In-Situ UV-Visible and Raman Spectroscopy for Gas-Liquid-Solid Systems
C. P. Stemmet, J. C. Schouten, and T. A. Nijhuis*
Laboratory of Chemical Reactor Engineering, Department of Chemical Engineering and Chemistry, EindhoVen
UniVersity of Technology, P.O. Box 513, 5600 MB EindhoVen, The Netherlands
This paper presents the use of UV-vis and Raman spectroscopy to measure the state of a solid in a multiphase
reactor. A slurry of particles and a packed bubble column were used. As this study is a proof of principle a
model for an active catalyst system, insoluble pH indicators deposited on the particles and on a solid foam
packing (used as the packing material in the packed bubble column) are used. This insoluble pH indicator
has different UV-vis and Raman spectra according to the liquid pH. The experimental results indicate that
for the slurry the spectra obtained from the moving particles can be used to characterize the state of the pH
indicator. Bubbles in the packed bubble column result in disturbances in the UV-vis and Raman spectra
collected from the pH indicator adsorbed to the solid surface and this interference is removed successfully
with a newly developed tolerance-and-averaging method.
Introduction
Approximately 85% of all chemical products (bulk and fine
chemicals) are produced using a heterogeneous catalyst to
increase the reaction rate and hence increase production capac-
ity.1
Most often the catalyst is deposited on a solid support to
aid its separation from the reaction products and examples of
such reactors include packed beds, fluidized beds, and slurry
reactors. Multiphase reactors, where gas and liquid reactants
are converted over a heterogeneous catalyst, have been used
widely in the chemical industry and examples of such reactors
are slurry bubble columns, trickle bed reactors, and packed
bubble columns. These multiphase reactors are used in a variety
of applications ranging from the production of pharmaceuticals
and hydrogenations and oxidations to produce bulk chemicals
(hydro-treating, desulphurization, oxidation of sulfur dioxide,
etc.).
In-situ spectroscopic techniques have been developed over
the years to provide much understanding into the mechanisms
of catalytic reactions. In-situ spectroscopic techniques using
spectroscopic probes have been applied to control gas-solid
reactors operating under reaction conditions based on the state
of the solid catalyst.2,3
The key advantage of control based on
spectroscopic information is that the actual catalytic site (catalyst
state) is being probed directly, rather than the end-of-pipe
response at the reactor exit. In multiphase flows an optical probe
has been used to evaluate the local volume fraction of each
phase in an oil-water-gas mixture,4
however, spectroscopic
techniques in characterizing catalysts operating in multiphase
reactors (gas-liquid-solid) have not been investigated. It is
the intention of this study to assess whether UV-vis and Raman
spectroscopy can be used to determine the state of a solid
catalyst in a multiphase reactor system. In a chemical reactor
changes to a catalyst are usually slow, as good catalyst stability
is very important for an industrial process. In this study we used
a model system of which the state can be easily be manipulated
which makes the model development more facile. Insoluble pH
indicators were adsorbed on solid surfaces in a slurry of particles
and in a packed bubble column. These adsorbed pH indicators
are used as a model system for the catalyst in a multiphase
reactor.
In-Situ Spectroscopy. Spectroscopic measurements (ex situ
or in situ) have been widely used in the development and
determination of the main parameters in the preparation affecting
the stability and the size of catalytically active sites and the
intrinsic reaction rate. These techniques include ultraviolet-visible
spectroscopy (UV-vis), infrared spectroscopy (IR), Raman
spectroscopy, X-ray absorption near edge spectroscopy (XANES),
and X-ray absorption fine structure (XAFS) spectroscopy. The
main techniques and their limitations in catalyst characterization
are given in the books edited by Weckhuysen,1
and Haw,5
as
well as the review papers by Weckhuysen,6
and Hunger and
Weitkamp.7
The use of in situ spectroscopy has seen significant
increase, largely due to the development of optical probes which
may be used under high temperature and high pressure reaction
conditions.
UV-vis and Raman spectroscopy can give quantitative
information regarding the oxidation state and co-ordination of
the metal oxide complexes serving as catalytically active species.
The main uses of oxides of transition metals are for a number
of chemical conversions such as hydrogenation, oxidation,
decomposition, and isomerization.8
A case study has been
presented of a fixed bed where UV-vis spectroscopy (diffuse
reflectance spectroscopy) and Raman spectroscopy have been
used to describe changes in the Cr/Al2O3 active catalyst species
by increases observed in specific peaks.9
These techniques have
also been used to track the formation of coke to maintain an
active catalyst and using this information to control the
regeneration.2
Combining spectroscopic techniques provides
complementary quantification of the state of the catalyst and
the changes occurring under reaction conditions.10
However,
these studies have all focused on fixed beds although the Raman
spectral information from moving particles in a fluidized bed
may be an attractive measure for catalysts and adsorbates to
monitor and avoid the formation of coke.11
An example using
spectroscopy techniques to characterize the state of a slurry
catalyst (Cr/SiO2) for the polymerization of ethylene, using
toluene as a solvent has been illustrated.12
Here the Raman
spectra of toluene dominate the spectra collected from the
reaction mixture and thus obscure the spectra from the catalytic
active material.
UV-vis spectroscopy using visible light gives information
regarding the electronic d-d and charge transfer transitions of
metal oxides. It provides a convenient method to probe the state
* To whom correspondence should be addressed. E-mail:
T.A.Nijhuis@tue.nl.
Ind. Eng. Chem. Res. XXXX, xxx, 000 A
10.1021/ie801984w CCC: $40.75  XXXX American Chemical Society
of the active metal oxide catalyst or an adsorbed pH indicator.
By the application of small optical probes, measurement is
relatively easy and may also be performed through the outer
wall of the reactor if made of glass.9
Raman spectroscopy using
laser light provides information regarding the vibrational spectra
of metal oxides and organic deposits, and optical probes based
on Raman spectroscopy are becoming readily available.
Multiphase Reactors. The added effect of gas-liquid and
liquid-solid interfaces in gas-liquid-solid reactors contributes
to the spectroscopic information collected and thus poses a
challenge in obtaining reliable spectroscopic information from
the catalyst deposited on the solid surface. Optical probes use
optical fibers to transmit light (from a laser or lamp) into the
reactor and another in the same fiber bundle to measure the
reflected light (connected to a spectrometer). The interfer-
ence of the liquid-solid and the gas-liquid interfaces with the
light reflected from the catalyst on the solid surface is described
schematically for slurry reactors and packed bubble columns
in Figures 1a and 1b, respectively. These interferences need to
be removed from the spectroscopic information collected to
successfully determine the state of the catalyst.
Model System. The aim is to develop a methodology for
characterization of the solid catalyst in a multiphase system.
Since transitions in the state of a real catalyst are usually very
slow and occur at high temperature or pressure, a model system
is used as a more convenient and controllable way to demon-
strate the concept. An insoluble pH indicator adsorbed on the
surface of a solid support is used in this study to simulate an
active catalyst to determine UV-vis and Raman spectra in
multiphase systems with relative ease. The pH indicator exhibits
different UV-vis and Raman spectra depending on the pH of
the liquid with which it is in contact. This system thus provides
a method of studying spectroscopic measurements at mild
conditions (ambient temperature and pressure). The spectra
collected are only from the surface of the solid particles or solid
support and the contribution of the liquid to the spectra is low
due the low solubility of the pH indicator. The characteristic
peaks in the UV-vis and Raman are specific to the current
system. The origin of these characteristic peaks will not be
discussed in detail here, since this paper focuses on the
description of a methodology to do spectroscopy on a (moving)
catalyst in a multiphase reactor system. The chemical origin of
the bands/vibrations are less relevant and the discussion thereof
is beyond the scope of this paper.
Experimental Details
The UV-vis experiments were performed using a reflection
probe consisting of a bundle of seven optical fibers (each 200
µm thick) encased in a steel tube (120 mm in length and 6 mm
in diameter). The central optical fiber is connected to a UV-vis
spectrometer (AvaSpec-2048-USB2, Avantes) and the remaining
six fibers are used for illumination and are connected to a light
source (AvaLight-D(H)-S Deuterium-Halogen light source,
215-2000 nm, Avantes). The calibration of the reflectance
spectrometer is performed using a white reference tile (WS-2,
Avantes). The UV-vis spectra were recorded in LabView
(LabView 7.0, National Instruments) at an integration time of
30 ms to decrease the noise to peak ratio.
The Raman experiments were performed using a reflection
probe consisting of a bundle of two optical fibers (200 µm
Figure 1. Schematic representation during operating conditions in (a) a
slurry reactor and (b) a packed bubble column, showing the interferences
at the gas-liquid and liquid-solid interfaces that scatter the light away
from the optical probe.
Figure 2. Images of a 40 ppi solid foam covered in TiO2 and insoluble pH indicator, bromocresol green, showing the acidity of the liquid medium, (a) acidic
and (b) basic. (c) Structure of bromocresol green in the acidic and basic states.
B Ind. Eng. Chem. Res., Vol. xxx, No. xx, XXXX
connected to spectrometer and 100 µm connected to the laser
source) encased in a steel tube (76 mm in length with 12.7 mm
diameter). This probe is connected to an integrated system
combining a laser source (532 nm) and a Raman spectrometer
(R-3000, Ocean Optics). The Raman spectra were recorded in
LabView (LabView 7.0, National Instruments) at integration
times between 1 and 5 s.
All chemicals used were laboratory grade supplied from
Merck unless specified otherwise. The pH of the liquid was
decreased using a solution of HCl (0.6 M) which was dosed
using a titrator (TIM854, Radiometer) at a rate of 0.2-5 mL/
min. The titrator also allowed measurement of the pH of the
liquid using a pH-electrode (pHC2001, Radiometer). The pH
indicator, bromocresol green (pH range 3.8-5.4), was dried onto
the solid particles (γ-Al2O3, 63-150 µm) or solid foam support
(40 ppi, ERG. Aerospace Corp.) from solution. This pH indicator
is usually dissolved in a liquid (a mixture of water and ethanol)
and exhibits different UV-vis and Raman spectra depending
on the liquid pH.
Slurry of Particles. Alumina particles (63-150 µm) covered
with bromocresol green were prepared using 5 wt % ethanol as
a solvent and dried in an oven at 90 °C from solution at a liquid
pH of approximately 5. The particles were filtered and washed
with distilled water until the filtrate remained clear. The pH
indicator has a low solubility in water and thus remains on the
surface of the particles and responds in the UV-vis and Raman
spectrum depending on the pH of the liquid surrounding the
particles.
Packed Bubble Column. Aluminum foam pieces (40 ppi,
ERG. Aerospace Corp.) were used as the packing material in
the packed bubble column experiments. Solid foams are
convenient to use due to the open structure and high surface
area for deposition of the pH indicator.13
