Toxicological Testing for Devices
Versus Drugs & Biologics

          L. Bruce Pearce, Ph.D.
                November 7, 2012
          American College of Toxicology
                 Orlando, Florida
Device

     Any thing which is ….

     intended for use in the diagnosis of disease …or in the
     cure, mitigation, treatment, or prevention of disease…, or intended to
     affect the structure or any function of the body of man…,

     and which does not achieve its primary intended purposes through

     chemical action …and which is not dependent upon being
     metabolized for the achievement of its primary intended
     purposes”




ACT Annual Meeting                                                            Slide 2
Scope of Medical Devices

     Extremely broad category of items, such as bandages, crutches and
     wheel chairs, sutures, needles and syringes, catheters, contact
     lenses and lens care solutions, orthopedic implants, dental
     fillings, pacemakers, thermometers, glucose meters, surgical tools
     and instruments, respirators, infusion pumps that deliver drugs, and
     equipment that monitor and measure blood pressure and heart
     rate, surgical meshes, vascular stents, etc.




ACT Annual Meeting                                                          Slide 3
Drugs vs. Devices: Fundamental Differences


  Drugs                                 Devices
  Pure molecules                        Complex components

  Toxicology                            Biocompatibility

  Short half-life                       Durable equipment

  Long market life                      Rapid product cycles

  Drug Interactions                     Malfunction

  Wrong Drug/Dose                       User Error

  Clinically studied                    Bench studied

  Good Manufacturing Practices          Quality Systems (ISO 9000)

  Karanian. J.W. MD-TIP UVA 9-26-2011




ACT Annual Meeting                                                   Slide 4
Drugs vs Devices: Toxicology

     The toxicological effects of devices are not an
       extension of the pharmacological effects at higher
       doses..this does not exist

     The toxicological effects of devices are related to the
       materials devices are made from, their breakdown
       products, precursor molecules and contaminants
       produced during manufacturing

     “Chemistry” is only responsible for toxicological and

       not the beneficial effects of devices

ACT Annual Meeting                                              Slide 5
Context of the Evaluation is Very Different

   If a toxicologist is involved in evaluating a device this
     most likely means that the product is going the 510(k)
     route or a PMA

   Only 2% are PMAs so 98% of the time we are talking a
     510(k)

   If a 510(k) then there is at least one predicate
     device, this means there is prior exposure of humans to
     many if not all the chemicals




ACT Annual Meeting                                              Slide 6
Context of the Evaluation is Very Different

   FDA generally relies, in part, on FDA’s prior
     determination that a reasonable assurance of safety
     and effectiveness exists for the predicate device

   The toxicological evaluation can be the last evaluation
     of biologic safety before the device becomes available
     for human use




ACT Annual Meeting                                            Slide 7
Evaluation of Class II or III Devices Requiring
Clinical Trials

        Investigational Device Exemption (IDE), analogous to
           IND and requires preclinical evidence of safety

        If a PMA, the extent of information required to support
           safety can be much greater and clinical testing is
           required



       .




 ACT Annual Meeting                                                Slide 8
What Are We Assessing?


  Biological Hazards and Risks

   Physical and chemical characteristics of the materials
   Any existing toxicology and other biological safety data on product
     and component materials

   History of clinical use or human exposure data
   Test procedures, in vitro and in vivo
  ISO 14971:2009 (Annex I)




ACT Annual Meeting                                                  Slide 9
Biological Hazard and Risk Assessment: Where Do
  We Begin?

Risk assessment has only recently been integrated into international standards and
endorsed as an integral part of chemical characterization and biocompatibility studies for
medical devices.

The first step for a biological hazard/safety assessment is to characterize the potential
toxicants.

Manufacturing, assembling, packaging, and sterilization of medical devices tend to
result in a multiplicity of process chemicals that can potentially migrate into surrounding
tissues and body fluids.

Many of these chemicals are complex mixtures, often with poorly defined toxicological
profiles. The profiles can become increasingly important because moving from a chemical
with well-established risks to a chemical we know less about can make it difficult to define
the hazard, so a higher risk is assigned.

ACT Annual Meeting                                                                     Slide 10
Elements of Risk Analysis

   Identification of all the chemicals that can leach out of the device and into the
     patient

   Review of the existing toxicology data with emphasis on NOAELs in multiple
     animal species and information regarding toxicological effects in humans and
     exposure limits such as PEL if available—provide a complete toxicological
     profile

   Estimation of Tolerable Exposure (TE) based on information like NOAELs from
     animal studies and uncertainty factors to provide an appropriate margin of
     safety

   Estimation of exposure to all the leachables and compare to TE to assess
     margin of safety


ACT Annual Meeting                                                                      Slide 11
Risk Assessment/Analysis Report

   • The guidelines for the preparation of a risk
      analysis report can be found in:

       • ISO 10993 Part 17 Establishment of Allowable Limits for
          Leachable Substances


   • When do you consider evaluation of inherent
      toxicity of materials?

       • This should begin in the design phase to aid in the
          selection of materials, manufacturing process design
          and sterilization, etc.

ACT Annual Meeting                                                 Slide 12
There are Two Elements to Risk Assessment

 • Chemical Data: Inherent toxicity and risk analysis

 • Biological Data: Biocompatibility Testing
     • Specified by international standards




ACT Annual Meeting                                      Slide 13
Biocompatibility Testing

•    Toxicological testing of devices comes under the rubric of
     “Biocompatibility Testing”

•    Safety assessment of medical devices is guided by toxicology studies
     recommended by the Biological Evaluation of Medical Devices
     Technical Committee of the International Organization for
     Standardization (ISO) documents 10993 (ISO 10993 Technical
     Committee).

