Introduction
The United States Environmental Protection
Agency (USEPA) Photochemical Assessment
Monitoring Stations (PAMS) are air quality
monitoring locations that are used to
monitor pollutant compounds in air that contribute to ground level (bad) ozone. These nonpolar
compounds are typically found in the 0.1 to tens ppbv, if not greater in some cases. Since these
levels are too low to detect directly, a concentration technique such as cryotrapping or chemical
adsorbent trapping needs to be used. One of the main issues with any sub-ambient temperature
method is the negative effect of moisture in the air sample. Quantities up to tens of milligrams of
water may be present and this can cause havoc with the chromatographic system, such as the
water freezing with the formation of ice causing a flow restriction or blockage. Water can also
cause poor peak shape/retention of analytes on certain columns such as Alumina PLOT. Several
different techniques can be used for the effective removal of water from a gas sampling stream,
prior to preconcentration, such as, drying agents, semipermeable membranes and reduced
temperatures (used to condense out the water). One well accepted technique uses a Nafion®
(TM,
DuPont™
) membrane to dry the incoming sample stream, and is very effective. The membrane
permeability is very dependent on the analyte, generally speaking, and nonpolars are retained while
water and certain polars are lost across the membrane.
This work is focused on humidity effects on low level sub-ppbv (0.3-1 ppbv) PAMS targets compared
to the work done by G. Broadway et al¹ who studied humidity effects at a higher 10 ppbv.
PAMS: A Study of
Performance in Low and
High Humidity Environments
A P P L I C A T I O N N O T E
Author:
PerkinElmer Canadian
Chromatographic Application Lab
PerkinElmer, Inc.
Woodbridge, ON Canada
Gas Chromatography
2
Experimental
PAMS standard was purchased from Linde Canada at the USEPA
recommended concentration of 3-10 ppbv and was used as the
stock standard. The PAMS components dilutions were dynamically
created in real time using a custom made on-line diluter, designed
and built by the author, using UHP Nitrogen (LN2) as the diluent.
The gas was purified using an advanced triple bed carrier gas
purifier filter.
The on-line diluter utilized two 2000 mL/min. MKS Mass-Flo Mass
flow controllers (MFC). One was used to control pure nitrogen
diluent and the other was to control the flow to a 10 um stainless
steel frit, inserted into a 2L HPLC solvent bottle, ¾ filled with 18
megohm DI water. The bubbler MFC outlet was teed in to the
pure nitrogen MFC output. To eliminate the possibility of trace
level contaminates in the water and/or associated tubing/fittings/
MFC, since we are running at sub-ppbv levels, the total diluent
flow (both Humid and Dry) was scrubbed of organics by the use
of two Soil Vapour Intrusion tubes (SVI, thermal desorption) in
series post blending of the diluent. The now clean diluent was
blended with the PAMS standard (via a 200 mL/min. MKS MFC)
into a ¼” SS tubing. To reduce the effects of laminar flow and
promote mixing, the tubing was reduced to 1/8” SS (6 ft). This
tubing was terminated into a SS 1/8” Tee with one arm going
toward the on-line accessory approx. 1 ft. away with a 15 mL/
min. flow. The remaining arm was plumbed with ¼” Bev-A-Line,
as it was downstream of the sampling accessory, and at the end
of this line a custom designed and manufactured humidity
accessory was installed. A block diagram of the system is shown
in Figure 1.
The custom designed humidity accessory utilized a Honeywell
Humidcon™ HIH9130-021 with a 0-100% RH range, with
+/-1.7 accuracy, +/-0.6 °C temperature reading communication
via I2C. The I2C data was read via a customized Atmel®
328P
microprocessor and the resulting data was displayed via I2C with
a 16 x 2 LCD screen. The % RH and temperature was updated at
10 Hz. Similarly, the MFC were controlled by a custom programmed
Atmel 328P, using SPI controlled MCP4821 12-bit DAC. Since the
MCP4821 max output is 4096 mV and the MFC requires 0-5000
mV, a TLC2272 Rail to Rail Op Amp was used to scale the DAC
0-4096 mV output by 1.25. A rotary encoder with momentary
push buttons was used for the setpoint and MFC selection. The
MFC readback was read by the ATMega328P 10-bit internal ADC
directly without buffering. The readbacks/setpoints were displayed
on a 16 x 2 LCD via I2C. An electronic block diagram showing
the basic components of the MFC control is shown in Figure 2.
Figure1.Diagramofadynamicdiluter.
Figure2.SchematicofMFCcontrolelectronics.
Table1.SystemConditionsforOzonePrecursorAnalysis.
GC Clarus®
580 GC with heartcut option
Column Ozone
Precursor Column Set
50 m x 0.22 mm x 1 μm dimethyl siloxane column
Part number 04978179
50 m x 0.32 mm alumina/Na2SO4 PLOT column
Part number N6301107
Oven Conditions
46 °C for 15 min., then 5 °C/min. to 170 °C,
then 15 °C/minute to 200 °C and hold for 6 min.
(48-min. chromatogram)
Detector Dual FID at 250 °C
Carrier Gas 42 psig helium at the ATD
16 psig helium at the midpoint
ATD
Thermal Desorber TurboMatrix™
TD with online
sampling accessory
Trap
Air Toxics trap packed with carbonaceous adsorbents
Low -30 °C
High 325 °C
Transfer Line 200 °C
Operating Mode On-Line at 15 mL/min. for 40 min.
Split
Inlet Off
Outlet 2 mL/min.
Data Handling TotalChrom™
3
MFC Settings and Flows
MFC settings were 10 bit (0-1023) for the 0-5 V analog input
for the MFC via the 12 bit DAC. The MFC flows were calibrated
using a NIST traceable PE1000 electronic flow meter. The MFCs
used were identified as MFC#1 (for Humid Flow), MFC#2 (Dry
Flow) and MFC#3 (Gas Standard). Unless otherwise noted the
MFC were set as follows:
MFC#1, set to 100, produced a measured flow of 228 mL/min.
direct at MFC output, and 263 mL/min. of humid flow at 77% RH
after humidification. Since the extra volume of the water vapor
at RT contributes, it is believed the difference in flow with the
humidifier attached is a result of the extra backpressure due
to humidifier.
MFC#2, set to 100, produced a flow of 193 mL/min.
MFC#3, set to 100, produced a flow of 34.7 mL/min.
Dilution of standard with dry nitrogen =
(193+34.7)/34.7 = 1:6.56
Dilution of standard with humidified nitrogen =
(263+34.7)/34.7 = 1:8.58
Humidity Notes
Since the generated humidity is passed through two SVI tubes
in series, the humidity needs time to stabilize as each tube has a
moisture capacity of approx. 35 mg each. This effect is described
in detail by Snow et al². In this study, typically one hour
stabilization time was used prior to sampling.
Results
Both the 77% RH and dry dilutions of the PAMS standard give
similar chromatograms in both the PLOT and BP-1 separations as
shown in Figures 3 and 4 respectively. Figure 5 is an overlay of
four sequential chromatographic runs highlighting excellent
repeatability at low levels.
Figure3.PLOTchromatogramofPAMSstandardatapprox.0.3-1ppbv.Topchromatogramis77%RHandbottomisapproximately0%RH.
4
Figure4.BP-1chromatogramofPAMSstandardatapprox.0.3-1ppbv.Topchromatogramis77%RHandbottomisapproximately0%RH.
Figure5.ExpandedsectionofBP-1columnseparationwithfouroverlaid77%RHsamples.
For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs
Copyright ©2015, PerkinElmer, Inc. All rights reserved. PerkinElmer®
is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners.
012297_01	PKI
PerkinElmer, Inc.
940 Winter Street
Waltham, MA 02451 USA	
P: (800) 762-4000 or
(+1) 203-925-4602
www.perkinelmer.com
References
1.	“Ozone Precursor Analysis Using a Thermal Desorption-GC
System” by G. Broadway and A. Tipler available for download at
www.perkinelmer.com.
2.	Patient Application US 20130180315 A1 Devices, systems and
methods for purging and loading sorbent tubes by Snow et al.
Conclusion
The ozone precursor system performance at the 0.3 ppbv-1 ppbv,
with a PAMS standard under high humidity conditions, are virtually
identical to that under low/zero humidity. The chromatographic
peak shapes and intensities show that high humidity has little/no
negative effect on the analysis of PAMS target compounds at low
levels. This enables the use of dry gaseous standards to be used
without the laborious preparation of humidified standards.

