Appealing to Users
Developing Usable & Desirable Devices




Presenter: Julian Dixon
Date: 20th September 2012


Confidential © Team Consulting 2011: World PFS Summit 2012, slide 1   www.team-consulting.com
Overview




• What is human factors (HF)?
• Why so much current interest in HF?

• Best practice in patient-use, drug delivery device HF
     – what is different about patient-use, drug delivery device usability?
• Practitioners’ experiences – tips and pitfalls
• Realistic usability targets for self-injection devices

• Looking beyond usability - safety and effectiveness – towards desirability




  Confidential © Team Consulting 2011: World PFS Summit 2012, slide 2         www.team-consulting.com
What is Human Factors Engineering?




         “Understanding and optimizing how people interact with
                               technology
                            is the subject of
                   human factors engineering (HFE)
                                   and
                      usability engineering (UE).”




Ref: FDA Draft Guidance, June 2011: Applying Human Factors and Usability Engineering to Optimize Medical Device Design

Confidential © Team Consulting 2011: World PFS Summit 2012, slide 3                                                www.team-consulting.com
Why current interest in HF?


• medical devices pose use-related risks
• risks need to be identified & addressed during device development …
• … to ensure devices can be used safely and effectively by their intended
  users

• why now?

• in the past, as an industry, we didn’t pay sufficient attention to use-related
  risk or to its management through the application of human factors
  engineering
• events have recently focused regulatory attention on the importance of
  the topic



  Confidential © Team Consulting 2011: World PFS Summit 2012, slide 4        www.team-consulting.com
Confidential © Team Consulting 2011: World PFS Summit 2012, slide 5   www.team-consulting.com
Patient-use, drug delivery devices pose a unique HF challenge


• disposability = very low cost target

• patient use
     – potentially untrained users
     – used by anyone, anywhere
• limited functionality / limited number of distinct tasks
     – all use-steps are ‘essential’ or core
• tasks go beyond the device
     – task completion depends on some user behaviour the device cannot enforce
     – dependence on user
     – e.g. correct storage / pre-injection checks / disposal
• simple task structure / simple device interface
     – perhaps their apparent simplicity/familiarity may promote guessing?


  Confidential © Team Consulting 2011: World PFS Summit 2012, slide 6        www.team-consulting.com
Statements of HF Objective




             “… intended users of a medical device can safely and effectively
                      perform critical tasks for the intended uses
                          in the expected use environments.”

                       “… no patterns of use failure or difficulties
                          that could be eliminated or reduced
            through further modification of the design of the user interface.”

      “… residual risk is outweighed by the advantages offered by the device.”



Ref: FDA Draft Guidance, June 2011: Applying Human Factors and Usability Engineering to Optimize Medical Device Design

Confidential © Team Consulting 2011: World PFS Summit 2012, slide 7                                                www.team-consulting.com
User Interface Design                                  Optimized
                            (e.g. device + IFU + packaging)                            User Interface
                                                                      mitigate
                                                                                                           
       Device               Opportunities
      Concept               for Use Error
                                                        Critical                 Formative      HF Validation   HFE Summary
      Drug/                    Hazards &               Use Errors                 Studies           Study          Report
   Formulation              Clinical Impact
                                                                       verify
                                         Use-Related Risk Assessment                                       
                                                                                      Residual Risk
                                                (e.g. uFMEA)




Confidential © Team Consulting 2011: World PFS Summit 2012, slide 8                                                   www.team-consulting.com
User Interface Design                                  Optimized
                            (e.g. device + IFU + packaging)                            User Interface
                                                                      mitigate
                                                                                                           
       Device               Opportunities
      Concept               for Use Error
                                                        Critical                 Formative      HF Validation   HFE Summary
      Drug/                    Hazards &               Use Errors                 Studies           Study          Report
   Formulation              Clinical Impact
                                                                       verify
                                         Use-Related Risk Assessment                                       
                                                                                      Residual Risk
                                                (e.g. uFMEA)

                                                Intended Use
                                        (users, tasks, environments)




Confidential © Team Consulting 2011: World PFS Summit 2012, slide 9                                                   www.team-consulting.com
start early

                                User Interface Design                          Optimized
                           (e.g. device + IFU + packaging)                    User Interface

       Device               Opportunities
      Concept               for Use Error
                                                        Critical        Formative      HF Validation   HFE Summary
      Drug/                    Hazards &               Use Errors        Studies           Study          Report
   Formulation              Clinical Impact

                                         Use-Related Risk Assessment
                                                                             Residual Risk
                                                (e.g. uFMEA)

                                                Intended Use
                                        (users, tasks, environments)




Confidential © Team Consulting 2011: World PFS Summit 2012, slide 10                                         www.team-consulting.com
our understanding on day 1 was/is good

                                User Interface Design                         Optimized
                           (e.g. device + IFU + packaging)                   User Interface

       Device               Opportunities
      Concept               for Use Error
                                                        Critical       Formative      HF Validation   HFE Summary
      Drug/                    Hazards &               Use Errors       Studies           Study          Report
   Formulation              Clinical Impact

                                         Use-Related Risk Assessment
                                                                            Residual Risk
                                                (e.g. uFMEA)

                                                Intended Use
                                        (users, tasks, environments)




Confidential © Team Consulting 2011: World PFS Summit 2012, slide 11                                        www.team-consulting.com
be clear about who are not intended users

