SlideShare a Scribd company logo
1 of 16
Download to read offline
TI066D/06/en
71008913
Technical Information
Dosimag
Electromagnetic Flow Measuring System
Volume flow measuring system for filling applications
Application
Electromagnetic flowmeter for measuring in filling
applications.
• Flow up to 1.66 l/s
• Fluid temperature up to +130 °C
• Process pressures up to 16 bar
• CIP/SIP cleanable
• Stainless steel housing
All liquids with a conductivity of ≥5 µS/cm can be
measured in the following sectors for example:
• Food industry
• Cosmetics industry
• Pharmaceuticals industry
• Chemicals industry
Approvals in the food industry/hygiene sector:
• 3-A approval, EHEDG-tested, in conformity with FDA
Application-specific lining material:
• PFA
Your benefits
Dosimag guarantees the highest level of accuracy and
repeatability even for short measuring times. The com-
pact housing shape means the units can be arranged very
close together in filling plants.
The “Batchline” concept comprises additionally:
• Identical process connections enable an uncomplica-
ted exchange between “Dosimag” and the Coriolis
mass flowmeter “Dosimass”
• Uniform operating concept using the “ToF Tool -
Fieldtool Package” operating software:
– Graphic display of measured values for detailed
trend analysis and optimization of the filling process
– Complete plant documentation with device config-
uration and filling diagrams can be created
The Dosimag flowmeter offers:
• Simple installation and commissioning
• Insensitivity to pipe vibrations
Dosimag
2 Endress+Hauser
Table of contents
Function and system design. . . . . . . . . . . . . . . . . . . . . 3
Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Operable flow range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Signal on alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Low flow cutoff . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Galvanic isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Switching output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Power supply. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Electrical connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Supply voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Cable connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Cable specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Power consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Power supply failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Potential equalization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Performance characteristics. . . . . . . . . . . . . . . . . . . . . 4
Reference operating conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Maximum measured error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Repeatability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Operating conditions: Installation . . . . . . . . . . . . . . . . 5
Installation instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Inlet and outlet runs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Adapters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Operating conditions: Environment. . . . . . . . . . . . . . . 9
Ambient temperature range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Shock and vibration resistance. . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Operating conditions: Process . . . . . . . . . . . . . . . . . . . 9
Medium temperature range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Conductivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Limiting medium pressure range (nominal pressure)
9
Pressure tightness (liner) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Limiting flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Pressure loss . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Mechanical construction . . . . . . . . . . . . . . . . . . . . . . 11
Design / dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Material load diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Process connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Surface roughness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
User interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Display elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Remote operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Certificates and approvals . . . . . . . . . . . . . . . . . . . . . 13
CE mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
C-tick mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Sanitary compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Other approvals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Pressure measuring device approval . . . . . . . . . . . . . . . . . . . . . . 14
Other standards and guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Ordering information. . . . . . . . . . . . . . . . . . . . . . . . . 14
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Supplementary Documentation . . . . . . . . . . . . . . . . . 15
Registered trademarks. . . . . . . . . . . . . . . . . . . . . . . . 15
Dosimag
Endress+Hauser 3
Function and system design
Measuring principle In accordance with Faraday’s law of magnetic induction a voltage is induced in a conductor which is moved
through a magnetic field. In the electromagnetic measuring principle, the flowing medium corresponds to the
moving conductor. The induced voltage is proportional to the flow velocity and is directed to the amplifier by
means of two measuring electrodes. The flow volume is calculated from the pipe cross-sectional area. The DC
magnetic field is generated by a switched direct current of alternating polarity.
a0003191
Ue = B · L · v
Q = A · v
Ue Induced voltage
B Magnetic induction (magnetic field)
L Electrode spacing
v Flow velocity
Q Volume flow
A Pipe cross-section
I Intensity of current
Measuring system The measuring system is a compact unit consisting of a sensor and transmitter.
Input
Measured variable Flow rate (proportional to induced voltage)
Measuring range Typically v = 0.01 to 10 m/s with the specified measuring accuracy
Operable flow range Over 1000:1
Output
Output signal Pulse output:
Passive, open emitter, max. 30 V DC / 25 mA, galvanically isolated, pulse value and pulse polarity can be
selected, adjustable pulse width (0.04 to 4 ms).
Signal on alarm Pulse output → failsafe mode can be selected
Status output → transistor non-conductive in the event of a fault or if the power supply fails
Low flow cutoff Switch-on point for low flow cutoff selectable.
Galvanic isolation The circuits of the pulse/status output are galvanically isolated on the device side from the communication and
power supply.
Ue
I
I
B
L
V
Dosimag
4 Endress+Hauser
Switching output Status output:
Passive, open emitter, max. 30 V DC / 25 mA, galvanically isolated
Power supply
Electrical connections The device electrical connection is by means of a Lumberg connector (type RSE8, M12x1).
a0003833
Wiring diagram (direct connection without an adapter)
A Socket on device
B Cable connector
1 (+), power supply (24 V DC nominal voltage (20 to 30 V DC), 6 W)
4 (–), power supply (24 V DC nominal voltage (20 to 30 V DC), 6 W)
5 (+), pulse output, status output (max. 30 V)
6 (–), pulse output (max. 25 mA)
7 (–), status output (max. 25 mA)
2 Service interface (may not be connected during normal operation)
3 Service interface (may not be connected during normal operation)
8 Service interface (may not be connected during normal operation)
Supply voltage 24 V DC nominal voltage (20 to 30 V DC), PELV or SELV
When installing the Dosimag based on the CAN/CSA-C22.2 No. 1010.1-92 safety standard for Canada, the
power must be supplied via a SELV power supply with maximum 30 V DC.
Cable connection Lumberg plug (RSE 8, M12x1) for power supply and signal outputs.
Cable specifications Use connecting cables with a cross-section of at least 0.25 mm2
(e.g. AWG23). The temperature specification
of the cable must be at least 20 °C higher than the ambient temperature in the application.
Power consumption DC: <6 W (incl. sensor)
Switch-on current: max. 1.9 A (< 5 ms) at 24 V DC
Power supply failure At least up to 20 milliseconds. All sensor and measuring point data remain in the M-DAT.
Potential equalization No potential matching is needed for grounded steel pipings.
Performance characteristics
Reference operating condi-
tions
To DIN EN 29104 and VDI/VDE 2641:
• Fluid temperature: +28 °C ± 2 K
• Ambient temperature: +22 °C ± 2 K
A
1
2
3
4
5
6
7
8
1
2
3
4
5
6
7
8
B
RSE8
A
B
Dosimag
Endress+Hauser 5
• Warm-up period: 30 minutes
Installation:
• Inlet run >10 x DN
• Outlet run >5 x DN
• Sensor and transmitter grounded.
• Sensor centered relative to the pipe.
Maximum measured error Volume flow:
• ±0.25% o.r. (1 to 4 m/s) or
• ±0.5% o.r. ± 1 mm/s or
• ± 5% o.r.
o.r. = of reading
Repeatability
Operating conditions: Installation
Installation instructions Mounting location
Correct measuring is possible only if the pipe is completely full. For this reason, we recommend you run trial
fillings before starting productive batching.
a0003761
Rotary filling system (example)
1 = Dosimag
2 = Tank
3 = Filler valve
4 = Container
Batch time “ta”
[s]
Relative standard deviation in relation to the batched volume
[%]
1.5 s < ta < 3 s 0.4
3 s < ta < 5 s 0.2
5 s < ta 0.1
2
1
3
4
Dosimag
6 Endress+Hauser
a0003762
Line filling system (example)
1 = Dosimag
2 = Tank
3 = Filler valve
4 = Container
Installation near valves
Do not install the sensor downstream of a filler valve. If you do, the measuring tube of the sensor would be
completely emptied after every filling cycle which would greatly distort the measured value during the next
flow measurement.
a0003768
Installation near valves (✓ = recommended, x = not recommended)
1 = Dosimag
2 = Filler valve
3 = Container
Orientation
An optimum orientation position helps avoid gas and air accumulations, and deposits in the measuring tube.
Vertical orientation
Optimum measurement takes place when the pipe system is completely filled with the fluid.
a0003795
Dosimag orientations
1 = Dosimag
2 = Filler valve
3 = Container
1 3
4
2
✘✔
1
2
2
1
3 3
1
2 1 2
1
2
3 3 3
Dosimag
Endress+Hauser 7
Horizontal orientation
The measuring electrode plane should be horizontal. This prevents brief insulation of the two electrodes by
entrained air bubbles.
a0003829
Horizontal installation with Dosimag
1 Measuring electrodes for signal acquisition
2 Measuring tube liner
Caution!
In the event of extreme heating (e.g. for CIP or SIP cleaning processes), we recommend you install the mea-
suring device in such a way that the transmitter part is pointing downwards. This reduces the risk of the elec-
tronic components overheating.
a0003830
Recommended orientation in the event of extreme heating (✓ = recommended, x = not recommended)
Vibrations
Secure the piping and the sensor if vibration is severe.
a0003831
Measures to prevent vibration of the measuring device
A
A
1 2
✘✔
Dosimag
8 Endress+Hauser
Inlet and outlet runs If possible, install the sensor before fittings such as valves, T-pieces, elbows, etc. Compliance with the following
requirements for the inlet and outlet runs is necessary in order to ensure measuring accuracy.
• Inlet run ≥ 5 x DN
• Outlet run ≥ 2 x DN
