SlideShare a Scribd company logo
1 of 1
Download to read offline
LTQ, Xcalibur and BioBasic are trademarks of Thermo Electron Corporation. TurboSEQUEST is a trademark of University of Washington. PepMap is a trademark of Dionex Corporation.
Ettan and CAF are trademarks of Amersham Biosciences Limited.
Amersham & Amersham Biosciences are trademarks of Amersham PLC, Amersham Biosciences UK Limited, Amersham Place, Little Chalfont, Buckinghamshire, HP7 9NA, England
Amersham Biosciences AB, SE-751 84 Uppsala, Sweden, Amersham Biosciences Corp., 800 Centennial Avenue, PO Box 1327, Piscataway, NJ 0855, USA, Amersham Biosciences Europe GmbH, Munzinger Strasse 9, D-79111, Freiburg, Germany
All goods and services are sold subject to the terms and conditions of sale of the company within the Amersham Biosciences group which supplies them. A copy of these terms and conditions is available on request. © Amersham Biosciences AB, 2004-All rights reserved.
* To whom correspondence should be addressed: Tel: +46-18-612 02 81, Fax: +46-18-612 18 26, e-mail: anders.tangen@amersham.com
The LC-MS Workflow: Is MALDI or ESI the Preferable Ionization Method?
Anders Tangen,*a Hanno Ehring,a John Flensburg,a Maria C. Prieto Conawayb and Henrik Wadenstena
aAmersham Biosciences AB, a GE Healthcare Company, Björkgatan 30, SE-751 84 Uppsala, Sweden
bThermo Electron Corp. 355 River Oaks Parkway, San Jose, CA, USA, 95134
µLC Configuration
Introduction
The use of the LC-MS workflow as a complement to 2DE-MS
workflow is growing in popularity.1, 2 Some of the 2DE limitations
are more easily overcome with the use of LC (hydrophobic or
extremely small/large or basic proteins), in part because the typical
sample is a tryptic digest – not a full length protein. Traditionally,
ESI mass spectrometers have been used as detectors for LC, due
to the simplicity of the interface. Today, many researchers also do
LC with MALDI-MS detection, either as a complement or an
alternative to LC-ESI-MS. Analysis speed and sensitivity are the
most frequently used arguments for the proper choice of instrument
configuration. However, from theoretical considerations one could
argue that the two ionization techniques are complementary by
nature.
Methods
MALDI-TOF, ESI-TOF, MALDI-LTQ and ESI-LTQ instruments have
been evaluated as detectors for the LC-MS workflow. Parameters
such as sensitivity, analysis time, ease of use and applicability for
peptides with different physiochemical characteristics have been
considered. Multivariate analysis has been used to find correlations
between fundamental peptide characteristics and probability of ESI-
MS or MALDI-MS detection.
Summary
Higher sequence coverage is obtained with the use of ESI.
Compared to ESI, relative MALDI detectability is correlated with
larger, more hydrophobic peptides, having lower pI values. The
optimum choice of MS detector is also depending on the type of
application.
µLC-ESI-TOF vs. MALDI-TOF:
Which peptides are detected?
MALDI only: 8 peptides
ESI & MALDI common: 15 peptides
LC-ESI only: 18 peptides
Total: 41 peptides detected (ESI + MALDI)
Correlation Between Peptide
Characteristics and ESI or MALDI
Detectability
The following peptide parameters were tested for influence upon
ESI and MALDI detectability: Hydrophobicity, pI, mass, amino acid
coverage and cystein content
