MedGem Training Module 
This course is designed to teach you how to use the 
MedGem device to personalize your clientsā€™ nutrition 
and exercise programs. At the completion of this 
training, you will be able to: 
ā€¢ Define and understand resting metabolic rate (RMR). 
ā€¢ Be able to perform a measurement using the MedGem device. 
ā€¢ Interpret and apply the measurement. 
ā€¢ Answer frequently asked questions. 
Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
Resting Metabolic Rate 
The Missing Link to Weight Management 
Your resting metabolic rate is the number of calories 
your body burns each day with no activity. It makes 
up the majority (60-75% in healthy individuals and 
100% in critically ill) of calories burned each day. 
Resting metabolic rate is unique to each individual 
and is affected by a number of factors: 
ā€¢ Age 
ā€¢ Body Weight 
ā€¢ Medications 
ā€¢ Fever and/or Infection 
ā€¢ Weight Loss 
1 Danforth, E. LifeSci 28:1891-1897, 1981. 
2 Fung EB, AACN Clinical Issues 11(4): 480-497, 2000. 
ā€¢ Gender 
ā€¢ Genetics 
ā€¢ Hormones 
ā€¢ Exercise 
ā€¢ Nutritional Supplements 
Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
Predictive Equations are inaccurate 
for Healthy and III Individuals 
Healthy Individuals 
RMR can vary significantly from person to person ā€“ up to 
900 calories per day in healthy people. 
Ill Individuals 
Several studies have indicated predictive equations are 
significantly different than measured RMR. Equations can 
overestimate RMR by as much as 59% and underestimate 
as much as 30% in sick individuals. 
This variability can result in very different outcomes for individuals. 
ADAā€™s Adult & Pediatric Weight Management Guidelines: 
ā€œIf possible, Indirect Calorimetry, should be used for assessing RMR when 
developing energy plans.ā€ 
1. Foster, G, et al. Metabolism, 1988; 37(5), 467-472 
2. Mann et al, Crit Care Med 1985: 13: 173-177 
3. Weissman et al, Journal of Anesthesiology 1986: 64: 673 679 
4. ABA-EBL: www.ADAevidencelibrary.com 
Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
RMR Varies Among Healthy and Ill Individuals 
Foster, G, et al. Metabolism, 37(5): 467-472, 1988 
Hunter D, et al. Br J Surg, 75: 875-878,1988, 
Welis, P. Am J Clin, 88; 959-970, 2008, 
Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
Clinical Benefits of Measuring RMR 
Measuring RMR enables you to gauge the effects of the 
clientā€™s condition on his/her calorie needs and adjust 
your nutrition recommendations accordingly. Conditions 
that impact your RMR include: 
ā€¢ disease severity and point of time in disease process 
ā€¢ change in medications/dosage 
ā€¢ change in dietary regimen 
ā€¢ change in weight (+/- 5%) 
Benefits 
Providing appropriate feeding minimizes the percentage of clients who 
are overfed and underfed. 
Patients following a diet plan established from measured RMR lose more 
weight compared to diet plans that estimate RMR. 
Measuring RMR to establish appropriate nutritional needs can result in 
direct cost savings by avoiding unnecessary nutritional support. 
1. Alberda et al., Nutri Clin Pract 17: 38-42, 2002 
2. McDoniel, et al., Obes Facts (1)6; 298-304, 2008. 
3. McClave et al, JPEN 22; 375-381, 1998. 
Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
Candidates for Metabolic Measurement 
Some facilities use the device on every individual as 
part of their initial nutrition assessment to determine 
baseline calorie needs; others choose to use the device 
on individuals that meet certain criteria that put them at 
high risk for complications associated with overweight 
and obesity. 
Sample criteria for determining medical necessity for diagnostic 
procedure (CPT 94690). Average payer reimbursement is $50-$80. 
ā€¢ 414 Chronic Pulmonary HT Disease 
ā€¢ 428 Heart Failure* 
ā€¢ 493 Asthma* 
ā€¢ 496 COPD* 
ā€¢ 51.8 Lung Involvement in other disease* 
ā€¢ 780.5 Obstructive Sleep Apnea 
ā€¢ 786.5 Dyspnea* 
ā€¢ 786.9 Hypoventilation Syndrome* 
ā€¢ 244.9 Hypothyroidism 
ā€¢ 250 Diabetes Mellitus 
ā€¢ 272.1 Hypercholesterolemia 
ā€¢ 22.4 Hyperlipidemia 
ā€¢ 277.7 Metabolic Syndrome 
ā€¢ 278 Obesity 
ā€¢ 401 Hypertension 
ā€¢ 414 Cardiovascular Disease 
*CMS indicates medical necessity 
Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
Performing a Measurement 
About the MedGem Device 
The MedGem is a hand-held indirect calorimeter that 
measures oxygen consumption to determine resting 
metabolic rate. 
The MedGem device is easy to use, providing 
measurements quickly and accurately. Useful as a tool 
for monitoring changes in metabolism, the MedGem 
gives you important information to help your clients 
achieve their personal weight management, nutrition, 
and exercise goals. 
ā€œThe validation results suggest the MedGem can be used to accurately measure oxygen 
consumption and RMR in long-term healthcare facilities, low-capacity hospitals, athletic training 
centers, health clubs, and weight loss clinics.ā€ 
Richard Branson, MS, RTT 
University of Cincinnati Medical Center, Cincinnati, OH 
1. RMR is calculated using the Weir equation and a constant RQ value of 0.85 (RMR=6.931 X VO2), 
Weir,JB., New Methods for Calculating Metabolic Rate with Special Reference to Protein Metabolism. J Physiology, 1949, 109; 1-9. 
Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
Getting Acquainted with MedGem 
ā€¢ The start button and LED indicator light are 
located on top. 
ā€¢ The air vent is located on the bottom of the 
device. 
ā€¢ The device is powered by a DC power 
supply, which plugs into the side of the 
MedGem. 
Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
MedGem Measurement Protocol 
Acceptable Conditions: 
1. Measurement is performed in quiet environment. 
2. Individual rests for 10-15 minutes prior to measurement. 
3. Individual has not exercised for at least 4 hours. 
4. Individual has not consumed caffeine for at least 4 hours. 
5. Individual has not consumed nutrition supplements for at least 4 hours. 
6. Individual has not used nicotine for at least 1 hour. 
7. Individual is not taking acute medications. Chronic medications are OK. 
8. Individual may have water prior to measurement. 
Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
1. Plug cord in wall outlet. 
2. Plug cord in device. 
3. Device will beep. LED light turns from red to green to orange. 
Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
4. Place MedGem on a table away from fans or vents. 
5. Insert mouthpiece into the MedGem. 
6. Push orange button to begin self-calibration process. 
Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
7. Self-calibration completed when 
device beeps and LED light turns to 
green. 
8. Hand device to the client and ensure there is 
a tight seal around the mouthpiece. 
9. Client should be seated or in a supine 
position and breathing normally into device. 
10. The MedGem will beep when the 
measurement is completed. 
11. Record RMR and discard mouthpiece and 
noseclip. 
Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
Tracking the Measurements 
Codes beginning with ā€œC-ā€ (ie C-10) refer to the 
remaining number of measurements that may be performed 
with a particular MedGem device. 
(C-10 means you have 10 remaining 
Ā© 2014 All Rights Reserved. No duplication and distribution without permission! 
measurements). 
A two-tone audible alert, combined with a flashing 
amber and green light indicates the MedGem is low on 
measurements. To acknowledge this warning and begin 
self-calibrating the device, press the Start button twice. 
When you receive a low measurement warning, 
contact Microlife Medical Home Solutions to arrange for 
delivery of a replacement.
Error Codes 
Errors beginning with ā€œ0ā€ (ie 01, 05) indicate user errors. 
Refer to the user manual for instructions. 
Solutions: 
1. Ensure the flowtube is firmly placed into the device. 
2. Ensure the single-use mouthpiece is firmly inserted into the flowtube. 
3. Ensure the client has a proper seal around the mouthpiece and noseclip 
is properly positioned on their nose. The client should only be able to 
breathe through their mouth. 
4. Repeat the measurement. 
Errors beginning with any other number (ie 11) may indicate possible 
internal error. Contact Microlife Medical Home Solutions for instructions. 
Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
Caring for the MedGem Device 
Here are some guidelines for properly caring for the 
Ā© 2014 All Rights Reserved. No duplication and distribution without permission! 
device. 
Do not expose the inside of the MedGem to cleaning solutions 
or water. If necessary, the outside of the MedGem can be 
cleaned with any of the following: 
ā€¢ Soft, clean cloth 
ā€¢ Clorox disinfecting wipes 
ā€¢ Soap and water 
Store you MedGem in its case when not in use.
The Missing Link 
Through self-monitoring, health professionals have 
been able to help clients track how many calories they 
eat. But thatā€™s only half of the energy balance equation. 
Your clients need to know how many calories they burn. 
Purposeful exercise and lifestyle/occupation 
activities contribute to the number of calories burned by 
an individual. 
Resting metabolic rate, however, accounts for the 
majority of calories burned and has been the missing link 
in nutrition programs. 
Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
ā€œIs my RMR good or bad?ā€ 
Unlike other health assessments (ie: blood pressure, 
etc), RMR assessments are not good or bad. RMR is unique to 
your clients physiology. Therefore, RMR may be higher or 
lower than other like individuals. 
The MedGem Analyzer will provide you metabolic range 
for each client. The metabolic range is based on the level of 
oxygen consumption and current body weight of your client. 
Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
ā€œWhat do I do with the RMR number?ā€ 
Now you have information you need to develop a 
balanced, personalized nutrition and exercise plan for your 
clients. 
Microlife Medical Home Solutions, Inc. offers several 
options for developing a calorie budget for individuals. Please 
visit our website at www.mimhs.com to learn how the 
MedGem Analyzer can help you develop a nutritional plan for 
your clients. 
Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
ā€œWhat do I do with the measurement result?ā€ 
Weight Loss Case Study 
Brian has a BMI of 32.3 kg/m. He was recently diagnosed with 
non-insulin dependent diabetes and hypertension. His physician 
recommends Brian lose 25 pounds by his next appointment in ~3 
months. The MedGem Analzyer will generate an individual 
treatment plan for his weight loss goal. 
ā€¢ Current weight: 245 lbs 
ā€¢ Goal weight: 220 lbs 
ā€¢ RMR: 2300 kcals VO2: 332 ml/min 
ā€¢ Occupation: Computer Programmer (~45 hours/wk) 
ā€¢ Goal date: 12 weeks 
ā€¢ Exercise plan: Elliptical Trainer and/or walking moderate-intensity for 30 
minutes, 5 days per week 
Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
ā€œWhat do I do with the measurement result?ā€ 
Weight Loss Case Study 
Personalized Program: For Brian to lose 25 lbs, he will need to consume 
1685 calories per day. 
ā€¢ Daily Energy Expenditure: 2685 kcals (determined by Analyzer) 
ā€¢ Rate of Weight Loss: -1000 kcal (2 lbs per week) 
ā€¢ Daily Calorie Needs for weight loss: 1685 kcals 
With Brianā€™s exercise plan, he will burn an additional 245 kcals. Therefore, Brian 
could consume 1930 kcals on the days he exercises and still reach his weight goal 
by his goal date. 
ā€¢ Daily Calorie Needs for weight loss 1685 kcals 
ā€¢ Exercise Program: +245 kcals 
ā€¢ Daily Calorie Needs with exercise 1930 kcals 
Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
Why Re-Measure? 
Resting metabolic rate can change significantly during a weight 
management program. Re-measuring RMR will ensure that 
your clients are on a successful track for long-term weight 
management success or appropriate recovery. 
