TECH TIP #6
Pierce Biotechnology PO Box 117 (815) 968-0747 www.thermo.com/pierce
3747 N. Meridian Road Rockford, lL 61105 USA (815) 968-7316 fax
Introduction
In many applications involving peptides or proteins it is important either to identify fractions containing protein or to estimate
the concentration of a purified sample. Amino acids containing aromatic side chains (i.e., tyrosine, tryptophan and phenyl-
alanine) exhibit strong UV-light absorption. Consequently, proteins and peptides absorb UV-light in proportion to their
aromatic amino acid content and total concentration. Once an absorptivity coefficient has been established for a given protein
(with its fixed amino acid composition), the protein’s concentration in solution can be calculated from its absorbance.
For most proteins, UV-light absorption allows detection of concentration down to 100µg/mL. Nevertheless, estimation of
protein concentration by UV-light absorption is not accurate for complex protein solutions (e.g., cell lysates) because the
composition of proteins with different absorption coefficients is not known. In addition, proteins are not the only molecules
that absorb UV-light, and complex solutions will usually contain compounds like nucleic acids that interfere with protein
concentration determination by this method. However, for aqueous protein solutions commonly used in the research
laboratory setting, interference from other compounds is minimized by measuring absorbances at 280nm.
Only the amino acids tryptophan (Trp, W) and tyrosine (Tyr, Y) and to a lesser extent cysteine (Cys, C) contribute
significantly to peptide or protein absorbance at 280nm. Phenylalanine (Phe, F), which was mentioned above, absorbs only at
lower wavelengths (240-265nm).
Absorbance and Extinction Coefficients
The ratio of radiant power transmitted (P) by a sample to the radiant power incident (P0) on the sample is called the
transmittance, T:
T = P/P0
Absorbance (A), then, is defined as the logarithm (base 10) of the reciprocal of the transmittance:
A = -log T = log (1/T)
In a spectrophotometer, monochromatic plane-parallel light enters a sample at right angles to the plane-surface of the sample.
In these conditions, the transmittance and absorbance of a sample depends on the molar concentration (c), light path length in
centimeters (L), and molar absorptivity (ε) for the dissolved substance at the specified wavelength (λ).1
Tλ = 10εcL
or Aλ = ε c L
Beer’s Law states that molar absorptivity is constant (and the absorbance is proportional to concentration) for a given
substance dissolved in a given solute and measured at a given wavelength.2
For this reason, molar absorptivities are called
molar absorption coefficients or molar extinction coefficients. Because transmittance and absorbance are unitless, the units
for molar absorptivity must cancel with units of measure in concentration and light path. Therefore, molar absorptivities have
units of M-1
cm-1
. Standard laboratory spectrophotometers are fitted for use with 1cm-width sample cuvettes; hence, the path
length is generally assumed to be equal to one and the term is dropped altogether in most calculations.
Aλ = ε c L = ε c when L = 1cm
The molar absorption coefficient of a peptide or protein is related to its tryptophan (W), tyrosine (Y) and cysteine (C) amino
acid composition. At 280nm, this value is approximated by the weighted sum of the 280nm molar absorption coefficients of
these three constituent amino acids, as described in the following equation:3,4
)
125
C
(
)
1490
Y
(
)
5500
W
(
ε ×
+
×
+
×
= n
n
n
where n is the number of each residue and the stated values are the amino acid molar absorptivities at 280nm.
TR0006.4
Extinction Coefficients
A guide to understanding extinction coefficients, with emphasis on
spectrophotometric determination of protein concentration
Pierce Biotechnology PO Box 117 (815) 968-0747 www.thermo.com/pierce
3747 N. Meridian Road Rockford, lL 61105 USA (815) 968-7316 fax
2
Determining the Protein Concentration of a Solution from its Absorbance
Solving the expression of Beer’s law for concentration, one can easily see what values are needed to determine the
concentration of a peptide or protein solution:
c = A / ε L ( = A / ε when L = 1 cm)
Dividing the measured absorbance of a peptide or protein solution by the calculated or known molar extinction coefficient
yields the molar concentration of the peptide or protein solution. The peptide or protein amino acid composition must be
known to calculate the molar extinction coefficient using the formula stated in the previous section.
A brief reading of the cited articles3,4
makes one appreciate that there is no single correct extinction coefficient value for a
complex molecule like a peptide or protein. Even minor differences in buffer type, ionic strength and pH affects absorptivity
values at least slightly. Most protein preparations, even those of equal purity, differ slightly in conformation and extent of
modifications, such as oxidation, and these also affect absorptivity. Therefore, the best extinction coefficient value is one that
is determined empirically using a solution of the study protein of known concentration dissolved in the same buffer as the
sample (e.g., see discussion of Pierce Albumin Standards to follow).
Alternatively, absorption coefficients (i.e., extinction coefficients) for many proteins have been compiled from the literature
and reported in the Practical Handbook of Biochemistry and Molecular Biology.5
These values provide sufficient accuracy
for most routine laboratory applications that require an assessment of protein concentration. Most sources report extinction
coefficients for proteins measured at or near a wavelength of 280nm in phosphate or other physiologic buffer.
Molar Extinction Coefficients vs. Absorbances for 1% Solutions
Application of a molar extinction coefficient in the calculation yields an expression of concentration in terms of molarity:
A / εmolar = molar concentration
However, many sources, including the reference cited above, do not provide molar extinction coefficients. Instead, they
