SlideShare a Scribd company logo
1 of 109
LOW FLOW ANAESTHESIA SYSTEM/
CO2 ABSORBER/ HUMIDIFICATION/
SCAVENGING SYSTEM
History of Low flow Anaesthesia
 As early as 1850 John Snow (1813-53) recognized that only a
small fraction that inhalation anaesthetics gets metabolized
and most of it exhaled unchanged in the expired air of the
anaesthetized patients.
He also concluded that the narcotic effect of the volatile
anaesthetics could be markedly prolonged by re-inhaling
these unused vapours.
 About 75 years later, in 1924, the rebreathing technique with
carbon dioxide absorption was introduced into routine clinical
practice by Ralph Waters (1883-1979), the to-and-fro
absorption system.
 Brian C. Sword introduced the circle (CO2 absorption) system in
the 1930s.
LOW FLOW ANESTHESIA
• Any technique that utilises a fresh gas flow (FGF) that is less than the
alveolar ventilation can be classified as „Low flow anaesthesia‟.
• Baum et al : had defined it as a technique wherein at least 50% of the
expired gases had been returned to the lungs after carbon dioxide
absorption in the next inspiration.
• Baker had classified the FGF used in anaesthetic practice into
the following categories:
• Metabolic flow : about 250 ml /min
• Minimal flow : 250-500 ml/min.
• Low flow : 500- 1000 ml/min.
• Medium flow : 1 - 2 l/min.
• For most practical considerations, utilization of a fresh gas flow less than 2
litres/min may be considered as low flow anaesthesia.
CONT.
• there exists a need for a system that provided the advantages of
the completely closed circuit and at the same time, reduced the
dangers associated with it.
• Low flow anaesthesia fulfilled these requirements.
• Low flow anaesthesia involves utilizing a fresh gas flow which is
higher than the metabolic flows but which is considerably lesser
than the conventional flows.
• The larger than metabolic flows provides for considerably greater
margin of safety and satisfactory maintenance of gas composition
in the inspired mixture.
• THE NEED FOR LOW FLOW ANAESTHESIA
• Completely closed circuit anaesthesia is based upon the reasoning that
anaesthesia can be safely maintained if the gases which are taken up
by the body alone are replaced into the circuit taking care to remove
the expired carbon dioxide with sodalime. No gas escapes out of the
circuit and would provide for maximal efficiency for the utilisation of
the fresh gas flows.
Requirements for LFA
1. A circle system with CO2 absorption.
2. A flow meter for the adjustments of FGF below 1 l/min.
3. Gas-tight breathing system, recommended test leakage
should be below 150 ml/min.
4. The breathing system must be having minimal internal
volume and a minimum number of components and
connections.
5. Continuous gas monitoring must be employed, essential in
controlling the patient's alveolar gas concentrations.
6. The vaporizers delivering the high concentrations must be
calibrated and must be accurate at low fresh gas flow.
Circle System
• The minimum requirement for conduct of low
flow anaesthesia is the effective absorption of
CO2 from the expired gas, so that the CO2 free
expired gas can be reutilized for alveolar
ventilation.
• In the past, two systems were in use; “To and
Fro” system introduced by Waters and the “circle
system” introduced by Brian Sword.
• But because of its bulkiness near the patient and other
disadvantages the “To and Fro” has gone out of favour
and the “circle system” using large sodalime canisters is
in use.
Describing a Circle System.
The circle system should have the
basic configuration with two
unidirectional valves on either side
of the sodalime canister, fresh gas
entry, reservoir bag, pop off valve,
and corrugated tubes and „Y‟ piece
to connect to the patient.
The relative position of fresh gas
entry, pop off valve, and reservoir
bag are immaterial as long as they
are positioned between the
expiratory and the inspiratory
unidirectional valves that functions
properly and CO2 absorption is
efficient at all times.
THE PRACTICE OF LOW FLOW
ANAESTHESIA:
The practice of low flow anaesthesia can be
dealt with under the following three
headings:
o 1. Initiation of Low flow anaesthesia
o 2. Maintenance of Low flow anaesthesia
o 3. Termination of Low flow anaesthesia.
INITIATION OF LOW FLOW ANAESTHESIA
Primary aim at the start of low flow anaesthesia is to
achieve an alveolar concentration of the anaesthetic
agent that is adequate for producing surgical
anaesthesia (approximately 1.3 MAC).
 Factors influencing to buildup of alveolar concentration should be
considered . These factors can broadly be classified into three
groups :
 1) Factors governing the inhaled tension of the anaesthetic agent,
 2) Factors responsible for rise in alveolar tension,
 3) Factors responsible for reducing the alveolar tension, uptake
from the lungs.
Factors governing the inhaled tension of
the anaesthetic.
Three most important factors responsible
for inhaled tension of anesthetics are :
 1.BREATHING CIRCUIT VOLUME
 2. RUBBER GAS SOLUBILITY
3. SET INSPIRED CONCENTRATION
Breathing Circuit Volume
 Circle system is often bulky and volume roughly equal to 6-7 liters, FRC of the
patient, which is roughly 3 litres, together constitutes a reserve volume of 10
liters; to which the anaesthetic gases and vapours have to be added.
• With the addition of FGF, the rate of change of composition of the reserve
volume is exponential, time required for the changes to occur is governed by
the time constant, which is equal to this reserve volume divided by the fresh
gas flow.
• Three time constants are needed for a 95% change in the gas
concentration to occur.
• Hence, if a FGF of 1L/min is used, then 30 minutes will be required for the
circuit concentration to reflect the gas concentration of the FGF.
• If the FGF is still lower, then correspondingly longer time will be
required.
Rubber Gas Solubility of the agent
• The anaesthetic agent could be lost from
the breathing system due to solubility of the
agent in rubber, and permeability through
the corrugated tubes.
• The concentration will take a longer time if
solubility is more.
• Though the amount of loss will be minimal,
it should be considered at the start if the
aimed anaesthetic concentration is low.
SET INSPIRED CONCENTRATION
 The rate of rise of alveolar partial pressure of the
anaesthetic agent must bear a direct relationship
to the inspired concentration.
 Higher the inspired concentration, the more
rapid is the rise in alveolar concentration.
 The Inspired concentration of the agent
depends on several factors.
– Denitrogenation of the circuit volume.
– Alveolar ventilation.
– Concentration effect.
Denitrogenation of FRC & Breathing
Circuit Volume
The functional residual capacity of the lung
and the body as a whole contain nitrogen
which will try to equilibrate with the circuit
volume and alter the gas concentration if
satisfactory denitrogenation is not achieved at
the start of anaesthesia.
 Hence, as a prelude to the initiation of closed or low flow
anaesthesia, thorough denitrogenation must be achieved with
either a non-rebreathing circuit or the closed circuit with a
large flow of oxygen and a tight fitting facemask.
Alveolar ventilation:
The second factor governing the delivery of
anaesthetic agent to the lung is the level of
alveolar ventilation.
The greater the alveolar ventilation, the
more rapid is the rise of alveolar
concentration towards the inspired
concentration.
• This effect is limited only by the lung volume, the larger
the functional residual capacity, the slower the wash. in
of the new anaesthetic gas.
Factors responsible for uptake from the lungs
thus reducing the alveolar tension
• 1. CARDIAC OUTPUT .
• 2. BLOOD GAS SOLUBILITY .
• 3. ALV – VENOUS GRADIENT.
Cardiac output
• Because blood carries anaesthetic away from the
lungs, the greater the cardiac output, the
greater the uptake, and consequently the slower
the rate of rise of alveolar tension.
• At low inspired concentration, the alveolar
concentration results from a balance between
the ventilatory input and circulatory uptake.
• If the later removes half the anaesthetic
introduced by ventilation, then the alveolar
concentration is half that inspired.
Blood gas solubility
• If other things are equal, the greater the blood gas
solubility coefficient, the greater the uptake of
anaesthetic, and slower the rate of rise of alveolar
concentration.
Alveolar to venous partial pressure gradient:
• During induction the tissues remove all the
anaesthetic brought to them by the blood.
Tissue uptake lowers the venous anaesthetic
partial pressure far below that of the arterial
blood.
• If tissue uptake is more then it results in a large alveolar
to venous anaesthetic partial pressure difference,
which promotes maximum anaesthetic uptake and
hence lowers the alveolar partial pressure.
Concentration effects
 The concentration effect modifies this influence of uptake.
o When appreciable volumes are taken up rapidly, the lungs do not
collapse; instead the sub atmospheric pressure created in the lung by the
anaesthetic uptake causes passive inspiration of an additional volume of
gas to replace that lost by uptake, thus increasing the alveolar
concentration and offsetting the mathematical calculations.
o Similarly, if an insoluble gas (e.g., nitrogen) is present in the inspired
mixture, as the blood takes up the anaesthetic gas, the concentration of
the insoluble gas will go up in the alveoli, reducing the concentration of
the anaesthetic agent.
o Concentration effect: The concentration effect helps in raising the
alveolar tension towards the inspired tension, but hinders with it if an
insoluble gas is present in the mixture.
Methods to achieve desired gas and
agent concentration.
• High flow for short time
• Use of Prefilled Circuits with gases
• Use of large doses of anaesthetic agents.
• Injector technique.
Use of high flows for a short time
 This is by and far the commonest and the most effective technique of initiating
closed circuit.
• By using high flows for a short time, the time constant is reduced thereby
bringing the circuit
concentration to the desired concentration rapidly.
• Often, a fresh gas flow of 10L of the desired gas concentration and 2 MAC agent
concentration is used so that by the end of three minutes (three time
constants) the circuit would be brought to the desired concentration.
 Advantages:
--- Rapidity with which the desired concentration is achieved, the ability to
prevent unexpected raise in the agent concentration and the ability to
plenum vaporizers to achieve the desired concentration.
– This also has the added advantage of achieving better denitrogenation, so
vital to the conduct of the low
flow anaesthesia.
 The chief disadvantage:
- Not economical & need scavenging systems to prevent theatre pollution
 Mapleson3 using a spreadsheet model of a circle breathing system has
calculated that, by using a FGF equal to minute ventilation and setting the
anaesthetic agent partial pressure to 3 MAC, the end expired partial pressure
of halothane will reach 1 MAC in 4 minutes and that of isoflurane in
1.5 minutes.
Use of Prefilled circuit
• The second method is utilizing a
different circuit like Magill’s for pre-
oxygenation.
• Simultaneously, the circle is fitted with a test
lung and the entire circuit is filled with the gas
mixture of the desired concentration.
• Following intubation, the patient is connected to the circuit
thereby ensuring rapid achievement of the desired
concentration in the circuit.
Use of large doses of anaesthetic
agents.
 The third method consists of adding large amounts of anaesthetic agent
into the circuit so that the circuit volume + FRC rapidly achieves the
desired concentration as well as compensates for the initial large
anaesthetic gas uptake.
 To execute this, Pre-oxygenation after intubation, fresh gas flow is started
with metabolic flows of oxygen and a large amount of nitrous oxide often
in the range of 3-5 litres per minute.
 Oxygen concentration in the circuit gradually falls, is continuously
monitor Oxygen and the nitrous oxide concentration, N2O flow is reduced
once the desired oxygen concentration (33 - 40%) is achieved .
 Disadvantage of this method : is hypoxia if the oxygen monitor were to
malfunction.
o Hence this method is seldom used for N2O.
o In the contemporary anaesthesia machines it is not possible to administer
lesser than 25% of oxygen and the described technique cannot be
executed.
Injection techniques.
 An alternative method for administering the large amounts of the
agents is by directly injecting the agent into the circuit. The high
volatility coupled with the high temperature in the circle results in
instantaneous vaporization of the agent.
 The injection is made through a self sealing rubber diaphragm
covering one limb of a metal t piece or a sampling port, inserted into
either the inspiratory or the expiratory limb.
o This is an old, time-tested method and is extremely reliable.
o Each ml of the liquid halothane, on vaporization yields 226 ml of
vapour and each ml of liquid isoflurane yields 196 ml of vapour at
20oC. Hence, the requirement of about 2ml of the agent is
injected in small increments into the circuit.
THE MAINTENANCE OF LOW FLOW
ANAESTHESIA
This is the most important phase as this is
stretched over a long period of time and financial
savings result directly from this; this phase is
characterized by :
1. Need a steady alveolar concentration of respiratory gases.
2. Minimal uptake of the anaesthetic agents by the body.
3. Need to prevent delivery of hypoxic gas mixtures.
 Since the uptake of the anaesthetic agent is small in this phase,
adding small amounts of the anaesthetic gases to match the uptake
and providing oxygen for the basal metabolism should suffice.
Management of the oxygen and nitrous oxide flow
during the maintenance phase
 The need to discuss the flow rates of N2O and O2 arises specifically because
of the possible danger of administration of a hypoxic mixture.
 Let us analyze the following example. 33% oxygen is set using a flow of 500
ml of O2 and 1000 ml of N2O.
o Oxygen is taken up from the lungs at a constant rate of about 4 ml/kg/min.
o N2O is a relatively insoluble gas and after the initial equilibration with the FRC
and vessel rich group of tissues, the up take is considerably reduced.
o In this situation, there is a constant removal of O2 at a rate of 200 - 250
ml/min, where as the insoluble gas N2O uptake is minimal. Hence the gas that
is partly vented and partly returning to the circuit will have more N2O and less
of O2.
o Over a period of time, due to the mixing of fresh gas that has 66% N2O and the
expired CO2 free gas that has N2O much higher than that, the percentage of
N2O will go up and that of O2 will fall, sometimes dangerously to produce
hypoxic mixtures.
A high flow of 10 lit/min at the start, for a period of 3 minutes, is followed by a
flow of 400 ml of O2 and 600 ml of N2O for the initial 20 minutes and a flow of
500 ml of O2 and 500 ml of N2O thereafter.
 This has been shown to maintain the oxygen concentration between 33 and 40 %
at all times.
Management of the potent anaesthetic agents during
maintenance phase
 This is easily accomplished by dialling in the calculated
concentration on the plenum vaporizer for the flow being used.
 For example, suppose the anaesthetic uptake for a desired
concentration of 0.5% halothane is 7.5ml/min.
o If a FGF of 500ml/min is being used, then the dial setting should be
1.5% for at this setting and for the used flow, the total vapour
output would be 7.5ml/min.
o If a flow of 1000ml/min is being used, then the dial setting should
be 0.8%.
 In practice the actual dial setting often over estimates the actual
output since the plenum vaporizer under delivers the agent at low
flows.
Hence, the dial setting is fine-tuned depending on the endpoints
being achieved.
Weir and Kennedy recommend
infusion of halothane
Weir and Kennedy recommend infusion of
halothane (in liquid ml/hr) at the following rates