PLS Loading Map (Only detected peptides included in model)
-”Maximum correlation between X- and Y-variables”
Comments to Relative Detectability
Compared to MALDI-TOF, µLC-ESI-TOF detectability is correlated
with smaller peptides having higher pI values and lower RPC
retention times.
Compared to µLC-ESI-TOF, MALDI-TOF detectability is correlated
with larger peptides, having lower pI values and higher RPC
retention times.
For a tryptic digest of BSA, most of the cysteines are located in the
larger peptides. In general, the larger peptides are harder to detect.
Experimental Conditions
Samples.
For ESI-TOF and MALDI-TOF experiments :
Tryptic digest of Bovine Serum Albumin (Michrom, USA).
For ESI-LTQ and MALDI-LTQ experiments:
Tryptic Digest (LC Packings, USA) of 6 Proteins (Cytochrome c,
Lysozyme, Alcohol dehydrogenase, Bovine serum albumin, Apo-
transferrin, β-Galactosidase), lyophilized, CAF modified using
standard protocol.
Chromatography Systems.
For ESI-TOF and MALDI-TOF experiments :
Ettan™ microLC (Amersham Biosciences, UK).
For ESI-LTQ and MALDI-LTQ experiments:
Ettan™ MDLC (Amersham Biosciences, UK). Controlled by
Xcalibur (Thermo Electron, USA) 1.4 via the VI interface.
MDLC dimension 1:
Column: 0.3 x 30mm BioBasic SCX, 5 µm, 300 Å (Thermo Electron,
USA). Mobile phase: 0.1% FA in H2O. Flow Rate: 10 µl/min
Flow through fraction analysed
MDLC dimension 2:
Trap column: 0.3 x 5mm, PepMap C18, 5µm, 100Å (Dionex/LC
Packings, USA). RPC column: 75µm x 150mm, PepMap C18, 3µm,
100Å (Dionex/LC Packings, USA). Buffer A: 0.1% formic acid.
Buffer B: 0.1% formic acid in 84% ACN. Gradient: 0–40% B in 20
min. Flow rate: 300 nl/min
MDLC-MALDI Interface: Probot (Dionex/LC Packings, USA).
MS.
For ESI-TOF and MALDI-TOF experiments :
Enterprise ESI-TOF (Analytica of Branford, USA).
Ettan™ MALDI-ToF (Amersham Biosciences, UK).
For ESI-LTQ and MALDI-LTQ experiments:
Finnigan ESI-LTQ and Finnigan vMALDI-LTQ (Thermo Electron,
USA). Dynamic exclusion activated
Multivariate Analysis.
SIMCA-P 8.0 (Umetrics, Sweden)
Phe-Gly-Glu-Arg-
Val-Ala-Ser-Leu-Arg-
Gln-Arg-Leu-Arg-
Ala-Phe-Asp-Glu-Lys-
Gly-Val-Phe-Arg-Arg-
Lys-Phe-Trp-Gly-Lys-
Ala-Trp-Ser-Val-Ala-Arg-
Ser-Glu-Ile-Ala-His-Arg-
Leu-Val-Thr-Asp-Leu-Thr-Lys-
Tyr-Leu-Tyr-Glu-Ile-Ala-Arg-
Asp-Leu-Gly-Glu-Glu-His-Phe-Lys-
Ser-His-Cys-Ile-Ala-Glu-Val-Glu-Lys-
Cys-Cys-Thr-Glu-Ser-Leu-Val-Asn-Arg-
Leu-Val-Asn-Glu-Leu-Thr-Glu-Phe-Ala-Lys-
Cys-Cys-Thr-Lys-Pro-Glu-Ser-Glu-Arg-
Phe-Lys-Asp-Leu-Gly-Glu-Glu-His-Phe-Lys-
His-Leu-Val-Asp-Glu-Pro-Gln-Asn-Leu-Ile-Lys-
Ser-Leu-His-Thr-Leu-Phe-Gly-Asp-Glu-Leu-Cys-Lys-
Arg-His-Pro-Glu-Tyr-Ala-Val-Ser-Val-Leu-Leu-Arg-
Tyr-Ile-Cys-Asp-Asn-Gln-Asp-Thr-Ile-Ser-Ser-Lys-
Thr-Cys-Val-Ala-Asp-Glu-Ser-His-Ala-Gly-Cys-Glu-Lys-
Leu-Arg-Cys-Ala-Ser-Ile-Gln-Lys-Phe-Gly-Glu-Arg-
Leu-Gly-Glu-Tyr-Gly-Phe-Gln-Asn-Ala-Leu-Ile-Val-Arg-
Leu-Lys-Glu-Cys-Cys-Asp-Lys-Pro-Leu-Leu-Glu-Lys-
Asp-Asp-Pro-His-Ala-Cys-Tyr-Ser-Thr-Val-Phe-Asp-Lys-
Asp-Ala-Phe-Leu-Gly-Ser-Phe-Leu-Tyr-Glu-Tyr-Ser-Arg-
Leu-Lys-Pro-Asp-Pro-Asn-Thr-Leu-Cys-Asp-Glu-Phe-Lys-
Lys-Val-Pro-Gln-Val-Ser-Thr-Pro-Thr-Leu-Val-Glu-Val-Ser-Arg-
Met-Pro-Cys-Thr-Glu-Asp-Tyr-Leu-Ser-Leu-Ile-Leu-Asn-Arg-
Tyr-Asn-Gly-Val-Phe-Gln-Glu-Cys-Cys-Gln-Ala-Glu-Asp-Lys-
Arg-Pro-Cys-Phe-Ser-Ala-Leu-Thr-Pro-Asp-Glu-Thr-Tyr-Val-Pro-Lys-
Leu-Phe-Thr-Phe-His-Ala-Asp-Ile-Cys-Thr-Leu-Pro-Asp-Thr-Glu-Lys-
