Call Girl Surat Madhuri 7001305949 Independent Escort Service Surat
Calidad del agua en el laboratorio
1. 1
Water in the Clinical Laboratory
Mikael Cleverstam
WW Clinical Business Manager
Role of water in Clinical Diagnostic
Purification technologies basics
Delivering water to the clinical analyzer
2. 2
Putting it all together
PatientPatient
ResultsResults
WaterWater
QualityQuality
DiagnosticDiagnostic
InstrumentsInstruments
MedicalMedical
TechnologistTechnologist
Quality ControlQuality Control
Diagnostic instruments
Assay developmentPatient results
Troubleshooting Your analyzer
Water quality as part
of the quality process
CLSI New standards
Water purification
Quality system
3. 3
Normal seen Problems
Frequent Calibrations
High CV%
Fluctuation in quality results over the day/week/month
Interfered assays
Calcium Interfered by rocks, leaves, geology
ALP Interfered by biofilm, detergent, rocks
CK Interfered by water treatment
Amylase Interfered by citrus fruit, detergents, leaves
LD Interfered by effluent, leaves, H2O2
Phosphorus Interfered by citrus fruit, leaves
Iron Interfered by rocks, leaves, detergents
Magnesium Interfered by citrus fruit
Triglycerides Interfered by plastics, chemicals
Urea Interfered by citrus fruit, water treatment
Troponin I Interfered by biofilm
4. 4
Water for Clinical Analysers
Cuvette washing
Tubing and probe rinsing
Reagent and buffer reconstitution
Dilution
Water Baths
5. 5
Clinical Analyzers
Features and benefits of automation
Precision optical systems for accuracy in testing
Automatic sampling and dilutions modes
Real time alerts to patient and QC failures
Improved software alerts end user to mechanical failures
Cost benefits
Workflow efficiency and high speed through put
Instrumentation targeted to reduce operating cost with more
efficient technology
Reduced operator interface
DiagnosticDiagnostic
InstrumentsInstruments
6. 6
Assay Development
Measuring chemical changes in the body for diagnosis, therapy and
prognosis has resulted in new assay development
Multiple method testing on a single analyzer
Current research methodologies for infectious disease and tumor
marker’s are moving from research labs in universities to the clinical
laboratory
Complex methodologies are being fully automated for more routine use
DiagnosticDiagnostic
InstrumentsInstruments
7. 7
Unique Challenges for Medical Technologist
Verification of final clinical results to be accurate and precise are
determined by Medical Technologist
Clinical decisions are not solely made on the test result, but in
conjunction with the patient’s history and symptoms
Software alerts, QC reviews, calibration must all be within stated limits
before results are released
Troubleshooting instrument problems result in production delays, are
costly and non-productive activities that must be performed and
documented
➙ Try to avoid diagnostic instrument service because it is expensive
MedicalMedical
TechnologistTechnologist
8. 8
Reviewing Patient Results
All analytical and pre-analytical factors must be reviewed and
documented
Medical Technologist must review all test results
If results are flagged, troubleshooting the cause is necessary
MedicalMedical
TechnologistTechnologist
PatientPatient
ResultsResults
Quality ControlQuality Control
DiagnosticDiagnostic
InstrumentsInstruments
9. 9
Troubleshooting procedures
Sample handling procedures confirmed
Quality control must be reviewed
Shifts and trends
Peer group
Previous data
Assay
Reagent issue
Calibrator stability
Mechanical
Instrument malfunction
Error codes
If above solutions do not correct the erroneous result, further
troubleshooting must identify cause before results can be released to
physicians
➙ Delayed patient treatment.
MedicalMedical
TechnologistTechnologist
PatientPatient
ResultsResults
Quality ControlQuality Control
DiagnosticDiagnostic
InstrumentsInstruments
11. 11
Water Quality
Quality results are dependent upon reliable instrumentation and
known water quality
Analytical factors need to be controlled and optimized to reduce the
number of test failures, failed calibrations, and high blanks that can
contribute to erroneous patient results
Maintenance of high purity water system is essential to reliable results
WaterWater
QualityQuality
12. 12
Understanding Water Quality
and Methodology
Water should be considered a bulk reagent on any analyzer
The high purity water system is a separate unit, not monitored by
diagnostic software on the clinical analyzer
The unique properties of water if not processed and monitored can
produce subtle changes in assay methods
These changes in water quality can lead to erratic and inconsistent
results
The quality of water required or its impact on the testing method
is often not considered until the purchase is complete
WaterWater
QualityQuality
13. 13
Diagnostic Dilemma
Smaller sample size and reaction vessel are subjected
to harsher environment
Inevitable build-up of biofilm in instruments, manifolds and tubing
require more frequent decontaminations
but
Less and less time available for maintenance of the instruments
Some sensitive assays can become contaminated with bacteria and
ions
Bacteria release enzymes and ions whose behavior is similar to the
enzymes targeted in the assay method
➙ Increased need to monitor water quality as closely as any other
instrument malfunction
DiagnosticDiagnostic
InstrumentsInstruments
WaterWater
QualityQuality
15. 15
Demonstration of ALP release from bacteria
Correlation between bacteria concentrations and levels of ALP
in water
Bacteria Strain
(identification by 16S rDNA sequencing)
Bacteria level
(x 106
cfu/mL)
ALP concentration
(µUnit/µL)
Sphingomonas paucimobilis
Caulobacter crescentus
Ralstonia pickettii
29.2
9.7
29.5
6.22
9.95
8.29
DiagnosticDiagnostic
InstrumentsInstruments
WaterWater
QualityQuality
17. 17
CLSI Water Quality Standards
New Standards released July 2006 (C3-A4 Vol. 26 No. 22)
Nomenclature Type I,II,III has been replaced with purity types that
provide more meaningful parameters
CLRW (Clinical Laboratory reagent Water) replaces Type I,II for most applications
IFW (Instrument Feed Water) allows instrument manufacturers to clarify specifications
for their particular methods
SRW ( Special Reagent Water) may be specified for specific applications when
additional parameter are needed to insure water quality
Autoclave and wash water will meet the requirements of previously classified Type III
Complete review of the document should be done when considering
new applications to insure the contaminants found in the source water
do not become an issue
WaterWater
QualityQuality
Quality ControlQuality Control
18. 18
Water Contaminants
Water: H2O …. and some other things
Purification technologies
H
H
H
H
Presence of contaminants
Particles
Gases
Microorganisms
Ions
Organics
19. 19
Protecting the Water Purification Unit:
Pretreatment cartridge
Due to the difference in water
quality around the world, additional
pretreatment cartridges are
required.
