Chapter-8 Coulometry and Electrogravimetric Analysis.pptx
1.
1
Chapter 8
Coulometry andElectrogravimetric Analysis
Coulometry is an electrochemical technique used to determine the
amount of a substance by measuring the total electric charge (Q)
required for its complete redox conversion (electrolysis).
It is an accurate, absolute (bulk) method based on Faraday's laws of
electrolysis where the charge is directly proportional to the amount of
chemical change.
The total charge, Q, in coulombs, passed during an electrolysis is
related to the absolute amount of analyte by Faraday’s law
Q = nFN
where n is the number of electrons transferred per mole of analyte, F
is Faraday’s constant (96487 C mol–1
), and N is the moles of analyte.
2.
2
There aretwo main types of coulometry
1. Constant-current coulometry (coulometric titration): A
constant current is applied, and the time it takes for the reaction to
complete is measured, which is then used to calculate the total
charge (Q = I t
⋅ e).
where te is the electrolysis time.
2. Controlled-potential coulometry: A constant potential is
applied, and the resulting current is measured over time as it
decreases, providing a measure of the total charge passed.
In this case, then the total charge is given by
3.
3
To obtainan accurate value for N, therefore, all the current must result in the
analyte’s oxidation or reduction.
In other words, coulometry requires 100% current efficiency i.e the
reaction consumes all the current.
Controlled-Potential Coulometry
The easiest method for ensuring 100% current efficiency is to maintain the
working electrode at a constant potential that allows for the analyte’s
quantitative oxidation or reduction, without simultaneously oxidizing or
reducing an interfering species.
The current flowing through an electrochemical cell under a constant
potential is proportional to the analyte’s concentration.
As electrolysis progresses the analyte’s concentration decreases, as does the
current.
4.
4
The resultingcurrent-versus-time profile for controlled-potential
coulometry, which also is known as potentiostatic coulometry,
looks like;
Integrating the area under the curve, from t = 0 until t = te,
gives the total charge.
5.
5
Selecting a ConstantPotential
In controlled-potential coulometry, the potential is selected so that
the desired oxidation or reduction reaction goes to completion
without interference from redox reactions involving other
components of the sample matrix.
For example, a constant-potential coulometric method for Cu2+
based on its reduction to copper metal at a Pt cathode working
electrode.
This reaction is favored when the working electrode’s potential is
more negative than +0.342 V versus the SHE or +0.093 V versus
the SCE.
6.
6
To maintaina 100% current efficiency, however, the potential must be
selected so that the reduction of H3O+
to H2 does not contribute
significantly to the total charge passed at the electrode.
The potential needed for a quantitative reduction of Cu2+
can be
calculated using the Nernst equation.
Minimizing Electrolysis Time
Since time is an important consideration in choosing and designing
analytical methods, the factors that determine the analysis time need to
be considered.
7.
7
For thisreason controlled-potential coulometry is carried out in
small-volume electrochemical cells, using electrodes with large
surface areas and high stirring rates.
A quantitative electrolysis typically requires approximately 30–60
min, although shorter or longer times are possible.
Instrumentation
Controlled-potential coulometry uses three-electrode potentiostat
(the device that used to control voltage and measure current).
Pt and Hg are commonly used as working electrodes.
However, the large overpotential for reducing H3O+
at mercury
makes it the electrode of choice for analytes requiring negative
8.
8
But itcan’t be used at potentials that are positive with respect to the SHE, Pt
is used when positive potentials are required.
A saturated calomel or Ag/AgCl electrode serves as the reference electrode.
The auxiliary electrode, which is often a Pt wire, is separated by a salt bridge
from the solution containing the analyte.
This is necessary to prevent electrolysis products generated at the auxiliary
electrode from reacting with the analyte and interfering in the analysis.
The other essential feature of instrumentation for controlled-potential
coulometry is a means of determining the total charge passed during
electrolysis.
9.
Analytical Chemistry andInstrumentation 9
One method is to monitor the current as a function of time and
determine the area under the curve.
However, modern instruments use electronic integration to
monitor charge as a function of time.
The total charge at the end of the electrolysis then can be read
directly from a digital readout or from a plot of charge versus
time.
10.
Controlled-Current Coulometry
Asecond approach to coulometry is to use a constant current in place
of a constant potential.
It is also known as amperostatic coulometry or coulometric titrimetry.
Even if it is not applicable for the analysis of mixture of ions (not
selective), it has two advantages over controlled-potential coulometry,
First, using a constant current makes for a more rapid analysis since the
current does not decrease over time.
Thus, a typical analysis time for controlled-current coulometry is less
than 10 min.
Second, the total charge is simply the product of current and time, i.e a
method for integrating the current–time curve is not necessary.
10
11.
11
• However, ithas two important problems that must be solved.
