InWave

User’s Manual
InWave
Table of Contents
Contact Information / Customer Service / Supplies / Support ........................ 4
About the InWave ........................................................................................... 5
Safety Information .......................................................................................... 6
Sensor and Lead Wire Setup ........................................................................... 9
Programming Instructions ............................................................................... 12
Preprogrammed Modes ................................................................................... 13
Controls........................................................................................................... 15
Operating Instructions..................................................................................... 16
Indications, Contraindications and Warnings………………………..17, 19, 21
Precautions and Adverse Reactions…………………………………..18, 20, 22
Troubleshooting .............................................................................................. 23
Specifications and Accessories ....................................................................... 24
Sensor, Batteries, and Cleaning ...................................................................... 25
Warranty ......................................................................................................... 26

3
Zynex Medical
Painezee
Contact Information
1-866-940-7030
0091-9030022402

CUSTOMER SERVICE
Supplies:

To order more electrodes or other accessories

Technical Support:

Questions or problems with using your device

Device Return:

Order a postage paid return envelope to return your
device at no charge

MAIN OFFICE
Billing Questions:

1-800-495-6670
0091-9030022402
Questions regarding insurance benefits and
covered benefits for Durable Medical Equipment
(DME) or questions about an Explanation of
Benefits form you received in the mail

FAX NUMBER

1-800-495-6695

MAILING ADDRESS
Zynex Medical, Inc.
Plot No: 6, Green Avenues,
9990 Park Meadows DriveNizampet Road,
Opp: Bhavysa Anandham,
Kukatpally, Hyderabad - 500090
Lone Tree, CO 80124-6739
Andhra
USA Pradesh, India
info@zynexmed.com
info@painezee.com

EMAIL

WEBSITE

www.zynexmed.com
www.painezee.com

4
About the InWave
The InWave is a multiple-mode stimulator which allows users a choice of
treatment options. The InWave causes muscle contractions which exercise and
strengthen the pelvic floor muscles.
Stress Incontinence Treatment
The output delivers electrical impulses at a frequency of 50 Hz, which produce
smooth, even muscle contractions. This type of stimulation exercises and
strengthens the pelvic floor muscles, which aids in the control of stress
incontinence.
Stimulation can be set to 3 or 5 second duration, with a resting period of 5 or
10 seconds between contractions.
Default treatment time is 15 minutes.
Urge Incontinence Treatment
The output delivers electrical impulses at a frequency of 12.5 Hz, which
produce a series of short, rapid contractions. This type of stimulation aids in
the relaxation of involuntary muscle contractions which cause urge
incontinence.
Stimulation can be set to 6 or 10 second duration, with a resting period of 3 or
10 seconds. Default treatment time is 20 minutes.
Mixed Incontinence Treatment
Mixed treatment modes consist of a 2, 5, or 10 minute stress treatment
followed by a 2, 5, or 10 minute urge treatment. This combination of
stimulation types aids in both stress and urge incontinence in the manners
mentioned above.
Caution: Federal law restricts this device to sale by or on the order of a
licensed healthcare practitioner.

5
Safety Information
 Equipment Classification
As per the International Electrotechnical Commission standard IEC
60601-1, and the European standard EN 60601-1, Medical Electrical
Equipment, General Requirements for Basic Safety and Essential
Performance, the InWave is classified as follows:


Type BF Equipment
A Type BF piece of equipment is one that provides a particular degree
of protection against electric shock, particularly regarding allowable
leakage current. Type BF applied part is one that is floating above
ground and is isolated from all other parts of the equipment.



Medical Device Directive (MDD) Classification
The InWave is a Class IIa medical device per the European Medical
Device Directive.



Internally Powered Equipment
The InWave is powered by an internal, 9 VDC, MN1604, battery, so
it is classified as Internally Powered Equipment.



Electromagnetic Compatibility
Conforms to IEC 60601-1-2.



Temperature
Operating temperature range: 0° - 40° C (32° - 104° F)
Storage temperature range: -20° - 70° C (-4° - 158° F)



Water Ingress
Ordinary equipment. This device does not have protection against
ingress of water.



Flammable Anesthetics
This device is not suitable for use in the presence of a flammable
anesthetic mixture with air, or in the presence of a flammable
anesthetic mixture with oxygen or nitrous oxide.

6
Safety Information


Mode of Operation
This device is suitable for continuous operation.

 Symbols

Safety symbols shown on the device above are defined below.

On/Standby. This symbol indicates that the labeled switch
electronically cycles the DC power on and off for part of the
equipment.
General Warning Sign.
Follow warnings stated in the
instruction manual to prevent potential hazards.
Refer to Instruction Manual/Booklet. The operator must read,
understand, and follow all instructions in the accompanying
document including all warnings, cautions, and precautions
before using this medical device.
Type BF Equipment. This symbol indicates that the patient
applied parts (electrodes) are Type BF (floating from ground)
offering the user a specific level of safety.

7
Waste Electrical and Electronic Equipment (WEEE). This
product may contain substances known to be hazardous to the
environment or to human health. It should be disposed of
properly (for example, at your local waste collection
administration or recycling plant) and in accordance with local
ordinances.

 Service and Calibration


Do not remove the cover. There are no user serviceable parts. Refer all
service to authorized personnel. No modification of the equipment is
allowed.



No preventative inspections are required. Factory testing and calibration
ensure equipment accuracy and response. Contact Zynex Medical for
factory re-calibration if necessary.

8
InWave
Sensor and Lead Wire Setup
Step 1

Open the sensor package and remove the sensor. Remove the
electrode wire from the packaging, and unbundle the wire.
Note: Only Zynex Medical sensors are approved for use with the
InWave. P/N 400015 are provided standard.

Step 2

Insert the lead wire pin connectors into the sensor connectors as
shown below.

Sensor
Connections

Lead Wire

9
InWave
Electrode and Lead Wire Setup (continued)
Step 3

Insert vaginal sensor as shown below. If needed, a small amount of
water-based lubricant can be applied to the tip of the sensor to aid
in insertion. Initial placement of the vaginal sensor and selection of
the treatment program should be performed under the guidance of a
physician. Note: Overuse of a lubricant can cause lack of
stimulation, and lead to ineffective treatment. In order to avoid
treatment interruption, it may be necessary to empty the bladder and
bowels before starting a treatment session.
See the directions included with the sensor for further information.
Note: The vaginal sensors are for single patient use only.

10
InWave
Electrode and Lead Wire Setup (continued)
Step 4

Plug the lead wire into the top of the device.

Step 5

Proceed to page 16 to start treatment or page 12 to program device.

Electrode
Connector

11
InWave
Device Programming Instructions
Turn device on by pressing the black On/Standby button
1.
2.
3.

Select the desired Modality by pressing the “STRESS”, “URGE”, or
“MIX” button once.
Press the “+” or “-” portions of the Program button until the arrow on the
LCD is next to the desired program.
Press the Treatment Timer button until the desired treatment time is set on
the screen.

The InWave is now ready to use.

On /
Standby
Button
Modality
Buttons

Treatment
Timer Button

12
InWave
Preprogrammed Modes
Stress Incontinence Treatment (50 Hz Frequency, 200 µSecond Pulse Width)
Program

Description

5/10

The output signal ramps up to the level set by the user over a
period of 1 second. The output stays on for the remainder of
the 5 second on time, then ramps down to zero over a period of
1 second. The output stays off for the remainder of the 10
second off time. This cycle repeats for the duration of the
treatment time. (Default is 15 minutes.)

3/10

The output signal ramps up to the level set by the user over a
period of 1 second. The output stays on for the remainder of
the 3 second on time, then ramps down to zero over a period of
1 second. The output stays off for the remainder of the 10
second off time. This cycle repeats for the duration of the
treatment time. (Default is 15 minutes.)

5/5

The output signal ramps up to the level set by the user over a
period of 1 second. The output stays on for the remainder of
the 5 second on time, then ramps down to zero over a period of
1 second. The output stays off for the remainder of the 5
second off time. This cycle repeats for the duration of the
treatment time. (Default is 15 minutes.)

Urge Incontinence Treatment (12.5 Hz Frequency, 600 µSecond Pulse Width)
Program

Description

10/3

The output signal ramps up to the level set by the user over a
period of 1 second. The output stays on for the remainder of
the 10 second on time, then ramps down to zero over a period
of 0.5 seconds. The output stays off for the remainder of the 3
second off time. This cycle repeats for the duration of the
treatment time. (Default is 20 minutes.)

6/3

The output signal ramps up to the level set by the user over a
period of 1 second. The output stays on for the remainder of
the 6 second on time, then ramps down to zero over a period
of 0.5 seconds. The output stays off for the remainder of the 3
second off time. This cycle repeats for the duration of the
treatment time. (Default is 20 minutes.)

10/10

The output signal ramps up to the level set by the user over a
period of 1 second. The output stays on for the remainder of
the 6 second on time, then ramps down to zero over a period
of 0.5 seconds. The output stays off for the remainder of the 3
second off time. This cycle repeats for the duration of the
treatment time. (Default is 20 minutes.)

13
Mixed Incontinence Treatment
Program

Description

10:10

A 10 minute “5/10” Stress program is run, followed by a 10
minute “10/3” Urge program. This cycle repeats for the
duration of the treatment time. Default treatment time is 20
minutes.

5:5

A 5 minute “5/10” Stress program is run, followed by a 5
minute “10/3” Urge program. This cycle repeats for the
duration of the treatment time. Default treatment time is 20
minutes.

2:2

A 2 minute “5/10” Stress program is run, followed by a 2
minute “10/3” Urge program. This cycle repeats for the
duration of the treatment time. Default treatment time is 20
minutes.

The InWave should be used daily for 15 to 20 minutes per session, for 1 to 3 months,
or as directed by a physician.