The cartridges provide protection
and insure good performance of the
reverse osmosis membrane
The pretreatment packs include 0. 5
micron filter (1) to remove particles
and activated carbon (2) to remove
chlorine
The activated carbon is
impregnated with a small level of
silver to prevent bacterial growth.
Example of a
pretreatment cartridge
21. 21
Technology Insight: Electro-Deionization
Resistivity: > 10 MΩ.cm
TOC: < 30 ppb
No need for regeneration
A - Anionic Membrane
C - Cationic Membrane
A C A C
CathodeCathode
Na+
Na+
H+
H+
OH-
Cl-
Na+
ProductProduct
+ Cl-
Cl-
Cl-
Cl-
Reverse Osmosis WaterReverse Osmosis Water
10 - 2010 - 20 µµS/cmS/cm
Na+
-AnodAnod
ee
RejectReject
Na+
OH-
EDI module
- Ion selective membranes
- Ion exchange resins
- Continuous current
22. 22
Filters – Bacteria Removal
Screen 0.2 µm filters
Designed for the removal of particles and microorganisms from
liquids and gases.
Use of PVDF membranes, provide high flow rates and
throughputs, low extractables, broad chemical compatibility and
the lowest protein binding of any membrane available.
23. 23
Ultrafiltration
Cut-off: 5 KDa to 20 KDa
Removes bacterial by-products such as most proteins and
macromolecules (e.g. endotoxins)
Utilized for immunochemistry assays
Immunoenzyme assays based on reporter enzymes (alkaline
phosphatase, ALP) are sensitive to ALP released by bacteria
Also filters bacteria
24. 24
Storage
CLRW water with a resistivity >10 megohm-cm cannot be stored
because ionic and organic contamination will leach from the
atmosphere and container materials in which it is stored.
CLRW water should be used as it is produced
Stored water is never as pure as when it is made
Storage of water enhances bacterial contamination
Containers need to be cleaned thoroughly between refilling.
Carboys, tanks, bottles
Notorious source of contamination since we often refill them without
thoroughly cleaning them when they are emptied
Some plastic materials out-gas polymers and plasticizers, and
these end up in the water
25. 25
Water Purification Unit
Feed
water
To analyzer
Pretreatment
cartridge
Pump
Reverse Osmosis
cartridge
Electrodeionization
module
Drain
UV
Germicidal
Ion exchange
resins
Tank
Simplified flow schematic combining purification technologies
The electrodeionization module is not present in some purification units
Resistivity
cell
26. 26
Connecting the Water Purification Unit to the
Clinical Analyzer
Water is delivered in its purified state to a harsh environment
within the chemistry analyzer bottle
Water bottles inside analyzer are not frequently decontaminated
Electronics, mechanical hardware, pumps all create heat within the
analyzer cabinet, thus raising the interior temperature of the water
bottle.
Increased temperatures enhance the growth of bacteria and biofilm
within the instruments manifolds and tubings.
A biofilm is an agglomeration of particulates, organic substances and bacteria on the inner surface of the reservoir.
As indicated earlier, bacteria do not “like” to grow alone, in the middle of water . They rather tend to stick to a support . This is what happens ususally inside reservoirs : some partciulates and organic substances bind to the surface; bacteria attach themselves to the same point, and a biofilm grows and becomes thicker with time, releasing from time to time bacteria in the stored water.
NB : just to give an order of magnitude about what represents 1 cfu/ml. For a simple calculation, let’s consider that a bacteria occupies 0.1 um 3 (it usually occupies a smaller volume : Pseudomonas Diminuta volume is about 0.01 um3- this is just to make the demonstration easier).A 0.1 um3 volume corresponds to 0.1 x (10*-6m)3 = 10*-19 m3. Therefore, the volume occupied by 1 billion bacteria would be 10*-19 x 10*9 = 10*-10 m3 = 10*-7L = 10*-4 ml= 0.0001 ml . This means that 1ppb (1 part per billion ) bacteria corresponds to 1 bacteria per 0.0001 ml , or 10,000 bacteria per ml.