A Single Ventilator for Multiple Simulated
Patients to Meet Disaster Surge
Greg Neyman, MD, Charlene Babcock Irvin, MD
Abstract
Objectives: To determine if a ventilator available in an emergency department could quickly be modified to
provide ventilation for four adults simultaneously.
Methods: Using lung simulators, readily available plastic tubing, and ventilators (840 Series Ventilator;
Puritan-Bennett), human lung simulators were added in parallel until the ventilator was ventilating the
equivalent of four adults. Data collected included peak pressure, positive end-expiratory pressure, total
tidal volume, and total minute ventilation. Any obvious asymmetry in the delivery of gas to the lung simu-
lators was also documented. The ventilator was run for almost 12 consecutive hours (5.5 hours of pressure
control and more than six hours of volume control).
Results: Using readily available plastic tubing set up to minimize dead space volume, the four lung simu-
lators were easily ventilated for 12 hours using one ventilator. In pressure control (set at 25 mm H2O),
the mean tidal volume was 1,884 mL (approximately 471 mL/lung simulator) with an average minute ven-
tilation of 30.2 L/min (or 7.5 L/min/lung simulator). In volume control (set at 2 L), the mean peak pressure
was 28 cm H2O and the minute ventilation was 32.5 L/min total (8.1 L/min/lung simulator).
Conclusions: A single ventilator may be quickly modified to ventilate four simulated adults for a limited
time. The volumes delivered in this simulation should be able to sustain four 70-kg individuals. While
further study is necessary, this pilot study suggests significant potential for the expanded use of a single
ventilator during cases of disaster surge involving multiple casualties with respiratory failure.
ACADEMIC EMERGENCY MEDICINE 2006; 13:1246–1249 ª 2006 by the Society for Academic Emergency
Medicine
Keywords: disaster, ventilator, respiratory failure, surge capacity
A
fter the events of September 11, 2001, and the
recent hurricanes in the Gulf Coast, there has
been a focus on anticipating the need for medi-
cal care for large numbers of victims.1
Addressing surge
capacity requires a multitiered approach involving local
and federal agencies as well as resource management
(i.e., personnel, patient space, supplies, and special
equipment). Recent experiences have shown that hospi-
tals may be rapidly extended to operate at 120%–130%
capacity.2
Depending on the nature of the disaster, many other-
wise plentiful hospital supplies, such as ventilators, may
suddenly become insufficient to support the demand.3
In the event of a large influx of patients in respiratory
distress (e.g., a large outbreak of botulism), the number
of ventilators available may not be enough to support
all of the patients.
While government resources would eventually be
available, there may be a time when hospitals will need
to provide ventilatory support to a greater number of pa-
tients than the available number of ventilators. Manual
ventilation (‘‘bagging’’) is possible, but it is possible that
the additional personnel required would not be available.
Some institutions have begun to stockpile disposable
automatic ventilators for use in the event of a disaster.4
One commercially available device is Vortran’s Auto-
matic Resuscitator, a single-patient, single-use, pressure-
powered ventilator. It could be kept in stock and deployed
when necessary. It runs on wall oxygen (50 psi) and
is pressure cycled off the wall oxygen source, and seven
can be run simultaneously off a single oxygen supply
line by using the Vortran E-Vent Case multioutlet manifold
device. One advantage is that this type of automatic venti-
lator is gas driven (from the oxygen source) and requires
no electricity (this may be an advantage in certain types
of disasters), is disposable, and is modifiable to provide
ventilation for up to seven individuals off one oxygen
From the Department of Emergency Medicine, St. John’s Hospi-
tal and Medical Center (GN, CBI), Detroit, MI.
Received March 3, 2006; revisions received May 4, 2006, and
May 7, 2006; accepted May 8, 2006.
Lung simulators were donated by Puritan-Bennett.
Address for correspondence and reprints: Greg Neyman, MD,
Department of Emergency Medicine, St. John Hospital and
Medical Center, 22101 Moross Road, Detroit, MI 48236. E-mail:
greg.neyman@stjohn.org.
ISSN 1069-6563 ª 2006 by the Society for Academic Emergency Medicine
PII ISSN 1069-6563583 doi: 10.1197/j.aem.2006.05.0091246
source. However, this device requires anticipatory pur-
chase, and it lacks the computerized monitoring that stan-
dard ventilators have, necessitating more intensive staff
support.5
Another option for providing increased ventilation ca-
pacity includes a proprietary device (patent pending) that
is essentially a control system for splitting one ventilator
to provide ventilation for two patients.6
This does not
have the advantages of the previously described dispos-