Cys-Cys-Ala-Ala-Asp-Asp-Lys-Glu-Ala-Cys-Phe-Ala-Val-Glu-Gly-Pro-Lys-
Leu-Lys-Pro-Asp-Pro-Asn-Thr-Leu-Cys-Asp-Glu-Phe-Lys-Ala-Asp-Glu-Lys-
Glu-Ala-Cys-Phe-Ala-Val-Glu-Gly-Pro-Lys-Leu-Val-Val-Ser-Thr-Gln-Thr-Ala-Leu-Ala-
Val-His-Lys-Glu-Cys-Cys-His-Gly-Asp-Leu-Leu-Glu-Cys-Ala-Asp-Asp-Arg-
Glu-Cys-Cys-His-Gly-Asp-Leu-Leu-Glu-Cys-Ala-Asp-Asp-Arg-Ala-Asp-Leu-Ala-Lys-
Tyr-Asn-Gly-Val-Phe-Gln-Glu-Cys-Cys-Gln-Ala-Glu-Asp-Lys-Gly-Ala-Cys-Leu-Leu-Pro-Lys-
Gln-Glu-Pro-Glu-Arg-Asn-Glu-Cys-Phe-Leu-Ser-His-Lys-Asp-Asp-Ser-Pro-Asp-Leu-Pro-Lys-
Val-His-Lys-Glu-Cys-Cys-His-Gly-Asp-Leu-Leu-Glu-Cys-Ala-Asp-Asp-Arg-Ala-Asp-Leu-Ala-Lys-
Glu-Tyr-Glu-Ala-Thr-Leu-Glu-Glu-Cys-Cys-Ala-Lys-Asp-Asp-Pro-His-Ala-Cys-Tyr-Ser-Thr-Val-Phe-Asp-Lys-
ESI
MALDI
ESI & MALDI
Discussion
Analysis speed and sensitivity are the most frequently used
arguments for the proper choice of instrument configuration. For a
single sample analysis, less time is obviously needed for MDLC-
ESI than for MDLC-MALDI. This is the other way round for the
analysis of large sample amounts. However, MALDI seem to
provide significantly lower sequence coverage than ESI (using TOF
or LTQ mass analyzers) for the samples tested in this study. The
difference is bigger for high sample amounts. For laboratories
handling a large number of samples, a small fraction of each
sample may be analysed by MALDI for screening purposes, while
the greater fraction of each sample is analysed by ESI.
The optimum choice of acid for the reversed phase chromatography
eluent is different for ESI and MALDI ionization, respectively. In the
case of nano LC-ESI, formic acid (FA) is the preferable acid (Tri
Fluoro acetic acid, TFA, is frequently avoided due to reported
problems with arcing at flow rates in the nl/min range). However,
TFA is reported to provide higher chromatographic resolution than
FA. In the case of MALDI, TFA is one of the preferred acids.
Consequently, for the nano LC-MALDI interface, the optimum acid
can be used regarding both chromatographic resolution and MS
sensitivity.
From theoretical considerations one could argue that the two
ionization techniques are complementary by nature. While the
relatively high mobility of low mass peptides make them easy to
detect using ESI, these peptides are generally harder to detect
using MALDI because the peptide mass is relatively close to the
mass of the typical matrix molecules used.
References
1.Johan Axelman, Henrik Wadensten, Staffan Renlund, Anders
Tangen, Axel Parbel, Hans-Rudolf Höpker, “The use of 2DLC-
MS/MS in proteome characterization – optimization of the LC
step”, ASMS 2003, Montreal
2.Anders Tangen, Henrik Wadensten, Axel Parbel, Matthias Berg
and Hans-Rudolf Höpker, “The LC-MS Workflow: A Comparative
Study of High Throughput Nano LC and MDLC Designs”, ABRF
2004, Portland
-0.60
-0.40
-0.20
0.00
0.20
0.40
0.60
0.80
-0.60 -0.50 -0.40 -0.30 -0.20 -0.10 0.00 0.10 0.20 0.30 0.40
wc[2]
wc[1]
AA Cover
MH+
HPLC
pI
Cystein
LC-ESI-T
MALDI-TO
MDLC-ESI-LTQ vs. MDLC-MALDI-LTQ:
Improvement factor for sequence coverage using ESI-LTQ instead of MALDI-LTQ.
Sample: CAF modified tryptic peptides from six proteins purified with salt step MDLC
0
2
4
6
8
10
BS
A
Transferrin
Lysozym
e
AlcoholdehydrogenaseB-G
alactosidase
C
ytochrom
e
C
ESI/MALDISeq.Cov.
MDLC Salt Step Configuration