                                User Interface Design                         Optimized
                           (e.g. device + IFU + packaging)                   User Interface

       Device               Opportunities
      Concept               for Use Error
                                                        Critical       Formative      HF Validation   HFE Summary
      Drug/                    Hazards &               Use Errors       Studies           Study          Report
   Formulation              Clinical Impact

                                         Use-Related Risk Assessment
                                                                            Residual Risk
                                                (e.g. uFMEA)

                                                Intended Use
                                        (users, tasks, environments)




Confidential © Team Consulting 2011: World PFS Summit 2012, slide 12                                        www.team-consulting.com
an important
              instructions for use are part of the user interface

                                User Interface Design                         Optimized
                           (e.g. device + IFU + packaging)                   User Interface

       Device               Opportunities
      Concept               for Use Error
                                                        Critical       Formative      HF Validation   HFE Summary
      Drug/                    Hazards &               Use Errors       Studies           Study          Report
   Formulation              Clinical Impact

                                         Use-Related Risk Assessment
                                                                            Residual Risk
                                                (e.g. uFMEA)

                                                Intended Use
                                        (users, tasks, environments)




Confidential © Team Consulting 2011: World PFS Summit 2012, slide 13                                        www.team-consulting.com
analytical and empirical HF

                                User Interface Design                         Optimized
                           (e.g. device + IFU + packaging)                   User Interface

       Device               Opportunities
      Concept               for Use Error
                                                        Critical       Formative      HF Validation   HFE Summary
      Drug/                    Hazards &               Use Errors       Studies           Study          Report
   Formulation              Clinical Impact

                                         Use-Related Risk Assessment
                                                                            Residual Risk
                                                (e.g. uFMEA)

                                                Intended Use
                                        (users, tasks, environments)




Confidential © Team Consulting 2011: World PFS Summit 2012, slide 14                                        www.team-consulting.com
expect to iterate

                                User Interface Design                         Optimized
                           (e.g. device + IFU + packaging)                   User Interface

       Device               Opportunities
      Concept               for Use Error
                                                        Critical       Formative      HF Validation   HFE Summary
      Drug/                    Hazards &               Use Errors       Studies           Study          Report
   Formulation              Clinical Impact

                                         Use-Related Risk Assessment
                                                                            Residual Risk
                                                (e.g. uFMEA)

                                                Intended Use
                                        (users, tasks, environments)




Confidential © Team Consulting 2011: World PFS Summit 2012, slide 15                                        www.team-consulting.com
don’t leave root cause analysis until validation

                                User Interface Design                         Optimized
                           (e.g. device + IFU + packaging)                   User Interface

       Device               Opportunities
      Concept               for Use Error
                                                        Critical       Formative      HF Validation   HFE Summary
      Drug/                    Hazards &               Use Errors       Studies           Study          Report
   Formulation              Clinical Impact

                                         Use-Related Risk Assessment
                                                                            Residual Risk
                                                (e.g. uFMEA)

                                                Intended Use
                                        (users, tasks, environments)




Confidential © Team Consulting 2011: World PFS Summit 2012, slide 16                                        www.team-consulting.com
‘critical’ and ‘successful’ should be applied with care

                                User Interface Design                         Optimized
                           (e.g. device + IFU + packaging)                   User Interface

       Device               Opportunities
      Concept               for Use Error
                                                        Critical       Formative      HF Validation   HFE Summary
      Drug/                    Hazards &               Use Errors       Studies           Study          Report
   Formulation              Clinical Impact

                                         Use-Related Risk Assessment
                                                                            Residual Risk
                                                (e.g. uFMEA)

                                                Intended Use
                                        (users, tasks, environments)




Confidential © Team Consulting 2011: World PFS Summit 2012, slide 17                                        www.team-consulting.com
we should learn nothing important during validation

                                User Interface Design                         Optimized
                           (e.g. device + IFU + packaging)                   User Interface

       Device               Opportunities
      Concept               for Use Error
                                                        Critical       Formative      HF Validation   HFE Summary
      Drug/                    Hazards &               Use Errors       Studies           Study          Report
   Formulation              Clinical Impact

                                         Use-Related Risk Assessment
                                                                            Residual Risk
                                                (e.g. uFMEA)

                                                Intended Use
                                        (users, tasks, environments)




Confidential © Team Consulting 2011: World PFS Summit 2012, slide 18                                        www.team-consulting.com
don’t hope for zero

                                User Interface Design                         Optimized
                           (e.g. device + IFU + packaging)                   User Interface

       Device               Opportunities
      Concept               for Use Error
                                                        Critical       Formative      HF Validation   HFE Summary
      Drug/                    Hazards &               Use Errors       Studies           Study          Report
   Formulation              Clinical Impact

                                         Use-Related Risk Assessment
                                                                            Residual Risk
                                                (e.g. uFMEA)

                                                Intended Use
                                        (users, tasks, environments)




Confidential © Team Consulting 2011: World PFS Summit 2012, slide 19                                        www.team-consulting.com
usability acceptance criteria are risk acceptance criteria

                                User Interface Design                         Optimized
                           (e.g. device + IFU + packaging)                   User Interface

       Device               Opportunities
      Concept               for Use Error
                                                        Critical       Formative      HF Validation   HFE Summary
      Drug/                    Hazards &               Use Errors       Studies           Study          Report
   Formulation              Clinical Impact

