Quality in the
Coagulation Laboratory
8/28/2019
Donna D. Castellone, MS,MT(ASCP)SH
New York Presbyterian Hospital-Columbia
QA Manager, Specialty Testing
Supervisor, Special Coagulation & Hematology
2 Ā© 2019 Cardinal Health. All Rights Reserved.
Quality in the Coagulation Laboratory
Objectives
ā€¢ Identify solutions in areas of coagulation that can be enhanced by
implementing good laboratory practices and standard operating procedures
ā€¢ Analyze real case studies and how to troubleshoot testing to determine root
cause and what steps to prevent reoccurrence.
ā€¢ Enhance general knowledge of coagulation, standards and methods.
3 Ā© 2019 Cardinal Health. All Rights Reserved.
Quality in the Coagulation Laboratory
WHY do we need this?
ā€¢ Enhance diagnostic accuracy
ā€¢ Improve patient care and safety
ā€¢ Commutability of healthcare information
ā€¢ Coagulation is comprised of enzymes and unstable proteins ā€“ a unique
challenge
4 Ā© 2019 Cardinal Health. All Rights Reserved.
What do you want from your Coagulation assays?
How do you do this?
Coagulation Assays
ā€¢ Precise results
ā€¢ Robust assays
ā€¢ Accurate results that can be used as
an aid in diagnosis and treatment
ā€¢ Reproducible results
How do you do this?
ā€¢ Good SOPā€™s
ā€¢ Effective QC practices
ā€¢ Enroll in proficiency testing programs
ā€¢ Good Laboratory Practices
ā€¢ Understand coagulation results and
their impact on patient care
5 Ā© 2019 Cardinal Health. All Rights Reserved.
Healthcare quality is
getting the right care to
the right patient at the
right time ā€“ every time!
6 Ā© 2019 Cardinal Health. All Rights Reserved.
Quality in the Coagulation Laboratory
Quality Management System
ā€¢ Quality Management Plan - There must be a document that describes the
overall QM program
ā€¢ It can be a broad-range plan can cover all of the CLIA Quality System
requirements
ā€¢ Should contain monitors of key indicators of quality
ā€¢ These indicators should be during the pre-analytical, analytical and post
analytical stage
7 Ā© 2019 Cardinal Health. All Rights Reserved.
What Key Indicators can be Monitored in
Coagulation?
THRESHOLDS SHOULD BE MEANINGFUL AND ACHIEVABLE:
Pre-analytical: ~64% errors in this phase
ā€¢ Short draw samples, hemolysis, time to be processed
ā€¢ Who are your repeat offender-ed, pediatrics
Analytical: Test Order Accuracy
ā€¢ Why are certain test being ordered? Excess amount of Factor X testing being performed. Investigate, contact
physicians
ā€¢ Rejection logs; data mining- found wrong tubes being processed for testing, platelet testing without
scheduling, testing getting cancelled
ā€¢ Turnaround Time: poor turn around time, may be due to add on testing; how do you set your benchmarks-
look to guidelines or clinician expectations - in particular for factor assays
Post-analytical:
ā€¢ Reports ā€“ flagging values
8 Ā© 2019 Cardinal Health. All Rights Reserved.
Quality Assessment Formal Event Recording
45%
33%
22%
Pre-analytical
Analytical
Post-analytical
JANUARY ā€“ JUNE 2019
9 Ā© 2019 Cardinal Health. All Rights Reserved.
Quality Assessment Recorded Issues
0% 5% 10% 15% 20% 25%
Delayed results
Collection issue
result not called
incorrect results
Sample not received
Tests not performed
10 Ā© 2019 Cardinal Health. All Rights Reserved.
TAT for Coagulation Scoreboard
Factor VIII Jan Feb March April May
< 24 hours 90% 96% 98% 85% 88%
> 24 hours 10% 4% 2% 15% 12%
TAT 0.5 0.4 0.4 0.9 1.1
% compliance 100% 100% 100% 100% 100%
BENCHMARK: 2 DAYS
11 Ā© 2019 Cardinal Health. All Rights Reserved.
Evaluate Errors
Root Cause Analysis
ā€¢ Want to identify and evaluate errors, incidents and other problems that may
interfere with patient care services. Can no longer say, ā€œI canā€™t control thatā€
ā€¢ Need a mechanism to capture internal and external sources such as
complaints, including mistakes and near misses
ā€¢ Conduct root cause analysis of occurrences, issues, errors and gaps in quality
ā€¢ Outcome is to demonstrate risk reduction activities based on such root cause
analyses ā€“ do not band-aid situations
12 Ā© 2019 Cardinal Health. All Rights Reserved.
Plan. Do. Check. Act
HTTP://ASQ.ORG/LEARN-ABOUT-QUALITY/PROJECT-PLANNING-TOOLS/OVERVIEW/PDCA-CYCLE.HTML
ā€¢ PDCA model of continuous improvement
ā€¢ Deming cycle or wheel of improvement
ā€¢ When starting a new improvement project
ā€¢ When developing a new or improved design of a process, product or service.
ā€¢ When defining a repetitive work process
ā€¢ When planning data collection and analysis in order to verify and prioritize
problems or root causes
ā€¢ When implementing any change
13 Ā© 2019 Cardinal Health. All Rights Reserved.
Continuous Improvement
ā€¢ PLAN: Specimen rejection log for Special coagulation P2Y12 and Aspirin. 40%
of the samples rejected are due to improper tube being collected.
ā€¢ DO: Track where the frequent offenders are, and follow up for re-training,
posted reminders of what tubes to use.
ā€¢ CHECK: Monitor the amount of samples that continue to be collected
improperly.
ā€¢ ACT: After 1 month, 50% reduction in the number of rejected tubes, of those,
28% were due to improper tube collection. Continue to evaluate where the
issues are coming from, and retrain. Most amount occurred in 1 day and were
from phlebotomy, they were alerted and retrained.
14 Ā© 2019 Cardinal Health. All Rights Reserved.
QI-11 Continuous Improvement
ā€¢ PLAN: Implement less order entry errors in particular for factor X
ā€¢ DO: monitor the number of factor X assays and patients also on heparin
ā€¢ CHECK: appeared most of the factor X were in reality anti-xa assays. Caused
patient treatment delays, possible under or over anticoagulated
ā€¢ ACT: Our LIS has a pop-up window when ordering this test: Factor X
assay looks at the activity levels of factor X and does not monitor heparin
levels
15 Ā© 2019 Cardinal Health. All Rights Reserved.
Internal Audit ā€“ Tracer Method
LOOK AT THE ENTIRE SYSTEM FOR A SPECIFIC ANALYTE FROM START TO FINISH
Must have a purposeful document that follows the pre-analytic, analytic, and
post-analytical process
Any deficiencies must have a corrective action and date by which this should be
completed. For more critical findings, TAT for correction should be ASAP ā€“
Looks at:
ā€¢ Are we adhering to our own quality system?
ā€¢ Is this process effective?
ā€¢ Is the system as a whole effective?
ā€¢ Look for opportunities to improve
16 Ā© 2019 Cardinal Health. All Rights Reserved.
Findings of a Tracer Audit: Bethesda VIII ā€“
Questionable Patient Results
DISCOVERY:
Pre-analytical:
Inhibitors were being performed on all requests, high FVIIIā€™s
Analytical:
1. Using the same material for calibration as control
2. Patient dilutions were not clear
3. Protocol did not follow process
4. Analyzer settings did not allow proper identification of test being ordered
Post-analytical:
Test is linear between 25-75%, results were being extrapolated; needed to change our dilutions
results within the curve
17 Ā© 2019 Cardinal Health. All Rights Reserved.
Corrective Actions:
METHODICALLY GOING THROUGH THE ENTIRE PROCESS UNCOVERED SEVERAL
PRACTICES THAT NEED TO BE CLARIFIED
ā€¢ Flow chart posted to understand inhibitor process
ā€¢ Protocol re-written ā€“ everyone re-trained
ā€¢ Implemented different material for controls and calibrators
ā€¢ Set up clear testing parameters on the analyzer
ā€¢ Updated dilutions and method of calculations
18 Ā© 2019 Cardinal Health. All Rights Reserved.
Inspection Readiness
ļ± Reduce number of inspection
deficiencies
ļ± Need a process to keep up with
changing requirements and
regulations
ļ± Engage staff in the quality
process
19 Ā© 2019 Cardinal Health. All Rights Reserved.
Top 10 CAP Deficiencies
2016 COLLEGE OF AMERICAN PATHOLOGISTS
Directive Requirement # of deficiencies
GEN.55500 Competency 1979
COM.01200 Activity Menu 1810
COM. 10000 Procedures 1345
COM. 01700 PT evaluation 1178
COM.10100 Procedure review 1137
GEN. 20375 Document control 1036
COM.30300 Reagent Labeling 1032
COM. 01400 PT Attestation 968
COM. 04200 Monthly Review 919
COM. 30450 New Lot Confirmation 897
20 Ā© 2019 Cardinal Health. All Rights Reserved.