for a 50 kg adult at different time intervals.
• 0-5 min 27 ml/hr
• 5-30 min 5.71 ml/hr
• 30-60 min 3.33ml/hr
• 60-120 min 2.36 ml/hr
These infusion rates had been derived from the
Lowe's theory of the uptake of anaesthetic agent.
Salient points to be considered during the
maintenance phase
The other salient points to be considered
during the maintenance phase are the
following:
o a) Leaks must be meticulously sought for and prevented, flows must
be adjusted to compensate for the gas lost in the leaks.
o b) Most of the gas monitors sample gases at the rate of 200 ml/min,
which may be sometimes as high as half the FGF. Hence, care must be
taken to return the sample back to the circuit to maximize the economy of
FGF utilisation.
TERMINATION OF LOW FLOW
ANAESTHESIA.
 Unlike the initiation or the maintenance of the closed circuit,
termination is less controversial. There are only two recognized
methods of termination of the closed circuit.
 A. Use of high flow of gases: the circuit is opened and a high flow of gas is
used to flush out the anaesthetic agents .
 B . Use of activated charcoal.
o Activated charcoal when heated to 220oC adsorbs the potent vapours
almost completely. Hence, a charcoal-containing canister with a bypass is
placed in the circuit, towards the end of the anaesthesia, the gas is
directed through the activated charcoal canister.
o This results in the activated charcoal adsorbing the anaesthetic agent
resulting in rapid recovery and at the same time, reducing theatre
pollution. Nitrous oxide, due to its low solubility is washed off towards the
end by using 100% oxygen.
Advantages of LFA system.
o 1.Enormous financial savings due to use of low fresh gas
flows as well as the agent.
o 2. High humidity in the system leads to fewer post
anaesthetic complications.
o 3.Maintenance of body temperature during prolonged
procedures due to conservation of heat.
o 4. Reduction in the theatre pollution.
o 5. Environmental Pollution ( Ozone depletion/green house
effect)
Disadvantages of LFA
• Inability to quickly alter the inspired concentrations.
• More attention is required
• Danger of hypercarbia.
• Uncertainty about inspired concentrations
• Faster absorbent exhaustion
• Accumulation of undesirable trace gases in the system like CO,
acetone, methane, hydrogen, ethanol, argon, nitrogen,
compound A, etc.
Concerns about Safety in LFA
Points of concern in providing safe
anaesthesia are:
o Hypoxia.
o Misdosage of the volatiles.
o Exhaustion of the absorbent.
o Gas volume deficiency.
o Reduced controllability.
Absorber assembly :
An absorber assembly consists of
► A canister containing an absorbent.
► two ports for connection to breathing
tubes
► a fresh gas inlet.
► Other components that may be
mounted are inspiratory and expiratory
unidirectional valves , an adjustable
pressure limiting (APL) valve, and a bag
mount.
Canisters :
The absorbent is held in canisters.
► Single or two canisters in series can be used.
► Side walls are transparent so that color changes can be easily recognised.
► Comes in large and small sizes.
Absorbent
Types
► High alkali absorbents
High amount of NaOH or KOH
When desiccated react with anesthetic agent to form CO or compound-A
Do not change color when dry.
► Low alkali absorbents
Less amount of KOH or NaOH
Less amount of CO or compound-A when desiccated.
► Alkali free absorbent
Consists mainly Ca(OH)2 and other added agents No evidence of CO
formation
Little or no compound-A formation
Different formulations
► 1. Sodalime
2. Baralyme
3. Sofnolime
4. Amsorb plus
5. Ca(OH)2 lime etc,.
6. LiOH lime
► Sodalime is made of
► Ca(OH)2 - 80%
► NaOH – 4%
► KOH – 1%
► Water – 14% to 19 %
Silica or Kieselguhr – helps in hardening and reduced dust formation.
Indicator
► CO2 absorption is exothermic reaction
► Principle – acid base neutralization
►End product – carbonic acid and water.
CO2 + H2O → H2CO3
H2CO3 + 2NaOH → Na2CO3 (sodium carbonate)+ 2H2O
Na2CO3 + Ca(OH)2 → 2NaOH or 2KOH + CaCO3 .
Pattern of CO2
absorption
A.Unused absorbent in the canister.
B.After limited use- absorption of CO2 has occurred primarily at the inlet and to lesser
extent along the sides.
C.After extensive use – granules at the inlet and along the sides are exhausted.
D.Exhausted sodalime – few granules at the distal 1/3rd are still capable of absorbing
CO2.
E.Channeling effect
Indicators
color When fresh when exhausted
Phenolphthalein white Pink
Ethyl violet White Purple
Clayton yellow Red Yellow
Ethyl orange Orange Yellow
Mimosa Z Red White
► Shape and size of absorbent
pellets or granules form
Advantage - small granules increases the surface area, decreases the
gas channelling along low resistance pathway.
Disadvantage –
increased
resistance,caking.
4 to 8 mesh more commonly used.
REACTIONS BETWEEN ABSORBENTS AND
ANESTHETIC AGENTS
► Haloalkene Formation
Halothane degradation most often occurs during closed-circuit anesthesia and
produces the haloalkene 2-bromo-2- chloro-1, 1-difluoroethene
(BCDFE).Nephrotoxic in rats.
►Compound A Formation
Sevoflurane decomposes in the presence of carbon dioxide absorbents to
compound A. Compound A is vinyl ether and has dose dependent
nephrotoxic effects in rats.
Carbon monoxide formation
► when desflurane, enflurane, or isoflurane is passed through
desiccated absorbent containing a strong alkali (potassium or
sodium hydroxide)
► When sevoflurane is degraded by desiccated absorbent, carbon
monoxide is formed if the temperature exceeds 80°C.
Factors associated with carbon
monoxide formation:
► Absorbent Composition – NaOH , KOH
► Absorbent Desiccation ( dehydrated absorbent).
► Anesthetic Agent – Desflurane > Enflurane>Isoflurane .Halothane produces
very less amount of CO.
► Increased temperature Inside the Absorber.
► Fresh Gas Flow - FGF leads to more dessication of absorbent and more CO
production.
► Carbon Dioxide Absorption – Increases absorption leads to decreased CO
production.
The APSF recommendations to prevent absorbent
desiccation if the department continues to use
strong alkali absorbents with volatile anesthetic
agents:
► All gas flows should be turned OFF after each case.
► Vaporizers should be turned OFF when not in use.
► Canisters on an anesthesia machine that is commonly not used for a long
period of time should not be filled with absorbent that contains strong alkali
or should be filled with fresh absorbent before each use.
► The absorbent should be changed routinely, at least once a week, preferably
on a Monday morning, and whenever fresh gas has been flowing for an
extensive or indeterminate period of time.
► The canister should be labelled with the filling date.
► Checking this date should be part of the daily machine checklist.
► If a double-chamber absorber is used, the absorbent in both canisters
should be changed at the same time.
► The practice of supplying oxygen for administration to a patient who is not
receiving general anesthesia through the circle system should be strongly
discouraged.
► Using fresh gas to dry breathing system components should be
discouraged.
WHEN AND HOW TO CHANGE THE ABSORBENT
► Inspired Carbon Dioxide - Most reliable method ( indicated by capnograph)
► Indicator Color Change
► Heat in the Canister
 HUMIDITY
general term used to describe the amount of water vapor in a gas
may be expressed several ways:-
Absolute Humidity
- the mass of water vapor present in a volume of gas
- commonly expressed in milligrams of water per liter of gas.
Humidity at Saturation
-maximum amount of water vapor that a volume of gas can hold
- varies with the temperature.The warmer the temperature, the more water
vapor can be held in a gas
HUMIDIFICATION
 RELATIVE HUMIDITY
the amount of water vapor at a particular temperature expressed as a
percentage of the amount that would be held if the gas were saturated.
 DEW POINT
The temperature at which a gas is 100% saturated
 WATER VAPOR PRESSURE
Humidity may also be expressed as the pressure exerted by water vapor
in a gas mixture.
 normal breathing :upper respiratory tract warms humidifies, and filters
inspired gases, primarily in the nasopharynx
 Even at extremes of inspired temperature and humidity, gas that
reaches the alveolar level is 100% saturated at body temperature.
 ISOTHERMIC SATURATION BOUNDARY (ISB) : The point at which gases
reach alveolar conditions (37°C and 100% relative humidity)
 Over the course of a normal day, the respiratory tract loses
approximately 1470 J of heat and 250 mL of water.
IMPORTANCE OF HUMIDIFICATION
 Water is intentionally removed from medical gases so that gases delivered
from anaesthesia machine are dry and at room temperature
 Gases must therefore be warmed to body temperature and saturated
with water by upper respiratory tract
 normal nasal breathing : temperature in upper trachea is between 30-
33o C with RH-98% & AH-33 mg/L
: at alveolar level 37 degreeC , RH- 100% ,AH – 44 mg/L
 Tracheal intubation and high fresh gas flows bypass this normal
humidification by upper airways and expose lower airways to dry (< 10 mg
H2O/L), room temperature gases
EFFECTS OF INHALING DRY GASES
 DAMAGETOTHE RESPIRATORYTRACT
 Mucosa dries
 Temperature drops
 Secretions thicken  if not cleared atelectasis / airway obstruction
 Thick plugs  loci for infection
 Ciliary function reduced
 Surfactant activity impaired
 Mucosa more susceptible to injury
 Dry gases  bronchoconstriction
 Body heat loss
Absorbent desiccation
Trachealtube obstruction
SOURCES OF HUMIDITY
 CO2 absorbent – reaction ofCO2 with absorbent releases water
-- absorbent granules also contain water
 Exhaled gases – some rebreathing is present in tracheal tube, supraglottic
airway device, and connections to breathing system
-almost half of humidity in expired gases is preserved in this manner
 Moistening/Rinsing breathing tubes and reservoir bag before use
 Low fresh gas flows – conserve moisture
 Coaxial breathing circuits – increase humidity more quickly than a system with
2 separate limbs, when combined with low flows
-not very efficient : Bain system (coaxial Version of Mapleson D) does not meet
optimal humidification requirements because of high fresh gas flow required
 HEAT AND MOISTURE
EXCHANGERS –
 hydrophobic
 hygroscopic
 HUMIDIFIERS – pass-over/blow- by/bubble-
through/vapor-phase humidifier
 HEATED
 UNHEATED
 Nebulisers
SOURCESOFHUMIDITY
HME
•SYNONYMS :- condenser
humidifier, swedish nose,
artificial nose, nose humidifier,
passive humidifier, regenerative
humidifier, moisture exchanger,
and vapor condenser
•When combined with filter for
bacteria & viruses :- HMEF
 disposable devices
 exchanging medium enclosed in a plastic housing
 vary in size and shape
 Each has a 15-mm female connection port at the patient
end and a 15-mm male port at the other end; patient port
may also have a concentric 22-mm male fitting
 May have a port to attach the gas sampling line for a
respiratory gas monitor or an oxygen line
 One type utilizes a ceramic heating element with a water
input port, a membrane, and an aluminum grid that
vaporizes the water
 Preferred for short term use.
Hydrophobic
 hydrophobic membrane with small
pores
 membrane is pleated to increase the
surface area
 provides moderately good inspired
humidity
 may be impaired by high ambient
temperatures
 efficient bacterial and viral filters
 prevent the HCV from passing
 allow the passage of water vapor but
not liquid water at usual ventilatory
pressure
 associated with small increases in
resistance even when wet
Hygroscopic
 contain a wool/foam/paperlike material
coated with moisture-retaining chemicals
 medium may be impregnated with a
bactericide
 Composite hygroscopic HMEs -- a
hygroscopic layer plus a layer of thin,
nonwoven fiber membrane that has been
subjected to an electrical field to increase its
polarity -- improves filtration efficiency and
hydrophobicity.
 composite hygroscopic HMEs are more
efficient than hydrophobic ones
 lose their airborne filtration efficiency if they
become wet; microorganisms held by the
filter medium can be washed through the
device
 Their resistance can increase greatly when
wet
Type
Heat and moisture
exchangingefficiency
Hygroscopic
Excellent
Hydrophobic
Good
Effect of increased tidal
volume on heat and
moisture exchange
Slightdecrease Significantdecrease
Filtrationefficiency
when dry
Good Excellent
Filtrationefficiency
when wet
Poor Excellent
Resistance when dry Low Low
Resistance when wet Significantlyincreased Slightlyincreased
Effect of nebulized
medications
Greatly increased
resistance
Little effect
 CONTAINDICATIONS
 thick, copious, or bloody secretions
 expired tidal volume less than 70% of the delivered tidal volume (eg, those
with large bronchopleurocutaneous fistulas or incompetent or absent
endotracheal tube cuffs)
 body temperatures less than 32°C
 high spontaneous minute volumes (> 10L/min)
 Spontaneously breathing, difficult to wean patients – relativeC/I
 MOISTUREOUTPUT
•Amount of Heat & Moisture provided.
•Currently no standards for minimum MO for HME’s
•Depends upon –TV, insp time, RR, system leak, initial humidity.
ADVANTAGES DISADVANTAGES
 Temp preservation
 Dead space rebreathing
 IncreasedWOB
 Resistance
 ETT obstruction
 Not for all patients
 Disconnection
 CO2 absorbent dessication
 Avoid using with nebulizers
 Short term use
 Inexpensive
 Easy to use
 Portable,
 light weight
 Less risk of contamination
 Efficient filter
 Do not require external heat,
water, temp monitoring
 No condensate- thus no
ventilator malfunction
HAZARDS
 ineffective low-pressure alarm during disconnection due to
resistance through HME.
 Suspected high resistance  Peak airway pressure should be
measured withAND without HME in place
 possible increased resistive work of breathing due to mucus plugging
of HME
 In Mapleson system – resistance causes diversion of FGF down the
expiratory limb.
 hypothermia
 airway obstruction– ETT blockage/ ETT kinking/ detachment
 Should not be used with Heated Humidifiers
 nebulized medications increase resistance of hygroscopic HME.
•Ineffective filtration:- liquid can break through a hygroscopic HME
•Foreign body aspiration – detached parts inhaled
•Rebreathing due to dead space (more in paed)
•Hypothermia
•Leaks & disconnections
•dryCO2 absorbent- absorb volatile agents, hence delay in induction.
POSITION INCIRCUIT : -
Patient – HME – filter – sampling line – machine/ ventilator
HUMIDIFIERS
- Add water to gas by passing the gas over a water chamber
(passover) / through a saturated wick (wick humidifier) /bubbling
it through water (bubble-through) / mixing it with vaporized
water (vapor-phase)
- do not filter respiratory gases
 UNHEATED
 Disposable, bubble-through
devices used in oxygen supplied to
patients via facemask or nasal
canula
 Simple containers containing
distilled water through which
oxygen is passed and it gets
humidified
 Maximum humidity that can be
achieved is 9mg H2O/L
 HEATED
 Incorporate a device to warm
water in the humidifier, some
also heat inspiratory tube
 Humidification chamber
 Heat source
 Temperature monitor
 Thermostat
 Inspiratorytube
UNHEATED HEATED
HEATED HUMIDIFIERS
 Humidification chamber – contains liquid water, disposable/
reusable, clear (easy to check water level)
 Heat source – heated rods immersed in water/ plate at bottom of
humidification chamber
 Temperature monitor – to measure gas temperature at patient end
of breathing system
 Thermostat –
1. Servo-controlled units – automatically regulates power to heating
element in response to temperature sensed by a probe near
patient connection/ humidifier outlet
2. Non servo-controlled units – provides power to heating element
according to setting of a control, irrespective of delivered
temperature
 Inspiratory tube – conveys humidified gas from humidifier outlet
to patient
 If unheated  gas will cool and lose some of its moisture as it
travels to the patient, water trap necessary to collect condensed
water
 If Heated or insulated  more precise control of temperature
and humidity delivered to patient, avoids moisture rainout
 Controls – most allow temperature selection at end of delivery