                                         Use-Related Risk Assessment
                                                                            Residual Risk
                                                (e.g. uFMEA)

                                                Intended Use
                                        (users, tasks, environments)




Confidential © Team Consulting 2011: World PFS Summit 2012, slide 20                                        www.team-consulting.com
qualitative not quantitative

                                User Interface Design                         Optimized
                           (e.g. device + IFU + packaging)                   User Interface

       Device               Opportunities
      Concept               for Use Error
                                                        Critical       Formative      HF Validation   HFE Summary
      Drug/                    Hazards &               Use Errors       Studies           Study          Report
   Formulation              Clinical Impact

                                         Use-Related Risk Assessment
                                                                            Residual Risk
                                                (e.g. uFMEA)

                                                Intended Use
                                        (users, tasks, environments)




Confidential © Team Consulting 2011: World PFS Summit 2012, slide 21                                        www.team-consulting.com
demonstrate understanding

                                User Interface Design                         Optimized
                           (e.g. device + IFU + packaging)                   User Interface

       Device               Opportunities
      Concept               for Use Error
                                                        Critical       Formative      HF Validation   HFE Summary
      Drug/                    Hazards &               Use Errors       Studies           Study          Report
   Formulation              Clinical Impact

                                         Use-Related Risk Assessment
                                                                            Residual Risk
                                                (e.g. uFMEA)

                                                Intended Use
                                        (users, tasks, environments)




Confidential © Team Consulting 2011: World PFS Summit 2012, slide 22                                        www.team-consulting.com
What might be a realistic and sound target for self-injector usability?


• What might we agree we should be committed to designing out?
     – too easy to repeatedly overdose/no-dose inadvertently
     – too easy to damage device through ‘reasonable misuse’
     – procedure of use too complicated
             • hard to learn/hard to train
             • easy to get wrong even if paying attention
     – too easy to misinterpret instructions for use (IFU)
     – too difficult to follow instructions for use (IFU)
     – poor physical usability - some tasks too difficult for intended users
     – opportunities for critical harm to result from predictable negative transfer of prior
       experience with other injectors
     – …


• What usability issues might we recognise have justifiable residual risks associated
  with them?
     – requirement for user to attend to simple-to-follow IFU in order to acquire adequate
       technique …

  Confidential © Team Consulting 2011: World PFS Summit 2012, slide 23                         www.team-consulting.com
design research – going beyond usability
  Opportunity                                                          Certainty




                                                  Summative - validatory
                              Formative - confirmatory

          Generative - exploratory



   Projective - discovery



Confidential © Team Consulting 2011: World PFS Summit 2012, slide 24               www.team-consulting.com
narrative – tell a coherent story

                                User Interface Design                         Optimized
                           (e.g. device + IFU + packaging)                   User Interface

       Device               Opportunities
      Concept               for Use Error
                                                        Critical       Formative      HF Validation   HFE Summary
      Drug/                    Hazards &               Use Errors       Studies           Study          Report
   Formulation              Clinical Impact

                                         Use-Related Risk Assessment
                                                                            Residual Risk
                                                (e.g. uFMEA)

                                                Intended Use
                                        (users, tasks, environments)




Confidential © Team Consulting 2011: World PFS Summit 2012, slide 25                                        www.team-consulting.com
Confidential © Team Consulting 2011: World PFS Summit 2012, slide 26   www.team-consulting.com
For further information, please contact

  Julian Dixon
  Tel: +44 (0)1799 532 746
  Email: jrd@team-consulting.com




                                                                                   Team Consulting Ltd.
                                                                              Abbey Barns, Duxford Road
                                                                                    Ickleton, Cambridge
                                                                                           CB10 1SX, UK

                                                                                   Tel:+44 (0)1799 532 700
                                                                                  Fax:+44 (0)1799 532 701
                                                                        Email: info@team-consulting.com



Confidential © Team Consulting 2011: World PFS Summit 2012, slide 27   www.team-consulting.com

More Related Content

Viewers also liked

Thinking Human by Julian Dixon, PMPS Inhalation Technology Supplement
Thinking Human by Julian Dixon, PMPS Inhalation Technology SupplementThinking Human by Julian Dixon, PMPS Inhalation Technology Supplement
Thinking Human by Julian Dixon, PMPS Inhalation Technology SupplementTeam Consulting Ltd
 
Designing safety critical medical devices
Designing safety critical medical devicesDesigning safety critical medical devices
Designing safety critical medical devicesTeam Consulting Ltd
 
Are electronically enabled delivery devices (EEDDs) the future?
Are electronically enabled delivery devices (EEDDs) the future?Are electronically enabled delivery devices (EEDDs) the future?
Are electronically enabled delivery devices (EEDDs) the future?Team Consulting Ltd
 
Choosing the right device: the case for DPIs (DDL22)
Choosing the right device: the case for DPIs (DDL22)Choosing the right device: the case for DPIs (DDL22)
Choosing the right device: the case for DPIs (DDL22)Team Consulting Ltd
 
Parenteral Drug Delivery in the Future: A View of Developments, Implications ...
Parenteral Drug Delivery in the Future: A View of Developments, Implications ...Parenteral Drug Delivery in the Future: A View of Developments, Implications ...
Parenteral Drug Delivery in the Future: A View of Developments, Implications ...Team Consulting Ltd
 