14
InWave
Device Controls

Display

Attach Lead Wire Here

Lead Wire
Connector

Turns Unit On/Off

On/Standby
Button

Cycles Through Programs

Adjusts Treatment Intensity

Program Select Button

Up and Down Button

Unlocks Control Buttons

Unlock Button
STRESS, URGE, MIX

Modality Buttons
Adjusts Treatment Times

Treatment Timer Button

15
InWave
Device Operating Instructions
Start Treatment:
Before starting treatment, the sensor must be inserted and the lead wire must
be connected to the InWave. See pages 9-11 for instructions.
1.

Turn the InWave on by pressing the On/Standby button once, and set the
desired program using the Modality and Program buttons.

2.

Increase the stimulation level by pressing the upper portion of the Intensity
button until a strong but comfortable stimulation level is felt.
NOTE: Output increases in 1/2 mA increments up to the 4 mA level, then
increases in increments of 1 mA.

3.

Once the desired level of stimulation is set, the InWave will automatically
shut off after the preset treatment time (shown on the display.) If the
Treatment Timer has been set to, “No Timer,” then the InWave will need
to be shut off manually. Refer to Programming Instructions on page 12 to
adjust the Treatment Timer.

4.

To turn off the InWave manually, press the On/Standby button.
Note: If “Check Connections!” message is displayed on the screen, see “Display Alerts”
below.

During Treatment:
IMPORTANT: Button controls lock after 20 seconds of inactivity. To unlock
button controls, press the Unlock button.
Increase Intensity:
To increase the intensity, press the upper potion of the Intensity button until
the desired level of stimulation is felt.
Decrease Intensity:
To decrease the intensity, press the lower portion of the Intensity button until
the desired level of stimulation is felt.
Display Alerts:
Check Connections: The lead wire and/or the sensor may not be attached
properly. Check all connections and try again (see page 23). If the problem
persists, call Technical Support (see page 4).
Low Battery: Replace the battery immediately.
Locked: The stimulation level and mode cannot be changed until the Unlock
button is pressed.

16
InWave
Indications for Use
Stress Incontinence Mode
Please read the InWave User’s Manual before using the InWave pelvic
muscle stimulator.
CAUTION: Federal law restricts this device to sale by or on the order of a
licensed healthcare practitioner. This device should only be used under medical
supervision for adjunctive therapy for the treatment of medical diseases and
conditions.
Indications


The Zynex Medical InWave is intended to provide electrical stimulation
and neuromuscular re-education for the purpose of rehabilitation of weak
pelvic floor muscles for the treatment of stress, urge, and mixed urinary
incontinence in women.

Contraindications


This InWave should not be used in the presence of the following:
-Where there is a tendency to hemorrhage following acute trauma or
fracture.
-Following recent surgical procedures when muscle contraction may
disrupt the healing process.
-Over the menstruating or pregnant uterus.
-Over areas of the skin which lack normal sensation.
-Irregular menstrual cycles.
-Urinary or vaginal infection.
-History of urinary retention or current symptoms.

Warnings








The long-term effects of chronic electrical stimulation are unknown.
Safety has not been established for the use of the InWave during
pregnancy.
The device is capable of generating currents and voltages in excess of 10
mA RMS and 10 V RMS, respectively. A yellow LED indicator is
provided next to the lead wire connector to show that the unit is delivering
energy for any non-zero value of stimulation output (0.5 to 80 mA).
Discontinue use of this device in the event of pain or bleeding.
Do not use this device in water.
Do not insert the lead wires into a mains power supply.

17
InWave
Precautions and Adverse Reactions
Stress Incontinence Mode
Precautions













Stimulation settings should be based on the guidance of the prescribing
practitioner.
This device should be kept out of the reach of children.
This device should only be used with the leads and vaginal sensor
recommended for use by the manufacturer.
This device should not be used while driving, operating machinery, or
during any activity in which involuntary muscle contractions may put the
user at risk of injury.
Caution should be exercised when this device is used by patients with
pacemakers and heart rhythm problems.
Do not use this device while under the influence of drugs or alcohol.
Do not insert or remove the vaginal sensor while the device is ON.
Do not use this device while in the close proximity of wireless devices.
Use this device only as directed.
Do not immerse the unit in water or any other liquid.
The vaginal sensor should not remain inserted in the vagina when the unit
is not in use.
This device should not be used while sleeping or when anything else is in
the vagina, such as a diaphragm or tampon.

Adverse Reactions


Skin irritation and electrode burns are potential adverse reactions.

Safety References
Zynex Medical, Inc. is only responsible for the safety, reliability and function
of the device when repairs and adjustments have been made by persons
authorized by Zynex Medical, Inc., and the device is used in accordance with
the user’s manual. Repairs and technical safety tests shall only be performed
by authorized personnel.

18
InWave
Indications for Use
Urge Incontinence Mode
Please read the InWave User’s Manual before using the InWave pelvic
muscle stimulator.
CAUTION: Federal law restricts this device to sale by or on the order of a
licensed healthcare practitioner. This device should only be used under medical
supervision for adjunctive therapy for the treatment of medical diseases and
conditions.
Indications


The Zynex Medical InWave is intended to provide electrical stimulation
and neuromuscular re-education for the purpose of rehabilitation of weak
pelvic floor muscles for the treatment of stress, urge, and mixed urinary
incontinence in women.

Contraindications


This InWave should not be used in the presence of the following:
-Where there is a tendency to hemorrhage following acute trauma or
fracture.
-Following recent surgical procedures when muscle contraction may
disrupt the healing process.
-Over the menstruating or pregnant uterus.
-Over areas of the skin which lack normal sensation.
-Irregular menstrual cycles.
-Urinary or vaginal infection.
-History of urinary retention or current symptoms.

Warnings








The long-term effects of chronic electrical stimulation are unknown.
Safety has not been established for the use of the InWave during
pregnancy.
The device is capable of generating currents and voltages in excess of 10
mA RMS and 10 V RMS, respectively. A yellow LED indicator is
provided next to the lead wire connector to show that the unit is delivering
energy for any non-zero value of stimulation output (0.5 to 80 mA).
Discontinue use of this device in the event of pain or bleeding.
Do not use this device in water.
Do not insert the lead wires into a mains power supply.

19
InWave
Precautions and Adverse Reactions
Urge Incontinence Mode
Precautions













Stimulation settings should be based on the guidance of the prescribing
practitioner.
This device should be kept out of the reach of children.
This device should only be used with the leads and vaginal sensor
recommended for use by the manufacturer.
This device should not be used while driving, operating machinery, or
during any activity in which involuntary muscle contractions may put the
user at risk of injury.
Caution should be exercised when this device is used by patients with
pacemakers and heart rhythm problems.
Do not use this device while under the influence of drugs or alcohol.
Do not insert or remove the vaginal sensor while the device is ON.
Do not use this device while in the close proximity of wireless devices.
Use this device only as directed.
Do not immerse the unit in water or any other liquid.
The vaginal sensor should not remain inserted in the vagina when the unit
is not in use.
This device should not be used while sleeping or when anything else is in
the vagina, such as a diaphragm or tampon.

Adverse Reactions


Skin irritation and electrode burns are potential adverse reactions.

Safety References
Zynex Medical, Inc. is only responsible for the safety, reliability and function
of the device when repairs and adjustments have been made by persons
authorized by Zynex Medical, Inc., and the device is used in accordance with
the user’s manual. Repairs and technical safety tests shall only be performed
by authorized personnel.

20
InWave
Indications for Use
Mixed Incontinence Mode
Please read the InWave User’s Manual before using the InWave pelvic
muscle stimulator.
CAUTION: Federal law restricts this device to sale by or on the order of a
licensed healthcare practitioner. This device should only be used under medical
supervision for adjunctive therapy for the treatment of medical diseases and
conditions.
Indications


The Zynex Medical InWave is intended to provide electrical stimulation
and neuromuscular re-education for the purpose of rehabilitation of weak
pelvic floor muscles for the treatment of stress, urge, and mixed urinary
incontinence in women.

Contraindications


This InWave should not be used in the presence of the following:
-Where there is a tendency to hemorrhage following acute trauma or
fracture.
-Following recent surgical procedures when muscle contraction may
disrupt the healing process.
-Over the menstruating or pregnant uterus.
-Over areas of the skin which lack normal sensation.
-Irregular menstrual cycles.
-Urinary or vaginal infection.
-History of urinary retention or current symptoms.

Warnings








The long-term effects of chronic electrical stimulation are unknown.
Safety has not been established for the use of the InWave during
pregnancy.
The device is capable of generating currents and voltages in excess of 10
mA RMS and 10 V RMS, respectively. A yellow LED indicator is
provided next to the lead wire connector to show that the unit is delivering
energy for any non-zero value of stimulation output (0.5 to 80 mA).
Discontinue use of this device in the event of pain or bleeding.
Do not use this device in water.
Do not insert the lead wires into a mains power supply.

21
InWave
Precautions and Adverse Reactions
Mixed Incontinence Mode
Precautions













Stimulation settings should be based on the guidance of the prescribing
practitioner.
This device should be kept out of the reach of children.
This device should only be used with the leads and vaginal sensor
recommended for use by the manufacturer.
This device should not be used while driving, operating machinery, or
during any activity in which involuntary muscle contractions may put the
user at risk of injury.
Caution should be exercised when this device is used by patients with
pacemakers and heart rhythm problems.
Do not use this device while under the influence of drugs or alcohol.
Do not insert or remove the vaginal sensor while the device is ON.
Do not use this device while in the close proximity of wireless devices.
Use this device only as directed.
Do not immerse the unit in water or any other liquid.
The vaginal sensor should not remain inserted in the vagina when the unit
is not in use.
This device should not be used while sleeping or when anything else is in
the vagina, such as a diaphragm or tampon.