The aluminum foam
was dipped in a slurry of TiO2 (17 wt %, P25, Degussa) and
dried while rotating at 60 rpm. Air was blown through the
samples to avoid pores of the solid foam material remaining
filled with the slurry. The solid foam pieces were calcined at
400 °C resulting in washcoats of approximately 7-10 wt %.
The bromocresol green pH indicator was applied to the surface
of the solid foam structure using the same preparation technique
described for the slurry particles, resulting in solid foams which
respond according to pH of the liquid medium, see Figure 2.
The bubble column used for the experiments is a cylindrical
glass vessel (diameter of 10 cm) where nitrogen is distributed
through a fritted glass disk from the bottom. The solid foam
was placed on the distributor and the UV-vis and Raman probes
focused on the solid foam packing from the top (approximately
2-5 mm focal distance). As gas bubbles travel through the solid
foam they interfere with the signal collected by the UV-vis
and Raman probes. Bubble interference was removed with an
in-house Labview application designed to discard compromised
UV-vis and Raman spectra. This method of tolerance-and-
averaging is described schematically in Figure 3. An average
spectrum (AS) is created over m (e.g., 10) number of spectra.
In the next cycle, the newly recorded spectrum (NS) is compared
to this averaged spectrum (ASold) for the UV-vis spectra over
wavelengths 300-800 nm and Raman spectra over Raman shifts
800-4300 cm-1
. The measurement is discarded if the sum of
the absolute difference between the spectra is larger than a set
tolerance value. This is repeated for the next nine (m - 1)
measurements. If, for example, three spectra were discarded, a
new average spectrum (ASnew) is created from the remaining
seven (n) spectra and used for comparison in the next cycle.
This procedure is repeated, producing a new average spectrum
for every 10 measurements. A spectrum averaged over a too
large number of spectra (e.g., m equal to 100) will have less
noise, but also rapid changes in the state of the pH indicator
are possibly evaluated as the interference of a bubble and thus
could be lost.
Results and Discussion
Slurry of Particles. A slurry of 30 wt % alumina particles
with the bromocresol green pH indicator adsorbed to the
particles’ surface was used to evaluate if UV-vis and Raman
spectroscopy could be used to determine changes in the state
of the pH indicator. UV-vis spectra were recorded every 200
ms while the pH of the liquid was decreased from 4 to 1 over
a time period of 300 s. The UV-vis spectra of the slurry of
bromocresol green particles are given in Figure 4 for a select
number of liquid pH values. The reflectance at 620 and 450
nm increases and decreases, respectively, as the pH is decreased.
Raman spectra were recorded every 15 s (integration time 5 s),
while the pH of the liquid was decreased from 4 to 1 over a
time period of 300 s. The Raman spectra of the slurry of
bromocresol green particles are given in Figure 5. Raman
intensities are observed to increase (at approximately 833, 1570,
2210, 2680, 3040, and 4000 cm-1
) as a result of the bromocresol
green changing from the basic to the acidic state with the
decrease in the liquid pH (addition of H+
) (see Figure 2c for
the structural changes). Thus, the UV-vis and Raman spectra
Figure 3. Schematic representation of the tolerance-and-averaging method
to remove bubble interference from UV-vis spectra. New spectra (NS)
from the spectrometer are compared to an average spectrum (AS) determined
over m spectra over the previous cycle. The sum of the absolute difference
is compared to a tolerance value (Tol) and if larger, the spectrum is
discarded. A new average is created from the n number of spectra remaining
and is used in the next cycle.
Figure 4. Observed UV-vis reflectance spectra for a 30% slurry of particles
covered with bromocresol green, while the pH of the liquid is decreased.
Ind. Eng. Chem. Res., Vol. xxx, No. xx, XXXX C
can be obtained from moving particles, and transients from one
state to another can be observed.
Packed Bubble Column. The experiments for the UV-vis
and Raman spectra resulting from the adsorbed bromocresol
green pH indicator operating in a packed bubble column are
shown in Figure 6-Figure 8. Fifty UV-vis spectra are recorded
every 2 s and the tolerance-and-averaging method is used to
assess the validity of the recorded spectra. The Raman spectra
measurements are performed every 2 s and 10 are used for the
tolerance-and-averaging method. In Figure 6b the average
recorded UV-vis spectra for the wavelength 620 nm is given
while the liquid pH is decreased, see Figure 6c. As the bubble
flow starts and bubbles interfere with the UV-vis spectra and
Raman spectra obtained, the tolerance-and-averaging method
maintains the average spectral measurement while discarding
spectra compromised by bubbles. The average UV-vis spectra
determined using the tolerance-and-averaging method shows less
noise in the recorded spectra at 620 nm. The relative difference
between the average reflectance determined using all the spectra
and the average reflectance determined using the tolerance-and-
averaging method can be as high as 30-40% (see Figure 6d).
An increase in the UV-vis reflectance is observed as the liquid
pH is decreased (see Figure 4). Similarly, the changing Raman
intensity recorded at a Raman shift of 4000 cm-1
as a result of
decreasing liqiud pH is presented in Figure 7. A figure created
in a similar manner for one of the other characteristic bands
visible in the Raman spectrum for the slurry particles (at
approximately 833, 1570, 2210, 2680, 3040 cm-1
) yielded a
graph with a comparable trend. The average Raman intensity
increases as the pH of the liquid is decreased over time. As
much as 80% of spectra may be discarded using the tolerance-
and-averaging method resulting in an averaged Raman spectrum
Figure 5. (a) Observed Raman spectra for a 30 wt % slurry of particles covered with bromocresol green, while the pH of the liquid is decreased over time.
(b) Liquid pH-profile over time.
Figure 6. The UV-vis spectra obtained for the bromocresol green pH indicator deposited on the surface of a solid foam packing operating in a packed
bubble column. (a) Reflectance at a wavelength of 620 nm is presented. The spectra which have been discarded for averaging are given by the open symbols
(O) and used for determining the average in closed symbols (b). (b) The average reflectance of all data points (---) and the reflectance determined by the
tolerance-and-averaging method (s) are given. (c) Liquid pH-profile over time. (d) Relative error between the average for all the data points and the average
determined by the tolerance-and-averaging method.
D Ind. Eng. Chem. Res., Vol. xxx, No. xx, XXXX
with less noise when compared when all measurement are used.
In Figure 8 the averaged Raman spectra is given for a decrease
in the liquid pH over time. Comparing Figure 8 and Figure 5
shows that similar Raman intensities are observed to increase
as the bromocresol green pH indicator makes a transition to its
acidic form. The differences in the observed Raman spectra for
the slurry of particles and packed bubble column is due the pH
indicator being adsorbed onto two different materials, Al2O3
and TiO2, respectively.
Conclusions
UV-vis and Raman spectra recorded for a slurry of particles,
and a solid foam packed bubble column are shown for insoluble
pH indicators adsorbed onto the solid surface of the particles
and the solid foam packing. A change in the adsorbed pH
indicator is accomplished by a change in the liquid pH and
results in a clear change in the UV-vis and Raman spectra.
These examples indicate that, although the UV-vis and Raman
spectra are recorded under multiphase conditions, changes in
the solid surface conditions can be observed. In a packed bubble
column the averaging-and-tolerance method can be used to
remove spectra that have been compromised by gas-liquid
interfaces. The results from this study indicate that in situ
UV-vis and Raman spectroscopy can be used to characterize
a solid surface operating under multiphase conditions where
bubbles interfere with the measured spectra.
Acknowledgment
The authors gratefully acknowledge the Dutch Technology
Foundation STW (STW-NWO Vidi Project 06186) for the
financial support. Dow Benelux B.V., BASF Nederland B.V.,
Avantium Technologies, and Shell Global Solutions Interna-
tional B.V. are acknowledged for their input during the course
of the project.
Literature Cited
(1) Weckhuysen, B. M. In-Situ Spectroscopy of Catalysts; American
Scientific Publishers: Stevenson Ranch, CA, 2004.
(2) Bennici, S. M.; Vogelaar, B. M.; Nijhuis, T. A.; Weckhuysen, B. M.
Real-time control of a catalytic solid in a fixed-bed reactor based on in-situ
spectroscopy. Angew Chem., Int. Ed. 2007, 46, 5412–5416.
(3) Nijhuis, T. A.; Tinnemans, S. J.; Visser, T.; Weckhuysen, B. M.
Operando spectroscopic investigation of supported metal oxide catalysts
by combined time-resolved UV-VIS/Raman/on-line mass spectrometry.
Phys. Chem. Chem. Phys. 2003, 5, 4361–4365.
(4) Ramos, R. T.; Holmes, A.; Wu, X.; Dussan, E. A local optical probe
using fluorescence and reflectance for measurment of volume fractions in
multi-phase flows. Meas. Sci. Technol. 2001, 12, 871–876.
Figure 7. The Raman spectra obtained for the bromocresol green pH indicator deposited on the surface of a solid foam packing operating in a packed bubble
column. (a) Raman intensity at a Raman shift of 4000 cm-
1 is presented. The spectra which have been discarded for averaging are given by the open
symbols (O) and used for determining the average in closed symbols (b). (b) The average Raman spectra of all data points (---) and the average determined
by the tolerance-and-averaging method (s) are given. (c) Liquid pH-profile over time.
Figure 8. The average Raman spectra obtained for bromocresol green pH
indicator deposited on the surface of a solid foam packing operating in a
packed bubble column as the liquid pH is decreased over time. Peaks in
the Raman spectra are observed as the pH indicator undergoes transition to
its acidic form.
Ind. Eng. Chem. Res., Vol. xxx, No. xx, XXXX E
(5) Haw, J. F. In In-Situ Spectroscopy in Heterogeneous Catalysis;
Wiley-VCH: Weinheim, Germany, 2002.
(6) Weckhuysen, B. M. Snapshots of a working catalyst: possibilities
and limitations of in situ spectroscopy in the field of heterogeneous catalysis.
Chem. Commun. 2002, 97–110.
(7) Hunger, M.; Weitkamp, J. In situ IR, NMR, EPR, and UV-vis
Spectroscopy: Tools for New Insight into the Mechanisms of Heterogeneous
Catalysis. Angew Chem., Int. Ed. 2001, 40, 2954–2971.
(8) Weckhuysen, B. M.; Wachs, I. E. Catalysis by supported metal
oxides. In Handbook of Surface Interfaces of Materials; Academic Press:
San Diego, CA, 2001; Vol. 1, Chapter 11, pp 613-648.
(9) Nijhuis, T. A.; Tinnemans, S. J.; Visser, T.; Weckhuysen, B. M.
Towards real-time spectroscopic process control for the dehydrogenation
of propane over supported chromium oxide catalysts. Chem. Eng. Sci. 2004,
59, 5487–5492.
(10) Tinnemans, S. J.; Mesu, J. G.; Kervinen, K.; Visser, T.; Nijhuis,
T. A.; Beale, A. M.; Keller, D. E.; van der Eerden, A. M. J.; Weckhuysen,
B. M. Combining operando techniques in one spectroscopic-reaction cell:
oppertunities for elucidating the active site and related reaction mechanism
in catalysis. Catal. Today 2006, 113, 3–15.
(11) Chua, Y. T.; Stair, P. C. A novel fluidized bed technique for
measureing UV Raman spectra of catalysts and adsorbates. J. Catal. 2000,
196, 66–72.
(12) Weckhuysen, B. M. Determining the active site in a catalytic
process: Operando spectroscopy is more than a buzzword. Phys. Chem.
Chem. Phys. 2003, 5, 4351–4360.
(13) Stemmet, C. P.; Meeuwse, M.; van der Schaaf, J.; Kuster, B. F. M.;
Schouten, J. C. Gas-liquid mass transfer and axial dispersion in solid foam
packings. Chem. Eng. Sci. 2007, 62, 5444–5450.
ReceiVed for reView December 23, 2008
ReVised manuscript receiVed February 23, 2009
Accepted February 26, 2009
IE801984W
F Ind. Eng. Chem. Res., Vol. xxx, No. xx, XXXX