able ventilator, because it would rely on an electrically
driven ventilator. Additionally, it would require the pur-
chase of additional equipment (the proprietary control
system).
Although larger urban hospitals may be able to justify
the resources to stockpile disposable ventilators, smaller
hospitals may not. In the event of a need for more venti-
lators than are currently available in a hospital, it may be
valuable to explore methods to maximally utilize the ven-
tilators that are already available. Hospitals do have gen-
erators and would likely be able to support electrical
service for a short time after the disaster. Given this po-
tential, it may be appropriate to consider a simple modi-
fication of the currently available hospital ventilators to
provide more patients with ventilator support. Our hy-
pothesis was that using simple, rapidly deployable mod-
ifications, a single ventilator could be used to ventilate
multiple casualties when the number of victims exceeds
the number of ventilators.
METHODS
Study Design
This was a simulator-based pilot study. This study was
considered exempt after review by the institutional
review board.
Study Protocol
Four sets of standard ventilator tubing (Hudson) were
connected to a single ventilator (Puritan-Bennett, 840 se-
ries) via two flow splitters (one on the patient inflow limb
of the circuit, and one on the patient exhaust limb). Each
flow splitter was constructed of three Briggs T-tubes
with included connection adapters (Hudson) (Figure 1),
with the valves removed. The Briggs T-tube is utilized
clinically (and generally available) for flow-by oxygen or
humidity for a patient with an endotracheal or tracheos-
tomy tube, or for in-line aerosol treatments of ventilated
patients.
The T-tubes were arranged so that the two side ports
of a central T-tube were attached to the bottom ports
of the two side T-tubes via adapters that come with the
T-tube. The final configuration of the three T-tubes is
seen in Figure 2 (with a trimmed section of standard ven-
tilation tubing at the hub for connection to the ventila-
tor); it allowed for air flowing from the ventilator to be
split evenly to four simulated patients and for the air
returning from the four patients to flow back into the
one exhaust port on the ventilator.
The ventilator tubing was run from the inflow splitter
to the outflow splitter, with four test lungs (Puritan-Ben-
nett) in the center. The test lungs were used to simulate
one patient each on the modified ventilator circuit. The
final configuration was a simulation of four patients on
a single ventilator in parallel operation (Figure 3).
To test this circuit, a time frame was arbitrarily chosen
as approximately six hours. There were two reasons for
this. First, this is a simple feasibility study, and we would
expect someone to inspect the system at least once in six
hours if ever used in a real disaster. Second, we realize
that beyond this feasibility study, animal studies are
needed, and this could allow for observation of the func-
tion of this circuit for a longer period. Finally, in many
potential disaster situations, by six hours additional sup-
port may be available.
Pressure control operation was randomly selected (via
coin toss) to precede volume control. To approximate
physiologic parameters, the ventilator settings were di-
aled to a peak pressure of 25 cm H2O, 0 cm of positive
end-expiratory pressure, and a respiratory rate of 16
breaths/min. The ventilator software chose an inspira-
tory/expiratory ratio of 1:2 automatically. After cumula-
tive random interval inspections, total pressure control
Figure 1. Briggs T-tube with included connection adapter.
Figure 2. Flow splitter configuration of three T-tubes with
connection adapters (shown here with trimmed section of
ventilator tubing).
ACAD EMERG MED  November 2006, Vol. 13, No. 11  www.aemj.org 1247
operation was 5 hours 33 minutes. Volume control settings
of 2,000 mL tidal volume (500 mL per test lung) and a respi-
ratory rate of 16 breaths/min were chosen to approximate
physiologic parameters. The ventilator software chose
an inspiratory/expiratory ratio of 1:1 automatically. After
cumulative random interval inspections, total volume
control operation was 6 hours 11 minutes.
Measurements
During operation, the circuit was inspected at random in-
tervals in between examinations of patients in a busy
metropolitan emergency department to simulate the in-
tervals that the circuit would be inspected in a mass casu-
alty event. The inspections occurred approximately every
23 (Æ18) minutes, and the ventilator display readouts
were recorded. Simultaneously, the lungs were subjec-
tively inspected for symmetry of excursion and evidence
of respiratory stacking. Specifically, the examiner moni-
tored for asymmetric inflation of individual test lungs
and incomplete deflation before subsequent inflation.
RESULTS
