BIOMETRY&IOL POWER
CALCULATION
DR MD ALBAN UDDIN
OUTLINE
• HISTORY
• COMPONENTS OF IOL POWER CALCULATION
• BIOMETRY
• FORMULAE
• VARIABLES
• PROBLEMS AND ERRORS
HISTORY
• 1949-1st IOL Implantation-Sir Harold Ridley
• Implantation was done on Feb 8,1950-secondary IOL
implantation.
• Weight of the lens-17.4mg in water
• Patient had a refractive surprise of nearly 20 D.
• 1967-Fyodorov et al-IOL power using vergence formulae.
• 1970s-”A “scans available-theoretical vergence formulae.
• 1980s-IDEM lens attempted.
• 1982-Gernet and Zorkendorfer-avg refractive power of
lens-+23.70D.
Biometry for Intra-Ocular Lens(IOL) Power Calculation-AAO
GENERATIONS OF INTRAOCULAR LENSES
GENERATION CHARACTERISTIC EXAMPLE
1st-1948 Experimental lens by Ridley Biconvex
2nd-1952 Anterior chamber angle fixated lens Baron’s Anterior chamber lens
Strampelli rigid tripod anterior
chamber IOL
Dannheim flexible closed loop
ACIOL
Barraquer J-Loop anterior chamber
IOL
3rd-1957 Iris supported,Iris clip lens Binkhorst iris clip lens
Fyrodov type 2 “Sputnik”iris –clip
IOL
Worst platina IOL
Copeland iris-plane IOL
4th-1956 Anterior chamber one piece lens Choyce Mark VIII rigid ACIOL
Strampelli angle supported ACIOL
5th Posterior chamber lenses Shearing J-loop posterior chamber
IOL
CRUDE IOL POWER CALCULATIONS IN PAST
• 1980s-Guess work approach.
• Concept-IDEM lenses
Standard lenses
Emmetropia lenses
Biometry for Intra-Ocular Lens(IOL) Power Calculation-AAO
• IDEM LENSES:
• Ideal emmetropic lenses
• Pre-and post refractive refraction is same.
• Recommended for emmetropic patients.
• 20D artificial lens in posterior chamber will restore preoperative refractive error.
Disorders of lens and cataract surgery-A K Khurana-pg 108
• LIMITATION-Will restore the preoperative refractive error,only if the natural lens
had refractive power of 23.70 D which is not always true, as refractive power is a
combination of different factors-corneal curvature,depth of AC ,dioptric power of
lens and length of eye.
Disorders of lens and cataract surgery-A K Khurana-pg 108
• STANDARD LENS:
• 2D stronger than IDEM lens.
• Renders eye pseudophakic about 1.5D myopic as compared to preoperative
emmetropic refraction.
• 80% of eyes-less than 1D nearsighted or farsighted-nearly correct in majority of
patients.
• Balanced distance and near vision.
• Standard PCIOL-+22D
• Standard ACIOL-+20D
• Disorders of lens and cataract surgery-A K Khurana-pg 108
• EMMETROPIA LENS:
• For restoring emmetropia in previously ammetropic eyes.
• Power calculation-
HYPERMETROPIA-Power of IDEM lens+(Preoperative refractive error x 1.25).
MYOPIA-Power of IDEM lens-(Preoperative refractive error x 1.25).
• History to be taken-kind of glasses patient wore previously,rely on refractive
values only if they come from reliable records,or if they match clinical findings.
• Deviation of 2-3D is possible,can lead to significant errors.
• Disorders of lens and cataract surgery-A K Khurana-pg 108
• DRAWBACKS:
• Based on guess work approach.
• 1970s-’A’ scans were available-studies were conducted to establish various
theoretical vergence formulas.
• It was found IOL power is determined by
Axial length
Keratometry
Effective IOL position
Biometry for Intra-Ocular Lens(IOL) Power Calculation-AAO
IOL POWER CALCULATION
BIOMETRY FORMULAS CLINICAL VARIABLES
• AXIAL LENGTH
• CORNEAL POWER
• EFFECTIVE LENS
POSITION
• GENERATIONS
• PERSONALISATION • SPECIAL
CIRCUMSTANCES
• PROBLEMS AND
ERRORS
BIOMETRY
• AXIAL LENGTH(AL):
• Most important factor in biometric calculations.
• Distance from tear film to RPE.
• 1mm error in AL results in refractive error of 2.35D in a 23.5 mm eye,error being
less in case of longer eyes as compared to shorter eyes.
• Methods of measurement of axial length-
Ultrasonic measurement
Optical measurement
Disorders of lens and cataract surgery-A K Khurana-pg 109
• ULTRASONIC MEASUREMENT:
• Dr Karl Ossoinig-developed standardised A scans
• Methods-Applanation method
Immersion method
• Measures axial length from anterior surface of the cornea
to ILM of retina.
• Measures axial length along anatomic axis.
• Aphakic,pseudophakic and silicone oil filled eyes requires
adjustments for variation in velocity of sound in different
media.
• Can cause variable corneal compression.
Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-201.
• CHARACTERISTICS OF A GOOD SCAN:
• Corneal echo-single tall peak.
• Aqueous chamber-No echo.
• Anterior and posterior lens capsules-tall echoes.
• Vitreous cavity-Few to no echoes.
• Retina-tall,sharply rising echoes with no staircase at origin.
• Orbital fat-medium to no echoes.
• Anterior lens echoes-90% or more of maximum height.
• Posterior lens echoes-50%-75% of maximum height.
• Retinal echoes-75% or more of maximum height..
• Obtain atleast 3 scans(ideally 5-10)which are within 15mm of each other,average
5-10 most consistent results,giving lowest standard deviation(ideally <0.06).
• Measure axial length of BE if:
AL is less than 22mm or more than 25mm in either eye.
Difference between two eyes is more than 0.3mm.
• OPTICAL BIOMETRY:
• Highly accurate non invasive automated method for measuring anatomical details
of eye.
• Introduced in late 1990s.
• 1999-IOL Mater 500(Carl zeiss Meditec,Jena,Germany).
• Technologies used:1.Partial coherence interferometry (PCI)
2. Optical low coherence reflectometry (OLCR)
3. Swept-source optical coherence tomography(SS-OCT)
Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-201.
• Partial coherence interferometry :
• 1986-Fercher and Roth.
• Infrared light.
• Tissue surfaces with differing refractive indices reflect this light.
• Interferometric techniques are then used to measure ocular distances.
• Optical low-coherence reflectometry:
• Uses principle of a Michaelson interferometer.
• Low coherence infrared light produced by a superluminescent diode is split into
two beams by a coupler.
• One beam introduced into eye,other directed towards surface reference mirror.
• Tissue interfaces reflects light.
• Interference pattern detected by a detector.
• Exact location from which the light was reflected from within the eye is
determined by scanning the reference beam.
Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-201
• Swept-source optical coherence tomography( SS-OCT):
• Uses swift sweeping laser as the source.
• The reference mirror is kept static.
• The captured interference signal then undergoes Fourier transformation.
• Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-
201
• Nazm N, Chakrabarti A. Update on optical biometry and intraocular lens power calculation. TNOA J Ophthalmic Sci Res
2017;55:196-210
• SETTINGS OF A BIOMETER:
• Calibration:-Check with the model eye provided from time to time for accuracy.
• Gain/sensitivity settings:
Normal gain-70%
Dense cataracts,High Myopia,Ocular opacities-Increase gain
Silicone filled eyes,pseudophakic eyes-Decrease gain
• Sound velocity settings:To be adjusted appropriately in
Normal
Cataractous
Aphakic
Pseudophakic
Mastering IOL-Ashok Garg pg 33
• OPTICAL BIOMETERS:
• IOLMaster 500
• IOLMaster 700
• Lenstar LS 900
• Eyestar 900
• OA-2000(Tomey)
• Argos Advanced Optical Biometer(Movu)
• Aladdin
• AL-Scan (Nidek)
• Galilei G6 Lens
Professional
• IOLMaster 500:
• First commercially available optical biometer from Carl Zeiss.
• Uses 780nm infrared laser as source.
• Can measure axial length to within 0.02 mm accuracy which is
5x more than that of an immersion ultrasound biometer.
• Parameters measured:Axial length(AL)
Keratometry(K)
Anterior chamber depth(ACD)
White-to-white distance(WTW)
• Formulae onboard-SRK 2
SRK-T
Haigis
Hoffer Q
Holladay-1
Haigis-L
Holliday 2
• Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-
201
• LENSTAR LS-900:
• OLCR technology.
• One of the first devices which could measure
Lens thickness(LT).
• A low-coherent beam of light of 820 nm wavelength
produced from a superluminescent diode is used.
• Can perform corneal topography-useful in Toric IOL
planning.
• Parameters measured:Axial length(AL)
Keratometry(K)
Anterior chamber depth(ACD)
White-to-white distance(WTW)
Lens thickness(LT)
Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-201
• Formulae onboard:Holladay IOL Consultant Professional
Barrett Suite
Hill-radial basis function (RBF)
Masket
Modified Masket
Shammas
• Additional software-Okulix and Olsen formulae also available.
Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-201
Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J
Ophthalmol 2019;31:191-201
Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol
• INTRAOPERATIVE ABERROMETRY:
• Can be performed during cataract surgery.
• Aphakic and pseudophakic refractive measurements.
• Can be attached to surgical microscope.
• Provide real-time information on axis of placement of a toric IOL, position of
limbal relaxing incisions.
• Intraoperative aberrometers are particularly beneficial in toric, multifocal, and
accommodative IOL implantation and also in cataract surgery in eyes postcorneal
refractive surgery.
• The two commercially available intraoperative aberrometers are: Optiwave
Refractive Analysis (ORA) and Holos IntraOp.
• OTHER USES OF OPTICAL BIOMETERS:
• Toric IOL calculation.
• IOL power calculation in eyes post corneal refractive surgery.
• Lens centration
• Central corneal thickness.
• Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-
201
• ADVANTAGES OF OPTICAL BIOMETRY:
Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol
2019;31:191-201
• R CECH et al:
• The median of axial eye length measured using an optical biometer was 23.08
mm, and the median of axial eye length measured using ultrasound biometry was
22.93 mm. The difference between these values was 0.15 mm (150 microns).
• Average variation of dioptric power of an implanted IOL from retrospectively
established optimum value of the IOLs optical power was 0.40 D lower with
optical biometry and 0.16 D lower with ultrasound biometry.
• Unmbiguous preference of ultrasound biometry over optical biometry.
• ChiaTMT et al:
• Optical biometry AL and IOL power were not significantly different from
ultrasound measurements.
• Cech R, Utíkal T, Juhászová J. Srovnání optické a ultrazvukové biometrie a zhodnocení užívání obou metod
v praxi [Comparison of optical and ultrasound biometry and assessment of using both methods in practice].
Cesk Slov Oftalmol. 2014 Feb;70(1):3-9. Czech. PMID: 24862369.
• DISADVANTAGES OF OPTICAL BIOMETRY:
• The cost of optical biometers is significantly more than ultrasound biometers.
• Dense cataracts and corneal opacities may preclude accurate axial length
determination optical biometry.Immersion ultrasound biometry is preferred in
such cases, as it is more repeatable due to lack of variable corneal compression.
