More Related Content
Similar to Bioanalytical validation.pptx (20)
Bioanalytical validation.pptx
- 4. 4
©2023 Waters Corporation COMPANY CONFIDENTIAL
common techniques for sample preparation
Protein precipitation
Solid phase extraction
Liquid / liquid extraction
- 5. 5
©2023 Waters Corporation COMPANY CONFIDENTIAL
Drug reference standard
• A drug reference standard or pharmaceutical reference standard is a highly characterized material
suitable to test the identity, strength, quality and purity of substances for pharmaceutical use and
medicinal products.
• Generally available from USP IP BP EP
• Having purity more than 99.5%
Drug Working standard
• A drug substance of established quality and purity as shown by comparison to the reference
standards.
• Available from commercial source like Clearsynth.
internal standard
• An internal standard is a substance added in a known amount to standards, samples, and blanks during
an analysis.
• Added to the analyte sample to check the variability due to analyte loss during the sample processing.
- 6. 6
©2023 Waters Corporation COMPANY CONFIDENTIAL
internal standard
internal standard
Stable isotope labeled IS
Structural related compound
(non labeled)
Metronidazole d4
Metronidazole Ornidazole
Advantages:
Better accuracy and prison
Disadvantages:
More expensive
- 7. 7
©2023 Waters Corporation COMPANY CONFIDENTIAL
Calibration curve standards (CC Std.)
Relation between instrumental response and the drug standard within intended quantification range.
Minimum 6 to 8 standards.
Blank: processed matrix (plasma or serum) without analyte.
Zero: processed matrix (plasma or serum) with IS only.
How to decide the range
Cmax
Lower limit
LLOQ or STD-1
Upper limit
ULOQ or STD-8
At least 5% of Cmax or less
At least 1.5 to 2.0 times of Cmax
For example, If Cmax of drug is 50 ng/mL
LLOQ: 1 to 2.5 ng/mL (1.0 ng/mL)
ULOQ: 100 ng/mL
Range of the study is 1.0 to 100 ng/mL
- 8. 8
©2023 Waters Corporation COMPANY CONFIDENTIAL
QUALITY CONTROL STANDEREDS
LLOQQC LQC MQC HQC
Quality control standards
Half of LLOQ
0.5 ng/mL
Three times of LLOQ
3.0 ng/mL
Around 30 to 50% of ULOQ
50.0 ng/mL
At least 75% of ULOQ
85.0 ng/mL
- 9. 9
©2023 Waters Corporation COMPANY CONFIDENTIAL
CC & QC’S
STD-1 STD-2 STD-3 STD-4 STD-5 STD-6 STD-7 STD-8
CC & QC’S
LLOQ
1.0 ng/mL
LLOQC
0.5 ng/mL
2.0 ng/mL 4.0 ng/mL 10.0 ng/mL 20.0 ng/mL 30.0 ng/mL 70.0 ng/mL 100.0 ng/mL
LQC
3.0 ng/mL
MQC
50.0 ng/mL
HQC
85.0 ng/mL
ULOQ
Acceptance criteria for cc STD’s
At least 75% CC STD’s should pass above criteria.
Area in Blank at RT of analyte should be <20% of area in LLOQ.
Area in Blank at RT of IS should be less than <5% of area in IS in
LLOQ.
Acceptance criteria for QC STD’s
At least 67% QC should be ± 15% of nominal
value.
≥ 50% of QC at each level should be 15% of
nominal value.
CC and QC should be prepared from the
separate stock solution.
- 10. 10
©2023 Waters Corporation COMPANY CONFIDENTIAL
Bioanalytical method validation parameters
Selectivity
Specificity
Carryover
Prison and Accuracy
Robustness Ruggedness
Matrix effect
Recovery
Dilution integrity
Reinjection reproducibility
Stability
- 11. 11
©2023 Waters Corporation COMPANY CONFIDENTIAL
Selectivity
Is the extent to which the method can determine the analyte of interest and IS in the matrix without any interference
from the matrix component.
Purpose
To prove that our method can selectively determine our analyte of interest and IS only.
To prove that our method is selective even if we use different plasma lots.
Procedure
Six different lots of plasma ( Normal )
One Hemolysed and one Lipolysed plasma lot
Prepare Blank and (1.0 ng/mL) LLOQ samples in each plasma lots.
Acceptance criteria
Area in all blank samples at the RT of analyte <20% area of LLOQ & at the RT of IS <5% area of LLOQ.
- 12. 12
©2023 Waters Corporation COMPANY CONFIDENTIAL
Specificity
Ability of method to assess the analyte of interest in presence of other components ( eg.impurities,matrix components &
other comedication.
Purpose
To prove that our method is specific to identify the analyte of interest in the presence of other plasma components or
any comedication is present without any interference.
Commonly used medication are paracetamol, ibuprofen, chlorpheniramine, Nicotine & caffeine.
Procedure
Spike plasma lots six normal One Hemolysed and one Lipolysed plasma lot Cmax of comedication
Prepare six LLOQ samples in each plasma lots
Prepare six LQC samples and spike with Cmax of comedication
Acceptance criteria
Area in all blank samples at the RT of analyte <20% area of LLOQ & at the RT of IS <5% area of LLOQ.
Accuracy of LQC should be within ±15% nominal value.
- 13. 13
©2023 Waters Corporation COMPANY CONFIDENTIAL
Carry over
Is the appearance of analyte signal in Blank after the analysis of sample with high analyte concentration.
Reason
Inappropriate sample dilution or diluent.
Inappropriate mobile phase or mobile phase Ph.
Inappropriate washing duration.
Remedy
Extend the washing duration.