tube or at humidification chamber outlet
 Alarms – to indicate temperature deviation by a fixed amount,
displacement of temperature probe, disconnection of heater
wire, low water level in humidification chamber, faulty airway
temperature probe , lack of gas flow in the circuit
 Standard requirements -An international and aU.S. standard on
humidifiers have been published
 In circle system, heated humidifier is placed in the inspiratory limb
downstream of unidirectional valve by using an accessory
breathing tube
 Must not be placed in the expiratory limb
 Filter, if used, must be placed upstream of humidifier to prevent it
from becoming clogged
 In Mapleson systems, humidifier is usually placed in fresh gas
supply tube
 Humidifier must be lower than patient to avoid risk of water
running down the tubing into the patient
 Condensate must be drained periodically or a water trap
inserted in the most dependent part of the tubing to prevent
blockage or aspiration
 Heater wire in delivery tube should not be bunched, but strung
evenly along length of tube
 Delivery tube should not rest on other surfaces or be covered
with sheets, blankets, or other materials; a boom arm or tube
tree may be used for support
 Advantages –
1. Capable of delivering saturated gas at body temperature or
above, even with high flow rates
2. More effective humidification than an HME
 Disadvantages –
1. Bulky and somewhat complex
2. Involve high maintenance costs, electrical hazards, and
increased work (temperature control, refilling the reservoir,
draining condensate, cleaning, and sterilization)
3. Offers relatively little protection against heat loss during
anaesthesia as compared to circulating water and forced-air
warming
NEBULISERS
 Aerosol generators/ atomizers/ nebulizing humidifiers
 Emit water in the form of an aerosol mist (water vapor plus
particulate water)
 Used for producing humdification and delivery of drug directly into
respiratory tract
 Drugs delivered by nebulizers – Bronchodilators, decongestants,
mucolytic agents, steroids
 Optimal particle size of droplet (aerosol) = 0.5 to 5 µm
 Particles > 5 µm – unable to reach peripheral airways, deposited in
main airways
 Particles < 0.5 µm - very light, come back with expired gases
without being deposited in airways
 Most commonly used 2 types –
1. Pneumatically driven (gas-driven, jet, high pressure, compressed
gas)
2. Ultrasonic
 Pneumatically driven nebulizer works by pushing a jet of high-
pressure gas into a liquid, inducing shearing forces and breaking
the water up into fine particles
 Should be placed in the fresh gas line (high flow of gas must be
used with pneumatic nebulizer)
 Produces particles of size 5 to 30 µm (only 30 to 40% of particles
produced are in optimal range)  most of the particles get
deposited in wall of main airways
 Ultrasonic nebulizer
 produces a fine mist by subjecting the liquid to a high-frequency,
electrically driven resonator
 Can be used in the fresh gas line or the inspiratory limb (No need
for a driving gas)
 Frequency of oscillation determines the size of the droplets
 Creates a denser mist than pneumatic ones
 produces mist with aerosol size of 1 to 10 µm (95% of particles
produced are in optimal range)  particles get deposited directly
in airways  very useful for delivery of bronchodilators directly in
peripheral airways
 Can nebulize 6 mL of water or drug in 1 minute
HAZARDS
1. Nebulized drugs may obstruct an HME or filter in the
breathing system
2. Overhydration
3. Hypothermia
4. Transmission of infection
5. Case reports where a nebulizer was connected
directly to a tracheal tube without provision for
exhalation  resulted in pneumothorax in one case
 Advantage - can deliver gases saturated with water
without heat and, if desired, can produce gases carrying
more water
 Disadvantages –
1. Somewhat costly
2. Pneumatic nebulizers require high gas flows
3. Ultrasonic nebulizers require a source of electricity and
may present electrical hazards
4. May be considerable water deposition in the tubings,
requiring frequent draining, water traps in both the
inspiratory and exhalation tubes, and posing the dangers
of water draining into the patient or blocking the tubing
INTRODUCTION :
● Tracelevel- concofanestheticgasthat isfar belowtheone thatisneededfor clinical
anesthesia.
● Variesgreatly – dependingonFGF,ventilationsystem, thelengthoftime ofanesthesia,
measurementsite andother variables
● Degreeofpollutionismore duringpaediatricsurgeries
SCAVENGING SYSTEM
- Hospital-basedandstand-aloneoperatingrooms,
- Recoveryrooms,
- Dentaloperatories,
- V
eterinaryfacilities.
Workplace exposures:
● Twodifferent classesof
chemicals:
nitrousoxide and
halogenated agents
● In1977NationalInstitute for
Occupational SafetyandHealth
(NIOSH) –
recommendedexposurelimits
(RELs) for both nitrousoxide and
halogenatedagent
Scavenging and Waste Anesthetic
Gases (WAGs)
SCAVENGING
● Definition
● Scavengingisthe collectionand
removal ofvented anesthetic
gases from the OR.
● Since the amount ofanestheticgas
supplied usuallyfar exceedsthe
amount necessary for the patient,
OR pollution isdecreased by
scavenging.
● Scavengerandoperating room
ventilationefficiencyare the two
most important factors in reduction
ofwaste anestheticgases(WAGs).
● Theuseofscavengingdeviceswith anesthesiadeliverysystemsisthe
most effectivewayto decrease waste anestheticgases.
● Anefficientscavengingsystemiscapableof reducing ambient
concentrations ofwastegasesbyup to 90%.
● Anesthesiamachinesandbreathingsystemsdelivering
halogenatedhydrocarbonanestheticsand/ornitrous
oxideshouldnotbeoperatedunlesstheyareequipped
withafunctional scavengingsystem
.
● Tobe effective,ascavengingsystemmustnot leakandmustcontrol
theconcentrationoftrace anestheticsin ambient air.
● Wastegasesshould not bedischarged into the outsideair in an
areawhere reentryinto the buildingislikely.
Types
● Scavengingmaybe active(suctionapplied) or passive(waste gasesproceed
passivelydown corrugated tubing through the room ventilation exhaust
grill ofthe OR).
● Activesystemsrequire ameansto protect the patient's airwayfrom
the applicationofsuction, or buildup ofpositive pressure.
● Passivesystems require that the patient be protected from
positivepressure buildup only.
● The 2 majorcausesof wastegascontamination in the O.R :
1. The anesthetic technique used and
2. The equipment issues
Regardinganesthesiatechnique the followingwill increase anesthetic gas
pollution wheninhalantanestheticsare used:
● Administeringinhalantanestheticsbyopen drop (eg, periodically
drippingliquid volatileanesthetic onto agauzesponge) or
insufflation(eg, delivery ofarelativelyhighflowof anesthetic in
oxygeninto the tracheaor pharynx throughacatheter) techniques.
● Turning on flowmetersandvaporizersbefore attachingthe
breathingsystemto the patient.
● Allowingflowmetersandvaporizersto remain on after the patient
isdisconnected fromthe breathingsystem.
● Using uncuffed endotracheal tubes that do not create a
completely sealed airway or using cuffed tubes without inflating
the cuff.
● Disconnecting apatient from abreathing system without eliminating as
much of the residual gases as reasonably possible through the
scavenging system.
● Spillingliquidanesthetic duringthe filling ofvaporizers,especially
duringan anestheticprocedure.
● Poorlyfittingmasks.
● Flushingthe circuit.
● Use ofbreathing circuits that are difficult to scavenge, suchasJackson-
Rees.
Componentsof thescavengersystem:
● Gascollection assembly.
● Transfer tubing.
● Scavenginginterface .
● Gasdisposal tubing.
● Gasdisposal assembly.
1.Gas collecting assembly
● Capturesexcessanesthetic
gases and deliversit to the
transfer tubing.
● WAGare vented fromanesthesia
systemthrough either
adjustable pressure limiting
valveor ventilator reliefvalve.
● Gaspassingthroughthese valves
accumulatesin the gas
collecting assemblyand is
directed to the transfer
means.
2.Transfer tubing
● The transfer tubescarriesexcessgasfromgascollecting assembly
to scavenging interface.
● The tubes mustbe either 19 or 30 mm, sometimesyellowcolor-coded .
● The tubesshould be sufficientlyrigidto prevent kinkingand asshort as
possibleto minimizethe chanceofocclusion.
● Separatetubes from the APLvalveandventilator relief valvemerge into a
singlehosebefore they enter scavenginginterface.
● Ifthe transfer meansisoccluded, baselinebreathing circuit pressure
willincrease and barotrauma can occur.
3.Scavenging interface
● The scavengerinterfaceisthe most important component because
it protects the breathing circuit or ventilator from excesspositiveor
negativepressure.
● Positive-pressurereliefismandatory, irrespective ofthe typeof
disposable systemused, to vent excessgasin caseofocclusiondistal
to interface.
● scavenger interfacesmaybe open
closed
● B&Cclosed interfaces
● A&Bare active systems
● Cispassive system
OPEN INTERFACE:
● It containsno valvesand isopen to
the atmosphere, allowingpositive
and negativepressure relief.
● Open interfacesshould be used
only withactive disposable systems
that haveacentralevacuation
system.
● open interfaces require areservoir
becausewastegasesare
intermittently discharged insurges
whereasflow fromthe evacuation
systemis continuous.
● The efficiencyofit dependson several factors:
A. Thevacuum flowrate per min must equalor exceed
the minute volume ofexcessgasesto prevent spillage.
B. Spillagewill occur ifthe volume ofasingle exhaled
breathexceedsthe capacityofreservoir.
Openinterfaces
aresaferforthepatients.
● CLOSEDINTERFACE:
● It communicateswith the atmosphere throughvalves.
● Twotypesofclosedinterfacesare commerciallyavailable:
positive pressure reliefonly
positive and negativepressure relief
POSITIVEPRESSURERELIEFONLY:
● It hasasingle positive pressure relief valve and is designated to be used
onlywithpassivedisposable systems.
● Transferofthe wastegasfromthe interface to the disposablesystem
reliesonweightor pressureofthe wastegasitselfbecauseanegative
pressure evacuation systemisnot used.
● In thissystemreservoir bagisnot required.
POSITIVEANDNEGA
TIVEPRESSURERELIEF:
● It haspositive and negative pressure relief valvein addition to areservoir
bag.
● It isused with active disposable systems.
Theeffectivenessofaclosed systeminpreventing spillagedependson:
the rate ofwaste gasinflow
the evacuation flow
rate the size of the
reservoir
● Leakage occurs only when the reservoir bag becomes fully inflated and
pressure increases sufficiently to open the positive pressure relief valve
.
● In contrast, the effectiveness of aopen system in preventing spillage
depends not only on the volume of the reservoir but also on the
4.Gas disposal tubing
● The gas disposable tubing conducts waste gas from the Scavenging
interface to the gas disposable assembly.
● It should be collapse proof and should run overhead, if possible to
minimize the chance of accidental occlusion.
● It ultimately eliminates excess waste gas.
● It is of two types: Active and Passive
● Active assembly:
● most commonly used. It uses central evacuation system.
● A vacuum pump serves as mechanical flow inducing device that removes the
waste gases.
● An interface with a negative pressure relief valve is mandatory because the
pressure within the system is negative.
● A reservoir is very desirable and the larger the reservoir, the lower the
suction flow rate needed.
5.Gas disposal assembly
- PipedVacuum
- Active duct system - employsflow inducingdevices(fans,
pumps, blowersetc..) to move evenat low pressures
Advantages
● Convenientin largehospitals, where manymachinesare in
use in different locations
Disadvantages
● Vacuumsystemandpipe workisamajor expense
● Needlevalve mayneedcontinualadjustment
● Passivedisposablesystem
:
● doesnot use amechanical flowinducingdevice.
● The weightor pressure from the heavier-than-air anesthetic
gasesproducesflowthroughthe system.
● Positive pressure reliefismandatory, but anegative pressure reliefand
reservoir are unnecessary.
● Passive– simpler and lessexpensivebut not effective
- Roomventilation system
- Nonrecirculatingor – Recirculating
- Pipingdirect to atmosphere
- Adsorption devices
- Catalyst decomposition
Charcoal canisters :
● activated charcoal
● connected to the outlet ofthe breathing system
● removeshalogenated anaestheticsby filtration
CatalyticDecomposition :
• Catalyticdecompositioncanbe
used
to convert nitrousoxide to nitrogen
andoxygen, reducingits
contribution to the greenhouse
effect
Advantages
● No set-up cost
● Mobile - moveswiththe
machine
Disadvantages
● Continuingcost ofreplacement.
● Continuosreplacement ismessy
● Doesnot remove nitrousoxide
● Heatingthe canister causesthe release of
the
inhalational
agents
Evaluation of Anesthetic Equipment
● Eachpieceofequipment involved in the deliveryofinhalant
anestheticsshouldbe evaluated regularly to assureits function and
integrity.
Proceduresforcheckoutofanesthesiaequipment,dependingontheequipmentto
beused,shouldincludethefollowing:
● Statusofthehigh-pressuresystem,includingtheoxygensupplyandnitrousoxide
supply- The nitrous oxide supply should not leak when the cylinder valve
ison and the nitrousoxideflowmeter isoff.
● Statusofthelow-pressuresystem(flowmeterfunction)- Anegative-pressure
leaktest should be performed at the commongasoutlet or the outlet
ofthe vaporizer immediatelyupstreamfromthe breathingsystem.
● Statusofthebreathingsystem- Anappropriateleaktest for acircle system
andNoncirclesystemscanusuallybe done. The quantityofleakagecan
be measuredbydeterminingthe flowrate ofoxygen necessaryto
maintainaconstant pressureinthe system,andthe leakrate should be
lessthan300ml/ minat 30cmofh2O.
● Statusofthescavengingsystem- Thescavengingsystemshouldbeproperly
attachedat allconnectors, andthe appropriate vacuumshould be
assured for activesystems.Ifcharcoalcanistersare employedfor
scavenging,theyshould bechangedat appropriateintervals.
● Statusofmechanicalventilators- Ventilatorsshouldbeconnected properly
to the anesthesiamachine, andanabsence ofleaksshould be assured.
Monitoring of the Effectiveness of
Antipollution Techniques
● Monitoring trace-gas concentrations in the workplace provides a
quantitative assessment of the effectiveness of a waste-gas
control program.
● Measuringthe concentration ofanesthetic in the breathing zoneof
the most heavilyexposed workersisthe usual procedure.
● Anair-monitoring programismostappropriately startedafter
anesthesiadeliverysystemshavebeen equipped with scavenging
systemsandafter other techniquesfor minimizingwastegaspollution
are in place.
● Anidealapproachwouldincludefrequentairmonitoring,preferably at least
semiannual evaluations.
Effectiveness
● Unscavengedoperatingrooms show10-70 ppmhalothane, and 400-
3000ppmN2O.
● scavengingbrings these levels downto 1 and60 ppmrespectively.
● Addingcarefulattention to leaksandtechniquecanyieldlevelsaslowas
0.005 and 1 ppm.
HAZARDS
● Scavengingsystemfunctionallyextendsthe anesthesiacircuit all the way
fromthe anesthesiamachine to the disposable site.
● Obstruction of scavengingpathwaycancauseexcessivepositivepressure in
the breathing circuit andbarotraumacanoccur.
● excessivevacuum appliedcanresult in undesirablenegativepressures
withinbreathingsystem.
● LossofMonitoringInput – it maymaskthe strong odor ofavolatile
anesthetic agent, delaying recognition ofanoverdose
● Alarmfailure – neg. pressure from the scavengingsysteminterface
prevent the bellows from collapsingwhen breathing systemis
disconnected . Low
airwaypressure alarmnot activated.
Avoiding waste gas exposure
● Good maskfit.
● Avoidunscavengeable techniquesifpossible(insufflation).
● Prevent flowfrom breathing systeminto room air (onlyturn on agent
andnitrous oxide after maskison face, turn them offbefore suctioning).
● Washoutanesthetics(into the breathing circuit) at the end of
the anesthetia.
● Don’t spill liquidagent.
● Use low flows.
● Usecuffedtracheal tubes whenpossible.
● Checkthe machine regularlyfor leaks.
● Disconnectnitrous oxide pipeline connection at wallat the day’send.
● Total intravenousanesthesia.
THANK U….