Viewers also liked (6)

Beyond asthma and COPD
Beyond asthma and COPDBeyond asthma and COPD
Beyond asthma and COPD
 
Thinking Human by Julian Dixon, PMPS Inhalation Technology Supplement
Thinking Human by Julian Dixon, PMPS Inhalation Technology SupplementThinking Human by Julian Dixon, PMPS Inhalation Technology Supplement
Thinking Human by Julian Dixon, PMPS Inhalation Technology Supplement
 
Designing safety critical medical devices
Designing safety critical medical devicesDesigning safety critical medical devices
Designing safety critical medical devices
 
Are electronically enabled delivery devices (EEDDs) the future?
Are electronically enabled delivery devices (EEDDs) the future?Are electronically enabled delivery devices (EEDDs) the future?
Are electronically enabled delivery devices (EEDDs) the future?
 
Choosing the right device: the case for DPIs (DDL22)
Choosing the right device: the case for DPIs (DDL22)Choosing the right device: the case for DPIs (DDL22)
Choosing the right device: the case for DPIs (DDL22)
 
Parenteral Drug Delivery in the Future: A View of Developments, Implications ...
Parenteral Drug Delivery in the Future: A View of Developments, Implications ...Parenteral Drug Delivery in the Future: A View of Developments, Implications ...
Parenteral Drug Delivery in the Future: A View of Developments, Implications ...
 

Similar to Appealing to users: developing usable & desirable devices

There is No Such Thing as User Error
There is No Such Thing as User ErrorThere is No Such Thing as User Error
There is No Such Thing as User ErrorGfK User Centric
 
mHealth Israel_Human Factors for MedTech Ergonomics and Usability_Rebecca Moses
mHealth Israel_Human Factors for MedTech Ergonomics and Usability_Rebecca MosesmHealth Israel_Human Factors for MedTech Ergonomics and Usability_Rebecca Moses
mHealth Israel_Human Factors for MedTech Ergonomics and Usability_Rebecca MosesLevi Shapiro
 
Human Factors Engineering for Medical Devices
Human Factors Engineering for Medical DevicesHuman Factors Engineering for Medical Devices
Human Factors Engineering for Medical DevicesEMMAIntl
 
Essay On Reliability Of Visualization Tools
Essay On Reliability Of Visualization ToolsEssay On Reliability Of Visualization Tools
Essay On Reliability Of Visualization ToolsNatasha Barnett
 
Integrating Human Factors Engineering into the Product Development Lifecycle_...
Integrating Human Factors Engineering into the Product Development Lifecycle_...Integrating Human Factors Engineering into the Product Development Lifecycle_...
Integrating Human Factors Engineering into the Product Development Lifecycle_...BasemAbdo4
 
How to design Usable EMR/EHR Application
How to design Usable EMR/EHR ApplicationHow to design Usable EMR/EHR Application
How to design Usable EMR/EHR ApplicationSubhasish Karmakar
 
Experimental Study Using Functional Size Measurement in Building Estimation M...
Experimental Study Using Functional Size Measurement in Building Estimation M...Experimental Study Using Functional Size Measurement in Building Estimation M...
Experimental Study Using Functional Size Measurement in Building Estimation M...Luigi Buglione
 
Experimental Study Using Functional Size Measurement in Building Estimation M...
Experimental Study Using Functional Size Measurement in Building Estimation M...Experimental Study Using Functional Size Measurement in Building Estimation M...
Experimental Study Using Functional Size Measurement in Building Estimation M...Luigi Buglione
 
Usability mHIseminar.Miller-Jacobs
Usability mHIseminar.Miller-JacobsUsability mHIseminar.Miller-Jacobs
Usability mHIseminar.Miller-JacobsmHealth Initiative
 
2010 10 07 Octane Ofi Presentation
2010 10 07 Octane Ofi Presentation2010 10 07 Octane Ofi Presentation
2010 10 07 Octane Ofi Presentationblainemurakami
 
Structured development in BMC Remedy AR System
Structured development in BMC Remedy AR SystemStructured development in BMC Remedy AR System
Structured development in BMC Remedy AR Systemgramlin42
 
Usability Validation Testing of Medical Devices and Software
Usability Validation Testing of Medical Devices and SoftwareUsability Validation Testing of Medical Devices and Software
Usability Validation Testing of Medical Devices and SoftwareUXPA Boston
 

Similar to Appealing to users: developing usable & desirable devices (20)

There is No Such Thing as User Error
There is No Such Thing as User ErrorThere is No Such Thing as User Error
There is No Such Thing as User Error
 
mHealth Israel_Human Factors for MedTech Ergonomics and Usability_Rebecca Moses
mHealth Israel_Human Factors for MedTech Ergonomics and Usability_Rebecca MosesmHealth Israel_Human Factors for MedTech Ergonomics and Usability_Rebecca Moses
mHealth Israel_Human Factors for MedTech Ergonomics and Usability_Rebecca Moses
 
Introduction to Usability
Introduction to UsabilityIntroduction to Usability
Introduction to Usability
 
Introduction to Usability
Introduction to UsabilityIntroduction to Usability
Introduction to Usability
 
