This ppt includes different types of strain gauges which are used for pressure, temperature, force, acceleration etc measurement.
All types of strain gauges are included. Also temperature compensation is also explained.
1. UNIT NO 3
Strain Gauges and Pressure
Measurement Devices
By:- Prof. P.B. Borakhede
2. CONTENTS
Introduction
Resistance Strain Gauge
a) Bonded Gauges
b) Unbonded Gauges
Strain Gauge Circuits
a) Ballast Circuit
b) Wheatstone Bridge Circuit
Temperature compensation
3. STRAIN GAUGES
Introduction
A strain gauge is a strain transducer ie device for
measuring dimensional change or the surface of a
structural member under test.
Measurement of strain is indispensable in a variety of
applications due to:
i) Utility of strain measurement as a means of
determining maximum stress values or to measure
force, pressure, acceleration, torque etc.
ii) Desirability to avoid the use of large factors of safety
in design of aircraft and automatic control equipment
due to mass/inertia considerations.
Prof. P.B. Borakhede, MGI-COET, Shegaon
5. Requirements of a strain gauge
Following points need consideration while designing any
strain gauge so that it gives an accurate measure of
strain:
• Extremely small size and negligible mass
• Simple and easy attachment to the specimen under test
• High sensitivity in the direction of measured strain but
low sensitivity in the transverse plane
• High speed of response: negligible time lag
• Capability to indicate static, transient and dynamic strain
• Insensitiveness to ambient conditions such as
temperature, humidity, vibration etc.
Prof. P.B. Borakhede, MGI-COET, Shegaon
6. Resistance strain gauge
Gauge Factor
When a bar is subjected to a simple tensile loading,
there occurs an increase in length of the bar in the
direction of load.
Strain refers to relative change in dimensions of bar
under load and it is the ratio of change in length to
original length of bar.
Strain € = change in length/ original length =(∆l/l)
Strain is usually dimensionless.
Consider a bar as shown in figure.
Prof. P.B. Borakhede, MGI-COET, Shegaon
7. An increase in length of the bar in the direction of applied
load is also accompanied by a decrease in lateral
dimension perpendicular to the load.
The ratio of strain in the lateral to lateral to in longitudinal is
called as Poisson’s Ratio.
Poissons Ratio (ũ) = (-∆D/D)/ ∆l/l
Consider appropriate length of a suitable material strained
within the elastic limits.
The change in physical dimension in conductor will cause a
change in its resistance.
The resistance of a conductor of length L, uniform cross-
section A and uniform resistivity ς is given by
R= (ςL/A)
Prof. P.B. Borakhede, MGI-COET, Shegaon
9. Where F represents fractional change in resistance
divided by a unit strain, and is called gauge factor.
The piezoelectric term is very small so it is neglected.
F= 1+ 2 ũ
Prof. P.B. Borakhede, MGI-COET, Shegaon
11. Unbonded Strain Gauges
These gauges are not directly
bonded bonded onto the surface of specimen
which is being examined.
In this the fine wire filaments ie res
resistance wires are stretched around around
rigid and electrically insulated insulated pins on
two frames A & B.
which can move relative to each other.
The frames are held close with a spring loaded mechanism.
When the force is applied on speciman the wires are
stretched.
When the frame A moves relative to frame B, the wire
filaments are strained.
Prof. P.B. Borakhede, MGI-COET, Shegaon
12. The strain can be detected through measurement of
change in resistance by electrical circuit.
The range of unbonded strain gauge is 0.15% strain
with accuracy better than 0.1%.
Since the gauges are not cemented but are simply
screwed at the desired location, they can be detached
and used again.
These gauges are used mainly for measurement of
force, pressure, acceleration rather than for
measurement of displacement because of its massive
structure.
Prof. P.B. Borakhede, MGI-COET, Shegaon
13. Bonded Strain Gauges
These gauges are bonded or cemented directly onto the
surface of the structural member which is being
examined.