• Many of these difficulties have, however, been overcome with the advent of
swept-source optical coherence tomography (SS-OCT) technology.
• Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-
201
• CORNEAL POWER:
• Second most important factor.
• 1.0 D error in corneal power results in 1.0D
postrefractive error.
• Methods of measurement:
Keratometry
Pentacam
Galilei
BAUSCH AND LOMB KERATOMETER
PENTACAM
GALILIE
• POINTS TO BE CONSIDERED FOR KERATOMETRY:
• Ideally be recalibrated after 20 cases for single observer.
• Always do keratometry first before tonometry or any other corneal contact
procedures.
• Re measure corneal curvature of both eyes if:
Corneal curvature is less than 40D or more than 47D.
The difference in corneal cylinder is more than 1D between eyes.
The corneal cylinder correlates poorly with the refraction cylinder
Mastering IOL-Ashok garg-pg36
• EFFECTIVE LENS POSITION:
• Historically referred as the anterior chamber depth(ACD).
• ACD was found to be inaccurate for “in the bag” IOL.
• Denotes the distance that the principal plane of IOL will sit behind the cornea.
• 1mm error in ELP leads to-0.25-0.5D of error.
• Factors influencing effective IOL position:
Anatomical factors:Axial length,steepness of cornea,limbal white to white
measurement,preoperative anterior chamber depth and lens thickness.
IOL-related factors:Shape,length,flexibility,anterior angulation,material of haptic of
IOL
Surgeon-related factors:Surgeon’s individual surgical technique
Bag to sulcus shift:PCR,Loss of anterior capsule integrity-IOL needs to be placed in
ciliary sulcus-needs deduction (0.50-0.75D) from calculated IOL power .
• Biometry for IOL-Power calculation-AAO
• Using gaussian reduction equation,formula for calculating IOL power can be
expressed in accordance with refractive index-
• P-Power of target IOL(In D)
• K-Average dioptric power of central cornea(In D)
• AL-Axial length(in mm)
• C-ELP(in mm)
• nv-refractive index of vitreous
• Na-refractive index of aqueous
Biometry for IOL-Power calculation-AAO
FORMULAE
• Based upon their derivation,the various IOL power formulae have been grouped
into-
• THEORETICAL FORMULAE:Based on mathematical principles revolving around
‘schematic eye’.
• REGRESSION FORMULAE:Arrived by looking at postoperative outcomes
retrospectively.
• Artificial intelligence-based formulae: These formulae do use a form of
regression, but they rely on huge databases and employ artificial intelligence-
based complex statistical models to arrive identify relationships between
variables.
• Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-
201
• Vergence formulae:Vergence formulae arrive at an IOL power using Gaussian
optics.
• Gaussian optics makes an assumption that image vergence = object vergence +
lens vergence.
• Most modern-day IOL power calculation formulae are based on the following
equation formulated by Fyodorov.
• IOL power = (1336/[AL-ELP]) – (1336/[1336/{1000/([1000/ DPostRx] – V) + K} –
ELP])
• K-Net corneal power ELP-Effective lens position V-Vertex distance
• AL-Axial length DPostRx-Desired refraction
• Vergence formulae are subclassified into 2-variable,3-variable,5-variable and
7-variable.
Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-201
• Ray tracing:
• Ray tracing is a strategy to calculate the path that a light ray will travel when passed
through an optical system.
• Gaussian optics assumes presence of a single optical axis and all rays of interest make
only very small angles to this.
• 3rd Generation vergence formulae assume that all components of optical system are thin
lenses
• To describe the IOL position in this method, the radii of curvature of both anterior and
posterior corneal surfaces, their asphericity, central IOL thickness, and refractive index of
refraction are all used.
• Gaussian optics formulae rely on postoperative refraction to back calculate ELP-less
accurate due to errors in postoperative refraction and incorrect IOL power labelling.
• The final IOL position is a true geometrical position described by the distance between
the apex of the posterior surface of cornea and the apex of the anterior surface of the
IOL (postoperative ACD).
• Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-
201
• Advantages:
• Postoperative ACD can be measured directly using PCI and compared with
preoperative estimated ACD.
• Predict IOL power more accurately in abnormally short,long eyes and in eyes that
have undergone corneal refractive surgery.
• Examples-Okulix,Olsen formula and PhacoOptics
• Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-
201
CLASSIFICATION OF FORMULAE BASED ON GENERATIONS
1ST GENERATION 2ND GENERATION 3RD
GENERATION
4TH
GENERATION
5TH
GENERATION
THEORETICAL
FORMULAE
REGRESSI
ON
FORMULA
E
THEORETICA
L
FORMULAE
REGRESSION
FORMULAE
BINKHORST SRK-1 MODIFIED
BINKHORST
SRK-2 HOLLADAY-1 HOLLADAY-2 HAIGIS
COLENBRAND
ER-HOFFER
SRK-T OLSEN
GILL’S HOFFER’S Q
CLAYMAN’S
FYODOROV
• 1ST GENERATION FORMULAE:
• 1-THEORETICAL FORMULAE:
• Based on 3 variables-AL of eye
K-reading
Estimated postoperative ACD
• A-BINKHORST FORMULA: P=
𝟏𝟑𝟑𝟔(𝟒𝒓−𝒂)
(𝒂−𝒅)(𝟒𝒓−𝒅)
• P-IOL power in diopters
• r-corneal radius in millimeters(average)
• a-AL in mm
• d-assumed postoperative ACD plus corneal thickness
• Disorders of lens and cataract surgery-A K Khurana-pg 112
• B-COLENBRANDER-HOFFER FORMULA:
• P=
1336
𝑎−𝑑−0.05
-
1336
1336
𝐾
−𝑑−0.05
• K-average keratometry in diopters
• GILL’S FORMULA:
• P=129.40+(-108 x K)+(-2.79 x AL)+(0.26 x LCL)+(0.38 x Ref)
• P-desired IOL power
• K-refractive power of cornea in diopters
• AL-axial length in mm
• LCL-distance of apex of anterior corneal surface to the apex of IOL in mm
• Ref-desired postoperative refraction
• Disorders of lens and cataract surgery-A K Khurana-pg 113
• CLAYMAN’S FORMULA:
• Assume-
• Emmetropizing iol=18D
• Emmetropic AL=24mm
• Emmetropic average keratometry value-42 D
• 1mm change in AL=3D of IOL power
• 1D in keratometry=1D of IOL power
• FYODOROV FORMULA:P =
1336−𝐿𝐾
𝐿−𝐶 −
𝐶𝐾
1336
• P-implant power for emmetropia
• L-axial length(AL)in mm
• K-corneal curvature in diopters
• C-estimated postoperative anterior chamber depth(ACD)
• Disorders of lens and cataract surgery-A K Khurana-pg 113
• ALGEBRAIC TRANSFORMATION OF THEORETICAL FORMULAE-
• P=
𝑁
𝐿−𝐶
-
𝑁𝐾
𝑁−𝐾𝐶
• P-Implant power for emmetropia
• N-aqueous and vitreous RI
• C-estimated postoperative ACD in millimeters
• L-AL in mm
• K-corneal curvature in diopters
• DRAWBACKS:
• Short eyes-large emmetropic value,Long eyes-too small value
• Too cumbersome to apply
• Guess of ACD is required
• Were developed when iris-supported lenses were used
• Based on theoretical simplistic assumption about optics of eye
• Disorders of lens and cataract surgery-A K Khurana-pg 114
• 2-REGRESSION FORMULAE:
• Based on actual regression analysis of actual postoperative results of implant
power as a function of variables of corneal power and AL.
• SRK-1 FORMULA:
• Introduced by Sanders,Retzlaff and Kraff
• Based on regression analysis,taking into account the retrospective analysis of
postoperative refractions.
• Postoperative ACD was not included,replaced with A constant which was unique
for different type of IOL,determined by the manufacturer depending upon
material,position of eye,optic and haptic design.
• Disorders of lens and cataract surgery-A K Khurana-pg 114
• FORMULA:
• P=A-2.5L-0.9K
P-IOL power
A-constant specific for each lens
L-AL in mm
K-avg keratometry in diopters
• Performs well for eyes with AL between 22.0 and 24.5mm
• Short eyes-too small value,Long eyes-too large value.
• Disorders of lens and cataract surgery-A K Khurana-pg 114
• SECOND-GENERATION FORMULAE:
• 1-THEORETICAL FORMULAE:
• Modified Binkhorst formulae:1981-Used AL as scaling factor for effective lens
position and presented a formula to better predict ACD.
• 2-REGRESSION FORMULAE:
• SRK-2 FORMULA: P=A-2.5L-0.9K
• L<20mm A+3.0
• L is 20.00-20.99mm A+2.0
• L is 21.00-21.99mm A+1.0
• L is 22.0-24.5mm A
• L>24.5mm A-0.5
• Disorders of lens and cataract surgery-A K Khurana-pg 114
• Modified SRK-2 Formula:
• Based on AL,A-Constant is modified as:
• If L<20mm A+1.5
• L is 20-21mm A+1.0
• L is 21-22mm A+0.5
• L is 22.0-24.5mm A
• L is 24.5-26.0mm A-1.0
• L is >26mm A-1.5
• Disorders of lens and cataract surgery-A K Khurana-pg 114
• 3rd GENERATION FORMULAE:Requires 2 variables-K and AL
• Holladay-1 formula:
• Introduced in 1988
• 2 variable predictor(AL and Keratometry)could significantly improve the predictor
of effective lens position,particularly in unusual eyes.
• Formula was proposed based on geometric relationship of the anterior segment.
• Had better outcome for eyes between 22.00mm -26.00mm compared to other 3rd
generation.
• SRK-T formula:
• Non-linear theoretical optical formula,empirically optimised for postoperative
ACD,retinal thickness and corneal refractive index.
• Combines advantages of both theoretical and empirical analysis.
• Better outcomes for eyes>26.0 mm compared to other 3rd generation formulae
• Disorders of lens and cataract surgery-A K Khurana-pg 115
• Hoffer-Q formula:
• Variables required-K and AL
• Better outcome for eyes <22.00mm,compared to other 3rd generation
• Disorders of lens and cataract surgery-A K Khurana-pg 115
• 4TH GENERATION FORMULAE:
• Holladay-2 formula:
• 1993-Dr.Holladay led a worldwide study involving 34 cataract surgeons to determine
which of the 7 variables were relevant as predictors of ELP.
• Led to concept of 9 types of eyes,not 3(short,medium and long)
• 7 Variables required-
1-K-Average K reading in diopters.
2-AL-Measured ultrasonic axial length in mm
3-ACD-Distance from corneal
4-LT-Lens thickness
5-Horizontal WTW-White to white corneal diameter
6-Age
7-Preoperative refraction
• Allows calculation of IOL power in many different types of eyes but also honing of
individual result by personalising A constant
• Available as a part of Holladay IOL consultant/surgical outcomes assessment programme
• Olsen formula:
• Variables required-
K
AL
ACD
LT
Horizontal WTW
• Improved performance over 3rd generation formulae for eyes with AL 20.00-
26.00mm
• Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-
201
• 5TH GENERATION FORMULAE:
• Haigis:
• Variables required
K
AL
ACD
• Very good outcomes for eyes across the AL range and best reported outcomes for
eyes >28.00mm.For best results,constants need to be optimised requiring data
from atleast 500 eyes.
• Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-
201
• MODERN IOL FORMULAE:
• LADAS SUPER FORMULA:
• Devloped by Dr John G.Ladas,Albert Jun,Aazim Siddiqui,Uday Devgan.
• Points of agreement and disagreement between Hoffer Q,Holladay 1,Holladay 1
with Wang-Koch adjustment,Haigis and SRK-T formulae for various ranges of axial