a0003861
Inlet and outlet runs
Adapters Suitable adapters can be used to install the sensor in larger-diameter pipes (see also DIN EN 545). The resultant
increase in the rate of flow improves measuring accuracy with very slow-moving fluids.
The nomogram shown here can be used to calculate the pressure loss caused by cross-section reduction.
Note!
The nomogram applies to fluids of viscosity similar to water.
1. Calculate the ratio of the diameters d/D.
2. From the nomogram, read off the pressure loss as a function of fluid velocity (downstream from the
reduction) and the d/D ratio.
a0003832
Pressure loss due to adapters
³5xDN³2xDN
Dd
max. 8°
100
10
0.5
d / D
[mbar]
0.6 0.7 0.8 0.9
1 m/s
2 m/s
3 m/s
4 m/s
5 m/s
6 m/s
7 m/s
8 m/s
1
Dosimag
Endress+Hauser 9
Operating conditions: Environment
Ambient temperature range –20 to +60 °C (sensor, transmitter)
Do not install the measuring device directly over equipment generating heat.
Storage temperature The storage temperature corresponds to the permitted ambient temperature range of the transmitter and sen-
sor.
Degree of protection Standard: IP 67 (NEMA 4X) for transmitter and sensor
Shock and vibration resistance Acceleration up to 2 g by analogy with IEC 60068-2-6
Operating conditions: Process
Medium temperature range Sensor:
• Process: –20 to +130 °C
• Cleaning: +150 °C/60 min for CIP and SIP processes
Seals:
• EPDM: –20 to +130 °C (max. +150 °C for cleaning)
• Silicone: –20 to +150 °C
• Viton: 0 to +150 °C
a0004805
TU = ambient temperature, TM = medium temperature
Light gray = standard medium temperature range, dark gray = medium temperature range for cleaning
Conductivity Minimum conductivity:
• 5 µS/cm for liquids generally
• 20 µS/cm for demineralized water
Limiting medium pressure
range (nominal pressure)
16 bar
Pressure tightness (liner) <1 mbar at a temperature range of +25 to +150 °C
0
-20
20
40
60
0-20 40 80 120 160
T [°C]M
T [°C]U
130 150
Dosimag
10 Endress+Hauser
Limiting flow The diameter of the pipe and the flow rate determine the nominal diameter of the sensor. The optimum velocity
of flow is 1 to 4 m/s. The velocity of flow (v), moreover, has to be matched to the physical properties of the
fluid:
• v < 2 m/s: for abrasive fluids such as cleaning agents, etc.
• v > 2 m/s: for fluids producing buildup such as liquids containing oil and sugar
Note!
Flow velocity can be increased, if necessary, by reducing the nominal diameter of the sensor.
Pressure loss • No pressure loss at DN 8 and DN 15 if the sensor is installed in a pipe of the same nominal diameter.
• Pressure losses for configurations incorporating adapters according to DIN EN 545 → Page 8.
Flow rate characteristic values - Dosimag (SI units)
Nominal diameter Recommended flow rate Factory settings
[mm] [inch] Max. full scale value Pulse value Low flow cutoff
(v ~ 0.04 m/s)
4 5/32" 0.14 l/sec 5 µl 2 l/h
8 5/16" 0.5 l/sec 20 µl 8 l/h
15* 1/2"* 1.2 l/sec 100 µl 26 l/h
15 1/2" 1.66 l/sec 100 µl 26 l/h
* = Conical version
Flow rate characteristic values - Dosimag (US units)
Nominal diameter Recommended flow rate Factory settings
[inch] [mm] Max. full scale value Pulse value Low flow cutoff
(v ~ 0.13 ft/s)
5/32" 4 0.035 gal/sec 0.0002 oz fl 0.009 gal/min
5/16" 8 0.13 gal/sec 0.001 oz fl 0.035 gal/min
1/2"* 15* 0.32 gal/sec 0.004 oz fl 0.12 gal/min
1/2" 15 0.44 gal/sec 0.004 oz fl 0.12 gal/min
* = Conical version
Dosimag
Endress+Hauser 11
Mechanical construction
Design / dimensions Dosimag dimensions
a0003864
Measuring tube dimensions
a0004874
DN L1 L2 Di di
DIN
[mm]
ANSI
[inch]
DIN
[mm] [mm] [mm] [mm]
4 5/32" 44 90 9 4.5
8 5/16" – 90 9 9
15* 1/2"* 20 90 16 12
15 1/2" – 90 16 16
Fitting length depends on process connections
* = Conical version
133
100
80
86
90
210
42
33.4
M6x4
12
7050
10
L1
L2
di
Di
Dosimag
12 Endress+Hauser
Process connection dimensions (with aseptic molded seal)
Weld nipples
1.4404 / 316L
**H**-U***********
Sensor Suits pipe di G L H x B
DN [mm] DIN 11850 [mm] [mm] [mm] [mm]
a0003870
4 / 8 14 x 2 9 14 23.3 60 x 42
15 / 15* 20 x 2 16 20 23.3 60 x 42
– Fitting length = (2 x L) + 86 mm
– If pigs are used for cleaning, it is essential to take the inside diameters of the measuring tube (→ Page 11) and process
connection (di) into account.
– * = Conical version
G
di
L
HxB
Weld nipple ODT/SMS
1.4404 / 316L
**H**-V***********
Sensor Suits pipe di G L H x B
DN [mm] OD/SMS [mm] [mm] [mm] [mm]
a0003871
4 / 8 12.7 x 1.65 9 12.7 16.1 60 x 42
15 / 15* 19.1 x 1.65 16 19.1 16.1 60 x 42
– Fitting length = (2 x L) + 86 mm
– If pigs are used for cleaning, it is essential to take the inside diameters of the measuring tube (→ Page 11) and process
connection (di) into account.
– * = Conical version
G
di
L
HxB
Tri-Clamp L14 AM7
1.4404 / 316L
**H**-1***********
Sensor Suits pipe di G L H x B
DN [mm] OD [mm] [mm] [mm] [mm]
a0003872
4 / 8 Tube 12.7 x 1.65 (ODT 1/2") 9.4 25.0 28.5 60 x 42
15 / 15* Tube 19.1 x 1.65 (ODT 3/4") 15.8 25.0 28.5 60 x 42
– Fitting length = (2 x L) + 86 mm
– If pigs are used for cleaning, it is essential to take the inside diameters of the measuring tube (→ Page 11) and process
connection (di) into account.
– * = Conical version
G
di
L
HxB
Tri-Clamp L14 AM7
1.4404 / 316L
**H**-2***********
Sensor Suits pipe d1 d2 G L H x B
DN [mm] OD [mm] [mm] [mm] [mm] [mm]
a0003878
4 / 8 Tube 19.1 x 1.65 (ODT 3/4") 9 15.8 25.0 28.5 60 x 42
– Fitting length = (2 x L) + 86 mm
– If pigs are used for cleaning, it is essential to take the inside diameters of the measuring tube and process connection
(d1 or d2) into account!
G
d2
L
HxB
d1
Dosimag
Endress+Hauser 13
Weight Approx. 2.8 kg
Material Transmitter housing:
1.4308/304
Sensor housing:
Acid and alkali-resistant outer surface; stainless steel 1.4308/304
Measuring tube:
Stainless steel 1.4301/304 with PFA lining
Process connection:
• Weld nipples → stainless steel 1.4404/316L
• Aseptic weld nipples → stainless steel 1.4404/316L
• Tri-Clamp → stainless steel 1.4404/316L
Measuring electrodes:
Standard: 1.4435 (AISI 316L); option: Alloy C-22
Seals:
Molded seal (EPDM, silicone, Viton)
Material load diagram Weld nipple (DIN 11850, ODT/SMS), Clamp L14 AM7
Material: 1.4404 / 316L (with molded seal)
A0005121
Process connection Dairy fittings: weld nipple (DIN 11850, ODT/SMS), Tri-Clamp L14 AM7
Surface roughness • Measuring tube lining with PFA: ≤0.4 µm
• Electrodes (1.4435 (AISI 316L), Alloy C-22): 0.3 to 0.5 µm
• Process connection Dosimag: ≤0.8 µm
(all information refers to wetted parts)
User interface
Display elements Dosimag does not have a display or display elements.
Remote operation Operation takes place via the “ToF-Tool - FieldTool Package” configuration and service program from
Endress+Hauser. This can be used to configure functions and read off measured values.
Certificates and approvals
CE mark The measuring system is in conformity with the statutory requirements of the EC Directives. Endress+Hauser
confirms successful testing of the device by affixing to it the CE mark.
0
10
5
20
15 PN16
[bar]
-60 -40 -20 0 20 40 60 80 100 120 140 160 180 [°C]
Dosimag
14 Endress+Hauser
C-tick mark The measuring system meets the EMC requirements of the Australian Communication and Media Authority
(ACMA).
Sanitary compatibility 3-A
EHEDG
Seals in conformity with FDA
Other approvals Additional approvals from the certification bodies FM (USA) and CSA (Canada) are available for all Dosimag
devices for use in non-explosion protected atmospheres. Contact your Endress+Hauser sales office for more
information.
Pressure measuring device
approval
All Dosimag devices correspond to Article 3(3) of the EC Directive 97/23/EC (Pressure Equipment Directive)
and have been designed and manufactured according to good engineering practice.
Other standards and guide-
lines
EN 60529:
Degrees of protection by housing (IP code)
EN 61010-1:
Protection Measures for Electrical Equipment for Measurement, Control, Regulation and Laboratory Proce-
dures.
EN 61326 (IEC 1326):
Electromagnetic compatibility (EMC requirements)
CSA-C22.2 No. 142-M1987
Process Control Equipment
CAN/CSA-C22.2 No. 1010.1-92
Safety Requirements for Electrical Equipment for Measuring, Control and Laboratory Use. Pollution degree 2,
Installation Category I
ANSI/ISA-S82.01
Safety Standard for Electrical and Electronic Test, Measuring, Controlling and related Equipment - General
Requirements. Pollution degree 2, Installation Category I
Ordering information
The Endress+Hauser service organization can provide detailed ordering information and information on spe-
cific order codes on request.
Accessories
Various accessories and spare parts, which can be ordered separately from Endress+Hauser, are available for
the transmitter and sensor. The Endress+Hauser service organization can provide detailed information on the
specific order codes on request.
Accessories Description Ordering code
Dosimag transmitter Transmitter for replacement or for storage. The following
specifications can be indicated by means of the order
code:
– Approvals
– Housing
– Cables
– Cable connection
– Power supply
– Software functions
– Outputs/inputs
5BXXX-XXXXX*******
Housing seal To seal the transmitter 50102857
Dosimag
Endress+Hauser 15
Supplementary Documentation
❑ Flow Measurement (FA005/06/en)
❑ Operating Instructions Dosimag (BA098/06/en)
Registered trademarks
TRI-CLAMP ®
Registered trademark of Ladish & Co., Inc., Kenosha, USA
HistoROM™, M-DAT®
, ToF Tool - Fieldtool®
Package, Applicator®
Registered or registration-pending trademarks of Endress+Hauser Flowtec AG, Reinach, CH
Seal set For regular replacement of the seals on the process con-
nections.
DK5HS – ***
Mounting kit Mounting kit comprising:
– 2 process connections
– Threaded fasteners
– Seals
DKH** – ****
Adapter connection Adapter connections for installing Dosimag on other pro-
cess connections
DK5HA – ******
Conversion kit Kit for converting Dosimag A to Dosimag DK5UP – H
RSE8 adapter Lumberg RSE8 connection jack, 8-pole adapter (RSE8),
24 V DC, pulse, status
50107169
RSE5 adapter Lumberg RSE8 connection jack, 5-pole adapter (RSE5),
24 V DC, pulse, status
50107168
RSE4 adapter Lumberg RSE8 connection jack, 4-pole adapter (RSE4),
24 V DC, pulse
50107167
RSE8 supply cable Lumberg cable RKWTN8-56/5 P92 50107895
ToF Tool - FieldTool
Package
Configuration and service software for flowmeters in the
field:
– Commissioning, maintenance analysis
– Configuring measuring devices
– Service functions
– Visualization of process data
– Troubleshooting
Contact your Endress+Hauser representative for more
information.
DXS10 - * * * * *
FXA 193 Service interface connecting cable from the device to the
PC for using the “ToF Tool FieldTool Package” operating
software
FXA193 – *
Service adapter Adapter for connecting FXA 193 to the device. 50106443
Applicator Software for selecting and configuring flowmeters.
Applicator can be downloaded from the Internet or
ordered on CD-ROM for installation on a local PC.
Contact your Endress+Hauser representative for more
information.
DKA80 - *
Accessories Description Ordering code
Dosimag
International Headquarters
Endress+Hauser
GmbH+Co. KG
Instruments International
Colmarer Str. 6
79576 Weil am Rhein
Deutschland
Tel. +49 76 21 9 75 02
Fax +49 76 21 9 75 34 5
www.endress.com
info@ii.endress.com
TI066D/06/en/11.05
71008913
FM+SGML 6.0 ProMoDo