Human Factors Engineering for Medical Devices
Human Factors Engineering for Medical DevicesHuman Factors Engineering for Medical Devices
Human Factors Engineering for Medical Devices
 
Mobilizing Electronic Healthcare Records
Mobilizing Electronic Healthcare RecordsMobilizing Electronic Healthcare Records
Mobilizing Electronic Healthcare Records
 
Essay On Reliability Of Visualization Tools
Essay On Reliability Of Visualization ToolsEssay On Reliability Of Visualization Tools
Essay On Reliability Of Visualization Tools
 
Ijetr021224
Ijetr021224Ijetr021224
Ijetr021224
 
Ijetr021224
Ijetr021224Ijetr021224
Ijetr021224
 
Integrating Human Factors Engineering into the Product Development Lifecycle_...
Integrating Human Factors Engineering into the Product Development Lifecycle_...Integrating Human Factors Engineering into the Product Development Lifecycle_...
Integrating Human Factors Engineering into the Product Development Lifecycle_...
 
Risk Assessment: It's a Risky World Out There
Risk Assessment: It's a Risky World Out ThereRisk Assessment: It's a Risky World Out There
Risk Assessment: It's a Risky World Out There
 
How to design Usable EMR/EHR Application
How to design Usable EMR/EHR ApplicationHow to design Usable EMR/EHR Application
How to design Usable EMR/EHR Application
 
IBM involvement in adoption of EHR, health data standards and epSOS - Matej Adam
IBM involvement in adoption of EHR, health data standards and epSOS - Matej AdamIBM involvement in adoption of EHR, health data standards and epSOS - Matej Adam
IBM involvement in adoption of EHR, health data standards and epSOS - Matej Adam
 
Experimental Study Using Functional Size Measurement in Building Estimation M...
Experimental Study Using Functional Size Measurement in Building Estimation M...Experimental Study Using Functional Size Measurement in Building Estimation M...
Experimental Study Using Functional Size Measurement in Building Estimation M...
 
Experimental Study Using Functional Size Measurement in Building Estimation M...
Experimental Study Using Functional Size Measurement in Building Estimation M...Experimental Study Using Functional Size Measurement in Building Estimation M...
Experimental Study Using Functional Size Measurement in Building Estimation M...
 
Usability mHIseminar.Miller-Jacobs
Usability mHIseminar.Miller-JacobsUsability mHIseminar.Miller-Jacobs
Usability mHIseminar.Miller-Jacobs
 
2010 10 07 Octane Ofi Presentation
2010 10 07 Octane Ofi Presentation2010 10 07 Octane Ofi Presentation
2010 10 07 Octane Ofi Presentation
 
Ce Marking Ho
Ce Marking HoCe Marking Ho
Ce Marking Ho
 
Structured development in BMC Remedy AR System
Structured development in BMC Remedy AR SystemStructured development in BMC Remedy AR System
Structured development in BMC Remedy AR System
 
Usability Validation Testing of Medical Devices and Software
Usability Validation Testing of Medical Devices and SoftwareUsability Validation Testing of Medical Devices and Software
Usability Validation Testing of Medical Devices and Software
 

More from Team Consulting Ltd

Planning for the future sustainable medical devices Alastair Willoughby .pdf
Planning for the future sustainable medical devices Alastair Willoughby  .pdfPlanning for the future sustainable medical devices Alastair Willoughby  .pdf
Planning for the future sustainable medical devices Alastair Willoughby .pdfTeam Consulting Ltd
 
The 'Blood-Brain Barrier' for beginners
The 'Blood-Brain Barrier' for beginnersThe 'Blood-Brain Barrier' for beginners
The 'Blood-Brain Barrier' for beginnersTeam Consulting Ltd
 
Informing product design with analytical data
Informing product design with analytical dataInforming product design with analytical data
Informing product design with analytical dataTeam Consulting Ltd
 
It's easy to turn money into technology...
It's easy to turn money into technology...It's easy to turn money into technology...
It's easy to turn money into technology...Team Consulting Ltd
 
ONdrugDelivery - The advantages of designing high-resistance swirl chambers f...
ONdrugDelivery - The advantages of designing high-resistance swirl chambers f...ONdrugDelivery - The advantages of designing high-resistance swirl chambers f...
ONdrugDelivery - The advantages of designing high-resistance swirl chambers f...Team Consulting Ltd
 
Technological Inspiration - Occoris
Technological Inspiration - OccorisTechnological Inspiration - Occoris
Technological Inspiration - OccorisTeam Consulting Ltd
 
Fluid Dynamics: Modelling the Realities of Injection Performance
Fluid Dynamics: Modelling the Realities of Injection PerformanceFluid Dynamics: Modelling the Realities of Injection Performance
Fluid Dynamics: Modelling the Realities of Injection PerformanceTeam Consulting Ltd
 
Engineering the perfect click for drug delivery devices
Engineering the perfect click for drug delivery devicesEngineering the perfect click for drug delivery devices
Engineering the perfect click for drug delivery devicesTeam Consulting Ltd
 
Designing for battery-powered and battery-packed medical devices, EPD&T, Dec ...
Designing for battery-powered and battery-packed medical devices, EPD&T, Dec ...Designing for battery-powered and battery-packed medical devices, EPD&T, Dec ...
Designing for battery-powered and battery-packed medical devices, EPD&T, Dec ...Team Consulting Ltd
 