length were identified.
• Ideal portions from four of five formulae were used to create super surface.
• Dynamic formula-constantly being updated using inputs from surgeons
worldwide.
• Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-
201
• BARRETT UNIVERSAL 2:
• Devloped by Graham D Barrett.
• Based on theoretical model eye in which ACD is related to axial length and
keratometry.
• Can be unmodified in eyes of all axial lengths and for all IOL designs-universal.
• 5 Variables-AL,Keratometry,ACD,LT,Horizontal WTW corneal diameter to calculate
ELP along with A constant and desired postoperative refraction.
• Predicts IOL power accurately in eyes with extremely long axial lengths.
• Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-
201
• HILL-RADIAL BASIS FUNCTION CALCULATOR:
• Devloped by Warren E Hill with RBF calculator physician team.
• Complex mathematical model-artificial neural network.
• Has a property of adaptive learning.
• Dynamic formula-becomes more accurate in its prediction as more data is fed
into it.
• Optimised for Haag streit Lenstar in combination with Alcon SN60WF biconvex
IOL for powers from +30.00D to +6.00D and Alcon MA60MA meniscus design IOL
powers from +5.00D to -5.00D.
• Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-
201
• OLSEN FORMULA:
• Devloped by Dr Thomas Olsen.
• Paraxial and ray tracing optics.
• Regression method used to determine ACD.ACD and LT used to determine ELP.
• IOL position is described as a fraction of the capsular bag size using a new
constant called the C constant.
• Variables used-ACD,AL,LT and keratometry.
• power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-201
• PHACOPTICS:
• Ray tracing.
• Uses concept of Olsen’s C constant.
• WANG-KOCH MODIFICATION:
• Traditional IOL formulae for abnormally long eyes-postoperative residual
hyperopia.
• Modification for AL more than 25.2mm to avoid refractive surprise.
• Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-
201
• FULL MONTE INTRAOCULAR LENS:
• Uses a method of analysis called the Markov Chain Monte Carlo process.
• A probability instead of a single emmetropic power is displayed,surgeon can
choose appropriate IOL power.
• Evolves continuously optimizing itself as new data is received.
• Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-
201
• TORIC CALCULATORS:
• Correction of corneal astigmatism.
• Components-1)Accurate K
2)Reliable method to calculate power of toric IOL.
• Ideal characteristics:
• It is comprehensive, that is, it can be used for preoperative planning, as well as
for refractive surprises
• Software should preferably be a part of the biometric device
• It should be applicable to all IOL types
• There should be dynamic display of variables (e.g., dynamic adjustment of
astigmatic effects of phaco incision)
• It should provide an alert for “axis-flip”
• It should take into account posterior corneal astigmatism (PCA)
• Nazm N, Chakrabarti A. Update on optical biometry and intraocular lens power calculation. TNOA J Ophthalmic Sci Res
2017;55:196-210
• TORIC IOL CALCULATORS:
• ASSORT TORIC CALCULATOR
• BARRETT TORIC CALCULATOR
• VERION
• CALLISTO EYE
• ASTIGMATISMFIX.COM
• Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-
201
Nazm N, Chakrabarti A. Update on optical biometry and intraocular lens power calculation. TNOA J Ophthalmic Sci Res
• FUTURE OF INTRAOCULAR LENS POWER CALCULATION:
• UniversIOL CALCULATOR:
• Designed by Dr.Samir Sayegh.
• Web based
• Combines all high quality 3rd and 4th generation formulae with a toric IOL
calculator.
• Instead of proposing a new IOL power calculation formula, the surgeon is
provided with a choice of IOL power calculation formulae or combinations of
formulae, a choice of computational methods, and a choice of IOL ranking
criteria.
• Tells how much the formula differ from each other,gives surgeon an idea how
close she/he will be to the target.
• Also contains all IOLs made so that an appropriate power as well as IOL can be
chosen.
• Nazm N, Chakrabarti A. Update on optical biometry and intraocular lens power calculation. TNOA J Ophthalmic Sci Res
2017;55:196-210
• OKULIX:
• Ray tracing formula.
• Radius of curvature of anterior and posterior corneal surfaces need to be
measured using a corneal topographer.
• IOL is described by radii of curvature of its anterior and posterior
surfaces,thickness and refractive index.
• Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-
201
CLINICAL VARIABLES
PERSONALISATION:
• Introduced by Retzlaff.
• Used A constant to refine the formula based on surgeon’s past experience
and data.
• Parameters required:
Axial length(Pre-op)
Corneal Power(Pre-op)
IOL power
Postoperative refractive error
• Same lens style required by one manufacturer implanted by one surgeon.
• Same biometry and technician.
• Eyes with postoperative surprises or acuity worse than 20/40 not included-
poor accuracy.
• Mastering Intraocular Lenses-Ashok garg-pg no 42
• Involves backsolving for the exact IOL position that would produce the resultant
refractive error with AL and K.
• All the “ideal”IOL positions are averaged to arrive at personalised value for use in
future.
• Can be easily done using Hoffer Programs or Holladay IOL Consultant computer
programs.
• Mastering Intraocular Lenses-Ashok garg-pg no 42
• BIOMETRY IN SPECIAL CONDITIONS:
• APHAKIA:
• Required for secondary IOL implantation.
• In aphakes,sound travels at speed of 1532m/s.
• A Scan-2 lens spikes are replaced by anterior vitreous phase and posterior lens
capsule.
• Immersion technique of biometry-method of choice.
• Present biometers-aphakic mode is available
• Mastering Intraocular Lenses-Ashok garg-pg no 50
• Mackool et al:
• IOL power (D) = Aphakic refraction × 1.75
• This method was found to be more accurate when anterior chamber IOLs were
implanted.
• Ianchulev and Leccisotti:
• IOL power (D) =0.07x 2 + 1.27x + 1.22, where x = aphakic refraction
• This method was found to be more accurate for IOL placed posteriorly.
• Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-
201
• PSEUDOPHAKIA:
• IOL exchange.
• High spike at lens followed by artificial chain
of reduplication of echoes-can be misinterpreted as
spikes from retina.
• Lower gain-eliminates artificial spikes and increases
retinal spikes.
• Sound velocity depends upon type of IOL.
• Modern biometers-options are available for phakic,
pseudophakic or aphakic mode.
Disorders of lens and cataract surgery-A K Khurana-pg 116
• PIGGYBACK IOL FOR PSEUDOPHAKIA:
• Indicated in patients with significant residual refractive error following the
primary IOL implantation.
• Surgically easier and more predictable optically rather than IOL exchange.
• Formulae:
Holladay’s refractive formula
Gill’s nomogram-calculated from residence spherical based on axial length
• VITRECTOMIZED EYE:
• RI of silicon oil is much less than that of vitreous,acts as negative lens.
• Silicon oil-sound attenuation-986m/s.
• Increase the system gain and adjust sound velocity accordingly.
• Result obtained is Apparent axial length(AAL)
which has to be converted to True
axial length(TAL)
• TAL=1133/1550 AAL
• PEDIATRIC BIOMETRY:
• Dahan et al’ simplified approach based on axial length
AXIAL LENGTH IOL POWER
17mm 28D
18mm 27D
19mm 26D
20mm 24D
21mm 24D
22mm 23D
23mm 23D
>24mm Axial length-1D
• Immersion A-scan preferred.
• IOL power-SRK/T and Holladay 2 are preferred.
• Undercorrection-
According to Dahan and Drusedau undercorrection of 20% in children <2 years
10% in children between 2 and 8 years
• Prost suggested 20% undercorrection between 1 and 2 years of age,
15% undercorrection between 2 and 4 years
10% between 4 and 8 years of age.
AAO-management of pediatric cataract.
• POST KERATOREFRACTIVE SURGERY:
• Standard methods of IOL power calculation-Inadequate.
• Factors altered in corneal refractive surgeries:
Anterior corneal curvature
Ratio of anterior and posterior corneal curvature
Corneal asphericity
• Modern formulae use the above factors to calculate IOL power-adds to error.
• Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-
201
• HAIGIS-L FORMULA:
• The corneal radius of curvature (r corr) is modified as follows:
• r - corr =
331.5
−5.1625 𝑥 𝑟 𝑚𝑒𝑎𝑠+82.2603−0.35
• where r meas is the measured corneal radius of curvature.
• SHAMMAS FORMULA:
• Calculated corneal power, K = 1.14 ×TK −6.8
• where TK is the post-LASIK corneal topography central K.
• Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-
201
• MASKET METHOD:
• Regression formula-Samuel musket.
• AL>23mm-Holladay 1.
• AL<23mm-Hoffer Q.
• Adjustment of IOL power-
(LSE × −0.326) +0.101 = Post-LASIK IOL power adjustment
Where LSE is the laser corrected spherical equivalent adjusted for vertex distance
• MODIFIED MASKET METHOD:
• (LSE × −0.4385) +0.0295 = Post-LASIK IOL power adjustment
• Others:
• ASCRS postrefractive calculator
• Hoffer–Savini LASIK intraocular lens power tool
• Barrett true-K formula
• The McCarthy post refractive intraocular lens calculator
PROBLEMS AND ERRORS-IOL POWER
SURPRISE
• AXIAL LENGTH:Most common source of error
• Axial length >25mm,higher incidence of staphylomas.
• Artificial shortening due to contact applanation ultrasound.
• Technician inexperience.
• Wrong average ultrasound velocity.
• Poor IOL master readings-Dense cataracts.
• Silicone filled eye.
• CORNEAL POWER:Second most common cause.
• Eyes with previous corneal refractive surgery.
• Wrong calibration.
• Not setting index of refraction(IR) in IOLMaster-leads to error in Hoffer Q formula.
• Contact lens wear.
• Corneal scarring.
• EFFECTIVE POSITION OF IOL:Least effective factor
• IOL in deeper or shallower position than predicted by the formula
• ACD constants not personalised to individual IOL style,surgeon and clinic.
• FORMULAS:
• Regression formulas(SRK 1) for eyes outside the normal AL range of 22-24.5mm.
• MISCELLANEOUS:
• Manfacturer labelling error.
• Insertion of wrong IOL.
• PREVENTION OF ERRORS:
• Immersion A-scan/IOLMater to measure AL.
• Suspect staphylomas in eyes>25mm-Use IOLMaster.
• Well-trained experienced technician.
• Regularly calibrate manual keratometers.
• Evaluate Biometer scans for reliability.
• Keep contact lens out for 2 weeks prior to keratometry
• Silicone oil-optical biometers/AL x 0.71
• AL<22mm-Hoffer Q
• AL 22-26mm-Holladay 1
• AL >26mm-SRK/T
• Avoid using SRK regression formulas(SRK 1 and 2)
• Personalise ELP factors in the formulas.
• Personally select IOL power for individual patient.
• Set IR to 1.3375 in set-up screen of IOLMaster.
• Post-refractive surgery-
Shammas method
Masket method
Haigis formula
• Consider delaying IOL implantation until cornea has healed after a penetrating
keratoplasty rather than performing a “triple procedure”.
THANK YOU

