The document provides guidance on selecting and configuring a POLYMETRON 8810 Titrator instrument. It discusses the various parameters that can be measured by the titrator, including ammonia, alkalinity, calcium, chloride, cyanide, fluoride, hardness, nitrite and sulphide. For each parameter, it lists the typical applications, measuring range, standard configuration, options, operating materials needed and any applications where the titrator is not suitable. It emphasizes that the titrator has over 20 options and requiring ensuring the correct configuration for each customer's specific application.
2. 2
Why a selection guide for the 8810?
The purpose of this selection guide is to help define, identify and configure a POLYMETRON 8810
Titrator instrument during the quoting process. Due to its complexity and number of options, the 8810
Titrator is difficult to configure and it is quite often a challenge to make sure it is suitable for specific
applications or to make sure we have prepared all part numbers. With this guide, we have tried to bring
you the knowledge of our technical product specialists.
This guide is structured in a way that will enable users, step by step, to identify different factors related to
their applications (e.g. parameter, measuring range, hardware, options, service, etc). Not only are the
technical aspects described, but particular attention has been paid to the 20+ years this product has
been used in successful applications and the importance of having knowledgeable people installing the
8810 Titrator.
How to use this guide
This selection guide is divided in two sections. The first, (A), lists the general recommendations and
options that must be considered for any 8810 model. The second, (B), is specific to each parameter.
You need to walk through the two sections in order to validate that the 8810 is suitable for your
customer’s application.
DO NOT QUOTE
Section B contains, a section for each parameter called “NOT SUITABLE FOR!” This
section is a list of the applications or conditions under which the 8810 Titrator is not
recommended at all. In case your customer’s application is affected by these limitations,
DO NOT QUOTE this instrument. If you still have doubts, please contact your local
technical support team.
ORDER PROCESSING
All components ordered along with an 8810 unit will be factory mounted on the 8810 and
tested. If you want to order separate spare parts along with your 8810 and you do not
want them to be mounted on the unit, we highly recommend you to place separate orders
or at least clearly state your request on your order; otherwise, when possible, spare parts
will be mounted and shipped on the 8810.
3. 3
A) General recommendations & options (all models)
Pumps
In some configurations, two different types of pumps may be available: piston and peristaltic.
If a different method has not been suggested, the choice of pump, can be done based on the
following criteria:
o Peristaltic pump (or volumetric pump) needs to be selected when cost is a concern
for the customer. This pump is less expensive, but will require more maintenance.
o Piston pump (or pulse pump) needs to be selected when maintenance frequency is a
concern for the customer. This pump is more expensive, but requires less
maintenance
Sample temperature
Always respect the 8810 Titrator’s temperature range: 5°C – 55°C
Sample pressure
Always respect the pressure at the 8810 Titrator’s inlet: 0.5 – 6 bar
Sample flow
Always respect the flow at the 8810 Titrator’s inlet: 50 – 300 l/h
Air source and rinsing water
An air supply is required to run the 8810 Titrator : 5 – 7 bar
A rinsing water supply is required to run the 8810 Titrator : 1 – 6 bar
Particles
Maximum particles allowed in sample: 3% or 30 g/l, with a diameter maximum of 2 mm
Enclosure and Tropicalization
Use the flow chart on the next page, and use the suggested part numbers according to the
application:
4. 4
Training & Support
Only trained people are allowed to support and install the 8810 Titrator. Before quoting, make
sure that a dedicated team will be able to support the product in the country of installation. If a
training is required, please contact your local Sales Support team.