      • The ISO guidelines are compendial methods;
      • described in the U.S. Pharmacopoeia (USP) or
      • in the standards published by the American Society for Testing
         and Materials (ASTM).
    ACT Annual Meeting                                                      Slide 14
ISO 10993 Biological Evaluation of Medical Devices

     Part                                                Title                                   PublicationRevision Year

       1     Evaluation and Testing with A Risk Management System                                Fourth Edition 2009
       2     Animal Welfare Requirements                                                         Second Edition 2006
       3     Tests for Genotoxicity, Carcinogenicity, and Reproductive Toxicity                  Second Edition 2003
       4     Selection of Tests for Interactions with Blood and Amendment 1                      Second Edition 2002
                                                                                                 Amendment 1 2006
       5     Tests for In Vitro Cytotoxicity                                                     Third Edition 2009
       6     Test for Local Effects after Implantation                                           Second Edition 2007
       7     Ethylene Oxide Sterilization Residuals                                              Second Edition 2008
                                                                                                 Correction 2009
       8     Withdrawn                                                                                 --
       9     Framework for Identification and Quantification of Potential Degradation Products   Second Edition 2009


      10     Tests for irritation and Delayed Type Hypersensitivity and Amendment 1              Third Edition 2010
      11     Tests for Systemic Toxicity                                                         Second Edition 2006
      12     Sample Preparation and Reference Materials                                          Third Edition 2007
      13     Identification and Quantification of Degradation Products from Polymeric Medical    First Edition 1998
             Devices
      14     Identification and Quantification of Degradation Products from Ceramics             First Edition 2001
      15     Identification and Quantification of Degradation Products from Metals and Alloys    First Edition 2000
                                                                                                 Correction 2001
      16     Toxicokinetic Study Design for Degradation Products and Leachables                  Second Edition 2010
      17     Establishment of Allowable Limits for Leachable Substances                          First Edition 2002
      18     Chemical Characterization of Materials                                              First Edition 2005
      19     Physicochemical, Morphological, and Topographical Characterization of Materials     First Edition 2006


      20     Principles and Methods for Immunotoxicity Testing of Medical Devices                First Edition 2006




ACT Annual Meeting                                                                                                           Slide 15
FDA vs ISO Biocompatibility Testing Matrix

     • Prior to 1995 biocompatibility testing of medical devices in
        the US was usually conducted by the use of the #G87-1
        Tripartite Biocompatibility Guidance (1987).
     • FDA issued the #G95-1 guidance document in 1995 which
        was a modified version of ISO 10993, "Biological
        Evaluation of Medical Devices—Part 1.“
     • This document is really in agreement with the ISO testing
        matrix with a number of exceptions
     • Assume we are discussing Class II and III medical devices
     • The number of tests required for a device depends on the
        intended use and duration of exposure
         • Is this a device that is placed on the skin or implanted in the body and for how
            long?




ACT Annual Meeting                                                                            Slide 16
Intended Use vs Evaluations
    ISO/FDA Test Chart




ACT Annual Meeting              Slide 17
Toxicity Testing Devices Vs Drugs/Biologics
 Devices                                                                      Biologics   Drugs
 Cytotoxicity                                                                        -      -
 Skin Sensitization                                                                  -      -
 Irritation (dermal or ocular)                                                       -      -
 Intracutaneous Reactivity                                                           √*    √*
 Pyrogenicity                                                                        √     √
 Genotoxicity                                                                        √     √
 Acute Toxicity                                                                      √     √
 Subchronic Toxicity
 Chronic Toxicity                                                                    √     √
 Carcinogenicity                                                                     √     √
 Reproductive Toxicity                                                               √     √
 Implantation                                                                        -      -

 *with intracutaneous or intramuscular administration, especially for vaccines and
 Included in general tox study design not as a separate study like for devices




ACT Annual Meeting                                                                                Slide 18
Biocompatibility

 •   Biomaterials or devices interact with tissues and produce some
     level of reaction in contrast to drugs/biologics

 •   The major category of tissue reactions is the “Foreign Body
     Response” or FBR

 •   This can be modified by surgical trauma or implant procedure
     and also things like;
     •   sutures used in the placement of the device eg. tissue mesh

     • contaminates from manufacturing and pyrogens

 •   Can be very complicated and a good pathologist can be
     important to avoiding a bad outcome



ACT Annual Meeting                                                     Slide 19
Summary

  • Enormous diversity of devices; surgical gloves to stents,
  • Efficacy does not involve chemical action or metabolism—
     you are not thinking about other pharmacological sites of
     action as a source of adverse events or toxicity

  • 98% of the time a 510(k) which may lead to “FDA-release”,
     not approval, the latter reserved for successful PMA
      • A predicate device is the reference
      • The goal is to establish substantial equivalence
      • No clinical trials to prove safety and efficacy majority of the time
      • FDA generally relies, in part, on FDA’s prior determination that a reasonable
         assurance of safety and effectiveness exists for the predicate device.