More Related Content

Similar to lowflowsystem-200601004249.pptx

Low flow anesthesia 3 13-2017 (2)
Low flow anesthesia 3 13-2017 (2)Low flow anesthesia 3 13-2017 (2)
Low flow anesthesia 3 13-2017 (2)Helga Komen
 
High frequency oscillatory ventilation
High frequency oscillatory ventilationHigh frequency oscillatory ventilation
High frequency oscillatory ventilationgnivri1666
 
bains circuit copy copy.pptx
bains circuit copy copy.pptxbains circuit copy copy.pptx
bains circuit copy copy.pptxTushar Mankar
 
Oxygenation – Peri intubation, Apnoeic, THRIVE - Copy.pptx
Oxygenation – Peri intubation, Apnoeic, THRIVE - Copy.pptxOxygenation – Peri intubation, Apnoeic, THRIVE - Copy.pptx
Oxygenation – Peri intubation, Apnoeic, THRIVE - Copy.pptxRitesh Pandey
 
Respiratory changes during anesthesia and ippv
Respiratory changes during anesthesia and ippvRespiratory changes during anesthesia and ippv
Respiratory changes during anesthesia and ippvImran Sheikh
 
BREATHING CICUITS.pptx
BREATHING CICUITS.pptxBREATHING CICUITS.pptx
BREATHING CICUITS.pptxmohamedrifky10
 
Oxygen therapy principles_and_practice shahna ali
Oxygen therapy principles_and_practice shahna ali Oxygen therapy principles_and_practice shahna ali
Oxygen therapy principles_and_practice shahna ali Shahnaali
 
breathing system anaesthesia by Dr Prakriti Maiti
breathing system anaesthesia by Dr Prakriti Maitibreathing system anaesthesia by Dr Prakriti Maiti
breathing system anaesthesia by Dr Prakriti MaitiPrakriti88
 
Sam ppt on effect of anaesthesia on respiratory system
Sam  ppt on effect of anaesthesia on respiratory systemSam  ppt on effect of anaesthesia on respiratory system
Sam ppt on effect of anaesthesia on respiratory systemRanjana Meena
 
Anesthesia machine
Anesthesia machine Anesthesia machine
Anesthesia machine dtgvyukj
 
5) gas exchange and diffusion
5) gas exchange and diffusion5) gas exchange and diffusion
5) gas exchange and diffusionAyub Abdi
 
respiration notes.pdf
respiration notes.pdfrespiration notes.pdf
respiration notes.pdfDanjaarDasan
 
INHALATIONAL AGENTS power point presentation
INHALATIONAL  AGENTS power point presentationINHALATIONAL  AGENTS power point presentation
INHALATIONAL AGENTS power point presentationSANDEEPKOTA22
 

Similar to lowflowsystem-200601004249.pptx (20)

Lfa
LfaLfa
Lfa
 
Low flow anesthesia 3 13-2017 (2)
Low flow anesthesia 3 13-2017 (2)Low flow anesthesia 3 13-2017 (2)
Low flow anesthesia 3 13-2017 (2)
 
High frequency oscillatory ventilation
High frequency oscillatory ventilationHigh frequency oscillatory ventilation
High frequency oscillatory ventilation
 
bains circuit copy copy.pptx
bains circuit copy copy.pptxbains circuit copy copy.pptx
bains circuit copy copy.pptx
 
Oxygenation – Peri intubation, Apnoeic, THRIVE - Copy.pptx
Oxygenation – Peri intubation, Apnoeic, THRIVE - Copy.pptxOxygenation – Peri intubation, Apnoeic, THRIVE - Copy.pptx
Oxygenation – Peri intubation, Apnoeic, THRIVE - Copy.pptx
 
HFOV in Neonates.pptx
HFOV in Neonates.pptxHFOV in Neonates.pptx
HFOV in Neonates.pptx
 