More Related Content

What's hot (20)

Orbscan &amp; topo
Orbscan &amp; topoOrbscan &amp; topo
Orbscan &amp; topo
 
Phakic iols
Phakic iolsPhakic iols
Phakic iols
 
Diagnosis of pre perimetric glaucoma
Diagnosis of pre perimetric glaucomaDiagnosis of pre perimetric glaucoma
Diagnosis of pre perimetric glaucoma
 
IOL Master
IOL MasterIOL Master
IOL Master
 
AS-OCT
AS-OCTAS-OCT
AS-OCT
 
Pentacam
Pentacam Pentacam
Pentacam
 
Accommodating IOLs
Accommodating IOLsAccommodating IOLs
Accommodating IOLs
 
Surgical strategies for small pupils - Malyugin Ring
Surgical strategies for small pupils - Malyugin RingSurgical strategies for small pupils - Malyugin Ring
Surgical strategies for small pupils - Malyugin Ring
 
Pediatric refraction
Pediatric       refractionPediatric       refraction
Pediatric refraction
 
Biometry
BiometryBiometry
Biometry
 
Biometry & Iol calculations
Biometry & Iol calculationsBiometry & Iol calculations
Biometry & Iol calculations
 
collagen crosslinking.pptx
collagen crosslinking.pptxcollagen crosslinking.pptx
collagen crosslinking.pptx
 
Keratometer
KeratometerKeratometer
Keratometer
 
IOL power calculation special situations
IOL power calculation special situations IOL power calculation special situations
IOL power calculation special situations
 
Corneal Topography
Corneal TopographyCorneal Topography
Corneal Topography
 
Iol master
Iol masterIol master
Iol master
 
Orthoptic evaluation 1
Orthoptic evaluation 1Orthoptic evaluation 1
Orthoptic evaluation 1
 
Corneal collagen cross linking
Corneal collagen cross linkingCorneal collagen cross linking
Corneal collagen cross linking
 
Surgical induced astigmatism
Surgical induced astigmatismSurgical induced astigmatism
Surgical induced astigmatism
 
Phakic iol ppt
Phakic iol pptPhakic iol ppt
Phakic iol ppt
 

Similar to BIOMETRY AND IOL POWER CALCULATION

A-Scan Biometry.pptx
A-Scan Biometry.pptxA-Scan Biometry.pptx
A-Scan Biometry.pptxAsif469093
 
IOL CALCULATIONS IN SPECIAL CASES
IOL CALCULATIONS IN SPECIAL CASES IOL CALCULATIONS IN SPECIAL CASES
IOL CALCULATIONS IN SPECIAL CASES PGOPTHALEDUCATION
 
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation Anis Suzanna Mohamad
 
Minimizing wrong iol calculation
Minimizing wrong iol calculationMinimizing wrong iol calculation
Minimizing wrong iol calculationAbdelmonem Hamed
 
Biometry- Iol power and calculation final ppt.pptx
Biometry- Iol power and calculation final ppt.pptxBiometry- Iol power and calculation final ppt.pptx
Biometry- Iol power and calculation final ppt.pptxKervi Mehta
 
biometry for ON.ppt
biometry for ON.pptbiometry for ON.ppt
biometry for ON.pptmikaelgirum
 
Ocular Biometry- IOL calculation methods
Ocular Biometry- IOL calculation methodsOcular Biometry- IOL calculation methods
Ocular Biometry- IOL calculation methodsDrMadhumita Prasad
 
IOL CALCULATIONS BY MRS ANTHONIA.pptx
IOL CALCULATIONS BY MRS ANTHONIA.pptxIOL CALCULATIONS BY MRS ANTHONIA.pptx
IOL CALCULATIONS BY MRS ANTHONIA.pptxSamuel Medeludo
 
A scan biometry | How to Use A-scan? Types of A-Scan Biometry?
A scan biometry | How to Use A-scan? Types of A-Scan Biometry?A scan biometry | How to Use A-scan? Types of A-Scan Biometry?
A scan biometry | How to Use A-scan? Types of A-Scan Biometry?Naeem Ahmad
 
Iol power calculation normal and post lasik eyes
Iol power calculation normal and post lasik eyesIol power calculation normal and post lasik eyes
Iol power calculation normal and post lasik eyesDINESH and SONALEE
 
Optical Biometry Measurements For Future Iol’S
Optical Biometry Measurements For Future Iol’SOptical Biometry Measurements For Future Iol’S
Optical Biometry Measurements For Future Iol’Smeikocat
 
Biometry instruments & equipment
Biometry instruments & equipmentBiometry instruments & equipment
Biometry instruments & equipmentDevdutta Nayak
 

Similar to BIOMETRY AND IOL POWER CALCULATION (20)