You should re-measure individuals: 
ā€¢ +/- 5% - 10% weight change during a weight management 
program 
ā€¢ If no weight loss has occurred in the first few weeks 
ā€¢ If individual has reached a weight loss plateau 
ā€¢ When there is a change in medications or a medical condition 
ā€¢ When there is no response to nutritional therapy 
Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
Avoiding Common Pitfalls 
Problem: Lost Readings 
Tim, a weight loss client, arrives for his RMR appointment. You 
seat him in a quiet room and allow him to relax as you review 
the measurement procedures and set up the MedGem device. 
You begin the measurement after calibrating the device. About 
5 minutes later, the MedGem beeps indicating the 
measurement is complete. 
You remove the noseclip and discard the single use 
mouthpiece. As you reach for the MedGem, you accidentally 
disconnect the power cord and the device shuts off before you 
have a chance to record the measurement reading. 
Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
Avoiding Common Pitfalls 
Solution: Lost Readings 
To recover Timā€™s RMR reading, simply plug the power cord back into 
the MedGem. After the device powers up, the last RMR measurement will be 
displayed. Record the reading. 
Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
Avoiding Common Pitfalls 
Problem: Inaccurate Measurements 
Mona is anxious to lose the weight she gained from her first pregnancy. She 
decides to have her RMR measured. This allows you to develop a personalized 
program to help her reach her weight-loss goal. 
When she arrives, you seat her in a quiet room and allow her to relax as your 
review the measurement procedures and set up the MedGem device. Then, you 
perform a measurement. 
You make note of the reading, which is 2980, and discard the single-use 
mouthpiece and noseclip. The reading is suspiciously high. 
What could have gone wrong? 
Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
Avoiding Common Pitfalls 
Solution: Inaccurate Measurements 
Enter the clientā€™s RMR measurement into the MedGem Analyzer. Compare the 
clientā€™s RMR to the listed normal range. Refer to the listed normal range in the 
MedGem Analyzer. 
When using the MedGem to measure a clientā€™s RMR, it is important to have 
the client adhere to proper resting conditions. 
If the client has performed physical activity, taken acute medications, has 
eaten, or has not rested for 10-15 minutes prior to the measurement, the MedGem 
measurement will not be accurate to their actual RMR. Always determine that the 
resting conditions were followed before performing a measurement. Developing a 
weight loss plan based on this inaccurate reading will prevent the client from reaching 
their goals. 
Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
Avoiding Common Pitfalls 
Problem: Undetected Air Leak 
Brian enrolled in your weight management program. He is 25 pounds 
overweight and needs to lower his cholesterol levels and blood pressure. 
You seat him in a quiet room and perform the MedGem measurement under 
acceptable conditions. You make a note of the reading, which is 1400, then you 
discard this single-use mouthpiece and noseclip. 
Both you and Brian are suspicious that this RMR reading may be too low since 
he weighs 315 pounds. 
What could have gone wrong? 
Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
Avoiding Common Pitfalls 
Solution: Undetected Air Leak 
Brianā€™s reading may be inaccurate due to an air leak that did not cause an 
error message to display. If an air leak occurs, some of the air is not measured by the 
device and Brianā€™s actual RMR would therefore not be determined. 
When using the MedGem to measure a clientā€™s RMR, it is important to insure 
that all resting conditions are followed. Your client should rest for 10-15 minutes prior 
to starting the measurement. Measure your client if he/she has: 
ā€¢ Not eaten or had anything to drink (water is OK) 4 hours prior 
ā€¢ Not exercised 4 hour prior 
ā€¢ Not consumed caffeine or stimulatory nutritional supplements 4 hours prior 
ā€¢ Not used nicotine for 1 hour prior 
ā€¢ Not taken acute medications (ie: antibiotics, pain medication) 
Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
Your Clients Will Have Questions! 
Your clients will have questions about metabolism, the 
MedGem device, and RMR. 
Microlife Medical Home Solutions has anticipated many of the 
questions that will be asked by you and your clients. The answers are 
provided in the MedGem User Guide. 
Read the document: MedGem User Guide for answers to 
common questions. If you need further assistance, visit our website at 
www.mimhs.com. 
Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
Congratulations! Youā€™ve reached the end! 
Donā€™t Forget about our additional tools 
Ā© 2014 All Rights Reserved. No duplication and distribution without permission! 
Measure 
Use the MedGem 
and MedGem 
Analyzer software 
to determine RMR 
and assess energy 
balance for a 
personalized 
nutrition program. 
Monitor 
Have clients self-monitor 
their food 
intake and exercise 
using MiHealthLog 
health management 
application. 
Modify 
Help you patients 
and clients adjust 
their nutrition and 
exercise plan by 
monitoring their 
daily health, 
nutrition, and 
activity goals with 
MiHealthCoach.

More Related Content

Similar to MedGem Indirect Calorimeter Training Slides

Pharmacy news 2014
Pharmacy news 2014Pharmacy news 2014
Pharmacy news 2014munaoqal
Ā 
A new algorithmforcprtrainingfinal
A new algorithmforcprtrainingfinalA new algorithmforcprtrainingfinal
A new algorithmforcprtrainingfinalMarco Darby
Ā 
Assessment Methods For Nutritional Status
Assessment Methods For Nutritional StatusAssessment Methods For Nutritional Status
Assessment Methods For Nutritional StatusDrSindhuAlmas
Ā 
Dr John Duffy Polypharmacy and Frailty
Dr John Duffy   Polypharmacy and FrailtyDr John Duffy   Polypharmacy and Frailty
Dr John Duffy Polypharmacy and FrailtyHealth Innovation Wessex
Ā 
Purple cow 2012 (wellness button)
Purple cow 2012 (wellness button)Purple cow 2012 (wellness button)
Purple cow 2012 (wellness button)thepurplecowph
Ā 
patient safety and enviroments 2 (1).pptx
patient safety and enviroments 2 (1).pptxpatient safety and enviroments 2 (1).pptx
patient safety and enviroments 2 (1).pptxssuserfd7cc21
Ā 
The Surprising ROI of Employee Wellness
The Surprising ROI of Employee WellnessThe Surprising ROI of Employee Wellness
The Surprising ROI of Employee WellnessMedgate Inc.