Examples of bonded strain gauges are
i) Fine wire gauges cemented to a paper backing
ii) Metal foil gauges
iii) Semiconductor gauges
i) Wire type strain gauges
A very fine wire of 0.025 mm in diameter is arranged in
the form of a grid shape consisting of a series of long
parallel loops onto the mechanical part on which
measurements are to be made.
Prof. P.B. Borakhede, MGI-COET, Shegaon
14. Leads and connecting terminals are also provided for
electrically connecting the grid to the measurement
instrument.
To ensure satisfactory performance, the bond between
the resistance element and the cement joining it to the
work piece should be stronger than grid itself.
Further, for ease of handling, shipping, storing and
attaching to the specimen, the wire grid is first cemented
to a thin paper sheet or to a very thin bakelite sheet and
covered with a protective covering of paper, felt or thin
bakelite.
The size of grid varies with application.
Prof. P.B. Borakhede, MGI-COET, Shegaon
15. ii) Metal Foil Gauge
The gauge is produced by printed circuit technique and
consists of a foil grid on plastic backing.
The desired grid pattern is first printed on a thin sheet
of metal-alloy foil with an acid resistant ink and then the
unprinted portion is etched away.
This construction allows the use of varying sections
throughout the grid length.
Larger area can be provided at the ends where lead
constructions are made.
The gauge has been successfully employed to fillets
and sharply curved shapes because of its fine and
accurate construction.
Prof. P.B. Borakhede, MGI-COET, Shegaon
16. Advantages:
• Improved hysteresis
• Easy soldering of welding of leads
• Better fatigue life
• Very good lateral strain sensitivity
• Improved transmission of strain from the test surface to
the sensitive grid
• Stability at high temperature.
iii) Semiconductor or Piezoresistive Gauge:
These gauges are produced in wafers from silicon of
germanium crystals in which exact amount of special
impurities such as boron have been added to impart
certain desirable characteristics.
Prof. P.B. Borakhede, MGI-COET, Shegaon
17. The process is called doping and the crystals are known
as doped crystals.
The semiconductor gauges are classified as
i) negative or n-type whose resistance decrease in
response to tensile strain
ii) Positive or p-type whose resistance increase in
response to tensile strain.
The gauge is in the form of a single rectangular filament
about 0.05 mm thick by 0.25 mm wide and 1.5 and 12
mm in length.
Prof. P.B. Borakhede, MGI-COET, Shegaon
18. A single film semiconductor gauge with leads on either
side is shown in figure.
The semiconductor gauges are usually provided with a
plastic or stainless steel backing and are bonded to the
test surface by the same methods as wire and foil
gauges.
Advantages:
Very high sensitive in comparison to metal gauges
High gauge factor in range of 100 to 200.
Low hysteresis
Brittle and not suitable for large strain measurement
Prof. P.B. Borakhede, MGI-COET, Shegaon
19. Strain gauge Circuits
In strain gauges, the change in resistance is very small.
Such small changes of resistance occurring in strain
measuring systems are converted into voltage and
generally measured by employing potentiometric circuit.
The ballast circuit and Wheatstone bridge circuit .
1. Ballast circuit
Prof. P.B. Borakhede, MGI-COET, Shegaon
20. R= resistance of the unstrained resistance gauge
Rb= resistance of system outside transducer ( ballast
resistance)
Vs= supply voltage or input voltage.
Prof. P.B. Borakhede, MGI-COET, Shegaon
23. 2. Wheatstone bridge Circuit
There are two ways of using wheatstone bridge circuit.
a) Balanced (null) condition
b) Unbalanced (deflection) condition
The null technique is more accurate by means of
resistance change but can only be used to measure
static strains.
a) Balanced condition
Prof. P.B. Borakhede, MGI-COET, Shegaon
24. With no straining the resistances are so arranged that
potential at B equals to potential at D and galvanometer
gives zero deflection ie no current indication on
galvanometer.
Prof. P.B. Borakhede, MGI-COET, Shegaon
25. 2. Deflection Mode
In deflection mode initially bridge resistances are so
adjusted that the bridge is balanced. After gauges are
strained, the equilibrium gets disturbed.
However, the bridge is allowed to stay unbalanced and
the galvanometer output Vo is observed.