More Related Content

What's hot

AICHE Senior Design Project
AICHE Senior Design ProjectAICHE Senior Design Project
AICHE Senior Design ProjectJonathan Sherwin
 
Stamolys ca71 ph endress+hauser datasheet-phosphate analyzer
Stamolys ca71 ph endress+hauser datasheet-phosphate analyzerStamolys ca71 ph endress+hauser datasheet-phosphate analyzer
Stamolys ca71 ph endress+hauser datasheet-phosphate analyzerENVIMART
 
pharmacoepial methods-physical tests
pharmacoepial methods-physical testspharmacoepial methods-physical tests
pharmacoepial methods-physical testsSiva Chaitanya Addala
 
Stamolys ca71 no endress+hauser datasheet-nitrite analyzer
Stamolys ca71 no endress+hauser datasheet-nitrite analyzerStamolys ca71 no endress+hauser datasheet-nitrite analyzer
Stamolys ca71 no endress+hauser datasheet-nitrite analyzerENVIMART
 
Stamolys ca71 mn endress+hauser datasheet-manganese analyser
Stamolys ca71 mn endress+hauser datasheet-manganese analyserStamolys ca71 mn endress+hauser datasheet-manganese analyser
Stamolys ca71 mn endress+hauser datasheet-manganese analyserENVIMART
 
Stamolys ca71 fe endress+hauser datasheet-iron analyzer
Stamolys ca71 fe endress+hauser datasheet-iron analyzerStamolys ca71 fe endress+hauser datasheet-iron analyzer
Stamolys ca71 fe endress+hauser datasheet-iron analyzerENVIMART
 
Liquid chlorine incident
Liquid chlorine incidentLiquid chlorine incident
Liquid chlorine incidentUsama Khan
 
Rapid LC-MS/MS Analysis of PFCs in Non-drinking Water Matrices
Rapid LC-MS/MS Analysis of PFCs in Non-drinking Water MatricesRapid LC-MS/MS Analysis of PFCs in Non-drinking Water Matrices
Rapid LC-MS/MS Analysis of PFCs in Non-drinking Water MatricesShimadzu Scientific Instruments
 
AIChE Final Design Report Excerpt
AIChE Final Design Report ExcerptAIChE Final Design Report Excerpt
AIChE Final Design Report Excerptaarittha
 
Scion instruments refinery gas analyzer saranalab mandiri analitika
Scion instruments refinery gas analyzer saranalab mandiri analitikaScion instruments refinery gas analyzer saranalab mandiri analitika
Scion instruments refinery gas analyzer saranalab mandiri analitikaSaranaLaboratorium
 
Toc analyser, microprocessor controlled
Toc analyser, microprocessor controlledToc analyser, microprocessor controlled
Toc analyser, microprocessor controlledhydrologyproject0
 
Types of Industrial Process Analyzers
Types of Industrial Process AnalyzersTypes of Industrial Process Analyzers
Types of Industrial Process AnalyzersAttiya Rehman
 
Scion instruments transformer oil gas analyzer by sarana lab mandiri analitika
Scion instruments transformer oil gas analyzer by sarana lab mandiri analitikaScion instruments transformer oil gas analyzer by sarana lab mandiri analitika
Scion instruments transformer oil gas analyzer by sarana lab mandiri analitikaSaranaLaboratorium
 
A Quantitative Analysis of Ethanol in Beer using Gas Liquid Chromatography – ...
A Quantitative Analysis of Ethanol in Beer using Gas Liquid Chromatography – ...A Quantitative Analysis of Ethanol in Beer using Gas Liquid Chromatography – ...
A Quantitative Analysis of Ethanol in Beer using Gas Liquid Chromatography – ...Ryan Baker
 
Unit 6 Water Content Determination and Moisture analysis
Unit 6   Water Content Determination and Moisture analysisUnit 6   Water Content Determination and Moisture analysis
Unit 6 Water Content Determination and Moisture analysisMalou Mojares
 
PHARMACEUTICAL APPLICATIONS OF GAS CHROMATOGRAPHY(GC), PHARMACEUTICAL ANALYS...
 PHARMACEUTICAL APPLICATIONS OF GAS CHROMATOGRAPHY(GC), PHARMACEUTICAL ANALYS... PHARMACEUTICAL APPLICATIONS OF GAS CHROMATOGRAPHY(GC), PHARMACEUTICAL ANALYS...
PHARMACEUTICAL APPLICATIONS OF GAS CHROMATOGRAPHY(GC), PHARMACEUTICAL ANALYS...Dr. Ravi Sankar
 

What's hot (19)

AICHE Senior Design Project
AICHE Senior Design ProjectAICHE Senior Design Project
AICHE Senior Design Project
 
Stamolys ca71 ph endress+hauser datasheet-phosphate analyzer
Stamolys ca71 ph endress+hauser datasheet-phosphate analyzerStamolys ca71 ph endress+hauser datasheet-phosphate analyzer
Stamolys ca71 ph endress+hauser datasheet-phosphate analyzer
 
pharmacoepial methods-physical tests
pharmacoepial methods-physical testspharmacoepial methods-physical tests
pharmacoepial methods-physical tests
 
Stamolys ca71 no endress+hauser datasheet-nitrite analyzer
Stamolys ca71 no endress+hauser datasheet-nitrite analyzerStamolys ca71 no endress+hauser datasheet-nitrite analyzer
Stamolys ca71 no endress+hauser datasheet-nitrite analyzer
 