Ā 
Qualcomm Life Connect 2013 - Terrance H. Gregg
Qualcomm Life Connect 2013 - Terrance H. GreggQualcomm Life Connect 2013 - Terrance H. Gregg
Qualcomm Life Connect 2013 - Terrance H. GreggQualcomm Life
Ā 
Chapter 7_Health Screening Services in Community Pharmacy.pptx
Chapter 7_Health Screening Services in Community Pharmacy.pptxChapter 7_Health Screening Services in Community Pharmacy.pptx
Chapter 7_Health Screening Services in Community Pharmacy.pptxVinayGaikwad14
Ā 
Purple cow 2012 (wellness button)
Purple cow 2012 (wellness button)Purple cow 2012 (wellness button)
Purple cow 2012 (wellness button)thepurplecowph
Ā 
Redmond Student Orientation 2014
Redmond Student Orientation 2014Redmond Student Orientation 2014
Redmond Student Orientation 2014BlueRidgeAHEC
Ā 
Purple cow 2012 (wellness button)
Purple cow 2012 (wellness button)Purple cow 2012 (wellness button)
Purple cow 2012 (wellness button)thepurplecowph
Ā 
Egyptalum integrated diabetes center profile
Egyptalum integrated diabetes center profile Egyptalum integrated diabetes center profile
Egyptalum integrated diabetes center profile Emad Hamed
Ā 
obesity - Prevention guidelines
 obesity - Prevention guidelines    obesity - Prevention guidelines
obesity - Prevention guidelines Vinod Kumar
Ā 
10. assessment of nutritional status
10. assessment of nutritional status10. assessment of nutritional status
10. assessment of nutritional statusUsman Khan
Ā 

Similar to MedGem Indirect Calorimeter Training Slides (20)

ABBOTT.pptx
ABBOTT.pptxABBOTT.pptx
ABBOTT.pptx
Ā 
Pharmacy news 2014
Pharmacy news 2014Pharmacy news 2014
Pharmacy news 2014
Ā 
Strategies to Reduce Errors
Strategies to Reduce ErrorsStrategies to Reduce Errors
Strategies to Reduce Errors
Ā 
A new algorithmforcprtrainingfinal
A new algorithmforcprtrainingfinalA new algorithmforcprtrainingfinal
A new algorithmforcprtrainingfinal
Ā 
Assessment Methods For Nutritional Status
Assessment Methods For Nutritional StatusAssessment Methods For Nutritional Status
Assessment Methods For Nutritional Status
Ā 
Dr John Duffy Polypharmacy and Frailty
Dr John Duffy   Polypharmacy and FrailtyDr John Duffy   Polypharmacy and Frailty
Dr John Duffy Polypharmacy and Frailty
Ā 
Purple cow 2012 (wellness button)
Purple cow 2012 (wellness button)Purple cow 2012 (wellness button)
Purple cow 2012 (wellness button)
Ā 
patient safety and enviroments 2 (1).pptx
patient safety and enviroments 2 (1).pptxpatient safety and enviroments 2 (1).pptx
patient safety and enviroments 2 (1).pptx
Ā 
The Surprising ROI of Employee Wellness
The Surprising ROI of Employee WellnessThe Surprising ROI of Employee Wellness
The Surprising ROI of Employee Wellness
Ā 
Qualcomm Life Connect 2013 - Terrance H. Gregg
Qualcomm Life Connect 2013 - Terrance H. GreggQualcomm Life Connect 2013 - Terrance H. Gregg
Qualcomm Life Connect 2013 - Terrance H. Gregg
Ā 
Chapter 7_Health Screening Services in Community Pharmacy.pptx
Chapter 7_Health Screening Services in Community Pharmacy.pptxChapter 7_Health Screening Services in Community Pharmacy.pptx
Chapter 7_Health Screening Services in Community Pharmacy.pptx
Ā 
HC360-OurSystem
HC360-OurSystemHC360-OurSystem
HC360-OurSystem
Ā 
Purple cow 2012 (wellness button)
Purple cow 2012 (wellness button)Purple cow 2012 (wellness button)
Purple cow 2012 (wellness button)
Ā 
Redmond Student Orientation 2014
Redmond Student Orientation 2014Redmond Student Orientation 2014
Redmond Student Orientation 2014
Ā 
Purple cow 2012 (wellness button)
Purple cow 2012 (wellness button)Purple cow 2012 (wellness button)
Purple cow 2012 (wellness button)
Ā 
Egyptalum integrated diabetes center profile
Egyptalum integrated diabetes center profile Egyptalum integrated diabetes center profile
Egyptalum integrated diabetes center profile
Ā 
obesity - Prevention guidelines
 obesity - Prevention guidelines    obesity - Prevention guidelines
obesity - Prevention guidelines
Ā 
Md flux dx4 week 10
Md flux dx4 week 10Md flux dx4 week 10
Md flux dx4 week 10
Ā 
10. assessment of nutritional status
10. assessment of nutritional status10. assessment of nutritional status
10. assessment of nutritional status
Ā 
Use your Pharmacist Huentelman
Use your Pharmacist HuentelmanUse your Pharmacist Huentelman
Use your Pharmacist Huentelman
Ā 

Recently uploaded

Call Girl Number in Panvel MumbaišŸ“² 9833363713 šŸ’ž Full Night Enjoy
Call Girl Number in Panvel MumbaišŸ“² 9833363713 šŸ’ž Full Night EnjoyCall Girl Number in Panvel MumbaišŸ“² 9833363713 šŸ’ž Full Night Enjoy
Call Girl Number in Panvel MumbaišŸ“² 9833363713 šŸ’ž Full Night Enjoybabeytanya
Ā 
VIP Call Girls Indore Kirti šŸ’ššŸ˜‹ 9256729539 šŸš€ Indore Escorts
VIP Call Girls Indore Kirti šŸ’ššŸ˜‹  9256729539 šŸš€ Indore EscortsVIP Call Girls Indore Kirti šŸ’ššŸ˜‹  9256729539 šŸš€ Indore Escorts
VIP Call Girls Indore Kirti šŸ’ššŸ˜‹ 9256729539 šŸš€ Indore Escortsaditipandeya
Ā 
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...narwatsonia7
Ā 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escortsvidya singh
Ā 
CALL ON āž„9907093804 šŸ” Call Girls Baramati ( Pune) Girls Service
CALL ON āž„9907093804 šŸ” Call Girls Baramati ( Pune)  Girls ServiceCALL ON āž„9907093804 šŸ” Call Girls Baramati ( Pune)  Girls Service
CALL ON āž„9907093804 šŸ” Call Girls Baramati ( Pune) Girls ServiceMiss joya
Ā 
Call Girl Coimbatore Prishaā˜Žļø 8250192130 Independent Escort Service Coimbatore
Call Girl Coimbatore Prishaā˜Žļø  8250192130 Independent Escort Service CoimbatoreCall Girl Coimbatore Prishaā˜Žļø  8250192130 Independent Escort Service Coimbatore
Call Girl Coimbatore Prishaā˜Žļø 8250192130 Independent Escort Service Coimbatorenarwatsonia7
Ā 
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls ServiceKesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Servicemakika9823
Ā 
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...narwatsonia7
Ā 
(šŸ‘‘VVIP ISHAAN ) Russian Call Girls Service Navi MumbaišŸ–•9920874524šŸ–•Independent...