More Related Content

What's hot

Gc ms(mr.s)
Gc ms(mr.s)Gc ms(mr.s)
Gc ms(mr.s)22suresh
 
Lcms gcms and its applications
Lcms gcms and its applicationsLcms gcms and its applications
Lcms gcms and its applicationsNihal Calicut
 
Gas chromatography mass spectrometry
Gas chromatography mass spectrometryGas chromatography mass spectrometry
Gas chromatography mass spectrometryHina Qaiser
 
Use of gc and gcms in api analysis
Use of gc and gcms in api analysis Use of gc and gcms in api analysis
Use of gc and gcms in api analysis Dipak Shetty
 
Detector used in hplc chromatogrphy by pradeep jaiswal msc part1 mithibai col...
Detector used in hplc chromatogrphy by pradeep jaiswal msc part1 mithibai col...Detector used in hplc chromatogrphy by pradeep jaiswal msc part1 mithibai col...
Detector used in hplc chromatogrphy by pradeep jaiswal msc part1 mithibai col...Pradeep Jaiswal
 
Hyphenated techniques- GCMS, LCMS, GCIR
Hyphenated techniques- GCMS, LCMS, GCIRHyphenated techniques- GCMS, LCMS, GCIR
Hyphenated techniques- GCMS, LCMS, GCIRSanket Shinde
 
Gas chromatography mass spectrometry ppt
Gas chromatography mass spectrometry pptGas chromatography mass spectrometry ppt
Gas chromatography mass spectrometry pptArunsinghlohhra
 
HPLC - High Performance Liquid Chromatography
HPLC - High Performance Liquid ChromatographyHPLC - High Performance Liquid Chromatography
HPLC - High Performance Liquid ChromatographyDivya Basuti
 
GAS CHROMATOGRAPHY AND MASS SPECTROMETRY (GC-MS) BY P.RAVISANKAR.
GAS CHROMATOGRAPHY AND MASS SPECTROMETRY (GC-MS) BY P.RAVISANKAR.GAS CHROMATOGRAPHY AND MASS SPECTROMETRY (GC-MS) BY P.RAVISANKAR.
GAS CHROMATOGRAPHY AND MASS SPECTROMETRY (GC-MS) BY P.RAVISANKAR.Dr. Ravi Sankar
 
Detector of HPLC
Detector of HPLCDetector of HPLC
Detector of HPLCkhaterehz
 
Detectors hplc (1)
Detectors hplc (1)Detectors hplc (1)
Detectors hplc (1)ceutics1315
 

What's hot (20)

Lc nmr
Lc nmrLc nmr
Lc nmr
 
Gc ms(mr.s)
Gc ms(mr.s)Gc ms(mr.s)
Gc ms(mr.s)
 
Triple Quadrupole Gas Chromatography-Mass Spectrometry/Mass Spectrometry Re-i...
Triple Quadrupole Gas Chromatography-Mass Spectrometry/Mass Spectrometry Re-i...Triple Quadrupole Gas Chromatography-Mass Spectrometry/Mass Spectrometry Re-i...
Triple Quadrupole Gas Chromatography-Mass Spectrometry/Mass Spectrometry Re-i...
 