Taguchi and the tablecloth trick | Insight, issue 5
Taguchi and the tablecloth trick | Insight, issue 5Taguchi and the tablecloth trick | Insight, issue 5
Taguchi and the tablecloth trick | Insight, issue 5Team Consulting Ltd
 
The usability of usability | Insight, issue 5
The usability of usability | Insight, issue 5The usability of usability | Insight, issue 5
The usability of usability | Insight, issue 5Team Consulting Ltd
 

More from Team Consulting Ltd (20)

Planning for the future sustainable medical devices Alastair Willoughby .pdf
Planning for the future sustainable medical devices Alastair Willoughby  .pdfPlanning for the future sustainable medical devices Alastair Willoughby  .pdf
Planning for the future sustainable medical devices Alastair Willoughby .pdf
 
BEEA Team Consulting PR pdf.pdf
BEEA Team Consulting PR pdf.pdfBEEA Team Consulting PR pdf.pdf
BEEA Team Consulting PR pdf.pdf
 
To connect, or not to connect?
To connect, or not to connect?To connect, or not to connect?
To connect, or not to connect?
 
Honing Haemostats
Honing HaemostatsHoning Haemostats
Honing Haemostats
 
Design Drivers
Design DriversDesign Drivers
Design Drivers
 
Failure is a friend
Failure is a friendFailure is a friend
Failure is a friend
 
Unlucky for some!
Unlucky for some!Unlucky for some!
Unlucky for some!
 
The 'Blood-Brain Barrier' for beginners
The 'Blood-Brain Barrier' for beginnersThe 'Blood-Brain Barrier' for beginners
The 'Blood-Brain Barrier' for beginners
 
More than a walk on part
More than a walk on partMore than a walk on part
More than a walk on part
 
Life beyond the 3D print
Life beyond the 3D printLife beyond the 3D print
Life beyond the 3D print
 
Informing product design with analytical data
Informing product design with analytical dataInforming product design with analytical data
Informing product design with analytical data
 
It's easy to turn money into technology...
It's easy to turn money into technology...It's easy to turn money into technology...
It's easy to turn money into technology...
 
ONdrugDelivery - The advantages of designing high-resistance swirl chambers f...
ONdrugDelivery - The advantages of designing high-resistance swirl chambers f...ONdrugDelivery - The advantages of designing high-resistance swirl chambers f...
ONdrugDelivery - The advantages of designing high-resistance swirl chambers f...
 
Healthcare in 2030
Healthcare in 2030Healthcare in 2030
Healthcare in 2030
 
Technological Inspiration - Occoris
Technological Inspiration - OccorisTechnological Inspiration - Occoris
Technological Inspiration - Occoris
 
Fluid Dynamics: Modelling the Realities of Injection Performance
Fluid Dynamics: Modelling the Realities of Injection PerformanceFluid Dynamics: Modelling the Realities of Injection Performance
Fluid Dynamics: Modelling the Realities of Injection Performance
 
Engineering the perfect click for drug delivery devices
Engineering the perfect click for drug delivery devicesEngineering the perfect click for drug delivery devices
Engineering the perfect click for drug delivery devices
 
Designing for battery-powered and battery-packed medical devices, EPD&T, Dec ...
Designing for battery-powered and battery-packed medical devices, EPD&T, Dec ...Designing for battery-powered and battery-packed medical devices, EPD&T, Dec ...
Designing for battery-powered and battery-packed medical devices, EPD&T, Dec ...
 
Taguchi and the tablecloth trick | Insight, issue 5
Taguchi and the tablecloth trick | Insight, issue 5Taguchi and the tablecloth trick | Insight, issue 5
Taguchi and the tablecloth trick | Insight, issue 5
 
The usability of usability | Insight, issue 5
The usability of usability | Insight, issue 5The usability of usability | Insight, issue 5
The usability of usability | Insight, issue 5
 