6. 6
8810 AMMONIA ANALYZER
Typical applications: Power, Waste water, Drinking water, Surface water
Measuring range: 0-1 mg/l N-NH4 (programmable) LDL = 0.01 mg/l N-NH4
0-100 mg/l N-NH4 (programmable) LDL = 0.1 mg/l N-NH4
4-4 000 mg/l N-NH4 (with dilution - programmable) LDL = 4 mg/l N-NH4
NOT SUITABLE FOR!!!: Nuclear, Petrochemical and Water ice rink applications
Or if sample contains cyanide, sulphide or oils (see AMTAX)
STANDARD PRODUCT CONFIGURATION
PART NUMBER ITEM DESCRIPTION
368810,36XXX
XXX = 220 220V/50Hz
XXX = 226 220V/60Hz
XXX = 240 240V/50Hz
XXX = 115 110V/50Hz
XXX = 116 110V/60Hz
8810 AMMONIA ANALYZER basic unit 19" panel mounted includes:
- Titration vessel
- Sprinkler
- Measuring combination AMMONIA electrode
- Temperature sensor Pt 100
- One peristaltic reagent pump for sample conditioning with NaOH
- Chemical cleaning (368810,56000) with 10l canister and reagent level
detector
- Automatic heating device/controller (368810,76000)
- 8810 ISE basic user manual
621=997=101
OR
621=997=102
Installation procedure AMMONIA 0–100 mg/l
Installation procedure AMMONIA 0–1 mg/l
7. 7
OPTIONS
368810,00100 Reagent canister holder (up to 4 x 10-liter canisters): L600 x W220 x H200
mm, in PP
08810=A=3500 Second channel option on 8810 ISE analyzer
368810,65000 Manual sample entry system
368810,93006 Spare parts kit (2 years), 8810 ISE analyzer w/ ammonia electrode
08811=A=20X2
X = 1 110V/50 Hz
X = 3 220V/50 Hz
X = 4 240V/50 Hz
2 channels 8811 Sequencer complete
OR 08811=A=20X3 3 channels 8811 Sequencer complete
OR 08811=A=20X4 4 channels 8811 Sequencer complete
OR 08811=A=20X5 5 channels 8811 Sequencer complete
OR 08811=A=20X6 6 channels 8811 Sequencer complete
8. 8
OPERATING MATERIAL/YEAR
based on one analysis every 15 minutes
PART NUMBER QTY ITEM DESCRIPTION
590=020=130 2 Grey/grey ismaprene pump tube
125=000=203 1 Ammonia electrode filling solution
125=000=204 1 20 membranes for ammonia electrode
Not supplied by Hach
Lange: to be
purchased locally.
Solid state : 10 kg
OR
Liquid state : 21 l
Sodium hydroxide NaOH in pellets (conditioning)
OR
Sodium hydroxide NaOH solution > 32% (conditioning)
7 l Concentrated hydrochloric acid,
HCl ~ 37% (chemical cleaning)
1 kg Ammonium chloride
NH4Cl (calibration)
9. 9
FREE ALKALINITY (p value)
Typical applications: Decarbonation processes, Water treatments
Or applications where the pH is above 8.3
Measuring range: 0-500 mg/l CaCO3 LDL = 1 mg/l CaCO3
0-10 mval/l CaCO3
NOT SUITABLE FOR!!!: Bioreactor applications
STANDARD PRODUCT CONFIGURATION
PART NUMBER ITEM DESCRIPTION
368810,40XXX
XXX = 220 220V/50Hz
XXX = 240 240V/50Hz
XXX = 115 110V/50Hz
XXX = 116 110V/60Hz
8810 FREE ALKALINITY ANALYZER basic unit 19" panel mounted includes:
- Titration vessel
- Sprinkler
- Measuring 8404 pH electrode
- Pt 100
- Reference electrode 8483B
- One reagent piston pump for titration
- Chemical cleaning (368810,56000) with 10 l. canister and reagent level
detector
- 8810 pH basic user manual
621=997=108 Installation procedure Free/Total ALKALINITY
10. 10
OPTIONS
368810,00100 Reagent canister holder (up to 4 x 10-liter canisters):
L600 x W220 x H200 mm, in PP
368810,60000 Automatic calibration volumetric pump
(includes 10 l. canister and reagent level detector)
Add: 368810,75060: 10 l. dark canister (for 8810) equipped for Auto-
Calibration with a level detector and a cartridge for coloured self-indicator
soda-lime (CO2-trap) -- No chemical supplied --
368810,65000 Manual sample entry system
368810,91010 Spare parts kit (2 years), 8810 pH analyzer with pH electrode compatible with piston pump
08811=A=20X2
X = 1 110V/50 Hz
X = 3 220V/50 Hz
X = 4 240V/50 Hz
2 channels 8811 Sequencer complete
OR 08811=A=20X3 3 channels 8811 Sequencer complete
OR 08811=A=20X4 4 channels 8811 Sequencer complete
OR 08811=A=20X5 5 channels 8811 Sequencer complete
OR 08811=A=20X6 6 channels 8811 Sequencer complete
OPERATING MATERIAL/YEAR
based on one analysis every 15 minutes
PART NUMBER QTY ITEM DESCRIPTION
Not supplied by Hach
Lange: to be purchased
locally.