After the configuration was sealed, the ventilator system
did not alarm. Visual inspection showed roughly equiva-
lent excursion of all lung models. No respiratory stacking
was seen. Averages of ventilator display readout sam-
plings over thecourse of thestudy are presentedin Table 1.
DISCUSSION
A four-patient configuration operated successfully on a
single ventilator for almost 12 hours. Pressures did not
exceed 35 cm H2O. Airway pressures beyond 35 mm
H2O are associated with ventilator-induced lung injury.7
Individual tidal volumes reached 471–507 mL, which ap-
proximates 7 mL/kg for a 70-kg individual. Studies have
shown that ventilation with 6–8 mL/kg is associated
with improved outcome in injured lungs.8
No evidence
of respiratory stacking or preferential filling of individual
lung simulators was observed.
LIMITATIONS
The chief limitation of this study is that it is a simula-
tor study. Therefore, only successful physical ventilation
could be demonstrated. Adequate oxygenation and the
potential for ventilator-associated lung injury could not
be addressed. The presumption of equal ventilation to all
four lung simulators presumed equal lung physiology. A
patient with asthma with greater resistance to ventilations
may not receive equal ventilation with this system. Further
animal studies are necessary to address this concern.
The inability to directly measure volumes delivered to the
individual test lungs may bias the results and thus change
the actual method in which this ventilator configuration
would be deployed. Potential infectious complications
from sharing one ventilator were not investigated. Again,
further study in this area would be beneficial.
Because this was a pilot study, further research is in-
dicated to test the efficacy and safety of the modified
circuit. Replication of the study in an animal model is in-
dicated. Ventilator software may allow for ventilation of
more patients than was explored in this study. This may
be an important consideration in usefulness of ventila-
tors in the disaster situation. Finally, development of
quantitative measurement techniques of individual tidal
volumes transferred would enhance further research
efforts as well as clinical delivery.
CONCLUSIONS
This pilot study suggests that the physics of a ventilator/
patient circuit could accommodate more than one
Figure 3. Final configuration of multiple patient ventilation
(*outflow limbs darkly stained for purposes of graphic clar-
ity only).
Table 1
Pressure and Flow Characteristics of Lung Simulators in Multiple
Patient Ventilation Configuration as an Average of Random
Collections
Pressure
Control
Volume
Control
Total time (hr/min) 5:33 6:11
Peak pressure Æ SD (cm H2O) 25 Æ 0 28 Æ 2
Mean pressure Æ SD (cm H2O) 8.7 Æ 0.1 10.4 Æ 0.5
Peak expiratory end pressure Æ
SD (cm H2O)
1.3 Æ 0.1 1.8 Æ 0.1
Tidal volume Æ SD (mL) 1,884 Æ 90 2029 Æ 3
Estimated tidal volume per
simulator Æ SD (mL)
471 Æ 22 507 Æ 1
Minute ventilation Æ SD (L) 30.2 Æ 1.4 32.5 Æ 0.1
Estimated minute ventilation
per simulator Æ SD (L)
7.5 Æ 0.4 8.1 Æ 0.0
Respiratory rate (breaths/min) 16 16
Inspiratory/expiratory ratio 1:2 1:1
1248 Neyman and Irvin  SINGLE VENTILATOR FOR MULTIPLE SIMULATED PATIENTS
patient. In a catastrophic situation, when there are more
patients who require ventilators than there are ventila-
tors available, simple modification of the ventilator cir-
cuit could help absorb the extra burden.
The authors thank the entire Respiratory Therapy Department at
St. John’s Medical Center (Detroit, MI) and Dr. Peter Hoffmann
(Physics Department, Wayne State University, Detroit, MI) for
consultation regarding fundamental physics.
References
1. American College of Emergency Physicians. Health
care system surge capacity recognition, prepared-
ness, and response. Ann Emerg Med. 2005; 45:
239.
2. Hick JL, Hanfling D, Burstein JL, et al. Health care fa-
cility and community strategies for patient care surge
capacity. Ann Emerg Med. 2004; 44:253–61.
3. Hick JL, O’Laughlin DT. Concept of operations for
triage of mechanical ventilation in an epidemic. Acad
Emerg Med. 2006; 13:223–9.
4. Yale New Haven Health. Emergency Preparedness
Briefing. Available at: http://yalenewhavenhealth.org/
emergency/commu/briefings/EPBriefingV3I49.pdf.
Accessed Apr 25, 2006.
5. VORTRAN Medical Technology 1, Inc. VORTRAN
Automatic Resuscitator (VAR) User’s Guide. Available
at: http://www.vortran.com/VAR_User_Guide_2005A.
pdf. p 3. Accessed May 8, 2006.
6. Darowski M, Englisz M. Artificial ventilation of the lungs
for emergencies. Front Med Biol Eng. 2000; 10:177–83.
7. MacIntyre NR. Principles of mechanical ventilation. In:
Mason RJ, Murray JF, Broaddus VC, Nadel JA, eds.
Murray  Nadel’s Textbook of Respiratory Medicine.
4th ed. Philadelphia, PA: Harcourt Health Sciences
Group, 2005, p. 2342.
8. Michaels AJ. Management of post traumatic respira-
tory failure. Crit Care Clin. 2004; 20:83–99.
ACAD EMERG MED  November 2006, Vol. 13, No. 11  www.aemj.org 1249

