Journal of Diabetes Science and Technology                                                                ORIGINAL ARTICLE
Volume 6, Issue 5, September 2012
© Diabetes Technology Society




  System Accuracy Evaluation of 43 Blood Glucose Monitoring Systems
  for Self-Monitoring of Blood Glucose according to DIN EN ISO 15197

  Guido Freckmann, M.D., Christina Schmid, Ph.D., Annette Baumstark, Ph.D., Stefan Pleus, M.S.,
                        Manuela Link, M.E., and Cornelia Haug, M.D.



   Abstract
   Background:
   The accuracy of systems for self-monitoring of blood glucose is important, as reliable measurement results are
   a prerequisite for therapeutic decisions.

   Methods:
   This system accuracy evaluation study was performed according to DIN EN ISO 15197:2003 for 43 Conformité
   Européenne (CE)-labeled blood glucose (BG) monitoring systems. Measurement results of each system were
   compared with results of the designated comparison method (manufacturer’s measurement procedure): glucose
   oxidase method (YSI 2300 glucose analyzer) or hexokinase method (Hitachi 917/ cobas 501).

   Results:
   Complete assessment according to the International Organization for Standardization (ISO) standard was
   performed for 34 out of 43 systems, and 27 (79.4%) meet the requirements of the standard, i.e., ≥95% of their results
   showed at least the minimum acceptable accuracy. For 9 of the 43 systems, complete accuracy assessment was
   not performed due to an oxygen sensitivity (manufacturer’s labeling). The bias (according to Bland and Altman)
   of all 43 evaluated systems ranged from -14.1% to +12.4%.

   Conclusions:
   From the 34 systems completely assessed, 7 systems did not fulfill the minimal accuracy requirements of
   the ISO standard. The CE mark apparently does not guarantee that all BG systems provide accuracy according to
   the standard. Because inaccurate systems bear the risk of false therapeutic decisions, regular and standardized
   evaluation of BG meters and test strips should be requested in order to ensure adherence to quality standards.

   J Diabetes Sci Technol 2012;6(5):1060-1075




Author Affiliation: Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm, Ulm, Germany

Abbreviations: (BG) blood glucose, (CE) Conformité Européenne, (GOx) glucose oxidase, (HK) hexokinase, (ISO) International Organization for
Standardization, (NIST) National Institute of Standards and Technology, (SMBG) self-monitoring of blood glucose

Keywords: blood glucose monitoring systems, Conformité Européenne mark, DIN EN ISO 15197:2003, self-monitoring of blood glucose,
system accuracy

Corresponding Author: Guido Freckmann, M.D., Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH,
Helmholtzstrasse 20, 89081 Ulm, Germany; email address guido.freckmann@uni-ulm.de

                                                                   1060

More Related Content

Similar to System Accuracy Evaluation Of 43 Blood Glucose Monitoring For Self Monitoring Of Blood Glucose According To Din En Iso 15197

Reality Diabetes Care
Reality Diabetes CareReality Diabetes Care
Reality Diabetes Carerxmike
 
Efficacy Studies for Bioresonance
Efficacy Studies for Bioresonance Efficacy Studies for Bioresonance
Efficacy Studies for Bioresonance David Franklin
 
Internal 2013 - General practice clinical systems
Internal 2013 - General practice clinical systemsInternal 2013 - General practice clinical systems
Internal 2013 - General practice clinical systemsEvangelos Kontopantelis
 
An Analysis on the UV-Visible Spectrophotometry Method
An Analysis on the UV-Visible Spectrophotometry MethodAn Analysis on the UV-Visible Spectrophotometry Method
An Analysis on the UV-Visible Spectrophotometry MethodAI Publications
 
In tech quality-control_in_clinical_laboratories
In tech quality-control_in_clinical_laboratoriesIn tech quality-control_in_clinical_laboratories
In tech quality-control_in_clinical_laboratoriesMillat Sultan
 
Quality control clia
Quality control cliaQuality control clia
Quality control cliaJuan Méndez
 
Quality management systems qms tqm
Quality management systems qms tqmQuality management systems qms tqm
Quality management systems qms tqmDr Ajay Kumar Tiwari
 
Are we using the correct quality goals?
Are we using the correct quality goals?Are we using the correct quality goals?
Are we using the correct quality goals?Ola Elgaddar
 
CONTROL OF BLOOD GLUCOSE IN DIABETIC PATIENTS
CONTROL OF BLOOD GLUCOSE IN DIABETIC PATIENTSCONTROL OF BLOOD GLUCOSE IN DIABETIC PATIENTS
CONTROL OF BLOOD GLUCOSE IN DIABETIC PATIENTSVictoria Onifade
 