Appealing to users: developing usable & desirable devices

  • 1. Appealing to Users Developing Usable & Desirable Devices Presenter: Julian Dixon Date: 20th September 2012 Confidential © Team Consulting 2011: World PFS Summit 2012, slide 1 www.team-consulting.com
  • 2. Overview • What is human factors (HF)? • Why so much current interest in HF? • Best practice in patient-use, drug delivery device HF – what is different about patient-use, drug delivery device usability? • Practitioners’ experiences – tips and pitfalls • Realistic usability targets for self-injection devices • Looking beyond usability - safety and effectiveness – towards desirability Confidential © Team Consulting 2011: World PFS Summit 2012, slide 2 www.team-consulting.com
  • 3. What is Human Factors Engineering? “Understanding and optimizing how people interact with technology is the subject of human factors engineering (HFE) and usability engineering (UE).” Ref: FDA Draft Guidance, June 2011: Applying Human Factors and Usability Engineering to Optimize Medical Device Design Confidential © Team Consulting 2011: World PFS Summit 2012, slide 3 www.team-consulting.com
  • 4. Why current interest in HF? • medical devices pose use-related risks • risks need to be identified & addressed during device development … • … to ensure devices can be used safely and effectively by their intended users • why now? • in the past, as an industry, we didn’t pay sufficient attention to use-related risk or to its management through the application of human factors engineering • events have recently focused regulatory attention on the importance of the topic Confidential © Team Consulting 2011: World PFS Summit 2012, slide 4 www.team-consulting.com
  • 5. Confidential © Team Consulting 2011: World PFS Summit 2012, slide 5 www.team-consulting.com
  • 6. Patient-use, drug delivery devices pose a unique HF challenge • disposability = very low cost target • patient use – potentially untrained users – used by anyone, anywhere • limited functionality / limited number of distinct tasks – all use-steps are ‘essential’ or core • tasks go beyond the device – task completion depends on some user behaviour the device cannot enforce – dependence on user – e.g. correct storage / pre-injection checks / disposal • simple task structure / simple device interface – perhaps their apparent simplicity/familiarity may promote guessing? Confidential © Team Consulting 2011: World PFS Summit 2012, slide 6 www.team-consulting.com
  • 7. Statements of HF Objective “… intended users of a medical device can safely and effectively perform critical tasks for the intended uses in the expected use environments.” “… no patterns of use failure or difficulties that could be eliminated or reduced through further modification of the design of the user interface.” “… residual risk is outweighed by the advantages offered by the device.” Ref: FDA Draft Guidance, June 2011: Applying Human Factors and Usability Engineering to Optimize Medical Device Design Confidential © Team Consulting 2011: World PFS Summit 2012, slide 7 www.team-consulting.com
  • 8. User Interface Design Optimized (e.g. device + IFU + packaging) User Interface mitigate  Device Opportunities Concept for Use Error Critical Formative HF Validation HFE Summary Drug/ Hazards & Use Errors Studies Study Report Formulation Clinical Impact verify Use-Related Risk Assessment  Residual Risk (e.g. uFMEA) Confidential © Team Consulting 2011: World PFS Summit 2012, slide 8 www.team-consulting.com
  • 9. User Interface Design Optimized (e.g. device + IFU + packaging) User Interface mitigate  Device Opportunities Concept for Use Error Critical Formative HF Validation HFE Summary Drug/ Hazards & Use Errors Studies Study Report Formulation Clinical Impact verify Use-Related Risk Assessment  Residual Risk (e.g. uFMEA) Intended Use (users, tasks, environments) Confidential © Team Consulting 2011: World PFS Summit 2012, slide 9 www.team-consulting.com
  • 10. start early User Interface Design Optimized (e.g. device + IFU + packaging) User Interface Device Opportunities Concept for Use Error Critical Formative HF Validation HFE Summary Drug/ Hazards & Use Errors Studies Study Report Formulation Clinical Impact Use-Related Risk Assessment Residual Risk (e.g. uFMEA) Intended Use (users, tasks, environments) Confidential © Team Consulting 2011: World PFS Summit 2012, slide 10 www.team-consulting.com
  • 11. our understanding on day 1 was/is good User Interface Design Optimized (e.g. device + IFU + packaging) User Interface Device Opportunities Concept for Use Error Critical Formative HF Validation HFE Summary Drug/ Hazards & Use Errors Studies Study Report Formulation Clinical Impact Use-Related Risk Assessment Residual Risk (e.g. uFMEA) Intended Use (users, tasks, environments) Confidential © Team Consulting 2011: World PFS Summit 2012, slide 11 www.team-consulting.com
  • 12. be clear about who are not intended users User Interface Design Optimized (e.g. device + IFU + packaging) User Interface Device Opportunities Concept for Use Error Critical Formative HF Validation HFE Summary Drug/ Hazards & Use Errors Studies Study Report Formulation Clinical Impact Use-Related Risk Assessment Residual Risk (e.g. uFMEA) Intended Use (users, tasks, environments) Confidential © Team Consulting 2011: World PFS Summit 2012, slide 12 www.team-consulting.com
  • 13. an important instructions for use are part of the user interface User Interface Design Optimized (e.g. device + IFU + packaging) User Interface Device Opportunities Concept for Use Error Critical Formative HF Validation HFE Summary Drug/ Hazards & Use Errors Studies Study Report Formulation Clinical Impact Use-Related Risk Assessment Residual Risk (e.g. uFMEA) Intended Use (users, tasks, environments) Confidential © Team Consulting 2011: World PFS Summit 2012, slide 13 www.team-consulting.com
  • 14. analytical and empirical HF User Interface Design Optimized (e.g. device + IFU + packaging) User Interface Device Opportunities Concept for Use Error Critical Formative HF Validation HFE Summary Drug/ Hazards & Use Errors Studies Study Report Formulation Clinical Impact Use-Related Risk Assessment Residual Risk (e.g. uFMEA) Intended Use (users, tasks, environments) Confidential © Team Consulting 2011: World PFS Summit 2012, slide 14 www.team-consulting.com