More Related Content

Similar to Poster_ASMS_MALDI_vs_ESI

Dui d in urine poster itsp
Dui d in urine poster itspDui d in urine poster itsp
Dui d in urine poster itspMark Hayward
 
Chemiluminescence Immunoassay (CLIA) Label Alkaline Phosphatase (ALP).pdf
Chemiluminescence Immunoassay (CLIA) Label Alkaline Phosphatase (ALP).pdfChemiluminescence Immunoassay (CLIA) Label Alkaline Phosphatase (ALP).pdf
Chemiluminescence Immunoassay (CLIA) Label Alkaline Phosphatase (ALP).pdfDoriaFang
 
Theoretical evaluation of shotgun proteomic analysis strategies; Peptide obse...
Theoretical evaluation of shotgun proteomic analysis strategies; Peptide obse...Theoretical evaluation of shotgun proteomic analysis strategies; Peptide obse...
Theoretical evaluation of shotgun proteomic analysis strategies; Peptide obse...Keiji Takamoto
 
LIQUID CHROMATOGRAPHY-MASS SPECTROSCOPY (LC-MS)
LIQUID CHROMATOGRAPHY-MASS SPECTROSCOPY (LC-MS)LIQUID CHROMATOGRAPHY-MASS SPECTROSCOPY (LC-MS)
LIQUID CHROMATOGRAPHY-MASS SPECTROSCOPY (LC-MS)GOPALASATHEESKUMAR K
 
Proteomics 2009 V9p1683
Proteomics 2009 V9p1683Proteomics 2009 V9p1683
Proteomics 2009 V9p1683jcruzsilva
 
Bioanalytical Method Development and Validation for Simultaneous Estimation o...
Bioanalytical Method Development and Validation for Simultaneous Estimation o...Bioanalytical Method Development and Validation for Simultaneous Estimation o...
Bioanalytical Method Development and Validation for Simultaneous Estimation o...BRNSSPublicationHubI
 
ICP-MS for Elemental Analysis in Drug Development
ICP-MS for Elemental Analysis in Drug DevelopmentICP-MS for Elemental Analysis in Drug Development
ICP-MS for Elemental Analysis in Drug DevelopmentQPS Holdings, LLC
 
Proteomics_Chapter 3 Protein Identification.ppt
Proteomics_Chapter 3 Protein Identification.pptProteomics_Chapter 3 Protein Identification.ppt
Proteomics_Chapter 3 Protein Identification.pptZaldaaZaldaa
 
IRSAE aquatic ecology 28 June 2018 metabolomics
IRSAE aquatic ecology 28 June 2018 metabolomicsIRSAE aquatic ecology 28 June 2018 metabolomics
IRSAE aquatic ecology 28 June 2018 metabolomicsPanagiotis Arapitsas
 
Adsorptive stripping differential pulse voltammetry determination of rivastig...
Adsorptive stripping differential pulse voltammetry determination of rivastig...Adsorptive stripping differential pulse voltammetry determination of rivastig...
Adsorptive stripping differential pulse voltammetry determination of rivastig...Pramod Kalambate
 
SCU LC-MS Agilent 6100 training_060710
SCU LC-MS Agilent 6100 training_060710SCU LC-MS Agilent 6100 training_060710
SCU LC-MS Agilent 6100 training_060710pead
 
Expanding Surface Plasmon Resonance Capabilities with Reichert
Expanding Surface Plasmon Resonance Capabilities with ReichertExpanding Surface Plasmon Resonance Capabilities with Reichert
Expanding Surface Plasmon Resonance Capabilities with ReichertReichertSPR
 
MALDI-TOF: Pricinple and Its Application in Biochemistry and Biotechnology
MALDI-TOF: Pricinple  and Its Application in Biochemistry and BiotechnologyMALDI-TOF: Pricinple  and Its Application in Biochemistry and Biotechnology
MALDI-TOF: Pricinple and Its Application in Biochemistry and BiotechnologyDevakumar Jain
 
Proteomics 2009 V9p1696
Proteomics 2009 V9p1696Proteomics 2009 V9p1696
Proteomics 2009 V9p1696jcruzsilva
 
Saravanan Kumar,V.Siva Reddy - Workshop on Proteomics
Saravanan Kumar,V.Siva Reddy - Workshop on ProteomicsSaravanan Kumar,V.Siva Reddy - Workshop on Proteomics
Saravanan Kumar,V.Siva Reddy - Workshop on ProteomicsSARAVANAN KUMAR
 