More Related Content

Viewers also liked (7)

pH electrodes-pH sensors CPS41
pH electrodes-pH sensors CPS41pH electrodes-pH sensors CPS41
pH electrodes-pH sensors CPS41
 
Microwave barrier- Soliwave FQR56/FDR56
Microwave barrier- Soliwave FQR56/FDR56Microwave barrier- Soliwave FQR56/FDR56
Microwave barrier- Soliwave FQR56/FDR56
 
Microwave barrier-Soliwave M FQR50 / FDR50
Microwave barrier-Soliwave M FQR50 / FDR50Microwave barrier-Soliwave M FQR50 / FDR50
Microwave barrier-Soliwave M FQR50 / FDR50
 
pH.ORP Transmitter-Analog and digital glass
pH.ORP Transmitter-Analog and digital glasspH.ORP Transmitter-Analog and digital glass
pH.ORP Transmitter-Analog and digital glass
 
Flow In Pipes
Flow In PipesFlow In Pipes
Flow In Pipes
 
Flow charts for pipes
Flow charts for pipesFlow charts for pipes
Flow charts for pipes
 
Water flow pipe sizes
Water flow pipe sizesWater flow pipe sizes
Water flow pipe sizes
 

Similar to Flow Measurement for Filling Applications

Proline promag 10 l endress+hauser datasheet-electromagnetic flowmeter
Proline promag 10 l endress+hauser datasheet-electromagnetic flowmeterProline promag 10 l endress+hauser datasheet-electromagnetic flowmeter
Proline promag 10 l endress+hauser datasheet-electromagnetic flowmeterENVIMART
 
Proline promag 10 w endress+hauser datasheet-electromagnetic flowmeter
Proline promag 10 w endress+hauser datasheet-electromagnetic flowmeterProline promag 10 w endress+hauser datasheet-electromagnetic flowmeter
Proline promag 10 w endress+hauser datasheet-electromagnetic flowmeterENVIMART
 
Proline promag 10 p endress+hauser datasheet-electromagnetic flowmeter
Proline promag 10 p endress+hauser datasheet-electromagnetic flowmeterProline promag 10 p endress+hauser datasheet-electromagnetic flowmeter
Proline promag 10 p endress+hauser datasheet-electromagnetic flowmeterENVIMART
 
Proline promag 50 l endress+hauser datasheet-electromagnetic flowmeter
Proline promag 50 l endress+hauser datasheet-electromagnetic flowmeterProline promag 50 l endress+hauser datasheet-electromagnetic flowmeter
Proline promag 50 l endress+hauser datasheet-electromagnetic flowmeterENVIMART
 
Proline prosonic flow 91 w endress+hauser datasheet-ultrasonic flowmeter
Proline prosonic flow 91 w endress+hauser datasheet-ultrasonic flowmeterProline prosonic flow 91 w endress+hauser datasheet-ultrasonic flowmeter
Proline prosonic flow 91 w endress+hauser datasheet-ultrasonic flowmeterENVIMART
 
Proline promag 50 w endress+hauser datasheet-electromagnetic flowmeter
Proline promag 50 w endress+hauser datasheet-electromagnetic flowmeterProline promag 50 w endress+hauser datasheet-electromagnetic flowmeter
Proline promag 50 w endress+hauser datasheet-electromagnetic flowmeterENVIMART
 
Proline promag 55 s endress+hauser datasheet-electromagnetic flowmeter
Proline promag 55 s endress+hauser datasheet-electromagnetic flowmeterProline promag 55 s endress+hauser datasheet-electromagnetic flowmeter
Proline promag 55 s endress+hauser datasheet-electromagnetic flowmeterENVIMART
 
Proline t mass b 150-endress+hauser datasheet-thermal mass flow measuring system
Proline t mass b 150-endress+hauser datasheet-thermal mass flow measuring systemProline t mass b 150-endress+hauser datasheet-thermal mass flow measuring system
Proline t mass b 150-endress+hauser datasheet-thermal mass flow measuring systemENVIMART
 

Similar to Flow Measurement for Filling Applications (20)

Proline Promag 10L-Electromagnetic Flowmeter
Proline Promag 10L-Electromagnetic FlowmeterProline Promag 10L-Electromagnetic Flowmeter
Proline Promag 10L-Electromagnetic Flowmeter
 
Proline promag 10 l endress+hauser datasheet-electromagnetic flowmeter
Proline promag 10 l endress+hauser datasheet-electromagnetic flowmeterProline promag 10 l endress+hauser datasheet-electromagnetic flowmeter
Proline promag 10 l endress+hauser datasheet-electromagnetic flowmeter
 