Adverse Reactions


Skin irritation and electrode burns are potential adverse reactions.

Safety References
Zynex Medical, Inc. is only responsible for the safety, reliability and function
of the device when repairs and adjustments have been made by persons
authorized by Zynex Medical, Inc., and the device is used in accordance with
the user’s manual. Repairs and technical safety tests shall only be performed
by authorized personnel.

22
Troubleshooting
Problem

Solution

Unit stays on even
after treatment ends.

Hold Off button down for 2 seconds to shut unit off.
Otherwise unit will shut off automatically after 5
minutes of no stimulation. Alternatively you can start
a new treatment session after the last one ends.

Cannot increase level
from its current setting.

Press and release Unlock button to unlock this safety
feature. Then increase stimulation to the desired level
of intensity. Intensity level is always locked after 20
seconds of key press inactivity.

Display shows “Check Check to see that the sensor is properly inserted.
Connections”
(Remove and re-insert the sensor if necessary to
(Electrode alarm).
establish a better electrical contact.) Verify that the
sensor is correctly connected to the lead wire and that
the lead wire is correctly connected to the InWave.
Connect the bare metal pins of the lead wire together
to short-circuit the output. If the electrode alarm ceases
the problem is with the sensor connection integrity.
Removing and re-inserting the sensor may establish a
better electrical contact.

23
InWave
Specifications and Accessories
Stress Incontinence Mode
Amplitude:
Frequency:
Pulse Width:
Stress Modes:
On Time:
Off Time:
Ramp Up Time:
Ramp Down Time:
Waveforms:

0-80 mA
50 Hz
200 µseconds
5/10, 3/10, 5/5
3 or 5 seconds
5 or 10 seconds
1 second
1 second
Symmetrical biphasic

Urge Incontinence Mode
Amplitude:
Frequency:
Pulse Width:
Urge Modes:
On Time:
Off Time:
Ramp Up Time:
Ramp Down Time:
Waveform:

0-80 mA
12.5 Hz
600 µseconds
10/3, 6/3, 10/10
6 or 10 seconds
3 or 10 seconds
1 second
0.5 seconds
Symmetrical biphasic

Mixed Incontinence Mode
Amplitude:
Frequency:
Pulse Width:
Mix Modes:
On-Time:
Off-Time:
Ramp Up:
Ramp Down:
Waveforms:

0-80 mA
12.5 Hz and 50 Hz
200 µseconds and 600 µseconds
10:10, 5:5, 2:2
3, 5, 6, or 10 seconds
3, 5, or 10 seconds
1 second
0.5 or 1 second
Symmetrical biphasic

Other Specifications
Treatment timer:

5 to 25 minutes, in 5 minute increments.

Compliance meter:

Records total usage time in minutes and number of times
used.

Dimensions:

2.9 x 4.6 x 1.0 in.

Weight:

5.8 oz. including battery

Warranty:

5 year manufacturer’s warranty on materials and
workmanship. Accessories excluded.

Accessories
Lead wire:

104204, lead wire, gray.

Sensor:

400015, vaginal sensor.

Battery:

130010, battery, 9 Volt, alkaline, MN1604.

24
InWave
Sensor, Batteries, and Cleaning
Sensor Use
The InWave is intended to be used with a re-usable vaginal sensor. Before and
after each use, clean the sensor with mild soap and water, and dry the sensor.
Store the sensor in a safe location when not in use.
Follow directions of the prescribing physician.
Note: The vaginal sensors are for single patient use only.
Battery Power
One 9 Volt alkaline battery is used to power the InWave. The battery
compartment on the back of the device opens by sliding the cover downwards.
Rechargeable batteries are not recommended since they have only a short
usage time and cannot be charged while in the InWave.
Battery replacement is indicated by the battery icon in the display. Replace
with a new 9 Volt alkaline battery.
The InWave operates normally even if the 9 Volt battery is inserted with
reverse polarity.
Please dispose of used batteries properly.
Device Cleaning
The InWave case and display window can be cleaned by lightly applying mild
soap to a damp soft cloth or paper towel and using it to wipe the surfaces clean.
Allow the unit to dry thoroughly before using. Do not spray cleaning solutions
directly onto the unit, or immerse it in water or other liquids.

25
InWave
Warranty Information
The InWave device is warranted to be free from defects in material,
workmanship, and structural integrity when subjected to normal domestic use
and service for five years after the original purchase.
During that time, Zynex Medical, Inc will repair or replace, at its sole
discretion, the InWave device that has been used in a standard manner. This
warranty does not cover misuse or use contrary to the operating instructions
supplied.
Warranty obligations are limited to replacement or repair of defective parts and
components, at the option of Zynex Medical, Inc. This Warranty is only valid
for the original purchaser of this product and will not be honored if the product
ownership is transferred or is resold.
To obtain warranty service, please contact Technical Support at the number
listed on page 4 of this manual.

26

More Related Content

What's hot

ZMPCZM023000.12.01 User's manual
ZMPCZM023000.12.01 User's  manualZMPCZM023000.12.01 User's  manual
ZMPCZM023000.12.01 User's manualPainezee Specialist
 
Zmpcrm073000.12.01 User's Manual
Zmpcrm073000.12.01 User's  ManualZmpcrm073000.12.01 User's  Manual
Zmpcrm073000.12.01 User's ManualPainezee Specialist
 
ShuntCheck Operating Manual
ShuntCheck Operating ManualShuntCheck Operating Manual
ShuntCheck Operating Manualneurodx
 
ZMPCZM016000.10.02 Nexwave Training for Hyderabad, India.
ZMPCZM016000.10.02 Nexwave Training for Hyderabad, India.ZMPCZM016000.10.02 Nexwave Training for Hyderabad, India.
ZMPCZM016000.10.02 Nexwave Training for Hyderabad, India.Painezee Specialist
 
IRJET- Intravenous Fluid Level Indicator
IRJET- Intravenous Fluid Level IndicatorIRJET- Intravenous Fluid Level Indicator
IRJET- Intravenous Fluid Level IndicatorIRJET Journal
 
Neuro-Stim System Neurostim FDA K140530 Electro Acupuncture Device
Neuro-Stim System Neurostim FDA K140530 Electro Acupuncture DeviceNeuro-Stim System Neurostim FDA K140530 Electro Acupuncture Device
Neuro-Stim System Neurostim FDA K140530 Electro Acupuncture Devicemcalipo
 
Hướng dẫn sử dụng Testo 770
Hướng dẫn sử dụng Testo 770Hướng dẫn sử dụng Testo 770
Hướng dẫn sử dụng Testo 770Tenmars Việt Nam
 
Y1299-90001_StartupGuide
Y1299-90001_StartupGuideY1299-90001_StartupGuide
Y1299-90001_StartupGuideChris Muntzer
 
Implantable Cardiac Devices
Implantable Cardiac Devices Implantable Cardiac Devices
Implantable Cardiac Devices Mohamed Elwakil
 
Multiquip Rebar Cutters
Multiquip Rebar CuttersMultiquip Rebar Cutters
Multiquip Rebar Cuttersmultiquipuk
 
Ac drive altivar 12 user manual
Ac drive altivar 12 user manualAc drive altivar 12 user manual
Ac drive altivar 12 user manualToàn Huỳnh
 

What's hot (15)

ZMPCZM023000.12.01 User's manual
ZMPCZM023000.12.01 User's  manualZMPCZM023000.12.01 User's  manual
ZMPCZM023000.12.01 User's manual
 
U270
U270U270
U270
 
Zmpcrm073000.12.01 User's Manual
Zmpcrm073000.12.01 User's  ManualZmpcrm073000.12.01 User's  Manual
Zmpcrm073000.12.01 User's Manual
 
ShuntCheck Operating Manual
ShuntCheck Operating ManualShuntCheck Operating Manual
ShuntCheck Operating Manual
 
ZMPCZM016000.10.02 Nexwave Training for Hyderabad, India.
ZMPCZM016000.10.02 Nexwave Training for Hyderabad, India.ZMPCZM016000.10.02 Nexwave Training for Hyderabad, India.
ZMPCZM016000.10.02 Nexwave Training for Hyderabad, India.
 
IRJET- Intravenous Fluid Level Indicator
IRJET- Intravenous Fluid Level IndicatorIRJET- Intravenous Fluid Level Indicator
IRJET- Intravenous Fluid Level Indicator
 
Hantek6022 be manual english(v1.0.3)
Hantek6022 be manual english(v1.0.3)Hantek6022 be manual english(v1.0.3)
Hantek6022 be manual english(v1.0.3)
 
La estimulación sin cables: resultados a medio plazo
La estimulación sin cables: resultados a medio plazoLa estimulación sin cables: resultados a medio plazo
La estimulación sin cables: resultados a medio plazo
 
Neuro-Stim System Neurostim FDA K140530 Electro Acupuncture Device
Neuro-Stim System Neurostim FDA K140530 Electro Acupuncture DeviceNeuro-Stim System Neurostim FDA K140530 Electro Acupuncture Device
Neuro-Stim System Neurostim FDA K140530 Electro Acupuncture Device
 
Carelink 2090
Carelink 2090Carelink 2090
Carelink 2090
 
Hướng dẫn sử dụng Testo 770
Hướng dẫn sử dụng Testo 770Hướng dẫn sử dụng Testo 770
Hướng dẫn sử dụng Testo 770
 
Y1299-90001_StartupGuide
Y1299-90001_StartupGuideY1299-90001_StartupGuide
Y1299-90001_StartupGuide
 
Implantable Cardiac Devices
Implantable Cardiac Devices Implantable Cardiac Devices
Implantable Cardiac Devices
 