Main Issues in Coagulation:
1. Platelet poor plasma
2. INR verification
3. Correct reporting/use of your D-dimer
4. Linearity studies, standard curves
5. Evaluation of educational challenges ā€“ must be purposeful
6. Heparin therapeutic range
7. Providing information to clinicians regarding impact of
anticoagulant
21 Ā© 2019 Cardinal Health. All Rights Reserved.
PT: FVIII Assay
Result: 151 : Mean =123 SD= 15.1 SDI = 1.8 Limits: 98-148
ā€¢ No PT material left
ā€¢ Checked curves, checked QC, brand new lot of reagents- not a lot of historical QC
ā€¢ QC was in, but on the higher side
ā€¢ Calibrate assay daily- compared the curves- the highest point was most different
ā€¢ Re-ran some patients from that day
ā€¢ Slightly lower results, but not moving from normal to abnormal even on borderline
results (Reference range is 56-191)
ā€¢ Sometimes the result may be statistically significant, but NOT clinically significant-
e.g. investigate YES- impact on patient care ā€“ minimal
ā€¢ A report should be written up with all of your findings, possible reason and action.
Signed by the director or designee
22 Ā© 2019 Cardinal Health. All Rights Reserved.
Quality Control
Purpose: detect errors in laboratory that may lead to presence of clinical error
ā€¢ This error forces a change in diagnosis or treatment in a patient unnecessarily
ā€¢ Ensures consistency, accuracy and reliability of patient test results & reports
ā€¢ Levels should reflect the testing population
ā€¢ Ranges should be validated by running prior to being put into use
ā€¢ Standards, run as QC to determine recovery of assigned value
ā€¢ Donā€™t forget if you run low factor assays to run controls in that range
23 Ā© 2019 Cardinal Health. All Rights Reserved.
When to Recalibrate?
1 Level of Control is Out: Re-Run
ā€¢ Check calibration curve to see if a point is very different, or curve is very
different- may need to re-run either a point, or the curve
ā€¢ After recalibration- QC still out discard all reagents
ā€¢ Before calling service-
ā€¢ Try running PT material- peer evaluated material a lot of information
ā€¢ Look at independent controls
2 Levels of Control are out: Recalibrate
24 Ā© 2019 Cardinal Health. All Rights Reserved.
Action Protocol for Changes in Monthly Statistics
LABORATORY MUST HAVE AN ACTION PROTOCOL THAT DEFINES WHAT TO DO WHEN
STATISTICS FALL OUTSIDE OF TOLERANCE LIMITS
ā€¢ What are those limits- can use published CVā€™s, or manufacturerā€™s
recommendation
ā€¢ When do you change range? Procedure and actions must match.
ā€¢ State a level in your procedure that can prompt a change: e.g. a deviation in
monthly QC >10%
ā€¢ For factor assays, was their a change in the lot of deficient plasma
ā€¢ When you have reviewed all documentation and there are no outside sources
of error, and a shift has occurred- change the range
ā€¢ Document what and why it was done
25 Ā© 2019 Cardinal Health. All Rights Reserved.
Case Study: New Lot of Quality Control
Laboratory Regulations: run control 20x to demonstrate the ranges
ā€¢ Do I have to use these ranges?
ā€¢ How do I run my 20 points?
ā€¢ Over 20 days- or 5 times over 4 days
ā€¢ Can I use the manufacturers range?
26 Ā© 2019 Cardinal Health. All Rights Reserved.
Results for FX Normal Control:
MEAN 95% RANGE 85-105
FVIII Day 1 Day 2 Day 3 Day 4
95 93 95 96
94 94 96 95
95 95 95 95
94 93 96 96
94 94 95 96
MANUFACTURERS RANGE: 80 ā€“ 120 SEC
27 Ā© 2019 Cardinal Health. All Rights Reserved.
Case Study
Review of the 1st Month
ā€¢ Mean 102%
ā€¢ Out of 30 days- 10 days spent, re-running controls-
ā€¢ Wasting time and reagents to get controls within range
ā€¢ This was happening throughout all of the laboratories
ā€¢ If you review the 20 control runs, looks like a precision run
ā€¢ Not unique to coagulation- all disciplines
ā€¢ If we would enter these ranges, we would spend a lot of time and money and
effort in having techs re-run, re-calibrate and hold up the runs.
ā€¢ Running 20 controls was not working
28 Ā© 2019 Cardinal Health. All Rights Reserved.
Case Study
Solution
ā€¢ We still run the 20 points; however we use this as a verification of the
manufacturerā€™s ranges.
ā€¢ The first month, we enter the manufacturers ranges.
ā€¢ After a month of running in a real-world setting, we re-evaluate the ā€œtrueā€
ranges.
ā€¢ Much less time wasted on trying to hit a ā€œperfect targetā€ which did not really
reflect the testing environment- your analyzer, your reagents in your lab, with
different operators
29 Ā© 2019 Cardinal Health. All Rights Reserved.
Case Study: New Shipment for Lot of FVIII
Deficient Plasma
ā€¢ New lot of aPTT reagents had already been validated with this lot of FVIII
deficient plasma
ā€¢ However, a new shipment of factor VIII was being put in use
ā€¢ New shipment is evaluated:
FVIII Old lot New Lot % difference Acceptable
Abnormal control 42% 41% 1% y
Normal control 98% 100% 2% y
Abnormal patient 35% 22% 37% n
Normal patient 123% 97% 21% n
30 Ā© 2019 Cardinal Health. All Rights Reserved.
Can I use these Reagents?
ā€¢ Controls are within range- patient samples are not within range-
ā€¢ FVIII is heat labile, maybe the samples are bad?
ā€¢ Tested another normal and abnormal patient sample, still out of range
ā€¢ Check independent controls matrix mimics FFP
FVIII Control result
OLD
Control result
NEW
% difference Acceptable
Abnormal
control
34 21 38% N
Normal control 98 77 21% N
31 Ā© 2019 Cardinal Health. All Rights Reserved.
New Lot of Reagents, Minimal Historical Data
ā€¢ Recalibrated, no change
ā€¢ Could it be the reagents? All other aPTT based assays were working fine.
ā€¢ However, did reconstitute new reagents, still no change
ā€¢ Bad lot of deficient plasma- even though we have been using the lot
same as previously used-Possible shipment issue- got a new
lot/shipment-
ā€¢ All results were within acceptable levels
ā€¢ MUST test both patient and controls-
ā€¢ Helps to have independent controls on hand
ā€¢ Also can test PT material- can be very helpful in troubleshooting
32 Ā© 2019 Cardinal Health. All Rights Reserved.
Case Study: Understanding your Reagents
ā€¢ 16-year-old male for a hernia operation
ā€¢ Family has a positive history of bleeding
ā€¢ Results: PT=12.9 (10.2-13.5)
o APTT=33.0 (29.9-33.5)
I. Will this patient bleed?
II. Do we check for a factor deficiency?
III. What is the most important information when evaluating if a patient will
bleed?
33 Ā© 2019 Cardinal Health. All Rights Reserved.
The Coagulation Cascade
Intrinsic
Pathway
Extrinsic
Pathway
Common
Pathway
34 Ā© 2019 Cardinal Health. All Rights Reserved.
Results:
ā€¢ Look at APTT factors: VIII, IX and XI (nr 50-150%)
ā€¢ VIII = 102%
ā€¢ IX= 84%
ā€¢ XI=21%
ā€¢ Abnormal level of factor XI despite the normal APTT
ā€¢ Patient is deficient in factor XI
ā€¢ Shouldnā€™t the APTT have been abnormal
35 Ā© 2019 Cardinal Health. All Rights Reserved.
How are your Reagents?
ā€¢ Do your reagents truly reflect normal factor levels
ā€¢ Normal PT & APTT levels indicate patient have a minimum of ~ 30% of factor
levels present
ā€¢ If your reagent is insensitive to a factor, you may get a normal PT or APTT
ā€¢ The reagent may not be able to pick up a factor level below 30%
36 Ā© 2019 Cardinal Health. All Rights Reserved.
Test Reagent Sensitivity for Factors
This reagent does not reflect an abnormality until 12.5%
ā€¢ Dilute normal plasma with factor deficient plasma at different levels
ā€¢ Run either a PT or APTT on the sample
ā€¢ Compare the results to the upper limit of the normal range
ā€¢ You may get a normal PT or APTT with an abnormal % factor level
FACTOR XI SENSITIVITY: APTT= (29.5 ā€“ 33.5)
NORMAL + DEFICIENT = % ACTIVITY APTT
PLASMA PLASMA
500UL 0UL 100% 29.0
250UL 250UL 50% 31.0
125UL 375UL 25% 32.0
62.5UL 437.5 12.5% 36.0
37 Ā© 2019 Cardinal Health. All Rights Reserved.