ACT Annual Meeting                                                                      Slide 20
Summary

 • Biological Hazard/Risk Assessment
     • The toxicologist can have a unique role in the 510(k) process
     • Assessing the toxicological risk of the leachables and contaminants that are
        well defined-Risk Analysis : ISO-10992-17

     • In vitro and in vivo Biocompatibility testing including systemic tox in animals
         ISO-10993 the Holy Grail with additions by FDA
         Extent of testing depends on route and duration of exposure




ACT Annual Meeting                                                                       Slide 21

More Related Content

What's hot

ISO 10993 Series Part 1: Evaluation and Testing In The Risk Management Process
ISO 10993 Series Part 1: Evaluation and Testing In The Risk Management ProcessISO 10993 Series Part 1: Evaluation and Testing In The Risk Management Process
ISO 10993 Series Part 1: Evaluation and Testing In The Risk Management ProcessNAMSA
 
A Complete Guide about Biocompatibility for Medical Devices and Biocompatibil...
A Complete Guide about Biocompatibility for Medical Devices and Biocompatibil...A Complete Guide about Biocompatibility for Medical Devices and Biocompatibil...
A Complete Guide about Biocompatibility for Medical Devices and Biocompatibil...Biocomptesting, Inc.
 
ISO 10993-5 Biological Evaluation of Medical Devices - Tests for In Vitro Cyt...
ISO 10993-5 Biological Evaluation of Medical Devices - Tests for In Vitro Cyt...ISO 10993-5 Biological Evaluation of Medical Devices - Tests for In Vitro Cyt...
ISO 10993-5 Biological Evaluation of Medical Devices - Tests for In Vitro Cyt...NAMSA
 
Biocompatibility testing of "BIOMATERIALS"
Biocompatibility testing of "BIOMATERIALS"Biocompatibility testing of "BIOMATERIALS"
Biocompatibility testing of "BIOMATERIALS"Swapnil Singh
 
Scale up techniques in the production of sterile products & ophthalmic pr...
Scale up techniques in the production of sterile products & ophthalmic pr...Scale up techniques in the production of sterile products & ophthalmic pr...
Scale up techniques in the production of sterile products & ophthalmic pr...Dr. Jigar Vyas
 
Actcopypearcesympxv110712 130114160643-phpapp01
Actcopypearcesympxv110712 130114160643-phpapp01Actcopypearcesympxv110712 130114160643-phpapp01
Actcopypearcesympxv110712 130114160643-phpapp01A Sairam Kishore, Ph.D.,
 
Toxicological Risk Assessment For Medical Devices - ISO 10993-1
Toxicological Risk Assessment For Medical Devices - ISO 10993-1Toxicological Risk Assessment For Medical Devices - ISO 10993-1
Toxicological Risk Assessment For Medical Devices - ISO 10993-1Russell Sloboda
 
ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effec...
ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effec...ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effec...
ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effec...NAMSA
 
Integration of Risk Assessment and Chemical Characterization (MD&M Minn. 2017)
Integration of Risk Assessment and Chemical Characterization (MD&M Minn. 2017)Integration of Risk Assessment and Chemical Characterization (MD&M Minn. 2017)
Integration of Risk Assessment and Chemical Characterization (MD&M Minn. 2017)Russell Sloboda
 
Impurities in Drug Synthesis
Impurities in Drug SynthesisImpurities in Drug Synthesis
Impurities in Drug SynthesisDrSSreenivasa
 
In-House Microbial Isolates in Compendial Testing: Regulatory Requirements
In-House Microbial Isolates in Compendial Testing: Regulatory RequirementsIn-House Microbial Isolates in Compendial Testing: Regulatory Requirements
In-House Microbial Isolates in Compendial Testing: Regulatory RequirementsRobert Westney
 
ISO 10993-3: Biological Evaluation of Medical Devices - Tests for Genotoxicit...
ISO 10993-3: Biological Evaluation of Medical Devices - Tests for Genotoxicit...ISO 10993-3: Biological Evaluation of Medical Devices - Tests for Genotoxicit...
ISO 10993-3: Biological Evaluation of Medical Devices - Tests for Genotoxicit...NAMSA
 
Characterization of Microbial Isolates from the Manufacturing Process Stream
Characterization of Microbial Isolates from the Manufacturing Process StreamCharacterization of Microbial Isolates from the Manufacturing Process Stream
Characterization of Microbial Isolates from the Manufacturing Process StreamRobert Westney
 
ISO 10993-7 Biological Evaluation of Medical Devices - Ethylene Oxide Sterili...
ISO 10993-7 Biological Evaluation of Medical Devices - Ethylene Oxide Sterili...ISO 10993-7 Biological Evaluation of Medical Devices - Ethylene Oxide Sterili...
ISO 10993-7 Biological Evaluation of Medical Devices - Ethylene Oxide Sterili...NAMSA
 
Q3D Guideline For Elemental Impurities
Q3D Guideline For Elemental ImpuritiesQ3D Guideline For Elemental Impurities
Q3D Guideline For Elemental ImpuritiesMuhamad Abdalkader
 
Case Study: Sterilization of Surgical Instruments
Case Study: Sterilization of Surgical InstrumentsCase Study: Sterilization of Surgical Instruments
Case Study: Sterilization of Surgical InstrumentsJordi Labs
 
Potential elemental impurities ,chns analyser
Potential elemental impurities ,chns analyserPotential elemental impurities ,chns analyser
Potential elemental impurities ,chns analyserKUNDLAJAYALAKSHMI
 

What's hot (20)

Biocompatibility
BiocompatibilityBiocompatibility
Biocompatibility
 
ISO 10993 Series Part 1: Evaluation and Testing In The Risk Management Process
ISO 10993 Series Part 1: Evaluation and Testing In The Risk Management ProcessISO 10993 Series Part 1: Evaluation and Testing In The Risk Management Process
ISO 10993 Series Part 1: Evaluation and Testing In The Risk Management Process
 
A Complete Guide about Biocompatibility for Medical Devices and Biocompatibil...
A Complete Guide about Biocompatibility for Medical Devices and Biocompatibil...A Complete Guide about Biocompatibility for Medical Devices and Biocompatibil...
A Complete Guide about Biocompatibility for Medical Devices and Biocompatibil...
 