Multiquip Rebar Cutters
Multiquip Rebar CuttersMultiquip Rebar Cutters
Multiquip Rebar Cutters
 
Ac drive altivar 12 user manual
Ac drive altivar 12 user manualAc drive altivar 12 user manual
Ac drive altivar 12 user manual
 

Viewers also liked

Zmpcrm072000.12.01 User's manual
Zmpcrm072000.12.01 User's  manualZmpcrm072000.12.01 User's  manual
Zmpcrm072000.12.01 User's manualPainezee Specialist
 
Zmpczm022000.12.03 Operating instructions
Zmpczm022000.12.03 Operating instructionsZmpczm022000.12.03 Operating instructions
Zmpczm022000.12.03 Operating instructionsPainezee Specialist
 
Zmpcrm076000.12.01 Instruction manual
Zmpcrm076000.12.01 Instruction manualZmpcrm076000.12.01 Instruction manual
Zmpcrm076000.12.01 Instruction manualPainezee Specialist
 
ZMPCHW070000.13.04 US FDA approval letter K 120398
ZMPCHW070000.13.04 US FDA approval letter  K 120398ZMPCHW070000.13.04 US FDA approval letter  K 120398
ZMPCHW070000.13.04 US FDA approval letter K 120398Painezee Specialist
 
ZMPCML041000.11.01 Clinical Evaluations and medical literature review.
ZMPCML041000.11.01 Clinical Evaluations and medical literature review.ZMPCML041000.11.01 Clinical Evaluations and medical literature review.
ZMPCML041000.11.01 Clinical Evaluations and medical literature review.Painezee Specialist
 
ZMPCZM016000.10.02 Nexwave Training for Hyderabad, India.
ZMPCZM016000.10.02 Nexwave Training for Hyderabad, India.ZMPCZM016000.10.02 Nexwave Training for Hyderabad, India.
ZMPCZM016000.10.02 Nexwave Training for Hyderabad, India.Painezee Specialist
 
ZMPCZM017000.11.03 Carey Experimentation on brain research
ZMPCZM017000.11.03 Carey Experimentation on brain researchZMPCZM017000.11.03 Carey Experimentation on brain research
ZMPCZM017000.11.03 Carey Experimentation on brain researchPainezee Specialist
 

Viewers also liked (13)

Zmpcrm072000.12.01 User's manual
Zmpcrm072000.12.01 User's  manualZmpcrm072000.12.01 User's  manual
Zmpcrm072000.12.01 User's manual
 
Zmpczm022000.12.03 Operating instructions
Zmpczm022000.12.03 Operating instructionsZmpczm022000.12.03 Operating instructions
Zmpczm022000.12.03 Operating instructions
 
ZMPCZM017000.12.02
ZMPCZM017000.12.02ZMPCZM017000.12.02
ZMPCZM017000.12.02
 
ZMPCZM017000.11.02
ZMPCZM017000.11.02ZMPCZM017000.11.02
ZMPCZM017000.11.02
 
ZMPCZM017000.11.02
ZMPCZM017000.11.02 ZMPCZM017000.11.02
ZMPCZM017000.11.02
 
ZMPCZM017000.11.03
ZMPCZM017000.11.03ZMPCZM017000.11.03
ZMPCZM017000.11.03
 
ZMPCRM073000.12.01 users Manual
ZMPCRM073000.12.01 users ManualZMPCRM073000.12.01 users Manual
ZMPCRM073000.12.01 users Manual
 
Zmpcrm076000.12.01 Instruction manual
Zmpcrm076000.12.01 Instruction manualZmpcrm076000.12.01 Instruction manual
Zmpcrm076000.12.01 Instruction manual
 
ZMPCZM019000.10.01
ZMPCZM019000.10.01ZMPCZM019000.10.01
ZMPCZM019000.10.01
 
ZMPCHW070000.13.04 US FDA approval letter K 120398
ZMPCHW070000.13.04 US FDA approval letter  K 120398ZMPCHW070000.13.04 US FDA approval letter  K 120398
ZMPCHW070000.13.04 US FDA approval letter K 120398
 
ZMPCML041000.11.01 Clinical Evaluations and medical literature review.
ZMPCML041000.11.01 Clinical Evaluations and medical literature review.ZMPCML041000.11.01 Clinical Evaluations and medical literature review.
ZMPCML041000.11.01 Clinical Evaluations and medical literature review.
 
ZMPCZM016000.10.02 Nexwave Training for Hyderabad, India.
ZMPCZM016000.10.02 Nexwave Training for Hyderabad, India.ZMPCZM016000.10.02 Nexwave Training for Hyderabad, India.
ZMPCZM016000.10.02 Nexwave Training for Hyderabad, India.
 
ZMPCZM017000.11.03 Carey Experimentation on brain research
ZMPCZM017000.11.03 Carey Experimentation on brain researchZMPCZM017000.11.03 Carey Experimentation on brain research
ZMPCZM017000.11.03 Carey Experimentation on brain research
 

Similar to ZMPCZM018000.12 User's manual

Hướng dẫn sử dụng Testo 435
Hướng dẫn sử dụng Testo 435Hướng dẫn sử dụng Testo 435
Hướng dẫn sử dụng Testo 435Tenmars Việt Nam
 
Hướng dẫn sử dụng máy đo nhiệt độ bề mặt Testo 435-3
Hướng dẫn sử dụng máy đo nhiệt độ bề mặt Testo 435-3Hướng dẫn sử dụng máy đo nhiệt độ bề mặt Testo 435-3
Hướng dẫn sử dụng máy đo nhiệt độ bề mặt Testo 435-3Tenmars Việt Nam
 
Hướng dẫn sử dụng Testo 435
Hướng dẫn sử dụng Testo 435Hướng dẫn sử dụng Testo 435
Hướng dẫn sử dụng Testo 435Tenmars Việt Nam
 
Hướng dẫn sử dụng Testo 435-3
Hướng dẫn sử dụng Testo 435-3Hướng dẫn sử dụng Testo 435-3
Hướng dẫn sử dụng Testo 435-3Tenmars Việt Nam
 
Hướng dẫn sử dụng testo 435-3
Hướng dẫn sử dụng  testo 435-3 Hướng dẫn sử dụng  testo 435-3
Hướng dẫn sử dụng testo 435-3 Tenmars Việt Nam
 
Hướng dẫn sử dụng máy đo đa năng Testo 435
Hướng dẫn sử dụng máy đo đa năng Testo 435Hướng dẫn sử dụng máy đo đa năng Testo 435
Hướng dẫn sử dụng máy đo đa năng Testo 435Tenmars Việt Nam
 
ZMPCRM073000.12.01 User's Manual
ZMPCRM073000.12.01 User's ManualZMPCRM073000.12.01 User's Manual
ZMPCRM073000.12.01 User's ManualPainezee Specialist
 
Zmpczm023000.12.01 User's manual
Zmpczm023000.12.01 User's  manualZmpczm023000.12.01 User's  manual
Zmpczm023000.12.01 User's manualPainezee Specialist
 
Schwinn 430 elliptical trainer User Manual
Schwinn 430 elliptical trainer User ManualSchwinn 430 elliptical trainer User Manual
Schwinn 430 elliptical trainer User ManualEllipticalHUB
 
SCOTTBUCHHEIT 510(k) PAPER
SCOTTBUCHHEIT 510(k) PAPERSCOTTBUCHHEIT 510(k) PAPER
SCOTTBUCHHEIT 510(k) PAPERScott Buchheit
 

Similar to ZMPCZM018000.12 User's manual (20)

Hướng dẫn sử dụng Testo 435
Hướng dẫn sử dụng Testo 435Hướng dẫn sử dụng Testo 435
Hướng dẫn sử dụng Testo 435
 
Hướng dẫn sử dụng máy đo nhiệt độ bề mặt Testo 435-3
Hướng dẫn sử dụng máy đo nhiệt độ bề mặt Testo 435-3Hướng dẫn sử dụng máy đo nhiệt độ bề mặt Testo 435-3
Hướng dẫn sử dụng máy đo nhiệt độ bề mặt Testo 435-3
 
Hướng dẫn sử dụng Testo 435
Hướng dẫn sử dụng Testo 435Hướng dẫn sử dụng Testo 435
Hướng dẫn sử dụng Testo 435
 
Hướng dẫn sử dụng Testo 435-3
Hướng dẫn sử dụng Testo 435-3Hướng dẫn sử dụng Testo 435-3
Hướng dẫn sử dụng Testo 435-3
 
Hướng dẫn sử dụng testo 435-3
Hướng dẫn sử dụng  testo 435-3 Hướng dẫn sử dụng  testo 435-3
Hướng dẫn sử dụng testo 435-3
 
Hướng dẫn sử dụng máy đo đa năng Testo 435
Hướng dẫn sử dụng máy đo đa năng Testo 435Hướng dẫn sử dụng máy đo đa năng Testo 435
Hướng dẫn sử dụng máy đo đa năng Testo 435
 
ZMPCRM077000.12 User's manual
ZMPCRM077000.12 User's  manualZMPCRM077000.12 User's  manual
ZMPCRM077000.12 User's manual
 
Zmpcrm077000.12 User's manual
Zmpcrm077000.12 User's  manualZmpcrm077000.12 User's  manual
Zmpcrm077000.12 User's manual
 
ZMPCRM077000.12 User's Manual
ZMPCRM077000.12 User's ManualZMPCRM077000.12 User's Manual
ZMPCRM077000.12 User's Manual
 
ZMPCRM073000.12.01 User's Manual
ZMPCRM073000.12.01 User's ManualZMPCRM073000.12.01 User's Manual
ZMPCRM073000.12.01 User's Manual
 