Factor Sensitivity
ā€¢ It is important to understand how your reagents perform
ā€¢ Small investment for a lot of information
ā€¢ Should be performed even if you do not run factor assays to understand what
the approximate levels of factor activity will prolong screening tests
ā€¢ In this case the family history is positive:
o FXI deficiency occurs in up to 8% of Ashkenazi Jews
o Incidence is estimated at 1 in 100,000 in the general population
o FXI deficiency is inherited in an autosomal recessive pattern
o Men and women are affected equally
o The level of FXI doesnā€™t correlate to bleeding
38 Ā© 2019 Cardinal Health. All Rights Reserved.
Case Study:
PRE-OPERATIVE PATIENT APTT 49.1 (25 ā€“ 36 SECONDS)
ā€¢ No history of bleeding
ā€¢ No medications
ā€¢ What can prolong an APTT
ā€¢ Pre-analytical variables- short draw
ā€¢ Presence of heparin or DOACs- can perform a thrombin time for detection
ā€¢ What factors deficiencies may prolong the APTT?
ā€¢ Those in the Intrinsic pathway: VIII, IX, XI, XII
ā€¢ An APTT may also be prolonged due to an inhibitor
39 Ā© 2019 Cardinal Health. All Rights Reserved.
Prolonged Screening Test:
NEXT STEP IS TO PERFORM A MIXING STUDY
ļ‚§ Correction: factor deficiency ( patient may bleed, unless FXII)
ļ‚§ No correction: inhibitor ( more likely a risk for thrombosis, unless specific
factor inhibitor, then they can bleed)
I. Do you use FFP for the mixing study, can I use a lyophilized standard human
plasma?
II. Do I incubate? How long?
III. Do I incubate separately? Together? Both?
IV. Is this test cost effective? What information do I get from it?
40 Ā© 2019 Cardinal Health. All Rights Reserved.
Classical 1:1 Mixing Study
Step 1: Immediate Mix
Patient PNP
C
1:1 Mix
1. Test all.
2. Compare C to A & B.
Did C correct?
3. Incubate all.
A B
Step 2: Incubated Mix
1-2 hrs., 37C
Patient PNP
C
Incubated Tubes
A B D
Control
1. Test all.
2. Compare incubated C to D. Is it
prolonged? Yes, possible time
dependent inhibitor.
3. Interpret final results
Mix
41 Ā© 2019 Cardinal Health. All Rights Reserved.
Results
ā€¢ APTT = 42.1 seconds (25-36 seconds)
ā€¢ Pooled Normal Plasma =31.5 seconds
ā€¢ Mixing study 1:1 = 36.5
ā€¢ Is this a correction? What are the variables that can impact a mixing study?
1. PPP critical for proper id of AN INHIBITOR; platelets are phospholipids and
will falsely shorten your results, when reacting with reagents
2. Reagent Sensitivity: at what level of factor deficiency will prolong your aPTT
3. Normal Pool plasma: must be ppp, have adequate levels of factors
COMPRISED OF A MINIMUM OF NORMAL INDIVIDUALS, NOT NORMAL PATIENTS
42 Ā© 2019 Cardinal Health. All Rights Reserved.
Interpretation of a Mixing Study
NO RECOMMENDATIONS OR GUIDELINES FROM ISTH SCIENTIFIC & STANDARDIZATION COMMITTEE
CAP COAGULATION RESOURCE COMMITTEE
ā€¢ Based on PT or aPTT normal range
Within limits @ 2SD or 3SD
Within 5 seconds of the 2SD upper limit
ā€¢ Rosner Index
o 1:1 mix
o Index = CT of 1:1 mix ā€“ CT of NPP x 100
CT of patient <15 = FD >15 = Inhibitor
NPP tested with Mixing Study
o Within 5 seconds of NPP value
o 10% of NPP value
o Lack of correction @ >15% of NPP
43 Ā© 2019 Cardinal Health. All Rights Reserved.
What do we have? (In my Lab)
ā€¢ A repeat APTT that is still prolonged 42.1 seconds (25-36 seconds)
ā€¢ PNP =31.0 seconds, mixing study is 36.4 seconds 5.5 seconds from PNP
ā€¢ > 5 seconds from PNP
ā€¢ > normal range
ā€¢ Not a correction/partial correction
ā€¢ Proceed with inhibitor testing
44 Ā© 2019 Cardinal Health. All Rights Reserved.
How to Standardize this Test:
ā€¢ Adhere to consistent policies in your laboratory- define your criteria and stick to
it
ā€¢ Eliminate the possibility of heparin being on board, perform a thrombin time,
best test for residual heparin
ā€¢ Can report just numbers and let the clinicians interpret
ā€¢ Can report with an interpretation
ā€¢ Eliminate the test, many laboratories feel it is too confusing, and just proceed
with additional testing
45 Ā© 2019 Cardinal Health. All Rights Reserved.
Case Study
PATIENT PRESENTS WITH A DVT, PLACED ON UNFRACTIONATED HEPARIN
ā€¢ Given a bolus dose of heparin - monitored by the APTT
ā€¢ First test taken 4 hours post dose
ā€¢ APTT= 67.5 sec (25-35 sec)
ā€¢ Develops a PE
ā€¢ Was the patient properly anticoagulated?
46 Ā© 2019 Cardinal Health. All Rights Reserved.
Based on APTT Result
ā€¢ Prolonged aPTT
ā€¢ What is the sensitivity of the aPTT reagent to heparin
ā€¢ Not good enough to use 1.5-2.5 times the mean of the normal range
ā€¢ Performed an anti-Xa assay = 0.25U/ml anti-Xa (0.3-0.7u/Ml)
ā€¢ Patient is under anticoagulated
ā€¢ If results are inaccurate and anticoagulation is excessive there could be a
bleed
ā€¢ If results are inadequate patients could clot
47 Ā© 2019 Cardinal Health. All Rights Reserved.
APTT Monitoring - Disadvantages
ā€¢Numerous factors elevate APTT:
o Warfarin therapy
o Lupus anticoagulants
o Factor deficiency, liver disease
ā€¢APTT response to heparin exaggerated
ā€¢Potential to under-anticoagulate patients
ā€¢Elevated factor VIII and I, shorten APTT
ā€¢Potential to over-anticoagulate patients
ā€¢No standard to cross reference APTT among laboratories
ā€¢No dose response relationship with heparin and APTT
Winter, W and Harris, WE, (2008), Anti-coagulant Monitoring in the Core Lab, Advance for Laboratory Professionals, pg.54-61.
48 Ā© 2019 Cardinal Health. All Rights Reserved.
CAP Requirement
ā€¢ It is recommended that the first method be used initially to establish the
therapeutic range before starting patient testing with a new instrument or new
reagent, while the second method can be used for validation of the therapeutic
range with subsequent reagent lot changes
ā€¢ It is not best practice to use plasma samples spiked with heparin in vitro to
calculate the therapeutic range, as differences in heparin binding proteins in
vitro may lead to overestimation of the therapeutic range
49 Ā© 2019 Cardinal Health. All Rights Reserved.
Heparin Therapeutic Range
CAP recommends that each laboratory establishes its own heparin therapeutic
range for each new lot of aPTT reagent based on reagent: instrument
combination
ā€¢ Heparin Anti-Xa Method
o 50-60 plasma samples collected from patients treated with UFH
o Mix of samples should span the therapeutic range
o No more than two samples on the same patient
o The PT/INR should normal (INR <1.3)
o Assay aPTT, anti-factor Xa (UFH), and PT/INR
o Plot the aPTT on the y-axis and the anti-Xa units on the X-axis
o Draw a line through the middle of the points
o The UFH therapeutic range is equivalent to
ā€“ 0.3-0.7 units/mL Anti-Xa Heparin assay
50 Ā© 2019 Cardinal Health. All Rights Reserved.
Brill-Edwards Technique
THERAPEUTIC RANGE= 0.3-0.7U/ML 70-120 SECONDS
51 Ā© 2019 Cardinal Health. All Rights Reserved.
Alternative Method: Subsequent Years
Cumulative summation of reagent mean differences
ā€¢ Initial therapeutic range established using the anti-Xa method
ā€¢ Run aPTT on samples with both the old and new lot of reagent
ā€¢ Plot the old lot on the X-axis and the new lot on the Y-axis
ā€¢ Determine the sum, mean and difference of the results with each new lot
ā€¢ Record the difference in the means to compare with past studies
ā€¢ A change of <5 seconds between the differences in the means is acceptable
ā€¢ A change of >5 seconds between the differences in the means requires action
Year Mean
Old Lot
Mean
New Lot
Difference New-
Old
Cum Sum Action
2006 34.0 37.2 3.2 3.2
2007 37.2 33.0 -4.2 -1.2 Accept
2008 33.0 43.0 10.0 8.8 Reject
2008-B 33.0 36.0 3.0 1.8 Accept
52 Ā© 2019 Cardinal Health. All Rights Reserved.
The ā€œDoā€™s" and ā€œDonā€™tsā€ to Enhance Quality in the
Coagulation Laboratory
DO
ā€¢ DO ask questions: use the five whys
to perform a root cause analysis
ā€¢ DO Listen to the issues from your
front-line staff, have everyone work
on problem solving
DONā€™T
ā€¢ DONā€™T blame people: Focus on
systems, not individuals- 99% of
people want to do the right thing!