ISO 10993-5 Biological Evaluation of Medical Devices - Tests for In Vitro Cyt...
ISO 10993-5 Biological Evaluation of Medical Devices - Tests for In Vitro Cyt...ISO 10993-5 Biological Evaluation of Medical Devices - Tests for In Vitro Cyt...
ISO 10993-5 Biological Evaluation of Medical Devices - Tests for In Vitro Cyt...
 
Biocompatibility testing of "BIOMATERIALS"
Biocompatibility testing of "BIOMATERIALS"Biocompatibility testing of "BIOMATERIALS"
Biocompatibility testing of "BIOMATERIALS"
 
Scale up techniques in the production of sterile products & ophthalmic pr...
Scale up techniques in the production of sterile products & ophthalmic pr...Scale up techniques in the production of sterile products & ophthalmic pr...
Scale up techniques in the production of sterile products & ophthalmic pr...
 
Extractable and leachables
Extractable and leachablesExtractable and leachables
Extractable and leachables
 
Actcopypearcesympxv110712 130114160643-phpapp01
Actcopypearcesympxv110712 130114160643-phpapp01Actcopypearcesympxv110712 130114160643-phpapp01
Actcopypearcesympxv110712 130114160643-phpapp01
 
Toxicological Risk Assessment For Medical Devices - ISO 10993-1
Toxicological Risk Assessment For Medical Devices - ISO 10993-1Toxicological Risk Assessment For Medical Devices - ISO 10993-1
Toxicological Risk Assessment For Medical Devices - ISO 10993-1
 
ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effec...
ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effec...ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effec...
ISO 10993-6: Biological Evaluation of Medical Devices - Tests for local effec...
 
Integration of Risk Assessment and Chemical Characterization (MD&M Minn. 2017)
Integration of Risk Assessment and Chemical Characterization (MD&M Minn. 2017)Integration of Risk Assessment and Chemical Characterization (MD&M Minn. 2017)
Integration of Risk Assessment and Chemical Characterization (MD&M Minn. 2017)
 
Impurities in Drug Synthesis
Impurities in Drug SynthesisImpurities in Drug Synthesis
Impurities in Drug Synthesis
 
In-House Microbial Isolates in Compendial Testing: Regulatory Requirements
In-House Microbial Isolates in Compendial Testing: Regulatory RequirementsIn-House Microbial Isolates in Compendial Testing: Regulatory Requirements
In-House Microbial Isolates in Compendial Testing: Regulatory Requirements
 
ISO 10993-3: Biological Evaluation of Medical Devices - Tests for Genotoxicit...
ISO 10993-3: Biological Evaluation of Medical Devices - Tests for Genotoxicit...ISO 10993-3: Biological Evaluation of Medical Devices - Tests for Genotoxicit...
ISO 10993-3: Biological Evaluation of Medical Devices - Tests for Genotoxicit...
 
Characterization of Microbial Isolates from the Manufacturing Process Stream
Characterization of Microbial Isolates from the Manufacturing Process StreamCharacterization of Microbial Isolates from the Manufacturing Process Stream
Characterization of Microbial Isolates from the Manufacturing Process Stream
 
ISO 10993-7 Biological Evaluation of Medical Devices - Ethylene Oxide Sterili...
ISO 10993-7 Biological Evaluation of Medical Devices - Ethylene Oxide Sterili...ISO 10993-7 Biological Evaluation of Medical Devices - Ethylene Oxide Sterili...
ISO 10993-7 Biological Evaluation of Medical Devices - Ethylene Oxide Sterili...
 
Q3A(R2)
Q3A(R2)Q3A(R2)
Q3A(R2)
 
Q3D Guideline For Elemental Impurities
Q3D Guideline For Elemental ImpuritiesQ3D Guideline For Elemental Impurities
Q3D Guideline For Elemental Impurities
 
Case Study: Sterilization of Surgical Instruments
Case Study: Sterilization of Surgical InstrumentsCase Study: Sterilization of Surgical Instruments
Case Study: Sterilization of Surgical Instruments
 
Potential elemental impurities ,chns analyser
Potential elemental impurities ,chns analyserPotential elemental impurities ,chns analyser
Potential elemental impurities ,chns analyser
 

Viewers also liked

Woodbay Ltd Takes Initiative by Transitioning to ISO 9001:2015
Woodbay Ltd Takes Initiative by Transitioning to ISO 9001:2015Woodbay Ltd Takes Initiative by Transitioning to ISO 9001:2015
Woodbay Ltd Takes Initiative by Transitioning to ISO 9001:2015SGS
 
DESIGN OF IMPLANTABLE DEVICE – UNDERSTANDING THE PREMARKET REVIEW PROCESS AN...
DESIGN OF IMPLANTABLE DEVICE – UNDERSTANDING THE PREMARKET REVIEW  PROCESS AN...DESIGN OF IMPLANTABLE DEVICE – UNDERSTANDING THE PREMARKET REVIEW  PROCESS AN...
DESIGN OF IMPLANTABLE DEVICE – UNDERSTANDING THE PREMARKET REVIEW PROCESS AN...UBMCanon
 
TAG Farnborough Airport Achieves International Environmental Standard ISO 140...
TAG Farnborough Airport Achieves International Environmental Standard ISO 140...TAG Farnborough Airport Achieves International Environmental Standard ISO 140...
TAG Farnborough Airport Achieves International Environmental Standard ISO 140...SGS
 