Ga dat cam bien
Ga dat cam bien Ga dat cam bien
Ga dat cam bien
 
Proline promag 10 w endress+hauser datasheet-electromagnetic flowmeter
Proline promag 10 w endress+hauser datasheet-electromagnetic flowmeterProline promag 10 w endress+hauser datasheet-electromagnetic flowmeter
Proline promag 10 w endress+hauser datasheet-electromagnetic flowmeter
 
Proline promag 10 p endress+hauser datasheet-electromagnetic flowmeter
Proline promag 10 p endress+hauser datasheet-electromagnetic flowmeterProline promag 10 p endress+hauser datasheet-electromagnetic flowmeter
Proline promag 10 p endress+hauser datasheet-electromagnetic flowmeter
 
Proline Promag 10P-Electromagnetic Flowmeter
Proline Promag 10P-Electromagnetic FlowmeterProline Promag 10P-Electromagnetic Flowmeter
Proline Promag 10P-Electromagnetic Flowmeter
 
Electromagnetic flowmeter - Proline Promag 10H
Electromagnetic flowmeter - Proline Promag 10HElectromagnetic flowmeter - Proline Promag 10H
Electromagnetic flowmeter - Proline Promag 10H
 
Ultrasonic flowmeter-Proline Prosonic Flow 92F
Ultrasonic flowmeter-Proline Prosonic Flow 92FUltrasonic flowmeter-Proline Prosonic Flow 92F
Ultrasonic flowmeter-Proline Prosonic Flow 92F
 
Flowmeter Promag 10W-Electromagnetic-Water and Waste Water
Flowmeter Promag 10W-Electromagnetic-Water and Waste WaterFlowmeter Promag 10W-Electromagnetic-Water and Waste Water
Flowmeter Promag 10W-Electromagnetic-Water and Waste Water
 
Electromagnetic flowmeter - Proline promag 10D
Electromagnetic flowmeter - Proline promag 10DElectromagnetic flowmeter - Proline promag 10D
Electromagnetic flowmeter - Proline promag 10D
 
Electromagnetic flowmeter - Proline Promag 50D
Electromagnetic flowmeter - Proline Promag 50DElectromagnetic flowmeter - Proline Promag 50D
Electromagnetic flowmeter - Proline Promag 50D
 
Proline Promag 50L-Electromagnetic Flowmeter
Proline Promag 50L-Electromagnetic FlowmeterProline Promag 50L-Electromagnetic Flowmeter
Proline Promag 50L-Electromagnetic Flowmeter
 
Proline promag 50 l endress+hauser datasheet-electromagnetic flowmeter
Proline promag 50 l endress+hauser datasheet-electromagnetic flowmeterProline promag 50 l endress+hauser datasheet-electromagnetic flowmeter
Proline promag 50 l endress+hauser datasheet-electromagnetic flowmeter
 
Proline prosonic flow 91 w endress+hauser datasheet-ultrasonic flowmeter
Proline prosonic flow 91 w endress+hauser datasheet-ultrasonic flowmeterProline prosonic flow 91 w endress+hauser datasheet-ultrasonic flowmeter
Proline prosonic flow 91 w endress+hauser datasheet-ultrasonic flowmeter
 
Ultrasonic flowmeter-Drinking water and process water
Ultrasonic flowmeter-Drinking water and process waterUltrasonic flowmeter-Drinking water and process water
Ultrasonic flowmeter-Drinking water and process water
 
Electromagnetic flowmeter - Proline Promag 55H
Electromagnetic flowmeter - Proline Promag 55HElectromagnetic flowmeter - Proline Promag 55H
Electromagnetic flowmeter - Proline Promag 55H
 
Proline promag 50 w endress+hauser datasheet-electromagnetic flowmeter
Proline promag 50 w endress+hauser datasheet-electromagnetic flowmeterProline promag 50 w endress+hauser datasheet-electromagnetic flowmeter
Proline promag 50 w endress+hauser datasheet-electromagnetic flowmeter
 
Ultrasonic flowmeter- Drinking water, wastewater
Ultrasonic flowmeter- Drinking water, wastewaterUltrasonic flowmeter- Drinking water, wastewater
Ultrasonic flowmeter- Drinking water, wastewater
 
Proline promag 55 s endress+hauser datasheet-electromagnetic flowmeter
Proline promag 55 s endress+hauser datasheet-electromagnetic flowmeterProline promag 55 s endress+hauser datasheet-electromagnetic flowmeter
Proline promag 55 s endress+hauser datasheet-electromagnetic flowmeter
 
Proline t mass b 150-endress+hauser datasheet-thermal mass flow measuring system
Proline t mass b 150-endress+hauser datasheet-thermal mass flow measuring systemProline t mass b 150-endress+hauser datasheet-thermal mass flow measuring system
Proline t mass b 150-endress+hauser datasheet-thermal mass flow measuring system
 

More from Edress Hauser: Flow Meter, Level, Pressure, Temperature

More from Edress Hauser: Flow Meter, Level, Pressure, Temperature (20)

Liquiphant M Density and Density Computer FML621
Liquiphant M Density and Density Computer FML621Liquiphant M Density and Density Computer FML621
Liquiphant M Density and Density Computer FML621
 
Level Limit Switch-Soliswitch FTE31
Level Limit Switch-Soliswitch FTE31Level Limit Switch-Soliswitch FTE31
Level Limit Switch-Soliswitch FTE31
 
Level limit switch-Soliswitch FTE30
Level limit switch-Soliswitch FTE30Level limit switch-Soliswitch FTE30
Level limit switch-Soliswitch FTE30
 
Flow indicator for bulk solids-Solimotion FTR20
Flow indicator for bulk solids-Solimotion FTR20Flow indicator for bulk solids-Solimotion FTR20
Flow indicator for bulk solids-Solimotion FTR20
 
Level measurement with Prothermo NMT532
Level measurement with Prothermo NMT532 Level measurement with Prothermo NMT532
Level measurement with Prothermo NMT532
 
High accuracy liquid level measurement-Proservo NMS5
High accuracy liquid level measurement-Proservo NMS5High accuracy liquid level measurement-Proservo NMS5
High accuracy liquid level measurement-Proservo NMS5
 
Conductive Limit Detection-Rod Probe 11263
Conductive Limit Detection-Rod Probe 11263Conductive Limit Detection-Rod Probe 11263
Conductive Limit Detection-Rod Probe 11263
 
FTW360 compact pump protector
FTW360 compact pump protectorFTW360 compact pump protector
FTW360 compact pump protector
 
Conductive-Point level switch for multiple point
Conductive-Point level switch for multiple pointConductive-Point level switch for multiple point
Conductive-Point level switch for multiple point
 
Level limit switch compact design-Liquiphant T FTL20H
Level limit switch compact design-Liquiphant T FTL20HLevel limit switch compact design-Liquiphant T FTL20H
Level limit switch compact design-Liquiphant T FTL20H
 
Vibronic-Point level switch for all kinds of liquids
Vibronic-Point level switch for all kinds of liquidsVibronic-Point level switch for all kinds of liquids
Vibronic-Point level switch for all kinds of liquids
 
Vibronic-Point level switch-Liquiphant M FTL51C
Vibronic-Point level switch-Liquiphant M FTL51CVibronic-Point level switch-Liquiphant M FTL51C
Vibronic-Point level switch-Liquiphant M FTL51C
 
Level limit switch for liquids-Liquiphant T FTL20
Level limit switch for liquids-Liquiphant T FTL20Level limit switch for liquids-Liquiphant T FTL20
Level limit switch for liquids-Liquiphant T FTL20
 
Radiometric measuring technology-Gammapilot FTG20
Radiometric measuring technology-Gammapilot FTG20Radiometric measuring technology-Gammapilot FTG20
Radiometric measuring technology-Gammapilot FTG20
 
Differential pressure measurement-Level transmiter
Differential pressure measurement-Level transmiterDifferential pressure measurement-Level transmiter
Differential pressure measurement-Level transmiter
 
Hydrostatic Level Measurement-Deltapilot S DB50/50L/51/52/53
Hydrostatic Level Measurement-Deltapilot S DB50/50L/51/52/53Hydrostatic Level Measurement-Deltapilot S DB50/50L/51/52/53
Hydrostatic Level Measurement-Deltapilot S DB50/50L/51/52/53
 
Hydrostatic level measurement-Deltapilot M FMB50/51/52/53
Hydrostatic level measurement-Deltapilot M FMB50/51/52/53Hydrostatic level measurement-Deltapilot M FMB50/51/52/53
Hydrostatic level measurement-Deltapilot M FMB50/51/52/53
 
Capacitance level measurement-Liquicap M FMI51, FMI52
Capacitance level measurement-Liquicap M FMI51, FMI52Capacitance level measurement-Liquicap M FMI51, FMI52
Capacitance level measurement-Liquicap M FMI51, FMI52
 
Ultrasonic Level Measurement-Prosonic M FMU40/41/42/43/44
Ultrasonic Level Measurement-Prosonic M FMU40/41/42/43/44Ultrasonic Level Measurement-Prosonic M FMU40/41/42/43/44
Ultrasonic Level Measurement-Prosonic M FMU40/41/42/43/44
 