Stamolys ca71 mn endress+hauser datasheet-manganese analyser
Stamolys ca71 mn endress+hauser datasheet-manganese analyserStamolys ca71 mn endress+hauser datasheet-manganese analyser
Stamolys ca71 mn endress+hauser datasheet-manganese analyser
 
Stamolys ca71 fe endress+hauser datasheet-iron analyzer
Stamolys ca71 fe endress+hauser datasheet-iron analyzerStamolys ca71 fe endress+hauser datasheet-iron analyzer
Stamolys ca71 fe endress+hauser datasheet-iron analyzer
 
Liquid chlorine incident
Liquid chlorine incidentLiquid chlorine incident
Liquid chlorine incident
 
Rapid LC-MS/MS Analysis of PFCs in Non-drinking Water Matrices
Rapid LC-MS/MS Analysis of PFCs in Non-drinking Water MatricesRapid LC-MS/MS Analysis of PFCs in Non-drinking Water Matrices
Rapid LC-MS/MS Analysis of PFCs in Non-drinking Water Matrices
 
AIChE Final Design Report Excerpt
AIChE Final Design Report ExcerptAIChE Final Design Report Excerpt
AIChE Final Design Report Excerpt
 
Clora-Online
Clora-OnlineClora-Online
Clora-Online
 
Scion instruments refinery gas analyzer saranalab mandiri analitika
Scion instruments refinery gas analyzer saranalab mandiri analitikaScion instruments refinery gas analyzer saranalab mandiri analitika
Scion instruments refinery gas analyzer saranalab mandiri analitika
 
Toc analyser, microprocessor controlled
Toc analyser, microprocessor controlledToc analyser, microprocessor controlled
Toc analyser, microprocessor controlled
 
Karl fischer
Karl fischerKarl fischer
Karl fischer
 
Types of Industrial Process Analyzers
Types of Industrial Process AnalyzersTypes of Industrial Process Analyzers
Types of Industrial Process Analyzers
 
Scion instruments transformer oil gas analyzer by sarana lab mandiri analitika
Scion instruments transformer oil gas analyzer by sarana lab mandiri analitikaScion instruments transformer oil gas analyzer by sarana lab mandiri analitika
Scion instruments transformer oil gas analyzer by sarana lab mandiri analitika
 
A Quantitative Analysis of Ethanol in Beer using Gas Liquid Chromatography – ...
A Quantitative Analysis of Ethanol in Beer using Gas Liquid Chromatography – ...A Quantitative Analysis of Ethanol in Beer using Gas Liquid Chromatography – ...
A Quantitative Analysis of Ethanol in Beer using Gas Liquid Chromatography – ...
 
Unit 6 Water Content Determination and Moisture analysis
Unit 6   Water Content Determination and Moisture analysisUnit 6   Water Content Determination and Moisture analysis
Unit 6 Water Content Determination and Moisture analysis
 
PHARMACEUTICAL APPLICATIONS OF GAS CHROMATOGRAPHY(GC), PHARMACEUTICAL ANALYS...
 PHARMACEUTICAL APPLICATIONS OF GAS CHROMATOGRAPHY(GC), PHARMACEUTICAL ANALYS... PHARMACEUTICAL APPLICATIONS OF GAS CHROMATOGRAPHY(GC), PHARMACEUTICAL ANALYS...
PHARMACEUTICAL APPLICATIONS OF GAS CHROMATOGRAPHY(GC), PHARMACEUTICAL ANALYS...
 
Water analysis from Metrohm
Water analysis from MetrohmWater analysis from Metrohm
Water analysis from Metrohm
 

Viewers also liked

Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17PerkinElmer, Inc.
 
The british regulatory system – global mba brazil
The british regulatory system – global mba brazilThe british regulatory system – global mba brazil
The british regulatory system – global mba brazilGlobal MBA Brazil
 
Analysis of Additives in Biodegradable Polymers Using DSA/TOF Mass Spectrometry
Analysis of Additives in Biodegradable Polymers Using DSA/TOF Mass SpectrometryAnalysis of Additives in Biodegradable Polymers Using DSA/TOF Mass Spectrometry
Analysis of Additives in Biodegradable Polymers Using DSA/TOF Mass SpectrometryPerkinElmer, Inc.
 
Detection and Identification of Contaminations in the Manufacturing Process U...
Detection and Identification of Contaminations in the Manufacturing Process U...Detection and Identification of Contaminations in the Manufacturing Process U...
Detection and Identification of Contaminations in the Manufacturing Process U...PerkinElmer, Inc.
 
Cairo Brokerage Conference- Global MBA Brazil
Cairo Brokerage Conference- Global MBA BrazilCairo Brokerage Conference- Global MBA Brazil
Cairo Brokerage Conference- Global MBA BrazilGlobal MBA Brazil
 
Optimization of Cannabis Grows Using Fourier Transform Mid-Infrared Spectroscopy
Optimization of Cannabis Grows Using Fourier Transform Mid-Infrared SpectroscopyOptimization of Cannabis Grows Using Fourier Transform Mid-Infrared Spectroscopy
Optimization of Cannabis Grows Using Fourier Transform Mid-Infrared SpectroscopyPerkinElmer, Inc.
 

Viewers also liked (7)

Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
 
Bit Byte
Bit ByteBit Byte
Bit Byte
 
The british regulatory system – global mba brazil
The british regulatory system – global mba brazilThe british regulatory system – global mba brazil
The british regulatory system – global mba brazil
 
Analysis of Additives in Biodegradable Polymers Using DSA/TOF Mass Spectrometry
Analysis of Additives in Biodegradable Polymers Using DSA/TOF Mass SpectrometryAnalysis of Additives in Biodegradable Polymers Using DSA/TOF Mass Spectrometry
Analysis of Additives in Biodegradable Polymers Using DSA/TOF Mass Spectrometry
 
Detection and Identification of Contaminations in the Manufacturing Process U...
Detection and Identification of Contaminations in the Manufacturing Process U...Detection and Identification of Contaminations in the Manufacturing Process U...
Detection and Identification of Contaminations in the Manufacturing Process U...
 
Cairo Brokerage Conference- Global MBA Brazil
Cairo Brokerage Conference- Global MBA BrazilCairo Brokerage Conference- Global MBA Brazil
Cairo Brokerage Conference- Global MBA Brazil
 
Optimization of Cannabis Grows Using Fourier Transform Mid-Infrared Spectroscopy
Optimization of Cannabis Grows Using Fourier Transform Mid-Infrared SpectroscopyOptimization of Cannabis Grows Using Fourier Transform Mid-Infrared Spectroscopy
Optimization of Cannabis Grows Using Fourier Transform Mid-Infrared Spectroscopy
 

Similar to PAMS: A Study of Performance in Low and High Humidity Environments

Voc72 m compounds analyser datasheet 0
Voc72 m compounds analyser datasheet 0Voc72 m compounds analyser datasheet 0
Voc72 m compounds analyser datasheet 0a1-cbiss
 
ATI-Q46H-83-Datasheet.pdf
ATI-Q46H-83-Datasheet.pdfATI-Q46H-83-Datasheet.pdf
ATI-Q46H-83-Datasheet.pdfENVIMART
 
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17PerkinElmer, Inc.
 