2 l Concentrated sulphuric acid, H2SO4 ~95% (titrant solution)
7 l Concentrated hydrochloric acid, HCl ~ 37% (chemical cleaning)
100 g Sodium carbonate anhydrous Na2CO3 (calibration solution)
11. 11
8810 TOTAL ALKALINITY (m value) ANALYZER
Typical applications: Decarbonation processes, Water treatments, Cooling water
Or applications where the pH is below 8.3
Measuring range: 0-500 mg/l CaCO3 LDL = 1 mg/l CaCO3
0-10 mval/l CaCO3
NOT SUITABLE FOR!!!: Bioreactor applications
STANDARD PRODUCT CONFIGURATION
PART NUMBER ITEM DESCRIPTION
368810,40XXX
XXX = 220 220V/50Hz
XXX = 240 240V/50Hz
XXX = 115 110V/50Hz
XXX = 116 110V/60Hz
8810 TOTAL ALKALINITY ANALYZER basic unit 19" panel mounted includes:
- Titration vessel
- Sprinkler
- Measuring 8404 pH electrode
- Reference electrode 8483B
- One reagent piston pump for titration
- Chemical cleaning (368810,56000) with 10 l. canister and reagent level
detector
- 8810 pH basic user manual
621=997=108 Installation procedure Free/Total ALKALINITY
12. 12
OPTIONS
368810,00100 Reagent canister holder (up to 4 x 10-liter canisters):
L600 x W220 x H200 mm, in PP
368810,60000 Automatic calibration volumetric pump
(includes 10 l. canister and reagent level detector)
Add: 368810,75060: 10 l. dark canister (for 8810) equipped for Auto-Calibration
with a level detector and a cartridge for colored self-indicator soda-lime (CO2-
trap) -- No chemical supplied --
368810,65000 Manual sample entry system
368810,91010 Spare parts kit (2 years), 8810 pH analyzer with pH electrode compatible with
piston pump
08811=A=20X2
X = 1 110V/50 Hz
X = 3 220V/50 Hz
X = 4 240V/50 Hz
2 channels 8811 Sequencer complete
OR 08811=A=20X3 3 channels 8811 Sequencer complete
OR 08811=A=20X4 4 channels 8811 Sequencer complete
OR 08811=A=20X5 5 channels 8811 Sequencer complete
OR 08811=A=20X6 6 channels 8811 Sequencer complete
OPERATING MATERIAL/YEAR
based on one analysis every 15 minutes
PART NUMBER QTY ITEM DESCRIPTION
Not supplied by Hach
Lange: to be purchased
locally.
2 l Concentrated sulphuric acid, H2SO4 ~95% (titrant solution)
7 l Concentrated hydrochloric acid, HCl ~ 37% (chemical cleaning)
100 g Sodium carbonate anhydrous Na2CO3 (calibration solution)
13. 13
8810 FREE AND TOTAL ALKALINITY (m & p / 2p-m values) ANALYZER
Typical applications: Decarbonation processes, Water treatments, Softening processes
Or applications where the pH is above 8.3
Measuring range: 0-500 mg/l CaCO3 LDL = 1 mg/l CaCO3
0-10 mval/l CaCO3
NOT SUITABLE FOR!!!: Bioreactor applications
STANDARD PRODUCT CONFIGURATION
PART NUMBER ITEM DESCRIPTION
368810,40XXX
XXX = 220 220V/50Hz
XXX = 240 240V/50Hz
XXX = 115 110V/50Hz
XXX = 116 110V/60Hz
8810 FREE & TOTAL ALKALINITY ANALYZER basic unit 19" panel mounted
includes:
- Titration vessel
- Sprinkler
- Measuring 8404 pH electrode
- Reference electrode 8483B
- One reagent piston pump for titration
- Chemical cleaning (368810,56000) with 10 l. canister and reagent level
detector
- 8810 pH basic user manual
368810,71050 Additional reagent piston pump for total alkalinity titration
If m ≥ 5p 2 pumps, otherwise 1 pump is enough
m = Total Alkalinity
p = Free Alkalinity
621=997=108 Installation procedure Free/Total ALKALINITY
14. 14
OPTIONS
368810,00100 Reagent canister holder (up to 4 x 10-liter canisters):
L600 x W220 x H200 mm, in PP
368810,60000 Automatic calibration volumetric pump
(includes 10 l. canister and reagent level detector)
Add: 368810,75060: 10 l. dark canister (for 8810) equipped for Auto-Calibration
with a level detector and a cartridge for coloured self-indicator soda-lime (CO2-
trap) -- No chemical supplied --
368810,65000 Manual sample entry system
368810,91010 Spare parts kit (2 years), 8810 pH analyzer with pH electrode compatible with
piston pump
08811=A=20X2
X = 1 110V/50 Hz
X = 3 220V/50 Hz
X = 4 240V/50 Hz
2 channels 8811 Sequencer complete
OR 08811=A=20X3 3 channels 8811 Sequencer complete
OR 08811=A=20X4 4 channels 8811 Sequencer complete
OR 08811=A=20X5 5 channels 8811 Sequencer complete
OR 08811=A=20X6 6 channels 8811 Sequencer complete
OPERATING MATERIAL/YEAR
based on one analysis every 15 minutes
PART NUMBER QTY ITEM DESCRIPTION
Not supplied by Hach
Lange: to be purchased
locally.
