÓN
ENSAYO RESPONSABILIDAD DE LA ADMINISTRACIÓN
 
logotipos - isotipos -isologotipos
logotipos - isotipos -isologotiposlogotipos - isotipos -isologotipos
logotipos - isotipos -isologotipos
 

Similar to AF-50-ST-M(2).pdf

Calibration and Qualification of Analytical Instruments
Calibration and Qualification of Analytical InstrumentsCalibration and Qualification of Analytical Instruments
Calibration and Qualification of Analytical InstrumentsISF COLLEGE OF PHARMACY MOGA
 
Maa sharda pharmacy college ayodhya SOP
Maa sharda pharmacy college ayodhya SOPMaa sharda pharmacy college ayodhya SOP
Maa sharda pharmacy college ayodhya SOPJitendraKumarChauhan3
 
Schumacher BT-100 Battery Tester User Manual
Schumacher BT-100 Battery Tester User ManualSchumacher BT-100 Battery Tester User Manual
Schumacher BT-100 Battery Tester User ManualTim Miller
 
Rheological properties using model 900 viscometer
Rheological properties using model 900 viscometerRheological properties using model 900 viscometer
Rheological properties using model 900 viscometerKamal Abdurahman
 
Paradigm Training
Paradigm TrainingParadigm Training
Paradigm Trainingrjinks
 
Hướng dẫn sử dụng Testo 511
Hướng dẫn sử dụng Testo 511Hướng dẫn sử dụng Testo 511
Hướng dẫn sử dụng Testo 511Tenmars Việt Nam
 
Medidor digital de presion 680 UserManual.pdf
Medidor digital de presion 680 UserManual.pdfMedidor digital de presion 680 UserManual.pdf
Medidor digital de presion 680 UserManual.pdfJuanRios958368
 
BA101 vehicle battery tester user manual
BA101 vehicle battery tester user manualBA101 vehicle battery tester user manual
BA101 vehicle battery tester user manualBuyobdtoolShop
 
Hướng dẫn sử dụng máy đo khí CO Extech CO10
Hướng dẫn sử dụng máy đo khí CO Extech CO10Hướng dẫn sử dụng máy đo khí CO Extech CO10
Hướng dẫn sử dụng máy đo khí CO Extech CO10Tenmars Việt Nam
 
Qualification of laboratory equipments by Mayuri Soni
Qualification of laboratory equipments by Mayuri SoniQualification of laboratory equipments by Mayuri Soni
Qualification of laboratory equipments by Mayuri SoniMayuri Soni
 
Actron CP7529 Digital Timing Light User Manual
Actron CP7529 Digital Timing Light User ManualActron CP7529 Digital Timing Light User Manual
Actron CP7529 Digital Timing Light User ManualTim Miller
 

Similar to AF-50-ST-M(2).pdf (20)

Calibration and Qualification of Analytical Instruments
Calibration and Qualification of Analytical InstrumentsCalibration and Qualification of Analytical Instruments
Calibration and Qualification of Analytical Instruments
 
Bpm 210 ul blood pressure monitor
Bpm 210 ul blood pressure monitorBpm 210 ul blood pressure monitor
Bpm 210 ul blood pressure monitor
 
Bpm 210 ul blood pressure monitor
Bpm 210 ul blood pressure monitorBpm 210 ul blood pressure monitor
Bpm 210 ul blood pressure monitor
 
Bpm 210 um blood pressure monitor
Bpm 210 um blood pressure monitorBpm 210 um blood pressure monitor
Bpm 210 um blood pressure monitor
 
Manual acc
Manual accManual acc
Manual acc
 
Manual epack
Manual epackManual epack
Manual epack
 
Erbe icc-80-50-icc-service-manual
Erbe icc-80-50-icc-service-manualErbe icc-80-50-icc-service-manual
Erbe icc-80-50-icc-service-manual
 
Manual epack
Manual epackManual epack
Manual epack
 
Bw 210 blood pressure monitor
Bw 210 blood pressure monitorBw 210 blood pressure monitor
Bw 210 blood pressure monitor
 
Maa sharda pharmacy college ayodhya SOP
Maa sharda pharmacy college ayodhya SOPMaa sharda pharmacy college ayodhya SOP
Maa sharda pharmacy college ayodhya SOP
 