Gas Chromatography laboratory equipments
Gas Chromatography laboratory equipmentsGas Chromatography laboratory equipments
Gas Chromatography laboratory equipmentskimberlabtroninfo
 
4 b = 58 p 45 basic ur plant description 06.09.2021-1
4 b = 58 p  45 basic ur plant description 06.09.2021-14 b = 58 p  45 basic ur plant description 06.09.2021-1
4 b = 58 p 45 basic ur plant description 06.09.2021-1ameermudasar
 
Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...
Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...
Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...PerkinElmer, Inc.
 
Liquiline system ca80 no endress+hauser datasheet-colorimetric analyzer for n...
Liquiline system ca80 no endress+hauser datasheet-colorimetric analyzer for n...Liquiline system ca80 no endress+hauser datasheet-colorimetric analyzer for n...
Liquiline system ca80 no endress+hauser datasheet-colorimetric analyzer for n...ENVIMART
 
Comparison of the Performances of Two Commercial Membranes in Demineralizatio...
Comparison of the Performances of Two Commercial Membranes in Demineralizatio...Comparison of the Performances of Two Commercial Membranes in Demineralizatio...
Comparison of the Performances of Two Commercial Membranes in Demineralizatio...IJRES Journal
 
IRES_ Final_Presenation
IRES_ Final_PresenationIRES_ Final_Presenation
IRES_ Final_PresenationHolden Ranz
 
Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...
Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...
Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...PerkinElmer, Inc.
 
160729 ama instruments_datasheet_gc5000
160729 ama instruments_datasheet_gc5000160729 ama instruments_datasheet_gc5000
160729 ama instruments_datasheet_gc5000European Tech Serv
 
GRC_Poster_2015
GRC_Poster_2015GRC_Poster_2015
GRC_Poster_2015David Gunn
 
Chloromax CCS142 D endress+hauser datasheet free chlorine
Chloromax CCS142 D endress+hauser datasheet free chlorineChloromax CCS142 D endress+hauser datasheet free chlorine
Chloromax CCS142 D endress+hauser datasheet free chlorineENVIMART
 
Cwea Me Hg Presentation Er
Cwea Me Hg Presentation ErCwea Me Hg Presentation Er
Cwea Me Hg Presentation Ernoel_enoki
 
5-MEO-DALT produces short-lived psychedelic effects
5-MEO-DALT produces short-lived psychedelic effects5-MEO-DALT produces short-lived psychedelic effects
5-MEO-DALT produces short-lived psychedelic effectsBRC SHOP
 
Process variables & product recovery in Fermentation
Process variables & product recovery in FermentationProcess variables & product recovery in Fermentation
Process variables & product recovery in FermentationKRATIKA SINGHAM
 
Q46H79S-Total-Chlorine-1.pdf
Q46H79S-Total-Chlorine-1.pdfQ46H79S-Total-Chlorine-1.pdf
Q46H79S-Total-Chlorine-1.pdfENVIMART
 

Similar to PAMS: A Study of Performance in Low and High Humidity Environments (20)

Voc72 m compounds analyser datasheet 0
Voc72 m compounds analyser datasheet 0Voc72 m compounds analyser datasheet 0
Voc72 m compounds analyser datasheet 0
 
ATI-Q46H-83-Datasheet.pdf
ATI-Q46H-83-Datasheet.pdfATI-Q46H-83-Datasheet.pdf
ATI-Q46H-83-Datasheet.pdf
 
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
 
Recent advances in hplc and gc
Recent advances in hplc and gcRecent advances in hplc and gc
Recent advances in hplc and gc
 
Fruit juices and puree technology
Fruit juices and puree technologyFruit juices and puree technology
Fruit juices and puree technology
 
Gas Chromatography laboratory equipments
Gas Chromatography laboratory equipmentsGas Chromatography laboratory equipments
Gas Chromatography laboratory equipments
 
4 b = 58 p 45 basic ur plant description 06.09.2021-1
4 b = 58 p  45 basic ur plant description 06.09.2021-14 b = 58 p  45 basic ur plant description 06.09.2021-1
4 b = 58 p 45 basic ur plant description 06.09.2021-1
 
Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...
Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...
Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...
 
Liquiline system ca80 no endress+hauser datasheet-colorimetric analyzer for n...
Liquiline system ca80 no endress+hauser datasheet-colorimetric analyzer for n...Liquiline system ca80 no endress+hauser datasheet-colorimetric analyzer for n...
Liquiline system ca80 no endress+hauser datasheet-colorimetric analyzer for n...
 
Comparison of the Performances of Two Commercial Membranes in Demineralizatio...
Comparison of the Performances of Two Commercial Membranes in Demineralizatio...Comparison of the Performances of Two Commercial Membranes in Demineralizatio...
Comparison of the Performances of Two Commercial Membranes in Demineralizatio...
 
IRES_ Final_Presenation
IRES_ Final_PresenationIRES_ Final_Presenation
IRES_ Final_Presenation
 
Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...
Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...
Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the...
 
160729 ama instruments_datasheet_gc5000
160729 ama instruments_datasheet_gc5000160729 ama instruments_datasheet_gc5000
160729 ama instruments_datasheet_gc5000
 
GC Agilent .pptx
GC Agilent .pptxGC Agilent .pptx
GC Agilent .pptx
 
GRC_Poster_2015
GRC_Poster_2015GRC_Poster_2015
GRC_Poster_2015
 
Chloromax CCS142 D endress+hauser datasheet free chlorine
Chloromax CCS142 D endress+hauser datasheet free chlorineChloromax CCS142 D endress+hauser datasheet free chlorine
Chloromax CCS142 D endress+hauser datasheet free chlorine
 
Cwea Me Hg Presentation Er
Cwea Me Hg Presentation ErCwea Me Hg Presentation Er
Cwea Me Hg Presentation Er
 
5-MEO-DALT produces short-lived psychedelic effects
5-MEO-DALT produces short-lived psychedelic effects5-MEO-DALT produces short-lived psychedelic effects
5-MEO-DALT produces short-lived psychedelic effects
 
Process variables & product recovery in Fermentation
Process variables & product recovery in FermentationProcess variables & product recovery in Fermentation
Process variables & product recovery in Fermentation
 
Q46H79S-Total-Chlorine-1.pdf
Q46H79S-Total-Chlorine-1.pdfQ46H79S-Total-Chlorine-1.pdf
Q46H79S-Total-Chlorine-1.pdf
 

More from PerkinElmer, Inc.

Detectando a Pré Eclampsia no 1º Trimestre
Detectando a Pré Eclampsia no 1º TrimestreDetectando a Pré Eclampsia no 1º Trimestre
Detectando a Pré Eclampsia no 1º TrimestrePerkinElmer, Inc.
 
