Part II also includes admittance and demag testing and uses more advanced equipment to test the meter. See live results from today’s newest test equipment.
Artificial intelligence in the post-deep learning era
TESCO Tuesday: Traditional Ratio, Burden, Admittance and Demag Testing – Part II
1. 1
Today’s session will begin shortly
@tescometering
@TESCO_Metering
TESCO -The Eastern Specialty Company
TESCO -The Eastern Specialty Company
2. 2
10/02/2012 Slide 2
CT Testing - Ratio,
Burden, Admittance
Part 2
Prepared by Rob Reese, TESCO
The Eastern Specialty Company
For TESCO Tuesday
Tuesday, March 16, 2021
3. 3
Agenda
• A Quick Review
• Detail of Additional Considerations
• Questions
4. 4
Review – Ratio, Burden, Admittance
Three Test Methods
1. Ratio
2. Burden
3. Admittance
8. 8
Ratio Testing
Ratio of Primary Current to Secondary Current
5A
PHASE A
SOURCE LOAD
400A
400A
400A
5A
5A
Calculate
Ratio
9. 9
Burden Testing
Functionality with Burden Present on the Secondary Loop
Example Burden Spec:
0.3% @ B0.1, B0.2, B0.5
or
There should be less than the 0.3%
change in secondary current from initial
(“0” burden) reading, when up to
0.5Ohms of burden is applied
10. 10
Burden Testing
Functionality with Burden Present on the Secondary Loop
ANSI Burden Values
0.1 Ohms
0.2 Ohms
0.5 Ohms
1 Ohms
2 Ohms
4 Ohms
8 Ohms
14. 14
Admittance Testing
• What is Admittance?
• Admittance testing measures the
overall “health” of the secondary
loop of the CT.
• Measured in units of MiliSiemens
(mS)
• Admittance is the inverse of
impedance.
• Impedance is the opposition to
current.
• Therefore, admittance testing
measures the overall “health” of
the secondary loop of the CT.
15. 15
Admittance Testing
• Admittance testing devices inject an audio sine
wave signal into the secondary loop of the CT.
• The resulting current is measured.
• The voltage of the initial signal is known.
• From these two parameters, the impedance,
and thus the admittance can be calculated.
Cat 6330 – Meter Site Analyzer
16. 16
Additional Considerations
• Importance
• Multipliers
• Parallelograms
• Accuracy Graphs
• Shop vs. Field Testing
• Standards
• Pole Top CT’s
• No Test Switches
• Demag Function
18. 18
Additional Considerations
• Yes!
• The meter is only one
small part of the whole
system.
• CT errors can be much
bigger than meter
accuracy errors
SOURCE LOAD
19. 19
Additional Considerations
Scenario 1: Meter Error
If the customer’s bill should
be $10,000, and the meter
registers 98%, you are still
getting $9,800
SOURCE LOAD
98%
20. 20
Additional Considerations
Scenario 2: One CT is
reversed
If the customer’s bill should
be $10,000, and on of the
CT’s are reversed, you are
losing 2/3rds of your
revenue, or $3,300 left.
SOURCE LOAD
31. 31
Additional Considerations
Field Testing
• Equipment relatively low cost
• Tests the whole service
• Limited to testing at customer’s
current load
Shop Testing
• Sizeable Investment
• Will test at high currents –
8000amps – through a CT’s
complete operating range
• Only tests the CT, not the entire
service
33. 33
Additional Considerations
Governing Standards for CT’s
• ANSI C12.11
• IEEE C57.13 standard requirements for instrument transformers
• IEEE C57.13.6 standard for high-accuracy instrument transformers
• IEC 60044-1 current transformers
• IEC 60044-6 requirements for protective current transformers for transient
performance
• IEC 61869-2 additional requirements for current transformers
41. 41
De-magnetization
CT’s can become magnetized, due to a number of reasons, including
leaving the shorting clip open, near lightning strikes, and harmonic
content.
CT’s can be demagnetized by slowly and smoothly increasing the
secondary resistance until saturation occurs, and then slowly and
smoothly decreasing the secondary resistance.
43. 43
Metering Configurations
Presentation Date
Realizing Your Investment in Technology Through Analytics 3/23
Shop Testing 3/30
Preparing for AMI Deployment 4/6
Communication Equipment 4/13
Sign up: tescometering.com/tesco-tuesdays