  • 15. expect to iterate User Interface Design Optimized (e.g. device + IFU + packaging) User Interface Device Opportunities Concept for Use Error Critical Formative HF Validation HFE Summary Drug/ Hazards & Use Errors Studies Study Report Formulation Clinical Impact Use-Related Risk Assessment Residual Risk (e.g. uFMEA) Intended Use (users, tasks, environments) Confidential © Team Consulting 2011: World PFS Summit 2012, slide 15 www.team-consulting.com
  • 16. don’t leave root cause analysis until validation User Interface Design Optimized (e.g. device + IFU + packaging) User Interface Device Opportunities Concept for Use Error Critical Formative HF Validation HFE Summary Drug/ Hazards & Use Errors Studies Study Report Formulation Clinical Impact Use-Related Risk Assessment Residual Risk (e.g. uFMEA) Intended Use (users, tasks, environments) Confidential © Team Consulting 2011: World PFS Summit 2012, slide 16 www.team-consulting.com
  • 17. ‘critical’ and ‘successful’ should be applied with care User Interface Design Optimized (e.g. device + IFU + packaging) User Interface Device Opportunities Concept for Use Error Critical Formative HF Validation HFE Summary Drug/ Hazards & Use Errors Studies Study Report Formulation Clinical Impact Use-Related Risk Assessment Residual Risk (e.g. uFMEA) Intended Use (users, tasks, environments) Confidential © Team Consulting 2011: World PFS Summit 2012, slide 17 www.team-consulting.com
  • 18. we should learn nothing important during validation User Interface Design Optimized (e.g. device + IFU + packaging) User Interface Device Opportunities Concept for Use Error Critical Formative HF Validation HFE Summary Drug/ Hazards & Use Errors Studies Study Report Formulation Clinical Impact Use-Related Risk Assessment Residual Risk (e.g. uFMEA) Intended Use (users, tasks, environments) Confidential © Team Consulting 2011: World PFS Summit 2012, slide 18 www.team-consulting.com
  • 19. don’t hope for zero User Interface Design Optimized (e.g. device + IFU + packaging) User Interface Device Opportunities Concept for Use Error Critical Formative HF Validation HFE Summary Drug/ Hazards & Use Errors Studies Study Report Formulation Clinical Impact Use-Related Risk Assessment Residual Risk (e.g. uFMEA) Intended Use (users, tasks, environments) Confidential © Team Consulting 2011: World PFS Summit 2012, slide 19 www.team-consulting.com
  • 20. usability acceptance criteria are risk acceptance criteria User Interface Design Optimized (e.g. device + IFU + packaging) User Interface Device Opportunities Concept for Use Error Critical Formative HF Validation HFE Summary Drug/ Hazards & Use Errors Studies Study Report Formulation Clinical Impact Use-Related Risk Assessment Residual Risk (e.g. uFMEA) Intended Use (users, tasks, environments) Confidential © Team Consulting 2011: World PFS Summit 2012, slide 20 www.team-consulting.com
  • 21. qualitative not quantitative User Interface Design Optimized (e.g. device + IFU + packaging) User Interface Device Opportunities Concept for Use Error Critical Formative HF Validation HFE Summary Drug/ Hazards & Use Errors Studies Study Report Formulation Clinical Impact Use-Related Risk Assessment Residual Risk (e.g. uFMEA) Intended Use (users, tasks, environments) Confidential © Team Consulting 2011: World PFS Summit 2012, slide 21 www.team-consulting.com
  • 22. demonstrate understanding User Interface Design Optimized (e.g. device + IFU + packaging) User Interface Device Opportunities Concept for Use Error Critical Formative HF Validation HFE Summary Drug/ Hazards & Use Errors Studies Study Report Formulation Clinical Impact Use-Related Risk Assessment Residual Risk (e.g. uFMEA) Intended Use (users, tasks, environments) Confidential © Team Consulting 2011: World PFS Summit 2012, slide 22 www.team-consulting.com
  • 23. What might be a realistic and sound target for self-injector usability? • What might we agree we should be committed to designing out? – too easy to repeatedly overdose/no-dose inadvertently – too easy to damage device through ‘reasonable misuse’ – procedure of use too complicated • hard to learn/hard to train • easy to get wrong even if paying attention – too easy to misinterpret instructions for use (IFU) – too difficult to follow instructions for use (IFU) – poor physical usability - some tasks too difficult for intended users – opportunities for critical harm to result from predictable negative transfer of prior experience with other injectors – … • What usability issues might we recognise have justifiable residual risks associated with them? – requirement for user to attend to simple-to-follow IFU in order to acquire adequate technique … Confidential © Team Consulting 2011: World PFS Summit 2012, slide 23 www.team-consulting.com
  • 24. design research – going beyond usability Opportunity Certainty Summative - validatory Formative - confirmatory Generative - exploratory Projective - discovery Confidential © Team Consulting 2011: World PFS Summit 2012, slide 24 www.team-consulting.com
  • 25. narrative – tell a coherent story User Interface Design Optimized (e.g. device + IFU + packaging) User Interface Device Opportunities Concept for Use Error Critical Formative HF Validation HFE Summary Drug/ Hazards & Use Errors Studies Study Report Formulation Clinical Impact Use-Related Risk Assessment Residual Risk (e.g. uFMEA) Intended Use (users, tasks, environments) Confidential © Team Consulting 2011: World PFS Summit 2012, slide 25 www.team-consulting.com
  • 26. Confidential © Team Consulting 2011: World PFS Summit 2012, slide 26 www.team-consulting.com
  • 27. For further information, please contact Julian Dixon Tel: +44 (0)1799 532 746 Email: jrd@team-consulting.com Team Consulting Ltd. Abbey Barns, Duxford Road Ickleton, Cambridge CB10 1SX, UK Tel:+44 (0)1799 532 700 Fax:+44 (0)1799 532 701 Email: info@team-consulting.com Confidential © Team Consulting 2011: World PFS Summit 2012, slide 27 www.team-consulting.com