Breathing Circuits.pptx
Breathing Circuits.pptxBreathing Circuits.pptx
Breathing Circuits.pptx
 
Respiratory changes during anesthesia and ippv
Respiratory changes during anesthesia and ippvRespiratory changes during anesthesia and ippv
Respiratory changes during anesthesia and ippv
 
BREATHING CICUITS.pptx
BREATHING CICUITS.pptxBREATHING CICUITS.pptx
BREATHING CICUITS.pptx
 
Oxygen therapy principles_and_practice shahna ali
Oxygen therapy principles_and_practice shahna ali Oxygen therapy principles_and_practice shahna ali
Oxygen therapy principles_and_practice shahna ali
 
Breathing systems
Breathing systemsBreathing systems
Breathing systems
 
breathing system anaesthesia by Dr Prakriti Maiti
breathing system anaesthesia by Dr Prakriti Maitibreathing system anaesthesia by Dr Prakriti Maiti
breathing system anaesthesia by Dr Prakriti Maiti
 
Sam ppt on effect of anaesthesia on respiratory system
Sam  ppt on effect of anaesthesia on respiratory systemSam  ppt on effect of anaesthesia on respiratory system
Sam ppt on effect of anaesthesia on respiratory system
 
Anesthesia machine
Anesthesia machine Anesthesia machine
Anesthesia machine
 
5) gas exchange and diffusion
5) gas exchange and diffusion5) gas exchange and diffusion
5) gas exchange and diffusion
 
Final circuits
Final circuitsFinal circuits
Final circuits
 
Resp.pptx
Resp.pptxResp.pptx
Resp.pptx
 
respiration notes.pdf
respiration notes.pdfrespiration notes.pdf
respiration notes.pdf
 
INHALATIONAL AGENTS power point presentation
INHALATIONAL  AGENTS power point presentationINHALATIONAL  AGENTS power point presentation
INHALATIONAL AGENTS power point presentation
 
Vaporizers
VaporizersVaporizers
Vaporizers
 

Recently uploaded

Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...
Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...
Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...narwatsonia7
 
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiCall Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiNehru place Escorts
 
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.MiadAlsulami
 
Call Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls Service
Call Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls ServiceCall Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls Service
Call Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls Servicenarwatsonia7
 
Call Girl Coimbatore Prisha☎️ 8250192130 Independent Escort Service Coimbatore
Call Girl Coimbatore Prisha☎️  8250192130 Independent Escort Service CoimbatoreCall Girl Coimbatore Prisha☎️  8250192130 Independent Escort Service Coimbatore
Call Girl Coimbatore Prisha☎️ 8250192130 Independent Escort Service Coimbatorenarwatsonia7
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalorenarwatsonia7
 
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...Miss joya
 
(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...
(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...
(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...indiancallgirl4rent
 
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on DeliveryCall Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Deliverynehamumbai
 
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...astropune
 
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls ServiceKesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Servicemakika9823
 
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...Miss joya
 
Call Girl Number in Panvel Mumbai📲 9833363713 💞 Full Night Enjoy
Call Girl Number in Panvel Mumbai📲 9833363713 💞 Full Night EnjoyCall Girl Number in Panvel Mumbai📲 9833363713 💞 Full Night Enjoy
Call Girl Number in Panvel Mumbai📲 9833363713 💞 Full Night Enjoybabeytanya
 
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...Garima Khatri
 
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Miss joya
 
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...Taniya Sharma
 
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...narwatsonia7
 
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiRussian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiAlinaDevecerski
 

Recently uploaded (20)

Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...
Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...
Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...
 
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiCall Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
 
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
 
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
 
Call Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls Service
Call Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls ServiceCall Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls Service
Call Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls Service
 
Call Girl Coimbatore Prisha☎️ 8250192130 Independent Escort Service Coimbatore
Call Girl Coimbatore Prisha☎️  8250192130 Independent Escort Service CoimbatoreCall Girl Coimbatore Prisha☎️  8250192130 Independent Escort Service Coimbatore
Call Girl Coimbatore Prisha☎️ 8250192130 Independent Escort Service Coimbatore
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
 
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
Russian Call Girls in Pune Tanvi 9907093804 Short 1500 Night 6000 Best call g...
 
(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...
(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...
(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...
 
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on DeliveryCall Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
 
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
 
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls ServiceKesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
 
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
 
Call Girl Number in Panvel Mumbai📲 9833363713 💞 Full Night Enjoy
Call Girl Number in Panvel Mumbai📲 9833363713 💞 Full Night EnjoyCall Girl Number in Panvel Mumbai📲 9833363713 💞 Full Night Enjoy
Call Girl Number in Panvel Mumbai📲 9833363713 💞 Full Night Enjoy
 
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
 
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
 
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCREscort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
 
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
 
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
 
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiRussian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
 