Hplc
HplcHplc
Hplc
 
Lcms gcms and its applications
Lcms gcms and its applicationsLcms gcms and its applications
Lcms gcms and its applications
 
Hplc
HplcHplc
Hplc
 
Gas chromatography mass spectrometry
Gas chromatography mass spectrometryGas chromatography mass spectrometry
Gas chromatography mass spectrometry
 
Use of gc and gcms in api analysis
Use of gc and gcms in api analysis Use of gc and gcms in api analysis
Use of gc and gcms in api analysis
 
Detector used in hplc chromatogrphy by pradeep jaiswal msc part1 mithibai col...
Detector used in hplc chromatogrphy by pradeep jaiswal msc part1 mithibai col...Detector used in hplc chromatogrphy by pradeep jaiswal msc part1 mithibai col...
Detector used in hplc chromatogrphy by pradeep jaiswal msc part1 mithibai col...
 
GCMS
GCMSGCMS
GCMS
 
Gc ms
Gc msGc ms
Gc ms
 
GC-MS
GC-MSGC-MS
GC-MS
 
Gas Chromatography and Mass Spectrometry-GCMS
Gas Chromatography and Mass Spectrometry-GCMSGas Chromatography and Mass Spectrometry-GCMS
Gas Chromatography and Mass Spectrometry-GCMS
 
Hyphenated techniques- GCMS, LCMS, GCIR
Hyphenated techniques- GCMS, LCMS, GCIRHyphenated techniques- GCMS, LCMS, GCIR
Hyphenated techniques- GCMS, LCMS, GCIR
 
Gas chromatography mass spectrometry ppt
Gas chromatography mass spectrometry pptGas chromatography mass spectrometry ppt
Gas chromatography mass spectrometry ppt
 
HPLC - High Performance Liquid Chromatography
HPLC - High Performance Liquid ChromatographyHPLC - High Performance Liquid Chromatography
HPLC - High Performance Liquid Chromatography
 
GCMS
GCMSGCMS
GCMS
 
GAS CHROMATOGRAPHY AND MASS SPECTROMETRY (GC-MS) BY P.RAVISANKAR.
GAS CHROMATOGRAPHY AND MASS SPECTROMETRY (GC-MS) BY P.RAVISANKAR.GAS CHROMATOGRAPHY AND MASS SPECTROMETRY (GC-MS) BY P.RAVISANKAR.
GAS CHROMATOGRAPHY AND MASS SPECTROMETRY (GC-MS) BY P.RAVISANKAR.
 
Detector of HPLC
Detector of HPLCDetector of HPLC
Detector of HPLC
 
Detectors hplc (1)
Detectors hplc (1)Detectors hplc (1)
Detectors hplc (1)
 

Similar to Cat 1

Chromatography BY Hamunyare Ndwabe
Chromatography BY Hamunyare NdwabeChromatography BY Hamunyare Ndwabe
Chromatography BY Hamunyare NdwabeHamunyare Ndwabe
 
Laser Ablation Molecular Isotopic Spectrometry for rare isotopes of the light...
Laser Ablation Molecular Isotopic Spectrometry for rare isotopes of the light...Laser Ablation Molecular Isotopic Spectrometry for rare isotopes of the light...
Laser Ablation Molecular Isotopic Spectrometry for rare isotopes of the light...Alexander Bolshakov
 
Laser ablation - optical cavity isotopic spectrometer (LAOCIS) for Mars rovers
Laser ablation - optical cavity isotopic spectrometer (LAOCIS) for Mars roversLaser ablation - optical cavity isotopic spectrometer (LAOCIS) for Mars rovers
Laser ablation - optical cavity isotopic spectrometer (LAOCIS) for Mars roversAlexander Bolshakov
 
New approaches for stable isotope ratio measurements iaea tecdoc 1247
New approaches for stable isotope ratio measurements   iaea tecdoc 1247New approaches for stable isotope ratio measurements   iaea tecdoc 1247
New approaches for stable isotope ratio measurements iaea tecdoc 1247Mahbubul Hassan
 
High performance liquid chromatography
High performance liquid chromatographyHigh performance liquid chromatography
High performance liquid chromatographySTUTISRIVASTAVA27
 
Online Size-Exclusion High-Performance Liquid Chromatography Light Scattering...
Online Size-Exclusion High-Performance Liquid Chromatography Light Scattering...Online Size-Exclusion High-Performance Liquid Chromatography Light Scattering...
Online Size-Exclusion High-Performance Liquid Chromatography Light Scattering...KBI Biopharma
 
GC-MS, amp, lcms
GC-MS, amp, lcmsGC-MS, amp, lcms
GC-MS, amp, lcmsDr. Samia
 
Notes for The principle and performance of liquid chromatography–mass spectro...
Notes for The principle and performance of liquid chromatography–mass spectro...Notes for The principle and performance of liquid chromatography–mass spectro...
Notes for The principle and performance of liquid chromatography–mass spectro...improvemed
 
Analytical techniques ppt
Analytical techniques pptAnalytical techniques ppt
Analytical techniques pptNikita Dewangan
 
Uv visible-spectroscopy-6500 (3)
Uv visible-spectroscopy-6500 (3)Uv visible-spectroscopy-6500 (3)
Uv visible-spectroscopy-6500 (3)PromilaThakur4
 

Similar to Cat 1 (20)

Lc ms
Lc msLc ms
Lc ms
 
Cat 2
Cat 2Cat 2
Cat 2
 
Chromatography BY Hamunyare Ndwabe
Chromatography BY Hamunyare NdwabeChromatography BY Hamunyare Ndwabe
Chromatography BY Hamunyare Ndwabe
 
Laser Ablation Molecular Isotopic Spectrometry for rare isotopes of the light...
Laser Ablation Molecular Isotopic Spectrometry for rare isotopes of the light...Laser Ablation Molecular Isotopic Spectrometry for rare isotopes of the light...
Laser Ablation Molecular Isotopic Spectrometry for rare isotopes of the light...
 
Laser ablation - optical cavity isotopic spectrometer (LAOCIS) for Mars rovers
Laser ablation - optical cavity isotopic spectrometer (LAOCIS) for Mars roversLaser ablation - optical cavity isotopic spectrometer (LAOCIS) for Mars rovers
Laser ablation - optical cavity isotopic spectrometer (LAOCIS) for Mars rovers
 
New approaches for stable isotope ratio measurements iaea tecdoc 1247
New approaches for stable isotope ratio measurements   iaea tecdoc 1247New approaches for stable isotope ratio measurements   iaea tecdoc 1247
New approaches for stable isotope ratio measurements iaea tecdoc 1247
 
High performance liquid chromatography
High performance liquid chromatographyHigh performance liquid chromatography
High performance liquid chromatography
 
High Performance Liquid Chromatography
High Performance Liquid ChromatographyHigh Performance Liquid Chromatography
High Performance Liquid Chromatography
 
Online Size-Exclusion High-Performance Liquid Chromatography Light Scattering...
Online Size-Exclusion High-Performance Liquid Chromatography Light Scattering...Online Size-Exclusion High-Performance Liquid Chromatography Light Scattering...
Online Size-Exclusion High-Performance Liquid Chromatography Light Scattering...
 
Unit I Absorption Spectroscopy.pdf
Unit I Absorption Spectroscopy.pdfUnit I Absorption Spectroscopy.pdf
Unit I Absorption Spectroscopy.pdf
 
Prabhakar singh ii sem-paper v-hplc, fplc, uplc, rrlc
Prabhakar singh  ii sem-paper v-hplc, fplc, uplc, rrlcPrabhakar singh  ii sem-paper v-hplc, fplc, uplc, rrlc
Prabhakar singh ii sem-paper v-hplc, fplc, uplc, rrlc
 
Hannah poster
Hannah posterHannah poster
Hannah poster
 
INSTRUMENTATION
INSTRUMENTATION INSTRUMENTATION
INSTRUMENTATION
 
LC-NMR_spectroscopy[1].pptx
LC-NMR_spectroscopy[1].pptxLC-NMR_spectroscopy[1].pptx
LC-NMR_spectroscopy[1].pptx
 
GC-MS, amp, lcms
GC-MS, amp, lcmsGC-MS, amp, lcms
GC-MS, amp, lcms
 
INFRARED(IR) SPECTROSCOPY OF NATURAL PRODUCTS
INFRARED(IR) SPECTROSCOPY OF NATURAL PRODUCTSINFRARED(IR) SPECTROSCOPY OF NATURAL PRODUCTS
INFRARED(IR) SPECTROSCOPY OF NATURAL PRODUCTS
 
Degradation Analysis Using LC-MS/MS
Degradation Analysis Using LC-MS/MSDegradation Analysis Using LC-MS/MS
Degradation Analysis Using LC-MS/MS
 
Notes for The principle and performance of liquid chromatography–mass spectro...
Notes for The principle and performance of liquid chromatography–mass spectro...Notes for The principle and performance of liquid chromatography–mass spectro...
Notes for The principle and performance of liquid chromatography–mass spectro...
 