Ultrasonic Level Measurement-Prosonic T FMU30
Ultrasonic Level Measurement-Prosonic T FMU30Ultrasonic Level Measurement-Prosonic T FMU30
Ultrasonic Level Measurement-Prosonic T FMU30
 

Recently uploaded

大学学位办理《原版美国USD学位证书》圣地亚哥大学毕业证制作成绩单修改
大学学位办理《原版美国USD学位证书》圣地亚哥大学毕业证制作成绩单修改大学学位办理《原版美国USD学位证书》圣地亚哥大学毕业证制作成绩单修改
大学学位办理《原版美国USD学位证书》圣地亚哥大学毕业证制作成绩单修改atducpo
 
🔝|97111༒99012🔝 Call Girls In {Delhi} Cr Park ₹5.5k Cash Payment With Room De...
🔝|97111༒99012🔝 Call Girls In  {Delhi} Cr Park ₹5.5k Cash Payment With Room De...🔝|97111༒99012🔝 Call Girls In  {Delhi} Cr Park ₹5.5k Cash Payment With Room De...
🔝|97111༒99012🔝 Call Girls In {Delhi} Cr Park ₹5.5k Cash Payment With Room De...Diya Sharma
 
Albania Vs Spain Albania is Loaded with Defensive Talent on their Roster.docx
Albania Vs Spain Albania is Loaded with Defensive Talent on their Roster.docxAlbania Vs Spain Albania is Loaded with Defensive Talent on their Roster.docx
Albania Vs Spain Albania is Loaded with Defensive Talent on their Roster.docxWorld Wide Tickets And Hospitality
 
08448380779 Call Girls In Lajpat Nagar Women Seeking Men
08448380779 Call Girls In Lajpat Nagar Women Seeking Men08448380779 Call Girls In Lajpat Nagar Women Seeking Men
08448380779 Call Girls In Lajpat Nagar Women Seeking MenDelhi Call girls
 
TAM Sports_IPL 17 Till Match 37_Celebrity Endorsement _Report.pdf
TAM Sports_IPL 17 Till Match 37_Celebrity Endorsement _Report.pdfTAM Sports_IPL 17 Till Match 37_Celebrity Endorsement _Report.pdf
TAM Sports_IPL 17 Till Match 37_Celebrity Endorsement _Report.pdfSocial Samosa
 
Spain Vs Italy 20 players confirmed for Spain's Euro 2024 squad, and three po...
Spain Vs Italy 20 players confirmed for Spain's Euro 2024 squad, and three po...Spain Vs Italy 20 players confirmed for Spain's Euro 2024 squad, and three po...
Spain Vs Italy 20 players confirmed for Spain's Euro 2024 squad, and three po...World Wide Tickets And Hospitality
 
大学假文凭《原版英国Imperial文凭》帝国理工学院毕业证制作成绩单修改
大学假文凭《原版英国Imperial文凭》帝国理工学院毕业证制作成绩单修改大学假文凭《原版英国Imperial文凭》帝国理工学院毕业证制作成绩单修改
大学假文凭《原版英国Imperial文凭》帝国理工学院毕业证制作成绩单修改atducpo
 
ALL NFL NETWORK CONTACTS- April 29, 2024
ALL NFL NETWORK CONTACTS- April 29, 2024ALL NFL NETWORK CONTACTS- April 29, 2024
ALL NFL NETWORK CONTACTS- April 29, 2024Brian Slack
 
Call Girls in Dhaula Kuan 💯Call Us 🔝8264348440🔝
Call Girls in Dhaula Kuan 💯Call Us 🔝8264348440🔝Call Girls in Dhaula Kuan 💯Call Us 🔝8264348440🔝
Call Girls in Dhaula Kuan 💯Call Us 🔝8264348440🔝soniya singh
 
Chennai Call Girls Anna Nagar Phone 🍆 8250192130 👅 celebrity escorts service
Chennai Call Girls Anna Nagar Phone 🍆 8250192130 👅 celebrity escorts serviceChennai Call Girls Anna Nagar Phone 🍆 8250192130 👅 celebrity escorts service
Chennai Call Girls Anna Nagar Phone 🍆 8250192130 👅 celebrity escorts servicevipmodelshub1
 
Serbia vs England Vanja Milinkovic-Savic's Stellar Performance in Euro 2024 P...
Serbia vs England Vanja Milinkovic-Savic's Stellar Performance in Euro 2024 P...Serbia vs England Vanja Milinkovic-Savic's Stellar Performance in Euro 2024 P...
Serbia vs England Vanja Milinkovic-Savic's Stellar Performance in Euro 2024 P...Eticketing.co
 
Croatia vs Albania Clash of Euro Cup 2024 Squad Preparations and Euro Cup Dre...
Croatia vs Albania Clash of Euro Cup 2024 Squad Preparations and Euro Cup Dre...Croatia vs Albania Clash of Euro Cup 2024 Squad Preparations and Euro Cup Dre...
Croatia vs Albania Clash of Euro Cup 2024 Squad Preparations and Euro Cup Dre...Eticketing.co
 
Resultados del Campeonato mundial de Marcha por equipos Antalya 2024
Resultados del Campeonato mundial de Marcha por equipos Antalya 2024Resultados del Campeonato mundial de Marcha por equipos Antalya 2024
Resultados del Campeonato mundial de Marcha por equipos Antalya 2024Judith Chuquipul
 
VIP Kolkata Call Girl Liluah 👉 8250192130 Available With Room
VIP Kolkata Call Girl Liluah 👉 8250192130  Available With RoomVIP Kolkata Call Girl Liluah 👉 8250192130  Available With Room
VIP Kolkata Call Girl Liluah 👉 8250192130 Available With Roomdivyansh0kumar0
 
08448380779 Call Girls In International Airport Women Seeking Men
08448380779 Call Girls In International Airport Women Seeking Men08448380779 Call Girls In International Airport Women Seeking Men
08448380779 Call Girls In International Airport Women Seeking MenDelhi Call girls
 
Atlanta Dream Exec Dan Gadd on Driving Fan Engagement and Growth, Serving the...
Atlanta Dream Exec Dan Gadd on Driving Fan Engagement and Growth, Serving the...Atlanta Dream Exec Dan Gadd on Driving Fan Engagement and Growth, Serving the...
Atlanta Dream Exec Dan Gadd on Driving Fan Engagement and Growth, Serving the...Neil Horowitz
 
JORNADA 4 LIGA MURO 2024TUXTEPEC1234.pdf
JORNADA 4 LIGA MURO 2024TUXTEPEC1234.pdfJORNADA 4 LIGA MURO 2024TUXTEPEC1234.pdf
JORNADA 4 LIGA MURO 2024TUXTEPEC1234.pdfArturo Pacheco Alvarez
 

Recently uploaded (20)

大学学位办理《原版美国USD学位证书》圣地亚哥大学毕业证制作成绩单修改
大学学位办理《原版美国USD学位证书》圣地亚哥大学毕业证制作成绩单修改大学学位办理《原版美国USD学位证书》圣地亚哥大学毕业证制作成绩单修改
大学学位办理《原版美国USD学位证书》圣地亚哥大学毕业证制作成绩单修改
 
🔝|97111༒99012🔝 Call Girls In {Delhi} Cr Park ₹5.5k Cash Payment With Room De...
🔝|97111༒99012🔝 Call Girls In  {Delhi} Cr Park ₹5.5k Cash Payment With Room De...🔝|97111༒99012🔝 Call Girls In  {Delhi} Cr Park ₹5.5k Cash Payment With Room De...
🔝|97111༒99012🔝 Call Girls In {Delhi} Cr Park ₹5.5k Cash Payment With Room De...
 
Albania Vs Spain Albania is Loaded with Defensive Talent on their Roster.docx
Albania Vs Spain Albania is Loaded with Defensive Talent on their Roster.docxAlbania Vs Spain Albania is Loaded with Defensive Talent on their Roster.docx
Albania Vs Spain Albania is Loaded with Defensive Talent on their Roster.docx
 
08448380779 Call Girls In Lajpat Nagar Women Seeking Men
08448380779 Call Girls In Lajpat Nagar Women Seeking Men08448380779 Call Girls In Lajpat Nagar Women Seeking Men
08448380779 Call Girls In Lajpat Nagar Women Seeking Men
 
Call Girls Service Noida Extension @9999965857 Delhi 🫦 No Advance VVIP 🍎 SER...
Call Girls Service Noida Extension @9999965857 Delhi 🫦 No Advance  VVIP 🍎 SER...Call Girls Service Noida Extension @9999965857 Delhi 🫦 No Advance  VVIP 🍎 SER...
Call Girls Service Noida Extension @9999965857 Delhi 🫦 No Advance VVIP 🍎 SER...
 