SGS Sustainability Report 2014
SGS Sustainability Report 2014SGS Sustainability Report 2014
SGS Sustainability Report 2014SGS
 
SGS Group Profile 2015
SGS Group Profile 2015SGS Group Profile 2015
SGS Group Profile 2015SGS
 
Let’s Talk Bananas: Understanding Food Safety & Quality for Fresh Fruits & Ve...
Let’s Talk Bananas: Understanding Food Safety & Quality for Fresh Fruits & Ve...Let’s Talk Bananas: Understanding Food Safety & Quality for Fresh Fruits & Ve...
Let’s Talk Bananas: Understanding Food Safety & Quality for Fresh Fruits & Ve...SGS
 
SGS INTRON Bulletin 24
SGS INTRON Bulletin 24SGS INTRON Bulletin 24
SGS INTRON Bulletin 24SGS
 
Determination of Elemental Impurities – Challenges of a Screening Method
Determination of Elemental Impurities – Challenges of a Screening MethodDetermination of Elemental Impurities – Challenges of a Screening Method
Determination of Elemental Impurities – Challenges of a Screening MethodSGS
 
SGS 2014 Annual Report
SGS 2014 Annual ReportSGS 2014 Annual Report
SGS 2014 Annual ReportSGS
 
Drug degradation
Drug degradationDrug degradation
Drug degradationAmeenah
 
SGS 2016 Annual Report
SGS 2016 Annual ReportSGS 2016 Annual Report
SGS 2016 Annual ReportSGS
 
Final gap analysis and impact assessment for packaging development for pharma...
Final gap analysis and impact assessment for packaging development for pharma...Final gap analysis and impact assessment for packaging development for pharma...
Final gap analysis and impact assessment for packaging development for pharma...hncsaurav
 

Viewers also liked (14)

Woodbay Ltd Takes Initiative by Transitioning to ISO 9001:2015
Woodbay Ltd Takes Initiative by Transitioning to ISO 9001:2015Woodbay Ltd Takes Initiative by Transitioning to ISO 9001:2015
Woodbay Ltd Takes Initiative by Transitioning to ISO 9001:2015
 
DESIGN OF IMPLANTABLE DEVICE – UNDERSTANDING THE PREMARKET REVIEW PROCESS AN...
DESIGN OF IMPLANTABLE DEVICE – UNDERSTANDING THE PREMARKET REVIEW  PROCESS AN...DESIGN OF IMPLANTABLE DEVICE – UNDERSTANDING THE PREMARKET REVIEW  PROCESS AN...
DESIGN OF IMPLANTABLE DEVICE – UNDERSTANDING THE PREMARKET REVIEW PROCESS AN...
 
Stability techniques
Stability techniquesStability techniques
Stability techniques
 
TAG Farnborough Airport Achieves International Environmental Standard ISO 140...
TAG Farnborough Airport Achieves International Environmental Standard ISO 140...TAG Farnborough Airport Achieves International Environmental Standard ISO 140...
TAG Farnborough Airport Achieves International Environmental Standard ISO 140...
 
SGS Sustainability Report 2014
SGS Sustainability Report 2014SGS Sustainability Report 2014
SGS Sustainability Report 2014
 
SGS Group Profile 2015
SGS Group Profile 2015SGS Group Profile 2015
SGS Group Profile 2015
 
Let’s Talk Bananas: Understanding Food Safety & Quality for Fresh Fruits & Ve...
Let’s Talk Bananas: Understanding Food Safety & Quality for Fresh Fruits & Ve...Let’s Talk Bananas: Understanding Food Safety & Quality for Fresh Fruits & Ve...
Let’s Talk Bananas: Understanding Food Safety & Quality for Fresh Fruits & Ve...
 
SGS INTRON Bulletin 24
SGS INTRON Bulletin 24SGS INTRON Bulletin 24
SGS INTRON Bulletin 24
 
Determination of Elemental Impurities – Challenges of a Screening Method
Determination of Elemental Impurities – Challenges of a Screening MethodDetermination of Elemental Impurities – Challenges of a Screening Method
Determination of Elemental Impurities – Challenges of a Screening Method
 
SGS 2014 Annual Report
SGS 2014 Annual ReportSGS 2014 Annual Report
SGS 2014 Annual Report
 
Insulin Pump
Insulin PumpInsulin Pump
Insulin Pump
 
Drug degradation
Drug degradationDrug degradation
Drug degradation
 
SGS 2016 Annual Report
SGS 2016 Annual ReportSGS 2016 Annual Report
SGS 2016 Annual Report
 
Final gap analysis and impact assessment for packaging development for pharma...
Final gap analysis and impact assessment for packaging development for pharma...Final gap analysis and impact assessment for packaging development for pharma...
Final gap analysis and impact assessment for packaging development for pharma...
 