Master of pharmacy... Pharmaceutics ICH Giudelines
Master of pharmacy... Pharmaceutics ICH GiudelinesMaster of pharmacy... Pharmaceutics ICH Giudelines
Master of pharmacy... Pharmaceutics ICH Giudelinesomkarmandlik678
 
A PROPOSED APPROACH TO ANALYTICAL METHOD LONG-TERM STABILITY ASSESSMENT
A PROPOSED APPROACH TO ANALYTICAL METHOD LONG-TERM STABILITY ASSESSMENTA PROPOSED APPROACH TO ANALYTICAL METHOD LONG-TERM STABILITY ASSESSMENT
A PROPOSED APPROACH TO ANALYTICAL METHOD LONG-TERM STABILITY ASSESSMENTFernando Berlitz
 
How to Submit Non-Clinical Data to CBER Using SEND : Understanding New FDA Re...
How to Submit Non-Clinical Data to CBER Using SEND : Understanding New FDA Re...How to Submit Non-Clinical Data to CBER Using SEND : Understanding New FDA Re...
How to Submit Non-Clinical Data to CBER Using SEND : Understanding New FDA Re...MMS Holdings
 
Quality management systems - INDUSTRIAL PHARMACY ll
Quality management systems - INDUSTRIAL PHARMACY llQuality management systems - INDUSTRIAL PHARMACY ll
Quality management systems - INDUSTRIAL PHARMACY llJafarali Masi
 
ISO 15189 Accreditation Guide - Improving Laboratory Performance Through Qual...
ISO 15189 Accreditation Guide - Improving Laboratory Performance Through Qual...ISO 15189 Accreditation Guide - Improving Laboratory Performance Through Qual...
ISO 15189 Accreditation Guide - Improving Laboratory Performance Through Qual...Randox
 
Ruchika zalpouri, ae 163-d
Ruchika zalpouri, ae 163-dRuchika zalpouri, ae 163-d
Ruchika zalpouri, ae 163-dRuchikaZalpouri
 

Similar to System Accuracy Evaluation Of 43 Blood Glucose Monitoring For Self Monitoring Of Blood Glucose According To Din En Iso 15197 (20)

U.A. NADKARNI MSc THESIS - Copy
U.A. NADKARNI MSc THESIS - CopyU.A. NADKARNI MSc THESIS - Copy
U.A. NADKARNI MSc THESIS - Copy
 
Reality Diabetes Care
Reality Diabetes CareReality Diabetes Care
Reality Diabetes Care
 
Document 30
Document 30Document 30
Document 30
 
Efficacy Studies for Bioresonance
Efficacy Studies for Bioresonance Efficacy Studies for Bioresonance
Efficacy Studies for Bioresonance
 
Internal 2013 - General practice clinical systems
Internal 2013 - General practice clinical systemsInternal 2013 - General practice clinical systems
Internal 2013 - General practice clinical systems
 
อ้างอิง 1 standardisation mosca cclm_2007
อ้างอิง 1 standardisation mosca cclm_2007อ้างอิง 1 standardisation mosca cclm_2007
อ้างอิง 1 standardisation mosca cclm_2007
 
An Analysis on the UV-Visible Spectrophotometry Method
An Analysis on the UV-Visible Spectrophotometry MethodAn Analysis on the UV-Visible Spectrophotometry Method
An Analysis on the UV-Visible Spectrophotometry Method
 
Article on glucose metre
Article on glucose metreArticle on glucose metre
Article on glucose metre
 
In tech quality-control_in_clinical_laboratories
In tech quality-control_in_clinical_laboratoriesIn tech quality-control_in_clinical_laboratories
In tech quality-control_in_clinical_laboratories
 
Quality control clia
Quality control cliaQuality control clia
Quality control clia
 
Quality management systems qms tqm
Quality management systems qms tqmQuality management systems qms tqm
Quality management systems qms tqm
 
Are we using the correct quality goals?
Are we using the correct quality goals?Are we using the correct quality goals?
Are we using the correct quality goals?
 