As Tecnologias que Definem o Desempenho
As Tecnologias que Definem o DesempenhoAs Tecnologias que Definem o Desempenho
As Tecnologias que Definem o DesempenhoPerkinElmer, Inc.
 
GCMS Portátil: Análises em 5 minutos!
GCMS Portátil: Análises em 5 minutos!GCMS Portátil: Análises em 5 minutos!
GCMS Portátil: Análises em 5 minutos!PerkinElmer, Inc.
 
Nós transformamos complexidade em simplicidade
Nós transformamos complexidade em simplicidadeNós transformamos complexidade em simplicidade
Nós transformamos complexidade em simplicidadePerkinElmer, Inc.
 
Janus G3: Rapidez e Automação que você precisa.
Janus G3: Rapidez e Automação que você precisa.Janus G3: Rapidez e Automação que você precisa.
Janus G3: Rapidez e Automação que você precisa.PerkinElmer, Inc.
 
Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?PerkinElmer, Inc.
 
Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?PerkinElmer, Inc.
 
O Mercado da Carne no Brasil
O Mercado da Carne no BrasilO Mercado da Carne no Brasil
O Mercado da Carne no BrasilPerkinElmer, Inc.
 
KITS de Preparo para Biblioteca NGS
KITS de Preparo para Biblioteca NGSKITS de Preparo para Biblioteca NGS
KITS de Preparo para Biblioteca NGSPerkinElmer, Inc.
 
Escolha um parceiro e não um fornecedor
Escolha um parceiro e não um fornecedorEscolha um parceiro e não um fornecedor
Escolha um parceiro e não um fornecedorPerkinElmer, Inc.
 
Eletroforese Capilar Automatizada
Eletroforese Capilar Automatizada Eletroforese Capilar Automatizada
Eletroforese Capilar Automatizada PerkinElmer, Inc.
 
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OESAnalysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OESPerkinElmer, Inc.
 
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...PerkinElmer, Inc.
 
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...PerkinElmer, Inc.
 
High Quality DNA Isolation Suitable for Ultra Rapid Sequencing
High Quality DNA Isolation Suitable for Ultra Rapid SequencingHigh Quality DNA Isolation Suitable for Ultra Rapid Sequencing
High Quality DNA Isolation Suitable for Ultra Rapid SequencingPerkinElmer, Inc.
 

More from PerkinElmer, Inc. (20)

Detectando a Pré Eclampsia no 1º Trimestre
Detectando a Pré Eclampsia no 1º TrimestreDetectando a Pré Eclampsia no 1º Trimestre
Detectando a Pré Eclampsia no 1º Trimestre
 
É A VEZ DO PRIME
É A VEZ DO PRIMEÉ A VEZ DO PRIME
É A VEZ DO PRIME
 
TREC PARA A VIDA
TREC PARA A VIDATREC PARA A VIDA
TREC PARA A VIDA
 
As Tecnologias que Definem o Desempenho
As Tecnologias que Definem o DesempenhoAs Tecnologias que Definem o Desempenho
As Tecnologias que Definem o Desempenho
 
GCMS Portátil: Análises em 5 minutos!
GCMS Portátil: Análises em 5 minutos!GCMS Portátil: Análises em 5 minutos!
GCMS Portátil: Análises em 5 minutos!
 
Nós transformamos complexidade em simplicidade
Nós transformamos complexidade em simplicidadeNós transformamos complexidade em simplicidade
Nós transformamos complexidade em simplicidade
 
Janus G3: Rapidez e Automação que você precisa.
Janus G3: Rapidez e Automação que você precisa.Janus G3: Rapidez e Automação que você precisa.
Janus G3: Rapidez e Automação que você precisa.
 
Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?
 
Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?
 
Aspirina Vs Pré Eclampsia
Aspirina Vs Pré Eclampsia Aspirina Vs Pré Eclampsia
Aspirina Vs Pré Eclampsia
 
O Mercado da Carne no Brasil
O Mercado da Carne no BrasilO Mercado da Carne no Brasil
O Mercado da Carne no Brasil
 
KITS de Preparo para Biblioteca NGS
KITS de Preparo para Biblioteca NGSKITS de Preparo para Biblioteca NGS
KITS de Preparo para Biblioteca NGS
 
O que é X Frágil?
O que é X Frágil?O que é X Frágil?
O que é X Frágil?
 
Escolha um parceiro e não um fornecedor
Escolha um parceiro e não um fornecedorEscolha um parceiro e não um fornecedor
Escolha um parceiro e não um fornecedor
 
Eletroforese Capilar Automatizada
Eletroforese Capilar Automatizada Eletroforese Capilar Automatizada
Eletroforese Capilar Automatizada
 
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OESAnalysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
 
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
 
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...
 
CG Clarus® All In One
CG Clarus® All In OneCG Clarus® All In One
CG Clarus® All In One
 
High Quality DNA Isolation Suitable for Ultra Rapid Sequencing
High Quality DNA Isolation Suitable for Ultra Rapid SequencingHigh Quality DNA Isolation Suitable for Ultra Rapid Sequencing
High Quality DNA Isolation Suitable for Ultra Rapid Sequencing
 

Recently uploaded

Call Now ☎ Russian Call Girls Connaught Place @ 9899900591 # Russian Escorts ...
Call Now ☎ Russian Call Girls Connaught Place @ 9899900591 # Russian Escorts ...Call Now ☎ Russian Call Girls Connaught Place @ 9899900591 # Russian Escorts ...
Call Now ☎ Russian Call Girls Connaught Place @ 9899900591 # Russian Escorts ...kauryashika82
 
$ Love Spells 💎 (310) 882-6330 in Pennsylvania, PA | Psychic Reading Best Bla...
$ Love Spells 💎 (310) 882-6330 in Pennsylvania, PA | Psychic Reading Best Bla...$ Love Spells 💎 (310) 882-6330 in Pennsylvania, PA | Psychic Reading Best Bla...
$ Love Spells 💎 (310) 882-6330 in Pennsylvania, PA | Psychic Reading Best Bla...PsychicRuben LoveSpells
 
Hot Call Girls 🫤 Malviya Nagar ➡️ 9711199171 ➡️ Delhi 🫦 Whatsapp Number
Hot Call Girls 🫤 Malviya Nagar ➡️ 9711199171 ➡️ Delhi 🫦 Whatsapp NumberHot Call Girls 🫤 Malviya Nagar ➡️ 9711199171 ➡️ Delhi 🫦 Whatsapp Number
Hot Call Girls 🫤 Malviya Nagar ➡️ 9711199171 ➡️ Delhi 🫦 Whatsapp Numberkumarajju5765
 
VVIP Pune Call Girls Koregaon Park (7001035870) Pune Escorts Nearby with Comp...
VVIP Pune Call Girls Koregaon Park (7001035870) Pune Escorts Nearby with Comp...VVIP Pune Call Girls Koregaon Park (7001035870) Pune Escorts Nearby with Comp...
VVIP Pune Call Girls Koregaon Park (7001035870) Pune Escorts Nearby with Comp...Call Girls in Nagpur High Profile
 