Similar to Toxicological Testing for Devices Versus Drugs & Biologics

Alternative and integrated testing strategies
Alternative and integrated testing strategiesAlternative and integrated testing strategies
Alternative and integrated testing strategiesEFSA EU
 
Iso 10993 1
Iso 10993 1Iso 10993 1
Iso 10993 1Suhas R
 
Use of Chemical Characterization to Assess the Equivalency of Medical Devices...
Use of Chemical Characterization to Assess the Equivalency of Medical Devices...Use of Chemical Characterization to Assess the Equivalency of Medical Devices...
Use of Chemical Characterization to Assess the Equivalency of Medical Devices...NAMSA
 
Redacted bpsa particulate paper bpog 150115 (1)
Redacted bpsa particulate paper bpog 150115 (1)Redacted bpsa particulate paper bpog 150115 (1)
Redacted bpsa particulate paper bpog 150115 (1)BioPhorum Operations Group
 
Toxicological profile for endosulfan
Toxicological profile for endosulfanToxicological profile for endosulfan
Toxicological profile for endosulfanMumbaikar Le
 
Nanotechnology in Consumer Products: An Update on Regulatory Responses and Li...
Nanotechnology in Consumer Products: An Update on Regulatory Responses and Li...Nanotechnology in Consumer Products: An Update on Regulatory Responses and Li...
Nanotechnology in Consumer Products: An Update on Regulatory Responses and Li...kurfirst
 
Upcoming USP 665 - Level of Characterization of Single-Use Systems Today and ...
Upcoming USP 665 - Level of Characterization of Single-Use Systems Today and ...Upcoming USP 665 - Level of Characterization of Single-Use Systems Today and ...
Upcoming USP 665 - Level of Characterization of Single-Use Systems Today and ...Merck Life Sciences
 
Upcoming USP 665 - Level of Characterization of Single-Use Systems Today and ...
Upcoming USP 665 - Level of Characterization of Single-Use Systems Today and ...Upcoming USP 665 - Level of Characterization of Single-Use Systems Today and ...
Upcoming USP 665 - Level of Characterization of Single-Use Systems Today and ...MilliporeSigma
 
International Guidelines and Regulatory Agencies for Toxicity Studies
International Guidelines and Regulatory Agencies for Toxicity StudiesInternational Guidelines and Regulatory Agencies for Toxicity Studies
International Guidelines and Regulatory Agencies for Toxicity StudiesSuneal Saini
 
Pharmacuetical Analysis lecture notes
Pharmacuetical Analysis  lecture notesPharmacuetical Analysis  lecture notes
Pharmacuetical Analysis lecture notesDenekeAlemayehu
 
Guidance Document for Acute Lethality Testing.PDF
Guidance Document for Acute Lethality Testing.PDFGuidance Document for Acute Lethality Testing.PDF
Guidance Document for Acute Lethality Testing.PDFGuy Gilron
 
Antti haapalinna 10th december 08 oulu1
Antti haapalinna 10th december 08 oulu1Antti haapalinna 10th december 08 oulu1
Antti haapalinna 10th december 08 oulu1Antti Haapalinna
 
Regulatory complexities in hp drugs
Regulatory complexities in hp drugsRegulatory complexities in hp drugs
Regulatory complexities in hp drugsBhaswat Chakraborty
 
Ich guidelines
Ich guidelinesIch guidelines
Ich guidelinesAnshul2593
 
Considerations for Biocompatibility Evaluation
Considerations for Biocompatibility EvaluationConsiderations for Biocompatibility Evaluation
Considerations for Biocompatibility EvaluationEMMAIntl
 

Similar to Toxicological Testing for Devices Versus Drugs & Biologics (20)

Alternative and integrated testing strategies
Alternative and integrated testing strategiesAlternative and integrated testing strategies
Alternative and integrated testing strategies
 
Quality for Biologics
Quality for BiologicsQuality for Biologics
Quality for Biologics
 
Iso 10993 1
Iso 10993 1Iso 10993 1
Iso 10993 1
 
Use of Chemical Characterization to Assess the Equivalency of Medical Devices...
Use of Chemical Characterization to Assess the Equivalency of Medical Devices...Use of Chemical Characterization to Assess the Equivalency of Medical Devices...
Use of Chemical Characterization to Assess the Equivalency of Medical Devices...
 
Redacted bpsa particulate paper bpog 150115 (1)
Redacted bpsa particulate paper bpog 150115 (1)Redacted bpsa particulate paper bpog 150115 (1)
Redacted bpsa particulate paper bpog 150115 (1)
 
Toxicological profile for endosulfan
Toxicological profile for endosulfanToxicological profile for endosulfan
Toxicological profile for endosulfan
 
Nanotechnology in Consumer Products: An Update on Regulatory Responses and Li...
Nanotechnology in Consumer Products: An Update on Regulatory Responses and Li...Nanotechnology in Consumer Products: An Update on Regulatory Responses and Li...
Nanotechnology in Consumer Products: An Update on Regulatory Responses and Li...
 
839
839839
839
 
Upcoming USP 665 - Level of Characterization of Single-Use Systems Today and ...
Upcoming USP 665 - Level of Characterization of Single-Use Systems Today and ...Upcoming USP 665 - Level of Characterization of Single-Use Systems Today and ...
Upcoming USP 665 - Level of Characterization of Single-Use Systems Today and ...
 
Upcoming USP 665 - Level of Characterization of Single-Use Systems Today and ...
Upcoming USP 665 - Level of Characterization of Single-Use Systems Today and ...Upcoming USP 665 - Level of Characterization of Single-Use Systems Today and ...
Upcoming USP 665 - Level of Characterization of Single-Use Systems Today and ...
 