4 l Concentrated sulphuric acid, H2SO4 ~95% (titrant solution)
7 l Concentrated hydrochloric acid, HCl ~ 37% (chemical cleaning)
100 g Sodium hydrogen carbonate Na2HCO3 (calibration solution)
100 g Sodium carbonate anhydrous Na2CO3 (calibration solution)
15. 15
8810 CALCIUM ANALYZER
Typical applications: Cooling water in Power plants (towers, boilers), Water treatments
Measuring range: 0-500 mg/l CaCO3 LDL = 1 mg/l CaCO3
NOT SUITABLE FOR!!!: Calcium measurement in saturated brine
Steel plants
STANDARD PRODUCT CONFIGURATION
PART NUMBER ITEM DESCRIPTION
368810,53XXX
XXX = 220 220V/50Hz
XXX = 240 240V/50Hz
XXX = 115 110V/50Hz
XXX = 116 110V/60Hz
8810 CALCIUM ANALYZER basic unit 19" panel mounted includes:
- Titration vessel
- Sprinkler
- Measuring calcium electrode
- Reference electrode 8483B
- One reagent piston pump for titration
- Chemical cleaning (368810,56000) with 10 l. canister and reagent level
detector
- Automatic heating device/controller (368810,76000)
- 8810 ORP basic user manual
368810,71050 Additional reagent piston pump for conditioning with Triethanolamine/Nitric acid
621=997=106 Installation procedure ORP CALCIUM
16. 16
OPTIONS
368810,00100 Reagent canister holder (up to 4 x 10-liter canisters):
L600 x W220 x H200 mm, in PP
368810,60000 Automatic calibration volumetric pump
(includes 10 l. canister and reagent level detector)
368810,65000 Manual sample entry system
368810,91010 Spare parts kit (2 years), 8810 ORP analyzer with calcium electrode compatible with piston pump
08811=A=20X2
X = 1 110V/50 Hz
X = 3 220V/50 Hz
X = 4 240V/50 Hz
2 channels 8811 Sequencer complete
OR 08811=A=20X3 3 channels 8811 Sequencer complete
OR 08811=A=20X4 4 channels 8811 Sequencer complete
OR 08811=A=20X5 5 channels 8811 Sequencer complete
OR 08811=A=20X6 6 channels 8811 Sequencer complete
OPERATING MATERIAL/YEAR
based on one analysis every 15 minutes
PART NUMBER QTY ITEM DESCRIPTION
Not supplied by Hach
Lange: to be purchased
locally.
750 g Ethylene glycol-bis(2-aminoethyl)-N,N,N’,N’-tetraacetic acid EGTA
(titration)
7 l Concentrated hydrochloric acid, HCl ~ 37% (chemical cleaning)
7 l Triethanolamine (conditioning)
1 l Nitric acid (conditioning)
250 g Calcium chloride dehydrate CaCl2.2H2O (calibration solution)
17. 17
8810 CHLORIDE ANALYZER
Typical applications: Cooling water in Power plants (towers, boilers), Surface water
Industrial Waste water
Measuring range: 0 – 5 mg/l Cl-
0.5 – 500 mg/l Cl -
50 – 20 000 mg/l Cl -
LDL = 0.5 mg/l Cl - w/o dilution ; 50 mg/l Cl - w/ dilution
NOT SUITABLE FOR!!!: Mining applications or ore treatments (e.g. presence of Iron, Gold, Silver)
Or if sample contains Chlorite (ClO4)
Or if sample pH is below 4
Or if sample contains Ammonium Nitrate (NH4NO3) (Explosive factories)
Or if sample contains Sulphides (S-) (e.g. Refinery, Drinking water)
STANDARD PRODUCT CONFIGURATION
PART NUMBER ITEM DESCRIPTION
&
368810,31XXX
XXX = 220 220V/50Hz
XXX = 240 240V/50Hz
XXX = 115 110V/50Hz
XXX = 116 110V/60Hz
OR
368810,38XXX
8810 CHLORIDE ANALYZER basic unit 19" panel mounted includes:
- Titration vessel
- Sprinkler
- Measuring chloride 8471B ion selective electrode
- Reference electrode 8483B
- One reagent peristaltic pump for sample conditioning with Sodium
Hydroxide/Citric acid
- Chemical cleaning (368810,56000) with 10 l. canister and reagent level
detector
- Automatic heating device/controller (368810,76000)
- 8810 ISE basic user manual
Same analyzer but for 0.5 – 5 mg/l Cl-, low range applications, with an electric
sample valve instead of a pneumatic valve
621=997=135
OR
621=997=137
OR
621=997=116
Installation procedure CHLORIDE 0-20’000 mg/l Cl-
Installation procedure CHLORIDE 0-5 mg/l Cl-
Installation procedure CHLORIDE 0.5-500 mg/l Cl-
18. 18
OPTIONS
Autocal required?