TAM Sports_IPL 17 Till Match 37_Celebrity Endorsement _Report.pdf
TAM Sports_IPL 17 Till Match 37_Celebrity Endorsement _Report.pdfTAM Sports_IPL 17 Till Match 37_Celebrity Endorsement _Report.pdf
TAM Sports_IPL 17 Till Match 37_Celebrity Endorsement _Report.pdf
 
Spain Vs Italy 20 players confirmed for Spain's Euro 2024 squad, and three po...
Spain Vs Italy 20 players confirmed for Spain's Euro 2024 squad, and three po...Spain Vs Italy 20 players confirmed for Spain's Euro 2024 squad, and three po...
Spain Vs Italy 20 players confirmed for Spain's Euro 2024 squad, and three po...
 
大学假文凭《原版英国Imperial文凭》帝国理工学院毕业证制作成绩单修改
大学假文凭《原版英国Imperial文凭》帝国理工学院毕业证制作成绩单修改大学假文凭《原版英国Imperial文凭》帝国理工学院毕业证制作成绩单修改
大学假文凭《原版英国Imperial文凭》帝国理工学院毕业证制作成绩单修改
 
ALL NFL NETWORK CONTACTS- April 29, 2024
ALL NFL NETWORK CONTACTS- April 29, 2024ALL NFL NETWORK CONTACTS- April 29, 2024
ALL NFL NETWORK CONTACTS- April 29, 2024
 
Call Girls In Vasundhara 📱 9999965857 🤩 Delhi 🫦 HOT AND SEXY VVIP 🍎 SERVICE
Call Girls In Vasundhara 📱  9999965857  🤩 Delhi 🫦 HOT AND SEXY VVIP 🍎 SERVICECall Girls In Vasundhara 📱  9999965857  🤩 Delhi 🫦 HOT AND SEXY VVIP 🍎 SERVICE
Call Girls In Vasundhara 📱 9999965857 🤩 Delhi 🫦 HOT AND SEXY VVIP 🍎 SERVICE
 
Call Girls in Dhaula Kuan 💯Call Us 🔝8264348440🔝
Call Girls in Dhaula Kuan 💯Call Us 🔝8264348440🔝Call Girls in Dhaula Kuan 💯Call Us 🔝8264348440🔝
Call Girls in Dhaula Kuan 💯Call Us 🔝8264348440🔝
 
Chennai Call Girls Anna Nagar Phone 🍆 8250192130 👅 celebrity escorts service
Chennai Call Girls Anna Nagar Phone 🍆 8250192130 👅 celebrity escorts serviceChennai Call Girls Anna Nagar Phone 🍆 8250192130 👅 celebrity escorts service
Chennai Call Girls Anna Nagar Phone 🍆 8250192130 👅 celebrity escorts service
 
Serbia vs England Vanja Milinkovic-Savic's Stellar Performance in Euro 2024 P...
Serbia vs England Vanja Milinkovic-Savic's Stellar Performance in Euro 2024 P...Serbia vs England Vanja Milinkovic-Savic's Stellar Performance in Euro 2024 P...
Serbia vs England Vanja Milinkovic-Savic's Stellar Performance in Euro 2024 P...
 
Croatia vs Albania Clash of Euro Cup 2024 Squad Preparations and Euro Cup Dre...
Croatia vs Albania Clash of Euro Cup 2024 Squad Preparations and Euro Cup Dre...Croatia vs Albania Clash of Euro Cup 2024 Squad Preparations and Euro Cup Dre...
Croatia vs Albania Clash of Euro Cup 2024 Squad Preparations and Euro Cup Dre...
 
Resultados del Campeonato mundial de Marcha por equipos Antalya 2024
Resultados del Campeonato mundial de Marcha por equipos Antalya 2024Resultados del Campeonato mundial de Marcha por equipos Antalya 2024
Resultados del Campeonato mundial de Marcha por equipos Antalya 2024
 
VIP Kolkata Call Girl Liluah 👉 8250192130 Available With Room
VIP Kolkata Call Girl Liluah 👉 8250192130  Available With RoomVIP Kolkata Call Girl Liluah 👉 8250192130  Available With Room
VIP Kolkata Call Girl Liluah 👉 8250192130 Available With Room
 
Call Girls 🫤 Paharganj ➡️ 9999965857 ➡️ Delhi 🫦 Russian Escorts FULL ENJOY
Call Girls 🫤 Paharganj ➡️ 9999965857  ➡️ Delhi 🫦  Russian Escorts FULL ENJOYCall Girls 🫤 Paharganj ➡️ 9999965857  ➡️ Delhi 🫦  Russian Escorts FULL ENJOY
Call Girls 🫤 Paharganj ➡️ 9999965857 ➡️ Delhi 🫦 Russian Escorts FULL ENJOY
 
08448380779 Call Girls In International Airport Women Seeking Men
08448380779 Call Girls In International Airport Women Seeking Men08448380779 Call Girls In International Airport Women Seeking Men
08448380779 Call Girls In International Airport Women Seeking Men
 
Atlanta Dream Exec Dan Gadd on Driving Fan Engagement and Growth, Serving the...
Atlanta Dream Exec Dan Gadd on Driving Fan Engagement and Growth, Serving the...Atlanta Dream Exec Dan Gadd on Driving Fan Engagement and Growth, Serving the...
Atlanta Dream Exec Dan Gadd on Driving Fan Engagement and Growth, Serving the...
 
JORNADA 4 LIGA MURO 2024TUXTEPEC1234.pdf
JORNADA 4 LIGA MURO 2024TUXTEPEC1234.pdfJORNADA 4 LIGA MURO 2024TUXTEPEC1234.pdf
JORNADA 4 LIGA MURO 2024TUXTEPEC1234.pdf
 