ELECTROCHEMICAL SENSORS
ELECTROCHEMICAL SENSORS ELECTROCHEMICAL SENSORS
ELECTROCHEMICAL SENSORS Lavaraj Devkota
 
5 Reasons Aptamers for Lateral Flow
5 Reasons Aptamers for Lateral Flow5 Reasons Aptamers for Lateral Flow
5 Reasons Aptamers for Lateral FlowLisa Thurston
 

Similar to Poster_ASMS_MALDI_vs_ESI (20)

Dui d in urine poster itsp
Dui d in urine poster itspDui d in urine poster itsp
Dui d in urine poster itsp
 
Chemiluminescence Immunoassay (CLIA) Label Alkaline Phosphatase (ALP).pdf
Chemiluminescence Immunoassay (CLIA) Label Alkaline Phosphatase (ALP).pdfChemiluminescence Immunoassay (CLIA) Label Alkaline Phosphatase (ALP).pdf
Chemiluminescence Immunoassay (CLIA) Label Alkaline Phosphatase (ALP).pdf
 
Theoretical evaluation of shotgun proteomic analysis strategies; Peptide obse...
Theoretical evaluation of shotgun proteomic analysis strategies; Peptide obse...Theoretical evaluation of shotgun proteomic analysis strategies; Peptide obse...
Theoretical evaluation of shotgun proteomic analysis strategies; Peptide obse...
 
LIQUID CHROMATOGRAPHY-MASS SPECTROSCOPY (LC-MS)
LIQUID CHROMATOGRAPHY-MASS SPECTROSCOPY (LC-MS)LIQUID CHROMATOGRAPHY-MASS SPECTROSCOPY (LC-MS)
LIQUID CHROMATOGRAPHY-MASS SPECTROSCOPY (LC-MS)
 
Proteomics 2009 V9p1683
Proteomics 2009 V9p1683Proteomics 2009 V9p1683
Proteomics 2009 V9p1683
 
Bioanalytical Method Development and Validation for Simultaneous Estimation o...
Bioanalytical Method Development and Validation for Simultaneous Estimation o...Bioanalytical Method Development and Validation for Simultaneous Estimation o...
Bioanalytical Method Development and Validation for Simultaneous Estimation o...
 
ICP-MS for Elemental Analysis in Drug Development
ICP-MS for Elemental Analysis in Drug DevelopmentICP-MS for Elemental Analysis in Drug Development
ICP-MS for Elemental Analysis in Drug Development
 
Mayuri shitre
Mayuri shitreMayuri shitre
Mayuri shitre
 
Proteomics_Chapter 3 Protein Identification.ppt
Proteomics_Chapter 3 Protein Identification.pptProteomics_Chapter 3 Protein Identification.ppt
Proteomics_Chapter 3 Protein Identification.ppt
 
IRSAE aquatic ecology 28 June 2018 metabolomics
IRSAE aquatic ecology 28 June 2018 metabolomicsIRSAE aquatic ecology 28 June 2018 metabolomics
IRSAE aquatic ecology 28 June 2018 metabolomics
 
Adsorptive stripping differential pulse voltammetry determination of rivastig...
Adsorptive stripping differential pulse voltammetry determination of rivastig...Adsorptive stripping differential pulse voltammetry determination of rivastig...
Adsorptive stripping differential pulse voltammetry determination of rivastig...
 
SCU LC-MS Agilent 6100 training_060710
SCU LC-MS Agilent 6100 training_060710SCU LC-MS Agilent 6100 training_060710
SCU LC-MS Agilent 6100 training_060710
 
Expanding Surface Plasmon Resonance Capabilities with Reichert
Expanding Surface Plasmon Resonance Capabilities with ReichertExpanding Surface Plasmon Resonance Capabilities with Reichert
Expanding Surface Plasmon Resonance Capabilities with Reichert
 
Presentation 6
Presentation 6Presentation 6
Presentation 6
 
MALDI-TOF: Pricinple and Its Application in Biochemistry and Biotechnology
MALDI-TOF: Pricinple  and Its Application in Biochemistry and BiotechnologyMALDI-TOF: Pricinple  and Its Application in Biochemistry and Biotechnology
MALDI-TOF: Pricinple and Its Application in Biochemistry and Biotechnology
 
Proteomics 2009 V9p1696
Proteomics 2009 V9p1696Proteomics 2009 V9p1696
Proteomics 2009 V9p1696
 
Saravanan Kumar,V.Siva Reddy - Workshop on Proteomics
Saravanan Kumar,V.Siva Reddy - Workshop on ProteomicsSaravanan Kumar,V.Siva Reddy - Workshop on Proteomics
Saravanan Kumar,V.Siva Reddy - Workshop on Proteomics
 