(šŸ‘‘VVIP ISHAAN ) Russian Call Girls Service Navi MumbaišŸ–•9920874524šŸ–•Independent...(šŸ‘‘VVIP ISHAAN ) Russian Call Girls Service Navi MumbaišŸ–•9920874524šŸ–•Independent...
(šŸ‘‘VVIP ISHAAN ) Russian Call Girls Service Navi MumbaišŸ–•9920874524šŸ–•Independent...Taniya Sharma
Ā 
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...Miss joya
Ā 
Bangalore Call Girls Marathahalli šŸ“ž 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli šŸ“ž 9907093804 High Profile Service 100% SafeBangalore Call Girls Marathahalli šŸ“ž 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli šŸ“ž 9907093804 High Profile Service 100% Safenarwatsonia7
Ā 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalorenarwatsonia7
Ā 
Call Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls Service
Call Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls ServiceCall Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls Service
Call Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls Servicenarwatsonia7
Ā 
Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...
Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...
Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...Miss joya
Ā 
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy GirlsCall Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girlsnehamumbai
Ā 
VIP Call Girls Pune Sanjana 9907093804 Short 1500 Night 6000 Best call girls ...
VIP Call Girls Pune Sanjana 9907093804 Short 1500 Night 6000 Best call girls ...VIP Call Girls Pune Sanjana 9907093804 Short 1500 Night 6000 Best call girls ...
VIP Call Girls Pune Sanjana 9907093804 Short 1500 Night 6000 Best call girls ...Miss joya
Ā 
ā™›VVIP Hyderabad Call Girls ChintalkuntašŸ–•7001035870šŸ–•Riya Kappor Top Call Girl ...
ā™›VVIP Hyderabad Call Girls ChintalkuntašŸ–•7001035870šŸ–•Riya Kappor Top Call Girl ...ā™›VVIP Hyderabad Call Girls ChintalkuntašŸ–•7001035870šŸ–•Riya Kappor Top Call Girl ...
ā™›VVIP Hyderabad Call Girls ChintalkuntašŸ–•7001035870šŸ–•Riya Kappor Top Call Girl ...astropune
Ā 
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...Miss joya
Ā 

Recently uploaded (20)

Call Girl Number in Panvel MumbaišŸ“² 9833363713 šŸ’ž Full Night Enjoy
Call Girl Number in Panvel MumbaišŸ“² 9833363713 šŸ’ž Full Night EnjoyCall Girl Number in Panvel MumbaišŸ“² 9833363713 šŸ’ž Full Night Enjoy
Call Girl Number in Panvel MumbaišŸ“² 9833363713 šŸ’ž Full Night Enjoy
Ā 
VIP Call Girls Indore Kirti šŸ’ššŸ˜‹ 9256729539 šŸš€ Indore Escorts
VIP Call Girls Indore Kirti šŸ’ššŸ˜‹  9256729539 šŸš€ Indore EscortsVIP Call Girls Indore Kirti šŸ’ššŸ˜‹  9256729539 šŸš€ Indore Escorts
VIP Call Girls Indore Kirti šŸ’ššŸ˜‹ 9256729539 šŸš€ Indore Escorts
Ā 
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
Ā 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Ā 
sauth delhi call girls in Bhajanpura šŸ” 9953056974 šŸ” escort Service
sauth delhi call girls in Bhajanpura šŸ” 9953056974 šŸ” escort Servicesauth delhi call girls in Bhajanpura šŸ” 9953056974 šŸ” escort Service
sauth delhi call girls in Bhajanpura šŸ” 9953056974 šŸ” escort Service
Ā 
CALL ON āž„9907093804 šŸ” Call Girls Baramati ( Pune) Girls Service
CALL ON āž„9907093804 šŸ” Call Girls Baramati ( Pune)  Girls ServiceCALL ON āž„9907093804 šŸ” Call Girls Baramati ( Pune)  Girls Service
CALL ON āž„9907093804 šŸ” Call Girls Baramati ( Pune) Girls Service
Ā 
Call Girl Coimbatore Prishaā˜Žļø 8250192130 Independent Escort Service Coimbatore
Call Girl Coimbatore Prishaā˜Žļø  8250192130 Independent Escort Service CoimbatoreCall Girl Coimbatore Prishaā˜Žļø  8250192130 Independent Escort Service Coimbatore
Call Girl Coimbatore Prishaā˜Žļø 8250192130 Independent Escort Service Coimbatore
Ā 
Escort Service Call Girls In Sarita Vihar,, 99530Ā°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530Ā°56974 Delhi NCREscort Service Call Girls In Sarita Vihar,, 99530Ā°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530Ā°56974 Delhi NCR
Ā 
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls ServiceKesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
Ā 
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
Ā 
(šŸ‘‘VVIP ISHAAN ) Russian Call Girls Service Navi MumbaišŸ–•9920874524šŸ–•Independent...