• First, as electrolysis occurs the analyte’s concentration and, the
current due to its oxidation or reduction steadily decreases.
• Thus, to maintain a constant current the cell potential must
change until another oxidation or reduction reaction can
occur at the WE.
12.
12
Unless thesystem is carefully designed, these secondary reactions will
produce a current efficiency of less than 100%.
The second problem is the need for a method of determining when the
analyte has been exhaustively electrolyzed (end point).
In controlled-potential coulometry this is signaled by a decrease in the
current to a constant background or residual current.
In controlled-current coulometry, however, a constant current continues
to flow even when the analyte has been completely oxidized or reduced.
A suitable means of determining the end-point of the reaction, te, is
needed.
13.
13
Maintaining Current Efficiency
To illustrate why changing the working electrode’s potential can lead to less
than 100% current efficiency, let’s consider the coulometric analysis for
Fe2+
based on its oxidation to Fe3+
at a Pt working electrode in 1 M H2SO4.
Initially the potential of the working electrode remains nearly constant at a
level near the standard-state potential for the Fe3+
/Fe2+
redox couple.
As the concentration of Fe2+
decreases, however, the potential of the
working electrode shifts toward more positive values until another
oxidation reaction can provide the necessary current.
Thus, in this case the potential eventually increases to a level at which the
oxidation of H2O occurs.
14.
14
• Since thecurrent due to the oxidation of H2O does not contribute
to the oxidation of Fe2+
, the current efficiency of the analysis is
less than 100%.
• To maintain a 100% current efficiency the products of any
competing oxidation reactions must react both rapidly and
quantitatively with the remaining Fe2+
.
• This may be accomplished, for example, by adding an excess of
Ce3+
to the analytical solution .
• When the potential of the working electrode shifts to a more
positive potential, the first species to be oxidized is Ce3+
.
15.
15
The Ce4+
producedat the working electrode rapidly mixes with the solution, where it
reacts with any available Fe2+
.
Combining these reactions gives the desired overall reaction of
In this manner, a current efficiency of 100% is maintained.
Furthermore, since the concentration of Ce3+
remains at its initial level, the potential
of the working electrode remains constant as long as any Fe2+
is present.
This prevents other oxidation reactions, such as that for H2O, from interfering with
the analysis.
A species, such as Ce3+
, which is used to maintain 100% current efficiency, is called
a mediator.
16.
End Point Determination
Using the same example, once all the Fe2+
has been oxidized current continues
to flow as a result of the oxidation of Ce3+
and, eventually, the oxidation of H2O.
What is needed is a means of indicating when the oxidation of Fe2+
is complete.
In this respect it is convenient to treat a controlled-current coulometric analysis
as if electrolysis of the analyte occurs only as a result of its reaction with the
mediator.
A reaction between an analyte and a mediator is identical to that encountered in
a redox titration.
Thus, the same end points that are used in redox titrimetry such as visual
indicators, potentiometric and conductometric measurements, may be used.
16
17.
17
Instrumentation
Controlled-current coulometrynormally is carried out using a galvanostat
(the divice that controls current and measure the time) and an
electrochemical cell consisting of a working electrode and a counter-
electrode.
The working electrode, which often is constructed from Pt, is also
called the generator electrode since it is where the mediator reacts to
generate the species reacting with the analyte.
The counter-electrode is isolated from the analytical solution by a salt
bridge or porous frit to prevent its electrolysis products from reacting with
the analyte.
Alternatively, oxidizing or reducing the mediator can be carried out
externally, and the appropriate products flushed into the analytical solution.
18.
18
• A solutioncontaining the mediator flows under the influence of
gravity into a small-volume electrochemical cell.
• The products generated at the anode and cathode pass through
separate tubes, and the appropriate oxidizing or reducing reagent
can be selectively delivered to the analytical solution.
• For example, external generation of Ce4+
can be obtained using an
aqueous solution of Ce3+
and the products generated at the anode.
19.
19
Applications Coulometry
Coulometrymay be used for the quantitative analysis of both inorganic
and organic compounds.
The majority of controlled-potential coulometric analyses involve the
determination of inorganic cations and anions, including trace metals
and halides.
The ability to control selectivity by carefully selecting the working
electrode’s potential, makes controlled-potential coulometry
particularly useful for the analysis of alloys.
For example, the composition of an alloy containing Ag, Bi, Cd, and Sb
can be determined by dissolving the sample and placing it in a matrix of
0.2 M H2SO4.
20.
20
A platinumworking electrode is immersed in the solution and held at a constant
potential of +0.40 V versus the SCE.
At this potential Ag(I) deposits on the Pt electrode as Ag, and the other metal ions
remain in solution.
When electrolysis is complete, the total charge is used to determine the amount of
silver in the alloy.
The potential of the platinum electrode is then shifted to –0.08 V versus the SCE,
depositing Bi on the working electrode.