Range
mg/l Cl- ?
Autocal required?
Automatic dilution
08810=A=5300
+
Automatic calibration volumetric
pump
368810,60000
Automatic dilution
08810=A=5300
Automatic calibration pulse pump
368810,72000
Nothing to add
High Range
50 - 20'000 mg/l Cl-
Standard Range
0.5 - 500 mg/l Cl-
Yes No
No
Yes
Low Range
0 - 5 mg/l Cl-
Autocal required?
Automatic calibration volumetric
pump
368810,60000
Yes
368810,00100 Reagent canister holder (up to 4 x 10-liter canisters):
L600 x W220 x H200 mm, in PP
08810=A=3500 Second channel option on 8810 analyzer
368810,65000 Manual sample entry system
368810,93001 Spare parts kit (2 years), 8810 ISE analyzer with chloride electrode
08811=A=20X2
X = 1 110V/50 Hz
X = 3 220V/50 Hz
X = 4 240V/50 Hz
2 channels 8811 Sequencer complete
OR 08811=A=20X3 3 channels 8811 Sequencer complete
OR 08811=A=20X4 4 channels 8811 Sequencer complete
OR 08811=A=20X5 5 channels 8811 Sequencer complete
OR 08811=A=20X6 6 channels 8811 Sequencer complete
19. 19
OPERATING MATERIAL/YEAR
based on one analysis every 15 minutes
PART NUMBER QTY ITEM DESCRIPTION
590=020=130 2 Grey/grey ismaprene pump tube
Hach/Hach Lange: to
be purchased locally.
Liquid state : 14 l Sodium hydroxide NaOH solution > 32% (conditioning)
13 kg Citric acid monohydrate C6H8O7.H2O(conditioning)
7 l Nitric acid concentrated 65 % (cleaning)
1 kg Potassium chloride KCl (calibration)
NOTE: Conditioning and cleaning reagents are the same w/ or w/o dilution. Only Calibration solutions vary.
20. 20
8810 CYANIDE ANALYZER
Typical applications: Industrial Waste water
Measuring range: 0 – 5 mg/l (programmable)
LDL = 0.03 mg/l CN -
NOT SUITABLE FOR!!!: Mining applications or ore treatments (e.g. presence of Iron, Gold, Silver)
Or if sample contains any Heavy metals (Pb, Hg, Cd, As, Ni, Cu, Zn, Mn, etc.)
STANDARD PRODUCT CONFIGURATION
PART NUMBER ITEM DESCRIPTION
368810,30XXX
XXX = 220 220V/50Hz
XXX = 226 220V/60Hz
XXX = 240 240V/50Hz
XXX = 115 110V/50Hz
XXX = 116 110V/60Hz
8810 CYANIDE ANALYZER basic unit 19" panel mounted includes:
- Titration vessel
- Sprinkler
- Measuring cyanide 8474B ion selective electrode
- Reference electrode 8483B
- Temperature sensor PT100
- One reagent peristaltic pump for sample conditioning with Sodium
Hydroxide/EDTA
- Chemical cleaning (368810,56000) with 10 l. canister and reagent level
detector
- Automatic heating device/controller (368810,76000)
- 8810 ISE basic user manual
621=997=103 Installation procedure CYANIDE
21. 21
OPTIONS
NOTE: No automatic calibration available on this application because the compound is chemically unstable.