ā€¢ DONā€™T say this canā€™t be solved, get
someone who doesnā€™t know the
process, a new set of eyes, they may
see things you are missing
53 Ā© 2019 Cardinal Health. All Rights Reserved.
Thank you for your time!
ā€œQUALITY IS NOT A GIVEN BUT NEEDS TO
BE A PRACTICE; PATIENT CARE AND TEST
OUTCOMES DEPEND ON IT.ā€

More Related Content

Similar to Quality_in_the_Coag_Lab_082819.pdf

9 Quality Management System_EAT G H 2021.pptx
9 Quality Management System_EAT G H 2021.pptx9 Quality Management System_EAT G H 2021.pptx
9 Quality Management System_EAT G H 2021.pptxNagaraju94925
Ā 
CLO4 LEAN.pptx
CLO4 LEAN.pptxCLO4 LEAN.pptx
CLO4 LEAN.pptxSYEDRAZA56411
Ā 
Introduction types, Objectives, Management of audit, Responsibilities, Planni...
Introduction types, Objectives, Management of audit, Responsibilities, Planni...Introduction types, Objectives, Management of audit, Responsibilities, Planni...
Introduction types, Objectives, Management of audit, Responsibilities, Planni...Kunal10679
Ā 
Overview of Quality Control and its implementation in the laboratory.pptx
Overview of Quality Control and its implementation in the laboratory.pptxOverview of Quality Control and its implementation in the laboratory.pptx
Overview of Quality Control and its implementation in the laboratory.pptxDipesh Tamrakar
Ā 
Point of Care Testing for Enhancing Patient Centered Planned Care Delivery
Point of Care Testing for Enhancing Patient Centered Planned Care DeliveryPoint of Care Testing for Enhancing Patient Centered Planned Care Delivery
Point of Care Testing for Enhancing Patient Centered Planned Care DeliveryPAFP
Ā 
Clinical Audits and Process Improvement in Hospitals
Clinical Audits and Process Improvement in HospitalsClinical Audits and Process Improvement in Hospitals
Clinical Audits and Process Improvement in HospitalsLallu Joseph
Ā 
Module 11 Process Improvement.pptx
Module 11 Process Improvement.pptxModule 11 Process Improvement.pptx
Module 11 Process Improvement.pptxFatima Fasih
Ā 
Residents engaged clinical audits by Mihertab
Residents engaged clinical audits by MihertabResidents engaged clinical audits by Mihertab
Residents engaged clinical audits by MihertabMihertab Kedida
Ā 
Role of quality system and audits in pharmamaceutical
Role of quality system and audits in pharmamaceuticalRole of quality system and audits in pharmamaceutical
Role of quality system and audits in pharmamaceuticalganpat420
Ā 
Best Quality Practices for Biomedic Research: Rick Calabrese ASQ 2014 NE Conf...
Best Quality Practices for Biomedic Research: Rick Calabrese ASQ 2014 NE Conf...Best Quality Practices for Biomedic Research: Rick Calabrese ASQ 2014 NE Conf...
Best Quality Practices for Biomedic Research: Rick Calabrese ASQ 2014 NE Conf...akrumenaker
Ā 
08 Am09 Presentations Kliger
08 Am09 Presentations   Kliger08 Am09 Presentations   Kliger
08 Am09 Presentations KligerSimon Prince
Ā 
Designing an appropriate qc design procedure for your lab 5 mar15
Designing an appropriate qc design procedure for your lab 5 mar15Designing an appropriate qc design procedure for your lab 5 mar15
Designing an appropriate qc design procedure for your lab 5 mar15Randox
Ā 
Designing an appropriate QC procedure for your laboratory
Designing an appropriate QC procedure for your laboratoryDesigning an appropriate QC procedure for your laboratory
Designing an appropriate QC procedure for your laboratoryRandox
Ā 
Understanding and implementing quality management system in medical laboratories
Understanding and implementing quality management system in medical laboratoriesUnderstanding and implementing quality management system in medical laboratories
Understanding and implementing quality management system in medical laboratoriesPathKind Labs
Ā 
Dr Ayman Ewies - Clinical audit made easy
Dr Ayman Ewies - Clinical audit made easyDr Ayman Ewies - Clinical audit made easy
Dr Ayman Ewies - Clinical audit made easyAymanEwies
Ā 
Focus on FDA cGMPs inspections
Focus on FDA cGMPs inspections Focus on FDA cGMPs inspections
Focus on FDA cGMPs inspections JinalPrajapati13
Ā 
Ignace vallejo resume qa qc manager (2)
Ignace vallejo resume   qa qc manager (2)Ignace vallejo resume   qa qc manager (2)
Ignace vallejo resume qa qc manager (2)Ian Vallejo
Ā 
Quality Assurance Manager Anjali Malhotra Resume
Quality Assurance Manager Anjali Malhotra ResumeQuality Assurance Manager Anjali Malhotra Resume
Quality Assurance Manager Anjali Malhotra ResumeAnjali Malhotra
Ā 

Similar to Quality_in_the_Coag_Lab_082819.pdf (20)

9 Quality Management System_EAT G H 2021.pptx
9 Quality Management System_EAT G H 2021.pptx9 Quality Management System_EAT G H 2021.pptx
9 Quality Management System_EAT G H 2021.pptx
Ā 
CLO4 LEAN.pptx
CLO4 LEAN.pptxCLO4 LEAN.pptx
CLO4 LEAN.pptx
Ā 
Introduction types, Objectives, Management of audit, Responsibilities, Planni...
Introduction types, Objectives, Management of audit, Responsibilities, Planni...Introduction types, Objectives, Management of audit, Responsibilities, Planni...
Introduction types, Objectives, Management of audit, Responsibilities, Planni...
Ā 
Overview of Quality Control and its implementation in the laboratory.pptx
Overview of Quality Control and its implementation in the laboratory.pptxOverview of Quality Control and its implementation in the laboratory.pptx
Overview of Quality Control and its implementation in the laboratory.pptx
Ā 
Point of Care Testing for Enhancing Patient Centered Planned Care Delivery
Point of Care Testing for Enhancing Patient Centered Planned Care DeliveryPoint of Care Testing for Enhancing Patient Centered Planned Care Delivery
Point of Care Testing for Enhancing Patient Centered Planned Care Delivery
Ā 
Clinical Audits and Process Improvement in Hospitals
Clinical Audits and Process Improvement in HospitalsClinical Audits and Process Improvement in Hospitals
Clinical Audits and Process Improvement in Hospitals
Ā 
Module 11 Process Improvement.pptx
Module 11 Process Improvement.pptxModule 11 Process Improvement.pptx
Module 11 Process Improvement.pptx
Ā 
Residents engaged clinical audits by Mihertab
Residents engaged clinical audits by MihertabResidents engaged clinical audits by Mihertab
Residents engaged clinical audits by Mihertab
Ā 
Role of quality system and audits in pharmamaceutical
Role of quality system and audits in pharmamaceuticalRole of quality system and audits in pharmamaceutical
Role of quality system and audits in pharmamaceutical
Ā 
Best Quality Practices for Biomedic Research: Rick Calabrese ASQ 2014 NE Conf...
Best Quality Practices for Biomedic Research: Rick Calabrese ASQ 2014 NE Conf...Best Quality Practices for Biomedic Research: Rick Calabrese ASQ 2014 NE Conf...
Best Quality Practices for Biomedic Research: Rick Calabrese ASQ 2014 NE Conf...
Ā 
08 Am09 Presentations Kliger
08 Am09 Presentations   Kliger08 Am09 Presentations   Kliger
08 Am09 Presentations Kliger
Ā 
Designing an appropriate qc design procedure for your lab 5 mar15
Designing an appropriate qc design procedure for your lab 5 mar15Designing an appropriate qc design procedure for your lab 5 mar15
Designing an appropriate qc design procedure for your lab 5 mar15
Ā 
Designing an appropriate QC procedure for your laboratory
Designing an appropriate QC procedure for your laboratoryDesigning an appropriate QC procedure for your laboratory
Designing an appropriate QC procedure for your laboratory
Ā 
nursing audit
nursing auditnursing audit
nursing audit
Ā 
Understanding and implementing quality management system in medical laboratories
Understanding and implementing quality management system in medical laboratoriesUnderstanding and implementing quality management system in medical laboratories
Understanding and implementing quality management system in medical laboratories
Ā 
Medical Laboratory Quality Assurance
Medical Laboratory Quality AssuranceMedical Laboratory Quality Assurance
Medical Laboratory Quality Assurance
Ā 
Dr Ayman Ewies - Clinical audit made easy
Dr Ayman Ewies - Clinical audit made easyDr Ayman Ewies - Clinical audit made easy
Dr Ayman Ewies - Clinical audit made easy
Ā 
Focus on FDA cGMPs inspections
Focus on FDA cGMPs inspections Focus on FDA cGMPs inspections
Focus on FDA cGMPs inspections
Ā 
Ignace vallejo resume qa qc manager (2)
Ignace vallejo resume   qa qc manager (2)Ignace vallejo resume   qa qc manager (2)
Ignace vallejo resume qa qc manager (2)
Ā 
Quality Assurance Manager Anjali Malhotra Resume
Quality Assurance Manager Anjali Malhotra ResumeQuality Assurance Manager Anjali Malhotra Resume
Quality Assurance Manager Anjali Malhotra Resume
Ā 

Recently uploaded

College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...perfect solution
Ā 
Night 7k to 12k Chennai City Center Call Girls šŸ‘‰šŸ‘‰ 7427069034ā­ā­ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls šŸ‘‰šŸ‘‰ 7427069034ā­ā­ 100% Genuine E...Night 7k to 12k Chennai City Center Call Girls šŸ‘‰šŸ‘‰ 7427069034ā­ā­ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls šŸ‘‰šŸ‘‰ 7427069034ā­ā­ 100% Genuine E...hotbabesbook
Ā 
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
Ā 
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
Ā 
Call Girls Mumbai Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Mumbai Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Mumbai Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Mumbai Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
Ā 
šŸ’ŽVVIP Kolkata Call Girls ParganasšŸ©±7001035870šŸ©±Independent Girl ( Ac Rooms Avai...