Biometry
BiometryBiometry
Biometry
 
BIOMETERY
BIOMETERYBIOMETERY
BIOMETERY
 
Biometery
Biometery Biometery
Biometery
 
Biometry-1, rifat.pptx
Biometry-1, rifat.pptxBiometry-1, rifat.pptx
Biometry-1, rifat.pptx
 
A-Scan Biometry.pptx
A-Scan Biometry.pptxA-Scan Biometry.pptx
A-Scan Biometry.pptx
 
IOL CALCULATIONS IN SPECIAL CASES
IOL CALCULATIONS IN SPECIAL CASES IOL CALCULATIONS IN SPECIAL CASES
IOL CALCULATIONS IN SPECIAL CASES
 
Iol
IolIol
Iol
 
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
National Ocular Biometry Course (NOBC) 2015 An echoslide presentation
 
Minimizing wrong iol calculation
Minimizing wrong iol calculationMinimizing wrong iol calculation
Minimizing wrong iol calculation
 
Biometry- Iol power and calculation final ppt.pptx
Biometry- Iol power and calculation final ppt.pptxBiometry- Iol power and calculation final ppt.pptx
Biometry- Iol power and calculation final ppt.pptx
 
biometry for ON.ppt
biometry for ON.pptbiometry for ON.ppt
biometry for ON.ppt
 
Biometry made easy
Biometry made easy Biometry made easy
Biometry made easy
 
Ocular Biometry- IOL calculation methods
Ocular Biometry- IOL calculation methodsOcular Biometry- IOL calculation methods
Ocular Biometry- IOL calculation methods
 
IOL CALCULATIONS BY MRS ANTHONIA.pptx
IOL CALCULATIONS BY MRS ANTHONIA.pptxIOL CALCULATIONS BY MRS ANTHONIA.pptx
IOL CALCULATIONS BY MRS ANTHONIA.pptx
 
Biometryy
BiometryyBiometryy
Biometryy
 
A project on
A project onA project on
A project on
 
A scan biometry | How to Use A-scan? Types of A-Scan Biometry?
A scan biometry | How to Use A-scan? Types of A-Scan Biometry?A scan biometry | How to Use A-scan? Types of A-Scan Biometry?
A scan biometry | How to Use A-scan? Types of A-Scan Biometry?
 
Iol power calculation normal and post lasik eyes
Iol power calculation normal and post lasik eyesIol power calculation normal and post lasik eyes
Iol power calculation normal and post lasik eyes
 
Optical Biometry Measurements For Future Iol’S
Optical Biometry Measurements For Future Iol’SOptical Biometry Measurements For Future Iol’S
Optical Biometry Measurements For Future Iol’S
 
Biometry instruments & equipment
Biometry instruments & equipmentBiometry instruments & equipment
Biometry instruments & equipment
 

Recently uploaded

Call Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
Call Girl Lucknow Mallika 7001305949 Independent Escort Service LucknowCall Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
Call Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknownarwatsonia7
 
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceCollege Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceNehru place Escorts
 
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowKolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowNehru place Escorts
 
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Miss joya
 
Call Girl Surat Madhuri 7001305949 Independent Escort Service Surat
Call Girl Surat Madhuri 7001305949 Independent Escort Service SuratCall Girl Surat Madhuri 7001305949 Independent Escort Service Surat
Call Girl Surat Madhuri 7001305949 Independent Escort Service Suratnarwatsonia7
 
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls ServiceCall Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Servicesonalikaur4
 
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...Garima Khatri
 
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...narwatsonia7
 
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Call Girl Indore Vrinda 9907093804 Independent Escort Service Indore
Call Girl Indore Vrinda 9907093804 Independent Escort Service IndoreCall Girl Indore Vrinda 9907093804 Independent Escort Service Indore
Call Girl Indore Vrinda 9907093804 Independent Escort Service IndoreRiya Pathan
 
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Miss joya
 
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...narwatsonia7
 
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...narwatsonia7
 
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...CALL GIRLS
 
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original PhotosCall Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photosnarwatsonia7
 
Aspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliAspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliRewAs ALI
 
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls JaipurCall Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipurparulsinha
 

Recently uploaded (20)

Call Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
Call Girl Lucknow Mallika 7001305949 Independent Escort Service LucknowCall Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
Call Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
 
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
 
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceCollege Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
 
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowKolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
 
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
 
Call Girl Surat Madhuri 7001305949 Independent Escort Service Surat
Call Girl Surat Madhuri 7001305949 Independent Escort Service SuratCall Girl Surat Madhuri 7001305949 Independent Escort Service Surat
Call Girl Surat Madhuri 7001305949 Independent Escort Service Surat
 
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls ServiceCall Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Service
 
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
 
Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...
Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...
Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...
 
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
 
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
 
Call Girl Indore Vrinda 9907093804 Independent Escort Service Indore
Call Girl Indore Vrinda 9907093804 Independent Escort Service IndoreCall Girl Indore Vrinda 9907093804 Independent Escort Service Indore
Call Girl Indore Vrinda 9907093804 Independent Escort Service Indore
 
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
 
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
 
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
 
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
 
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
 
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original PhotosCall Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
 
Aspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliAspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas Ali
 
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls JaipurCall Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
 