provide absorbance (A280nm) values for 1% (= 1g/100mL) solutions measured in a 1 cm cuvette. These values can be
understood as percent solution extinction coefficients (εpercent) having units of (g/100mL)-1
cm-1
instead of M-1
cm-1
.
Consequently, when these values are applied as extinction coefficients in the general formula, the units for concentration, c,
are percent solution (i.e., 1% = 1g/100mL = 10mg/mL).
A / εpercent = percent concentration
If one wishes to report concentration in terms of mg/ml, then an adjustment factor of 10 must be made when using these
percent solution extinction coefficients (i.e., one must convert from 10 mg/ml units to 1 mg/ml concentration units).
(A / εpercent) 10 = concentration in mg/ml
The relationship between molar extinction coefficient (εmolar) and percent extinction coefficient (εpercent) is as follows:
(εmolar) 10 = (εpercent) × (molecular weight of protein)
Still other sources provide protein absorbance values for 0.1% (= mg/mL) solutions, as this unit of measure is more
convenient and common for protein work than percent solution. This variation in reporting style underscores the importance
of carefully reading stated values to be sure that the unit of measure is understood and applied correctly.
Examples
A. Proteins and Protein Mixtures with Unknown Extinction Coefficients
If no extinction coefficient information exists for a protein or protein mixture of interest, and a rough estimate of protein
concentration is required for a solution that has no other interfering substances, assume εpercent = 10. Most protein extinction
coefficients (εpercent) range from 4.0 to 24.0.5
Therefore, although any given protein can vary significantly from εpercent = 10,
the average for a mixture of many different proteins likely will be approximately 10.
B. Immunoglobulins
Most mammalian antibodies (i.e., immunoglobulins) have protein extinction coefficients (εpercent) in the range of 12 to 15.
Therefore, for typical antibody solutions, assume 14
A %
1
nm
280 = or 4
.
1
A
A mg/ml
1
nm
280
%
1
.
0
nm
280 =
= .
For a typical IgG with MW = 150,000, this value corresponds to a molar extinction coefficient (ε) equal to 210,000M-1
cm-1
.
Pierce Biotechnology PO Box 117 (815) 968-0747 www.thermo.com/pierce
3747 N. Meridian Road Rockford, lL 61105 USA (815) 968-7316 fax
3
C. Bovine Serum Albumin (BSA)
Thermo Scientific™ Pierce™ Albumin Standard Ampules (Product No. 23209) are provided as 2mg/mL solutions of purified
bovine serum albumin (BSA) in 0.9% NaCl. The product is calibrated by direct comparison of the absorbance at 280nm to a
known concentration of a BSA standard from the National Institute of Standards and Technology (NIST). Numerous values
for the absorptivity of BSA have been reported in the literature but are generally ~6.6 for a 1% solution at 280nm.
Therefore, the predicted absorbance at 280nm for the Albumin Standard, assuming exactly 2mg/mL and εpercent = 6.6 is
εpercent c L / 10 = A
[(6.6)(2.000)(1)] / 10 = 1.320
Suppose that relative to a water reference a researcher obtains a 280nm absorbance reading of 1.346 for the Albumin
Standard. The calculated concentration, assuming the stated percent absorptivity value, is as follows:
(A / εpercent) × 10 = cmg/ml
(1.346 / 6.6) × 10 = 2.039mg/mL
Assuming a MW = 66,400, the molar extinction coefficient at 280nm for BSA is approximately 43,824M-1
cm-1
.
Using Thermo Scientific Pierce Albumin Standards
If you plan to use a Pierce Albumin Solution (Product No. 23209 or 23210) as an absorbance standard, assume its
concentration to be accurate (that’s the whole point of a standard!) and use it to calculate your own “system-specific”
extinction coefficient. Do this by measuring the absorbance of the provided solution (or several dilutions thereof, ideally
prepared in duplicate or triplicate) and then applying the formula A / c L = ε. As in the above examples, this ε that you
calculate will be in terms of the units you used for c. The resulting “system-specific” extinction coefficient will be accurate
for your particular buffer, spectrophotometer and cuvette, etc., allowing the albumin to function as an accurate reference
standard for protein samples of unknown concentration. Without also knowing the extinction coefficient for the proteins in
the sample, you will not know whether the same concentrations of BSA and sample protein will have the same absorbance.
(For example, 1mg/mL IgG has nearly twice the absorbance of 1mg/mL BSA.) However, you will be able to use the BSA
standard as a uniform reference to compare and normalize multiple samples to each other.
Related Thermo Scientific Pierce Protein Research Products
23209 Albumin Standard Ampules, 2mg/mL, 10 × 1mL ampules containing bovine serum albumin (BSA)
at a concentration of 2.0mg/mL in 0.9% saline and 0.05% sodium azide
23210 Albumin Standard Ampules, 2mg/mL, 50mL, containing bovine serum albumin (BSA) at a
concentration of 2.0mg/mL in 0.9% saline and 0.05% sodium azide
23212 Bovine Gamma Globulin Standard Ampules, 2mg/mL, 10 × 1mL
23225 BCA Protein Assay Kit, sufficient reagents for 500 test tube or 5000 microplate assays
23235 Micro BCA Protein Assay Kit, working range of 0.5-20µg/mL
23236 Coomassie Plus (Bradford) Assay Kit, working range of 1-1500µg/mL
References
1. Lange’s Handbook of Chemistry, 14th
Edition, Dean, J.A., Ed. (1992). McGraw-Hill, Inc., New York.
2. Handbook of Chemistry and Physics, 56th
Edition, Weast, R.C., Ed. (1975). CRC Press, Cleveland.
3. Gill, S.C. and von Hippel, P.H. (1989). Calculation of protein extinction coefficients from amino acid sequence data. Anal. Biochem. 182:319-26.
4. Pace, C.N., et al. (1995). How to measure and predict the molar absorption coefficient of a protein. Protein Sci. 4:2411-23.
5. Practical Handbook of Biochemistry and Molecular Biology, Fasman, D.G., Ed. (1992). CRC Press, Boston.
Current product instructions are available at www.thermoscientific.com/pierce. For a faxed copy, call 800-874-3723 or contact your local distributor.
© 2013 Thermo Fisher Scientific Inc. All rights reserved. Unless otherwise indicated, all trademarks are property of Thermo Fisher Scientific Inc. and its
subsidiaries. Printed in the USA.