šŸ’ŽVVIP Kolkata Call Girls ParganasšŸ©±7001035870šŸ©±Independent Girl ( Ac Rooms Avai...šŸ’ŽVVIP Kolkata Call Girls ParganasšŸ©±7001035870šŸ©±Independent Girl ( Ac Rooms Avai...
šŸ’ŽVVIP Kolkata Call Girls ParganasšŸ©±7001035870šŸ©±Independent Girl ( Ac Rooms Avai...Taniya Sharma
Ā 
The Most Attractive Hyderabad Call Girls Kothapet š– ‹ 6297143586 š– ‹ Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet š– ‹ 6297143586 š– ‹ Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet š– ‹ 6297143586 š– ‹ Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet š– ‹ 6297143586 š– ‹ Will You Mis...chandars293
Ā 
All Time Service Available Call Girls Marine Drive šŸ“³ 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive šŸ“³ 9820252231 For 18+ VIP C...All Time Service Available Call Girls Marine Drive šŸ“³ 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive šŸ“³ 9820252231 For 18+ VIP C...Arohi Goyal
Ā 
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeTop Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeCall Girls Delhi
Ā 
Top Rated Bangalore Call Girls Mg Road āŸŸ 8250192130 āŸŸ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road āŸŸ 8250192130 āŸŸ Call Me For Genuine Sex...Top Rated Bangalore Call Girls Mg Road āŸŸ 8250192130 āŸŸ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road āŸŸ 8250192130 āŸŸ Call Me For Genuine Sex...narwatsonia7
Ā 
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
Ā 
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual Needs
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual NeedsBangalore Call Girl Whatsapp Number 100% Complete Your Sexual Needs
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual NeedsGfnyt
Ā 
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort ServicePremium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Servicevidya singh
Ā 
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
Ā 
Call Girls Bareilly Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bareilly Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Bareilly Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bareilly Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
Ā 
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
Ā 
Top Rated Bangalore Call Girls Richmond Circle āŸŸ 8250192130 āŸŸ Call Me For Gen...
Top Rated Bangalore Call Girls Richmond Circle āŸŸ 8250192130 āŸŸ Call Me For Gen...Top Rated Bangalore Call Girls Richmond Circle āŸŸ 8250192130 āŸŸ Call Me For Gen...
Top Rated Bangalore Call Girls Richmond Circle āŸŸ 8250192130 āŸŸ Call Me For Gen...narwatsonia7
Ā 
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...Garima Khatri
Ā 
ā™›VVIP Hyderabad Call Girls ChintalkuntašŸ–•7001035870šŸ–•Riya Kappor Top Call Girl ...
ā™›VVIP Hyderabad Call Girls ChintalkuntašŸ–•7001035870šŸ–•Riya Kappor Top Call Girl ...ā™›VVIP Hyderabad Call Girls ChintalkuntašŸ–•7001035870šŸ–•Riya Kappor Top Call Girl ...
ā™›VVIP Hyderabad Call Girls ChintalkuntašŸ–•7001035870šŸ–•Riya Kappor Top Call Girl ...astropune
Ā 
Best Rate (Hyderabad) Call Girls Jahanuma āŸŸ 8250192130 āŸŸ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma āŸŸ 8250192130 āŸŸ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma āŸŸ 8250192130 āŸŸ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma āŸŸ 8250192130 āŸŸ High Class Call Girl...astropune
Ā 

Recently uploaded (20)

College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
Ā 
Night 7k to 12k Chennai City Center Call Girls šŸ‘‰šŸ‘‰ 7427069034ā­ā­ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls šŸ‘‰šŸ‘‰ 7427069034ā­ā­ 100% Genuine E...Night 7k to 12k Chennai City Center Call Girls šŸ‘‰šŸ‘‰ 7427069034ā­ā­ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls šŸ‘‰šŸ‘‰ 7427069034ā­ā­ 100% Genuine E...
Ā 
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
Ā 
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
Ā 
Call Girls Mumbai Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Mumbai Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Mumbai Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Mumbai Just Call 9907093804 Top Class Call Girl Service Available
Ā 
šŸ’ŽVVIP Kolkata Call Girls ParganasšŸ©±7001035870šŸ©±Independent Girl ( Ac Rooms Avai...
šŸ’ŽVVIP Kolkata Call Girls ParganasšŸ©±7001035870šŸ©±Independent Girl ( Ac Rooms Avai...šŸ’ŽVVIP Kolkata Call Girls ParganasšŸ©±7001035870šŸ©±Independent Girl ( Ac Rooms Avai...
šŸ’ŽVVIP Kolkata Call Girls ParganasšŸ©±7001035870šŸ©±Independent Girl ( Ac Rooms Avai...
Ā 
The Most Attractive Hyderabad Call Girls Kothapet š– ‹ 6297143586 š– ‹ Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet š– ‹ 6297143586 š– ‹ Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet š– ‹ 6297143586 š– ‹ Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet š– ‹ 6297143586 š– ‹ Will You Mis...
Ā 
All Time Service Available Call Girls Marine Drive šŸ“³ 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive šŸ“³ 9820252231 For 18+ VIP C...All Time Service Available Call Girls Marine Drive šŸ“³ 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive šŸ“³ 9820252231 For 18+ VIP C...
Ā 
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeTop Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Ā 
Top Rated Bangalore Call Girls Mg Road āŸŸ 8250192130 āŸŸ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road āŸŸ 8250192130 āŸŸ Call Me For Genuine Sex...Top Rated Bangalore Call Girls Mg Road āŸŸ 8250192130 āŸŸ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road āŸŸ 8250192130 āŸŸ Call Me For Genuine Sex...
Ā 
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
Ā 
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual Needs
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual NeedsBangalore Call Girl Whatsapp Number 100% Complete Your Sexual Needs
Bangalore Call Girl Whatsapp Number 100% Complete Your Sexual Needs
Ā 
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort ServicePremium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Ā 
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Ā 
Call Girls Bareilly Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bareilly Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Bareilly Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bareilly Just Call 9907093804 Top Class Call Girl Service Available
Ā 
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Ā 
Top Rated Bangalore Call Girls Richmond Circle āŸŸ 8250192130 āŸŸ Call Me For Gen...
Top Rated Bangalore Call Girls Richmond Circle āŸŸ 8250192130 āŸŸ Call Me For Gen...Top Rated Bangalore Call Girls Richmond Circle āŸŸ 8250192130 āŸŸ Call Me For Gen...
Top Rated Bangalore Call Girls Richmond Circle āŸŸ 8250192130 āŸŸ Call Me For Gen...
Ā 
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
Ā 
ā™›VVIP Hyderabad Call Girls ChintalkuntašŸ–•7001035870šŸ–•Riya Kappor Top Call Girl ...
ā™›VVIP Hyderabad Call Girls ChintalkuntašŸ–•7001035870šŸ–•Riya Kappor Top Call Girl ...ā™›VVIP Hyderabad Call Girls ChintalkuntašŸ–•7001035870šŸ–•Riya Kappor Top Call Girl ...
ā™›VVIP Hyderabad Call Girls ChintalkuntašŸ–•7001035870šŸ–•Riya Kappor Top Call Girl ...
Ā 
Best Rate (Hyderabad) Call Girls Jahanuma āŸŸ 8250192130 āŸŸ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma āŸŸ 8250192130 āŸŸ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma āŸŸ 8250192130 āŸŸ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma āŸŸ 8250192130 āŸŸ High Class Call Girl...