CONTROL OF BLOOD GLUCOSE IN DIABETIC PATIENTS
CONTROL OF BLOOD GLUCOSE IN DIABETIC PATIENTSCONTROL OF BLOOD GLUCOSE IN DIABETIC PATIENTS
CONTROL OF BLOOD GLUCOSE IN DIABETIC PATIENTS
 
18. Dr. Andrey Vasiliev - Ministry of Health (Russian Federation)
18. Dr. Andrey Vasiliev - Ministry of Health (Russian Federation)18. Dr. Andrey Vasiliev - Ministry of Health (Russian Federation)
18. Dr. Andrey Vasiliev - Ministry of Health (Russian Federation)
 
Master of pharmacy... Pharmaceutics ICH Giudelines
Master of pharmacy... Pharmaceutics ICH GiudelinesMaster of pharmacy... Pharmaceutics ICH Giudelines
Master of pharmacy... Pharmaceutics ICH Giudelines
 
A PROPOSED APPROACH TO ANALYTICAL METHOD LONG-TERM STABILITY ASSESSMENT
A PROPOSED APPROACH TO ANALYTICAL METHOD LONG-TERM STABILITY ASSESSMENTA PROPOSED APPROACH TO ANALYTICAL METHOD LONG-TERM STABILITY ASSESSMENT
A PROPOSED APPROACH TO ANALYTICAL METHOD LONG-TERM STABILITY ASSESSMENT
 
How to Submit Non-Clinical Data to CBER Using SEND : Understanding New FDA Re...
How to Submit Non-Clinical Data to CBER Using SEND : Understanding New FDA Re...How to Submit Non-Clinical Data to CBER Using SEND : Understanding New FDA Re...
How to Submit Non-Clinical Data to CBER Using SEND : Understanding New FDA Re...
 
Quality management systems - INDUSTRIAL PHARMACY ll
Quality management systems - INDUSTRIAL PHARMACY llQuality management systems - INDUSTRIAL PHARMACY ll
Quality management systems - INDUSTRIAL PHARMACY ll
 
ISO 15189 Accreditation Guide - Improving Laboratory Performance Through Qual...
ISO 15189 Accreditation Guide - Improving Laboratory Performance Through Qual...ISO 15189 Accreditation Guide - Improving Laboratory Performance Through Qual...
ISO 15189 Accreditation Guide - Improving Laboratory Performance Through Qual...
 
Ruchika zalpouri, ae 163-d
Ruchika zalpouri, ae 163-dRuchika zalpouri, ae 163-d
Ruchika zalpouri, ae 163-d
 

System Accuracy Evaluation Of 43 Blood Glucose Monitoring For Self Monitoring Of Blood Glucose According To Din En Iso 15197

  • 1. Journal of Diabetes Science and Technology ORIGINAL ARTICLE Volume 6, Issue 5, September 2012 © Diabetes Technology Society System Accuracy Evaluation of 43 Blood Glucose Monitoring Systems for Self-Monitoring of Blood Glucose according to DIN EN ISO 15197 Guido Freckmann, M.D., Christina Schmid, Ph.D., Annette Baumstark, Ph.D., Stefan Pleus, M.S., Manuela Link, M.E., and Cornelia Haug, M.D. Abstract Background: The accuracy of systems for self-monitoring of blood glucose is important, as reliable measurement results are a prerequisite for therapeutic decisions. Methods: This system accuracy evaluation study was performed according to DIN EN ISO 15197:2003 for 43 Conformité Européenne (CE)-labeled blood glucose (BG) monitoring systems. Measurement results of each system were compared with results of the designated comparison method (manufacturer’s measurement procedure): glucose oxidase method (YSI 2300 glucose analyzer) or hexokinase method (Hitachi 917/ cobas 501). Results: Complete assessment according to the International Organization for Standardization (ISO) standard was performed for 34 out of 43 systems, and 27 (79.4%) meet the requirements of the standard, i.e., ≥95% of their results showed at least the minimum acceptable accuracy. For 9 of the 43 systems, complete accuracy assessment was not performed due to an oxygen sensitivity (manufacturer’s labeling). The bias (according to Bland and Altman) of all 43 evaluated systems ranged from -14.1% to +12.4%. Conclusions: From the 34 systems completely assessed, 7 systems did not fulfill the minimal accuracy requirements of the ISO standard. The CE mark apparently does not guarantee that all BG systems provide accuracy according to the standard. Because inaccurate systems bear the risk of false therapeutic decisions, regular and standardized evaluation of BG meters and test strips should be requested in order to ensure adherence to quality standards. J Diabetes Sci Technol 2012;6(5):1060-1075 Author Affiliation: Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm, Ulm, Germany Abbreviations: (BG) blood glucose, (CE) Conformité Européenne, (GOx) glucose oxidase, (HK) hexokinase, (ISO) International Organization for Standardization, (NIST) National Institute of Standards and Technology, (SMBG) self-monitoring of blood glucose Keywords: blood glucose monitoring systems, Conformité Européenne mark, DIN EN ISO 15197:2003, self-monitoring of blood glucose, system accuracy Corresponding Author: Guido Freckmann, M.D., Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH, Helmholtzstrasse 20, 89081 Ulm, Germany; email address guido.freckmann@uni-ulm.de 1060