More Related Content

Similar to TR0006-Extinction-coefficients.pdf

Pharmaceutical Assay and Multicomponent Analysis
Pharmaceutical Assay and Multicomponent AnalysisPharmaceutical Assay and Multicomponent Analysis
Pharmaceutical Assay and Multicomponent AnalysisPerkinElmer, Inc.
 
Challenges in measuring comparability of aggregation and conformation for bio...
Challenges in measuring comparability of aggregation and conformation for bio...Challenges in measuring comparability of aggregation and conformation for bio...
Challenges in measuring comparability of aggregation and conformation for bio...KBI Biopharma
 
Molar extinction coefficient
Molar extinction coefficientMolar extinction coefficient
Molar extinction coefficientDhruv Sharma
 
SIMONA CAVALU_Rotational Correlation Times of proteins
SIMONA CAVALU_Rotational Correlation Times of proteinsSIMONA CAVALU_Rotational Correlation Times of proteins
SIMONA CAVALU_Rotational Correlation Times of proteinsSimona Cavalu
 
UV-visible spectrophotometry ppt
UV-visible spectrophotometry ppt UV-visible spectrophotometry ppt
UV-visible spectrophotometry ppt rdeepthi1
 
Class 2 - Basic calculations and physicochemical.pdf
Class 2 - Basic calculations and physicochemical.pdfClass 2 - Basic calculations and physicochemical.pdf
Class 2 - Basic calculations and physicochemical.pdfBulchaTolcha
 
Application of UV-vis in natural product research
Application of UV-vis in natural product researchApplication of UV-vis in natural product research
Application of UV-vis in natural product researchZarTaj2
 
General techniques for identification and characterization of protein and Nuc...
General techniques for identification and characterization of protein and Nuc...General techniques for identification and characterization of protein and Nuc...
General techniques for identification and characterization of protein and Nuc...JohamSarfrazAli1
 
UV-VISIBLE SPECTROPHOTOMETRY AND INORGANIC PHOSPHATE DETERMINATION.pdf
UV-VISIBLE SPECTROPHOTOMETRY AND INORGANIC PHOSPHATE DETERMINATION.pdfUV-VISIBLE SPECTROPHOTOMETRY AND INORGANIC PHOSPHATE DETERMINATION.pdf
UV-VISIBLE SPECTROPHOTOMETRY AND INORGANIC PHOSPHATE DETERMINATION.pdfTatendaMageja
 
UV- Visible Spectroscopy
UV- Visible SpectroscopyUV- Visible Spectroscopy
UV- Visible SpectroscopyROHIT
 
Online Size-Exclusion High-Performance Liquid Chromatography Light Scattering...
Online Size-Exclusion High-Performance Liquid Chromatography Light Scattering...Online Size-Exclusion High-Performance Liquid Chromatography Light Scattering...
Online Size-Exclusion High-Performance Liquid Chromatography Light Scattering...KBI Biopharma
 
Analysis of bioreactor parameters online and offline
Analysis of bioreactor parameters online and offlineAnalysis of bioreactor parameters online and offline
Analysis of bioreactor parameters online and offlinevikash_94
 
Uv vis spectroscopy practical.
Uv vis spectroscopy practical.Uv vis spectroscopy practical.
Uv vis spectroscopy practical.Salum Mkata
 
Unknown Salt Essay
Unknown Salt EssayUnknown Salt Essay
Unknown Salt EssayLaura Martin
 
Factor affecting uv vis absorption lalit
Factor affecting uv vis absorption lalitFactor affecting uv vis absorption lalit
Factor affecting uv vis absorption lalitLalit Singh
 
absolute-molar-mass-uhplc
absolute-molar-mass-uhplcabsolute-molar-mass-uhplc
absolute-molar-mass-uhplcAym Berges
 

Similar to TR0006-Extinction-coefficients.pdf (20)

Pharmaceutical Assay and Multicomponent Analysis
Pharmaceutical Assay and Multicomponent AnalysisPharmaceutical Assay and Multicomponent Analysis
Pharmaceutical Assay and Multicomponent Analysis
 
Challenges in measuring comparability of aggregation and conformation for bio...
Challenges in measuring comparability of aggregation and conformation for bio...Challenges in measuring comparability of aggregation and conformation for bio...
Challenges in measuring comparability of aggregation and conformation for bio...
 
Molar extinction coefficient
Molar extinction coefficientMolar extinction coefficient
Molar extinction coefficient
 
SIMONA CAVALU_Rotational Correlation Times of proteins
SIMONA CAVALU_Rotational Correlation Times of proteinsSIMONA CAVALU_Rotational Correlation Times of proteins
SIMONA CAVALU_Rotational Correlation Times of proteins
 
C5OB00465A (1)
C5OB00465A (1)C5OB00465A (1)
C5OB00465A (1)
 
UV-visible spectrophotometry ppt
UV-visible spectrophotometry ppt UV-visible spectrophotometry ppt
UV-visible spectrophotometry ppt
 
Class 2 - Basic calculations and physicochemical.pdf
Class 2 - Basic calculations and physicochemical.pdfClass 2 - Basic calculations and physicochemical.pdf
Class 2 - Basic calculations and physicochemical.pdf
 
Application of UV-vis in natural product research
Application of UV-vis in natural product researchApplication of UV-vis in natural product research
Application of UV-vis in natural product research
 
B37
B37B37
B37
 
General techniques for identification and characterization of protein and Nuc...
General techniques for identification and characterization of protein and Nuc...General techniques for identification and characterization of protein and Nuc...
General techniques for identification and characterization of protein and Nuc...
 