Ā 

Quality_in_the_Coag_Lab_082819.pdf

  • 1. Quality in the Coagulation Laboratory 8/28/2019 Donna D. Castellone, MS,MT(ASCP)SH New York Presbyterian Hospital-Columbia QA Manager, Specialty Testing Supervisor, Special Coagulation & Hematology
  • 2. 2 Ā© 2019 Cardinal Health. All Rights Reserved. Quality in the Coagulation Laboratory Objectives ā€¢ Identify solutions in areas of coagulation that can be enhanced by implementing good laboratory practices and standard operating procedures ā€¢ Analyze real case studies and how to troubleshoot testing to determine root cause and what steps to prevent reoccurrence. ā€¢ Enhance general knowledge of coagulation, standards and methods.
  • 3. 3 Ā© 2019 Cardinal Health. All Rights Reserved. Quality in the Coagulation Laboratory WHY do we need this? ā€¢ Enhance diagnostic accuracy ā€¢ Improve patient care and safety ā€¢ Commutability of healthcare information ā€¢ Coagulation is comprised of enzymes and unstable proteins ā€“ a unique challenge
  • 4. 4 Ā© 2019 Cardinal Health. All Rights Reserved. What do you want from your Coagulation assays? How do you do this? Coagulation Assays ā€¢ Precise results ā€¢ Robust assays ā€¢ Accurate results that can be used as an aid in diagnosis and treatment ā€¢ Reproducible results How do you do this? ā€¢ Good SOPā€™s ā€¢ Effective QC practices ā€¢ Enroll in proficiency testing programs ā€¢ Good Laboratory Practices ā€¢ Understand coagulation results and their impact on patient care
  • 5. 5 Ā© 2019 Cardinal Health. All Rights Reserved. Healthcare quality is getting the right care to the right patient at the right time ā€“ every time!
  • 6. 6 Ā© 2019 Cardinal Health. All Rights Reserved. Quality in the Coagulation Laboratory Quality Management System ā€¢ Quality Management Plan - There must be a document that describes the overall QM program ā€¢ It can be a broad-range plan can cover all of the CLIA Quality System requirements ā€¢ Should contain monitors of key indicators of quality ā€¢ These indicators should be during the pre-analytical, analytical and post analytical stage
  • 7. 7 Ā© 2019 Cardinal Health. All Rights Reserved. What Key Indicators can be Monitored in Coagulation? THRESHOLDS SHOULD BE MEANINGFUL AND ACHIEVABLE: Pre-analytical: ~64% errors in this phase ā€¢ Short draw samples, hemolysis, time to be processed ā€¢ Who are your repeat offender-ed, pediatrics Analytical: Test Order Accuracy ā€¢ Why are certain test being ordered? Excess amount of Factor X testing being performed. Investigate, contact physicians ā€¢ Rejection logs; data mining- found wrong tubes being processed for testing, platelet testing without scheduling, testing getting cancelled ā€¢ Turnaround Time: poor turn around time, may be due to add on testing; how do you set your benchmarks- look to guidelines or clinician expectations - in particular for factor assays Post-analytical: ā€¢ Reports ā€“ flagging values
  • 8. 8 Ā© 2019 Cardinal Health. All Rights Reserved. Quality Assessment Formal Event Recording 45% 33% 22% Pre-analytical Analytical Post-analytical JANUARY ā€“ JUNE 2019
  • 9. 9 Ā© 2019 Cardinal Health. All Rights Reserved. Quality Assessment Recorded Issues 0% 5% 10% 15% 20% 25% Delayed results Collection issue result not called incorrect results Sample not received Tests not performed
  • 10. 10 Ā© 2019 Cardinal Health. All Rights Reserved. TAT for Coagulation Scoreboard Factor VIII Jan Feb March April May < 24 hours 90% 96% 98% 85% 88% > 24 hours 10% 4% 2% 15% 12% TAT 0.5 0.4 0.4 0.9 1.1 % compliance 100% 100% 100% 100% 100% BENCHMARK: 2 DAYS
  • 11. 11 Ā© 2019 Cardinal Health. All Rights Reserved. Evaluate Errors Root Cause Analysis ā€¢ Want to identify and evaluate errors, incidents and other problems that may interfere with patient care services. Can no longer say, ā€œI canā€™t control thatā€ ā€¢ Need a mechanism to capture internal and external sources such as complaints, including mistakes and near misses ā€¢ Conduct root cause analysis of occurrences, issues, errors and gaps in quality ā€¢ Outcome is to demonstrate risk reduction activities based on such root cause analyses ā€“ do not band-aid situations
  • 12. 12 Ā© 2019 Cardinal Health. All Rights Reserved. Plan. Do. Check. Act HTTP://ASQ.ORG/LEARN-ABOUT-QUALITY/PROJECT-PLANNING-TOOLS/OVERVIEW/PDCA-CYCLE.HTML ā€¢ PDCA model of continuous improvement ā€¢ Deming cycle or wheel of improvement ā€¢ When starting a new improvement project ā€¢ When developing a new or improved design of a process, product or service. ā€¢ When defining a repetitive work process ā€¢ When planning data collection and analysis in order to verify and prioritize problems or root causes ā€¢ When implementing any change
  • 13. 13 Ā© 2019 Cardinal Health. All Rights Reserved. Continuous Improvement ā€¢ PLAN: Specimen rejection log for Special coagulation P2Y12 and Aspirin. 40% of the samples rejected are due to improper tube being collected. ā€¢ DO: Track where the frequent offenders are, and follow up for re-training, posted reminders of what tubes to use. ā€¢ CHECK: Monitor the amount of samples that continue to be collected improperly. ā€¢ ACT: After 1 month, 50% reduction in the number of rejected tubes, of those, 28% were due to improper tube collection. Continue to evaluate where the issues are coming from, and retrain. Most amount occurred in 1 day and were from phlebotomy, they were alerted and retrained.
  • 14. 14 Ā© 2019 Cardinal Health. All Rights Reserved. QI-11 Continuous Improvement ā€¢ PLAN: Implement less order entry errors in particular for factor X ā€¢ DO: monitor the number of factor X assays and patients also on heparin ā€¢ CHECK: appeared most of the factor X were in reality anti-xa assays. Caused patient treatment delays, possible under or over anticoagulated ā€¢ ACT: Our LIS has a pop-up window when ordering this test: Factor X assay looks at the activity levels of factor X and does not monitor heparin levels
  • 15. 15 Ā© 2019 Cardinal Health. All Rights Reserved. Internal Audit ā€“ Tracer Method LOOK AT THE ENTIRE SYSTEM FOR A SPECIFIC ANALYTE FROM START TO FINISH Must have a purposeful document that follows the pre-analytic, analytic, and post-analytical process Any deficiencies must have a corrective action and date by which this should be completed. For more critical findings, TAT for correction should be ASAP ā€“ Looks at: ā€¢ Are we adhering to our own quality system? ā€¢ Is this process effective? ā€¢ Is the system as a whole effective? ā€¢ Look for opportunities to improve
  • 16. 16 Ā© 2019 Cardinal Health. All Rights Reserved. Findings of a Tracer Audit: Bethesda VIII ā€“ Questionable Patient Results DISCOVERY: Pre-analytical: Inhibitors were being performed on all requests, high FVIIIā€™s Analytical: 1. Using the same material for calibration as control 2. Patient dilutions were not clear 3. Protocol did not follow process 4. Analyzer settings did not allow proper identification of test being ordered Post-analytical: Test is linear between 25-75%, results were being extrapolated; needed to change our dilutions results within the curve
  • 17. 17 Ā© 2019 Cardinal Health. All Rights Reserved. Corrective Actions: METHODICALLY GOING THROUGH THE ENTIRE PROCESS UNCOVERED SEVERAL PRACTICES THAT NEED TO BE CLARIFIED ā€¢ Flow chart posted to understand inhibitor process ā€¢ Protocol re-written ā€“ everyone re-trained ā€¢ Implemented different material for controls and calibrators ā€¢ Set up clear testing parameters on the analyzer ā€¢ Updated dilutions and method of calculations
  • 18. 18 Ā© 2019 Cardinal Health. All Rights Reserved. Inspection Readiness ļ± Reduce number of inspection deficiencies ļ± Need a process to keep up with changing requirements and regulations ļ± Engage staff in the quality process
  • 19. 19 Ā© 2019 Cardinal Health. All Rights Reserved. Top 10 CAP Deficiencies 2016 COLLEGE OF AMERICAN PATHOLOGISTS Directive Requirement # of deficiencies GEN.55500 Competency 1979 COM.01200 Activity Menu 1810 COM. 10000 Procedures 1345 COM. 01700 PT evaluation 1178 COM.10100 Procedure review 1137 GEN. 20375 Document control 1036 COM.30300 Reagent Labeling 1032 COM. 01400 PT Attestation 968 COM. 04200 Monthly Review 919 COM. 30450 New Lot Confirmation 897
  • 20. 20 Ā© 2019 Cardinal Health. All Rights Reserved. Main Issues in Coagulation: 1. Platelet poor plasma 2. INR verification 3. Correct reporting/use of your D-dimer 4. Linearity studies, standard curves 5. Evaluation of educational challenges ā€“ must be purposeful 6. Heparin therapeutic range 7. Providing information to clinicians regarding impact of anticoagulant