Flow Measurement for Filling Applications

  • 1. TI066D/06/en 71008913 Technical Information Dosimag Electromagnetic Flow Measuring System Volume flow measuring system for filling applications Application Electromagnetic flowmeter for measuring in filling applications. • Flow up to 1.66 l/s • Fluid temperature up to +130 °C • Process pressures up to 16 bar • CIP/SIP cleanable • Stainless steel housing All liquids with a conductivity of ≥5 µS/cm can be measured in the following sectors for example: • Food industry • Cosmetics industry • Pharmaceuticals industry • Chemicals industry Approvals in the food industry/hygiene sector: • 3-A approval, EHEDG-tested, in conformity with FDA Application-specific lining material: • PFA Your benefits Dosimag guarantees the highest level of accuracy and repeatability even for short measuring times. The com- pact housing shape means the units can be arranged very close together in filling plants. The “Batchline” concept comprises additionally: • Identical process connections enable an uncomplica- ted exchange between “Dosimag” and the Coriolis mass flowmeter “Dosimass” • Uniform operating concept using the “ToF Tool - Fieldtool Package” operating software: – Graphic display of measured values for detailed trend analysis and optimization of the filling process – Complete plant documentation with device config- uration and filling diagrams can be created The Dosimag flowmeter offers: • Simple installation and commissioning • Insensitivity to pipe vibrations
  • 2. Dosimag 2 Endress+Hauser Table of contents Function and system design. . . . . . . . . . . . . . . . . . . . . 3 Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Operable flow range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Signal on alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Low flow cutoff . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Galvanic isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Switching output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Power supply. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Electrical connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Supply voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Cable connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Cable specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Power consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Power supply failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Potential equalization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Performance characteristics. . . . . . . . . . . . . . . . . . . . . 4 Reference operating conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Maximum measured error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Repeatability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Operating conditions: Installation . . . . . . . . . . . . . . . . 5 Installation instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Inlet and outlet runs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Adapters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Operating conditions: Environment. . . . . . . . . . . . . . . 9 Ambient temperature range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Shock and vibration resistance. . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Operating conditions: Process . . . . . . . . . . . . . . . . . . . 9 Medium temperature range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Conductivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Limiting medium pressure range (nominal pressure) 9 Pressure tightness (liner) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Limiting flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Pressure loss . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Mechanical construction . . . . . . . . . . . . . . . . . . . . . . 11 Design / dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Material load diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Process connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Surface roughness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 User interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Display elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Remote operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Certificates and approvals . . . . . . . . . . . . . . . . . . . . . 13 CE mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 C-tick mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Sanitary compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Other approvals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Pressure measuring device approval . . . . . . . . . . . . . . . . . . . . . . 14 Other standards and guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Ordering information. . . . . . . . . . . . . . . . . . . . . . . . . 14 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Supplementary Documentation . . . . . . . . . . . . . . . . . 15 Registered trademarks. . . . . . . . . . . . . . . . . . . . . . . . 15
  • 3. Dosimag Endress+Hauser 3 Function and system design Measuring principle In accordance with Faraday’s law of magnetic induction a voltage is induced in a conductor which is moved through a magnetic field. In the electromagnetic measuring principle, the flowing medium corresponds to the moving conductor. The induced voltage is proportional to the flow velocity and is directed to the amplifier by means of two measuring electrodes. The flow volume is calculated from the pipe cross-sectional area. The DC magnetic field is generated by a switched direct current of alternating polarity. a0003191 Ue = B · L · v Q = A · v Ue Induced voltage B Magnetic induction (magnetic field) L Electrode spacing v Flow velocity Q Volume flow A Pipe cross-section I Intensity of current Measuring system The measuring system is a compact unit consisting of a sensor and transmitter. Input Measured variable Flow rate (proportional to induced voltage) Measuring range Typically v = 0.01 to 10 m/s with the specified measuring accuracy Operable flow range Over 1000:1 Output Output signal Pulse output: Passive, open emitter, max. 30 V DC / 25 mA, galvanically isolated, pulse value and pulse polarity can be selected, adjustable pulse width (0.04 to 4 ms). Signal on alarm Pulse output → failsafe mode can be selected Status output → transistor non-conductive in the event of a fault or if the power supply fails Low flow cutoff Switch-on point for low flow cutoff selectable. Galvanic isolation The circuits of the pulse/status output are galvanically isolated on the device side from the communication and power supply. Ue I I B L V
  • 4. Dosimag 4 Endress+Hauser Switching output Status output: Passive, open emitter, max. 30 V DC / 25 mA, galvanically isolated Power supply Electrical connections The device electrical connection is by means of a Lumberg connector (type RSE8, M12x1). a0003833 Wiring diagram (direct connection without an adapter) A Socket on device B Cable connector 1 (+), power supply (24 V DC nominal voltage (20 to 30 V DC), 6 W) 4 (–), power supply (24 V DC nominal voltage (20 to 30 V DC), 6 W) 5 (+), pulse output, status output (max. 30 V) 6 (–), pulse output (max. 25 mA) 7 (–), status output (max. 25 mA) 2 Service interface (may not be connected during normal operation) 3 Service interface (may not be connected during normal operation) 8 Service interface (may not be connected during normal operation) Supply voltage 24 V DC nominal voltage (20 to 30 V DC), PELV or SELV When installing the Dosimag based on the CAN/CSA-C22.2 No. 1010.1-92 safety standard for Canada, the power must be supplied via a SELV power supply with maximum 30 V DC. Cable connection Lumberg plug (RSE 8, M12x1) for power supply and signal outputs. Cable specifications Use connecting cables with a cross-section of at least 0.25 mm2 (e.g. AWG23). The temperature specification of the cable must be at least 20 °C higher than the ambient temperature in the application. Power consumption DC: <6 W (incl. sensor) Switch-on current: max. 1.9 A (< 5 ms) at 24 V DC Power supply failure At least up to 20 milliseconds. All sensor and measuring point data remain in the M-DAT. Potential equalization No potential matching is needed for grounded steel pipings. Performance characteristics Reference operating condi- tions To DIN EN 29104 and VDI/VDE 2641: • Fluid temperature: +28 °C ± 2 K • Ambient temperature: +22 °C ± 2 K A 1 2 3 4 5 6 7 8 1 2 3 4 5 6 7 8 B RSE8 A B
  • 5. Dosimag Endress+Hauser 5 • Warm-up period: 30 minutes Installation: • Inlet run >10 x DN • Outlet run >5 x DN • Sensor and transmitter grounded. • Sensor centered relative to the pipe. Maximum measured error Volume flow: • ±0.25% o.r. (1 to 4 m/s) or • ±0.5% o.r. ± 1 mm/s or • ± 5% o.r. o.r. = of reading Repeatability Operating conditions: Installation Installation instructions Mounting location Correct measuring is possible only if the pipe is completely full. For this reason, we recommend you run trial fillings before starting productive batching. a0003761 Rotary filling system (example) 1 = Dosimag 2 = Tank 3 = Filler valve 4 = Container Batch time “ta” [s] Relative standard deviation in relation to the batched volume [%] 1.5 s < ta < 3 s 0.4 3 s < ta < 5 s 0.2 5 s < ta 0.1 2 1 3 4
  • 6. Dosimag 6 Endress+Hauser a0003762 Line filling system (example) 1 = Dosimag 2 = Tank 3 = Filler valve 4 = Container Installation near valves Do not install the sensor downstream of a filler valve. If you do, the measuring tube of the sensor would be completely emptied after every filling cycle which would greatly distort the measured value during the next flow measurement. a0003768 Installation near valves (✓ = recommended, x = not recommended) 1 = Dosimag 2 = Filler valve 3 = Container Orientation An optimum orientation position helps avoid gas and air accumulations, and deposits in the measuring tube. Vertical orientation Optimum measurement takes place when the pipe system is completely filled with the fluid. a0003795 Dosimag orientations 1 = Dosimag 2 = Filler valve 3 = Container 1 3 4 2 ✘✔ 1 2 2 1 3 3 1 2 1 2 1 2 3 3 3
  • 7. Dosimag Endress+Hauser 7 Horizontal orientation The measuring electrode plane should be horizontal. This prevents brief insulation of the two electrodes by entrained air bubbles. a0003829 Horizontal installation with Dosimag 1 Measuring electrodes for signal acquisition 2 Measuring tube liner Caution! In the event of extreme heating (e.g. for CIP or SIP cleaning processes), we recommend you install the mea- suring device in such a way that the transmitter part is pointing downwards. This reduces the risk of the elec- tronic components overheating. a0003830 Recommended orientation in the event of extreme heating (✓ = recommended, x = not recommended) Vibrations Secure the piping and the sensor if vibration is severe. a0003831 Measures to prevent vibration of the measuring device A A 1 2 ✘✔