UV-VISIBLE SPECTROPHOTOMETRY AND INORGANIC PHOSPHATE DETERMINATION.pdf
UV-VISIBLE SPECTROPHOTOMETRY AND INORGANIC PHOSPHATE DETERMINATION.pdfUV-VISIBLE SPECTROPHOTOMETRY AND INORGANIC PHOSPHATE DETERMINATION.pdf
UV-VISIBLE SPECTROPHOTOMETRY AND INORGANIC PHOSPHATE DETERMINATION.pdf
 
UV- Visible Spectroscopy
UV- Visible SpectroscopyUV- Visible Spectroscopy
UV- Visible Spectroscopy
 
Online Size-Exclusion High-Performance Liquid Chromatography Light Scattering...
Online Size-Exclusion High-Performance Liquid Chromatography Light Scattering...Online Size-Exclusion High-Performance Liquid Chromatography Light Scattering...
Online Size-Exclusion High-Performance Liquid Chromatography Light Scattering...
 
Analysis of bioreactor parameters online and offline
Analysis of bioreactor parameters online and offlineAnalysis of bioreactor parameters online and offline
Analysis of bioreactor parameters online and offline
 
Uv vis spectroscopy practical.
Uv vis spectroscopy practical.Uv vis spectroscopy practical.
Uv vis spectroscopy practical.
 
poster template 2015 4501
poster template 2015 4501poster template 2015 4501
poster template 2015 4501
 
Unknown Salt Essay
Unknown Salt EssayUnknown Salt Essay
Unknown Salt Essay
 
Unknown Salt Essay
Unknown Salt EssayUnknown Salt Essay
Unknown Salt Essay
 
Factor affecting uv vis absorption lalit
Factor affecting uv vis absorption lalitFactor affecting uv vis absorption lalit
Factor affecting uv vis absorption lalit
 
absolute-molar-mass-uhplc
absolute-molar-mass-uhplcabsolute-molar-mass-uhplc
absolute-molar-mass-uhplc
 

Recently uploaded

Vikas Nagar #Dating Call Girls Lucknow Get 50% Off On VIP Escorts Service 🍇 8...
Vikas Nagar #Dating Call Girls Lucknow Get 50% Off On VIP Escorts Service 🍇 8...Vikas Nagar #Dating Call Girls Lucknow Get 50% Off On VIP Escorts Service 🍇 8...
Vikas Nagar #Dating Call Girls Lucknow Get 50% Off On VIP Escorts Service 🍇 8...akbard9823
 
Low Rate Call Girls Nagpur Esha Call 7001035870 Meet With Nagpur Escorts
Low Rate Call Girls Nagpur Esha Call 7001035870 Meet With Nagpur EscortsLow Rate Call Girls Nagpur Esha Call 7001035870 Meet With Nagpur Escorts
Low Rate Call Girls Nagpur Esha Call 7001035870 Meet With Nagpur Escortsranjana rawat
 
Jp Nagar Call Girls Bangalore WhatsApp 8250192130 High Profile Service
Jp Nagar Call Girls Bangalore WhatsApp 8250192130 High Profile ServiceJp Nagar Call Girls Bangalore WhatsApp 8250192130 High Profile Service
Jp Nagar Call Girls Bangalore WhatsApp 8250192130 High Profile ServiceHigh Profile Call Girls
 
Call Girl Nashik Khushi 7001305949 Independent Escort Service Nashik
Call Girl Nashik Khushi 7001305949 Independent Escort Service NashikCall Girl Nashik Khushi 7001305949 Independent Escort Service Nashik
Call Girl Nashik Khushi 7001305949 Independent Escort Service Nashikranjana rawat
 
BPP NC II Lesson 3 - Pastry Products.pptx
BPP NC II Lesson 3 - Pastry Products.pptxBPP NC II Lesson 3 - Pastry Products.pptx
BPP NC II Lesson 3 - Pastry Products.pptxmaricel769799
 
Grade Eight Quarter 4_Week 6_Cookery.pptx
Grade Eight Quarter 4_Week 6_Cookery.pptxGrade Eight Quarter 4_Week 6_Cookery.pptx
Grade Eight Quarter 4_Week 6_Cookery.pptxKurtGardy
 
Dubai Call Girls Drilled O525547819 Call Girls Dubai (Raphie)
Dubai Call Girls Drilled O525547819 Call Girls Dubai (Raphie)Dubai Call Girls Drilled O525547819 Call Girls Dubai (Raphie)
Dubai Call Girls Drilled O525547819 Call Girls Dubai (Raphie)kojalkojal131
 
(PRIYANKA) Katraj Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune E...
(PRIYANKA) Katraj Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune E...(PRIYANKA) Katraj Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune E...
(PRIYANKA) Katraj Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune E...ranjana rawat
 
Pesticide Calculation Review 2013 post.pptx
Pesticide Calculation Review 2013 post.pptxPesticide Calculation Review 2013 post.pptx
Pesticide Calculation Review 2013 post.pptxalfordglenn
 
VIP Kolkata Call Girl Jadavpur 👉 8250192130 Available With Room
VIP Kolkata Call Girl Jadavpur 👉 8250192130  Available With RoomVIP Kolkata Call Girl Jadavpur 👉 8250192130  Available With Room
VIP Kolkata Call Girl Jadavpur 👉 8250192130 Available With Roomdivyansh0kumar0
 