Analytical techniques ppt
Analytical techniques pptAnalytical techniques ppt
Analytical techniques ppt
 
Uv visible-spectroscopy-6500 (3)
Uv visible-spectroscopy-6500 (3)Uv visible-spectroscopy-6500 (3)
Uv visible-spectroscopy-6500 (3)
 

Recently uploaded

VIP Call Girl Sector 25 Gurgaon Just Call Me 9899900591
VIP Call Girl Sector 25 Gurgaon Just Call Me 9899900591VIP Call Girl Sector 25 Gurgaon Just Call Me 9899900591
VIP Call Girl Sector 25 Gurgaon Just Call Me 9899900591adityaroy0215
 
❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...
❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...
❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...Gfnyt.com
 
College Call Girls Hyderabad Sakshi 9907093804 Independent Escort Service Hyd...
College Call Girls Hyderabad Sakshi 9907093804 Independent Escort Service Hyd...College Call Girls Hyderabad Sakshi 9907093804 Independent Escort Service Hyd...
College Call Girls Hyderabad Sakshi 9907093804 Independent Escort Service Hyd...delhimodelshub1
 
Call Girls LB Nagar 7001305949 all area service COD available Any Time
Call Girls LB Nagar 7001305949 all area service COD available Any TimeCall Girls LB Nagar 7001305949 all area service COD available Any Time
Call Girls LB Nagar 7001305949 all area service COD available Any Timedelhimodelshub1
 
💚😋Mumbai Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Mumbai Escort Service Call Girls, ₹5000 To 25K With AC💚😋💚😋Mumbai Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Mumbai Escort Service Call Girls, ₹5000 To 25K With AC💚😋Sheetaleventcompany
 
(Jessica) Call Girl in Jaipur- 9521753030 Escorts Service 50% Off with Cash O...
(Jessica) Call Girl in Jaipur- 9521753030 Escorts Service 50% Off with Cash O...(Jessica) Call Girl in Jaipur- 9521753030 Escorts Service 50% Off with Cash O...
(Jessica) Call Girl in Jaipur- 9521753030 Escorts Service 50% Off with Cash O...indiancallgirl4rent
 
Udaipur Call Girls 📲 9999965857 Call Girl in Udaipur
Udaipur Call Girls 📲 9999965857 Call Girl in UdaipurUdaipur Call Girls 📲 9999965857 Call Girl in Udaipur
Udaipur Call Girls 📲 9999965857 Call Girl in Udaipurseemahedar019
 
Vip sexy Call Girls Service In Sector 137,9999965857 Young Female Escorts Ser...
Vip sexy Call Girls Service In Sector 137,9999965857 Young Female Escorts Ser...Vip sexy Call Girls Service In Sector 137,9999965857 Young Female Escorts Ser...
Vip sexy Call Girls Service In Sector 137,9999965857 Young Female Escorts Ser...Call Girls Noida
 
Leading transformational change: inner and outer skills
Leading transformational change: inner and outer skillsLeading transformational change: inner and outer skills
Leading transformational change: inner and outer skillsHelenBevan4
 
Hot Call Girl In Ludhiana 👅🥵 9053'900678 Call Girls Service In Ludhiana
Hot  Call Girl In Ludhiana 👅🥵 9053'900678 Call Girls Service In LudhianaHot  Call Girl In Ludhiana 👅🥵 9053'900678 Call Girls Service In Ludhiana
Hot Call Girl In Ludhiana 👅🥵 9053'900678 Call Girls Service In LudhianaRussian Call Girls in Ludhiana
 
💚😋Chandigarh Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Chandigarh Escort Service Call Girls, ₹5000 To 25K With AC💚😋💚😋Chandigarh Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Chandigarh Escort Service Call Girls, ₹5000 To 25K With AC💚😋Sheetaleventcompany
 
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591adityaroy0215
 
indian Call Girl Panchkula ❤️🍑 9907093804 Low Rate Call Girls Ludhiana Tulsi
indian Call Girl Panchkula ❤️🍑 9907093804 Low Rate Call Girls Ludhiana Tulsiindian Call Girl Panchkula ❤️🍑 9907093804 Low Rate Call Girls Ludhiana Tulsi
indian Call Girl Panchkula ❤️🍑 9907093804 Low Rate Call Girls Ludhiana TulsiHigh Profile Call Girls Chandigarh Aarushi
 
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real MeetCall Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meetpriyashah722354
 
pOOJA sexy Call Girls In Sector 49,9999965857 Young Female Escorts Service In...
pOOJA sexy Call Girls In Sector 49,9999965857 Young Female Escorts Service In...pOOJA sexy Call Girls In Sector 49,9999965857 Young Female Escorts Service In...
pOOJA sexy Call Girls In Sector 49,9999965857 Young Female Escorts Service In...Call Girls Noida
 
hyderabad call girl.pdfRussian Call Girls in Hyderabad Amrita 9907093804 Inde...
hyderabad call girl.pdfRussian Call Girls in Hyderabad Amrita 9907093804 Inde...hyderabad call girl.pdfRussian Call Girls in Hyderabad Amrita 9907093804 Inde...
hyderabad call girl.pdfRussian Call Girls in Hyderabad Amrita 9907093804 Inde...delhimodelshub1
 
Call Girl Price Amritsar ❤️🍑 9053900678 Call Girls in Amritsar Suman
Call Girl Price Amritsar ❤️🍑 9053900678 Call Girls in Amritsar SumanCall Girl Price Amritsar ❤️🍑 9053900678 Call Girls in Amritsar Suman
Call Girl Price Amritsar ❤️🍑 9053900678 Call Girls in Amritsar SumanCall Girls Service Chandigarh Ayushi
 

Recently uploaded (20)

VIP Call Girl Sector 25 Gurgaon Just Call Me 9899900591
VIP Call Girl Sector 25 Gurgaon Just Call Me 9899900591VIP Call Girl Sector 25 Gurgaon Just Call Me 9899900591
VIP Call Girl Sector 25 Gurgaon Just Call Me 9899900591
 
❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...
❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...
❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...
 
College Call Girls Hyderabad Sakshi 9907093804 Independent Escort Service Hyd...
College Call Girls Hyderabad Sakshi 9907093804 Independent Escort Service Hyd...College Call Girls Hyderabad Sakshi 9907093804 Independent Escort Service Hyd...
College Call Girls Hyderabad Sakshi 9907093804 Independent Escort Service Hyd...
 
Call Girls LB Nagar 7001305949 all area service COD available Any Time
Call Girls LB Nagar 7001305949 all area service COD available Any TimeCall Girls LB Nagar 7001305949 all area service COD available Any Time
Call Girls LB Nagar 7001305949 all area service COD available Any Time
 
Call Girl Guwahati Aashi 👉 7001305949 👈 🔝 Independent Escort Service Guwahati
Call Girl Guwahati Aashi 👉 7001305949 👈 🔝 Independent Escort Service GuwahatiCall Girl Guwahati Aashi 👉 7001305949 👈 🔝 Independent Escort Service Guwahati
Call Girl Guwahati Aashi 👉 7001305949 👈 🔝 Independent Escort Service Guwahati
 
💚😋Mumbai Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Mumbai Escort Service Call Girls, ₹5000 To 25K With AC💚😋💚😋Mumbai Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Mumbai Escort Service Call Girls, ₹5000 To 25K With AC💚😋
 
VIP Call Girls Lucknow Isha 🔝 9719455033 🔝 🎶 Independent Escort Service Lucknow
VIP Call Girls Lucknow Isha 🔝 9719455033 🔝 🎶 Independent Escort Service LucknowVIP Call Girls Lucknow Isha 🔝 9719455033 🔝 🎶 Independent Escort Service Lucknow
VIP Call Girls Lucknow Isha 🔝 9719455033 🔝 🎶 Independent Escort Service Lucknow
 
(Jessica) Call Girl in Jaipur- 9521753030 Escorts Service 50% Off with Cash O...
(Jessica) Call Girl in Jaipur- 9521753030 Escorts Service 50% Off with Cash O...(Jessica) Call Girl in Jaipur- 9521753030 Escorts Service 50% Off with Cash O...
(Jessica) Call Girl in Jaipur- 9521753030 Escorts Service 50% Off with Cash O...
 