Schumacher BT-100 Battery Tester User Manual
Schumacher BT-100 Battery Tester User ManualSchumacher BT-100 Battery Tester User Manual
Schumacher BT-100 Battery Tester User Manual
 
Rheological properties using model 900 viscometer
Rheological properties using model 900 viscometerRheological properties using model 900 viscometer
Rheological properties using model 900 viscometer
 
Paradigm Training
Paradigm TrainingParadigm Training
Paradigm Training
 
Hướng dẫn sử dụng Testo 511
Hướng dẫn sử dụng Testo 511Hướng dẫn sử dụng Testo 511
Hướng dẫn sử dụng Testo 511
 
Medidor digital de presion 680 UserManual.pdf
Medidor digital de presion 680 UserManual.pdfMedidor digital de presion 680 UserManual.pdf
Medidor digital de presion 680 UserManual.pdf
 
BA101 vehicle battery tester user manual
BA101 vehicle battery tester user manualBA101 vehicle battery tester user manual
BA101 vehicle battery tester user manual
 
Hướng dẫn sử dụng máy đo khí CO Extech CO10
Hướng dẫn sử dụng máy đo khí CO Extech CO10Hướng dẫn sử dụng máy đo khí CO Extech CO10
Hướng dẫn sử dụng máy đo khí CO Extech CO10
 
Viscometer model 900
Viscometer model 900Viscometer model 900
Viscometer model 900
 
Qualification of laboratory equipments by Mayuri Soni
Qualification of laboratory equipments by Mayuri SoniQualification of laboratory equipments by Mayuri Soni
Qualification of laboratory equipments by Mayuri Soni
 
Actron CP7529 Digital Timing Light User Manual
Actron CP7529 Digital Timing Light User ManualActron CP7529 Digital Timing Light User Manual
Actron CP7529 Digital Timing Light User Manual
 