Call Girls Moshi Call Me 7737669865 Budget Friendly No Advance Booking
Call Girls Moshi Call Me 7737669865 Budget Friendly No Advance BookingCall Girls Moshi Call Me 7737669865 Budget Friendly No Advance Booking
Call Girls Moshi Call Me 7737669865 Budget Friendly No Advance Bookingroncy bisnoi
 
Call Girls Ramtek Call Me 7737669865 Budget Friendly No Advance Booking
Call Girls Ramtek Call Me 7737669865 Budget Friendly No Advance BookingCall Girls Ramtek Call Me 7737669865 Budget Friendly No Advance Booking
Call Girls Ramtek Call Me 7737669865 Budget Friendly No Advance Bookingroncy bisnoi
 
CSR_Tested activities in the classroom -EN
CSR_Tested activities in the classroom -ENCSR_Tested activities in the classroom -EN
CSR_Tested activities in the classroom -ENGeorgeDiamandis11
 
VIP Call Girls Valsad 7001035870 Whatsapp Number, 24/07 Booking
VIP Call Girls Valsad 7001035870 Whatsapp Number, 24/07 BookingVIP Call Girls Valsad 7001035870 Whatsapp Number, 24/07 Booking
VIP Call Girls Valsad 7001035870 Whatsapp Number, 24/07 Bookingdharasingh5698
 
VIP Model Call Girls Hadapsar ( Pune ) Call ON 8005736733 Starting From 5K to...
VIP Model Call Girls Hadapsar ( Pune ) Call ON 8005736733 Starting From 5K to...VIP Model Call Girls Hadapsar ( Pune ) Call ON 8005736733 Starting From 5K to...
VIP Model Call Girls Hadapsar ( Pune ) Call ON 8005736733 Starting From 5K to...SUHANI PANDEY
 
The Most Attractive Pune Call Girls Shirwal 8250192130 Will You Miss This Cha...
The Most Attractive Pune Call Girls Shirwal 8250192130 Will You Miss This Cha...The Most Attractive Pune Call Girls Shirwal 8250192130 Will You Miss This Cha...
The Most Attractive Pune Call Girls Shirwal 8250192130 Will You Miss This Cha...ranjana rawat
 
VIP Model Call Girls Viman Nagar ( Pune ) Call ON 8005736733 Starting From 5K...
VIP Model Call Girls Viman Nagar ( Pune ) Call ON 8005736733 Starting From 5K...VIP Model Call Girls Viman Nagar ( Pune ) Call ON 8005736733 Starting From 5K...
VIP Model Call Girls Viman Nagar ( Pune ) Call ON 8005736733 Starting From 5K...SUHANI PANDEY
 
Call Girls In Okhla DELHI ~9654467111~ Short 1500 Night 6000
Call Girls In Okhla DELHI ~9654467111~ Short 1500 Night 6000Call Girls In Okhla DELHI ~9654467111~ Short 1500 Night 6000
Call Girls In Okhla DELHI ~9654467111~ Short 1500 Night 6000Sapana Sha
 
Booking open Available Pune Call Girls Parvati Darshan 6297143586 Call Hot I...
Booking open Available Pune Call Girls Parvati Darshan  6297143586 Call Hot I...Booking open Available Pune Call Girls Parvati Darshan  6297143586 Call Hot I...
Booking open Available Pune Call Girls Parvati Darshan 6297143586 Call Hot I...Call Girls in Nagpur High Profile
 
Call Girl Nagpur Roshni Call 7001035870 Meet With Nagpur Escorts
Call Girl Nagpur Roshni Call 7001035870 Meet With Nagpur EscortsCall Girl Nagpur Roshni Call 7001035870 Meet With Nagpur Escorts
Call Girl Nagpur Roshni Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur High Profile
 
Call Girls In Bloom Boutique | GK-1 ☎ 9990224454 High Class Delhi NCR 24 Hour...
Call Girls In Bloom Boutique | GK-1 ☎ 9990224454 High Class Delhi NCR 24 Hour...Call Girls In Bloom Boutique | GK-1 ☎ 9990224454 High Class Delhi NCR 24 Hour...
Call Girls In Bloom Boutique | GK-1 ☎ 9990224454 High Class Delhi NCR 24 Hour...rajputriyana310
 
Get Premium Hoskote Call Girls (8005736733) 24x7 Rate 15999 with A/c Room Cas...
Get Premium Hoskote Call Girls (8005736733) 24x7 Rate 15999 with A/c Room Cas...Get Premium Hoskote Call Girls (8005736733) 24x7 Rate 15999 with A/c Room Cas...
Get Premium Hoskote Call Girls (8005736733) 24x7 Rate 15999 with A/c Room Cas...MOHANI PANDEY
 
NO1 Verified kala jadu karne wale ka contact number kala jadu karne wale baba...
NO1 Verified kala jadu karne wale ka contact number kala jadu karne wale baba...NO1 Verified kala jadu karne wale ka contact number kala jadu karne wale baba...
NO1 Verified kala jadu karne wale ka contact number kala jadu karne wale baba...Amil baba
 

Recently uploaded (20)

Call Now ☎ Russian Call Girls Connaught Place @ 9899900591 # Russian Escorts ...
Call Now ☎ Russian Call Girls Connaught Place @ 9899900591 # Russian Escorts ...Call Now ☎ Russian Call Girls Connaught Place @ 9899900591 # Russian Escorts ...
Call Now ☎ Russian Call Girls Connaught Place @ 9899900591 # Russian Escorts ...
 
$ Love Spells 💎 (310) 882-6330 in Pennsylvania, PA | Psychic Reading Best Bla...
$ Love Spells 💎 (310) 882-6330 in Pennsylvania, PA | Psychic Reading Best Bla...$ Love Spells 💎 (310) 882-6330 in Pennsylvania, PA | Psychic Reading Best Bla...
$ Love Spells 💎 (310) 882-6330 in Pennsylvania, PA | Psychic Reading Best Bla...
 
Hot Call Girls 🫤 Malviya Nagar ➡️ 9711199171 ➡️ Delhi 🫦 Whatsapp Number
Hot Call Girls 🫤 Malviya Nagar ➡️ 9711199171 ➡️ Delhi 🫦 Whatsapp NumberHot Call Girls 🫤 Malviya Nagar ➡️ 9711199171 ➡️ Delhi 🫦 Whatsapp Number
Hot Call Girls 🫤 Malviya Nagar ➡️ 9711199171 ➡️ Delhi 🫦 Whatsapp Number
 
VVIP Pune Call Girls Koregaon Park (7001035870) Pune Escorts Nearby with Comp...
VVIP Pune Call Girls Koregaon Park (7001035870) Pune Escorts Nearby with Comp...VVIP Pune Call Girls Koregaon Park (7001035870) Pune Escorts Nearby with Comp...
VVIP Pune Call Girls Koregaon Park (7001035870) Pune Escorts Nearby with Comp...
 