Grab sample « PROCESS » calibration is recommended.
368810,00100 Reagent canister holder (up to 4 x 10-liter canisters):
L600 x W220 x H200 mm, in PP
08810=A=3500 Second channel option on 8810 analyzer
368810,65000 Manual sample entry system
368810,93000 Spare parts kit (2 years), 8810 ISE analyzer with cyanide electrode
08811=A=20X2
X = 1 110V/50 Hz
X = 3 220V/50 Hz
X = 4 240V/50 Hz
2 channels 8811 Sequencer complete
OR 08811=A=20X3 3 channels 8811 Sequencer complete
OR 08811=A=20X4 4 channels 8811 Sequencer complete
OR 08811=A=20X5 5 channels 8811 Sequencer complete
OR 08811=A=20X6 6 channels 8811 Sequencer complete
OPERATING MATERIAL/YEAR
based on one analysis every 15 minutes
PART NUMBER QTY ITEM DESCRIPTION
590=020=130 2 Grey/grey ismaprene pump tube
Not supplied by
Hach/Hach Lange: to
be purchased locally.
Solid state : 10 kg
OR
Liquid state : 21 l
Sodium hydroxide NaOH in pellets (conditioning)
Sodium hydroxide NaOH solution > 32% (conditioning)
10 kg EDTA tetra sodium tetra hydrate Na4EDTA.4H2O (conditioning)
7 l Concentrated hydrochloric acid HCl ~ 37% (chemical cleaning)
100 g Potassium Cyanide KCN (calibration standards)
22. 22
8810 FLUORIDE ANALYZER
Typical applications: Semi-Conductors, Industrial Waste water, Drinking water
Measuring range: 0-1 000 mg/l F- (programmable) LDL = 0.1 mg/l F-
4-4 000 mg/l F- (programmable) LDL = 4.0 mg/l F-
NOT SUITABLE FOR!!!: Samples with many particles (e.g. sludges)
STANDARD PRODUCT CONFIGURATION
PART NUMBER ITEM DESCRIPTION
368810,32XXX
XXX = 220
220V/50Hz
XXX = 226
220V/60Hz
XXX = 240
240V/50Hz
XXX = 115
110V/50Hz
XXX = 116
110V/60Hz
8810 FLUORIDE ANALYZER basic unit 19" panel mounted includes:
- Titration vessel
- Sprinkler
- Measuring fluoride 8470B ion selective electrode
- Reference electrode 8483B
- Temperature sensor PT100
- One reagent peristaltic pump for sample conditioning
- Chemical cleaning (368810,56000) with 10 l. canister and reagent level
detector
- Automatic heating device/controller (368810,76000)
- 8810 ISE basic user manual
621=997=104 Installation procedure FLUORIDE
25. 25
08810=A=3500 Second channel option on 8810 analyzer
368810,00100 Reagent canister holder (up to 4 x 10-liter canisters):
L600 x W220 x H200 mm, in PP
368810,93002 Spare parts kit (2 years), 8810 ISE analyzer with fluoride electrode
08811=A=20X2
X = 1 110V/50 Hz
X = 3 220V/50 Hz
X = 4 240V/50 Hz
2 channels 8811 Sequencer complete
OR 08811=A=20X3 3 channels 8811 Sequencer complete
OR 08811=A=20X4 4 channels 8811 Sequencer complete
OR 08811=A=20X5 5 channels 8811 Sequencer complete
OR 08811=A=20X6 6 channels 8811 Sequencer complete
OPERATING MATERIAL/YEAR
based on one analysis every 15 minutes
PART NUMBER QTY ITEM DESCRIPTION
590=020=130 2 or 4 Grey/grey ismaprene pump tube
Not supplied by
Hach/Hach Lange: to be
purchased locally.