Udaipur Call Girls 📲 9999965857 Call Girl in Udaipur
Udaipur Call Girls 📲 9999965857 Call Girl in UdaipurUdaipur Call Girls 📲 9999965857 Call Girl in Udaipur
Udaipur Call Girls 📲 9999965857 Call Girl in Udaipur
 
Vip sexy Call Girls Service In Sector 137,9999965857 Young Female Escorts Ser...
Vip sexy Call Girls Service In Sector 137,9999965857 Young Female Escorts Ser...Vip sexy Call Girls Service In Sector 137,9999965857 Young Female Escorts Ser...
Vip sexy Call Girls Service In Sector 137,9999965857 Young Female Escorts Ser...
 
Leading transformational change: inner and outer skills
Leading transformational change: inner and outer skillsLeading transformational change: inner and outer skills
Leading transformational change: inner and outer skills
 
Hot Call Girl In Ludhiana 👅🥵 9053'900678 Call Girls Service In Ludhiana
Hot  Call Girl In Ludhiana 👅🥵 9053'900678 Call Girls Service In LudhianaHot  Call Girl In Ludhiana 👅🥵 9053'900678 Call Girls Service In Ludhiana
Hot Call Girl In Ludhiana 👅🥵 9053'900678 Call Girls Service In Ludhiana
 
💚😋Chandigarh Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Chandigarh Escort Service Call Girls, ₹5000 To 25K With AC💚😋💚😋Chandigarh Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Chandigarh Escort Service Call Girls, ₹5000 To 25K With AC💚😋
 
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591
VIP Call Girl Sector 88 Gurgaon Delhi Just Call Me 9899900591
 
indian Call Girl Panchkula ❤️🍑 9907093804 Low Rate Call Girls Ludhiana Tulsi
indian Call Girl Panchkula ❤️🍑 9907093804 Low Rate Call Girls Ludhiana Tulsiindian Call Girl Panchkula ❤️🍑 9907093804 Low Rate Call Girls Ludhiana Tulsi
indian Call Girl Panchkula ❤️🍑 9907093804 Low Rate Call Girls Ludhiana Tulsi
 
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real MeetCall Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
 
pOOJA sexy Call Girls In Sector 49,9999965857 Young Female Escorts Service In...
pOOJA sexy Call Girls In Sector 49,9999965857 Young Female Escorts Service In...pOOJA sexy Call Girls In Sector 49,9999965857 Young Female Escorts Service In...
pOOJA sexy Call Girls In Sector 49,9999965857 Young Female Escorts Service In...
 
hyderabad call girl.pdfRussian Call Girls in Hyderabad Amrita 9907093804 Inde...
hyderabad call girl.pdfRussian Call Girls in Hyderabad Amrita 9907093804 Inde...hyderabad call girl.pdfRussian Call Girls in Hyderabad Amrita 9907093804 Inde...
hyderabad call girl.pdfRussian Call Girls in Hyderabad Amrita 9907093804 Inde...
 
Call Girl Price Amritsar ❤️🍑 9053900678 Call Girls in Amritsar Suman
Call Girl Price Amritsar ❤️🍑 9053900678 Call Girls in Amritsar SumanCall Girl Price Amritsar ❤️🍑 9053900678 Call Girls in Amritsar Suman
Call Girl Price Amritsar ❤️🍑 9053900678 Call Girls in Amritsar Suman
 
Call Girl Lucknow Gauri 🔝 8923113531 🔝 🎶 Independent Escort Service Lucknow
Call Girl Lucknow Gauri 🔝 8923113531  🔝 🎶 Independent Escort Service LucknowCall Girl Lucknow Gauri 🔝 8923113531  🔝 🎶 Independent Escort Service Lucknow
Call Girl Lucknow Gauri 🔝 8923113531 🔝 🎶 Independent Escort Service Lucknow
 