ZMPCRM073000.12.01 Users Manual
ZMPCRM073000.12.01 Users ManualZMPCRM073000.12.01 Users Manual
ZMPCRM073000.12.01 Users Manual
 
ZMPCRM073000.12.01 users manual
ZMPCRM073000.12.01 users manualZMPCRM073000.12.01 users manual
ZMPCRM073000.12.01 users manual
 
ZMPCRM073000.12.01 Users manual
ZMPCRM073000.12.01 Users manualZMPCRM073000.12.01 Users manual
ZMPCRM073000.12.01 Users manual
 
ZMPCRM073000.12.01 Users Manual
ZMPCRM073000.12.01 Users  ManualZMPCRM073000.12.01 Users  Manual
ZMPCRM073000.12.01 Users Manual
 
Zmpczm023000.12.01 User's manual
Zmpczm023000.12.01 User's  manualZmpczm023000.12.01 User's  manual
Zmpczm023000.12.01 User's manual
 
User Instruction Manual
User Instruction ManualUser Instruction Manual
User Instruction Manual
 
Defibrillator ppt.pptx
Defibrillator ppt.pptxDefibrillator ppt.pptx
Defibrillator ppt.pptx
 
Schwinn 430 elliptical trainer User Manual
Schwinn 430 elliptical trainer User ManualSchwinn 430 elliptical trainer User Manual
Schwinn 430 elliptical trainer User Manual
 
Operation manual m503_e027a
Operation manual m503_e027aOperation manual m503_e027a
Operation manual m503_e027a
 
SCOTTBUCHHEIT 510(k) PAPER
SCOTTBUCHHEIT 510(k) PAPERSCOTTBUCHHEIT 510(k) PAPER
SCOTTBUCHHEIT 510(k) PAPER
 

More from Painezee Specialist

ZMGYPP100000 Pressure right design_specifications
ZMGYPP100000 Pressure right design_specificationsZMGYPP100000 Pressure right design_specifications
ZMGYPP100000 Pressure right design_specificationsPainezee Specialist
 
ZMGYPP100000 Pressure right product_instructions
ZMGYPP100000 Pressure right product_instructionsZMGYPP100000 Pressure right product_instructions
ZMGYPP100000 Pressure right product_instructionsPainezee Specialist
 
Neuromove – pathway to recovery
Neuromove – pathway to recoveryNeuromove – pathway to recovery
Neuromove – pathway to recoveryPainezee Specialist
 
Zmpczm0170001003 ZMPCZM017000.10.03 Neuromove clinical presentation from Pain...
Zmpczm0170001003 ZMPCZM017000.10.03 Neuromove clinical presentation from Pain...Zmpczm0170001003 ZMPCZM017000.10.03 Neuromove clinical presentation from Pain...
Zmpczm0170001003 ZMPCZM017000.10.03 Neuromove clinical presentation from Pain...Painezee Specialist
 
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzee
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzeeZMPCZM017000.10.03 Neuromove clinical presentation from PainEzee
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzeePainezee Specialist
 
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzee
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzeeZMPCZM017000.10.03 Neuromove clinical presentation from PainEzee
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzeePainezee Specialist
 
ZMPCZM016000.13.04 EC Declaration of Conformity NexWave
ZMPCZM016000.13.04 EC Declaration of Conformity NexWaveZMPCZM016000.13.04 EC Declaration of Conformity NexWave
ZMPCZM016000.13.04 EC Declaration of Conformity NexWavePainezee Specialist
 
ZMPCZM017000.12.03 Neuromove Electrode placement chart
ZMPCZM017000.12.03 Neuromove Electrode placement chartZMPCZM017000.12.03 Neuromove Electrode placement chart
ZMPCZM017000.12.03 Neuromove Electrode placement chartPainezee Specialist
 
ZMPCZM019000.12.07 NeuroGraph User's manual
ZMPCZM019000.12.07 NeuroGraph User's manualZMPCZM019000.12.07 NeuroGraph User's manual
ZMPCZM019000.12.07 NeuroGraph User's manualPainezee Specialist
 
ZMPCZM019000.12.06 NeuroGames User's manual
ZMPCZM019000.12.06 NeuroGames User's manualZMPCZM019000.12.06 NeuroGames User's manual
ZMPCZM019000.12.06 NeuroGames User's manualPainezee Specialist
 
ZMPCZM019000.12.05 Incontinence user's manual
ZMPCZM019000.12.05 Incontinence user's manualZMPCZM019000.12.05 Incontinence user's manual
ZMPCZM019000.12.05 Incontinence user's manualPainezee Specialist
 
ZMPCZM019000.12.04 Head Ache User's Manual
ZMPCZM019000.12.04 Head Ache User's ManualZMPCZM019000.12.04 Head Ache User's Manual
ZMPCZM019000.12.04 Head Ache User's ManualPainezee Specialist
 
ZMPCZM019000.12.03 Chiro Scan User's guide
ZMPCZM019000.12.03 Chiro Scan User's guideZMPCZM019000.12.03 Chiro Scan User's guide
ZMPCZM019000.12.03 Chiro Scan User's guidePainezee Specialist
 
ZMPCZM016000.12.16 Electrode placement & Protocol Guidelines
ZMPCZM016000.12.16 Electrode placement & Protocol GuidelinesZMPCZM016000.12.16 Electrode placement & Protocol Guidelines
ZMPCZM016000.12.16 Electrode placement & Protocol GuidelinesPainezee Specialist
 
ZMPCZM016000.12.14 Hand,Wrist,Elbow & upperarm electrode placement
ZMPCZM016000.12.14 Hand,Wrist,Elbow & upperarm electrode placementZMPCZM016000.12.14 Hand,Wrist,Elbow & upperarm electrode placement
ZMPCZM016000.12.14 Hand,Wrist,Elbow & upperarm electrode placementPainezee Specialist
 
ZMPCZM016000.12.14 Hand,Wrist,Elbow & upperarm electrode placement
ZMPCZM016000.12.14 Hand,Wrist,Elbow & upperarm electrode placementZMPCZM016000.12.14 Hand,Wrist,Elbow & upperarm electrode placement
ZMPCZM016000.12.14 Hand,Wrist,Elbow & upperarm electrode placementPainezee Specialist
 

More from Painezee Specialist (20)

ZMGYPP100000 Pressure right design_specifications
ZMGYPP100000 Pressure right design_specificationsZMGYPP100000 Pressure right design_specifications
ZMGYPP100000 Pressure right design_specifications
 
ZMGYPP100000 Pressure right product_instructions
ZMGYPP100000 Pressure right product_instructionsZMGYPP100000 Pressure right product_instructions
ZMGYPP100000 Pressure right product_instructions
 
Neuromove – pathway to recovery
Neuromove – pathway to recoveryNeuromove – pathway to recovery
Neuromove – pathway to recovery
 
Zmpczm0170001003 ZMPCZM017000.10.03 Neuromove clinical presentation from Pain...
Zmpczm0170001003 ZMPCZM017000.10.03 Neuromove clinical presentation from Pain...Zmpczm0170001003 ZMPCZM017000.10.03 Neuromove clinical presentation from Pain...
Zmpczm0170001003 ZMPCZM017000.10.03 Neuromove clinical presentation from Pain...
 
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzee
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzeeZMPCZM017000.10.03 Neuromove clinical presentation from PainEzee
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzee
 
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzee
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzeeZMPCZM017000.10.03 Neuromove clinical presentation from PainEzee
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzee
 
ZMPCCM111000.03
ZMPCCM111000.03 ZMPCCM111000.03
ZMPCCM111000.03
 
ZMPCZM016000.13.04 EC Declaration of Conformity NexWave
ZMPCZM016000.13.04 EC Declaration of Conformity NexWaveZMPCZM016000.13.04 EC Declaration of Conformity NexWave
ZMPCZM016000.13.04 EC Declaration of Conformity NexWave
 
ZMPCZM017000.12.03 Neuromove Electrode placement chart
ZMPCZM017000.12.03 Neuromove Electrode placement chartZMPCZM017000.12.03 Neuromove Electrode placement chart
ZMPCZM017000.12.03 Neuromove Electrode placement chart
 
Zywie
ZywieZywie
Zywie
 
ZMPCZM016000.12.07
ZMPCZM016000.12.07ZMPCZM016000.12.07
ZMPCZM016000.12.07
 
ZMPCZM016000.12.06
ZMPCZM016000.12.06ZMPCZM016000.12.06
ZMPCZM016000.12.06
 
ZMPCZM019000.12.07 NeuroGraph User's manual
ZMPCZM019000.12.07 NeuroGraph User's manualZMPCZM019000.12.07 NeuroGraph User's manual
ZMPCZM019000.12.07 NeuroGraph User's manual
 
ZMPCZM019000.12.06 NeuroGames User's manual
ZMPCZM019000.12.06 NeuroGames User's manualZMPCZM019000.12.06 NeuroGames User's manual
ZMPCZM019000.12.06 NeuroGames User's manual
 
ZMPCZM019000.12.05 Incontinence user's manual
ZMPCZM019000.12.05 Incontinence user's manualZMPCZM019000.12.05 Incontinence user's manual
ZMPCZM019000.12.05 Incontinence user's manual
 
ZMPCZM019000.12.04 Head Ache User's Manual
ZMPCZM019000.12.04 Head Ache User's ManualZMPCZM019000.12.04 Head Ache User's Manual
ZMPCZM019000.12.04 Head Ache User's Manual
 
ZMPCZM019000.12.03 Chiro Scan User's guide
ZMPCZM019000.12.03 Chiro Scan User's guideZMPCZM019000.12.03 Chiro Scan User's guide
ZMPCZM019000.12.03 Chiro Scan User's guide
 
