The Elcometer 146 SaltSmart™ Salt Contamination System determines the amount of soluble salts present in a test sample. Using both the Elcometer SaltSmart™ Test Strips and the Elcometer 146 SaltSmart™ Contamination Meter, this system offers a fast, easy-to-use alternative to the Bresle Test and has been proven to be equivalent to ISO 8502-9.
A de-ionised water bottle is attached to the end of the Elcometer 146 SaltSmart™ Test Strip which is then fixed to the substrate using non-residue tape. The test strip is positioned at a 10° angle to allow the water to flow on to the foam sensor which is in contact with the substrate.
After approximately 8 minutes the test strip is removed and the amount of soluble salts present is measured using the Elcometer 146 SaltSmart™ Contamination Meter, which displays and stores results in µS/cm, ug/cm² or mg/m².
Measuring salt contamination using the SaltSmart™ Contamination Meter
Taking care not to touch the white foam sensor, remove the protective packaging and place the de-ionized water bottle onto the SaltSmart™ test strip.
Place the test strip & bottle on to the substrate and hold in place with painters tape. On vertical surfaces, make sure that the red line on the test strip is horizontal - creating a 10° incline to allow water flow.
Whilst multiple test strips can be set up at the same time, after approximately 8 minutes remove each test strip in order from the surface.
Place the test strip into the SaltSmart™ meter, and read the value on the display.
Environmentally discard the test strip, bottle and tape.
Part Number
Description
E146----1
Elcometer 146 SaltSmart™ Soluble Salt Contamination Meter
Measuring Range
0-150μS/cm (0-155 mg/m2)
Resolution
1 μS (1.03 mg/m2)
Operating Temperature
0—50 °C (32—122 °F)
Accuracy
±1%
Temperature Coeffient
2.0 % per °C (F)
Battery Type
2 AA Batteries; suitable for approx 100+ tests
Sample Time
8 minutes for strip development, 15 seconds for meter analysis
Sampling Size
20.6 x 15mm (309 mm2)
Dimensions
160 x 80 x 30mm
Weight
1.2 kg (2.5 lb)
Packing List
Elcometer 146 SaltSmart™ Contamination Meter, Rubber Boot for SaltSmart™ Contamination Meter, USB Cable, SaltSmart™ Validation Strip , 2 x AA Batteries, Belt Pouch, Getting Started CD
Can be used in accordance with:
NACE SP0508, IMO PSPC, ISO 8502-9, SSPC, US Navy 009-32 FY12