Cat 1

  • 1. In-Situ UV-Visible and Raman Spectroscopy for Gas-Liquid-Solid Systems C. P. Stemmet, J. C. Schouten, and T. A. Nijhuis* Laboratory of Chemical Reactor Engineering, Department of Chemical Engineering and Chemistry, EindhoVen UniVersity of Technology, P.O. Box 513, 5600 MB EindhoVen, The Netherlands This paper presents the use of UV-vis and Raman spectroscopy to measure the state of a solid in a multiphase reactor. A slurry of particles and a packed bubble column were used. As this study is a proof of principle a model for an active catalyst system, insoluble pH indicators deposited on the particles and on a solid foam packing (used as the packing material in the packed bubble column) are used. This insoluble pH indicator has different UV-vis and Raman spectra according to the liquid pH. The experimental results indicate that for the slurry the spectra obtained from the moving particles can be used to characterize the state of the pH indicator. Bubbles in the packed bubble column result in disturbances in the UV-vis and Raman spectra collected from the pH indicator adsorbed to the solid surface and this interference is removed successfully with a newly developed tolerance-and-averaging method. Introduction Approximately 85% of all chemical products (bulk and fine chemicals) are produced using a heterogeneous catalyst to increase the reaction rate and hence increase production capac- ity.1 Most often the catalyst is deposited on a solid support to aid its separation from the reaction products and examples of such reactors include packed beds, fluidized beds, and slurry reactors. Multiphase reactors, where gas and liquid reactants are converted over a heterogeneous catalyst, have been used widely in the chemical industry and examples of such reactors are slurry bubble columns, trickle bed reactors, and packed bubble columns. These multiphase reactors are used in a variety of applications ranging from the production of pharmaceuticals and hydrogenations and oxidations to produce bulk chemicals (hydro-treating, desulphurization, oxidation of sulfur dioxide, etc.). In-situ spectroscopic techniques have been developed over the years to provide much understanding into the mechanisms of catalytic reactions. In-situ spectroscopic techniques using spectroscopic probes have been applied to control gas-solid reactors operating under reaction conditions based on the state of the solid catalyst.2,3 The key advantage of control based on spectroscopic information is that the actual catalytic site (catalyst state) is being probed directly, rather than the end-of-pipe response at the reactor exit. In multiphase flows an optical probe has been used to evaluate the local volume fraction of each phase in an oil-water-gas mixture,4 however, spectroscopic techniques in characterizing catalysts operating in multiphase reactors (gas-liquid-solid) have not been investigated. It is the intention of this study to assess whether UV-vis and Raman spectroscopy can be used to determine the state of a solid catalyst in a multiphase reactor system. In a chemical reactor changes to a catalyst are usually slow, as good catalyst stability is very important for an industrial process. In this study we used a model system of which the state can be easily be manipulated which makes the model development more facile. Insoluble pH indicators were adsorbed on solid surfaces in a slurry of particles and in a packed bubble column. These adsorbed pH indicators are used as a model system for the catalyst in a multiphase reactor. In-Situ Spectroscopy. Spectroscopic measurements (ex situ or in situ) have been widely used in the development and determination of the main parameters in the preparation affecting the stability and the size of catalytically active sites and the intrinsic reaction rate. These techniques include ultraviolet-visible spectroscopy (UV-vis), infrared spectroscopy (IR), Raman spectroscopy, X-ray absorption near edge spectroscopy (XANES), and X-ray absorption fine structure (XAFS) spectroscopy. The main techniques and their limitations in catalyst characterization are given in the books edited by Weckhuysen,1 and Haw,5 as well as the review papers by Weckhuysen,6 and Hunger and Weitkamp.7 The use of in situ spectroscopy has seen significant increase, largely due to the development of optical probes which may be used under high temperature and high pressure reaction conditions. UV-vis and Raman spectroscopy can give quantitative information regarding the oxidation state and co-ordination of the metal oxide complexes serving as catalytically active species. The main uses of oxides of transition metals are for a number of chemical conversions such as hydrogenation, oxidation, decomposition, and isomerization.8 A case study has been presented of a fixed bed where UV-vis spectroscopy (diffuse reflectance spectroscopy) and Raman spectroscopy have been used to describe changes in the Cr/Al2O3 active catalyst species by increases observed in specific peaks.9 These techniques have also been used to track the formation of coke to maintain an active catalyst and using this information to control the regeneration.2 Combining spectroscopic techniques provides complementary quantification of the state of the catalyst and the changes occurring under reaction conditions.10 However, these studies have all focused on fixed beds although the Raman spectral information from moving particles in a fluidized bed may be an attractive measure for catalysts and adsorbates to monitor and avoid the formation of coke.11 An example using spectroscopy techniques to characterize the state of a slurry catalyst (Cr/SiO2) for the polymerization of ethylene, using toluene as a solvent has been illustrated.12 Here the Raman spectra of toluene dominate the spectra collected from the reaction mixture and thus obscure the spectra from the catalytic active material. UV-vis spectroscopy using visible light gives information regarding the electronic d-d and charge transfer transitions of metal oxides. It provides a convenient method to probe the state * To whom correspondence should be addressed. E-mail: T.A.Nijhuis@tue.nl. Ind. Eng. Chem. Res. XXXX, xxx, 000 A 10.1021/ie801984w CCC: $40.75  XXXX American Chemical Society
  • 2. of the active metal oxide catalyst or an adsorbed pH indicator. By the application of small optical probes, measurement is relatively easy and may also be performed through the outer wall of the reactor if made of glass.9 Raman spectroscopy using laser light provides information regarding the vibrational spectra of metal oxides and organic deposits, and optical probes based on Raman spectroscopy are becoming readily available. Multiphase Reactors. The added effect of gas-liquid and liquid-solid interfaces in gas-liquid-solid reactors contributes to the spectroscopic information collected and thus poses a challenge in obtaining reliable spectroscopic information from the catalyst deposited on the solid surface. Optical probes use optical fibers to transmit light (from a laser or lamp) into the reactor and another in the same fiber bundle to measure the reflected light (connected to a spectrometer). The interfer- ence of the liquid-solid and the gas-liquid interfaces with the light reflected from the catalyst on the solid surface is described schematically for slurry reactors and packed bubble columns in Figures 1a and 1b, respectively. These interferences need to be removed from the spectroscopic information collected to successfully determine the state of the catalyst. Model System. The aim is to develop a methodology for characterization of the solid catalyst in a multiphase system. Since transitions in the state of a real catalyst are usually very slow and occur at high temperature or pressure, a model system is used as a more convenient and controllable way to demon- strate the concept. An insoluble pH indicator adsorbed on the surface of a solid support is used in this study to simulate an active catalyst to determine UV-vis and Raman spectra in multiphase systems with relative ease. The pH indicator exhibits different UV-vis and Raman spectra depending on the pH of the liquid with which it is in contact. This system thus provides a method of studying spectroscopic measurements at mild conditions (ambient temperature and pressure). The spectra collected are only from the surface of the solid particles or solid support and the contribution of the liquid to the spectra is low due the low solubility of the pH indicator. The characteristic peaks in the UV-vis and Raman are specific to the current system. The origin of these characteristic peaks will not be discussed in detail here, since this paper focuses on the description of a methodology to do spectroscopy on a (moving) catalyst in a multiphase reactor system. The chemical origin of the bands/vibrations are less relevant and the discussion thereof is beyond the scope of this paper. Experimental Details The UV-vis experiments were performed using a reflection probe consisting of a bundle of seven optical fibers (each 200 µm thick) encased in a steel tube (120 mm in length and 6 mm in diameter). The central optical fiber is connected to a UV-vis spectrometer (AvaSpec-2048-USB2, Avantes) and the remaining six fibers are used for illumination and are connected to a light source (AvaLight-D(H)-S Deuterium-Halogen light source, 215-2000 nm, Avantes). The calibration of the reflectance spectrometer is performed using a white reference tile (WS-2, Avantes). The UV-vis spectra were recorded in LabView (LabView 7.0, National Instruments) at an integration time of 30 ms to decrease the noise to peak ratio. The Raman experiments were performed using a reflection probe consisting of a bundle of two optical fibers (200 µm Figure 1. Schematic representation during operating conditions in (a) a slurry reactor and (b) a packed bubble column, showing the interferences at the gas-liquid and liquid-solid interfaces that scatter the light away from the optical probe. Figure 2. Images of a 40 ppi solid foam covered in TiO2 and insoluble pH indicator, bromocresol green, showing the acidity of the liquid medium, (a) acidic and (b) basic. (c) Structure of bromocresol green in the acidic and basic states. B Ind. Eng. Chem. Res., Vol. xxx, No. xx, XXXX
  • 3. connected to spectrometer and 100 µm connected to the laser source) encased in a steel tube (76 mm in length with 12.7 mm diameter). This probe is connected to an integrated system combining a laser source (532 nm) and a Raman spectrometer (R-3000, Ocean Optics). The Raman spectra were recorded in LabView (LabView 7.0, National Instruments) at integration times between 1 and 5 s. All chemicals used were laboratory grade supplied from Merck unless specified otherwise. The pH of the liquid was decreased using a solution of HCl (0.6 M) which was dosed using a titrator (TIM854, Radiometer) at a rate of 0.2-5 mL/ min. The titrator also allowed measurement of the pH of the liquid using a pH-electrode (pHC2001, Radiometer). The pH indicator, bromocresol green (pH range 3.8-5.4), was dried onto the solid particles (γ-Al2O3, 63-150 µm) or solid foam support (40 ppi, ERG. Aerospace Corp.) from solution. This pH indicator is usually dissolved in a liquid (a mixture of water and ethanol) and exhibits different UV-vis and Raman spectra depending on the liquid pH. Slurry of Particles. Alumina particles (63-150 µm) covered with bromocresol green were prepared using 5 wt % ethanol as a solvent and dried in an oven at 90 °C from solution at a liquid pH of approximately 5. The particles were filtered and washed with distilled water until the filtrate remained clear. The pH indicator has a low solubility in water and thus remains on the surface of the particles and responds in the UV-vis and Raman spectrum depending on the pH of the liquid surrounding the particles. Packed Bubble Column. Aluminum foam pieces (40 ppi, ERG. Aerospace Corp.) were used as the packing material in the packed bubble column experiments. Solid foams are convenient to use due to the open structure and high surface area for deposition of the pH indicator.13 