(šŸ‘‘VVIP ISHAAN ) Russian Call Girls Service Navi MumbaišŸ–•9920874524šŸ–•Independent...(šŸ‘‘VVIP ISHAAN ) Russian Call Girls Service Navi MumbaišŸ–•9920874524šŸ–•Independent...
(šŸ‘‘VVIP ISHAAN ) Russian Call Girls Service Navi MumbaišŸ–•9920874524šŸ–•Independent...
Ā 
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
Ā 
Bangalore Call Girls Marathahalli šŸ“ž 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli šŸ“ž 9907093804 High Profile Service 100% SafeBangalore Call Girls Marathahalli šŸ“ž 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli šŸ“ž 9907093804 High Profile Service 100% Safe
Ā 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Ā 
Call Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls Service
Call Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls ServiceCall Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls Service
Call Girls Service Bellary Road Just Call 7001305949 Enjoy College Girls Service
Ā 
Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...
Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...
Call Girls Service Pune Vaishnavi 9907093804 Short 1500 Night 6000 Best call ...
Ā 
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy GirlsCall Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Ā 
VIP Call Girls Pune Sanjana 9907093804 Short 1500 Night 6000 Best call girls ...
VIP Call Girls Pune Sanjana 9907093804 Short 1500 Night 6000 Best call girls ...VIP Call Girls Pune Sanjana 9907093804 Short 1500 Night 6000 Best call girls ...
VIP Call Girls Pune Sanjana 9907093804 Short 1500 Night 6000 Best call girls ...
Ā 
ā™›VVIP Hyderabad Call Girls ChintalkuntašŸ–•7001035870šŸ–•Riya Kappor Top Call Girl ...
ā™›VVIP Hyderabad Call Girls ChintalkuntašŸ–•7001035870šŸ–•Riya Kappor Top Call Girl ...ā™›VVIP Hyderabad Call Girls ChintalkuntašŸ–•7001035870šŸ–•Riya Kappor Top Call Girl ...
ā™›VVIP Hyderabad Call Girls ChintalkuntašŸ–•7001035870šŸ–•Riya Kappor Top Call Girl ...
Ā 
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
Ā 

MedGem Indirect Calorimeter Training Slides

  • 1. MedGem Training Module This course is designed to teach you how to use the MedGem device to personalize your clientsā€™ nutrition and exercise programs. At the completion of this training, you will be able to: ā€¢ Define and understand resting metabolic rate (RMR). ā€¢ Be able to perform a measurement using the MedGem device. ā€¢ Interpret and apply the measurement. ā€¢ Answer frequently asked questions. Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
  • 2. Resting Metabolic Rate The Missing Link to Weight Management Your resting metabolic rate is the number of calories your body burns each day with no activity. It makes up the majority (60-75% in healthy individuals and 100% in critically ill) of calories burned each day. Resting metabolic rate is unique to each individual and is affected by a number of factors: ā€¢ Age ā€¢ Body Weight ā€¢ Medications ā€¢ Fever and/or Infection ā€¢ Weight Loss 1 Danforth, E. LifeSci 28:1891-1897, 1981. 2 Fung EB, AACN Clinical Issues 11(4): 480-497, 2000. ā€¢ Gender ā€¢ Genetics ā€¢ Hormones ā€¢ Exercise ā€¢ Nutritional Supplements Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
  • 3. Predictive Equations are inaccurate for Healthy and III Individuals Healthy Individuals RMR can vary significantly from person to person ā€“ up to 900 calories per day in healthy people. Ill Individuals Several studies have indicated predictive equations are significantly different than measured RMR. Equations can overestimate RMR by as much as 59% and underestimate as much as 30% in sick individuals. This variability can result in very different outcomes for individuals. ADAā€™s Adult & Pediatric Weight Management Guidelines: ā€œIf possible, Indirect Calorimetry, should be used for assessing RMR when developing energy plans.ā€ 1. Foster, G, et al. Metabolism, 1988; 37(5), 467-472 2. Mann et al, Crit Care Med 1985: 13: 173-177 3. Weissman et al, Journal of Anesthesiology 1986: 64: 673 679 4. ABA-EBL: www.ADAevidencelibrary.com Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
  • 4. RMR Varies Among Healthy and Ill Individuals Foster, G, et al. Metabolism, 37(5): 467-472, 1988 Hunter D, et al. Br J Surg, 75: 875-878,1988, Welis, P. Am J Clin, 88; 959-970, 2008, Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
  • 5. Clinical Benefits of Measuring RMR Measuring RMR enables you to gauge the effects of the clientā€™s condition on his/her calorie needs and adjust your nutrition recommendations accordingly. Conditions that impact your RMR include: ā€¢ disease severity and point of time in disease process ā€¢ change in medications/dosage ā€¢ change in dietary regimen ā€¢ change in weight (+/- 5%) Benefits Providing appropriate feeding minimizes the percentage of clients who are overfed and underfed. Patients following a diet plan established from measured RMR lose more weight compared to diet plans that estimate RMR. Measuring RMR to establish appropriate nutritional needs can result in direct cost savings by avoiding unnecessary nutritional support. 1. Alberda et al., Nutri Clin Pract 17: 38-42, 2002 2. McDoniel, et al., Obes Facts (1)6; 298-304, 2008. 3. McClave et al, JPEN 22; 375-381, 1998. Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