More Related Content

What's hot

AIRWAYS-2: Effect of a Strategy of a Supraglottic Airway Device vs Tracheal I...
AIRWAYS-2: Effect of a Strategy of a Supraglottic Airway Device vs Tracheal I...AIRWAYS-2: Effect of a Strategy of a Supraglottic Airway Device vs Tracheal I...
AIRWAYS-2: Effect of a Strategy of a Supraglottic Airway Device vs Tracheal I...Intensive Care Society
 
The menagerie of monitoring tools by Professor Jean-Louis Teboul
The menagerie of monitoring tools by Professor Jean-Louis TeboulThe menagerie of monitoring tools by Professor Jean-Louis Teboul
The menagerie of monitoring tools by Professor Jean-Louis TeboulCICM 2019 Annual Scientific Meeting
 
2013 oems protocol_changes_gutman
2013 oems protocol_changes_gutman2013 oems protocol_changes_gutman
2013 oems protocol_changes_gutmanCraigomal
 
Monitoring Fluid Responsiveness in ICU
Monitoring Fluid Responsiveness in ICUMonitoring Fluid Responsiveness in ICU
Monitoring Fluid Responsiveness in ICUYazan Kherallah
 
HFOV Gas Exchange
HFOV Gas ExchangeHFOV Gas Exchange
HFOV Gas ExchangeJPriest4
 
Letter 2 Dr. Paul W. Corey
Letter 2 Dr. Paul W. CoreyLetter 2 Dr. Paul W. Corey
Letter 2 Dr. Paul W. Coreyfast.track
 
Ultrasonography evaluation during the weaning process
Ultrasonography evaluation during the  weaning processUltrasonography evaluation during the  weaning process
Ultrasonography evaluation during the weaning processFadel Omar
 
Norsk et al. Fluid Shifts 2015
Norsk et al. Fluid Shifts 2015Norsk et al. Fluid Shifts 2015
Norsk et al. Fluid Shifts 2015Peter Norsk
 
Controversies in critical care
Controversies in critical careControversies in critical care
Controversies in critical careBalwinderRekhi
 
ARDS Dr. MADHU KIRAN, MD. PULMONOLOGY
ARDS  Dr. MADHU KIRAN, MD. PULMONOLOGYARDS  Dr. MADHU KIRAN, MD. PULMONOLOGY
ARDS Dr. MADHU KIRAN, MD. PULMONOLOGYDr. Madhu Kiran
 
Trasnscatheter Aortic Valve Replacement and Anesthesia
Trasnscatheter Aortic Valve Replacement and AnesthesiaTrasnscatheter Aortic Valve Replacement and Anesthesia
Trasnscatheter Aortic Valve Replacement and AnesthesiaCarlos D A Bersot
 
Chest compression alone cardiopulmonary resuscitation
Chest compression alone cardiopulmonary resuscitationChest compression alone cardiopulmonary resuscitation
Chest compression alone cardiopulmonary resuscitationEmergency Live
 

What's hot (20)

Misclassification of Pulmonary Arterial Hypertension 4.16.08
Misclassification of Pulmonary Arterial Hypertension 4.16.08Misclassification of Pulmonary Arterial Hypertension 4.16.08
Misclassification of Pulmonary Arterial Hypertension 4.16.08
 
AIRWAYS-2: Effect of a Strategy of a Supraglottic Airway Device vs Tracheal I...
AIRWAYS-2: Effect of a Strategy of a Supraglottic Airway Device vs Tracheal I...AIRWAYS-2: Effect of a Strategy of a Supraglottic Airway Device vs Tracheal I...
AIRWAYS-2: Effect of a Strategy of a Supraglottic Airway Device vs Tracheal I...
 
