TSQ Endura Mass Position Check and Calibration Procedure
1. Standard Operating
Procedure
TSQ Endura Mass Position Check and
Calibration Procedure
Revision : A
Author: Jerry Rattanong Published Date: 12-2-15 Page 1 of 6
TSQ ENDURA MASS POSITIONCHECK AND
CALIBRATION PROCEDURE
What youwill need (Figure 1):
Syringe
Infusionadapterandtubing
Syringe Infuser
Methanol forCleaning
Pierce Triple Quadrupole CalibrationSolution
(PolyT)
Figure 1: Syringe, Calibration Solution, Infusion
adapter and Tubing
1- Fill yoursyringe withcleaningmethanol
and dispense intowastebeakertorinse
out the syringe. Dothistwo times.
2- Fill yoursyringe withthe Triple
Quadrupole CalibrationSolution(Poly
T)
3- Attach syringe byinsertingthe tipinto
the clear plasticattachedtothe
adapter. You will meetsome
resistance,thenpushina little more bit
to seat.
4- Place Syringe intoinfuser. Pressthe
small buttonandslide the blockto
adjust. Pull roundtab androtate clipto
holdsyringe inplace (Figure 2). Attach
the endof tubingintoESI (Figure 3).
Figure 2: Syringe Infuser
Figure 3: Complete Set up
Tuning
OpenEndura Tune iconby double clicking.
Beginbysettingselectingthe DefineScanTab,
Ensuringthat positive modeisselectedand
makingsure the Full Scan Q1 is enabledandthe
scan range is 150-1050 (Figure 4).
2. Standard Operating
Procedure
TSQ Endura Mass Position Check and
Calibration Procedure
Revision : A
Author: Jerry Rattanong Published Date: 12-2-15 Page 2 of 6
Figure 4: Define Scan Settings
Next,setthe ionsource settingsforpositive
mode. Clickonthe IonSource tab, thenclickon
the leftarrowshownhere on the rightside
(Figure5) . Thisopensthe custom user settings
panel. All customionsource settingsare
locatedhere
Figure 5: Arrow Location
Next,selectthe PositiveMode PolyTsettings
underUser Settings,rightclick andselectload
(Figure 6).
Figure 6: User Settings
The Ion source Settingsshouldnow be setto
the following(Figure 7):
Figure 7: Ion Source Settings
3. Standard Operating
Procedure
TSQ Endura Mass Position Check and
Calibration Procedure
Revision : A
Author: Jerry Rattanong Published Date: 12-2-15 Page 3 of 6
Start Infusionof PolyT
Selectthe Syringe iconatthe top to turn iton and setthe flow rate to 3ul/min (Figure 8).
Figure 8: Syringe flow rate
The screenshouldnowlooklike this (Figure 9):
Figure 9: The three Poly T masses in Positive mode.
*note:you mayset the syringe rate to 10uL or more to getthe flow started. Setthe flow backto 3uL
once the three ionsare detected.
4. Standard Operating
Procedure
TSQ Endura Mass Position Check and
Calibration Procedure
Revision : A
Author: Jerry Rattanong Published Date: 12-2-15 Page 4 of 6
BeginCalibration:
Selectthe CalibrationtabandclickCalibrate
(Figure 10).
Make sure that bothQuads are selectedand
selectCheckMass PositionandResolutionand
clickCalibrate tobegin (Figure 11):
Figure 11: Calibration Selection
Figure 10: Calibration Tab
Checkmass andposition inprogress (Figure 12):
5. Standard Operating
Procedure
TSQ Endura Mass Position Check and
Calibration Procedure
Revision : A
Author: Jerry Rattanong Published Date: 12-2-15 Page 5 of 6
Figure 12: In
progress
Screenshot of check mass and position
Afterthe CheckMass and Resolutioniscomplete,pressokaytogenerate areportand save it (Figure
13).
Figure 13: Generate Report dialog box
If the CheckMass and Resolutionfails,runthe Calibrate MassandResolution,thendothe SystemTune
and Checkand thenfinallythe Electron MultiplierCalibrationlast.
Afterthat all positive modecalibrationsare completeperformthe negative modecalibrationby
switchingtonegative mode andloadingthe appropriateionsource usersettings. Setthe syringe to
8uL/minFlowand verifythe three negative ions,182.11, 508.33 and 997.44 are present(Figure 13).
6. Standard Operating
Procedure
TSQ Endura Mass Position Check and
Calibration Procedure
Revision : A
Author: Jerry Rattanong Published Date: 12-2-15 Page 6 of 6
Figure 13: Negative mode poly t source settings and Ions
Thenproceedto calibrate the Mass andResolution,thenSystemTune andCheck. Donot worryabout
the Electron MultiplierCalibrationforNegative mode.
Once the negative calibrationiscomplete,stopthe syringe infusion andrinse the barrel outwith
Methanol,andalsomanuallyrinse outthe line andadapterusingthe syringe andstore the parts. Set
startingconditionsof abatch and run a few samples,suchasQCs are Calibratorstodetermine if
everythingisworkingproperly.