  • 8. Dosimag 8 Endress+Hauser Inlet and outlet runs If possible, install the sensor before fittings such as valves, T-pieces, elbows, etc. Compliance with the following requirements for the inlet and outlet runs is necessary in order to ensure measuring accuracy. • Inlet run ≥ 5 x DN • Outlet run ≥ 2 x DN a0003861 Inlet and outlet runs Adapters Suitable adapters can be used to install the sensor in larger-diameter pipes (see also DIN EN 545). The resultant increase in the rate of flow improves measuring accuracy with very slow-moving fluids. The nomogram shown here can be used to calculate the pressure loss caused by cross-section reduction. Note! The nomogram applies to fluids of viscosity similar to water. 1. Calculate the ratio of the diameters d/D. 2. From the nomogram, read off the pressure loss as a function of fluid velocity (downstream from the reduction) and the d/D ratio. a0003832 Pressure loss due to adapters ³5xDN³2xDN Dd max. 8° 100 10 0.5 d / D [mbar] 0.6 0.7 0.8 0.9 1 m/s 2 m/s 3 m/s 4 m/s 5 m/s 6 m/s 7 m/s 8 m/s 1
  • 9. Dosimag Endress+Hauser 9 Operating conditions: Environment Ambient temperature range –20 to +60 °C (sensor, transmitter) Do not install the measuring device directly over equipment generating heat. Storage temperature The storage temperature corresponds to the permitted ambient temperature range of the transmitter and sen- sor. Degree of protection Standard: IP 67 (NEMA 4X) for transmitter and sensor Shock and vibration resistance Acceleration up to 2 g by analogy with IEC 60068-2-6 Operating conditions: Process Medium temperature range Sensor: • Process: –20 to +130 °C • Cleaning: +150 °C/60 min for CIP and SIP processes Seals: • EPDM: –20 to +130 °C (max. +150 °C for cleaning) • Silicone: –20 to +150 °C • Viton: 0 to +150 °C a0004805 TU = ambient temperature, TM = medium temperature Light gray = standard medium temperature range, dark gray = medium temperature range for cleaning Conductivity Minimum conductivity: • 5 µS/cm for liquids generally • 20 µS/cm for demineralized water Limiting medium pressure range (nominal pressure) 16 bar Pressure tightness (liner) <1 mbar at a temperature range of +25 to +150 °C 0 -20 20 40 60 0-20 40 80 120 160 T [°C]M T [°C]U 130 150
  • 10. Dosimag 10 Endress+Hauser Limiting flow The diameter of the pipe and the flow rate determine the nominal diameter of the sensor. The optimum velocity of flow is 1 to 4 m/s. The velocity of flow (v), moreover, has to be matched to the physical properties of the fluid: • v < 2 m/s: for abrasive fluids such as cleaning agents, etc. • v > 2 m/s: for fluids producing buildup such as liquids containing oil and sugar Note! Flow velocity can be increased, if necessary, by reducing the nominal diameter of the sensor. Pressure loss • No pressure loss at DN 8 and DN 15 if the sensor is installed in a pipe of the same nominal diameter. • Pressure losses for configurations incorporating adapters according to DIN EN 545 → Page 8. Flow rate characteristic values - Dosimag (SI units) Nominal diameter Recommended flow rate Factory settings [mm] [inch] Max. full scale value Pulse value Low flow cutoff (v ~ 0.04 m/s) 4 5/32" 0.14 l/sec 5 µl 2 l/h 8 5/16" 0.5 l/sec 20 µl 8 l/h 15* 1/2"* 1.2 l/sec 100 µl 26 l/h 15 1/2" 1.66 l/sec 100 µl 26 l/h * = Conical version Flow rate characteristic values - Dosimag (US units) Nominal diameter Recommended flow rate Factory settings [inch] [mm] Max. full scale value Pulse value Low flow cutoff (v ~ 0.13 ft/s) 5/32" 4 0.035 gal/sec 0.0002 oz fl 0.009 gal/min 5/16" 8 0.13 gal/sec 0.001 oz fl 0.035 gal/min 1/2"* 15* 0.32 gal/sec 0.004 oz fl 0.12 gal/min 1/2" 15 0.44 gal/sec 0.004 oz fl 0.12 gal/min * = Conical version
  • 11. Dosimag Endress+Hauser 11 Mechanical construction Design / dimensions Dosimag dimensions a0003864 Measuring tube dimensions a0004874 DN L1 L2 Di di DIN [mm] ANSI [inch] DIN [mm] [mm] [mm] [mm] 4 5/32" 44 90 9 4.5 8 5/16" – 90 9 9 15* 1/2"* 20 90 16 12 15 1/2" – 90 16 16 Fitting length depends on process connections * = Conical version 133 100 80 86 90 210 42 33.4 M6x4 12 7050 10 L1 L2 di Di
  • 12. Dosimag 12 Endress+Hauser Process connection dimensions (with aseptic molded seal) Weld nipples 1.4404 / 316L **H**-U*********** Sensor Suits pipe di G L H x B DN [mm] DIN 11850 [mm] [mm] [mm] [mm] a0003870 4 / 8 14 x 2 9 14 23.3 60 x 42 15 / 15* 20 x 2 16 20 23.3 60 x 42 – Fitting length = (2 x L) + 86 mm – If pigs are used for cleaning, it is essential to take the inside diameters of the measuring tube (→ Page 11) and process connection (di) into account. – * = Conical version G di L HxB Weld nipple ODT/SMS 1.4404 / 316L **H**-V*********** Sensor Suits pipe di G L H x B DN [mm] OD/SMS [mm] [mm] [mm] [mm] a0003871 4 / 8 12.7 x 1.65 9 12.7 16.1 60 x 42 15 / 15* 19.1 x 1.65 16 19.1 16.1 60 x 42 – Fitting length = (2 x L) + 86 mm – If pigs are used for cleaning, it is essential to take the inside diameters of the measuring tube (→ Page 11) and process connection (di) into account. – * = Conical version G di L HxB Tri-Clamp L14 AM7 1.4404 / 316L **H**-1*********** Sensor Suits pipe di G L H x B DN [mm] OD [mm] [mm] [mm] [mm] a0003872 4 / 8 Tube 12.7 x 1.65 (ODT 1/2") 9.4 25.0 28.5 60 x 42 15 / 15* Tube 19.1 x 1.65 (ODT 3/4") 15.8 25.0 28.5 60 x 42 – Fitting length = (2 x L) + 86 mm – If pigs are used for cleaning, it is essential to take the inside diameters of the measuring tube (→ Page 11) and process connection (di) into account. – * = Conical version G di L HxB Tri-Clamp L14 AM7 1.4404 / 316L **H**-2*********** Sensor Suits pipe d1 d2 G L H x B DN [mm] OD [mm] [mm] [mm] [mm] [mm] a0003878 4 / 8 Tube 19.1 x 1.65 (ODT 3/4") 9 15.8 25.0 28.5 60 x 42 – Fitting length = (2 x L) + 86 mm – If pigs are used for cleaning, it is essential to take the inside diameters of the measuring tube and process connection (d1 or d2) into account! G d2 L HxB d1
  • 13. Dosimag Endress+Hauser 13 Weight Approx. 2.8 kg Material Transmitter housing: 1.4308/304 Sensor housing: Acid and alkali-resistant outer surface; stainless steel 1.4308/304 Measuring tube: Stainless steel 1.4301/304 with PFA lining Process connection: • Weld nipples → stainless steel 1.4404/316L • Aseptic weld nipples → stainless steel 1.4404/316L • Tri-Clamp → stainless steel 1.4404/316L Measuring electrodes: Standard: 1.4435 (AISI 316L); option: Alloy C-22 Seals: Molded seal (EPDM, silicone, Viton) Material load diagram Weld nipple (DIN 11850, ODT/SMS), Clamp L14 AM7 Material: 1.4404 / 316L (with molded seal) A0005121 Process connection Dairy fittings: weld nipple (DIN 11850, ODT/SMS), Tri-Clamp L14 AM7 Surface roughness • Measuring tube lining with PFA: ≤0.4 µm • Electrodes (1.4435 (AISI 316L), Alloy C-22): 0.3 to 0.5 µm • Process connection Dosimag: ≤0.8 µm (all information refers to wetted parts) User interface Display elements Dosimag does not have a display or display elements. Remote operation Operation takes place via the “ToF-Tool - FieldTool Package” configuration and service program from Endress+Hauser. This can be used to configure functions and read off measured values. Certificates and approvals CE mark The measuring system is in conformity with the statutory requirements of the EC Directives. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark. 0 10 5 20 15 PN16 [bar] -60 -40 -20 0 20 40 60 80 100 120 140 160 180 [°C]
  • 14. Dosimag 14 Endress+Hauser C-tick mark The measuring system meets the EMC requirements of the Australian Communication and Media Authority (ACMA). Sanitary compatibility 3-A EHEDG Seals in conformity with FDA Other approvals Additional approvals from the certification bodies FM (USA) and CSA (Canada) are available for all Dosimag devices for use in non-explosion protected atmospheres. Contact your Endress+Hauser sales office for more information. Pressure measuring device approval All Dosimag devices correspond to Article 3(3) of the EC Directive 97/23/EC (Pressure Equipment Directive) and have been designed and manufactured according to good engineering practice. Other standards and guide- lines EN 60529: Degrees of protection by housing (IP code) EN 61010-1: Protection Measures for Electrical Equipment for Measurement, Control, Regulation and Laboratory Proce- dures. EN 61326 (IEC 1326): Electromagnetic compatibility (EMC requirements) CSA-C22.2 No. 142-M1987 Process Control Equipment CAN/CSA-C22.2 No. 1010.1-92 Safety Requirements for Electrical Equipment for Measuring, Control and Laboratory Use. Pollution degree 2, Installation Category I ANSI/ISA-S82.01 Safety Standard for Electrical and Electronic Test, Measuring, Controlling and related Equipment - General Requirements. Pollution degree 2, Installation Category I Ordering information The Endress+Hauser service organization can provide detailed ordering information and information on spe- cific order codes on request. Accessories Various accessories and spare parts, which can be ordered separately from Endress+Hauser, are available for the transmitter and sensor. The Endress+Hauser service organization can provide detailed information on the specific order codes on request. Accessories Description Ordering code Dosimag transmitter Transmitter for replacement or for storage. The following specifications can be indicated by means of the order code: – Approvals – Housing – Cables – Cable connection – Power supply – Software functions – Outputs/inputs 5BXXX-XXXXX******* Housing seal To seal the transmitter 50102857
  • 15. Dosimag Endress+Hauser 15 Supplementary Documentation ❑ Flow Measurement (FA005/06/en) ❑ Operating Instructions Dosimag (BA098/06/en) Registered trademarks TRI-CLAMP ® Registered trademark of Ladish & Co., Inc., Kenosha, USA HistoROM™, M-DAT® , ToF Tool - Fieldtool® Package, Applicator® Registered or registration-pending trademarks of Endress+Hauser Flowtec AG, Reinach, CH Seal set For regular replacement of the seals on the process con- nections. DK5HS – *** Mounting kit Mounting kit comprising: – 2 process connections – Threaded fasteners – Seals DKH** – **** Adapter connection Adapter connections for installing Dosimag on other pro- cess connections DK5HA – ****** Conversion kit Kit for converting Dosimag A to Dosimag DK5UP – H RSE8 adapter Lumberg RSE8 connection jack, 8-pole adapter (RSE8), 24 V DC, pulse, status 50107169 RSE5 adapter Lumberg RSE8 connection jack, 5-pole adapter (RSE5), 24 V DC, pulse, status 50107168 RSE4 adapter Lumberg RSE8 connection jack, 4-pole adapter (RSE4), 24 V DC, pulse 50107167 RSE8 supply cable Lumberg cable RKWTN8-56/5 P92 50107895 ToF Tool - FieldTool Package Configuration and service software for flowmeters in the field: – Commissioning, maintenance analysis – Configuring measuring devices – Service functions – Visualization of process data – Troubleshooting Contact your Endress+Hauser representative for more information. DXS10 - * * * * * FXA 193 Service interface connecting cable from the device to the PC for using the “ToF Tool FieldTool Package” operating software FXA193 – * Service adapter Adapter for connecting FXA 193 to the device. 50106443 Applicator Software for selecting and configuring flowmeters. Applicator can be downloaded from the Internet or ordered on CD-ROM for installation on a local PC. Contact your Endress+Hauser representative for more information. DKA80 - * Accessories Description Ordering code
  • 16. Dosimag International Headquarters Endress+Hauser GmbH+Co. KG Instruments International Colmarer Str. 6 79576 Weil am Rhein Deutschland Tel. +49 76 21 9 75 02 Fax +49 76 21 9 75 34 5 www.endress.com info@ii.endress.com TI066D/06/en/11.05 71008913 FM+SGML 6.0 ProMoDo