lowflowsystem-200601004249.pptx

  • 1. LOW FLOW ANAESTHESIA SYSTEM/ CO2 ABSORBER/ HUMIDIFICATION/ SCAVENGING SYSTEM
  • 2. History of Low flow Anaesthesia  As early as 1850 John Snow (1813-53) recognized that only a small fraction that inhalation anaesthetics gets metabolized and most of it exhaled unchanged in the expired air of the anaesthetized patients. He also concluded that the narcotic effect of the volatile anaesthetics could be markedly prolonged by re-inhaling these unused vapours.  About 75 years later, in 1924, the rebreathing technique with carbon dioxide absorption was introduced into routine clinical practice by Ralph Waters (1883-1979), the to-and-fro absorption system.  Brian C. Sword introduced the circle (CO2 absorption) system in the 1930s.
  • 3. LOW FLOW ANESTHESIA • Any technique that utilises a fresh gas flow (FGF) that is less than the alveolar ventilation can be classified as „Low flow anaesthesia‟. • Baum et al : had defined it as a technique wherein at least 50% of the expired gases had been returned to the lungs after carbon dioxide absorption in the next inspiration. • Baker had classified the FGF used in anaesthetic practice into the following categories: • Metabolic flow : about 250 ml /min • Minimal flow : 250-500 ml/min. • Low flow : 500- 1000 ml/min. • Medium flow : 1 - 2 l/min. • For most practical considerations, utilization of a fresh gas flow less than 2 litres/min may be considered as low flow anaesthesia.
  • 4. CONT. • there exists a need for a system that provided the advantages of the completely closed circuit and at the same time, reduced the dangers associated with it. • Low flow anaesthesia fulfilled these requirements. • Low flow anaesthesia involves utilizing a fresh gas flow which is higher than the metabolic flows but which is considerably lesser than the conventional flows. • The larger than metabolic flows provides for considerably greater margin of safety and satisfactory maintenance of gas composition in the inspired mixture. • THE NEED FOR LOW FLOW ANAESTHESIA • Completely closed circuit anaesthesia is based upon the reasoning that anaesthesia can be safely maintained if the gases which are taken up by the body alone are replaced into the circuit taking care to remove the expired carbon dioxide with sodalime. No gas escapes out of the circuit and would provide for maximal efficiency for the utilisation of the fresh gas flows.
  • 5. Requirements for LFA 1. A circle system with CO2 absorption. 2. A flow meter for the adjustments of FGF below 1 l/min. 3. Gas-tight breathing system, recommended test leakage should be below 150 ml/min. 4. The breathing system must be having minimal internal volume and a minimum number of components and connections. 5. Continuous gas monitoring must be employed, essential in controlling the patient's alveolar gas concentrations. 6. The vaporizers delivering the high concentrations must be calibrated and must be accurate at low fresh gas flow.
  • 6. Circle System • The minimum requirement for conduct of low flow anaesthesia is the effective absorption of CO2 from the expired gas, so that the CO2 free expired gas can be reutilized for alveolar ventilation. • In the past, two systems were in use; “To and Fro” system introduced by Waters and the “circle system” introduced by Brian Sword. • But because of its bulkiness near the patient and other disadvantages the “To and Fro” has gone out of favour and the “circle system” using large sodalime canisters is in use.
  • 7. Describing a Circle System. The circle system should have the basic configuration with two unidirectional valves on either side of the sodalime canister, fresh gas entry, reservoir bag, pop off valve, and corrugated tubes and „Y‟ piece to connect to the patient. The relative position of fresh gas entry, pop off valve, and reservoir bag are immaterial as long as they are positioned between the expiratory and the inspiratory unidirectional valves that functions properly and CO2 absorption is efficient at all times.
  • 8. THE PRACTICE OF LOW FLOW ANAESTHESIA: The practice of low flow anaesthesia can be dealt with under the following three headings: o 1. Initiation of Low flow anaesthesia o 2. Maintenance of Low flow anaesthesia o 3. Termination of Low flow anaesthesia.
  • 9. INITIATION OF LOW FLOW ANAESTHESIA Primary aim at the start of low flow anaesthesia is to achieve an alveolar concentration of the anaesthetic agent that is adequate for producing surgical anaesthesia (approximately 1.3 MAC).  Factors influencing to buildup of alveolar concentration should be considered . These factors can broadly be classified into three groups :  1) Factors governing the inhaled tension of the anaesthetic agent,  2) Factors responsible for rise in alveolar tension,  3) Factors responsible for reducing the alveolar tension, uptake from the lungs.
  • 10. Factors governing the inhaled tension of the anaesthetic. Three most important factors responsible for inhaled tension of anesthetics are :  1.BREATHING CIRCUIT VOLUME  2. RUBBER GAS SOLUBILITY 3. SET INSPIRED CONCENTRATION
  • 11. Breathing Circuit Volume  Circle system is often bulky and volume roughly equal to 6-7 liters, FRC of the patient, which is roughly 3 litres, together constitutes a reserve volume of 10 liters; to which the anaesthetic gases and vapours have to be added. • With the addition of FGF, the rate of change of composition of the reserve volume is exponential, time required for the changes to occur is governed by the time constant, which is equal to this reserve volume divided by the fresh gas flow. • Three time constants are needed for a 95% change in the gas concentration to occur. • Hence, if a FGF of 1L/min is used, then 30 minutes will be required for the circuit concentration to reflect the gas concentration of the FGF. • If the FGF is still lower, then correspondingly longer time will be required.
  • 12. Rubber Gas Solubility of the agent • The anaesthetic agent could be lost from the breathing system due to solubility of the agent in rubber, and permeability through the corrugated tubes. • The concentration will take a longer time if solubility is more. • Though the amount of loss will be minimal, it should be considered at the start if the aimed anaesthetic concentration is low.
  • 13. SET INSPIRED CONCENTRATION  The rate of rise of alveolar partial pressure of the anaesthetic agent must bear a direct relationship to the inspired concentration.  Higher the inspired concentration, the more rapid is the rise in alveolar concentration.  The Inspired concentration of the agent depends on several factors. – Denitrogenation of the circuit volume. – Alveolar ventilation. – Concentration effect.
  • 14. Denitrogenation of FRC & Breathing Circuit Volume The functional residual capacity of the lung and the body as a whole contain nitrogen which will try to equilibrate with the circuit volume and alter the gas concentration if satisfactory denitrogenation is not achieved at the start of anaesthesia.  Hence, as a prelude to the initiation of closed or low flow anaesthesia, thorough denitrogenation must be achieved with either a non-rebreathing circuit or the closed circuit with a large flow of oxygen and a tight fitting facemask.
  • 15. Alveolar ventilation: The second factor governing the delivery of anaesthetic agent to the lung is the level of alveolar ventilation. The greater the alveolar ventilation, the more rapid is the rise of alveolar concentration towards the inspired concentration. • This effect is limited only by the lung volume, the larger the functional residual capacity, the slower the wash. in of the new anaesthetic gas.
  • 16. Factors responsible for uptake from the lungs thus reducing the alveolar tension • 1. CARDIAC OUTPUT . • 2. BLOOD GAS SOLUBILITY . • 3. ALV – VENOUS GRADIENT.
  • 17. Cardiac output • Because blood carries anaesthetic away from the lungs, the greater the cardiac output, the greater the uptake, and consequently the slower the rate of rise of alveolar tension. • At low inspired concentration, the alveolar concentration results from a balance between the ventilatory input and circulatory uptake. • If the later removes half the anaesthetic introduced by ventilation, then the alveolar concentration is half that inspired.
  • 18. Blood gas solubility • If other things are equal, the greater the blood gas solubility coefficient, the greater the uptake of anaesthetic, and slower the rate of rise of alveolar concentration.
  • 19. Alveolar to venous partial pressure gradient: • During induction the tissues remove all the anaesthetic brought to them by the blood. Tissue uptake lowers the venous anaesthetic partial pressure far below that of the arterial blood. • If tissue uptake is more then it results in a large alveolar to venous anaesthetic partial pressure difference, which promotes maximum anaesthetic uptake and hence lowers the alveolar partial pressure.
  • 20. Concentration effects  The concentration effect modifies this influence of uptake. o When appreciable volumes are taken up rapidly, the lungs do not collapse; instead the sub atmospheric pressure created in the lung by the anaesthetic uptake causes passive inspiration of an additional volume of gas to replace that lost by uptake, thus increasing the alveolar concentration and offsetting the mathematical calculations. o Similarly, if an insoluble gas (e.g., nitrogen) is present in the inspired mixture, as the blood takes up the anaesthetic gas, the concentration of the insoluble gas will go up in the alveoli, reducing the concentration of the anaesthetic agent. o Concentration effect: The concentration effect helps in raising the alveolar tension towards the inspired tension, but hinders with it if an insoluble gas is present in the mixture.
  • 21. Methods to achieve desired gas and agent concentration. • High flow for short time • Use of Prefilled Circuits with gases • Use of large doses of anaesthetic agents. • Injector technique.
  • 22. Use of high flows for a short time  This is by and far the commonest and the most effective technique of initiating closed circuit. • By using high flows for a short time, the time constant is reduced thereby bringing the circuit concentration to the desired concentration rapidly. • Often, a fresh gas flow of 10L of the desired gas concentration and 2 MAC agent concentration is used so that by the end of three minutes (three time constants) the circuit would be brought to the desired concentration.  Advantages: --- Rapidity with which the desired concentration is achieved, the ability to prevent unexpected raise in the agent concentration and the ability to plenum vaporizers to achieve the desired concentration. – This also has the added advantage of achieving better denitrogenation, so vital to the conduct of the low flow anaesthesia.  The chief disadvantage: - Not economical & need scavenging systems to prevent theatre pollution  Mapleson3 using a spreadsheet model of a circle breathing system has calculated that, by using a FGF equal to minute ventilation and setting the anaesthetic agent partial pressure to 3 MAC, the end expired partial pressure of halothane will reach 1 MAC in 4 minutes and that of isoflurane in 1.5 minutes.
  • 23. Use of Prefilled circuit • The second method is utilizing a different circuit like Magill’s for pre- oxygenation. • Simultaneously, the circle is fitted with a test lung and the entire circuit is filled with the gas mixture of the desired concentration. • Following intubation, the patient is connected to the circuit thereby ensuring rapid achievement of the desired concentration in the circuit.
  • 24. Use of large doses of anaesthetic agents.  The third method consists of adding large amounts of anaesthetic agent into the circuit so that the circuit volume + FRC rapidly achieves the desired concentration as well as compensates for the initial large anaesthetic gas uptake.  To execute this, Pre-oxygenation after intubation, fresh gas flow is started with metabolic flows of oxygen and a large amount of nitrous oxide often in the range of 3-5 litres per minute.  Oxygen concentration in the circuit gradually falls, is continuously monitor Oxygen and the nitrous oxide concentration, N2O flow is reduced once the desired oxygen concentration (33 - 40%) is achieved .  Disadvantage of this method : is hypoxia if the oxygen monitor were to malfunction. o Hence this method is seldom used for N2O. o In the contemporary anaesthesia machines it is not possible to administer lesser than 25% of oxygen and the described technique cannot be executed.
  • 25. Injection techniques.  An alternative method for administering the large amounts of the agents is by directly injecting the agent into the circuit. The high volatility coupled with the high temperature in the circle results in instantaneous vaporization of the agent.  The injection is made through a self sealing rubber diaphragm covering one limb of a metal t piece or a sampling port, inserted into either the inspiratory or the expiratory limb. o This is an old, time-tested method and is extremely reliable. o Each ml of the liquid halothane, on vaporization yields 226 ml of vapour and each ml of liquid isoflurane yields 196 ml of vapour at 20oC. Hence, the requirement of about 2ml of the agent is injected in small increments into the circuit.
  • 26. THE MAINTENANCE OF LOW FLOW ANAESTHESIA This is the most important phase as this is stretched over a long period of time and financial savings result directly from this; this phase is characterized by : 1. Need a steady alveolar concentration of respiratory gases. 2. Minimal uptake of the anaesthetic agents by the body. 3. Need to prevent delivery of hypoxic gas mixtures.  Since the uptake of the anaesthetic agent is small in this phase, adding small amounts of the anaesthetic gases to match the uptake and providing oxygen for the basal metabolism should suffice.
  • 27. Management of the oxygen and nitrous oxide flow during the maintenance phase  The need to discuss the flow rates of N2O and O2 arises specifically because of the possible danger of administration of a hypoxic mixture.  Let us analyze the following example. 33% oxygen is set using a flow of 500 ml of O2 and 1000 ml of N2O. o Oxygen is taken up from the lungs at a constant rate of about 4 ml/kg/min. o N2O is a relatively insoluble gas and after the initial equilibration with the FRC and vessel rich group of tissues, the up take is considerably reduced. o In this situation, there is a constant removal of O2 at a rate of 200 - 250 ml/min, where as the insoluble gas N2O uptake is minimal. Hence the gas that is partly vented and partly returning to the circuit will have more N2O and less of O2. o Over a period of time, due to the mixing of fresh gas that has 66% N2O and the expired CO2 free gas that has N2O much higher than that, the percentage of N2O will go up and that of O2 will fall, sometimes dangerously to produce hypoxic mixtures. A high flow of 10 lit/min at the start, for a period of 3 minutes, is followed by a flow of 400 ml of O2 and 600 ml of N2O for the initial 20 minutes and a flow of 500 ml of O2 and 500 ml of N2O thereafter.  This has been shown to maintain the oxygen concentration between 33 and 40 % at all times.
  • 28. Management of the potent anaesthetic agents during maintenance phase  This is easily accomplished by dialling in the calculated concentration on the plenum vaporizer for the flow being used.  For example, suppose the anaesthetic uptake for a desired concentration of 0.5% halothane is 7.5ml/min. o If a FGF of 500ml/min is being used, then the dial setting should be 1.5% for at this setting and for the used flow, the total vapour output would be 7.5ml/min. o If a flow of 1000ml/min is being used, then the dial setting should be 0.8%.  In practice the actual dial setting often over estimates the actual output since the plenum vaporizer under delivers the agent at low flows. Hence, the dial setting is fine-tuned depending on the endpoints being achieved.
  • 29. Weir and Kennedy recommend infusion of halothane Weir and Kennedy recommend infusion of halothane (in liquid ml/hr) at the following rates for a 50 kg adult at different time intervals. • 0-5 min 27 ml/hr • 5-30 min 5.71 ml/hr • 30-60 min 3.33ml/hr • 60-120 min 2.36 ml/hr These infusion rates had been derived from the Lowe's theory of the uptake of anaesthetic agent.