  • 6. Candidates for Metabolic Measurement Some facilities use the device on every individual as part of their initial nutrition assessment to determine baseline calorie needs; others choose to use the device on individuals that meet certain criteria that put them at high risk for complications associated with overweight and obesity. Sample criteria for determining medical necessity for diagnostic procedure (CPT 94690). Average payer reimbursement is $50-$80. ā€¢ 414 Chronic Pulmonary HT Disease ā€¢ 428 Heart Failure* ā€¢ 493 Asthma* ā€¢ 496 COPD* ā€¢ 51.8 Lung Involvement in other disease* ā€¢ 780.5 Obstructive Sleep Apnea ā€¢ 786.5 Dyspnea* ā€¢ 786.9 Hypoventilation Syndrome* ā€¢ 244.9 Hypothyroidism ā€¢ 250 Diabetes Mellitus ā€¢ 272.1 Hypercholesterolemia ā€¢ 22.4 Hyperlipidemia ā€¢ 277.7 Metabolic Syndrome ā€¢ 278 Obesity ā€¢ 401 Hypertension ā€¢ 414 Cardiovascular Disease *CMS indicates medical necessity Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
  • 7. Performing a Measurement About the MedGem Device The MedGem is a hand-held indirect calorimeter that measures oxygen consumption to determine resting metabolic rate. The MedGem device is easy to use, providing measurements quickly and accurately. Useful as a tool for monitoring changes in metabolism, the MedGem gives you important information to help your clients achieve their personal weight management, nutrition, and exercise goals. ā€œThe validation results suggest the MedGem can be used to accurately measure oxygen consumption and RMR in long-term healthcare facilities, low-capacity hospitals, athletic training centers, health clubs, and weight loss clinics.ā€ Richard Branson, MS, RTT University of Cincinnati Medical Center, Cincinnati, OH 1. RMR is calculated using the Weir equation and a constant RQ value of 0.85 (RMR=6.931 X VO2), Weir,JB., New Methods for Calculating Metabolic Rate with Special Reference to Protein Metabolism. J Physiology, 1949, 109; 1-9. Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
  • 8. Getting Acquainted with MedGem ā€¢ The start button and LED indicator light are located on top. ā€¢ The air vent is located on the bottom of the device. ā€¢ The device is powered by a DC power supply, which plugs into the side of the MedGem. Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
  • 9. MedGem Measurement Protocol Acceptable Conditions: 1. Measurement is performed in quiet environment. 2. Individual rests for 10-15 minutes prior to measurement. 3. Individual has not exercised for at least 4 hours. 4. Individual has not consumed caffeine for at least 4 hours. 5. Individual has not consumed nutrition supplements for at least 4 hours. 6. Individual has not used nicotine for at least 1 hour. 7. Individual is not taking acute medications. Chronic medications are OK. 8. Individual may have water prior to measurement. Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
  • 10. 1. Plug cord in wall outlet. 2. Plug cord in device. 3. Device will beep. LED light turns from red to green to orange. Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
  • 11. 4. Place MedGem on a table away from fans or vents. 5. Insert mouthpiece into the MedGem. 6. Push orange button to begin self-calibration process. Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
  • 12. 7. Self-calibration completed when device beeps and LED light turns to green. 8. Hand device to the client and ensure there is a tight seal around the mouthpiece. 9. Client should be seated or in a supine position and breathing normally into device. 10. The MedGem will beep when the measurement is completed. 11. Record RMR and discard mouthpiece and noseclip. Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
  • 13. Tracking the Measurements Codes beginning with ā€œC-ā€ (ie C-10) refer to the remaining number of measurements that may be performed with a particular MedGem device. (C-10 means you have 10 remaining Ā© 2014 All Rights Reserved. No duplication and distribution without permission! measurements). A two-tone audible alert, combined with a flashing amber and green light indicates the MedGem is low on measurements. To acknowledge this warning and begin self-calibrating the device, press the Start button twice. When you receive a low measurement warning, contact Microlife Medical Home Solutions to arrange for delivery of a replacement.
  • 14. Error Codes Errors beginning with ā€œ0ā€ (ie 01, 05) indicate user errors. Refer to the user manual for instructions. Solutions: 1. Ensure the flowtube is firmly placed into the device. 2. Ensure the single-use mouthpiece is firmly inserted into the flowtube. 3. Ensure the client has a proper seal around the mouthpiece and noseclip is properly positioned on their nose. The client should only be able to breathe through their mouth. 4. Repeat the measurement. Errors beginning with any other number (ie 11) may indicate possible internal error. Contact Microlife Medical Home Solutions for instructions. Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
  • 15. Caring for the MedGem Device Here are some guidelines for properly caring for the Ā© 2014 All Rights Reserved. No duplication and distribution without permission! device. Do not expose the inside of the MedGem to cleaning solutions or water. If necessary, the outside of the MedGem can be cleaned with any of the following: ā€¢ Soft, clean cloth ā€¢ Clorox disinfecting wipes ā€¢ Soap and water Store you MedGem in its case when not in use.
