Redford D et al (University of Arizona); Absolute and Trend
Accuracy of Masimo O3 Regional Oximetry in healthy Volunteers
During Controlled Hypoxia; Abstract 2014
• 23 of 27 subjects enrolled used for data collection
• Ages 18-35 years, 12 Males and 11 Females
• 202 paired measurements
• Reference arterial and venous blood samples (SavO2), using a
70:30 ratio of SjvO2 and SavO2
• Reference blood samples were taken from both radial artery and
internal jugular bulb
• Obtained at baseline and after a series of increasingly hypoxic
states
• Accuracy:
• Absolute 4.0% ARMS
• Trending 2.1 % ARMS

Redford Synopsis123145617891010112124.pptx

  • 1.
    1 ©2013 Nonin Medical,Inc. All trademarks are the property of Nonin Medical, Inc. unless otherwise noted. M-10913 CONFIDENTIAL Redford D et al (University of Arizona); Absolute and Trend Accuracy of Masimo O3 Regional Oximetry in healthy Volunteers During Controlled Hypoxia; Abstract 2014 • 23 of 27 subjects enrolled used for data collection • Ages 18-35 years, 12 Males and 11 Females • 202 paired measurements • Reference arterial and venous blood samples (SavO2), using a 70:30 ratio of SjvO2 and SavO2 • Reference blood samples were taken from both radial artery and internal jugular bulb • Obtained at baseline and after a series of increasingly hypoxic states • Accuracy: • Absolute 4.0% ARMS • Trending 2.1 % ARMS