HOW TO DISPLAY CARDIAC OUTPUT AS
LIVE-CHANNEL DURING PRESSURE-
VOLUME RECORDING
Using AD Instrument LabChart
Software
THIS IS FINAL SCREEN SET UP WITH CO BEING 5TH CALCULATED CHANNEL THAT
COULD BE SEEN LIVE DURING THE PROCEDURE.
BEFORE STARTING TO RECORD& COLLECT THE DATA, GO TO VOLUME CHANNEL>
CLICK ON UNIT CONVERSION AND CHECK IN 2 POINT CAL, AND POINT 2 BEING
300ML
NEXT GO INTO SETUP>CHANNEL SETTING>AND CREATE 5TH
CHANNEL AND CALL IT CARDIAC OUTPUT
AFTER CHANNEL NAME IS CREATED, CLICK OK TO CONFIRM;
AND 5TH CHANNELS SHOWS UP
CHANNEL 5 SET UP CONTINUED
CLICK ON 5TH CHANNEL (INSET) AND SELECT PV LOOP FROM THE
MENU
CHANNEL 5 SET UP CONTINUED
CLICK ON PV LOOP AND SELECT FROM THE DROP DOWN MENU
CARDIAC OUTPUT >THEN OK
WHEN COLLECTING DATA YOU WILL HAVE THIS CHANNEL LIVE, DISPLAYING
THE CO (SEE THE ARROW)
You can also pre-set the
Hemodynamic table and
select only CO to be
displayed.
1) Go to Hemodynamic
Table
2) Options
3) Uncheck all
4) Reselect CO
5) Hemodynamic Table
display only one
value, CO.

How to create cardiac output live channel using pv loop recording

  • 1.
    HOW TO DISPLAYCARDIAC OUTPUT AS LIVE-CHANNEL DURING PRESSURE- VOLUME RECORDING Using AD Instrument LabChart Software
  • 2.
    THIS IS FINALSCREEN SET UP WITH CO BEING 5TH CALCULATED CHANNEL THAT COULD BE SEEN LIVE DURING THE PROCEDURE.
  • 3.
    BEFORE STARTING TORECORD& COLLECT THE DATA, GO TO VOLUME CHANNEL> CLICK ON UNIT CONVERSION AND CHECK IN 2 POINT CAL, AND POINT 2 BEING 300ML
  • 4.
    NEXT GO INTOSETUP>CHANNEL SETTING>AND CREATE 5TH CHANNEL AND CALL IT CARDIAC OUTPUT
  • 5.
    AFTER CHANNEL NAMEIS CREATED, CLICK OK TO CONFIRM; AND 5TH CHANNELS SHOWS UP
  • 6.
    CHANNEL 5 SETUP CONTINUED CLICK ON 5TH CHANNEL (INSET) AND SELECT PV LOOP FROM THE MENU
  • 7.
    CHANNEL 5 SETUP CONTINUED CLICK ON PV LOOP AND SELECT FROM THE DROP DOWN MENU CARDIAC OUTPUT >THEN OK
  • 8.
    WHEN COLLECTING DATAYOU WILL HAVE THIS CHANNEL LIVE, DISPLAYING THE CO (SEE THE ARROW) You can also pre-set the Hemodynamic table and select only CO to be displayed. 1) Go to Hemodynamic Table 2) Options 3) Uncheck all 4) Reselect CO 5) Hemodynamic Table display only one value, CO.