International Guidelines and Regulatory Agencies for Toxicity Studies
International Guidelines and Regulatory Agencies for Toxicity StudiesInternational Guidelines and Regulatory Agencies for Toxicity Studies
International Guidelines and Regulatory Agencies for Toxicity Studies
 
Pharmacuetical Analysis lecture notes
Pharmacuetical Analysis  lecture notesPharmacuetical Analysis  lecture notes
Pharmacuetical Analysis lecture notes
 
Guidance Document for Acute Lethality Testing.PDF
Guidance Document for Acute Lethality Testing.PDFGuidance Document for Acute Lethality Testing.PDF
Guidance Document for Acute Lethality Testing.PDF
 
Antti haapalinna 10th december 08 oulu1
Antti haapalinna 10th december 08 oulu1Antti haapalinna 10th december 08 oulu1
Antti haapalinna 10th december 08 oulu1
 
Toxicology endpoints
Toxicology endpointsToxicology endpoints
Toxicology endpoints
 
Regulatory complexities in hp drugs
Regulatory complexities in hp drugsRegulatory complexities in hp drugs
Regulatory complexities in hp drugs
 
Ich guidelines
Ich guidelinesIch guidelines
Ich guidelines
 
Tiêu chuẩn phòng sạch trong sản xuất vaccine
Tiêu chuẩn phòng sạch trong sản xuất vaccine Tiêu chuẩn phòng sạch trong sản xuất vaccine
Tiêu chuẩn phòng sạch trong sản xuất vaccine
 
Considerations for Biocompatibility Evaluation
Considerations for Biocompatibility EvaluationConsiderations for Biocompatibility Evaluation
Considerations for Biocompatibility Evaluation
 