(PRIYA) Call Girls Budhwar Peth ( 7001035870 ) HI-Fi Pune Escorts Service
(PRIYA) Call Girls Budhwar Peth ( 7001035870 ) HI-Fi Pune Escorts Service(PRIYA) Call Girls Budhwar Peth ( 7001035870 ) HI-Fi Pune Escorts Service
(PRIYA) Call Girls Budhwar Peth ( 7001035870 ) HI-Fi Pune Escorts Serviceranjana rawat
 
High Class Call Girls Nashik Priya 7001305949 Independent Escort Service Nashik
High Class Call Girls Nashik Priya 7001305949 Independent Escort Service NashikHigh Class Call Girls Nashik Priya 7001305949 Independent Escort Service Nashik
High Class Call Girls Nashik Priya 7001305949 Independent Escort Service Nashikranjana rawat
 
(ANJALI) Shikrapur Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune ...
(ANJALI) Shikrapur Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune ...(ANJALI) Shikrapur Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune ...
(ANJALI) Shikrapur Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune ...ranjana rawat
 
(ISHITA) Call Girls Manchar ( 7001035870 ) HI-Fi Pune Escorts Service
(ISHITA) Call Girls Manchar ( 7001035870 ) HI-Fi Pune Escorts Service(ISHITA) Call Girls Manchar ( 7001035870 ) HI-Fi Pune Escorts Service
(ISHITA) Call Girls Manchar ( 7001035870 ) HI-Fi Pune Escorts Serviceranjana rawat
 
VVIP Pune Call Girls Viman Nagar (7001035870) Pune Escorts Nearby with Comple...
VVIP Pune Call Girls Viman Nagar (7001035870) Pune Escorts Nearby with Comple...VVIP Pune Call Girls Viman Nagar (7001035870) Pune Escorts Nearby with Comple...
VVIP Pune Call Girls Viman Nagar (7001035870) Pune Escorts Nearby with Comple...Call Girls in Nagpur High Profile
 
Book Paid Chakan Call Girls Pune 8250192130Low Budget Full Independent High P...
Book Paid Chakan Call Girls Pune 8250192130Low Budget Full Independent High P...Book Paid Chakan Call Girls Pune 8250192130Low Budget Full Independent High P...
Book Paid Chakan Call Girls Pune 8250192130Low Budget Full Independent High P...ranjana rawat
 
(SUNAINA) Call Girls Alandi Road ( 7001035870 ) HI-Fi Pune Escorts Service
(SUNAINA) Call Girls Alandi Road ( 7001035870 ) HI-Fi Pune Escorts Service(SUNAINA) Call Girls Alandi Road ( 7001035870 ) HI-Fi Pune Escorts Service
(SUNAINA) Call Girls Alandi Road ( 7001035870 ) HI-Fi Pune Escorts Serviceranjana rawat
 
Top Rated Pune Call Girls JM road ⟟ 6297143586 ⟟ Call Me For Genuine Sex Ser...
Top Rated  Pune Call Girls JM road ⟟ 6297143586 ⟟ Call Me For Genuine Sex Ser...Top Rated  Pune Call Girls JM road ⟟ 6297143586 ⟟ Call Me For Genuine Sex Ser...
Top Rated Pune Call Girls JM road ⟟ 6297143586 ⟟ Call Me For Genuine Sex Ser...Call Girls in Nagpur High Profile
 
(KRITIKA) Balaji Nagar Call Girls Just Call 7001035870 [ Cash on Delivery ] P...
(KRITIKA) Balaji Nagar Call Girls Just Call 7001035870 [ Cash on Delivery ] P...(KRITIKA) Balaji Nagar Call Girls Just Call 7001035870 [ Cash on Delivery ] P...
(KRITIKA) Balaji Nagar Call Girls Just Call 7001035870 [ Cash on Delivery ] P...ranjana rawat
 

Recently uploaded (20)

Vikas Nagar #Dating Call Girls Lucknow Get 50% Off On VIP Escorts Service 🍇 8...
Vikas Nagar #Dating Call Girls Lucknow Get 50% Off On VIP Escorts Service 🍇 8...Vikas Nagar #Dating Call Girls Lucknow Get 50% Off On VIP Escorts Service 🍇 8...
Vikas Nagar #Dating Call Girls Lucknow Get 50% Off On VIP Escorts Service 🍇 8...
 
Low Rate Call Girls Nagpur Esha Call 7001035870 Meet With Nagpur Escorts
Low Rate Call Girls Nagpur Esha Call 7001035870 Meet With Nagpur EscortsLow Rate Call Girls Nagpur Esha Call 7001035870 Meet With Nagpur Escorts
Low Rate Call Girls Nagpur Esha Call 7001035870 Meet With Nagpur Escorts
 
Jp Nagar Call Girls Bangalore WhatsApp 8250192130 High Profile Service
Jp Nagar Call Girls Bangalore WhatsApp 8250192130 High Profile ServiceJp Nagar Call Girls Bangalore WhatsApp 8250192130 High Profile Service
Jp Nagar Call Girls Bangalore WhatsApp 8250192130 High Profile Service
 
Call Girl Nashik Khushi 7001305949 Independent Escort Service Nashik
Call Girl Nashik Khushi 7001305949 Independent Escort Service NashikCall Girl Nashik Khushi 7001305949 Independent Escort Service Nashik
Call Girl Nashik Khushi 7001305949 Independent Escort Service Nashik
 
BPP NC II Lesson 3 - Pastry Products.pptx
BPP NC II Lesson 3 - Pastry Products.pptxBPP NC II Lesson 3 - Pastry Products.pptx
BPP NC II Lesson 3 - Pastry Products.pptx
 