  • 30. Salient points to be considered during the maintenance phase The other salient points to be considered during the maintenance phase are the following: o a) Leaks must be meticulously sought for and prevented, flows must be adjusted to compensate for the gas lost in the leaks. o b) Most of the gas monitors sample gases at the rate of 200 ml/min, which may be sometimes as high as half the FGF. Hence, care must be taken to return the sample back to the circuit to maximize the economy of FGF utilisation.
  • 31. TERMINATION OF LOW FLOW ANAESTHESIA.  Unlike the initiation or the maintenance of the closed circuit, termination is less controversial. There are only two recognized methods of termination of the closed circuit.  A. Use of high flow of gases: the circuit is opened and a high flow of gas is used to flush out the anaesthetic agents .  B . Use of activated charcoal. o Activated charcoal when heated to 220oC adsorbs the potent vapours almost completely. Hence, a charcoal-containing canister with a bypass is placed in the circuit, towards the end of the anaesthesia, the gas is directed through the activated charcoal canister. o This results in the activated charcoal adsorbing the anaesthetic agent resulting in rapid recovery and at the same time, reducing theatre pollution. Nitrous oxide, due to its low solubility is washed off towards the end by using 100% oxygen.
  • 32. Advantages of LFA system. o 1.Enormous financial savings due to use of low fresh gas flows as well as the agent. o 2. High humidity in the system leads to fewer post anaesthetic complications. o 3.Maintenance of body temperature during prolonged procedures due to conservation of heat. o 4. Reduction in the theatre pollution. o 5. Environmental Pollution ( Ozone depletion/green house effect)
  • 33. Disadvantages of LFA • Inability to quickly alter the inspired concentrations. • More attention is required • Danger of hypercarbia. • Uncertainty about inspired concentrations • Faster absorbent exhaustion • Accumulation of undesirable trace gases in the system like CO, acetone, methane, hydrogen, ethanol, argon, nitrogen, compound A, etc.
  • 34. Concerns about Safety in LFA Points of concern in providing safe anaesthesia are: o Hypoxia. o Misdosage of the volatiles. o Exhaustion of the absorbent. o Gas volume deficiency. o Reduced controllability.
  • 35. Absorber assembly : An absorber assembly consists of ► A canister containing an absorbent. ► two ports for connection to breathing tubes ► a fresh gas inlet. ► Other components that may be mounted are inspiratory and expiratory unidirectional valves , an adjustable pressure limiting (APL) valve, and a bag mount.
  • 36. Canisters : The absorbent is held in canisters. ► Single or two canisters in series can be used. ► Side walls are transparent so that color changes can be easily recognised. ► Comes in large and small sizes.
  • 37. Absorbent Types ► High alkali absorbents High amount of NaOH or KOH When desiccated react with anesthetic agent to form CO or compound-A Do not change color when dry. ► Low alkali absorbents Less amount of KOH or NaOH Less amount of CO or compound-A when desiccated. ► Alkali free absorbent Consists mainly Ca(OH)2 and other added agents No evidence of CO formation Little or no compound-A formation
  • 38. Different formulations ► 1. Sodalime 2. Baralyme 3. Sofnolime 4. Amsorb plus 5. Ca(OH)2 lime etc,. 6. LiOH lime
  • 39. ► Sodalime is made of ► Ca(OH)2 - 80% ► NaOH – 4% ► KOH – 1% ► Water – 14% to 19 % Silica or Kieselguhr – helps in hardening and reduced dust formation. Indicator
  • 40. ► CO2 absorption is exothermic reaction ► Principle – acid base neutralization ►End product – carbonic acid and water. CO2 + H2O → H2CO3 H2CO3 + 2NaOH → Na2CO3 (sodium carbonate)+ 2H2O Na2CO3 + Ca(OH)2 → 2NaOH or 2KOH + CaCO3 .
  • 41. Pattern of CO2 absorption A.Unused absorbent in the canister. B.After limited use- absorption of CO2 has occurred primarily at the inlet and to lesser extent along the sides. C.After extensive use – granules at the inlet and along the sides are exhausted. D.Exhausted sodalime – few granules at the distal 1/3rd are still capable of absorbing CO2. E.Channeling effect
  • 42. Indicators color When fresh when exhausted Phenolphthalein white Pink Ethyl violet White Purple Clayton yellow Red Yellow Ethyl orange Orange Yellow Mimosa Z Red White
  • 43. ► Shape and size of absorbent pellets or granules form Advantage - small granules increases the surface area, decreases the gas channelling along low resistance pathway. Disadvantage – increased resistance,caking. 4 to 8 mesh more commonly used.
  • 44. REACTIONS BETWEEN ABSORBENTS AND ANESTHETIC AGENTS ► Haloalkene Formation Halothane degradation most often occurs during closed-circuit anesthesia and produces the haloalkene 2-bromo-2- chloro-1, 1-difluoroethene (BCDFE).Nephrotoxic in rats. ►Compound A Formation Sevoflurane decomposes in the presence of carbon dioxide absorbents to compound A. Compound A is vinyl ether and has dose dependent nephrotoxic effects in rats.
  • 45. Carbon monoxide formation ► when desflurane, enflurane, or isoflurane is passed through desiccated absorbent containing a strong alkali (potassium or sodium hydroxide) ► When sevoflurane is degraded by desiccated absorbent, carbon monoxide is formed if the temperature exceeds 80°C.
  • 46. Factors associated with carbon monoxide formation: ► Absorbent Composition – NaOH , KOH ► Absorbent Desiccation ( dehydrated absorbent). ► Anesthetic Agent – Desflurane > Enflurane>Isoflurane .Halothane produces very less amount of CO. ► Increased temperature Inside the Absorber. ► Fresh Gas Flow - FGF leads to more dessication of absorbent and more CO production. ► Carbon Dioxide Absorption – Increases absorption leads to decreased CO production.
  • 47. The APSF recommendations to prevent absorbent desiccation if the department continues to use strong alkali absorbents with volatile anesthetic agents: ► All gas flows should be turned OFF after each case. ► Vaporizers should be turned OFF when not in use. ► Canisters on an anesthesia machine that is commonly not used for a long period of time should not be filled with absorbent that contains strong alkali or should be filled with fresh absorbent before each use. ► The absorbent should be changed routinely, at least once a week, preferably on a Monday morning, and whenever fresh gas has been flowing for an extensive or indeterminate period of time.
  • 48. ► The canister should be labelled with the filling date. ► Checking this date should be part of the daily machine checklist. ► If a double-chamber absorber is used, the absorbent in both canisters should be changed at the same time. ► The practice of supplying oxygen for administration to a patient who is not receiving general anesthesia through the circle system should be strongly discouraged. ► Using fresh gas to dry breathing system components should be discouraged.
  • 49. WHEN AND HOW TO CHANGE THE ABSORBENT ► Inspired Carbon Dioxide - Most reliable method ( indicated by capnograph) ► Indicator Color Change ► Heat in the Canister
  • 50.  HUMIDITY general term used to describe the amount of water vapor in a gas may be expressed several ways:- Absolute Humidity - the mass of water vapor present in a volume of gas - commonly expressed in milligrams of water per liter of gas. Humidity at Saturation -maximum amount of water vapor that a volume of gas can hold - varies with the temperature.The warmer the temperature, the more water vapor can be held in a gas HUMIDIFICATION
  • 51.  RELATIVE HUMIDITY the amount of water vapor at a particular temperature expressed as a percentage of the amount that would be held if the gas were saturated.  DEW POINT The temperature at which a gas is 100% saturated  WATER VAPOR PRESSURE Humidity may also be expressed as the pressure exerted by water vapor in a gas mixture.
  • 52.  normal breathing :upper respiratory tract warms humidifies, and filters inspired gases, primarily in the nasopharynx  Even at extremes of inspired temperature and humidity, gas that reaches the alveolar level is 100% saturated at body temperature.  ISOTHERMIC SATURATION BOUNDARY (ISB) : The point at which gases reach alveolar conditions (37°C and 100% relative humidity)  Over the course of a normal day, the respiratory tract loses approximately 1470 J of heat and 250 mL of water. IMPORTANCE OF HUMIDIFICATION
  • 53.  Water is intentionally removed from medical gases so that gases delivered from anaesthesia machine are dry and at room temperature  Gases must therefore be warmed to body temperature and saturated with water by upper respiratory tract  normal nasal breathing : temperature in upper trachea is between 30- 33o C with RH-98% & AH-33 mg/L : at alveolar level 37 degreeC , RH- 100% ,AH – 44 mg/L  Tracheal intubation and high fresh gas flows bypass this normal humidification by upper airways and expose lower airways to dry (< 10 mg H2O/L), room temperature gases
  • 54. EFFECTS OF INHALING DRY GASES  DAMAGETOTHE RESPIRATORYTRACT  Mucosa dries  Temperature drops  Secretions thicken  if not cleared atelectasis / airway obstruction  Thick plugs  loci for infection  Ciliary function reduced  Surfactant activity impaired  Mucosa more susceptible to injury  Dry gases  bronchoconstriction  Body heat loss Absorbent desiccation Trachealtube obstruction
  • 55.
  • 56. SOURCES OF HUMIDITY  CO2 absorbent – reaction ofCO2 with absorbent releases water -- absorbent granules also contain water  Exhaled gases – some rebreathing is present in tracheal tube, supraglottic airway device, and connections to breathing system -almost half of humidity in expired gases is preserved in this manner  Moistening/Rinsing breathing tubes and reservoir bag before use  Low fresh gas flows – conserve moisture  Coaxial breathing circuits – increase humidity more quickly than a system with 2 separate limbs, when combined with low flows -not very efficient : Bain system (coaxial Version of Mapleson D) does not meet optimal humidification requirements because of high fresh gas flow required
  • 57.  HEAT AND MOISTURE EXCHANGERS –  hydrophobic  hygroscopic  HUMIDIFIERS – pass-over/blow- by/bubble- through/vapor-phase humidifier  HEATED  UNHEATED  Nebulisers SOURCESOFHUMIDITY
  • 58. HME •SYNONYMS :- condenser humidifier, swedish nose, artificial nose, nose humidifier, passive humidifier, regenerative humidifier, moisture exchanger, and vapor condenser •When combined with filter for bacteria & viruses :- HMEF
  • 59.
  • 60.  disposable devices  exchanging medium enclosed in a plastic housing  vary in size and shape  Each has a 15-mm female connection port at the patient end and a 15-mm male port at the other end; patient port may also have a concentric 22-mm male fitting  May have a port to attach the gas sampling line for a respiratory gas monitor or an oxygen line  One type utilizes a ceramic heating element with a water input port, a membrane, and an aluminum grid that vaporizes the water  Preferred for short term use.
  • 61. Hydrophobic  hydrophobic membrane with small pores  membrane is pleated to increase the surface area  provides moderately good inspired humidity  may be impaired by high ambient temperatures  efficient bacterial and viral filters  prevent the HCV from passing  allow the passage of water vapor but not liquid water at usual ventilatory pressure  associated with small increases in resistance even when wet Hygroscopic  contain a wool/foam/paperlike material coated with moisture-retaining chemicals  medium may be impregnated with a bactericide  Composite hygroscopic HMEs -- a hygroscopic layer plus a layer of thin, nonwoven fiber membrane that has been subjected to an electrical field to increase its polarity -- improves filtration efficiency and hydrophobicity.  composite hygroscopic HMEs are more efficient than hydrophobic ones  lose their airborne filtration efficiency if they become wet; microorganisms held by the filter medium can be washed through the device  Their resistance can increase greatly when wet
  • 62. Type Heat and moisture exchangingefficiency Hygroscopic Excellent Hydrophobic Good Effect of increased tidal volume on heat and moisture exchange Slightdecrease Significantdecrease Filtrationefficiency when dry Good Excellent Filtrationefficiency when wet Poor Excellent Resistance when dry Low Low Resistance when wet Significantlyincreased Slightlyincreased Effect of nebulized medications Greatly increased resistance Little effect
  • 63.  CONTAINDICATIONS  thick, copious, or bloody secretions  expired tidal volume less than 70% of the delivered tidal volume (eg, those with large bronchopleurocutaneous fistulas or incompetent or absent endotracheal tube cuffs)  body temperatures less than 32°C  high spontaneous minute volumes (> 10L/min)  Spontaneously breathing, difficult to wean patients – relativeC/I  MOISTUREOUTPUT •Amount of Heat & Moisture provided. •Currently no standards for minimum MO for HME’s •Depends upon –TV, insp time, RR, system leak, initial humidity.
  • 64. ADVANTAGES DISADVANTAGES  Temp preservation  Dead space rebreathing  IncreasedWOB  Resistance  ETT obstruction  Not for all patients  Disconnection  CO2 absorbent dessication  Avoid using with nebulizers  Short term use  Inexpensive  Easy to use  Portable,  light weight  Less risk of contamination  Efficient filter  Do not require external heat, water, temp monitoring  No condensate- thus no ventilator malfunction
  • 65. HAZARDS  ineffective low-pressure alarm during disconnection due to resistance through HME.  Suspected high resistance  Peak airway pressure should be measured withAND without HME in place  possible increased resistive work of breathing due to mucus plugging of HME  In Mapleson system – resistance causes diversion of FGF down the expiratory limb.  hypothermia  airway obstruction– ETT blockage/ ETT kinking/ detachment  Should not be used with Heated Humidifiers  nebulized medications increase resistance of hygroscopic HME.
  • 66. •Ineffective filtration:- liquid can break through a hygroscopic HME •Foreign body aspiration – detached parts inhaled •Rebreathing due to dead space (more in paed) •Hypothermia •Leaks & disconnections •dryCO2 absorbent- absorb volatile agents, hence delay in induction. POSITION INCIRCUIT : - Patient – HME – filter – sampling line – machine/ ventilator
  • 67. HUMIDIFIERS - Add water to gas by passing the gas over a water chamber (passover) / through a saturated wick (wick humidifier) /bubbling it through water (bubble-through) / mixing it with vaporized water (vapor-phase) - do not filter respiratory gases  UNHEATED  Disposable, bubble-through devices used in oxygen supplied to patients via facemask or nasal canula  Simple containers containing distilled water through which oxygen is passed and it gets humidified  Maximum humidity that can be achieved is 9mg H2O/L  HEATED  Incorporate a device to warm water in the humidifier, some also heat inspiratory tube  Humidification chamber  Heat source  Temperature monitor  Thermostat  Inspiratorytube
  • 69. HEATED HUMIDIFIERS  Humidification chamber – contains liquid water, disposable/ reusable, clear (easy to check water level)  Heat source – heated rods immersed in water/ plate at bottom of humidification chamber  Temperature monitor – to measure gas temperature at patient end of breathing system  Thermostat – 1. Servo-controlled units – automatically regulates power to heating element in response to temperature sensed by a probe near patient connection/ humidifier outlet 2. Non servo-controlled units – provides power to heating element according to setting of a control, irrespective of delivered temperature
  • 70.  Inspiratory tube – conveys humidified gas from humidifier outlet to patient  If unheated  gas will cool and lose some of its moisture as it travels to the patient, water trap necessary to collect condensed water  If Heated or insulated  more precise control of temperature and humidity delivered to patient, avoids moisture rainout  Controls – most allow temperature selection at end of delivery tube or at humidification chamber outlet  Alarms – to indicate temperature deviation by a fixed amount, displacement of temperature probe, disconnection of heater wire, low water level in humidification chamber, faulty airway temperature probe , lack of gas flow in the circuit  Standard requirements -An international and aU.S. standard on humidifiers have been published