  • 21. 21 Ā© 2019 Cardinal Health. All Rights Reserved. PT: FVIII Assay Result: 151 : Mean =123 SD= 15.1 SDI = 1.8 Limits: 98-148 ā€¢ No PT material left ā€¢ Checked curves, checked QC, brand new lot of reagents- not a lot of historical QC ā€¢ QC was in, but on the higher side ā€¢ Calibrate assay daily- compared the curves- the highest point was most different ā€¢ Re-ran some patients from that day ā€¢ Slightly lower results, but not moving from normal to abnormal even on borderline results (Reference range is 56-191) ā€¢ Sometimes the result may be statistically significant, but NOT clinically significant- e.g. investigate YES- impact on patient care ā€“ minimal ā€¢ A report should be written up with all of your findings, possible reason and action. Signed by the director or designee
  • 22. 22 Ā© 2019 Cardinal Health. All Rights Reserved. Quality Control Purpose: detect errors in laboratory that may lead to presence of clinical error ā€¢ This error forces a change in diagnosis or treatment in a patient unnecessarily ā€¢ Ensures consistency, accuracy and reliability of patient test results & reports ā€¢ Levels should reflect the testing population ā€¢ Ranges should be validated by running prior to being put into use ā€¢ Standards, run as QC to determine recovery of assigned value ā€¢ Donā€™t forget if you run low factor assays to run controls in that range
  • 23. 23 Ā© 2019 Cardinal Health. All Rights Reserved. When to Recalibrate? 1 Level of Control is Out: Re-Run ā€¢ Check calibration curve to see if a point is very different, or curve is very different- may need to re-run either a point, or the curve ā€¢ After recalibration- QC still out discard all reagents ā€¢ Before calling service- ā€¢ Try running PT material- peer evaluated material a lot of information ā€¢ Look at independent controls 2 Levels of Control are out: Recalibrate
  • 24. 24 Ā© 2019 Cardinal Health. All Rights Reserved. Action Protocol for Changes in Monthly Statistics LABORATORY MUST HAVE AN ACTION PROTOCOL THAT DEFINES WHAT TO DO WHEN STATISTICS FALL OUTSIDE OF TOLERANCE LIMITS ā€¢ What are those limits- can use published CVā€™s, or manufacturerā€™s recommendation ā€¢ When do you change range? Procedure and actions must match. ā€¢ State a level in your procedure that can prompt a change: e.g. a deviation in monthly QC >10% ā€¢ For factor assays, was their a change in the lot of deficient plasma ā€¢ When you have reviewed all documentation and there are no outside sources of error, and a shift has occurred- change the range ā€¢ Document what and why it was done
  • 25. 25 Ā© 2019 Cardinal Health. All Rights Reserved. Case Study: New Lot of Quality Control Laboratory Regulations: run control 20x to demonstrate the ranges ā€¢ Do I have to use these ranges? ā€¢ How do I run my 20 points? ā€¢ Over 20 days- or 5 times over 4 days ā€¢ Can I use the manufacturers range?
  • 26. 26 Ā© 2019 Cardinal Health. All Rights Reserved. Results for FX Normal Control: MEAN 95% RANGE 85-105 FVIII Day 1 Day 2 Day 3 Day 4 95 93 95 96 94 94 96 95 95 95 95 95 94 93 96 96 94 94 95 96 MANUFACTURERS RANGE: 80 ā€“ 120 SEC
  • 27. 27 Ā© 2019 Cardinal Health. All Rights Reserved. Case Study Review of the 1st Month ā€¢ Mean 102% ā€¢ Out of 30 days- 10 days spent, re-running controls- ā€¢ Wasting time and reagents to get controls within range ā€¢ This was happening throughout all of the laboratories ā€¢ If you review the 20 control runs, looks like a precision run ā€¢ Not unique to coagulation- all disciplines ā€¢ If we would enter these ranges, we would spend a lot of time and money and effort in having techs re-run, re-calibrate and hold up the runs. ā€¢ Running 20 controls was not working
  • 28. 28 Ā© 2019 Cardinal Health. All Rights Reserved. Case Study Solution ā€¢ We still run the 20 points; however we use this as a verification of the manufacturerā€™s ranges. ā€¢ The first month, we enter the manufacturers ranges. ā€¢ After a month of running in a real-world setting, we re-evaluate the ā€œtrueā€ ranges. ā€¢ Much less time wasted on trying to hit a ā€œperfect targetā€ which did not really reflect the testing environment- your analyzer, your reagents in your lab, with different operators
  • 29. 29 Ā© 2019 Cardinal Health. All Rights Reserved. Case Study: New Shipment for Lot of FVIII Deficient Plasma ā€¢ New lot of aPTT reagents had already been validated with this lot of FVIII deficient plasma ā€¢ However, a new shipment of factor VIII was being put in use ā€¢ New shipment is evaluated: FVIII Old lot New Lot % difference Acceptable Abnormal control 42% 41% 1% y Normal control 98% 100% 2% y Abnormal patient 35% 22% 37% n Normal patient 123% 97% 21% n
  • 30. 30 Ā© 2019 Cardinal Health. All Rights Reserved. Can I use these Reagents? ā€¢ Controls are within range- patient samples are not within range- ā€¢ FVIII is heat labile, maybe the samples are bad? ā€¢ Tested another normal and abnormal patient sample, still out of range ā€¢ Check independent controls matrix mimics FFP FVIII Control result OLD Control result NEW % difference Acceptable Abnormal control 34 21 38% N Normal control 98 77 21% N
  • 31. 31 Ā© 2019 Cardinal Health. All Rights Reserved. New Lot of Reagents, Minimal Historical Data ā€¢ Recalibrated, no change ā€¢ Could it be the reagents? All other aPTT based assays were working fine. ā€¢ However, did reconstitute new reagents, still no change ā€¢ Bad lot of deficient plasma- even though we have been using the lot same as previously used-Possible shipment issue- got a new lot/shipment- ā€¢ All results were within acceptable levels ā€¢ MUST test both patient and controls- ā€¢ Helps to have independent controls on hand ā€¢ Also can test PT material- can be very helpful in troubleshooting
  • 32. 32 Ā© 2019 Cardinal Health. All Rights Reserved. Case Study: Understanding your Reagents ā€¢ 16-year-old male for a hernia operation ā€¢ Family has a positive history of bleeding ā€¢ Results: PT=12.9 (10.2-13.5) o APTT=33.0 (29.9-33.5) I. Will this patient bleed? II. Do we check for a factor deficiency? III. What is the most important information when evaluating if a patient will bleed?
  • 33. 33 Ā© 2019 Cardinal Health. All Rights Reserved. The Coagulation Cascade Intrinsic Pathway Extrinsic Pathway Common Pathway
  • 34. 34 Ā© 2019 Cardinal Health. All Rights Reserved. Results: ā€¢ Look at APTT factors: VIII, IX and XI (nr 50-150%) ā€¢ VIII = 102% ā€¢ IX= 84% ā€¢ XI=21% ā€¢ Abnormal level of factor XI despite the normal APTT ā€¢ Patient is deficient in factor XI ā€¢ Shouldnā€™t the APTT have been abnormal
  • 35. 35 Ā© 2019 Cardinal Health. All Rights Reserved. How are your Reagents? ā€¢ Do your reagents truly reflect normal factor levels ā€¢ Normal PT & APTT levels indicate patient have a minimum of ~ 30% of factor levels present ā€¢ If your reagent is insensitive to a factor, you may get a normal PT or APTT ā€¢ The reagent may not be able to pick up a factor level below 30%
  • 36. 36 Ā© 2019 Cardinal Health. All Rights Reserved. Test Reagent Sensitivity for Factors This reagent does not reflect an abnormality until 12.5% ā€¢ Dilute normal plasma with factor deficient plasma at different levels ā€¢ Run either a PT or APTT on the sample ā€¢ Compare the results to the upper limit of the normal range ā€¢ You may get a normal PT or APTT with an abnormal % factor level FACTOR XI SENSITIVITY: APTT= (29.5 ā€“ 33.5) NORMAL + DEFICIENT = % ACTIVITY APTT PLASMA PLASMA 500UL 0UL 100% 29.0 250UL 250UL 50% 31.0 125UL 375UL 25% 32.0 62.5UL 437.5 12.5% 36.0
  • 37. 37 Ā© 2019 Cardinal Health. All Rights Reserved. Factor Sensitivity ā€¢ It is important to understand how your reagents perform ā€¢ Small investment for a lot of information ā€¢ Should be performed even if you do not run factor assays to understand what the approximate levels of factor activity will prolong screening tests ā€¢ In this case the family history is positive: o FXI deficiency occurs in up to 8% of Ashkenazi Jews o Incidence is estimated at 1 in 100,000 in the general population o FXI deficiency is inherited in an autosomal recessive pattern o Men and women are affected equally o The level of FXI doesnā€™t correlate to bleeding
  • 38. 38 Ā© 2019 Cardinal Health. All Rights Reserved. Case Study: PRE-OPERATIVE PATIENT APTT 49.1 (25 ā€“ 36 SECONDS) ā€¢ No history of bleeding ā€¢ No medications ā€¢ What can prolong an APTT ā€¢ Pre-analytical variables- short draw ā€¢ Presence of heparin or DOACs- can perform a thrombin time for detection ā€¢ What factors deficiencies may prolong the APTT? ā€¢ Those in the Intrinsic pathway: VIII, IX, XI, XII ā€¢ An APTT may also be prolonged due to an inhibitor
  • 39. 39 Ā© 2019 Cardinal Health. All Rights Reserved. Prolonged Screening Test: NEXT STEP IS TO PERFORM A MIXING STUDY ļ‚§ Correction: factor deficiency ( patient may bleed, unless FXII) ļ‚§ No correction: inhibitor ( more likely a risk for thrombosis, unless specific factor inhibitor, then they can bleed) I. Do you use FFP for the mixing study, can I use a lyophilized standard human plasma? II. Do I incubate? How long? III. Do I incubate separately? Together? Both? IV. Is this test cost effective? What information do I get from it?