ELECTROCHEMICAL SENSORS
ELECTROCHEMICAL SENSORS ELECTROCHEMICAL SENSORS
ELECTROCHEMICAL SENSORS
 
JClinChem_2003
JClinChem_2003JClinChem_2003
JClinChem_2003
 
5 Reasons Aptamers for Lateral Flow
5 Reasons Aptamers for Lateral Flow5 Reasons Aptamers for Lateral Flow
5 Reasons Aptamers for Lateral Flow
 

Poster_ASMS_MALDI_vs_ESI

  • 1. LTQ, Xcalibur and BioBasic are trademarks of Thermo Electron Corporation. TurboSEQUEST is a trademark of University of Washington. PepMap is a trademark of Dionex Corporation. Ettan and CAF are trademarks of Amersham Biosciences Limited. Amersham & Amersham Biosciences are trademarks of Amersham PLC, Amersham Biosciences UK Limited, Amersham Place, Little Chalfont, Buckinghamshire, HP7 9NA, England Amersham Biosciences AB, SE-751 84 Uppsala, Sweden, Amersham Biosciences Corp., 800 Centennial Avenue, PO Box 1327, Piscataway, NJ 0855, USA, Amersham Biosciences Europe GmbH, Munzinger Strasse 9, D-79111, Freiburg, Germany All goods and services are sold subject to the terms and conditions of sale of the company within the Amersham Biosciences group which supplies them. A copy of these terms and conditions is available on request. © Amersham Biosciences AB, 2004-All rights reserved. * To whom correspondence should be addressed: Tel: +46-18-612 02 81, Fax: +46-18-612 18 26, e-mail: anders.tangen@amersham.com The LC-MS Workflow: Is MALDI or ESI the Preferable Ionization Method? Anders Tangen,*a Hanno Ehring,a John Flensburg,a Maria C. Prieto Conawayb and Henrik Wadenstena aAmersham Biosciences AB, a GE Healthcare Company, Björkgatan 30, SE-751 84 Uppsala, Sweden bThermo Electron Corp. 355 River Oaks Parkway, San Jose, CA, USA, 95134 µLC Configuration Introduction The use of the LC-MS workflow as a complement to 2DE-MS workflow is growing in popularity.1, 2 Some of the 2DE limitations are more easily overcome with the use of LC (hydrophobic or extremely small/large or basic proteins), in part because the typical sample is a tryptic digest – not a full length protein. Traditionally, ESI mass spectrometers have been used as detectors for LC, due to the simplicity of the interface. Today, many researchers also do LC with MALDI-MS detection, either as a complement or an alternative to LC-ESI-MS. Analysis speed and sensitivity are the most frequently used arguments for the proper choice of instrument configuration. However, from theoretical considerations one could argue that the two ionization techniques are complementary by nature. Methods MALDI-TOF, ESI-TOF, MALDI-LTQ and ESI-LTQ instruments have been evaluated as detectors for the LC-MS workflow. Parameters such as sensitivity, analysis time, ease of use and applicability for peptides with different physiochemical characteristics have been considered. Multivariate analysis has been used to find correlations between fundamental peptide characteristics and probability of ESI- MS or MALDI-MS detection. Summary Higher sequence coverage is obtained with the use of ESI. Compared to ESI, relative MALDI detectability is correlated with larger, more hydrophobic peptides, having lower pI values. The optimum choice of MS detector is also depending on the type of application. µLC-ESI-TOF vs. MALDI-TOF: Which peptides are detected? MALDI only: 8 peptides ESI & MALDI common: 15 peptides LC-ESI only: 18 peptides Total: 41 peptides detected (ESI + MALDI) Correlation Between Peptide Characteristics