  • 71.  In circle system, heated humidifier is placed in the inspiratory limb downstream of unidirectional valve by using an accessory breathing tube  Must not be placed in the expiratory limb  Filter, if used, must be placed upstream of humidifier to prevent it from becoming clogged  In Mapleson systems, humidifier is usually placed in fresh gas supply tube  Humidifier must be lower than patient to avoid risk of water running down the tubing into the patient
  • 72.  Condensate must be drained periodically or a water trap inserted in the most dependent part of the tubing to prevent blockage or aspiration  Heater wire in delivery tube should not be bunched, but strung evenly along length of tube  Delivery tube should not rest on other surfaces or be covered with sheets, blankets, or other materials; a boom arm or tube tree may be used for support
  • 73.  Advantages – 1. Capable of delivering saturated gas at body temperature or above, even with high flow rates 2. More effective humidification than an HME  Disadvantages – 1. Bulky and somewhat complex 2. Involve high maintenance costs, electrical hazards, and increased work (temperature control, refilling the reservoir, draining condensate, cleaning, and sterilization) 3. Offers relatively little protection against heat loss during anaesthesia as compared to circulating water and forced-air warming
  • 74. NEBULISERS  Aerosol generators/ atomizers/ nebulizing humidifiers  Emit water in the form of an aerosol mist (water vapor plus particulate water)  Used for producing humdification and delivery of drug directly into respiratory tract  Drugs delivered by nebulizers – Bronchodilators, decongestants, mucolytic agents, steroids  Optimal particle size of droplet (aerosol) = 0.5 to 5 µm  Particles > 5 µm – unable to reach peripheral airways, deposited in main airways  Particles < 0.5 µm - very light, come back with expired gases without being deposited in airways
  • 75.  Most commonly used 2 types – 1. Pneumatically driven (gas-driven, jet, high pressure, compressed gas) 2. Ultrasonic
  • 76.  Pneumatically driven nebulizer works by pushing a jet of high- pressure gas into a liquid, inducing shearing forces and breaking the water up into fine particles  Should be placed in the fresh gas line (high flow of gas must be used with pneumatic nebulizer)  Produces particles of size 5 to 30 µm (only 30 to 40% of particles produced are in optimal range)  most of the particles get deposited in wall of main airways
  • 77.  Ultrasonic nebulizer  produces a fine mist by subjecting the liquid to a high-frequency, electrically driven resonator  Can be used in the fresh gas line or the inspiratory limb (No need for a driving gas)  Frequency of oscillation determines the size of the droplets  Creates a denser mist than pneumatic ones  produces mist with aerosol size of 1 to 10 µm (95% of particles produced are in optimal range)  particles get deposited directly in airways  very useful for delivery of bronchodilators directly in peripheral airways  Can nebulize 6 mL of water or drug in 1 minute
  • 78.
  • 79. HAZARDS 1. Nebulized drugs may obstruct an HME or filter in the breathing system 2. Overhydration 3. Hypothermia 4. Transmission of infection 5. Case reports where a nebulizer was connected directly to a tracheal tube without provision for exhalation  resulted in pneumothorax in one case
  • 80.  Advantage - can deliver gases saturated with water without heat and, if desired, can produce gases carrying more water  Disadvantages – 1. Somewhat costly 2. Pneumatic nebulizers require high gas flows 3. Ultrasonic nebulizers require a source of electricity and may present electrical hazards 4. May be considerable water deposition in the tubings, requiring frequent draining, water traps in both the inspiratory and exhalation tubes, and posing the dangers of water draining into the patient or blocking the tubing
  • 81. INTRODUCTION : ● Tracelevel- concofanestheticgasthat isfar belowtheone thatisneededfor clinical anesthesia. ● Variesgreatly – dependingonFGF,ventilationsystem, thelengthoftime ofanesthesia, measurementsite andother variables ● Degreeofpollutionismore duringpaediatricsurgeries SCAVENGING SYSTEM - Hospital-basedandstand-aloneoperatingrooms, - Recoveryrooms, - Dentaloperatories, - V eterinaryfacilities. Workplace exposures:
  • 82. ● Twodifferent classesof chemicals: nitrousoxide and halogenated agents ● In1977NationalInstitute for Occupational SafetyandHealth (NIOSH) – recommendedexposurelimits (RELs) for both nitrousoxide and halogenatedagent
  • 83. Scavenging and Waste Anesthetic Gases (WAGs) SCAVENGING ● Definition ● Scavengingisthe collectionand removal ofvented anesthetic gases from the OR. ● Since the amount ofanestheticgas supplied usuallyfar exceedsthe amount necessary for the patient, OR pollution isdecreased by scavenging. ● Scavengerandoperating room ventilationefficiencyare the two most important factors in reduction ofwaste anestheticgases(WAGs).
  • 84. ● Theuseofscavengingdeviceswith anesthesiadeliverysystemsisthe most effectivewayto decrease waste anestheticgases. ● Anefficientscavengingsystemiscapableof reducing ambient concentrations ofwastegasesbyup to 90%. ● Anesthesiamachinesandbreathingsystemsdelivering halogenatedhydrocarbonanestheticsand/ornitrous oxideshouldnotbeoperatedunlesstheyareequipped withafunctional scavengingsystem . ● Tobe effective,ascavengingsystemmustnot leakandmustcontrol theconcentrationoftrace anestheticsin ambient air. ● Wastegasesshould not bedischarged into the outsideair in an areawhere reentryinto the buildingislikely.
  • 85. Types ● Scavengingmaybe active(suctionapplied) or passive(waste gasesproceed passivelydown corrugated tubing through the room ventilation exhaust grill ofthe OR). ● Activesystemsrequire ameansto protect the patient's airwayfrom the applicationofsuction, or buildup ofpositive pressure. ● Passivesystems require that the patient be protected from positivepressure buildup only.
  • 86. ● The 2 majorcausesof wastegascontamination in the O.R : 1. The anesthetic technique used and 2. The equipment issues Regardinganesthesiatechnique the followingwill increase anesthetic gas pollution wheninhalantanestheticsare used: ● Administeringinhalantanestheticsbyopen drop (eg, periodically drippingliquid volatileanesthetic onto agauzesponge) or insufflation(eg, delivery ofarelativelyhighflowof anesthetic in oxygeninto the tracheaor pharynx throughacatheter) techniques. ● Turning on flowmetersandvaporizersbefore attachingthe breathingsystemto the patient. ● Allowingflowmetersandvaporizersto remain on after the patient isdisconnected fromthe breathingsystem.
  • 87. ● Using uncuffed endotracheal tubes that do not create a completely sealed airway or using cuffed tubes without inflating the cuff. ● Disconnecting apatient from abreathing system without eliminating as much of the residual gases as reasonably possible through the scavenging system. ● Spillingliquidanesthetic duringthe filling ofvaporizers,especially duringan anestheticprocedure. ● Poorlyfittingmasks. ● Flushingthe circuit. ● Use ofbreathing circuits that are difficult to scavenge, suchasJackson- Rees.
  • 88. Componentsof thescavengersystem: ● Gascollection assembly. ● Transfer tubing. ● Scavenginginterface . ● Gasdisposal tubing. ● Gasdisposal assembly.
  • 89. 1.Gas collecting assembly ● Capturesexcessanesthetic gases and deliversit to the transfer tubing. ● WAGare vented fromanesthesia systemthrough either adjustable pressure limiting valveor ventilator reliefvalve. ● Gaspassingthroughthese valves accumulatesin the gas collecting assemblyand is directed to the transfer means.
  • 90. 2.Transfer tubing ● The transfer tubescarriesexcessgasfromgascollecting assembly to scavenging interface. ● The tubes mustbe either 19 or 30 mm, sometimesyellowcolor-coded . ● The tubesshould be sufficientlyrigidto prevent kinkingand asshort as possibleto minimizethe chanceofocclusion. ● Separatetubes from the APLvalveandventilator relief valvemerge into a singlehosebefore they enter scavenginginterface. ● Ifthe transfer meansisoccluded, baselinebreathing circuit pressure willincrease and barotrauma can occur.
  • 91. 3.Scavenging interface ● The scavengerinterfaceisthe most important component because it protects the breathing circuit or ventilator from excesspositiveor negativepressure. ● Positive-pressurereliefismandatory, irrespective ofthe typeof disposable systemused, to vent excessgasin caseofocclusiondistal to interface. ● scavenger interfacesmaybe open closed
  • 92. ● B&Cclosed interfaces ● A&Bare active systems ● Cispassive system
  • 93. OPEN INTERFACE: ● It containsno valvesand isopen to the atmosphere, allowingpositive and negativepressure relief. ● Open interfacesshould be used only withactive disposable systems that haveacentralevacuation system. ● open interfaces require areservoir becausewastegasesare intermittently discharged insurges whereasflow fromthe evacuation systemis continuous.
  • 94. ● The efficiencyofit dependson several factors: A. Thevacuum flowrate per min must equalor exceed the minute volume ofexcessgasesto prevent spillage. B. Spillagewill occur ifthe volume ofasingle exhaled breathexceedsthe capacityofreservoir. Openinterfaces aresaferforthepatients.
  • 95. ● CLOSEDINTERFACE: ● It communicateswith the atmosphere throughvalves. ● Twotypesofclosedinterfacesare commerciallyavailable: positive pressure reliefonly positive and negativepressure relief POSITIVEPRESSURERELIEFONLY: ● It hasasingle positive pressure relief valve and is designated to be used onlywithpassivedisposable systems. ● Transferofthe wastegasfromthe interface to the disposablesystem reliesonweightor pressureofthe wastegasitselfbecauseanegative pressure evacuation systemisnot used. ● In thissystemreservoir bagisnot required.
  • 96.
  • 97. POSITIVEANDNEGA TIVEPRESSURERELIEF: ● It haspositive and negative pressure relief valvein addition to areservoir bag. ● It isused with active disposable systems. Theeffectivenessofaclosed systeminpreventing spillagedependson: the rate ofwaste gasinflow the evacuation flow rate the size of the reservoir ● Leakage occurs only when the reservoir bag becomes fully inflated and pressure increases sufficiently to open the positive pressure relief valve . ● In contrast, the effectiveness of aopen system in preventing spillage depends not only on the volume of the reservoir but also on the
  • 98. 4.Gas disposal tubing ● The gas disposable tubing conducts waste gas from the Scavenging interface to the gas disposable assembly. ● It should be collapse proof and should run overhead, if possible to minimize the chance of accidental occlusion. ● It ultimately eliminates excess waste gas. ● It is of two types: Active and Passive ● Active assembly: ● most commonly used. It uses central evacuation system. ● A vacuum pump serves as mechanical flow inducing device that removes the waste gases. ● An interface with a negative pressure relief valve is mandatory because the pressure within the system is negative. ● A reservoir is very desirable and the larger the reservoir, the lower the suction flow rate needed. 5.Gas disposal assembly
  • 99. - PipedVacuum - Active duct system - employsflow inducingdevices(fans, pumps, blowersetc..) to move evenat low pressures Advantages ● Convenientin largehospitals, where manymachinesare in use in different locations Disadvantages ● Vacuumsystemandpipe workisamajor expense ● Needlevalve mayneedcontinualadjustment
  • 100. ● Passivedisposablesystem : ● doesnot use amechanical flowinducingdevice. ● The weightor pressure from the heavier-than-air anesthetic gasesproducesflowthroughthe system. ● Positive pressure reliefismandatory, but anegative pressure reliefand reservoir are unnecessary. ● Passive– simpler and lessexpensivebut not effective - Roomventilation system - Nonrecirculatingor – Recirculating - Pipingdirect to atmosphere - Adsorption devices - Catalyst decomposition
  • 101. Charcoal canisters : ● activated charcoal ● connected to the outlet ofthe breathing system ● removeshalogenated anaestheticsby filtration CatalyticDecomposition : • Catalyticdecompositioncanbe used to convert nitrousoxide to nitrogen andoxygen, reducingits contribution to the greenhouse effect
  • 102. Advantages ● No set-up cost ● Mobile - moveswiththe machine Disadvantages ● Continuingcost ofreplacement. ● Continuosreplacement ismessy ● Doesnot remove nitrousoxide ● Heatingthe canister causesthe release of the inhalational agents
  • 103. Evaluation of Anesthetic Equipment ● Eachpieceofequipment involved in the deliveryofinhalant anestheticsshouldbe evaluated regularly to assureits function and integrity. Proceduresforcheckoutofanesthesiaequipment,dependingontheequipmentto beused,shouldincludethefollowing: ● Statusofthehigh-pressuresystem,includingtheoxygensupplyandnitrousoxide supply- The nitrous oxide supply should not leak when the cylinder valve ison and the nitrousoxideflowmeter isoff. ● Statusofthelow-pressuresystem(flowmeterfunction)- Anegative-pressure leaktest should be performed at the commongasoutlet or the outlet ofthe vaporizer immediatelyupstreamfromthe breathingsystem.
  • 104. ● Statusofthebreathingsystem- Anappropriateleaktest for acircle system andNoncirclesystemscanusuallybe done. The quantityofleakagecan be measuredbydeterminingthe flowrate ofoxygen necessaryto maintainaconstant pressureinthe system,andthe leakrate should be lessthan300ml/ minat 30cmofh2O. ● Statusofthescavengingsystem- Thescavengingsystemshouldbeproperly attachedat allconnectors, andthe appropriate vacuumshould be assured for activesystems.Ifcharcoalcanistersare employedfor scavenging,theyshould bechangedat appropriateintervals. ● Statusofmechanicalventilators- Ventilatorsshouldbeconnected properly to the anesthesiamachine, andanabsence ofleaksshould be assured.
  • 105. Monitoring of the Effectiveness of Antipollution Techniques ● Monitoring trace-gas concentrations in the workplace provides a quantitative assessment of the effectiveness of a waste-gas control program. ● Measuringthe concentration ofanesthetic in the breathing zoneof the most heavilyexposed workersisthe usual procedure. ● Anair-monitoring programismostappropriately startedafter anesthesiadeliverysystemshavebeen equipped with scavenging systemsandafter other techniquesfor minimizingwastegaspollution are in place. ● Anidealapproachwouldincludefrequentairmonitoring,preferably at least semiannual evaluations.
  • 106. Effectiveness ● Unscavengedoperatingrooms show10-70 ppmhalothane, and 400- 3000ppmN2O. ● scavengingbrings these levels downto 1 and60 ppmrespectively. ● Addingcarefulattention to leaksandtechniquecanyieldlevelsaslowas 0.005 and 1 ppm.
  • 107. HAZARDS ● Scavengingsystemfunctionallyextendsthe anesthesiacircuit all the way fromthe anesthesiamachine to the disposable site. ● Obstruction of scavengingpathwaycancauseexcessivepositivepressure in the breathing circuit andbarotraumacanoccur. ● excessivevacuum appliedcanresult in undesirablenegativepressures withinbreathingsystem. ● LossofMonitoringInput – it maymaskthe strong odor ofavolatile anesthetic agent, delaying recognition ofanoverdose ● Alarmfailure – neg. pressure from the scavengingsysteminterface prevent the bellows from collapsingwhen breathing systemis disconnected . Low airwaypressure alarmnot activated.
  • 108. Avoiding waste gas exposure ● Good maskfit. ● Avoidunscavengeable techniquesifpossible(insufflation). ● Prevent flowfrom breathing systeminto room air (onlyturn on agent andnitrous oxide after maskison face, turn them offbefore suctioning). ● Washoutanesthetics(into the breathing circuit) at the end of the anesthetia. ● Don’t spill liquidagent. ● Use low flows. ● Usecuffedtracheal tubes whenpossible. ● Checkthe machine regularlyfor leaks. ● Disconnectnitrous oxide pipeline connection at wallat the day’send. ● Total intravenousanesthesia.