When the coulometric analysis for bismuth is complete, antimony is determined by
shifting the working electrode’s potential to –0.33 V versus the SCE, depositing Sb.
Finally, cadmium is determined following its electrodeposition on the Pt electrode
at a potential of –0.80 V versus the SCE.
21.
21
Another areawhere controlled-potential coulometry has found
application is in nuclear chemistry, in which elements such as
uranium and polonium can be determined at trace levels.
For example, microgram quantities of uranium in a medium of
H2SO4 can be determined by reducing U(VI) to U(IV) at a
mercury working electrode.
Controlled-potential coulometry also can be applied to the
quantitative analysis of organic compounds, although the
number of applications is significantly less than that for
inorganic analytes.
One example is the six-electron reduction of a nitro group,
–NO2, to a primary amine, –NH2, at a mercury electrode.
22.
22
Solutions ofpicric acid, for instance, can be analyzed by reducing
to triaminophenol.
• In comparison with conventional titrimetry, there are several
advantages to the coulometric titrations.
• One advantage is that the electrochemical generation of a “titrant”
that reacts immediately with the analyte allows the use of reagents
whose instability prevents their preparation and storage as a
standard solution.
• Thus, highly reactive reagents such as Ag2+
and Mn3+
can be used in
coulometric titrations.
23.
23
Because itis relatively easy to measure small quantities of charge, coulometric
titrations can be used to determine small quantities of analyte that cannot be
measured accurately by a conventional titration.
Example 1, the concentration of H2S in a wastewater sample can be determined
by a coulometric titration using KI as a mediator and I3
–
as the “titrant.”
H2S(aq) + I3
–
(aq) ↔ 2H+
+ 3I–
(aq) + S(s) or
H2S(aq) + I3
–
(aq) + 2H2O(l) ↔ 2H3O+
(aq) + 3I–
(aq) + S(s)
A 50.00-mL sample of wastewater is placed in a coulometric cell, along with an
excess of KI and a small amount of starch as an indicator.
24.
24
Electrolysis iscarried out at a constant current of 84.6 mA,
requiring 386 s to reach the starch end point.
Calculate the concentration of H2S in the sample in parts per
million.
Solution, MW = 34.08 g/mol.
Then calculate the total charge (Q) passed:
Q = I × t = (84.6 × 10 ³ A) × (386 s)
⁻ = 32.6556 C
Calculate the number of mole (N) = Q / nF = 32.6556 C / 96485
C/mol e⁻x2 =1.69225 × 10 mol H S
⁻⁴ ₂
25.
25
Calculate themass of H S
₂
Mass of H S = (1.69225 × 10 mol) × (34.0
₂ ⁴
⁻ 8 g/mol)
Mass of H S = 0.005766 g
₂ = 5.766 mg
Calculate the concentration of H S in mg/L:
₂
The sample volume is 50.00 mL = 0.05000 L.
Convert the concentration to parts per million (ppm):
For dilute aqueous solutions, 1 mg/L is approximately equal to 1
ppm.
Concentration (ppm) = 115.32 ppm
26.
26
Example 2,a 0.3619-g sample of pure tetrachloropicolinic acid,
C6HNO2Cl4, is dissolved in distilled water, transferred to a 1000-
mL volumetric flask, and diluted to volume. An exhaustive
controlled-potential electrolysis of a 10.00-mL portion of this
homogenous solution requires 5.374 C of charge.
What is the value of n for this reduction reaction?
Solution, MW = 260.88 g/mol. the 10.00-mL portion of sample
contains 3.619 mg, or 1.39×10–5
mol of tetrachloropicolinic acid.
Solving for n and making appropriate substitutions gives
27.
27
Thus, reducinga molecule of tetrachloropicolinic acid requires
four electrons.
The overall reaction, which results in the selective formation of
3,6-dichloropicolinic acid,
28.
28
Electrogravimetry
If acceptablecurrent efficiency is not feasible during coulometric
analysis, it is possible to determine the mass of the analyte
deposited at the working electrode.
The analyte’s oxidation or reduction leads to its deposition on the
working electrode.
In this case the working electrode is weighed before beginning
the electrolysis and reweighed when electrolysis of the analyte is
complete.
The difference in the electrode’s weight gives the analyte’s mass.
This technique is known as electrogravimetry.
Mostly used for metal ions analysis.
29.
29
The oxidationof Pb2+
, and its deposition as PbO2 on a Pt anode is one
example of electrogravimetry.
Reduction also may be used in electrogravimetry.
The electrodeposition of Cu on a Pt cathode, for example, provides a
direct analysis for Cu2+
.
But this method has its own disadvantage;
1. Other metal ions reduced at similar potentials or interfering species
present, leading to impure deposits (e.g., chromium with copper).
2. Incomplete drying, contamination, or residual solution on the
electrode produce error.