Comparison of Reactive Rates Between EIA and Chemiluminescent Donor Screening
Assays
G. Martinez, N. Haubert, G. Leparc, P. Williamson, R. Spizman, G. Robertson, S. Caglioti,
Creative Testing Solutions, a large national
donor testing laboratory changed its supplier
of screening immunoassays for infectious
diseases (Viral Markers or VMs) during the
month of July in 2013. The new VMs
generate chemiluminescence signals
(ChLIA) to detect the targeted molecules.
Repeat reactive rates (RRR) and, when
applicable, confirmatory rates (CR) of VMs
were compared between Enzyme-Linked
Immunosorbent Assay (EIA) and ChLIA for a
period of six months before and after the
change in suppliers to assess comparative
performance and associated culling effect
on blood donors.
CTS observed differences in the RRR
performance of VMs with the two different
technologies. Only the HCV assays had
similar RRR and percentage of CR.
Four of the five ChLIA VM assays
demonstrated lower RRR than EIA.
The ChLIA HIV assay had a higher RRR
and a lower percentage of CR than the
corresponding EIA..
The combined VM results from the ChLIAs
are expected to result in 9 fewer
unnecessary discarded donations per
10,000 donations tested as compared to
the EIAs.
Given the 13.8% decrease in the HIV RRR
between the first and the sixth month of use
for the ChLIAs, we anticipate continued
improvement in the total RRR compared to
EIA, as culling of false positive donors is
completed.
Methods
Viral Marker RRRs and CRs were
calculated from a volume of at least 2
million samples received from across the
country over the six months period of time
before the change in supplier and over the
six months period of time immediately after
the change.
This comparison primarily focused on VMs
required to be performed on every donation.
T. cruzi was not fully included in this report
because it is only performed on donors who
presumably donated for the first time and
they would not help us see if and to what
degree the performance of the ChLIA assay
improved over time as the population of EIA
tested donors was culled by the new
methodology. However, representative EIA
and ChLIA rates will be briefly listed in the
results section.
A decrease of 0.093% in the combined RRR
with the ChLIA VM assays was observed by
CTS for every donation as compared to the
RRR of the corresponding set of required
EIA VM assays. The difference in the total
RRRs represents approximately 9 fewer
discarded donations per 10,000 donations
tested with the ChLIAs as compared to the
EIAs.
RRR Table for the Five Required VMs
EIA ChLIA ChLIA- EIA
VM Assay RRR CR RRR CR RRR CR
HIV 1 / 2 0.038% 30.2% 0.103% 9.6% 0.065% -20.6%
HCV 0.152% 56.8% 0.148% 55.1% -0.004% -1.7%
HBsAg 0.083% 27.8% 0.032% 64.1% -0.051% 36.3%
HBc 0.483% N/A 0.435% N/A -0.048% N/A
HTLV I / II 0.098% 13.0% 0.043% 26.2% -0.055% 13.2%
Sum 0.854% 0.761% -0.093%
Changes in VM Confirmed Results
During the Study Period
0%
5%
10%
15%
20%
25%
30%
35%
40%
Percent of Repeat Reactive HIV 1/2 Samples that Were
Positive with a Confirmatory Assay
0%
5%
10%
15%
20%
25%
30%
35%
Percent of Repeat Reactive HTLV I/II Samples that
Were Positive with a Confirmatory Assay
0%
10%
20%
30%
40%
50%
60%
70%
80%
Percent of Repeat Reactive HBsAg Samples that Were
Positive with a Confirmatory Assay
Results
0%
10%
20%
30%
40%
50%
60%
70%
Percent of Repeat Reactive HCV Samples that Were
Positive with a Confirmatory Assay
T. Cruzi EIA ChLIA
RRR 0.017% 0.079%

Reactive rate comparison AABB 2014

  • 1.
    Background Conclusion Insert Footer orCopyright Information Here Printed by Comparison of Reactive Rates Between EIA and Chemiluminescent Donor Screening Assays G. Martinez, N. Haubert, G. Leparc, P. Williamson, R. Spizman, G. Robertson, S. Caglioti, Creative Testing Solutions, a large national donor testing laboratory changed its supplier of screening immunoassays for infectious diseases (Viral Markers or VMs) during the month of July in 2013. The new VMs generate chemiluminescence signals (ChLIA) to detect the targeted molecules. Repeat reactive rates (RRR) and, when applicable, confirmatory rates (CR) of VMs were compared between Enzyme-Linked Immunosorbent Assay (EIA) and ChLIA for a period of six months before and after the change in suppliers to assess comparative performance and associated culling effect on blood donors. CTS observed differences in the RRR performance of VMs with the two different technologies. Only the HCV assays had similar RRR and percentage of CR. Four of the five ChLIA VM assays demonstrated lower RRR than EIA. The ChLIA HIV assay had a higher RRR and a lower percentage of CR than the corresponding EIA.. The combined VM results from the ChLIAs are expected to result in 9 fewer unnecessary discarded donations per 10,000 donations tested as compared to the EIAs. Given the 13.8% decrease in the HIV RRR between the first and the sixth month of use for the ChLIAs, we anticipate continued improvement in the total RRR compared to EIA, as culling of false positive donors is completed. Methods Viral Marker RRRs and CRs were calculated from a volume of at least 2 million samples received from across the country over the six months period of time before the change in supplier and over the six months period of time immediately after the change. This comparison primarily focused on VMs required to be performed on every donation. T. cruzi was not fully included in this report because it is only performed on donors who presumably donated for the first time and they would not help us see if and to what degree the performance of the ChLIA assay improved over time as the population of EIA tested donors was culled by the new methodology. However, representative EIA and ChLIA rates will be briefly listed in the results section. A decrease of 0.093% in the combined RRR with the ChLIA VM assays was observed by CTS for every donation as compared to the RRR of the corresponding set of required EIA VM assays. The difference in the total RRRs represents approximately 9 fewer discarded donations per 10,000 donations tested with the ChLIAs as compared to the EIAs. RRR Table for the Five Required VMs EIA ChLIA ChLIA- EIA VM Assay RRR CR RRR CR RRR CR HIV 1 / 2 0.038% 30.2% 0.103% 9.6% 0.065% -20.6% HCV 0.152% 56.8% 0.148% 55.1% -0.004% -1.7% HBsAg 0.083% 27.8% 0.032% 64.1% -0.051% 36.3% HBc 0.483% N/A 0.435% N/A -0.048% N/A HTLV I / II 0.098% 13.0% 0.043% 26.2% -0.055% 13.2% Sum 0.854% 0.761% -0.093% Changes in VM Confirmed Results During the Study Period 0% 5% 10% 15% 20% 25% 30% 35% 40% Percent of Repeat Reactive HIV 1/2 Samples that Were Positive with a Confirmatory Assay 0% 5% 10% 15% 20% 25% 30% 35% Percent of Repeat Reactive HTLV I/II Samples that Were Positive with a Confirmatory Assay 0% 10% 20% 30% 40% 50% 60% 70% 80% Percent of Repeat Reactive HBsAg Samples that Were Positive with a Confirmatory Assay Results 0% 10% 20% 30% 40% 50% 60% 70% Percent of Repeat Reactive HCV Samples that Were Positive with a Confirmatory Assay T. Cruzi EIA ChLIA RRR 0.017% 0.079%