Unger airway dem2014_final
Unger airway dem2014_finalUnger airway dem2014_final
Unger airway dem2014_final
 
The menagerie of monitoring tools by Professor Jean-Louis Teboul
The menagerie of monitoring tools by Professor Jean-Louis TeboulThe menagerie of monitoring tools by Professor Jean-Louis Teboul
The menagerie of monitoring tools by Professor Jean-Louis Teboul
 
2013 oems protocol_changes_gutman
2013 oems protocol_changes_gutman2013 oems protocol_changes_gutman
2013 oems protocol_changes_gutman
 
Copd 9-837
Copd 9-837Copd 9-837
Copd 9-837
 
FINAL PRESENTATION
FINAL PRESENTATIONFINAL PRESENTATION
FINAL PRESENTATION
 
Poster sta-2011-final
Poster sta-2011-finalPoster sta-2011-final
Poster sta-2011-final
 
Monitoring Fluid Responsiveness in ICU
Monitoring Fluid Responsiveness in ICUMonitoring Fluid Responsiveness in ICU
Monitoring Fluid Responsiveness in ICU
 
HFOV Gas Exchange
HFOV Gas ExchangeHFOV Gas Exchange
HFOV Gas Exchange
 
S436.full
S436.fullS436.full
S436.full
 
Letter 2 Dr. Paul W. Corey
Letter 2 Dr. Paul W. CoreyLetter 2 Dr. Paul W. Corey
Letter 2 Dr. Paul W. Corey
 
Ultrasonography evaluation during the weaning process
Ultrasonography evaluation during the  weaning processUltrasonography evaluation during the  weaning process
Ultrasonography evaluation during the weaning process
 
Norsk et al. Fluid Shifts 2015
Norsk et al. Fluid Shifts 2015Norsk et al. Fluid Shifts 2015
Norsk et al. Fluid Shifts 2015
 
24x31
24x3124x31
24x31
 
Jbp
JbpJbp
Jbp
 
Controversies in critical care
Controversies in critical careControversies in critical care
Controversies in critical care
 
ARDS Dr. MADHU KIRAN, MD. PULMONOLOGY
ARDS  Dr. MADHU KIRAN, MD. PULMONOLOGYARDS  Dr. MADHU KIRAN, MD. PULMONOLOGY
ARDS Dr. MADHU KIRAN, MD. PULMONOLOGY
 
Trasnscatheter Aortic Valve Replacement and Anesthesia
Trasnscatheter Aortic Valve Replacement and AnesthesiaTrasnscatheter Aortic Valve Replacement and Anesthesia
Trasnscatheter Aortic Valve Replacement and Anesthesia
 
Chest compression alone cardiopulmonary resuscitation
Chest compression alone cardiopulmonary resuscitationChest compression alone cardiopulmonary resuscitation
Chest compression alone cardiopulmonary resuscitation
 

Similar to Jaem

Effects of Heat and Humidification on Aerosol Delivery during Auto-CPAP nonin...
Effects of Heat and Humidification on Aerosol Delivery during Auto-CPAP nonin...Effects of Heat and Humidification on Aerosol Delivery during Auto-CPAP nonin...
Effects of Heat and Humidification on Aerosol Delivery during Auto-CPAP nonin...Ahmed Elberry
 
respiratory safety pharmacology ptsm2.pptx
respiratory safety pharmacology ptsm2.pptxrespiratory safety pharmacology ptsm2.pptx
respiratory safety pharmacology ptsm2.pptxPraveen kumar S
 
First year experience
First year experienceFirst year experience
First year experiencedebiemottin
 
Breathing systems open circuit- shoeib
Breathing systems  open circuit- shoeibBreathing systems  open circuit- shoeib
Breathing systems open circuit- shoeibhavalprit
 
CM Ventilator pp.pptx
CM Ventilator pp.pptxCM Ventilator pp.pptx
CM Ventilator pp.pptxSadorYonas
 
Wed risks and-complications_of_mv
Wed risks and-complications_of_mvWed risks and-complications_of_mv
Wed risks and-complications_of_mvangelo_rocha
 
The Oxylator—A Compact And Durable Patient Responsive Ventilation System
The Oxylator—A Compact And Durable Patient Responsive Ventilation SystemThe Oxylator—A Compact And Durable Patient Responsive Ventilation System
The Oxylator—A Compact And Durable Patient Responsive Ventilation SystemJames_DuCanto_MD
 
Embolectomy prior to lung transplant
Embolectomy prior to lung transplantEmbolectomy prior to lung transplant
Embolectomy prior to lung transplantmshihatasite
 