  • 40. 40 Ā© 2019 Cardinal Health. All Rights Reserved. Classical 1:1 Mixing Study Step 1: Immediate Mix Patient PNP C 1:1 Mix 1. Test all. 2. Compare C to A & B. Did C correct? 3. Incubate all. A B Step 2: Incubated Mix 1-2 hrs., 37C Patient PNP C Incubated Tubes A B D Control 1. Test all. 2. Compare incubated C to D. Is it prolonged? Yes, possible time dependent inhibitor. 3. Interpret final results Mix
  • 41. 41 Ā© 2019 Cardinal Health. All Rights Reserved. Results ā€¢ APTT = 42.1 seconds (25-36 seconds) ā€¢ Pooled Normal Plasma =31.5 seconds ā€¢ Mixing study 1:1 = 36.5 ā€¢ Is this a correction? What are the variables that can impact a mixing study? 1. PPP critical for proper id of AN INHIBITOR; platelets are phospholipids and will falsely shorten your results, when reacting with reagents 2. Reagent Sensitivity: at what level of factor deficiency will prolong your aPTT 3. Normal Pool plasma: must be ppp, have adequate levels of factors COMPRISED OF A MINIMUM OF NORMAL INDIVIDUALS, NOT NORMAL PATIENTS
  • 42. 42 Ā© 2019 Cardinal Health. All Rights Reserved. Interpretation of a Mixing Study NO RECOMMENDATIONS OR GUIDELINES FROM ISTH SCIENTIFIC & STANDARDIZATION COMMITTEE CAP COAGULATION RESOURCE COMMITTEE ā€¢ Based on PT or aPTT normal range Within limits @ 2SD or 3SD Within 5 seconds of the 2SD upper limit ā€¢ Rosner Index o 1:1 mix o Index = CT of 1:1 mix ā€“ CT of NPP x 100 CT of patient <15 = FD >15 = Inhibitor NPP tested with Mixing Study o Within 5 seconds of NPP value o 10% of NPP value o Lack of correction @ >15% of NPP
  • 43. 43 Ā© 2019 Cardinal Health. All Rights Reserved. What do we have? (In my Lab) ā€¢ A repeat APTT that is still prolonged 42.1 seconds (25-36 seconds) ā€¢ PNP =31.0 seconds, mixing study is 36.4 seconds 5.5 seconds from PNP ā€¢ > 5 seconds from PNP ā€¢ > normal range ā€¢ Not a correction/partial correction ā€¢ Proceed with inhibitor testing
  • 44. 44 Ā© 2019 Cardinal Health. All Rights Reserved. How to Standardize this Test: ā€¢ Adhere to consistent policies in your laboratory- define your criteria and stick to it ā€¢ Eliminate the possibility of heparin being on board, perform a thrombin time, best test for residual heparin ā€¢ Can report just numbers and let the clinicians interpret ā€¢ Can report with an interpretation ā€¢ Eliminate the test, many laboratories feel it is too confusing, and just proceed with additional testing
  • 45. 45 Ā© 2019 Cardinal Health. All Rights Reserved. Case Study PATIENT PRESENTS WITH A DVT, PLACED ON UNFRACTIONATED HEPARIN ā€¢ Given a bolus dose of heparin - monitored by the APTT ā€¢ First test taken 4 hours post dose ā€¢ APTT= 67.5 sec (25-35 sec) ā€¢ Develops a PE ā€¢ Was the patient properly anticoagulated?
  • 46. 46 Ā© 2019 Cardinal Health. All Rights Reserved. Based on APTT Result ā€¢ Prolonged aPTT ā€¢ What is the sensitivity of the aPTT reagent to heparin ā€¢ Not good enough to use 1.5-2.5 times the mean of the normal range ā€¢ Performed an anti-Xa assay = 0.25U/ml anti-Xa (0.3-0.7u/Ml) ā€¢ Patient is under anticoagulated ā€¢ If results are inaccurate and anticoagulation is excessive there could be a bleed ā€¢ If results are inadequate patients could clot
  • 47. 47 Ā© 2019 Cardinal Health. All Rights Reserved. APTT Monitoring - Disadvantages ā€¢Numerous factors elevate APTT: o Warfarin therapy o Lupus anticoagulants o Factor deficiency, liver disease ā€¢APTT response to heparin exaggerated ā€¢Potential to under-anticoagulate patients ā€¢Elevated factor VIII and I, shorten APTT ā€¢Potential to over-anticoagulate patients ā€¢No standard to cross reference APTT among laboratories ā€¢No dose response relationship with heparin and APTT Winter, W and Harris, WE, (2008), Anti-coagulant Monitoring in the Core Lab, Advance for Laboratory Professionals, pg.54-61.
  • 48. 48 Ā© 2019 Cardinal Health. All Rights Reserved. CAP Requirement ā€¢ It is recommended that the first method be used initially to establish the therapeutic range before starting patient testing with a new instrument or new reagent, while the second method can be used for validation of the therapeutic range with subsequent reagent lot changes ā€¢ It is not best practice to use plasma samples spiked with heparin in vitro to calculate the therapeutic range, as differences in heparin binding proteins in vitro may lead to overestimation of the therapeutic range
  • 49. 49 Ā© 2019 Cardinal Health. All Rights Reserved. Heparin Therapeutic Range CAP recommends that each laboratory establishes its own heparin therapeutic range for each new lot of aPTT reagent based on reagent: instrument combination ā€¢ Heparin Anti-Xa Method o 50-60 plasma samples collected from patients treated with UFH o Mix of samples should span the therapeutic range o No more than two samples on the same patient o The PT/INR should normal (INR <1.3) o Assay aPTT, anti-factor Xa (UFH), and PT/INR o Plot the aPTT on the y-axis and the anti-Xa units on the X-axis o Draw a line through the middle of the points o The UFH therapeutic range is equivalent to ā€“ 0.3-0.7 units/mL Anti-Xa Heparin assay
  • 50. 50 Ā© 2019 Cardinal Health. All Rights Reserved. Brill-Edwards Technique THERAPEUTIC RANGE= 0.3-0.7U/ML 70-120 SECONDS
  • 51. 51 Ā© 2019 Cardinal Health. All Rights Reserved. Alternative Method: Subsequent Years Cumulative summation of reagent mean differences ā€¢ Initial therapeutic range established using the anti-Xa method ā€¢ Run aPTT on samples with both the old and new lot of reagent ā€¢ Plot the old lot on the X-axis and the new lot on the Y-axis ā€¢ Determine the sum, mean and difference of the results with each new lot ā€¢ Record the difference in the means to compare with past studies ā€¢ A change of <5 seconds between the differences in the means is acceptable ā€¢ A change of >5 seconds between the differences in the means requires action Year Mean Old Lot Mean New Lot Difference New- Old Cum Sum Action 2006 34.0 37.2 3.2 3.2 2007 37.2 33.0 -4.2 -1.2 Accept 2008 33.0 43.0 10.0 8.8 Reject 2008-B 33.0 36.0 3.0 1.8 Accept
  • 52. 52 Ā© 2019 Cardinal Health. All Rights Reserved. The ā€œDoā€™s" and ā€œDonā€™tsā€ to Enhance Quality in the Coagulation Laboratory DO ā€¢ DO ask questions: use the five whys to perform a root cause analysis ā€¢ DO Listen to the issues from your front-line staff, have everyone work on problem solving DONā€™T ā€¢ DONā€™T blame people: Focus on systems, not individuals- 99% of people want to do the right thing! ā€¢ DONā€™T say this canā€™t be solved, get someone who doesnā€™t know the process, a new set of eyes, they may see things you are missing
  • 53. 53 Ā© 2019 Cardinal Health. All Rights Reserved. Thank you for your time! ā€œQUALITY IS NOT A GIVEN BUT NEEDS TO BE A PRACTICE; PATIENT CARE AND TEST OUTCOMES DEPEND ON IT.ā€