and ESI or MALDI Detectability The following peptide parameters were tested for influence upon ESI and MALDI detectability: Hydrophobicity, pI, mass, amino acid coverage and cystein content PLS Loading Map (Only detected peptides included in model) -”Maximum correlation between X- and Y-variables” Comments to Relative Detectability Compared to MALDI-TOF, µLC-ESI-TOF detectability is correlated with smaller peptides having higher pI values and lower RPC retention times. Compared to µLC-ESI-TOF, MALDI-TOF detectability is correlated with larger peptides, having lower pI values and higher RPC retention times. For a tryptic digest of BSA, most of the cysteines are located in the larger peptides. In general, the larger peptides are harder to detect. Experimental Conditions Samples. For ESI-TOF and MALDI-TOF experiments : Tryptic digest of Bovine Serum Albumin (Michrom, USA). For ESI-LTQ and MALDI-LTQ experiments: Tryptic Digest (LC Packings, USA) of 6 Proteins (Cytochrome c, Lysozyme, Alcohol dehydrogenase, Bovine serum albumin, Apo- transferrin, β-Galactosidase), lyophilized, CAF modified using standard protocol. Chromatography Systems. For ESI-TOF and MALDI-TOF experiments : Ettan™ microLC (Amersham Biosciences, UK). For ESI-LTQ and MALDI-LTQ experiments: Ettan™ MDLC (Amersham Biosciences, UK). Controlled by Xcalibur (Thermo Electron, USA) 1.4 via the VI interface. MDLC dimension 1: Column: 0.3 x 30mm BioBasic SCX, 5 µm, 300 Å (Thermo Electron, USA). Mobile phase: 0.1% FA in H2O. Flow Rate: 10 µl/min Flow through fraction analysed MDLC dimension 2: Trap column: 0.3 x 5mm, PepMap C18, 5µm, 100Å (Dionex/LC Packings, USA). RPC column: 75µm x 150mm, PepMap C18, 3µm, 100Å (Dionex/LC Packings, USA). Buffer A: 0.1% formic acid. Buffer B: 0.1% formic acid in 84% ACN. Gradient: 0–40% B in 20 min. Flow rate: 300 nl/min MDLC-MALDI Interface: Probot (Dionex/LC Packings, USA). MS. For ESI-TOF and MALDI-TOF experiments : Enterprise ESI-TOF (Analytica of Branford, USA). Ettan™ MALDI-ToF (Amersham Biosciences, UK). For ESI-LTQ and MALDI-LTQ experiments: Finnigan ESI-LTQ and Finnigan vMALDI-LTQ (Thermo Electron, USA). Dynamic exclusion activated Multivariate Analysis. SIMCA-P 8.0 (Umetrics, Sweden) Phe-Gly-Glu-Arg- Val-Ala-Ser-Leu-Arg- Gln-Arg-Leu-Arg- Ala-Phe-Asp-Glu-Lys- Gly-Val-Phe-Arg-Arg- Lys-Phe-Trp-Gly-Lys- Ala-Trp-Ser-Val-Ala-Arg- Ser-Glu-Ile-Ala-His-Arg- Leu-Val-Thr-Asp-Leu-Thr-Lys- Tyr-Leu-Tyr-Glu-Ile-Ala-Arg- Asp-Leu-Gly-Glu-Glu-His-Phe-Lys- Ser-His-Cys-Ile-Ala-Glu-Val-Glu-Lys- Cys-Cys-Thr-Glu-Ser-Leu-Val-Asn-Arg- Leu-Val-Asn-Glu-Leu-Thr-Glu-Phe-Ala-Lys- Cys-Cys-Thr-Lys-Pro-Glu-Ser-Glu-Arg- Phe-Lys-Asp-Leu-Gly-Glu-Glu-His-Phe-Lys- His-Leu-Val-Asp-Glu-Pro-Gln-Asn-Leu-Ile-Lys- Ser-Leu-His-Thr-Leu-Phe-Gly-Asp-Glu-Leu-Cys-Lys- Arg-His-Pro-Glu-Tyr-Ala-Val-Ser-Val-Leu-Leu-Arg- Tyr-Ile-Cys-Asp-Asn-Gln-Asp-Thr-Ile-Ser-Ser-Lys- Thr-Cys-Val-Ala-Asp-Glu-Ser-His-Ala-Gly-Cys-Glu-Lys- Leu-Arg-Cys-Ala-Ser-Ile-Gln-Lys-Phe-Gly-Glu-Arg- Leu-Gly-Glu-Tyr-Gly-Phe-Gln-Asn-Ala-Leu-Ile-Val-Arg- Leu-Lys-Glu-Cys-Cys-Asp-Lys-Pro-Leu-Leu-Glu-Lys- Asp-Asp-Pro-His-Ala-Cys-Tyr-Ser-Thr-Val-Phe-Asp-Lys- Asp-Ala-Phe-Leu-Gly-Ser-Phe-Leu-Tyr-Glu-Tyr-Ser-Arg- Leu-Lys-Pro-Asp-Pro-Asn-Thr-Leu-Cys-Asp-Glu-Phe-Lys- Lys-Val-Pro-Gln-Val-Ser-Thr-Pro-Thr-Leu-Val-Glu-Val-Ser-Arg- Met-Pro-Cys-Thr-Glu-Asp-Tyr-Leu-Ser-Leu-Ile-Leu-Asn-Arg- Tyr-Asn-Gly-Val-Phe-Gln-Glu-Cys-Cys-Gln-Ala-Glu-Asp-Lys- Arg-Pro-Cys-Phe-Ser-Ala-Leu-Thr-Pro-Asp-Glu-Thr-Tyr-Val-Pro-Lys- Leu-Phe-Thr-Phe-His-Ala-Asp-Ile-Cys-Thr-Leu-Pro-Asp-Thr-Glu-Lys- Cys-Cys-Ala-Ala-Asp-Asp-Lys-Glu-Ala-Cys-Phe-Ala-Val-Glu-Gly-Pro-Lys- Leu-Lys-Pro-Asp-Pro-Asn-Thr-Leu-Cys-Asp-Glu-Phe-Lys-Ala-Asp-Glu-Lys- Glu-Ala-Cys-Phe-Ala-Val-Glu-Gly-Pro-Lys-Leu-Val-Val-Ser-Thr-Gln-Thr-Ala-Leu-Ala- Val-His-Lys-Glu-Cys-Cys-His-Gly-Asp-Leu-Leu-Glu-Cys-Ala-Asp-Asp-Arg- Glu-Cys-Cys-His-Gly-Asp-Leu-Leu-Glu-Cys-Ala-Asp-Asp-Arg-Ala-Asp-Leu-Ala-Lys- Tyr-Asn-Gly-Val-Phe-Gln-Glu-Cys-Cys-Gln-Ala-Glu-Asp-Lys-Gly-Ala-Cys-Leu-Leu-Pro-Lys- Gln-Glu-Pro-Glu-Arg-Asn-Glu-Cys-Phe-Leu-Ser-His-Lys-Asp-Asp-Ser-Pro-Asp-Leu-Pro-Lys- Val-His-Lys-Glu-Cys-Cys-His-Gly-Asp-Leu-Leu-Glu-Cys-Ala-Asp-Asp-Arg-Ala-Asp-Leu-Ala-Lys- Glu-Tyr-Glu-Ala-Thr-Leu-Glu-Glu-Cys-Cys-Ala-Lys-Asp-Asp-Pro-His-Ala-Cys-Tyr-Ser-Thr-Val-Phe-Asp-Lys- ESI MALDI ESI & MALDI Discussion Analysis speed and sensitivity are the most frequently used arguments for the proper choice of instrument configuration. For a single sample analysis, less time is obviously needed for MDLC- ESI than for MDLC-MALDI. This is the other way round for the analysis of large sample amounts. However, MALDI seem to provide significantly lower sequence coverage than ESI (using TOF or LTQ mass analyzers) for the samples tested in this study. The difference is bigger for high sample amounts. For laboratories handling a large number of samples, a small fraction of each sample may be analysed by MALDI for screening purposes, while the greater fraction of each sample is analysed by ESI. The optimum choice of acid for the reversed phase chromatography eluent is different for ESI and MALDI ionization, respectively. In the case of nano LC-ESI, formic acid (FA) is the preferable acid (Tri Fluoro acetic acid, TFA, is frequently avoided due to reported problems with arcing at flow rates in the nl/min range). However, TFA is reported to provide higher chromatographic resolution than FA. In the case of MALDI, TFA is one of the preferred acids. Consequently, for the nano LC-MALDI interface, the optimum acid can be used regarding both chromatographic resolution and MS sensitivity. From theoretical considerations one could argue that the two ionization techniques are complementary by nature. While the relatively high mobility of low mass peptides make them easy to detect using ESI, these peptides are generally harder to detect using MALDI because the peptide mass is relatively close to the mass of the typical matrix molecules used. References 1.Johan Axelman, Henrik Wadensten, Staffan Renlund, Anders Tangen, Axel Parbel, Hans-Rudolf Höpker, “The use of 2DLC- MS/MS in proteome characterization – optimization of the LC step”, ASMS 2003, Montreal 2.Anders Tangen, Henrik Wadensten, Axel Parbel, Matthias Berg and Hans-Rudolf Höpker, “The LC-MS Workflow: A Comparative Study of High Throughput Nano LC and MDLC Designs”, ABRF 2004, Portland -0.60 -0.40 -0.20 0.00 0.20 0.40 0.60 0.80 -0.60 -0.50 -0.40 -0.30 -0.20 -0.10 0.00 0.10 0.20 0.30 0.40 wc[2] wc[1] AA Cover MH+ HPLC pI Cystein LC-ESI-T MALDI-TO MDLC-ESI-LTQ vs. MDLC-MALDI-LTQ: Improvement factor for sequence coverage using ESI-LTQ instead of MALDI-LTQ. Sample: CAF modified tryptic peptides from six proteins purified with salt step MDLC 0 2 4 6 8 10 BS A Transferrin Lysozym e AlcoholdehydrogenaseB-G alactosidase C ytochrom e C ESI/MALDISeq.Cov. MDLC Salt Step Configuration