Grade Eight Quarter 4_Week 6_Cookery.pptx
Grade Eight Quarter 4_Week 6_Cookery.pptxGrade Eight Quarter 4_Week 6_Cookery.pptx
Grade Eight Quarter 4_Week 6_Cookery.pptx
 
Dubai Call Girls Drilled O525547819 Call Girls Dubai (Raphie)
Dubai Call Girls Drilled O525547819 Call Girls Dubai (Raphie)Dubai Call Girls Drilled O525547819 Call Girls Dubai (Raphie)
Dubai Call Girls Drilled O525547819 Call Girls Dubai (Raphie)
 
(PRIYANKA) Katraj Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune E...
(PRIYANKA) Katraj Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune E...(PRIYANKA) Katraj Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune E...
(PRIYANKA) Katraj Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune E...
 
Pesticide Calculation Review 2013 post.pptx
Pesticide Calculation Review 2013 post.pptxPesticide Calculation Review 2013 post.pptx
Pesticide Calculation Review 2013 post.pptx
 
VIP Kolkata Call Girl Jadavpur 👉 8250192130 Available With Room
VIP Kolkata Call Girl Jadavpur 👉 8250192130  Available With RoomVIP Kolkata Call Girl Jadavpur 👉 8250192130  Available With Room
VIP Kolkata Call Girl Jadavpur 👉 8250192130 Available With Room
 
(PRIYA) Call Girls Budhwar Peth ( 7001035870 ) HI-Fi Pune Escorts Service
(PRIYA) Call Girls Budhwar Peth ( 7001035870 ) HI-Fi Pune Escorts Service(PRIYA) Call Girls Budhwar Peth ( 7001035870 ) HI-Fi Pune Escorts Service
(PRIYA) Call Girls Budhwar Peth ( 7001035870 ) HI-Fi Pune Escorts Service
 
High Class Call Girls Nashik Priya 7001305949 Independent Escort Service Nashik
High Class Call Girls Nashik Priya 7001305949 Independent Escort Service NashikHigh Class Call Girls Nashik Priya 7001305949 Independent Escort Service Nashik
High Class Call Girls Nashik Priya 7001305949 Independent Escort Service Nashik
 
(ANJALI) Shikrapur Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune ...
(ANJALI) Shikrapur Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune ...(ANJALI) Shikrapur Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune ...
(ANJALI) Shikrapur Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune ...
 
(ISHITA) Call Girls Manchar ( 7001035870 ) HI-Fi Pune Escorts Service
(ISHITA) Call Girls Manchar ( 7001035870 ) HI-Fi Pune Escorts Service(ISHITA) Call Girls Manchar ( 7001035870 ) HI-Fi Pune Escorts Service
(ISHITA) Call Girls Manchar ( 7001035870 ) HI-Fi Pune Escorts Service
 
VVIP Pune Call Girls Viman Nagar (7001035870) Pune Escorts Nearby with Comple...
VVIP Pune Call Girls Viman Nagar (7001035870) Pune Escorts Nearby with Comple...VVIP Pune Call Girls Viman Nagar (7001035870) Pune Escorts Nearby with Comple...
VVIP Pune Call Girls Viman Nagar (7001035870) Pune Escorts Nearby with Comple...
 
Book Paid Chakan Call Girls Pune 8250192130Low Budget Full Independent High P...
Book Paid Chakan Call Girls Pune 8250192130Low Budget Full Independent High P...Book Paid Chakan Call Girls Pune 8250192130Low Budget Full Independent High P...
Book Paid Chakan Call Girls Pune 8250192130Low Budget Full Independent High P...
 
(SUNAINA) Call Girls Alandi Road ( 7001035870 ) HI-Fi Pune Escorts Service
(SUNAINA) Call Girls Alandi Road ( 7001035870 ) HI-Fi Pune Escorts Service(SUNAINA) Call Girls Alandi Road ( 7001035870 ) HI-Fi Pune Escorts Service
(SUNAINA) Call Girls Alandi Road ( 7001035870 ) HI-Fi Pune Escorts Service
 
Top Rated Pune Call Girls JM road ⟟ 6297143586 ⟟ Call Me For Genuine Sex Ser...
Top Rated  Pune Call Girls JM road ⟟ 6297143586 ⟟ Call Me For Genuine Sex Ser...Top Rated  Pune Call Girls JM road ⟟ 6297143586 ⟟ Call Me For Genuine Sex Ser...
Top Rated Pune Call Girls JM road ⟟ 6297143586 ⟟ Call Me For Genuine Sex Ser...
 
(KRITIKA) Balaji Nagar Call Girls Just Call 7001035870 [ Cash on Delivery ] P...
(KRITIKA) Balaji Nagar Call Girls Just Call 7001035870 [ Cash on Delivery ] P...(KRITIKA) Balaji Nagar Call Girls Just Call 7001035870 [ Cash on Delivery ] P...
(KRITIKA) Balaji Nagar Call Girls Just Call 7001035870 [ Cash on Delivery ] P...
 