7 l Concentrated ammonium hydroxide NH4OH ~ 25 %
(conditioning)
If Al 3+, Fe 3+, Si 4+, or SiF6 are present in the sample
(like Semi-Con. / glass manufacturing /…/ applications)
12 kg Sodium Acetate trihydrate CH3COONa.3H2O (conditioning)
12 kg Citric acid monohydrate C6H8O7.H2O (conditioning)
7 l Concentrated hydrochloric acid HCl ~ 37% (chemical cleaning)
100 g Sodium Fluoride NaF (calibration standards)
26. 26
8810 TOTAL HARDNESS ANALYZER
Typical applications: Boiling water, Softening, Water treatment
(e.g. leakage detection of reverse osmosis water treatment)
Measuring range: 0-500 mg/l CaCO3 LDL = 5 mg/l CaCO3
0-10 mg/l CaCO3 LDL = 1 mg/l CaCO3
NOT SUITABLE FOR!!!: Sea water applications or where Cl- or NaCl levels are high
STANDARD PRODUCT CONFIGURATION
PART NUMBER ITEM DESCRIPTION
368810,52XXX
XXX = 220 220V/50Hz
XXX = 240 240V/50Hz
XXX = 115 110V/50Hz
XXX = 116 110V/60Hz
8810 HARDNESS ANALYZER basic unit 19" panel mounted includes:
- Titration vessel
- Sprinkler
- Measuring cupric Cu2+ electrode
- Reference electrode 8483B
- One reagent piston pump for titration
- Chemical cleaning (368810,56000) with 10 l. canister and reagent level
detector
- 8810 ORP basic user manual
368810,71050
Additional reagent piston pump for conditioning with Ammonium Hydroxide /
Ammonium Chloride / Cupric Cu2+ solution
621=997=107
OR
621=997=113
Installation procedure TOTAL HARDNESS 0-10 mg/l CaCO3
Installation procedure TOTAL HARDNESS 0-500 mg/l CaCO3
27. 27
OPTIONS
368810,00100 Reagent canister holder (up to 4 x 10-liter canisters):
L600 x W220 x H200 mm, in PP
368810,60000 Automatic calibration volumetric pump
(includes 10 l. canister and reagent level detector)
368810,92012 Spare parts kit (2 years), 8810 ORP analyzer with copper electrode compatible
with piston pump
08811=A=20X2
X = 1 110V/50 Hz
X = 3 220V/50 Hz
X = 4 240V/50 Hz
2 channels 8811 Sequencer complete
OR 08811=A=20X3 3 channels 8811 Sequencer complete
OR 08811=A=20X4 4 channels 8811 Sequencer complete
OR 08811=A=20X5 5 channels 8811 Sequencer complete
OR 08811=A=20X6 6 channels 8811 Sequencer complete
29. 29
8810 NITRITES ANALYZER
Typical applications: Waste water (nitrification/denitrification)
Measuring range: 0 - 600 mg/l NO2- (programmable)
LDL = 5 mg/l NO2-
NOT SUITABLE FOR!!!: Food processing (e.g. tannery, swine processing)
STANDARD PRODUCT CONFIGURATION
PART NUMBER ITEM DESCRIPTION
368810,20XXX
XXX = 220 220V/50Hz
XXX = 240 240V/50Hz
XXX = 115 110V/50Hz
XXX = 116 110V/60Hz
8810 NITRITES ANALYZER basic unit 19" panel mounted includes:
- Titration vessel
- Sprinkler
- Measuring platinum electrode
- Reference electrode 8483B
- Temperature sensor PT100
- One reagent peristaltic pump for titration
- Chemical cleaning (368810,56000) with 10 l. canister and reagent level
detector
- 8810 ORP basic user manual
368810,700XX
X = 50 24V/50Hz
X = 60 24V/60Hz
Additional reagent peristaltic pump for conditioning with sulphuric acid
621=997=138 Installation procedure NITRITES
30. 30
OPTIONS
NOTE: No automatic calibration available on this application because the compound is chemically unstable.
Grab sample « PROCESS » calibration is recommended.
368810,00100 Reagent canister holder (up to 4 x 10-liter canisters):
L600 x W220 x H200 mm, in PP
368810,65000 Manual sample entry system
368810,92000 Spare parts kit (2 years), 8810 ORP analyzer with platinum electrode
compatible with peristaltic pump
08811=A=20X2
X = 1 110V/50 Hz
X = 3 220V/50 Hz
X = 4 240V/50 Hz
2 channels 8811 Sequencer complete
OR 08811=A=20X3 3 channels 8811 Sequencer complete
OR 08811=A=20X4 4 channels 8811 Sequencer complete
OR 08811=A=20X5 5 channels 8811 Sequencer complete
OR 08811=A=20X6 6 channels 8811 Sequencer complete
OPERATING MATERIAL/YEAR
based on one analysis every 15 minutes
PART NUMBER QTY ITEM DESCRIPTION
590=020=130 2 Grey/grey ismaprene pump tube
(Not applicable if piston pump is used)
Not supplied by
Hach/Hach Lange: to
be purchased locally.