Prof. P.B. Borakhede, MGI-COET, Shegaon
26. 1. Quarter Bridge
Only one strain gauge is used and the other three elements of
the bridge are fixed resistors.
Prof. P.B. Borakhede, MGI-COET, Shegaon
28. ii) Half Bridge Circuit
Two of the bridge elements are strain gauges and the
other two are fixed resistors.
The gauge R1 is bonded to upper surface of cantilever
beam, and second gauge R2 is bonded to the lower
surface and located underneath the first gauge.
These gauges are connected electrically to form
adjacent limbs of the Wheatstone bridge circuit.
When no strain is applied,
Vab = Vad= Vs/2
The terminals B and D are at same potential.
The bridge is then balanced and the output voltage
Vo= 0.
When load is applied to beam, resistance of gauge R1
increases due tensile load while R2 decreases due to
equal compressive strain.
Prof. P.B. Borakhede, MGI-COET, Shegaon
31. iii) Full Bridge Circuit
All the four elements of bridge are strain gauges.
When no strain applied,
Vab = Vad = Vs/2
Terminals B and D are at same potential. Bridge is
balanced and output voltage V0 = 0.
Prof. P.B. Borakhede, MGI-COET, Shegaon
33. dVo = dR/R = Vs F€
Temperature compensation of Strain Gauges
Effect of ambient temperature variations on the strain
gauge output has to be removed or minimized.
Because gauge resistance changes both with strain
and temperature.
The temperature then considered as interfering input
and brings about a change in the gauge resistance in
two ways:
i) Resistance change in the wire filament or grid due to
change in its volume and resistivity,
ii) Resistance change due to differential expansion
Prof. P.B. Borakhede, MGI-COET, Shegaon
34. Existing between the gauge and test surface to which
gauge has been bonded.
Methods to compensate effect of temperature are:
1. Use of Dummy gauge:
Prof. P.B. Borakhede, MGI-COET, Shegaon
35. Measuring gauge is active gauge, is mounted on the
test-piece and constitutes resistance R1 of bridge circuit.
The gauge for temperature compensation called dummy
gauge, is connected to adjacent limb of bridge as
resistance R2.
Dummy gauge is identical to active strain gauge so that
they form a match pair.
Dummy gauge is bonded to separate, unstrained
component identical to that of loaded member.
The dummy gauge remains unstrained throughout
process and suffers change in resistance due to
temperature only.
Two gauges are placed close to each other so that they
are influnced equally by ambient temperature.
Prof. P.B. Borakhede, MGI-COET, Shegaon
36. R1 and R2 are initially equal and with temperature
increase both the gauges will be subjected to equal
temperature induced strain.
Strain in active gauge= €d + €t
Strain in dummy gauge = €t
Total strain = €d+ €t- €t = €d
2. Temperature compensation by using compensating
gauge
Prof. P.B. Borakhede, MGI-COET, Shegaon
37. In this the active gauge R1 is bonded to the top surface
and is in tension while compensating gauge R2 is
bonded to the bottom surface is in compression.
Any change in temperature effects both the equally and
so no errors are caused by resistance change due to
temperature variation.
Further, since the resistance changes due to applied
load are equal but of opposite sign in two gauges, the
arrangement has the added advantage in bridge output
being doubled.
3. Poission’s Method
In the poission’s method, gauge R2 is bonded at right
angles to the active gauge R1 on the test specimen.
Prof. P.B. Borakhede, MGI-COET, Shegaon
38. Gauge R1 is in tension but the gauge R2 is reduced in
length depending upon Poisson’s ratio of material.
In addition to temperature compensation, the
arrangement increases the bridge output in proportion to
mount the gauge in corresponding position on the
underside of the test member.
Prof. P.B. Borakhede, MGI-COET, Shegaon
39. Above considerations show that usually it is
advantageous to use more than one gauge in bridge
circuit for the following reasons:
i) To eliminate unwanted effects and to provide for
temperature compensation
ii) To increase the bridge output and thereby increase the
bridge sensitivity.
Prof. P.B. Borakhede, MGI-COET, Shegaon