Call Girls Moshi Call Me 7737669865 Budget Friendly No Advance Booking
Call Girls Moshi Call Me 7737669865 Budget Friendly No Advance BookingCall Girls Moshi Call Me 7737669865 Budget Friendly No Advance Booking
Call Girls Moshi Call Me 7737669865 Budget Friendly No Advance Booking
 
Call Girls Ramtek Call Me 7737669865 Budget Friendly No Advance Booking
Call Girls Ramtek Call Me 7737669865 Budget Friendly No Advance BookingCall Girls Ramtek Call Me 7737669865 Budget Friendly No Advance Booking
Call Girls Ramtek Call Me 7737669865 Budget Friendly No Advance Booking
 
CSR_Tested activities in the classroom -EN
CSR_Tested activities in the classroom -ENCSR_Tested activities in the classroom -EN
CSR_Tested activities in the classroom -EN
 
VIP Call Girls Valsad 7001035870 Whatsapp Number, 24/07 Booking
VIP Call Girls Valsad 7001035870 Whatsapp Number, 24/07 BookingVIP Call Girls Valsad 7001035870 Whatsapp Number, 24/07 Booking
VIP Call Girls Valsad 7001035870 Whatsapp Number, 24/07 Booking
 
(NEHA) Call Girls Navi Mumbai Call Now 8250077686 Navi Mumbai Escorts 24x7
(NEHA) Call Girls Navi Mumbai Call Now 8250077686 Navi Mumbai Escorts 24x7(NEHA) Call Girls Navi Mumbai Call Now 8250077686 Navi Mumbai Escorts 24x7
(NEHA) Call Girls Navi Mumbai Call Now 8250077686 Navi Mumbai Escorts 24x7
 
VIP Model Call Girls Hadapsar ( Pune ) Call ON 8005736733 Starting From 5K to...
VIP Model Call Girls Hadapsar ( Pune ) Call ON 8005736733 Starting From 5K to...VIP Model Call Girls Hadapsar ( Pune ) Call ON 8005736733 Starting From 5K to...
VIP Model Call Girls Hadapsar ( Pune ) Call ON 8005736733 Starting From 5K to...
 
The Most Attractive Pune Call Girls Shirwal 8250192130 Will You Miss This Cha...
The Most Attractive Pune Call Girls Shirwal 8250192130 Will You Miss This Cha...The Most Attractive Pune Call Girls Shirwal 8250192130 Will You Miss This Cha...
The Most Attractive Pune Call Girls Shirwal 8250192130 Will You Miss This Cha...
 
VIP Model Call Girls Viman Nagar ( Pune ) Call ON 8005736733 Starting From 5K...
VIP Model Call Girls Viman Nagar ( Pune ) Call ON 8005736733 Starting From 5K...VIP Model Call Girls Viman Nagar ( Pune ) Call ON 8005736733 Starting From 5K...
VIP Model Call Girls Viman Nagar ( Pune ) Call ON 8005736733 Starting From 5K...
 
Sustainable Packaging
Sustainable PackagingSustainable Packaging
Sustainable Packaging
 
Call Girls In Yamuna Vihar꧁❤ 🔝 9953056974🔝❤꧂ Escort ServiCe
Call Girls In Yamuna Vihar꧁❤ 🔝 9953056974🔝❤꧂ Escort ServiCeCall Girls In Yamuna Vihar꧁❤ 🔝 9953056974🔝❤꧂ Escort ServiCe
Call Girls In Yamuna Vihar꧁❤ 🔝 9953056974🔝❤꧂ Escort ServiCe
 
Call Girls In Okhla DELHI ~9654467111~ Short 1500 Night 6000
Call Girls In Okhla DELHI ~9654467111~ Short 1500 Night 6000Call Girls In Okhla DELHI ~9654467111~ Short 1500 Night 6000
Call Girls In Okhla DELHI ~9654467111~ Short 1500 Night 6000
 
Booking open Available Pune Call Girls Parvati Darshan 6297143586 Call Hot I...
Booking open Available Pune Call Girls Parvati Darshan  6297143586 Call Hot I...Booking open Available Pune Call Girls Parvati Darshan  6297143586 Call Hot I...
Booking open Available Pune Call Girls Parvati Darshan 6297143586 Call Hot I...
 
Call Girl Nagpur Roshni Call 7001035870 Meet With Nagpur Escorts
Call Girl Nagpur Roshni Call 7001035870 Meet With Nagpur EscortsCall Girl Nagpur Roshni Call 7001035870 Meet With Nagpur Escorts
Call Girl Nagpur Roshni Call 7001035870 Meet With Nagpur Escorts
 
Call Girls In Bloom Boutique | GK-1 ☎ 9990224454 High Class Delhi NCR 24 Hour...
Call Girls In Bloom Boutique | GK-1 ☎ 9990224454 High Class Delhi NCR 24 Hour...Call Girls In Bloom Boutique | GK-1 ☎ 9990224454 High Class Delhi NCR 24 Hour...
Call Girls In Bloom Boutique | GK-1 ☎ 9990224454 High Class Delhi NCR 24 Hour...
 
Get Premium Hoskote Call Girls (8005736733) 24x7 Rate 15999 with A/c Room Cas...
Get Premium Hoskote Call Girls (8005736733) 24x7 Rate 15999 with A/c Room Cas...Get Premium Hoskote Call Girls (8005736733) 24x7 Rate 15999 with A/c Room Cas...
Get Premium Hoskote Call Girls (8005736733) 24x7 Rate 15999 with A/c Room Cas...
 
NO1 Verified kala jadu karne wale ka contact number kala jadu karne wale baba...
NO1 Verified kala jadu karne wale ka contact number kala jadu karne wale baba...NO1 Verified kala jadu karne wale ka contact number kala jadu karne wale baba...
NO1 Verified kala jadu karne wale ka contact number kala jadu karne wale baba...
 