ZMPCZM016000.12.16 Electrode placement & Protocol Guidelines
ZMPCZM016000.12.16 Electrode placement & Protocol GuidelinesZMPCZM016000.12.16 Electrode placement & Protocol Guidelines
ZMPCZM016000.12.16 Electrode placement & Protocol Guidelines
 
ZMPCZM016000.12.14 Hand,Wrist,Elbow & upperarm electrode placement
ZMPCZM016000.12.14 Hand,Wrist,Elbow & upperarm electrode placementZMPCZM016000.12.14 Hand,Wrist,Elbow & upperarm electrode placement
ZMPCZM016000.12.14 Hand,Wrist,Elbow & upperarm electrode placement
 
ZMPCZM016000.12.14 Hand,Wrist,Elbow & upperarm electrode placement
ZMPCZM016000.12.14 Hand,Wrist,Elbow & upperarm electrode placementZMPCZM016000.12.14 Hand,Wrist,Elbow & upperarm electrode placement
ZMPCZM016000.12.14 Hand,Wrist,Elbow & upperarm electrode placement
 

Recently uploaded

Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Miss joya
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escortsaditipandeya
 
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...narwatsonia7
 
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment BookingHousewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Bookingnarwatsonia7
 
Low Rate Call Girls Ambattur Anika 8250192130 Independent Escort Service Amba...
Low Rate Call Girls Ambattur Anika 8250192130 Independent Escort Service Amba...Low Rate Call Girls Ambattur Anika 8250192130 Independent Escort Service Amba...
Low Rate Call Girls Ambattur Anika 8250192130 Independent Escort Service Amba...narwatsonia7
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalorenarwatsonia7
 
Call Girl Chennai Indira 9907093804 Independent Call Girls Service Chennai
Call Girl Chennai Indira 9907093804 Independent Call Girls Service ChennaiCall Girl Chennai Indira 9907093804 Independent Call Girls Service Chennai
Call Girl Chennai Indira 9907093804 Independent Call Girls Service ChennaiNehru place Escorts
 
Call Girls Doddaballapur Road Just Call 7001305949 Top Class Call Girl Servic...
Call Girls Doddaballapur Road Just Call 7001305949 Top Class Call Girl Servic...Call Girls Doddaballapur Road Just Call 7001305949 Top Class Call Girl Servic...
Call Girls Doddaballapur Road Just Call 7001305949 Top Class Call Girl Servic...narwatsonia7
 
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...CALL GIRLS
 
Call Girls Chennai Megha 9907093804 Independent Call Girls Service Chennai
Call Girls Chennai Megha 9907093804 Independent Call Girls Service ChennaiCall Girls Chennai Megha 9907093804 Independent Call Girls Service Chennai
Call Girls Chennai Megha 9907093804 Independent Call Girls Service ChennaiNehru place Escorts
 
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Miss joya
 
Call Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service NoidaCall Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service NoidaPooja Gupta
 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escortsvidya singh
 
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...narwatsonia7
 
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near MeHi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Menarwatsonia7
 
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.MiadAlsulami
 
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableVip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableNehru place Escorts
 
Call Girls Yelahanka Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Yelahanka Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Yelahanka Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Yelahanka Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Russian Call Girls in Chennai Pallavi 9907093804 Independent Call Girls Servi...
Russian Call Girls in Chennai Pallavi 9907093804 Independent Call Girls Servi...Russian Call Girls in Chennai Pallavi 9907093804 Independent Call Girls Servi...
Russian Call Girls in Chennai Pallavi 9907093804 Independent Call Girls Servi...Nehru place Escorts
 
Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...
Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...
Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...narwatsonia7
 

Recently uploaded (20)

Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
 
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
 
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment BookingHousewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
Housewife Call Girls Hoskote | 7001305949 At Low Cost Cash Payment Booking
 
Low Rate Call Girls Ambattur Anika 8250192130 Independent Escort Service Amba...
Low Rate Call Girls Ambattur Anika 8250192130 Independent Escort Service Amba...Low Rate Call Girls Ambattur Anika 8250192130 Independent Escort Service Amba...
Low Rate Call Girls Ambattur Anika 8250192130 Independent Escort Service Amba...
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
 
Call Girl Chennai Indira 9907093804 Independent Call Girls Service Chennai
Call Girl Chennai Indira 9907093804 Independent Call Girls Service ChennaiCall Girl Chennai Indira 9907093804 Independent Call Girls Service Chennai
Call Girl Chennai Indira 9907093804 Independent Call Girls Service Chennai
 
Call Girls Doddaballapur Road Just Call 7001305949 Top Class Call Girl Servic...
Call Girls Doddaballapur Road Just Call 7001305949 Top Class Call Girl Servic...Call Girls Doddaballapur Road Just Call 7001305949 Top Class Call Girl Servic...
Call Girls Doddaballapur Road Just Call 7001305949 Top Class Call Girl Servic...
 
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
 
Call Girls Chennai Megha 9907093804 Independent Call Girls Service Chennai
Call Girls Chennai Megha 9907093804 Independent Call Girls Service ChennaiCall Girls Chennai Megha 9907093804 Independent Call Girls Service Chennai
Call Girls Chennai Megha 9907093804 Independent Call Girls Service Chennai
 
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
 
Call Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service NoidaCall Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service Noida
 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
 
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
 
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near MeHi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
Hi,Fi Call Girl In Mysore Road - 7001305949 | 24x7 Service Available Near Me
 
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
 
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableVip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
 
Call Girls Yelahanka Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Yelahanka Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Yelahanka Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Yelahanka Just Call 7001305949 Top Class Call Girl Service Available
 
Russian Call Girls in Chennai Pallavi 9907093804 Independent Call Girls Servi...
Russian Call Girls in Chennai Pallavi 9907093804 Independent Call Girls Servi...Russian Call Girls in Chennai Pallavi 9907093804 Independent Call Girls Servi...
Russian Call Girls in Chennai Pallavi 9907093804 Independent Call Girls Servi...
 
Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...
Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...
Russian Call Girls in Bangalore Manisha 7001305949 Independent Escort Service...
 

