Biomechanics is concerned with the functions and structures of the musco-skeletal system and movements of biological systems. Medical products to be tested are prostheses, osteosynthetic implants, trauma products and orthopedic implants.
2. Standards:
ISO 7206-4:2002; ASTM F 1612-95
(determination of endurance properties of
stemmed femoral components, 5 million cycles,
15 Hz)
ISO 7206-8:1995
(specification for endurance of stemmed
femoral components with application of torsion)
simulates and measures the natural wear of the
joint in a shorter time for different body weights
and types/level of activities
allows designers to develop improved prostheses
with a reduced chance of implant failure for
patients of different weight and activity levels
10kN or 25kN servo-hydraulic machine, optionally
with temperature-conditioned water-bath
Biomechanics & orthopaedics
Durability of hip-joint implants
Amsler HC 10 Hip joint
2015Testing of biomechanics & orthopaedics 2
3. ISO 7206-9:1994
(method for determination of resistance to
torque of head fixation of stemmed femoral
components).
This test involves determining resistance to
torque between the femoral head and the
femoral stem.
Allows designers to develop improved fixation
between
the two components.
Zwick 20kN static biaxial testing machine with
500 Nm torque drive.
A static compressive preload is applied, then
the transverse torque is increased until the
head begins to rotate.
Biomechanics & orthopaedics
Zwick Z020 testing machine for compression and torque testing on
femoral head and stem of a hip implant
Component fixing in hip implants
2015Testing of biomechanics & orthopaedics 3
4. ASTM F 1717 (method for static and fatigue
testing of spinal implant assemblies in a
vertebrectomy model)
HC10 servo-hydraulic testing machine
integrated low-noise hydraulic pump
flexible fixture and specimen mounting via
T-slotted table
easy, safe operation via proven software-
controller combination
ASTM F 1717-04 fatigue test
6 samples
up to 5 million cycles
frequency up to 5 Hz
customer-specific: additional torsion
drive
Biomechanics & orthopaedics
Durability of spinal implants
Zwick HCT25 testing machine for compression tests on a spinal
vertebrectomy model
2015Testing of biomechanics & orthopaedics 4
5. Mmax: 50 Nm
Dynamic: +- 20°
Static: 90°
Biomechanics & orthopaedics
Torsion test kit for static tests Torsion test kit for fatigue tests
Testing methods on spinal implants ASTM F2077
Torsion (static & dynamic)
2015Testing of biomechanics & orthopaedics 5
7. Biomechanics & orthopaedics
Testing trauma products
Fatigue in nails (ASTM F1264)
Tightening/unscrewing test on bone screwsFatigue in angled plates (ASTM F384) Screw pull-out tests (ASTM F543)
2015Testing of biomechanics & orthopaedics 7
8. Biomechanics & orthopaedics
This specification includes two test
methods:
Single Cycle Bend Text
Bending Fatigue Test
optionally with displacement
transducer for direct
measurement of deflection
on the osteosynthesis specimen.
Fatigue test on plates acc. to ASTM F 382
2015Testing of biomechanics & orthopaedics 8
9. Biomechanics & orthopaedics
testing components,
implants, functions
lowest calibrated force
0.02 N
Combination with laserXtens Compact
PrecisionLine Vario Stapes prosthesis
(Heinz Kurz GmbH Medizintechnik)
Testing micro-components
2015Testing of biomechanics & orthopaedics 9
10. Contact
Zwick GmbH & Co. KG
August-Nagel-Str. 11
89079 Ulm, Germany
Tel: +49 7305 - 10-0