TruSpec Elemental Determinators
Specification Sheet
Carbon: 50 ppm or 0.005% to 50%
Instrument Range @ 500 mg* Hydrogen: 200 ppm or 0.02% to 50%
Nitrogen: 80 ppm or 0.008% to 100%
Carbon: 25 ppm or 0.5% RSD (whichever is greater)
Precision Range @ 500 mg Hydrogen: 100 ppm or 1% RSD (whichever is greater)
Nitrogen: 40 ppm or 0.5% RSD (whichever is greater)
Readability 0.0001
Analysis Time 4 minutes
Sample Size Up to 1 gram
Carbon/Hydrogen: Optimized, Low-Noise, Non-Dispersive Infrared Absorption
Detection Method Nitrogen: Optimized, Low-Drift Thermal Conductivity (TC Cell) Detector
Carrier: Helium (99.99% pure) @ 35 psi (2.4 bar) ±10%
Gases Required Combustion: Oxygen (99.99% pure) @ 35 psi (2.4 bar) ±10%
Pneumatic: Compressed air (source must be oil and water free); 40 psi (2.8 bar) ±10%
Furnace Resistance furnace; both primary and afterburner; up to 1050oC
Autoloader 30-position (stackable to 120 samples)
o o o o
Environmental Conditions Operating Temp: 15 C to 30 C (59 F to 86 F) Humidity: 20% to 80%, non-condensing
Dimensions 31 in H x 27 in. W x 28 in. D (79 x 69 x 71 cm)
Weight (approx.) 273 lb (124 kg) Shipping Weight (approx.): 324 lb. (147 kg)
Electrical Power Requirements 230 V~ (±10%; at max load), 50/60 Hz, single phase, 12 A; 9,500 BTU/hr
Part Numbers
®
TruSpec N (Nitrogen only) with PC tower, Windows -based operating software,
TRSNC flat panel monitor, and autoloader
®
TruSpec CN (Carbon/Nitrogen) with PC tower, Windows -based operating software,
TRSCNC flat panel monitor, and autoloader
TruSpec CHN (Carbon/Hydrogen/Nitrogen) with PC tower, Windows®-based operating
TRSCHNC software, flat panel monitor, and autoloader
*Adjusting sample size may extend instrument range.
†
Delivering the Right Results
Allow a 6-inch (15 cm) minimum access area around all units.
V~ denotes VAC.