BIOMETRY AND IOL POWER CALCULATION

  • 2. OUTLINE • HISTORY • COMPONENTS OF IOL POWER CALCULATION • BIOMETRY • FORMULAE • VARIABLES • PROBLEMS AND ERRORS
  • 3. HISTORY • 1949-1st IOL Implantation-Sir Harold Ridley • Implantation was done on Feb 8,1950-secondary IOL implantation. • Weight of the lens-17.4mg in water • Patient had a refractive surprise of nearly 20 D. • 1967-Fyodorov et al-IOL power using vergence formulae. • 1970s-”A “scans available-theoretical vergence formulae. • 1980s-IDEM lens attempted. • 1982-Gernet and Zorkendorfer-avg refractive power of lens-+23.70D. Biometry for Intra-Ocular Lens(IOL) Power Calculation-AAO
  • 4. GENERATIONS OF INTRAOCULAR LENSES GENERATION CHARACTERISTIC EXAMPLE 1st-1948 Experimental lens by Ridley Biconvex 2nd-1952 Anterior chamber angle fixated lens Baron’s Anterior chamber lens Strampelli rigid tripod anterior chamber IOL Dannheim flexible closed loop ACIOL Barraquer J-Loop anterior chamber IOL 3rd-1957 Iris supported,Iris clip lens Binkhorst iris clip lens Fyrodov type 2 “Sputnik”iris –clip IOL Worst platina IOL Copeland iris-plane IOL 4th-1956 Anterior chamber one piece lens Choyce Mark VIII rigid ACIOL Strampelli angle supported ACIOL 5th Posterior chamber lenses Shearing J-loop posterior chamber IOL
  • 5. CRUDE IOL POWER CALCULATIONS IN PAST • 1980s-Guess work approach. • Concept-IDEM lenses Standard lenses Emmetropia lenses Biometry for Intra-Ocular Lens(IOL) Power Calculation-AAO
  • 6. • IDEM LENSES: • Ideal emmetropic lenses • Pre-and post refractive refraction is same. • Recommended for emmetropic patients. • 20D artificial lens in posterior chamber will restore preoperative refractive error. Disorders of lens and cataract surgery-A K Khurana-pg 108
  • 7. • LIMITATION-Will restore the preoperative refractive error,only if the natural lens had refractive power of 23.70 D which is not always true, as refractive power is a combination of different factors-corneal curvature,depth of AC ,dioptric power of lens and length of eye. Disorders of lens and cataract surgery-A K Khurana-pg 108
  • 8. • STANDARD LENS: • 2D stronger than IDEM lens. • Renders eye pseudophakic about 1.5D myopic as compared to preoperative emmetropic refraction. • 80% of eyes-less than 1D nearsighted or farsighted-nearly correct in majority of patients. • Balanced distance and near vision. • Standard PCIOL-+22D • Standard ACIOL-+20D • Disorders of lens and cataract surgery-A K Khurana-pg 108
  • 9. • EMMETROPIA LENS: • For restoring emmetropia in previously ammetropic eyes. • Power calculation- HYPERMETROPIA-Power of IDEM lens+(Preoperative refractive error x 1.25). MYOPIA-Power of IDEM lens-(Preoperative refractive error x 1.25). • History to be taken-kind of glasses patient wore previously,rely on refractive values only if they come from reliable records,or if they match clinical findings. • Deviation of 2-3D is possible,can lead to significant errors. • Disorders of lens and cataract surgery-A K Khurana-pg 108
  • 10. • DRAWBACKS: • Based on guess work approach. • 1970s-’A’ scans were available-studies were conducted to establish various theoretical vergence formulas. • It was found IOL power is determined by Axial length Keratometry Effective IOL position Biometry for Intra-Ocular Lens(IOL) Power Calculation-AAO
  • 11. IOL POWER CALCULATION BIOMETRY FORMULAS CLINICAL VARIABLES • AXIAL LENGTH • CORNEAL POWER • EFFECTIVE LENS POSITION • GENERATIONS • PERSONALISATION • SPECIAL CIRCUMSTANCES • PROBLEMS AND ERRORS
  • 12. BIOMETRY • AXIAL LENGTH(AL): • Most important factor in biometric calculations. • Distance from tear film to RPE. • 1mm error in AL results in refractive error of 2.35D in a 23.5 mm eye,error being less in case of longer eyes as compared to shorter eyes. • Methods of measurement of axial length- Ultrasonic measurement Optical measurement Disorders of lens and cataract surgery-A K Khurana-pg 109
  • 13. • ULTRASONIC MEASUREMENT: • Dr Karl Ossoinig-developed standardised A scans • Methods-Applanation method Immersion method • Measures axial length from anterior surface of the cornea to ILM of retina. • Measures axial length along anatomic axis. • Aphakic,pseudophakic and silicone oil filled eyes requires adjustments for variation in velocity of sound in different media. • Can cause variable corneal compression. Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-201.
  • 14. • CHARACTERISTICS OF A GOOD SCAN: • Corneal echo-single tall peak. • Aqueous chamber-No echo. • Anterior and posterior lens capsules-tall echoes. • Vitreous cavity-Few to no echoes. • Retina-tall,sharply rising echoes with no staircase at origin. • Orbital fat-medium to no echoes. • Anterior lens echoes-90% or more of maximum height. • Posterior lens echoes-50%-75% of maximum height. • Retinal echoes-75% or more of maximum height.. • Obtain atleast 3 scans(ideally 5-10)which are within 15mm of each other,average 5-10 most consistent results,giving lowest standard deviation(ideally <0.06). • Measure axial length of BE if: AL is less than 22mm or more than 25mm in either eye. Difference between two eyes is more than 0.3mm.
  • 15. • OPTICAL BIOMETRY: • Highly accurate non invasive automated method for measuring anatomical details of eye. • Introduced in late 1990s. • 1999-IOL Mater 500(Carl zeiss Meditec,Jena,Germany). • Technologies used:1.Partial coherence interferometry (PCI) 2. Optical low coherence reflectometry (OLCR) 3. Swept-source optical coherence tomography(SS-OCT) Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-201.
  • 16. • Partial coherence interferometry : • 1986-Fercher and Roth. • Infrared light. • Tissue surfaces with differing refractive indices reflect this light. • Interferometric techniques are then used to measure ocular distances. • Optical low-coherence reflectometry: • Uses principle of a Michaelson interferometer. • Low coherence infrared light produced by a superluminescent diode is split into two beams by a coupler. • One beam introduced into eye,other directed towards surface reference mirror. • Tissue interfaces reflects light. • Interference pattern detected by a detector. • Exact location from which the light was reflected from within the eye is determined by scanning the reference beam. Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-201
  • 17. • Swept-source optical coherence tomography( SS-OCT): • Uses swift sweeping laser as the source. • The reference mirror is kept static. • The captured interference signal then undergoes Fourier transformation. • Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191- 201
  • 18. • Nazm N, Chakrabarti A. Update on optical biometry and intraocular lens power calculation. TNOA J Ophthalmic Sci Res 2017;55:196-210
  • 19. • SETTINGS OF A BIOMETER: • Calibration:-Check with the model eye provided from time to time for accuracy. • Gain/sensitivity settings: Normal gain-70% Dense cataracts,High Myopia,Ocular opacities-Increase gain Silicone filled eyes,pseudophakic eyes-Decrease gain • Sound velocity settings:To be adjusted appropriately in Normal Cataractous Aphakic Pseudophakic Mastering IOL-Ashok Garg pg 33
  • 20. • OPTICAL BIOMETERS: • IOLMaster 500 • IOLMaster 700 • Lenstar LS 900 • Eyestar 900 • OA-2000(Tomey) • Argos Advanced Optical Biometer(Movu) • Aladdin • AL-Scan (Nidek) • Galilei G6 Lens Professional
  • 21. • IOLMaster 500: • First commercially available optical biometer from Carl Zeiss. • Uses 780nm infrared laser as source. • Can measure axial length to within 0.02 mm accuracy which is 5x more than that of an immersion ultrasound biometer. • Parameters measured:Axial length(AL) Keratometry(K) Anterior chamber depth(ACD) White-to-white distance(WTW) • Formulae onboard-SRK 2 SRK-T Haigis Hoffer Q Holladay-1 Haigis-L Holliday 2 • Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191- 201
  • 22. • LENSTAR LS-900: • OLCR technology. • One of the first devices which could measure Lens thickness(LT). • A low-coherent beam of light of 820 nm wavelength produced from a superluminescent diode is used. • Can perform corneal topography-useful in Toric IOL planning. • Parameters measured:Axial length(AL) Keratometry(K) Anterior chamber depth(ACD) White-to-white distance(WTW) Lens thickness(LT) Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-201
  • 23. • Formulae onboard:Holladay IOL Consultant Professional Barrett Suite Hill-radial basis function (RBF) Masket Modified Masket Shammas • Additional software-Okulix and Olsen formulae also available. Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-201
  • 24. Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-201
  • 25. Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol
  • 26. • INTRAOPERATIVE ABERROMETRY: • Can be performed during cataract surgery. • Aphakic and pseudophakic refractive measurements. • Can be attached to surgical microscope. • Provide real-time information on axis of placement of a toric IOL, position of limbal relaxing incisions. • Intraoperative aberrometers are particularly beneficial in toric, multifocal, and accommodative IOL implantation and also in cataract surgery in eyes postcorneal refractive surgery. • The two commercially available intraoperative aberrometers are: Optiwave Refractive Analysis (ORA) and Holos IntraOp.
  • 27. • OTHER USES OF OPTICAL BIOMETERS: • Toric IOL calculation. • IOL power calculation in eyes post corneal refractive surgery. • Lens centration • Central corneal thickness. • Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191- 201
  • 28. • ADVANTAGES OF OPTICAL BIOMETRY: Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-201
  • 29. • R CECH et al: • The median of axial eye length measured using an optical biometer was 23.08 mm, and the median of axial eye length measured using ultrasound biometry was 22.93 mm. The difference between these values was 0.15 mm (150 microns). • Average variation of dioptric power of an implanted IOL from retrospectively established optimum value of the IOLs optical power was 0.40 D lower with optical biometry and 0.16 D lower with ultrasound biometry. • Unmbiguous preference of ultrasound biometry over optical biometry. • ChiaTMT et al: • Optical biometry AL and IOL power were not significantly different from ultrasound measurements. • Cech R, Utíkal T, Juhászová J. Srovnání optické a ultrazvukové biometrie a zhodnocení užívání obou metod v praxi [Comparison of optical and ultrasound biometry and assessment of using both methods in practice]. Cesk Slov Oftalmol. 2014 Feb;70(1):3-9. Czech. PMID: 24862369.
  • 30. • DISADVANTAGES OF OPTICAL BIOMETRY: • The cost of optical biometers is significantly more than ultrasound biometers. • Dense cataracts and corneal opacities may preclude accurate axial length determination optical biometry.Immersion ultrasound biometry is preferred in such cases, as it is more repeatable due to lack of variable corneal compression. • Many of these difficulties have, however, been overcome with the advent of swept-source optical coherence tomography (SS-OCT) technology. • Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191- 201