The aluminum foam was dipped in a slurry of TiO2 (17 wt %, P25, Degussa) and dried while rotating at 60 rpm. Air was blown through the samples to avoid pores of the solid foam material remaining filled with the slurry. The solid foam pieces were calcined at 400 °C resulting in washcoats of approximately 7-10 wt %. The bromocresol green pH indicator was applied to the surface of the solid foam structure using the same preparation technique described for the slurry particles, resulting in solid foams which respond according to pH of the liquid medium, see Figure 2. The bubble column used for the experiments is a cylindrical glass vessel (diameter of 10 cm) where nitrogen is distributed through a fritted glass disk from the bottom. The solid foam was placed on the distributor and the UV-vis and Raman probes focused on the solid foam packing from the top (approximately 2-5 mm focal distance). As gas bubbles travel through the solid foam they interfere with the signal collected by the UV-vis and Raman probes. Bubble interference was removed with an in-house Labview application designed to discard compromised UV-vis and Raman spectra. This method of tolerance-and- averaging is described schematically in Figure 3. An average spectrum (AS) is created over m (e.g., 10) number of spectra. In the next cycle, the newly recorded spectrum (NS) is compared to this averaged spectrum (ASold) for the UV-vis spectra over wavelengths 300-800 nm and Raman spectra over Raman shifts 800-4300 cm-1 . The measurement is discarded if the sum of the absolute difference between the spectra is larger than a set tolerance value. This is repeated for the next nine (m - 1) measurements. If, for example, three spectra were discarded, a new average spectrum (ASnew) is created from the remaining seven (n) spectra and used for comparison in the next cycle. This procedure is repeated, producing a new average spectrum for every 10 measurements. A spectrum averaged over a too large number of spectra (e.g., m equal to 100) will have less noise, but also rapid changes in the state of the pH indicator are possibly evaluated as the interference of a bubble and thus could be lost. Results and Discussion Slurry of Particles. A slurry of 30 wt % alumina particles with the bromocresol green pH indicator adsorbed to the particles’ surface was used to evaluate if UV-vis and Raman spectroscopy could be used to determine changes in the state of the pH indicator. UV-vis spectra were recorded every 200 ms while the pH of the liquid was decreased from 4 to 1 over a time period of 300 s. The UV-vis spectra of the slurry of bromocresol green particles are given in Figure 4 for a select number of liquid pH values. The reflectance at 620 and 450 nm increases and decreases, respectively, as the pH is decreased. Raman spectra were recorded every 15 s (integration time 5 s), while the pH of the liquid was decreased from 4 to 1 over a time period of 300 s. The Raman spectra of the slurry of bromocresol green particles are given in Figure 5. Raman intensities are observed to increase (at approximately 833, 1570, 2210, 2680, 3040, and 4000 cm-1 ) as a result of the bromocresol green changing from the basic to the acidic state with the decrease in the liquid pH (addition of H+ ) (see Figure 2c for the structural changes). Thus, the UV-vis and Raman spectra Figure 3. Schematic representation of the tolerance-and-averaging method to remove bubble interference from UV-vis spectra. New spectra (NS) from the spectrometer are compared to an average spectrum (AS) determined over m spectra over the previous cycle. The sum of the absolute difference is compared to a tolerance value (Tol) and if larger, the spectrum is discarded. A new average is created from the n number of spectra remaining and is used in the next cycle. Figure 4. Observed UV-vis reflectance spectra for a 30% slurry of particles covered with bromocresol green, while the pH of the liquid is decreased. Ind. Eng. Chem. Res., Vol. xxx, No. xx, XXXX C
  • 4. can be obtained from moving particles, and transients from one state to another can be observed. Packed Bubble Column. The experiments for the UV-vis and Raman spectra resulting from the adsorbed bromocresol green pH indicator operating in a packed bubble column are shown in Figure 6-Figure 8. Fifty UV-vis spectra are recorded every 2 s and the tolerance-and-averaging method is used to assess the validity of the recorded spectra. The Raman spectra measurements are performed every 2 s and 10 are used for the tolerance-and-averaging method. In Figure 6b the average recorded UV-vis spectra for the wavelength 620 nm is given while the liquid pH is decreased, see Figure 6c. As the bubble flow starts and bubbles interfere with the UV-vis spectra and Raman spectra obtained, the tolerance-and-averaging method maintains the average spectral measurement while discarding spectra compromised by bubbles. The average UV-vis spectra determined using the tolerance-and-averaging method shows less noise in the recorded spectra at 620 nm. The relative difference between the average reflectance determined using all the spectra and the average reflectance determined using the tolerance-and- averaging method can be as high as 30-40% (see Figure 6d). An increase in the UV-vis reflectance is observed as the liquid pH is decreased (see Figure 4). Similarly, the changing Raman intensity recorded at a Raman shift of 4000 cm-1 as a result of decreasing liqiud pH is presented in Figure 7. A figure created in a similar manner for one of the other characteristic bands visible in the Raman spectrum for the slurry particles (at approximately 833, 1570, 2210, 2680, 3040 cm-1 ) yielded a graph with a comparable trend. The average Raman intensity increases as the pH of the liquid is decreased over time. As much as 80% of spectra may be discarded using the tolerance- and-averaging method resulting in an averaged Raman spectrum Figure 5. (a) Observed Raman spectra for a 30 wt % slurry of particles covered with bromocresol green, while the pH of the liquid is decreased over time. (b) Liquid pH-profile over time. Figure 6. The UV-vis spectra obtained for the bromocresol green pH indicator deposited on the surface of a solid foam packing operating in a packed bubble column. (a) Reflectance at a wavelength of 620 nm is presented. The spectra which have been discarded for averaging are given by the open symbols (O) and used for determining the average in closed symbols (b). (b) The average reflectance of all data points (---) and the reflectance determined by the tolerance-and-averaging method (s) are given. (c) Liquid pH-profile over time. (d) Relative error between the average for all the data points and the average determined by the tolerance-and-averaging method. D Ind. Eng. Chem. Res., Vol. xxx, No. xx, XXXX
  • 5. with less noise when compared when all measurement are used. In Figure 8 the averaged Raman spectra is given for a decrease in the liquid pH over time. Comparing Figure 8 and Figure 5 shows that similar Raman intensities are observed to increase as the bromocresol green pH indicator makes a transition to its acidic form. The differences in the observed Raman spectra for the slurry of particles and packed bubble column is due the pH indicator being adsorbed onto two different materials, Al2O3 and TiO2, respectively. Conclusions UV-vis and Raman spectra recorded for a slurry of particles, and a solid foam packed bubble column are shown for insoluble pH indicators adsorbed onto the solid surface of the particles and the solid foam packing. A change in the adsorbed pH indicator is accomplished by a change in the liquid pH and results in a clear change in the UV-vis and Raman spectra. These examples indicate that, although the UV-vis and Raman spectra are recorded under multiphase conditions, changes in the solid surface conditions can be observed. In a packed bubble column the averaging-and-tolerance method can be used to remove spectra that have been compromised by gas-liquid interfaces. The results from this study indicate that in situ UV-vis and Raman spectroscopy can be used to characterize a solid surface operating under multiphase conditions where bubbles interfere with the measured spectra. Acknowledgment The authors gratefully acknowledge the Dutch Technology Foundation STW (STW-NWO Vidi Project 06186) for the financial support. Dow Benelux B.V., BASF Nederland B.V., Avantium Technologies, and Shell Global Solutions Interna- tional B.V. are acknowledged for their input during the course of the project. Literature Cited (1) Weckhuysen, B. M. In-Situ Spectroscopy of Catalysts; American Scientific Publishers: Stevenson Ranch, CA, 2004. (2) Bennici, S. M.; Vogelaar, B. M.; Nijhuis, T. A.; Weckhuysen, B. M. Real-time control of a catalytic solid in a fixed-bed reactor based on in-situ spectroscopy. Angew Chem., Int. Ed. 2007, 46, 5412–5416. (3) Nijhuis, T. A.; Tinnemans, S. J.; Visser, T.; Weckhuysen, B. M. Operando spectroscopic investigation of supported metal oxide catalysts by combined time-resolved UV-VIS/Raman/on-line mass spectrometry. Phys. Chem. Chem. Phys. 2003, 5, 4361–4365. (4) Ramos, R. T.; Holmes, A.; Wu, X.; Dussan, E. A local optical probe using fluorescence and reflectance for measurment of volume fractions in multi-phase flows. Meas. Sci. Technol. 2001, 12, 871–876. Figure 7. The Raman spectra obtained for the bromocresol green pH indicator deposited on the surface of a solid foam packing operating in a packed bubble column. (a) Raman intensity at a Raman shift of 4000 cm- 1 is presented. The spectra which have been discarded for averaging are given by the open symbols (O) and used for determining the average in closed symbols (b). (b) The average Raman spectra of all data points (---) and the average determined by the tolerance-and-averaging method (s) are given. (c) Liquid pH-profile over time. Figure 8. The average Raman spectra obtained for bromocresol green pH indicator deposited on the surface of a solid foam packing operating in a packed bubble column as the liquid pH is decreased over time. Peaks in the Raman spectra are observed as the pH indicator undergoes transition to its acidic form. Ind. Eng. Chem. Res., Vol. xxx, No. xx, XXXX E
  • 6. (5) Haw, J. F. In In-Situ Spectroscopy in Heterogeneous Catalysis; Wiley-VCH: Weinheim, Germany, 2002. (6) Weckhuysen, B. M. Snapshots of a working catalyst: possibilities and limitations of in situ spectroscopy in the field of heterogeneous catalysis. Chem. Commun. 2002, 97–110. (7) Hunger, M.; Weitkamp, J. In situ IR, NMR, EPR, and UV-vis Spectroscopy: Tools for New Insight into the Mechanisms of Heterogeneous Catalysis. Angew Chem., Int. Ed. 2001, 40, 2954–2971. (8) Weckhuysen, B. M.; Wachs, I. E. Catalysis by supported metal oxides. In Handbook of Surface Interfaces of Materials; Academic Press: San Diego, CA, 2001; Vol. 1, Chapter 11, pp 613-648. (9) Nijhuis, T. A.; Tinnemans, S. J.; Visser, T.; Weckhuysen, B. M. Towards real-time spectroscopic process control for the dehydrogenation of propane over supported chromium oxide catalysts. Chem. Eng. Sci. 2004, 59, 5487–5492. (10) Tinnemans, S. J.; Mesu, J. G.; Kervinen, K.; Visser, T.; Nijhuis, T. A.; Beale, A. M.; Keller, D. E.; van der Eerden, A. M. J.; Weckhuysen, B. M. Combining operando techniques in one spectroscopic-reaction cell: oppertunities for elucidating the active site and related reaction mechanism in catalysis. Catal. Today 2006, 113, 3–15. (11) Chua, Y. T.; Stair, P. C. A novel fluidized bed technique for measureing UV Raman spectra of catalysts and adsorbates. J. Catal. 2000, 196, 66–72. (12) Weckhuysen, B. M. Determining the active site in a catalytic process: Operando spectroscopy is more than a buzzword. Phys. Chem. Chem. Phys. 2003, 5, 4351–4360. (13) Stemmet, C. P.; Meeuwse, M.; van der Schaaf, J.; Kuster, B. F. M.; Schouten, J. C. Gas-liquid mass transfer and axial dispersion in solid foam packings. Chem. Eng. Sci. 2007, 62, 5444–5450. ReceiVed for reView December 23, 2008 ReVised manuscript receiVed February 23, 2009 Accepted February 26, 2009 IE801984W F Ind. Eng. Chem. Res., Vol. xxx, No. xx, XXXX