5th Annual International OMGD Grand Chapitre: A Spanish Wine Journey
5th Annual International OMGD Grand Chapitre: A Spanish Wine Journey5th Annual International OMGD Grand Chapitre: A Spanish Wine Journey
5th Annual International OMGD Grand Chapitre: A Spanish Wine Journey
 

TR0006-Extinction-coefficients.pdf

  • 1. TECH TIP #6 Pierce Biotechnology PO Box 117 (815) 968-0747 www.thermo.com/pierce 3747 N. Meridian Road Rockford, lL 61105 USA (815) 968-7316 fax Introduction In many applications involving peptides or proteins it is important either to identify fractions containing protein or to estimate the concentration of a purified sample. Amino acids containing aromatic side chains (i.e., tyrosine, tryptophan and phenyl- alanine) exhibit strong UV-light absorption. Consequently, proteins and peptides absorb UV-light in proportion to their aromatic amino acid content and total concentration. Once an absorptivity coefficient has been established for a given protein (with its fixed amino acid composition), the protein’s concentration in solution can be calculated from its absorbance. For most proteins, UV-light absorption allows detection of concentration down to 100µg/mL. Nevertheless, estimation of protein concentration by UV-light absorption is not accurate for complex protein solutions (e.g., cell lysates) because the composition of proteins with different absorption coefficients is not known. In addition, proteins are not the only molecules that absorb UV-light, and complex solutions will usually contain compounds like nucleic acids that interfere with protein concentration determination by this method. However, for aqueous protein solutions commonly used in the research laboratory setting, interference from other compounds is minimized by measuring absorbances at 280nm. Only the amino acids tryptophan (Trp, W) and tyrosine (Tyr, Y) and to a lesser extent cysteine (Cys, C) contribute significantly to peptide or protein absorbance at 280nm. Phenylalanine (Phe, F), which was mentioned above, absorbs only at lower wavelengths (240-265nm). Absorbance and Extinction Coefficients The ratio of radiant power transmitted (P) by a sample to the radiant power incident (P0) on the sample is called the transmittance, T: T = P/P0 Absorbance (A), then, is defined as the logarithm (base 10) of the reciprocal of the transmittance: A = -log T = log (1/T) In a spectrophotometer, monochromatic plane-parallel light enters a sample at right angles to the plane-surface of the sample. In these conditions, the transmittance and absorbance of a sample depends on the molar concentration (c), light path length in centimeters (L), and molar absorptivity (ε) for the dissolved substance at the specified wavelength (λ).1 Tλ = 10εcL or Aλ = ε c L Beer’s Law states that molar absorptivity is constant (and the absorbance is proportional to concentration) for a given substance dissolved in a given solute and measured at a given wavelength.2 For this reason, molar absorptivities are called molar absorption coefficients or molar extinction coefficients. Because transmittance and absorbance are unitless, the units for molar absorptivity must cancel with units of measure in concentration and light path. Therefore, molar absorptivities have units of M-1 cm-1 . Standard laboratory spectrophotometers are fitted for use with 1cm-width sample cuvettes; hence, the path length is generally assumed to be equal to one and the term is dropped altogether in most calculations. Aλ = ε c L = ε c when L = 1cm The molar absorption coefficient of a peptide or protein is related to its tryptophan (W), tyrosine (Y) and cysteine (C) amino acid composition. At 280nm, this value is approximated by the weighted sum of the 280nm molar absorption coefficients of these three constituent amino acids, as described in the following equation:3,4 ) 125 C ( ) 1490 Y ( ) 5500 W ( ε × + × + × = n n n where n is the number of each residue and the stated values are the amino acid molar absorptivities at 280nm. TR0006.4 Extinction Coefficients A guide to understanding extinction coefficients, with emphasis on spectrophotometric determination of protein concentration
  • 2. Pierce Biotechnology PO Box 117 (815) 968-0747 www.thermo.com/pierce 3747 N. Meridian Road Rockford, lL 61105 USA (815) 968-7316 fax 2 Determining the Protein Concentration of a Solution from its Absorbance Solving the expression of Beer’s law for concentration, one can easily see what values are needed to determine the concentration of a peptide or protein solution: c = A / ε L ( = A / ε when L = 1 cm) Dividing the measured absorbance of a peptide or protein solution by the calculated or known molar extinction coefficient yields the molar concentration of the peptide or protein solution. The peptide or protein amino acid composition must be known to calculate the molar extinction coefficient using the formula stated in the previous section. A brief reading of the cited articles3,4 makes one appreciate that there is no single correct extinction coefficient value for a complex molecule like a peptide or protein. Even minor differences in buffer type, ionic strength and pH affects absorptivity values at least slightly. Most protein preparations, even those of equal purity, differ slightly in conformation and extent of modifications, such as oxidation, and these also affect absorptivity. Therefore, the best extinction coefficient value is one that is determined empirically using a solution of the study protein of known concentration dissolved in the same buffer as the sample (e.g., see discussion of Pierce Albumin Standards to follow). Alternatively, absorption coefficients (i.e., extinction coefficients) for many proteins have been compiled from the literature and reported in the Practical Handbook of Biochemistry and Molecular Biology.5 These values provide sufficient accuracy for most routine laboratory applications that require an assessment of protein concentration. Most sources report extinction coefficients for proteins measured at or near a wavelength of 280nm in phosphate or other physiologic buffer. Molar Extinction Coefficients vs. Absorbances for 1% Solutions Application of a molar extinction coefficient in the calculation yields an expression of concentration in terms of molarity: A / εmolar = molar concentration