1.3 ICH .pdf
1.3 ICH .pdf1.3 ICH .pdf
1.3 ICH .pdf
 

Toxicological Testing for Devices Versus Drugs & Biologics

  • 1. Toxicological Testing for Devices Versus Drugs & Biologics L. Bruce Pearce, Ph.D. November 7, 2012 American College of Toxicology Orlando, Florida
  • 2. Device Any thing which is …. intended for use in the diagnosis of disease …or in the cure, mitigation, treatment, or prevention of disease…, or intended to affect the structure or any function of the body of man…, and which does not achieve its primary intended purposes through chemical action …and which is not dependent upon being metabolized for the achievement of its primary intended purposes” ACT Annual Meeting Slide 2
  • 3. Scope of Medical Devices Extremely broad category of items, such as bandages, crutches and wheel chairs, sutures, needles and syringes, catheters, contact lenses and lens care solutions, orthopedic implants, dental fillings, pacemakers, thermometers, glucose meters, surgical tools and instruments, respirators, infusion pumps that deliver drugs, and equipment that monitor and measure blood pressure and heart rate, surgical meshes, vascular stents, etc. ACT Annual Meeting Slide 3
  • 4. Drugs vs. Devices: Fundamental Differences Drugs Devices Pure molecules Complex components Toxicology Biocompatibility Short half-life Durable equipment Long market life Rapid product cycles Drug Interactions Malfunction Wrong Drug/Dose User Error Clinically studied Bench studied Good Manufacturing Practices Quality Systems (ISO 9000) Karanian. J.W. MD-TIP UVA 9-26-2011 ACT Annual Meeting Slide 4
  • 5. Drugs vs Devices: Toxicology  The toxicological effects of devices are not an extension of the pharmacological effects at higher doses..this does not exist  The toxicological effects of devices are related to the materials devices are made from, their breakdown products, precursor molecules and contaminants produced during manufacturing  “Chemistry” is only responsible for toxicological and not the beneficial effects of devices ACT Annual Meeting Slide 5
  • 6. Context of the Evaluation is Very Different  If a toxicologist is involved in evaluating a device this most likely means that the product is going the 510(k) route or a PMA  Only 2% are PMAs so 98% of the time we are talking a 510(k)  If a 510(k) then there is at least one predicate device, this means there is prior exposure of humans to many if not all the chemicals ACT Annual Meeting Slide 6
  • 7. Context of the Evaluation is Very Different  FDA generally relies, in part, on FDA’s prior determination that a reasonable assurance of safety and effectiveness exists for the predicate device  The toxicological evaluation can be the last evaluation of biologic safety before the device becomes available for human use ACT Annual Meeting Slide 7
  • 8. Evaluation of Class II or III Devices Requiring Clinical Trials  Investigational Device Exemption (IDE), analogous to IND and requires preclinical evidence of safety  If a PMA, the extent of information required to support safety can be much greater and clinical testing is required . ACT Annual Meeting Slide 8
  • 9. What Are We Assessing? Biological Hazards and Risks  Physical and chemical characteristics of the materials  Any existing toxicology and other biological safety data on product and component materials  History of clinical use or human exposure data  Test procedures, in vitro and in vivo ISO 14971:2009 (Annex I) ACT Annual Meeting Slide 9
  • 10. Biological Hazard and Risk Assessment: Where Do We Begin? Risk assessment has only recently been integrated into international standards and endorsed as an integral part of chemical characterization and biocompatibility studies for medical devices. The first step for a biological hazard/safety assessment is to characterize the potential toxicants. Manufacturing, assembling, packaging, and sterilization of medical devices tend to result in a multiplicity of process chemicals that can potentially migrate into surrounding tissues and body fluids. Many of these chemicals are complex mixtures, often with poorly defined toxicological profiles. The profiles can become increasingly important because moving from a chemical with well-established risks to a chemical we know less about can make it difficult to define the hazard, so a higher risk is assigned. ACT Annual Meeting Slide 10
  • 11. Elements of Risk Analysis  Identification of all the chemicals that can leach out of the device and into the patient  Review of the existing toxicology data with emphasis on NOAELs in multiple animal species and information regarding toxicological effects in humans and exposure limits such as PEL if available—provide a complete toxicological profile  Estimation of Tolerable Exposure (TE) based on information like NOAELs from animal studies and uncertainty factors to provide an appropriate margin of safety  Estimation of exposure to all the leachables and compare to TE to assess margin of safety ACT Annual Meeting Slide 11
  • 12. Risk Assessment/Analysis Report • The guidelines for the preparation of a risk analysis report can be found in: • ISO 10993 Part 17 Establishment of Allowable Limits for Leachable Substances • When do you consider evaluation of inherent toxicity of materials? • This should begin in the design phase to aid in the selection of materials, manufacturing process design and sterilization, etc. ACT Annual Meeting Slide 12
  • 13. There are Two Elements to Risk Assessment • Chemical Data: Inherent toxicity and risk analysis • Biological Data: Biocompatibility Testing • Specified by international standards ACT Annual Meeting Slide 13
  • 14. Biocompatibility Testing • Toxicological testing of devices comes under the rubric of “Biocompatibility Testing” • Safety assessment of medical devices is guided by toxicology studies recommended by the Biological Evaluation of Medical Devices Technical Committee of the International Organization for Standardization (ISO) documents 10993 (ISO 10993 Technical Committee). • The ISO guidelines are compendial methods; • described in the U.S. Pharmacopoeia (USP) or • in the standards published by the American Society for Testing and Materials (ASTM). ACT Annual Meeting Slide 14
  • 15. ISO 10993 Biological Evaluation of Medical Devices Part Title PublicationRevision Year 1 Evaluation and Testing with A Risk Management System Fourth Edition 2009 2 Animal Welfare Requirements Second Edition 2006 3 Tests for Genotoxicity, Carcinogenicity, and Reproductive Toxicity Second Edition 2003 4 Selection of Tests for Interactions with Blood and Amendment 1 Second Edition 2002 Amendment 1 2006 5 Tests for In Vitro Cytotoxicity Third Edition 2009 6 Test for Local Effects after Implantation Second Edition 2007 7 Ethylene Oxide Sterilization Residuals Second Edition 2008 Correction 2009 8 Withdrawn -- 9 Framework for Identification and Quantification of Potential Degradation Products Second Edition 2009 10 Tests for irritation and Delayed Type Hypersensitivity and Amendment 1 Third Edition 2010 11 Tests for Systemic Toxicity Second Edition 2006 12 Sample Preparation and Reference Materials Third Edition 2007 13 Identification and Quantification of Degradation Products from Polymeric Medical First Edition 1998 Devices 14 Identification and Quantification of Degradation Products from Ceramics First Edition 2001 15 Identification and Quantification of Degradation Products from Metals and Alloys First Edition 2000 Correction 2001 16 Toxicokinetic Study Design for Degradation Products and Leachables Second Edition 2010 17 Establishment of Allowable Limits for Leachable Substances First Edition 2002 18 Chemical Characterization of Materials First Edition 2005 19 Physicochemical, Morphological, and Topographical Characterization of Materials First Edition 2006 20 Principles and Methods for Immunotoxicity Testing of Medical Devices First Edition 2006 ACT Annual Meeting Slide 15
  • 16. FDA vs ISO Biocompatibility Testing Matrix • Prior to 1995 biocompatibility testing of medical devices in the US was usually conducted by the use of the #G87-1 Tripartite Biocompatibility Guidance (1987). • FDA issued the #G95-1 guidance document in 1995 which was a modified version of ISO 10993, "Biological Evaluation of Medical Devices—Part 1.“ • This document is really in agreement with the ISO testing matrix with a number of exceptions • Assume we are discussing Class II and III medical devices • The number of tests required for a device depends on the intended use and duration of exposure • Is this a device that is placed on the skin or implanted in the body and for how long? ACT Annual Meeting Slide 16
  • 17. Intended Use vs Evaluations ISO/FDA Test Chart ACT Annual Meeting Slide 17
  • 18. Toxicity Testing Devices Vs Drugs/Biologics Devices Biologics Drugs Cytotoxicity - - Skin Sensitization - - Irritation (dermal or ocular) - - Intracutaneous Reactivity √* √* Pyrogenicity √ √ Genotoxicity √ √ Acute Toxicity √ √ Subchronic Toxicity Chronic Toxicity √ √ Carcinogenicity √ √ Reproductive Toxicity √ √ Implantation - - *with intracutaneous or intramuscular administration, especially for vaccines and Included in general tox study design not as a separate study like for devices ACT Annual Meeting Slide 18
  • 19. Biocompatibility • Biomaterials or devices interact with tissues and produce some level of reaction in contrast to drugs/biologics • The major category of tissue reactions is the “Foreign Body Response” or FBR • This can be modified by surgical trauma or implant procedure and also things like; • sutures used in the placement of the device eg. tissue mesh • contaminates from manufacturing and pyrogens • Can be very complicated and a good pathologist can be important to avoiding a bad outcome ACT Annual Meeting Slide 19
  • 20. Summary • Enormous diversity of devices; surgical gloves to stents, • Efficacy does not involve chemical action or metabolism— you are not thinking about other pharmacological sites of action as a source of adverse events or toxicity • 98% of the time a 510(k) which may lead to “FDA-release”, not approval, the latter reserved for successful PMA • A predicate device is the reference • The goal is to establish substantial equivalence • No clinical trials to prove safety and efficacy majority of the time • FDA generally relies, in part, on FDA’s prior determination that a reasonable assurance of safety and effectiveness exists for the predicate device. ACT Annual Meeting Slide 20
  • 21. Summary • Biological Hazard/Risk Assessment • The toxicologist can have a unique role in the 510(k) process • Assessing the toxicological risk of the leachables and contaminants that are well defined-Risk Analysis : ISO-10992-17 • In vitro and in vivo Biocompatibility testing including systemic tox in animals  ISO-10993 the Holy Grail with additions by FDA  Extent of testing depends on route and duration of exposure ACT Annual Meeting Slide 21