PAMS: A Study of Performance in Low and High Humidity Environments

  • 1. Introduction The United States Environmental Protection Agency (USEPA) Photochemical Assessment Monitoring Stations (PAMS) are air quality monitoring locations that are used to monitor pollutant compounds in air that contribute to ground level (bad) ozone. These nonpolar compounds are typically found in the 0.1 to tens ppbv, if not greater in some cases. Since these levels are too low to detect directly, a concentration technique such as cryotrapping or chemical adsorbent trapping needs to be used. One of the main issues with any sub-ambient temperature method is the negative effect of moisture in the air sample. Quantities up to tens of milligrams of water may be present and this can cause havoc with the chromatographic system, such as the water freezing with the formation of ice causing a flow restriction or blockage. Water can also cause poor peak shape/retention of analytes on certain columns such as Alumina PLOT. Several different techniques can be used for the effective removal of water from a gas sampling stream, prior to preconcentration, such as, drying agents, semipermeable membranes and reduced temperatures (used to condense out the water). One well accepted technique uses a Nafion® (TM, DuPont™ ) membrane to dry the incoming sample stream, and is very effective. The membrane permeability is very dependent on the analyte, generally speaking, and nonpolars are retained while water and certain polars are lost across the membrane. This work is focused on humidity effects on low level sub-ppbv (0.3-1 ppbv) PAMS targets compared to the work done by G. Broadway et al¹ who studied humidity effects at a higher 10 ppbv. PAMS: A Study of Performance in Low and High Humidity Environments A P P L I C A T I O N N O T E Author: PerkinElmer Canadian Chromatographic Application Lab PerkinElmer, Inc. Woodbridge, ON Canada Gas Chromatography
  • 2. 2 Experimental PAMS standard was purchased from Linde Canada at the USEPA recommended concentration of 3-10 ppbv and was used as the stock standard. The PAMS components dilutions were dynamically created in real time using a custom made on-line diluter, designed and built by the author, using UHP Nitrogen (LN2) as the diluent. The gas was purified using an advanced triple bed carrier gas purifier filter. The on-line diluter utilized two 2000 mL/min. MKS Mass-Flo Mass flow controllers (MFC). One was used to control pure nitrogen diluent and the other was to control the flow to a 10 um stainless steel frit, inserted into a 2L HPLC solvent bottle, ¾ filled with 18 megohm DI water. The bubbler MFC outlet was teed in to the pure nitrogen MFC output. To eliminate the possibility of trace level contaminates in the water and/or associated tubing/fittings/ MFC, since we are running at sub-ppbv levels, the total diluent flow (both Humid and Dry) was scrubbed of organics by the use of two Soil Vapour Intrusion tubes (SVI, thermal desorption) in series post blending of the diluent. The now clean diluent was blended with the PAMS standard (via a 200 mL/min. MKS MFC) into a ¼” SS tubing. To reduce the effects of laminar flow and promote mixing, the tubing was reduced to 1/8” SS (6 ft). This tubing was terminated into a SS 1/8” Tee with one arm going toward the on-line accessory approx. 1 ft. away with a 15 mL/ min. flow. The remaining arm was plumbed with ¼” Bev-A-Line, as it was downstream of the sampling accessory, and at the end of this line a custom designed and manufactured humidity accessory was installed. A block diagram of the system is shown in Figure 1. The custom designed humidity accessory utilized a Honeywell Humidcon™ HIH9130-021 with a 0-100% RH range, with +/-1.7 accuracy, +/-0.6 °C temperature reading communication via I2C. The I2C data was read via a customized Atmel® 328P microprocessor and the resulting data was displayed via I2C with a 16 x 2 LCD screen. The % RH and temperature was updated at 10 Hz. Similarly, the MFC were controlled by a custom programmed Atmel 328P, using SPI controlled MCP4821 12-bit DAC. Since the MCP4821 max output is 4096 mV and the MFC requires 0-5000 mV, a TLC2272 Rail to Rail Op Amp was used to scale the DAC 0-4096 mV output by 1.25. A rotary encoder with momentary push buttons was used for the setpoint and MFC selection. The MFC readback was read by the ATMega328P 10-bit internal ADC directly without buffering. The readbacks/setpoints were displayed on a 16 x 2 LCD via I2C. An electronic block diagram showing the basic components of the MFC control is shown in Figure 2. Figure1.Diagramofadynamicdiluter. Figure2.SchematicofMFCcontrolelectronics. Table1.SystemConditionsforOzonePrecursorAnalysis. GC Clarus® 580 GC with heartcut option Column Ozone Precursor Column Set 50 m x 0.22 mm x 1 μm dimethyl siloxane column Part number 04978179 50 m x 0.32 mm alumina/Na2SO4 PLOT column Part number N6301107 Oven Conditions 46 °C for 15 min., then 5 °C/min. to 170 °C, then 15 °C/minute to 200 °C and hold for 6 min. (48-min. chromatogram) Detector Dual FID at 250 °C Carrier Gas 42 psig helium at the ATD 16 psig helium at the midpoint ATD Thermal Desorber TurboMatrix™ TD with online sampling accessory Trap Air Toxics trap packed with carbonaceous adsorbents Low -30 °C High 325 °C Transfer Line 200 °C Operating Mode On-Line at 15 mL/min. for 40 min. Split Inlet Off Outlet 2 mL/min. Data Handling TotalChrom™
  • 3. 3 MFC Settings and Flows MFC settings were 10 bit (0-1023) for the 0-5 V analog input for the MFC via the 12 bit DAC. The MFC flows were calibrated using a NIST traceable PE1000 electronic flow meter. The MFCs used were identified as MFC#1 (for Humid Flow), MFC#2 (Dry Flow) and MFC#3 (Gas Standard). Unless otherwise noted the MFC were set as follows: MFC#1, set to 100, produced a measured flow of 228 mL/min. direct at MFC output, and 263 mL/min. of humid flow at 77% RH after humidification. Since the extra volume of the water vapor at RT contributes, it is believed the difference in flow with the humidifier attached is a result of the extra backpressure due to humidifier. MFC#2, set to 100, produced a flow of 193 mL/min. MFC#3, set to 100, produced a flow of 34.7 mL/min. Dilution of standard with dry nitrogen = (193+34.7)/34.7 = 1:6.56 Dilution of standard with humidified nitrogen = (263+34.7)/34.7 = 1:8.58 Humidity Notes Since the generated humidity is passed through two SVI tubes in series, the humidity needs time to stabilize as each tube has a moisture capacity of approx. 35 mg each. This effect is described in detail by Snow et al². In this study, typically one hour stabilization time was used prior to sampling. Results Both the 77% RH and dry dilutions of the PAMS standard give similar chromatograms in both the PLOT and BP-1 separations as shown in Figures 3 and 4 respectively. Figure 5 is an overlay of four sequential chromatographic runs highlighting excellent repeatability at low levels. Figure3.PLOTchromatogramofPAMSstandardatapprox.0.3-1ppbv.Topchromatogramis77%RHandbottomisapproximately0%RH.
  • 5. For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs Copyright ©2015, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners. 012297_01 PKI PerkinElmer, Inc. 940 Winter Street Waltham, MA 02451 USA P: (800) 762-4000 or (+1) 203-925-4602 www.perkinelmer.com References 1. “Ozone Precursor Analysis Using a Thermal Desorption-GC System” by G. Broadway and A. Tipler available for download at www.perkinelmer.com. 2. Patient Application US 20130180315 A1 Devices, systems and methods for purging and loading sorbent tubes by Snow et al. Conclusion The ozone precursor system performance at the 0.3 ppbv-1 ppbv, with a PAMS standard under high humidity conditions, are virtually identical to that under low/zero humidity. The chromatographic peak shapes and intensities show that high humidity has little/no negative effect on the analysis of PAMS target compounds at low levels. This enables the use of dry gaseous standards to be used without the laborious preparation of humidified standards.