Change the mobile phase or diluent.
Acceptance criteria
Area in all blank samples at the RT of analyte <20% area of LLOQ & at the RT of IS <5% area of LLOQ.
Accuracy of LQC should be within ±15% nominal value.
Procedure
Inject the processed blank sample after higher concentration QC repeat the procedure for six times.
- 14. 14
©2023 Waters Corporation COMPANY CONFIDENTIAL
Precision and Accuracy
Freshly prepared CC and QC standard’s
At least two runs are required in two days
Each batch of PA contains : 1 Blank, 1 Zero, CC STD’S (1-8) QC (LLOQQC,LQC,MQC,HQC)
Acceptance criteria for CC
All CC STD (2-8) should be ±15% nominal value.
For STD-1 (LLOQ) should be ±20% nominal value.
At least 75% of and minimum of 6 STD. should meet above criteria.
Acceptance criteria for QC
Accuracy : The mean concentration of each level of QC ( L,M & H) should be within ±15% nominal value.
For LLOQQC it should be within ± 20% nominal value.
At least 67% of overall QC & 50 % at each QC level should meet acceptance criteria.
- 15. 15
©2023 Waters Corporation COMPANY CONFIDENTIAL
Robustness & Ruggedness
Ability of analytical method remains unaffected by the small variations in the method parameters like
column,analyst,instrument.
Evaluated by changing at least one of the factors during one of the P&A batch.
Acceptance criteria
As per P&A
Matrix Effect
Effect of the matrix on the method.
Matrix effects can be observed either as a loss in response (ion suppression) or as an increase in response (ion enhancement)
Can be minimized by selective extraction procedure changing extraction solvents and buffers.
Matrix Factor (MF) = Peak response in presence of matrix ions / Peak response in absence of matrix ions
- 16. 16
©2023 Waters Corporation COMPANY CONFIDENTIAL
Procedure to check matrix effect
Eight different matrix lot
Six normal one Haemolysed one lipemic
Set 1 set 2
Standard solvent without plasma At LQC,MQC,HQC Spiked post extracted sample with matrix
Matrix factor =
𝑃𝑒𝑎𝑘 𝑎𝑟𝑒𝑎 𝑖𝑛 𝑝𝑟𝑒𝑠𝑒𝑛𝑠𝑒 𝑜𝑓 𝑚𝑎𝑡𝑟𝑖𝑥
𝑝𝑒𝑎𝑘 𝑎𝑟𝑒𝑎 𝑖𝑛 𝑎𝑏𝑠𝑒𝑛𝑐𝑒 𝑜𝑓 𝑚𝑎𝑡𝑟𝑖𝑥
If matrix effect is =>1 ion enhancement
If matrix effect is =<1 ion Suppression
If matrix factor is =1 No matrix effect
- 17. 17
©2023 Waters Corporation COMPANY CONFIDENTIAL
Recovery
Extraction efficacy of an analytical process reported as % of known amount of analyte recovered through sample
extraction.
Recovery need not to be 100% always but it should be consistent and reproducible for analyte and IS
Determined at three levels LQC,MQC&HQC
6 Replicates of Extracted LQC,MQC&HQC
6 Replicates of Post spiked LQC,MQC&HQC
Procedure
Acceptance criteria for QC
Precision (%RSD) of overall % Recovery should be within 15% and % Recovery should be ≥ at individual quality control level
(i.e. LQC, MQC, HQC).
%Recovary =
𝑎𝑟𝑒𝑎 𝑜𝑓 𝑒𝑥𝑡𝑟𝑎𝑐𝑡𝑒𝑑 𝑠𝑎𝑚𝑝𝑙𝑒
𝑎𝑟𝑒𝑎 𝑖𝑛 𝑝𝑜𝑠𝑡 𝑠𝑝𝑖𝑘𝑒𝑑 𝑠𝑎𝑚𝑝𝑙𝑒
x 100
- 18. 18
©2023 Waters Corporation COMPANY CONFIDENTIAL
Dilution Integrity
This experiment is carried out to assess the ability of method to quantify analyte and yield accurate and precise result
even after sample dilution
Procedure
Prepare the six QC (Dilution QC) with concentration higher than ULOQ of the original concentration by at least two times;
Eg: If ULOQ is 100ng/mL & cc range is 0.5 to 100ng/mL Dilution QC should be 200ng/mL
In order to fit in the range of our CC 200ng/mL will be diluted for 4 times to get 50ng/mL
Six replicates of 50ng/mL will be injected
Acceptance criteria
% Accuracy & Precision should (%CV) should be within 15%.
- 19. 19
©2023 Waters Corporation COMPANY CONFIDENTIAL
Reinjection reproducibility
To access the ability of method to quantify analyte accurately and precisely upon reinjection.
If any sample is reinjected, we must get same result.
If any sample results goes wrong or any other error of LCMS or HPLC etc. then under those circumstances, then that
injection will be reinjected.
Procedure
Inject previously accepted P&A batch
Acceptance criteria
As per P&A
- 20. 20
©2023 Waters Corporation COMPANY CONFIDENTIAL
Stability
The chemical stability of analyte in a given matrix under specific condition in given time intervals is assessed in several
ways.
Freeze and thaw stability
During the freeze/thaw stability evaluation the freezing and thawing of stability should mimic the intended sample handling
conditions to be used during sample analysis. Stability should be accessed for minimum of 3 freeze thaw cycle.
Bench top Stability
Bench top stability experiments should be designed and conducted to cover the laboratory handling conditions that are
expected for the study samples.
Long term stability
Storage time should exceed the time between the date of first sample collection and the date of last sample analysis.