  • 16. The Missing Link Through self-monitoring, health professionals have been able to help clients track how many calories they eat. But thatā€™s only half of the energy balance equation. Your clients need to know how many calories they burn. Purposeful exercise and lifestyle/occupation activities contribute to the number of calories burned by an individual. Resting metabolic rate, however, accounts for the majority of calories burned and has been the missing link in nutrition programs. Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
  • 17. ā€œIs my RMR good or bad?ā€ Unlike other health assessments (ie: blood pressure, etc), RMR assessments are not good or bad. RMR is unique to your clients physiology. Therefore, RMR may be higher or lower than other like individuals. The MedGem Analyzer will provide you metabolic range for each client. The metabolic range is based on the level of oxygen consumption and current body weight of your client. Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
  • 18. ā€œWhat do I do with the RMR number?ā€ Now you have information you need to develop a balanced, personalized nutrition and exercise plan for your clients. Microlife Medical Home Solutions, Inc. offers several options for developing a calorie budget for individuals. Please visit our website at www.mimhs.com to learn how the MedGem Analyzer can help you develop a nutritional plan for your clients. Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
  • 19. ā€œWhat do I do with the measurement result?ā€ Weight Loss Case Study Brian has a BMI of 32.3 kg/m. He was recently diagnosed with non-insulin dependent diabetes and hypertension. His physician recommends Brian lose 25 pounds by his next appointment in ~3 months. The MedGem Analzyer will generate an individual treatment plan for his weight loss goal. ā€¢ Current weight: 245 lbs ā€¢ Goal weight: 220 lbs ā€¢ RMR: 2300 kcals VO2: 332 ml/min ā€¢ Occupation: Computer Programmer (~45 hours/wk) ā€¢ Goal date: 12 weeks ā€¢ Exercise plan: Elliptical Trainer and/or walking moderate-intensity for 30 minutes, 5 days per week Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
  • 20. ā€œWhat do I do with the measurement result?ā€ Weight Loss Case Study Personalized Program: For Brian to lose 25 lbs, he will need to consume 1685 calories per day. ā€¢ Daily Energy Expenditure: 2685 kcals (determined by Analyzer) ā€¢ Rate of Weight Loss: -1000 kcal (2 lbs per week) ā€¢ Daily Calorie Needs for weight loss: 1685 kcals With Brianā€™s exercise plan, he will burn an additional 245 kcals. Therefore, Brian could consume 1930 kcals on the days he exercises and still reach his weight goal by his goal date. ā€¢ Daily Calorie Needs for weight loss 1685 kcals ā€¢ Exercise Program: +245 kcals ā€¢ Daily Calorie Needs with exercise 1930 kcals Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
  • 21. Why Re-Measure? Resting metabolic rate can change significantly during a weight management program. Re-measuring RMR will ensure that your clients are on a successful track for long-term weight management success or appropriate recovery. You should re-measure individuals: ā€¢ +/- 5% - 10% weight change during a weight management program ā€¢ If no weight loss has occurred in the first few weeks ā€¢ If individual has reached a weight loss plateau ā€¢ When there is a change in medications or a medical condition ā€¢ When there is no response to nutritional therapy Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
  • 22. Avoiding Common Pitfalls Problem: Lost Readings Tim, a weight loss client, arrives for his RMR appointment. You seat him in a quiet room and allow him to relax as you review the measurement procedures and set up the MedGem device. You begin the measurement after calibrating the device. About 5 minutes later, the MedGem beeps indicating the measurement is complete. You remove the noseclip and discard the single use mouthpiece. As you reach for the MedGem, you accidentally disconnect the power cord and the device shuts off before you have a chance to record the measurement reading. Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
  • 23. Avoiding Common Pitfalls Solution: Lost Readings To recover Timā€™s RMR reading, simply plug the power cord back into the MedGem. After the device powers up, the last RMR measurement will be displayed. Record the reading. Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
  • 24. Avoiding Common Pitfalls Problem: Inaccurate Measurements Mona is anxious to lose the weight she gained from her first pregnancy. She decides to have her RMR measured. This allows you to develop a personalized program to help her reach her weight-loss goal. When she arrives, you seat her in a quiet room and allow her to relax as your review the measurement procedures and set up the MedGem device. Then, you perform a measurement. You make note of the reading, which is 2980, and discard the single-use mouthpiece and noseclip. The reading is suspiciously high. What could have gone wrong? Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
  • 25. Avoiding Common Pitfalls Solution: Inaccurate Measurements Enter the clientā€™s RMR measurement into the MedGem Analyzer. Compare the clientā€™s RMR to the listed normal range. Refer to the listed normal range in the MedGem Analyzer. When using the MedGem to measure a clientā€™s RMR, it is important to have the client adhere to proper resting conditions. If the client has performed physical activity, taken acute medications, has eaten, or has not rested for 10-15 minutes prior to the measurement, the MedGem measurement will not be accurate to their actual RMR. Always determine that the resting conditions were followed before performing a measurement. Developing a weight loss plan based on this inaccurate reading will prevent the client from reaching their goals. Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
  • 26. Avoiding Common Pitfalls Problem: Undetected Air Leak Brian enrolled in your weight management program. He is 25 pounds overweight and needs to lower his cholesterol levels and blood pressure. You seat him in a quiet room and perform the MedGem measurement under acceptable conditions. You make a note of the reading, which is 1400, then you discard this single-use mouthpiece and noseclip. Both you and Brian are suspicious that this RMR reading may be too low since he weighs 315 pounds. What could have gone wrong? Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
  • 27. Avoiding Common Pitfalls Solution: Undetected Air Leak Brianā€™s reading may be inaccurate due to an air leak that did not cause an error message to display. If an air leak occurs, some of the air is not measured by the device and Brianā€™s actual RMR would therefore not be determined. When using the MedGem to measure a clientā€™s RMR, it is important to insure that all resting conditions are followed. Your client should rest for 10-15 minutes prior to starting the measurement. Measure your client if he/she has: ā€¢ Not eaten or had anything to drink (water is OK) 4 hours prior ā€¢ Not exercised 4 hour prior ā€¢ Not consumed caffeine or stimulatory nutritional supplements 4 hours prior ā€¢ Not used nicotine for 1 hour prior ā€¢ Not taken acute medications (ie: antibiotics, pain medication) Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
  • 28. Your Clients Will Have Questions! Your clients will have questions about metabolism, the MedGem device, and RMR. Microlife Medical Home Solutions has anticipated many of the questions that will be asked by you and your clients. The answers are provided in the MedGem User Guide. Read the document: MedGem User Guide for answers to common questions. If you need further assistance, visit our website at www.mimhs.com. Ā© 2014 All Rights Reserved. No duplication and distribution without permission!
  • 29. Congratulations! Youā€™ve reached the end! Donā€™t Forget about our additional tools Ā© 2014 All Rights Reserved. No duplication and distribution without permission! Measure Use the MedGem and MedGem Analyzer software to determine RMR and assess energy balance for a personalized nutrition program. Monitor Have clients self-monitor their food intake and exercise using MiHealthLog health management application. Modify Help you patients and clients adjust their nutrition and exercise plan by monitoring their daily health, nutrition, and activity goals with MiHealthCoach.