ZMPCZM018000.12 User's manual

  • 2. InWave Table of Contents Contact Information / Customer Service / Supplies / Support ........................ 4 About the InWave ........................................................................................... 5 Safety Information .......................................................................................... 6 Sensor and Lead Wire Setup ........................................................................... 9 Programming Instructions ............................................................................... 12 Preprogrammed Modes ................................................................................... 13 Controls........................................................................................................... 15 Operating Instructions..................................................................................... 16 Indications, Contraindications and Warnings………………………..17, 19, 21 Precautions and Adverse Reactions…………………………………..18, 20, 22 Troubleshooting .............................................................................................. 23 Specifications and Accessories ....................................................................... 24 Sensor, Batteries, and Cleaning ...................................................................... 25 Warranty ......................................................................................................... 26 3
  • 3. Zynex Medical Painezee Contact Information 1-866-940-7030 0091-9030022402 CUSTOMER SERVICE Supplies: To order more electrodes or other accessories Technical Support: Questions or problems with using your device Device Return: Order a postage paid return envelope to return your device at no charge MAIN OFFICE Billing Questions: 1-800-495-6670 0091-9030022402 Questions regarding insurance benefits and covered benefits for Durable Medical Equipment (DME) or questions about an Explanation of Benefits form you received in the mail FAX NUMBER 1-800-495-6695 MAILING ADDRESS Zynex Medical, Inc. Plot No: 6, Green Avenues, 9990 Park Meadows DriveNizampet Road, Opp: Bhavysa Anandham, Kukatpally, Hyderabad - 500090 Lone Tree, CO 80124-6739 Andhra USA Pradesh, India info@zynexmed.com info@painezee.com EMAIL WEBSITE www.zynexmed.com www.painezee.com 4
  • 4. About the InWave The InWave is a multiple-mode stimulator which allows users a choice of treatment options. The InWave causes muscle contractions which exercise and strengthen the pelvic floor muscles. Stress Incontinence Treatment The output delivers electrical impulses at a frequency of 50 Hz, which produce smooth, even muscle contractions. This type of stimulation exercises and strengthens the pelvic floor muscles, which aids in the control of stress incontinence. Stimulation can be set to 3 or 5 second duration, with a resting period of 5 or 10 seconds between contractions. Default treatment time is 15 minutes. Urge Incontinence Treatment The output delivers electrical impulses at a frequency of 12.5 Hz, which produce a series of short, rapid contractions. This type of stimulation aids in the relaxation of involuntary muscle contractions which cause urge incontinence. Stimulation can be set to 6 or 10 second duration, with a resting period of 3 or 10 seconds. Default treatment time is 20 minutes. Mixed Incontinence Treatment Mixed treatment modes consist of a 2, 5, or 10 minute stress treatment followed by a 2, 5, or 10 minute urge treatment. This combination of stimulation types aids in both stress and urge incontinence in the manners mentioned above. Caution: Federal law restricts this device to sale by or on the order of a licensed healthcare practitioner. 5
  • 5. Safety Information  Equipment Classification As per the International Electrotechnical Commission standard IEC 60601-1, and the European standard EN 60601-1, Medical Electrical Equipment, General Requirements for Basic Safety and Essential Performance, the InWave is classified as follows:  Type BF Equipment A Type BF piece of equipment is one that provides a particular degree of protection against electric shock, particularly regarding allowable leakage current. Type BF applied part is one that is floating above ground and is isolated from all other parts of the equipment.  Medical Device Directive (MDD) Classification The InWave is a Class IIa medical device per the European Medical Device Directive.  Internally Powered Equipment The InWave is powered by an internal, 9 VDC, MN1604, battery, so it is classified as Internally Powered Equipment.  Electromagnetic Compatibility Conforms to IEC 60601-1-2.  Temperature Operating temperature range: 0° - 40° C (32° - 104° F) Storage temperature range: -20° - 70° C (-4° - 158° F)  Water Ingress Ordinary equipment. This device does not have protection against ingress of water.  Flammable Anesthetics This device is not suitable for use in the presence of a flammable anesthetic mixture with air, or in the presence of a flammable anesthetic mixture with oxygen or nitrous oxide. 6
  • 6. Safety Information  Mode of Operation This device is suitable for continuous operation.  Symbols Safety symbols shown on the device above are defined below. On/Standby. This symbol indicates that the labeled switch electronically cycles the DC power on and off for part of the equipment. General Warning Sign. Follow warnings stated in the instruction manual to prevent potential hazards. Refer to Instruction Manual/Booklet. The operator must read, understand, and follow all instructions in the accompanying document including all warnings, cautions, and precautions before using this medical device. Type BF Equipment. This symbol indicates that the patient applied parts (electrodes) are Type BF (floating from ground) offering the user a specific level of safety. 7
  • 7. Waste Electrical and Electronic Equipment (WEEE). This product may contain substances known to be hazardous to the environment or to human health. It should be disposed of properly (for example, at your local waste collection administration or recycling plant) and in accordance with local ordinances.  Service and Calibration  Do not remove the cover. There are no user serviceable parts. Refer all service to authorized personnel. No modification of the equipment is allowed.  No preventative inspections are required. Factory testing and calibration ensure equipment accuracy and response. Contact Zynex Medical for factory re-calibration if necessary. 8
  • 8. InWave Sensor and Lead Wire Setup Step 1 Open the sensor package and remove the sensor. Remove the electrode wire from the packaging, and unbundle the wire. Note: Only Zynex Medical sensors are approved for use with the InWave. P/N 400015 are provided standard. Step 2 Insert the lead wire pin connectors into the sensor connectors as shown below. Sensor Connections Lead Wire 9
  • 9. InWave Electrode and Lead Wire Setup (continued) Step 3 Insert vaginal sensor as shown below. If needed, a small amount of water-based lubricant can be applied to the tip of the sensor to aid in insertion. Initial placement of the vaginal sensor and selection of the treatment program should be performed under the guidance of a physician. Note: Overuse of a lubricant can cause lack of stimulation, and lead to ineffective treatment. In order to avoid treatment interruption, it may be necessary to empty the bladder and bowels before starting a treatment session. See the directions included with the sensor for further information. Note: The vaginal sensors are for single patient use only. 10
  • 10. InWave Electrode and Lead Wire Setup (continued) Step 4 Plug the lead wire into the top of the device. Step 5 Proceed to page 16 to start treatment or page 12 to program device. Electrode Connector 11
  • 11. InWave Device Programming Instructions Turn device on by pressing the black On/Standby button 1. 2. 3. Select the desired Modality by pressing the “STRESS”, “URGE”, or “MIX” button once. Press the “+” or “-” portions of the Program button until the arrow on the LCD is next to the desired program. Press the Treatment Timer button until the desired treatment time is set on the screen. The InWave is now ready to use. On / Standby Button Modality Buttons Treatment Timer Button 12
  • 12. InWave Preprogrammed Modes Stress Incontinence Treatment (50 Hz Frequency, 200 µSecond Pulse Width) Program Description 5/10 The output signal ramps up to the level set by the user over a period of 1 second. The output stays on for the remainder of the 5 second on time, then ramps down to zero over a period of 1 second. The output stays off for the remainder of the 10 second off time. This cycle repeats for the duration of the treatment time. (Default is 15 minutes.) 3/10 The output signal ramps up to the level set by the user over a period of 1 second. The output stays on for the remainder of the 3 second on time, then ramps down to zero over a period of 1 second. The output stays off for the remainder of the 10 second off time. This cycle repeats for the duration of the treatment time. (Default is 15 minutes.) 5/5 The output signal ramps up to the level set by the user over a period of 1 second. The output stays on for the remainder of the 5 second on time, then ramps down to zero over a period of 1 second. The output stays off for the remainder of the 5 second off time. This cycle repeats for the duration of the treatment time. (Default is 15 minutes.) Urge Incontinence Treatment (12.5 Hz Frequency, 600 µSecond Pulse Width) Program Description 10/3 The output signal ramps up to the level set by the user over a period of 1 second. The output stays on for the remainder of the 10 second on time, then ramps down to zero over a period of 0.5 seconds. The output stays off for the remainder of the 3 second off time. This cycle repeats for the duration of the treatment time. (Default is 20 minutes.) 6/3 The output signal ramps up to the level set by the user over a period of 1 second. The output stays on for the remainder of the 6 second on time, then ramps down to zero over a period of 0.5 seconds. The output stays off for the remainder of the 3 second off time. This cycle repeats for the duration of the treatment time. (Default is 20 minutes.) 10/10 The output signal ramps up to the level set by the user over a period of 1 second. The output stays on for the remainder of the 6 second on time, then ramps down to zero over a period of 0.5 seconds. The output stays off for the remainder of the 3 second off time. This cycle repeats for the duration of the treatment time. (Default is 20 minutes.) 13
  • 13. Mixed Incontinence Treatment Program Description 10:10 A 10 minute “5/10” Stress program is run, followed by a 10 minute “10/3” Urge program. This cycle repeats for the duration of the treatment time. Default treatment time is 20 minutes. 5:5 A 5 minute “5/10” Stress program is run, followed by a 5 minute “10/3” Urge program. This cycle repeats for the duration of the treatment time. Default treatment time is 20 minutes. 2:2 A 2 minute “5/10” Stress program is run, followed by a 2 minute “10/3” Urge program. This cycle repeats for the duration of the treatment time. Default treatment time is 20 minutes. The InWave should be used daily for 15 to 20 minutes per session, for 1 to 3 months, or as directed by a physician. 14
  • 14. InWave Device Controls Display Attach Lead Wire Here Lead Wire Connector Turns Unit On/Off On/Standby Button Cycles Through Programs Adjusts Treatment Intensity Program Select Button Up and Down Button Unlocks Control Buttons Unlock Button STRESS, URGE, MIX Modality Buttons Adjusts Treatment Times Treatment Timer Button 15
  • 15. InWave Device Operating Instructions Start Treatment: Before starting treatment, the sensor must be inserted and the lead wire must be connected to the InWave. See pages 9-11 for instructions. 1. Turn the InWave on by pressing the On/Standby button once, and set the desired program using the Modality and Program buttons. 2. Increase the stimulation level by pressing the upper portion of the Intensity button until a strong but comfortable stimulation level is felt. NOTE: Output increases in 1/2 mA increments up to the 4 mA level, then increases in increments of 1 mA. 3. Once the desired level of stimulation is set, the InWave will automatically shut off after the preset treatment time (shown on the display.) If the Treatment Timer has been set to, “No Timer,” then the InWave will need to be shut off manually. Refer to Programming Instructions on page 12 to adjust the Treatment Timer. 4. To turn off the InWave manually, press the On/Standby button. Note: If “Check Connections!” message is displayed on the screen, see “Display Alerts” below. During Treatment: IMPORTANT: Button controls lock after 20 seconds of inactivity. To unlock button controls, press the Unlock button. Increase Intensity: To increase the intensity, press the upper potion of the Intensity button until the desired level of stimulation is felt. Decrease Intensity: To decrease the intensity, press the lower portion of the Intensity button until the desired level of stimulation is felt. Display Alerts: Check Connections: The lead wire and/or the sensor may not be attached properly. Check all connections and try again (see page 23). If the problem persists, call Technical Support (see page 4). Low Battery: Replace the battery immediately. Locked: The stimulation level and mode cannot be changed until the Unlock button is pressed. 16