  • 31. • CORNEAL POWER: • Second most important factor. • 1.0 D error in corneal power results in 1.0D postrefractive error. • Methods of measurement: Keratometry Pentacam Galilei BAUSCH AND LOMB KERATOMETER PENTACAM GALILIE
  • 32. • POINTS TO BE CONSIDERED FOR KERATOMETRY: • Ideally be recalibrated after 20 cases for single observer. • Always do keratometry first before tonometry or any other corneal contact procedures. • Re measure corneal curvature of both eyes if: Corneal curvature is less than 40D or more than 47D. The difference in corneal cylinder is more than 1D between eyes. The corneal cylinder correlates poorly with the refraction cylinder Mastering IOL-Ashok garg-pg36
  • 33. • EFFECTIVE LENS POSITION: • Historically referred as the anterior chamber depth(ACD). • ACD was found to be inaccurate for “in the bag” IOL. • Denotes the distance that the principal plane of IOL will sit behind the cornea. • 1mm error in ELP leads to-0.25-0.5D of error. • Factors influencing effective IOL position: Anatomical factors:Axial length,steepness of cornea,limbal white to white measurement,preoperative anterior chamber depth and lens thickness. IOL-related factors:Shape,length,flexibility,anterior angulation,material of haptic of IOL Surgeon-related factors:Surgeon’s individual surgical technique Bag to sulcus shift:PCR,Loss of anterior capsule integrity-IOL needs to be placed in ciliary sulcus-needs deduction (0.50-0.75D) from calculated IOL power .
  • 34. • Biometry for IOL-Power calculation-AAO
  • 35. • Using gaussian reduction equation,formula for calculating IOL power can be expressed in accordance with refractive index- • P-Power of target IOL(In D) • K-Average dioptric power of central cornea(In D) • AL-Axial length(in mm) • C-ELP(in mm) • nv-refractive index of vitreous • Na-refractive index of aqueous Biometry for IOL-Power calculation-AAO
  • 36. FORMULAE • Based upon their derivation,the various IOL power formulae have been grouped into- • THEORETICAL FORMULAE:Based on mathematical principles revolving around ‘schematic eye’. • REGRESSION FORMULAE:Arrived by looking at postoperative outcomes retrospectively. • Artificial intelligence-based formulae: These formulae do use a form of regression, but they rely on huge databases and employ artificial intelligence- based complex statistical models to arrive identify relationships between variables. • Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191- 201
  • 37. • Vergence formulae:Vergence formulae arrive at an IOL power using Gaussian optics. • Gaussian optics makes an assumption that image vergence = object vergence + lens vergence. • Most modern-day IOL power calculation formulae are based on the following equation formulated by Fyodorov. • IOL power = (1336/[AL-ELP]) – (1336/[1336/{1000/([1000/ DPostRx] – V) + K} – ELP]) • K-Net corneal power ELP-Effective lens position V-Vertex distance • AL-Axial length DPostRx-Desired refraction • Vergence formulae are subclassified into 2-variable,3-variable,5-variable and 7-variable. Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-201
  • 38. • Ray tracing: • Ray tracing is a strategy to calculate the path that a light ray will travel when passed through an optical system. • Gaussian optics assumes presence of a single optical axis and all rays of interest make only very small angles to this. • 3rd Generation vergence formulae assume that all components of optical system are thin lenses • To describe the IOL position in this method, the radii of curvature of both anterior and posterior corneal surfaces, their asphericity, central IOL thickness, and refractive index of refraction are all used. • Gaussian optics formulae rely on postoperative refraction to back calculate ELP-less accurate due to errors in postoperative refraction and incorrect IOL power labelling. • The final IOL position is a true geometrical position described by the distance between the apex of the posterior surface of cornea and the apex of the anterior surface of the IOL (postoperative ACD). • Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191- 201
  • 39. • Advantages: • Postoperative ACD can be measured directly using PCI and compared with preoperative estimated ACD. • Predict IOL power more accurately in abnormally short,long eyes and in eyes that have undergone corneal refractive surgery. • Examples-Okulix,Olsen formula and PhacoOptics • Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191- 201
  • 40. CLASSIFICATION OF FORMULAE BASED ON GENERATIONS 1ST GENERATION 2ND GENERATION 3RD GENERATION 4TH GENERATION 5TH GENERATION THEORETICAL FORMULAE REGRESSI ON FORMULA E THEORETICA L FORMULAE REGRESSION FORMULAE BINKHORST SRK-1 MODIFIED BINKHORST SRK-2 HOLLADAY-1 HOLLADAY-2 HAIGIS COLENBRAND ER-HOFFER SRK-T OLSEN GILL’S HOFFER’S Q CLAYMAN’S FYODOROV
  • 41.
  • 42. • 1ST GENERATION FORMULAE: • 1-THEORETICAL FORMULAE: • Based on 3 variables-AL of eye K-reading Estimated postoperative ACD • A-BINKHORST FORMULA: P= 𝟏𝟑𝟑𝟔(𝟒𝒓−𝒂) (𝒂−𝒅)(𝟒𝒓−𝒅) • P-IOL power in diopters • r-corneal radius in millimeters(average) • a-AL in mm • d-assumed postoperative ACD plus corneal thickness • Disorders of lens and cataract surgery-A K Khurana-pg 112
  • 43. • B-COLENBRANDER-HOFFER FORMULA: • P= 1336 𝑎−𝑑−0.05 - 1336 1336 𝐾 −𝑑−0.05 • K-average keratometry in diopters • GILL’S FORMULA: • P=129.40+(-108 x K)+(-2.79 x AL)+(0.26 x LCL)+(0.38 x Ref) • P-desired IOL power • K-refractive power of cornea in diopters • AL-axial length in mm • LCL-distance of apex of anterior corneal surface to the apex of IOL in mm • Ref-desired postoperative refraction • Disorders of lens and cataract surgery-A K Khurana-pg 113
  • 44. • CLAYMAN’S FORMULA: • Assume- • Emmetropizing iol=18D • Emmetropic AL=24mm • Emmetropic average keratometry value-42 D • 1mm change in AL=3D of IOL power • 1D in keratometry=1D of IOL power • FYODOROV FORMULA:P = 1336−𝐿𝐾 𝐿−𝐶 − 𝐶𝐾 1336 • P-implant power for emmetropia • L-axial length(AL)in mm • K-corneal curvature in diopters • C-estimated postoperative anterior chamber depth(ACD) • Disorders of lens and cataract surgery-A K Khurana-pg 113
  • 45. • ALGEBRAIC TRANSFORMATION OF THEORETICAL FORMULAE- • P= 𝑁 𝐿−𝐶 - 𝑁𝐾 𝑁−𝐾𝐶 • P-Implant power for emmetropia • N-aqueous and vitreous RI • C-estimated postoperative ACD in millimeters • L-AL in mm • K-corneal curvature in diopters • DRAWBACKS: • Short eyes-large emmetropic value,Long eyes-too small value • Too cumbersome to apply • Guess of ACD is required • Were developed when iris-supported lenses were used • Based on theoretical simplistic assumption about optics of eye • Disorders of lens and cataract surgery-A K Khurana-pg 114
  • 46. • 2-REGRESSION FORMULAE: • Based on actual regression analysis of actual postoperative results of implant power as a function of variables of corneal power and AL. • SRK-1 FORMULA: • Introduced by Sanders,Retzlaff and Kraff • Based on regression analysis,taking into account the retrospective analysis of postoperative refractions. • Postoperative ACD was not included,replaced with A constant which was unique for different type of IOL,determined by the manufacturer depending upon material,position of eye,optic and haptic design. • Disorders of lens and cataract surgery-A K Khurana-pg 114
  • 47. • FORMULA: • P=A-2.5L-0.9K P-IOL power A-constant specific for each lens L-AL in mm K-avg keratometry in diopters • Performs well for eyes with AL between 22.0 and 24.5mm • Short eyes-too small value,Long eyes-too large value. • Disorders of lens and cataract surgery-A K Khurana-pg 114
  • 48. • SECOND-GENERATION FORMULAE: • 1-THEORETICAL FORMULAE: • Modified Binkhorst formulae:1981-Used AL as scaling factor for effective lens position and presented a formula to better predict ACD. • 2-REGRESSION FORMULAE: • SRK-2 FORMULA: P=A-2.5L-0.9K • L<20mm A+3.0 • L is 20.00-20.99mm A+2.0 • L is 21.00-21.99mm A+1.0 • L is 22.0-24.5mm A • L>24.5mm A-0.5 • Disorders of lens and cataract surgery-A K Khurana-pg 114
  • 49. • Modified SRK-2 Formula: • Based on AL,A-Constant is modified as: • If L<20mm A+1.5 • L is 20-21mm A+1.0 • L is 21-22mm A+0.5 • L is 22.0-24.5mm A • L is 24.5-26.0mm A-1.0 • L is >26mm A-1.5 • Disorders of lens and cataract surgery-A K Khurana-pg 114
  • 50. • 3rd GENERATION FORMULAE:Requires 2 variables-K and AL • Holladay-1 formula: • Introduced in 1988 • 2 variable predictor(AL and Keratometry)could significantly improve the predictor of effective lens position,particularly in unusual eyes. • Formula was proposed based on geometric relationship of the anterior segment. • Had better outcome for eyes between 22.00mm -26.00mm compared to other 3rd generation. • SRK-T formula: • Non-linear theoretical optical formula,empirically optimised for postoperative ACD,retinal thickness and corneal refractive index. • Combines advantages of both theoretical and empirical analysis. • Better outcomes for eyes>26.0 mm compared to other 3rd generation formulae • Disorders of lens and cataract surgery-A K Khurana-pg 115
  • 51. • Hoffer-Q formula: • Variables required-K and AL • Better outcome for eyes <22.00mm,compared to other 3rd generation • Disorders of lens and cataract surgery-A K Khurana-pg 115
  • 52. • 4TH GENERATION FORMULAE: • Holladay-2 formula: • 1993-Dr.Holladay led a worldwide study involving 34 cataract surgeons to determine which of the 7 variables were relevant as predictors of ELP. • Led to concept of 9 types of eyes,not 3(short,medium and long) • 7 Variables required- 1-K-Average K reading in diopters. 2-AL-Measured ultrasonic axial length in mm 3-ACD-Distance from corneal 4-LT-Lens thickness 5-Horizontal WTW-White to white corneal diameter 6-Age 7-Preoperative refraction • Allows calculation of IOL power in many different types of eyes but also honing of individual result by personalising A constant • Available as a part of Holladay IOL consultant/surgical outcomes assessment programme
  • 53.
  • 54. • Olsen formula: • Variables required- K AL ACD LT Horizontal WTW • Improved performance over 3rd generation formulae for eyes with AL 20.00- 26.00mm • Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191- 201
  • 55. • 5TH GENERATION FORMULAE: • Haigis: • Variables required K AL ACD • Very good outcomes for eyes across the AL range and best reported outcomes for eyes >28.00mm.For best results,constants need to be optimised requiring data from atleast 500 eyes. • Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191- 201
  • 56.
  • 57. • MODERN IOL FORMULAE: • LADAS SUPER FORMULA: • Devloped by Dr John G.Ladas,Albert Jun,Aazim Siddiqui,Uday Devgan. • Points of agreement and disagreement between Hoffer Q,Holladay 1,Holladay 1 with Wang-Koch adjustment,Haigis and SRK-T formulae for various ranges of axial length were identified. • Ideal portions from four of five formulae were used to create super surface. • Dynamic formula-constantly being updated using inputs from surgeons worldwide. • Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191- 201