However, many sources, including the reference cited above, do not provide molar extinction coefficients. Instead, they provide absorbance (A280nm) values for 1% (= 1g/100mL) solutions measured in a 1 cm cuvette. These values can be understood as percent solution extinction coefficients (εpercent) having units of (g/100mL)-1 cm-1 instead of M-1 cm-1 . Consequently, when these values are applied as extinction coefficients in the general formula, the units for concentration, c, are percent solution (i.e., 1% = 1g/100mL = 10mg/mL). A / εpercent = percent concentration If one wishes to report concentration in terms of mg/ml, then an adjustment factor of 10 must be made when using these percent solution extinction coefficients (i.e., one must convert from 10 mg/ml units to 1 mg/ml concentration units). (A / εpercent) 10 = concentration in mg/ml The relationship between molar extinction coefficient (εmolar) and percent extinction coefficient (εpercent) is as follows: (εmolar) 10 = (εpercent) × (molecular weight of protein) Still other sources provide protein absorbance values for 0.1% (= mg/mL) solutions, as this unit of measure is more convenient and common for protein work than percent solution. This variation in reporting style underscores the importance of carefully reading stated values to be sure that the unit of measure is understood and applied correctly. Examples A. Proteins and Protein Mixtures with Unknown Extinction Coefficients If no extinction coefficient information exists for a protein or protein mixture of interest, and a rough estimate of protein concentration is required for a solution that has no other interfering substances, assume εpercent = 10. Most protein extinction coefficients (εpercent) range from 4.0 to 24.0.5 Therefore, although any given protein can vary significantly from εpercent = 10, the average for a mixture of many different proteins likely will be approximately 10. B. Immunoglobulins Most mammalian antibodies (i.e., immunoglobulins) have protein extinction coefficients (εpercent) in the range of 12 to 15. Therefore, for typical antibody solutions, assume 14 A % 1 nm 280 = or 4 . 1 A A mg/ml 1 nm 280 % 1 . 0 nm 280 = = . For a typical IgG with MW = 150,000, this value corresponds to a molar extinction coefficient (ε) equal to 210,000M-1 cm-1 .
  • 3. Pierce Biotechnology PO Box 117 (815) 968-0747 www.thermo.com/pierce 3747 N. Meridian Road Rockford, lL 61105 USA (815) 968-7316 fax 3 C. Bovine Serum Albumin (BSA) Thermo Scientific™ Pierce™ Albumin Standard Ampules (Product No. 23209) are provided as 2mg/mL solutions of purified bovine serum albumin (BSA) in 0.9% NaCl. The product is calibrated by direct comparison of the absorbance at 280nm to a known concentration of a BSA standard from the National Institute of Standards and Technology (NIST). Numerous values for the absorptivity of BSA have been reported in the literature but are generally ~6.6 for a 1% solution at 280nm. Therefore, the predicted absorbance at 280nm for the Albumin Standard, assuming exactly 2mg/mL and εpercent = 6.6 is εpercent c L / 10 = A [(6.6)(2.000)(1)] / 10 = 1.320 Suppose that relative to a water reference a researcher obtains a 280nm absorbance reading of 1.346 for the Albumin Standard. The calculated concentration, assuming the stated percent absorptivity value, is as follows: (A / εpercent) × 10 = cmg/ml (1.346 / 6.6) × 10 = 2.039mg/mL Assuming a MW = 66,400, the molar extinction coefficient at 280nm for BSA is approximately 43,824M-1 cm-1 . Using Thermo Scientific Pierce Albumin Standards If you plan to use a Pierce Albumin Solution (Product No. 23209 or 23210) as an absorbance standard, assume its concentration to be accurate (that’s the whole point of a standard!) and use it to calculate your own “system-specific” extinction coefficient. Do this by measuring the absorbance of the provided solution (or several dilutions thereof, ideally prepared in duplicate or triplicate) and then applying the formula A / c L = ε. As in the above examples, this ε that you calculate will be in terms of the units you used for c. The resulting “system-specific” extinction coefficient will be accurate for your particular buffer, spectrophotometer and cuvette, etc., allowing the albumin to function as an accurate reference standard for protein samples of unknown concentration. Without also knowing the extinction coefficient for the proteins in the sample, you will not know whether the same concentrations of BSA and sample protein will have the same absorbance. (For example, 1mg/mL IgG has nearly twice the absorbance of 1mg/mL BSA.) However, you will be able to use the BSA standard as a uniform reference to compare and normalize multiple samples to each other. Related Thermo Scientific Pierce Protein Research Products 23209 Albumin Standard Ampules, 2mg/mL, 10 × 1mL ampules containing bovine serum albumin (BSA) at a concentration of 2.0mg/mL in 0.9% saline and 0.05% sodium azide 23210 Albumin Standard Ampules, 2mg/mL, 50mL, containing bovine serum albumin (BSA) at a concentration of 2.0mg/mL in 0.9% saline and 0.05% sodium azide 23212 Bovine Gamma Globulin Standard Ampules, 2mg/mL, 10 × 1mL 23225 BCA Protein Assay Kit, sufficient reagents for 500 test tube or 5000 microplate assays 23235 Micro BCA Protein Assay Kit, working range of 0.5-20µg/mL 23236 Coomassie Plus (Bradford) Assay Kit, working range of 1-1500µg/mL References 1. Lange’s Handbook of Chemistry, 14th Edition, Dean, J.A., Ed. (1992). McGraw-Hill, Inc., New York. 2. Handbook of Chemistry and Physics, 56th Edition, Weast, R.C., Ed. (1975). CRC Press, Cleveland. 3. Gill, S.C. and von Hippel, P.H. (1989). Calculation of protein extinction coefficients from amino acid sequence data. Anal. Biochem. 182:319-26. 4. Pace, C.N., et al. (1995). How to measure and predict the molar absorption coefficient of a protein. Protein Sci. 4:2411-23. 5. Practical Handbook of Biochemistry and Molecular Biology, Fasman, D.G., Ed. (1992). CRC Press, Boston. Current product instructions are available at www.thermoscientific.com/pierce. For a faxed copy, call 800-874-3723 or contact your local distributor. © 2013 Thermo Fisher Scientific Inc. All rights reserved. Unless otherwise indicated, all trademarks are property of Thermo Fisher Scientific Inc. and its subsidiaries. Printed in the USA.