250 g Potassium permanganate KMnO4 (titration)
10 l. Concentrated sulphuric acid H2SO4 95-97% (conditioning)
10 l. Concentrated sulphuric acid H2SO4 95-97% (chem. cleaning)
31. 31
8810 SULPHIDE ANALYZER
Typical applications: Industrial Waste water (Refinery/Petrochemistry, final effluent)
Measuring range: 0-1 mg/l S2- LDL = 0.1 mg/l S2-
0-150 mg/l S2- (programmable) LDL = 2.0 mg/l S2-
NOT SUITABLE FOR!!!: Food processing (e.g. tannery, swine processing)
Or if sample contains sea water (Cl- or NaCl)
STANDARD PRODUCT CONFIGURATION
PART NUMBER ITEM DESCRIPTION
32. 32
368810,24XXX
XXX = 220 220V/50Hz
XXX = 240 240V/50Hz
XXX = 115 110V/50Hz
XXX = 116 110V/60Hz
OR
368810,54XXX
8810 SULPHIDE ANALYZER basic unit 19" panel mounted includes:
- Titration vessel
- Sprinkler
- Measuring sulphide 8490K electrode
- Reference electrode 8483B
- One reagent peristaltic pump for titration
- Dark canister (368810,84000), set of 12 black Viton pump tubes and
black Viton connection tube Viton black
- Chemical cleaning (368810,56000) with 10 l. canister and reagent level
detector
- 8810 ORP basic user manual
Same analyzer with a piston pump instead of a peristaltic pump for titration
368810,700XX
X = 50 24V/50Hz
X = 60 24V/60Hz
OR
368810,71050
Additional reagent peristaltic pump for conditioning with Sodium hydroxide
Additional reagent piston pump for conditioning with Sodium hydroxide
621=997=125 Installation procedure SULPHIDE 0-1mg/l S2-
33. 33
OPTIONS
NOTE: No automatic calibration available on this application because the compound is chemically unstable.
Grab sample « PROCESS » calibration is recommended.
368810,00100 Reagent canister holder (up to 4 x 10-liter canisters):
L600 x W220 x H200 mm, in PP
368810,92004 Spare parts kit (2 years), 8810 ORP analyzer with sulphide electrode
08811=A=20X2
X = 1 110V/50 Hz
X = 3 220V/50 Hz
X = 4 240V/50 Hz
2 channels 8811 Sequencer complete
OR 08811=A=20X3 3 channels 8811 Sequencer complete
OR 08811=A=20X4 4 channels 8811 Sequencer complete
OR 08811=A=20X5 5 channels 8811 Sequencer complete
OR 08811=A=20X6 6 channels 8811 Sequencer complete
34. 34
368810,21XXX
XXX = 220 220V/50Hz
XXX = 240 240V/50Hz
XXX = 115 110V/50Hz
XXX = 116 110V/60Hz
OR
368810,51XXX
8810 SULPHIDE ANALYZER basic unit 19" panel mounted includes:
- Titration vessel
- Sprinkler
- Measuring Silver stick electrode
- Reference electrode 8483B
- One reagent peristaltic pump for titration
- Chemical cleaning (368810,56000) with 10 l. canister and reagent level
detector
- 8810 ORP basic user manual
Same analyzer with a piston pump instead of a peristaltic pump for titration
368810,700XX
X = 50 24V/50Hz
X = 60 24V/60Hz
OR
368810,71050
Additional reagent peristaltic pump for conditioning with Sodium hydroxide
Additional reagent piston pump for conditioning with Sodium hydroxide
621=997=117 Installation procedure SULPHIDE 0-100mg/l S2-
35. 35
OPTIONS
NOTE: No automatic calibration available on this application because the compound is chemically unstable.
Grab sample « PROCESS » calibration is recommended.
368810,00100 Reagent canister holder (up to 4 x 10-liter canisters):
L600 x W220 x H200 mm, in PP
368810,65000 Manual sample entry system
368810,920X1 Spare parts kit (2 years), 8810 ORP analyzer with silver stick electrode
compatible with pump X : 0 peristaltic; 1 piston.
08811=A=20X2
X = 1 110V/50 Hz
X = 3 220V/50 Hz
X = 4 240V/50 Hz
2 channels 8811 Sequencer complete
OR 08811=A=20X3 3 channels 8811 Sequencer complete
OR 08811=A=20X4 4 channels 8811 Sequencer complete
OR 08811=A=20X5 5 channels 8811 Sequencer complete
OR 08811=A=20X6 6 channels 8811 Sequencer complete