  • 58. • BARRETT UNIVERSAL 2: • Devloped by Graham D Barrett. • Based on theoretical model eye in which ACD is related to axial length and keratometry. • Can be unmodified in eyes of all axial lengths and for all IOL designs-universal. • 5 Variables-AL,Keratometry,ACD,LT,Horizontal WTW corneal diameter to calculate ELP along with A constant and desired postoperative refraction. • Predicts IOL power accurately in eyes with extremely long axial lengths. • Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191- 201
  • 59. • HILL-RADIAL BASIS FUNCTION CALCULATOR: • Devloped by Warren E Hill with RBF calculator physician team. • Complex mathematical model-artificial neural network. • Has a property of adaptive learning. • Dynamic formula-becomes more accurate in its prediction as more data is fed into it. • Optimised for Haag streit Lenstar in combination with Alcon SN60WF biconvex IOL for powers from +30.00D to +6.00D and Alcon MA60MA meniscus design IOL powers from +5.00D to -5.00D. • Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191- 201
  • 60. • OLSEN FORMULA: • Devloped by Dr Thomas Olsen. • Paraxial and ray tracing optics. • Regression method used to determine ACD.ACD and LT used to determine ELP. • IOL position is described as a fraction of the capsular bag size using a new constant called the C constant. • Variables used-ACD,AL,LT and keratometry. • power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191-201
  • 61. • PHACOPTICS: • Ray tracing. • Uses concept of Olsen’s C constant. • WANG-KOCH MODIFICATION: • Traditional IOL formulae for abnormally long eyes-postoperative residual hyperopia. • Modification for AL more than 25.2mm to avoid refractive surprise. • Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191- 201
  • 62. • FULL MONTE INTRAOCULAR LENS: • Uses a method of analysis called the Markov Chain Monte Carlo process. • A probability instead of a single emmetropic power is displayed,surgeon can choose appropriate IOL power. • Evolves continuously optimizing itself as new data is received. • Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191- 201
  • 63. • TORIC CALCULATORS: • Correction of corneal astigmatism. • Components-1)Accurate K 2)Reliable method to calculate power of toric IOL. • Ideal characteristics: • It is comprehensive, that is, it can be used for preoperative planning, as well as for refractive surprises • Software should preferably be a part of the biometric device • It should be applicable to all IOL types • There should be dynamic display of variables (e.g., dynamic adjustment of astigmatic effects of phaco incision) • It should provide an alert for “axis-flip” • It should take into account posterior corneal astigmatism (PCA) • Nazm N, Chakrabarti A. Update on optical biometry and intraocular lens power calculation. TNOA J Ophthalmic Sci Res 2017;55:196-210
  • 64. • TORIC IOL CALCULATORS: • ASSORT TORIC CALCULATOR • BARRETT TORIC CALCULATOR • VERION • CALLISTO EYE • ASTIGMATISMFIX.COM • Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191- 201
  • 65. Nazm N, Chakrabarti A. Update on optical biometry and intraocular lens power calculation. TNOA J Ophthalmic Sci Res
  • 66. • FUTURE OF INTRAOCULAR LENS POWER CALCULATION: • UniversIOL CALCULATOR: • Designed by Dr.Samir Sayegh. • Web based • Combines all high quality 3rd and 4th generation formulae with a toric IOL calculator. • Instead of proposing a new IOL power calculation formula, the surgeon is provided with a choice of IOL power calculation formulae or combinations of formulae, a choice of computational methods, and a choice of IOL ranking criteria. • Tells how much the formula differ from each other,gives surgeon an idea how close she/he will be to the target. • Also contains all IOLs made so that an appropriate power as well as IOL can be chosen. • Nazm N, Chakrabarti A. Update on optical biometry and intraocular lens power calculation. TNOA J Ophthalmic Sci Res 2017;55:196-210
  • 67. • OKULIX: • Ray tracing formula. • Radius of curvature of anterior and posterior corneal surfaces need to be measured using a corneal topographer. • IOL is described by radii of curvature of its anterior and posterior surfaces,thickness and refractive index. • Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191- 201
  • 68. CLINICAL VARIABLES PERSONALISATION: • Introduced by Retzlaff. • Used A constant to refine the formula based on surgeon’s past experience and data. • Parameters required: Axial length(Pre-op) Corneal Power(Pre-op) IOL power Postoperative refractive error • Same lens style required by one manufacturer implanted by one surgeon. • Same biometry and technician. • Eyes with postoperative surprises or acuity worse than 20/40 not included- poor accuracy. • Mastering Intraocular Lenses-Ashok garg-pg no 42
  • 69. • Involves backsolving for the exact IOL position that would produce the resultant refractive error with AL and K. • All the “ideal”IOL positions are averaged to arrive at personalised value for use in future. • Can be easily done using Hoffer Programs or Holladay IOL Consultant computer programs. • Mastering Intraocular Lenses-Ashok garg-pg no 42
  • 70. • BIOMETRY IN SPECIAL CONDITIONS: • APHAKIA: • Required for secondary IOL implantation. • In aphakes,sound travels at speed of 1532m/s. • A Scan-2 lens spikes are replaced by anterior vitreous phase and posterior lens capsule. • Immersion technique of biometry-method of choice. • Present biometers-aphakic mode is available • Mastering Intraocular Lenses-Ashok garg-pg no 50
  • 71. • Mackool et al: • IOL power (D) = Aphakic refraction × 1.75 • This method was found to be more accurate when anterior chamber IOLs were implanted. • Ianchulev and Leccisotti: • IOL power (D) =0.07x 2 + 1.27x + 1.22, where x = aphakic refraction • This method was found to be more accurate for IOL placed posteriorly. • Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191- 201
  • 72. • PSEUDOPHAKIA: • IOL exchange. • High spike at lens followed by artificial chain of reduplication of echoes-can be misinterpreted as spikes from retina. • Lower gain-eliminates artificial spikes and increases retinal spikes. • Sound velocity depends upon type of IOL. • Modern biometers-options are available for phakic, pseudophakic or aphakic mode. Disorders of lens and cataract surgery-A K Khurana-pg 116
  • 73. • PIGGYBACK IOL FOR PSEUDOPHAKIA: • Indicated in patients with significant residual refractive error following the primary IOL implantation. • Surgically easier and more predictable optically rather than IOL exchange. • Formulae: Holladay’s refractive formula Gill’s nomogram-calculated from residence spherical based on axial length
  • 74. • VITRECTOMIZED EYE: • RI of silicon oil is much less than that of vitreous,acts as negative lens. • Silicon oil-sound attenuation-986m/s. • Increase the system gain and adjust sound velocity accordingly. • Result obtained is Apparent axial length(AAL) which has to be converted to True axial length(TAL) • TAL=1133/1550 AAL
  • 75. • PEDIATRIC BIOMETRY: • Dahan et al’ simplified approach based on axial length AXIAL LENGTH IOL POWER 17mm 28D 18mm 27D 19mm 26D 20mm 24D 21mm 24D 22mm 23D 23mm 23D >24mm Axial length-1D
  • 76. • Immersion A-scan preferred. • IOL power-SRK/T and Holladay 2 are preferred. • Undercorrection- According to Dahan and Drusedau undercorrection of 20% in children <2 years 10% in children between 2 and 8 years • Prost suggested 20% undercorrection between 1 and 2 years of age, 15% undercorrection between 2 and 4 years 10% between 4 and 8 years of age. AAO-management of pediatric cataract.
  • 77. • POST KERATOREFRACTIVE SURGERY: • Standard methods of IOL power calculation-Inadequate. • Factors altered in corneal refractive surgeries: Anterior corneal curvature Ratio of anterior and posterior corneal curvature Corneal asphericity • Modern formulae use the above factors to calculate IOL power-adds to error. • Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191- 201
  • 78. • HAIGIS-L FORMULA: • The corneal radius of curvature (r corr) is modified as follows: • r - corr = 331.5 −5.1625 𝑥 𝑟 𝑚𝑒𝑎𝑠+82.2603−0.35 • where r meas is the measured corneal radius of curvature. • SHAMMAS FORMULA: • Calculated corneal power, K = 1.14 ×TK −6.8 • where TK is the post-LASIK corneal topography central K. • Mohan P, Chakrabarti A. Intraocular lens power calculation in 2019: The cutting edge. Kerala J Ophthalmol 2019;31:191- 201
  • 79. • MASKET METHOD: • Regression formula-Samuel musket. • AL>23mm-Holladay 1. • AL<23mm-Hoffer Q. • Adjustment of IOL power- (LSE × −0.326) +0.101 = Post-LASIK IOL power adjustment Where LSE is the laser corrected spherical equivalent adjusted for vertex distance • MODIFIED MASKET METHOD: • (LSE × −0.4385) +0.0295 = Post-LASIK IOL power adjustment • Others: • ASCRS postrefractive calculator • Hoffer–Savini LASIK intraocular lens power tool • Barrett true-K formula • The McCarthy post refractive intraocular lens calculator
  • 80. PROBLEMS AND ERRORS-IOL POWER SURPRISE • AXIAL LENGTH:Most common source of error • Axial length >25mm,higher incidence of staphylomas. • Artificial shortening due to contact applanation ultrasound. • Technician inexperience. • Wrong average ultrasound velocity. • Poor IOL master readings-Dense cataracts. • Silicone filled eye.
  • 81. • CORNEAL POWER:Second most common cause. • Eyes with previous corneal refractive surgery. • Wrong calibration. • Not setting index of refraction(IR) in IOLMaster-leads to error in Hoffer Q formula. • Contact lens wear. • Corneal scarring. • EFFECTIVE POSITION OF IOL:Least effective factor • IOL in deeper or shallower position than predicted by the formula • ACD constants not personalised to individual IOL style,surgeon and clinic.
  • 82. • FORMULAS: • Regression formulas(SRK 1) for eyes outside the normal AL range of 22-24.5mm. • MISCELLANEOUS: • Manfacturer labelling error. • Insertion of wrong IOL.
  • 83. • PREVENTION OF ERRORS: • Immersion A-scan/IOLMater to measure AL. • Suspect staphylomas in eyes>25mm-Use IOLMaster. • Well-trained experienced technician. • Regularly calibrate manual keratometers. • Evaluate Biometer scans for reliability. • Keep contact lens out for 2 weeks prior to keratometry • Silicone oil-optical biometers/AL x 0.71 • AL<22mm-Hoffer Q • AL 22-26mm-Holladay 1 • AL >26mm-SRK/T • Avoid using SRK regression formulas(SRK 1 and 2)
  • 84. • Personalise ELP factors in the formulas. • Personally select IOL power for individual patient. • Set IR to 1.3375 in set-up screen of IOLMaster. • Post-refractive surgery- Shammas method Masket method Haigis formula • Consider delaying IOL implantation until cornea has healed after a penetrating keratoplasty rather than performing a “triple procedure”.