AF-50-ST-M(2).pdf

  • 1. Please read the user's manual carefully before use. AF-50AF-50 Breath Alcohol Tester Professional Fuel cell sensor Personal & professional use alcohol tester Instruction Manual 50 sets of data memory Professional Fuel Cell Sensor Law enforcement grade accuracy 4-digit LCD display with a backlit Fast analyzing and result Automatic power off User adjustable settings (Blow time, Unit, Beep alarm, Blow detecting)
  • 2. Before using the alcohol tester, read all instruction manual carefully and practice the test. Maintenance and calibration should be performed by experts. Test should be performed after 20 minutes after eating, drinking or smoking, if failure to wait 20 minutes may result in invalid results or incorrect results. Residual alcohol in the mouth can distort the measurement. Aromatic drinks or mouth sprays containing alcohol can interfere with measurements. In this case, wait for 20 minutes or rinse the mouth with pure water. Do not permit to blow cigarette smoke into the tester which may damage the sensor. Do not drink and drive, even small amount of alcohol can result in driving impairment. Recommended to use calibration service every 12 months or after 500 times tests. Do not use this device as a tool to determine "drink and drive". The manufacturer and the sellers will not take any legal responsibility for the test results. When not used for a long time, keep the device after taking out the batteries. In case of testing after long-term storage, there might be high tolerance of test results for the first 3 tests. Replace the batteries when the battery indicator icon reaches empty for the accuracy reason. In case the test result is higher than 0.1% BAC (1.0 g/l) and you try again without interval, then warm up time might be delayed. Perform a test within the operating temperature, which is specified in the instruction manual. Thank you for purchasing our breathalyzer, ALCOFIND AF-50 which is a personal & professional use alcohol tester with extremely high-reliability. It can be used for in the hospital, school and many other industrial fields. It measures estimated B.A.C(Blood Alcohol Content) in human breath by means of a fuel cell sensor which is high durable sensors and providing accurate breath alcohol results. Important safety information Introduction
  • 3. Preparation Performing the measurement <Front> <Back> Install two AA batteries in the battery compartment by paying attention to the battery polarity. Insert a mouthpiece into the port on the side of the alcohol tester. Ensure whether it is securely fit or not. A new mouthpieces is to be used for every person tested. Two AA batteries (1.5V Alkaline) Press the power button The total number of test performed will be displayed briefly and start countdown. When the countdown reaches zero with a double beep sound, three dashes ( ) will be displayed and "START" will flash. Take a deep breath and blow the mouthpiece for five seconds until there is a double beep sound. Digital Breathalyzer ALCOFIND AF-50 Two AAsize Batteries 6 (six) Mouthpieces Carrying Case User's Manual Contents 1. Mouthpiece 2. Mouthpiece slot 3. Display Name of Each Part 4. Power & Start button 5. Mode button 6. Battery cover Press and pull up the cover to open
  • 4. Wait at least 20 minutes after eating, drinking or smoking before measurement. After a few seconds(depending on the alcohol level), the estimated BAC(Blood Alcohol Concentration) will be displayed. The value will flash for several seconds. If you want to test again, pressthe start button to restart the countdown cycle. The tester automatically will be turned offafter 15~20 seconds. When the sampling was successful, LCD outer lines will be turning along the edge of LCD as you can see the picture. User Settings & Memory User may adjust UNIT, BEEP ALARM, BLOW TIME, FLOW DETECTION through Menu mode. Turn on the tester which may begin a countdown. During the countdown, press and hold the mode button for more than 5 seconds, the tester may enter Menu mode & Memory mode Adjust measurement units When "Unit" appears on the display, press the start button to adjust the measurement unit. Adjust beep alarm setting When "bEEp" appears on the display, press the start button to set the beep alarm. If user set the beep alarm at 0.05%BAC, the tester will sound an audible warning alarm when the test result is over 0.05%BAC. If user set "none" the tester will not sound at all. Recommended to set the value at your country's BAC Limit. Adjust blow time When "bL-t" appears on the display, press the start button to adjust the blow time. A longer blow time may provide a deep lung air which may provide more accurate reading. Adjust flow detection setting When "FL-d" appears on the display, press the start button to adjust the flow detection. Use this option to adjust how hard a user is required to blow. LO Low level detection, this is default setting. HI High level flow detection. This level requires a stronger breath. Recommend if previous test attempts have been unsuccessful. USER User activated test mode. When testing in this mode, the user must press the start button during blowing to obtain breath sample and perform a test successfully. Enter the Memory Mode When "DAT" appears on the display, press the start button, the most recent test result will appear. By pressing start button, next test result will appear. The tester will recall up to 50 test results. If the memory is full, the oldest test result is deleted automatically.
  • 5. Replace new batteries. Replace new batteries immediately. Press button again to repeat the measurement. Press button to restart the countdown cycle and retest. Repeat the test after the unit has reached a temperature within the operation range. Display message & Problem-solving guide MESSAGE WHAT ITMEANS ACTION Full battery Empty battery : Warns when battery is low. The unit can be operated but recommended to change new batteries for correct result! If the LCD displays "BAt", the unit cannot operate. If the sampling volume is too low or not sufficiently blow. If there is no breath sample blown within 30 seconds, the LCD will display "OUT". If the unit is powered on beyond the acceptable temperature range( ) the unit will display it. 5~40 C or 41~104 F Full battery Technical specifications Indication of B.A.C. Accuracy Warm up time Sensor Power supply Battery Life Dimensions(WxHxD) Weight Calibration Operating temperature 0.000 ~ 0.500%BAC 0.00 ~ 5.00 0.00 ~ 2.50mg/l +/- 0.005%BAC at 0.1%BAC at 25 C +/- 0.05 at 1.0 at 25 C +/- 0.025mg/l at 0.50mg/l at 25 C Within 15 seconds at 0.50g/l( ) warm up time may vary depending on the BAC measured. Professional Fuel cell sensor Two AAsize alkaline batteries Approx. 1500 tests 55mm x 125mm x 20mm 123g including batteries every 12 months or after 500 tests 5 ~ 40 C The specifications are subject to change without prior notice for functional improvement.
  • 6. Warranty Customer service Distributor : For any questions, after-sales service, product calibration or to purchase additional mouthpieces, please contact nearest distributor. The manufacturer warrants the product to be free from defects in workmanship or material (excluding calibration) under normal use either for 1(one) year from the date of purchase or 500 breath tests whichever comes earlier. The manufacturer's obligation under this warranty is limited to replacing, adjusting or repairing the unit at manufacturer's option, if returned along with proof of purchase. For any defects due to natural disaster or customer's negligence, misuse, reconstruction and repair by unauthorized person, reasonable cost shall be charged for repair or part replacement.