  • 16. InWave Indications for Use Stress Incontinence Mode Please read the InWave User’s Manual before using the InWave pelvic muscle stimulator. CAUTION: Federal law restricts this device to sale by or on the order of a licensed healthcare practitioner. This device should only be used under medical supervision for adjunctive therapy for the treatment of medical diseases and conditions. Indications  The Zynex Medical InWave is intended to provide electrical stimulation and neuromuscular re-education for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of stress, urge, and mixed urinary incontinence in women. Contraindications  This InWave should not be used in the presence of the following: -Where there is a tendency to hemorrhage following acute trauma or fracture. -Following recent surgical procedures when muscle contraction may disrupt the healing process. -Over the menstruating or pregnant uterus. -Over areas of the skin which lack normal sensation. -Irregular menstrual cycles. -Urinary or vaginal infection. -History of urinary retention or current symptoms. Warnings       The long-term effects of chronic electrical stimulation are unknown. Safety has not been established for the use of the InWave during pregnancy. The device is capable of generating currents and voltages in excess of 10 mA RMS and 10 V RMS, respectively. A yellow LED indicator is provided next to the lead wire connector to show that the unit is delivering energy for any non-zero value of stimulation output (0.5 to 80 mA). Discontinue use of this device in the event of pain or bleeding. Do not use this device in water. Do not insert the lead wires into a mains power supply. 17
  • 17. InWave Precautions and Adverse Reactions Stress Incontinence Mode Precautions             Stimulation settings should be based on the guidance of the prescribing practitioner. This device should be kept out of the reach of children. This device should only be used with the leads and vaginal sensor recommended for use by the manufacturer. This device should not be used while driving, operating machinery, or during any activity in which involuntary muscle contractions may put the user at risk of injury. Caution should be exercised when this device is used by patients with pacemakers and heart rhythm problems. Do not use this device while under the influence of drugs or alcohol. Do not insert or remove the vaginal sensor while the device is ON. Do not use this device while in the close proximity of wireless devices. Use this device only as directed. Do not immerse the unit in water or any other liquid. The vaginal sensor should not remain inserted in the vagina when the unit is not in use. This device should not be used while sleeping or when anything else is in the vagina, such as a diaphragm or tampon. Adverse Reactions  Skin irritation and electrode burns are potential adverse reactions. Safety References Zynex Medical, Inc. is only responsible for the safety, reliability and function of the device when repairs and adjustments have been made by persons authorized by Zynex Medical, Inc., and the device is used in accordance with the user’s manual. Repairs and technical safety tests shall only be performed by authorized personnel. 18
  • 18. InWave Indications for Use Urge Incontinence Mode Please read the InWave User’s Manual before using the InWave pelvic muscle stimulator. CAUTION: Federal law restricts this device to sale by or on the order of a licensed healthcare practitioner. This device should only be used under medical supervision for adjunctive therapy for the treatment of medical diseases and conditions. Indications  The Zynex Medical InWave is intended to provide electrical stimulation and neuromuscular re-education for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of stress, urge, and mixed urinary incontinence in women. Contraindications  This InWave should not be used in the presence of the following: -Where there is a tendency to hemorrhage following acute trauma or fracture. -Following recent surgical procedures when muscle contraction may disrupt the healing process. -Over the menstruating or pregnant uterus. -Over areas of the skin which lack normal sensation. -Irregular menstrual cycles. -Urinary or vaginal infection. -History of urinary retention or current symptoms. Warnings       The long-term effects of chronic electrical stimulation are unknown. Safety has not been established for the use of the InWave during pregnancy. The device is capable of generating currents and voltages in excess of 10 mA RMS and 10 V RMS, respectively. A yellow LED indicator is provided next to the lead wire connector to show that the unit is delivering energy for any non-zero value of stimulation output (0.5 to 80 mA). Discontinue use of this device in the event of pain or bleeding. Do not use this device in water. Do not insert the lead wires into a mains power supply. 19
  • 19. InWave Precautions and Adverse Reactions Urge Incontinence Mode Precautions             Stimulation settings should be based on the guidance of the prescribing practitioner. This device should be kept out of the reach of children. This device should only be used with the leads and vaginal sensor recommended for use by the manufacturer. This device should not be used while driving, operating machinery, or during any activity in which involuntary muscle contractions may put the user at risk of injury. Caution should be exercised when this device is used by patients with pacemakers and heart rhythm problems. Do not use this device while under the influence of drugs or alcohol. Do not insert or remove the vaginal sensor while the device is ON. Do not use this device while in the close proximity of wireless devices. Use this device only as directed. Do not immerse the unit in water or any other liquid. The vaginal sensor should not remain inserted in the vagina when the unit is not in use. This device should not be used while sleeping or when anything else is in the vagina, such as a diaphragm or tampon. Adverse Reactions  Skin irritation and electrode burns are potential adverse reactions. Safety References Zynex Medical, Inc. is only responsible for the safety, reliability and function of the device when repairs and adjustments have been made by persons authorized by Zynex Medical, Inc., and the device is used in accordance with the user’s manual. Repairs and technical safety tests shall only be performed by authorized personnel. 20
  • 20. InWave Indications for Use Mixed Incontinence Mode Please read the InWave User’s Manual before using the InWave pelvic muscle stimulator. CAUTION: Federal law restricts this device to sale by or on the order of a licensed healthcare practitioner. This device should only be used under medical supervision for adjunctive therapy for the treatment of medical diseases and conditions. Indications  The Zynex Medical InWave is intended to provide electrical stimulation and neuromuscular re-education for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of stress, urge, and mixed urinary incontinence in women. Contraindications  This InWave should not be used in the presence of the following: -Where there is a tendency to hemorrhage following acute trauma or fracture. -Following recent surgical procedures when muscle contraction may disrupt the healing process. -Over the menstruating or pregnant uterus. -Over areas of the skin which lack normal sensation. -Irregular menstrual cycles. -Urinary or vaginal infection. -History of urinary retention or current symptoms. Warnings       The long-term effects of chronic electrical stimulation are unknown. Safety has not been established for the use of the InWave during pregnancy. The device is capable of generating currents and voltages in excess of 10 mA RMS and 10 V RMS, respectively. A yellow LED indicator is provided next to the lead wire connector to show that the unit is delivering energy for any non-zero value of stimulation output (0.5 to 80 mA). Discontinue use of this device in the event of pain or bleeding. Do not use this device in water. Do not insert the lead wires into a mains power supply. 21
  • 21. InWave Precautions and Adverse Reactions Mixed Incontinence Mode Precautions             Stimulation settings should be based on the guidance of the prescribing practitioner. This device should be kept out of the reach of children. This device should only be used with the leads and vaginal sensor recommended for use by the manufacturer. This device should not be used while driving, operating machinery, or during any activity in which involuntary muscle contractions may put the user at risk of injury. Caution should be exercised when this device is used by patients with pacemakers and heart rhythm problems. Do not use this device while under the influence of drugs or alcohol. Do not insert or remove the vaginal sensor while the device is ON. Do not use this device while in the close proximity of wireless devices. Use this device only as directed. Do not immerse the unit in water or any other liquid. The vaginal sensor should not remain inserted in the vagina when the unit is not in use. This device should not be used while sleeping or when anything else is in the vagina, such as a diaphragm or tampon. Adverse Reactions  Skin irritation and electrode burns are potential adverse reactions. Safety References Zynex Medical, Inc. is only responsible for the safety, reliability and function of the device when repairs and adjustments have been made by persons authorized by Zynex Medical, Inc., and the device is used in accordance with the user’s manual. Repairs and technical safety tests shall only be performed by authorized personnel. 22
  • 22. Troubleshooting Problem Solution Unit stays on even after treatment ends. Hold Off button down for 2 seconds to shut unit off. Otherwise unit will shut off automatically after 5 minutes of no stimulation. Alternatively you can start a new treatment session after the last one ends. Cannot increase level from its current setting. Press and release Unlock button to unlock this safety feature. Then increase stimulation to the desired level of intensity. Intensity level is always locked after 20 seconds of key press inactivity. Display shows “Check Check to see that the sensor is properly inserted. Connections” (Remove and re-insert the sensor if necessary to (Electrode alarm). establish a better electrical contact.) Verify that the sensor is correctly connected to the lead wire and that the lead wire is correctly connected to the InWave. Connect the bare metal pins of the lead wire together to short-circuit the output. If the electrode alarm ceases the problem is with the sensor connection integrity. Removing and re-inserting the sensor may establish a better electrical contact. 23
  • 23. InWave Specifications and Accessories Stress Incontinence Mode Amplitude: Frequency: Pulse Width: Stress Modes: On Time: Off Time: Ramp Up Time: Ramp Down Time: Waveforms: 0-80 mA 50 Hz 200 µseconds 5/10, 3/10, 5/5 3 or 5 seconds 5 or 10 seconds 1 second 1 second Symmetrical biphasic Urge Incontinence Mode Amplitude: Frequency: Pulse Width: Urge Modes: On Time: Off Time: Ramp Up Time: Ramp Down Time: Waveform: 0-80 mA 12.5 Hz 600 µseconds 10/3, 6/3, 10/10 6 or 10 seconds 3 or 10 seconds 1 second 0.5 seconds Symmetrical biphasic Mixed Incontinence Mode Amplitude: Frequency: Pulse Width: Mix Modes: On-Time: Off-Time: Ramp Up: Ramp Down: Waveforms: 0-80 mA 12.5 Hz and 50 Hz 200 µseconds and 600 µseconds 10:10, 5:5, 2:2 3, 5, 6, or 10 seconds 3, 5, or 10 seconds 1 second 0.5 or 1 second Symmetrical biphasic Other Specifications Treatment timer: 5 to 25 minutes, in 5 minute increments. Compliance meter: Records total usage time in minutes and number of times used. Dimensions: 2.9 x 4.6 x 1.0 in. Weight: 5.8 oz. including battery Warranty: 5 year manufacturer’s warranty on materials and workmanship. Accessories excluded. Accessories Lead wire: 104204, lead wire, gray. Sensor: 400015, vaginal sensor. Battery: 130010, battery, 9 Volt, alkaline, MN1604. 24
  • 24. InWave Sensor, Batteries, and Cleaning Sensor Use The InWave is intended to be used with a re-usable vaginal sensor. Before and after each use, clean the sensor with mild soap and water, and dry the sensor. Store the sensor in a safe location when not in use. Follow directions of the prescribing physician. Note: The vaginal sensors are for single patient use only. Battery Power One 9 Volt alkaline battery is used to power the InWave. The battery compartment on the back of the device opens by sliding the cover downwards. Rechargeable batteries are not recommended since they have only a short usage time and cannot be charged while in the InWave. Battery replacement is indicated by the battery icon in the display. Replace with a new 9 Volt alkaline battery. The InWave operates normally even if the 9 Volt battery is inserted with reverse polarity. Please dispose of used batteries properly. Device Cleaning The InWave case and display window can be cleaned by lightly applying mild soap to a damp soft cloth or paper towel and using it to wipe the surfaces clean. Allow the unit to dry thoroughly before using. Do not spray cleaning solutions directly onto the unit, or immerse it in water or other liquids. 25
  • 25. InWave Warranty Information The InWave device is warranted to be free from defects in material, workmanship, and structural integrity when subjected to normal domestic use and service for five years after the original purchase. During that time, Zynex Medical, Inc will repair or replace, at its sole discretion, the InWave device that has been used in a standard manner. This warranty does not cover misuse or use contrary to the operating instructions supplied. Warranty obligations are limited to replacement or repair of defective parts and components, at the option of Zynex Medical, Inc. This Warranty is only valid for the original purchaser of this product and will not be honored if the product ownership is transferred or is resold. To obtain warranty service, please contact Technical Support at the number listed on page 4 of this manual. 26