Modeling and optimization of nebulizers' performance in non-invasive ventilat...
Modeling and optimization of nebulizers' performance in non-invasive ventilat...Modeling and optimization of nebulizers' performance in non-invasive ventilat...
Modeling and optimization of nebulizers' performance in non-invasive ventilat...Ahmed Elberry
 
Aerosol therapy for mv patients by Bashaier A. Alyami
Aerosol therapy for mv patients by Bashaier A. AlyamiAerosol therapy for mv patients by Bashaier A. Alyami
Aerosol therapy for mv patients by Bashaier A. AlyamiMaher AlQuaimi
 
Nebulizers and spacers for aerosol delivery through adult nasal cannula at lo...
Nebulizers and spacers for aerosol delivery through adult nasal cannula at lo...Nebulizers and spacers for aerosol delivery through adult nasal cannula at lo...
Nebulizers and spacers for aerosol delivery through adult nasal cannula at lo...Ahmed Elberry
 
Anaesthesia ventilators
Anaesthesia ventilatorsAnaesthesia ventilators
Anaesthesia ventilatorssnigdhanaskar1
 
Respiration Pathway_Dräger Ventilation Story (1).pptx
Respiration Pathway_Dräger Ventilation Story (1).pptxRespiration Pathway_Dräger Ventilation Story (1).pptx
Respiration Pathway_Dräger Ventilation Story (1).pptxAthulRaj169213
 
Cirugia bariatrica reclutamiento y peep
Cirugia bariatrica reclutamiento y peepCirugia bariatrica reclutamiento y peep
Cirugia bariatrica reclutamiento y peepCORRALMTZ
 
3. breathing system bsc
3. breathing system bsc3. breathing system bsc
3. breathing system bscCHERUDUGASE
 
weaning 986.full.pdf
weaning 986.full.pdfweaning 986.full.pdf
weaning 986.full.pdfvignesh79445
 
Chapter 9 room ventilation systems (1)
Chapter 9 room ventilation systems (1)Chapter 9 room ventilation systems (1)
Chapter 9 room ventilation systems (1)Imran Sultan
 
Tracheostomy overview
Tracheostomy overviewTracheostomy overview
Tracheostomy overviewisakakinada
 

Similar to Jaem (20)

Effects of Heat and Humidification on Aerosol Delivery during Auto-CPAP nonin...
Effects of Heat and Humidification on Aerosol Delivery during Auto-CPAP nonin...Effects of Heat and Humidification on Aerosol Delivery during Auto-CPAP nonin...
Effects of Heat and Humidification on Aerosol Delivery during Auto-CPAP nonin...
 
respiratory safety pharmacology ptsm2.pptx
respiratory safety pharmacology ptsm2.pptxrespiratory safety pharmacology ptsm2.pptx
respiratory safety pharmacology ptsm2.pptx
 
First year experience
First year experienceFirst year experience
First year experience
 
Breathing systems open circuit- shoeib
Breathing systems  open circuit- shoeibBreathing systems  open circuit- shoeib
Breathing systems open circuit- shoeib
 
CM Ventilator pp.pptx
CM Ventilator pp.pptxCM Ventilator pp.pptx
CM Ventilator pp.pptx
 
Wed risks and-complications_of_mv
Wed risks and-complications_of_mvWed risks and-complications_of_mv
Wed risks and-complications_of_mv
 
The Oxylator—A Compact And Durable Patient Responsive Ventilation System
The Oxylator—A Compact And Durable Patient Responsive Ventilation SystemThe Oxylator—A Compact And Durable Patient Responsive Ventilation System
The Oxylator—A Compact And Durable Patient Responsive Ventilation System
 
Embolectomy prior to lung transplant
Embolectomy prior to lung transplantEmbolectomy prior to lung transplant
Embolectomy prior to lung transplant
 
Modeling and optimization of nebulizers' performance in non-invasive ventilat...
Modeling and optimization of nebulizers' performance in non-invasive ventilat...Modeling and optimization of nebulizers' performance in non-invasive ventilat...
Modeling and optimization of nebulizers' performance in non-invasive ventilat...
 
Aerosol therapy for mv patients by Bashaier A. Alyami
Aerosol therapy for mv patients by Bashaier A. AlyamiAerosol therapy for mv patients by Bashaier A. Alyami
Aerosol therapy for mv patients by Bashaier A. Alyami
 
A stitch in time saves nine
A stitch in time saves nineA stitch in time saves nine
A stitch in time saves nine
 
Fistulas
FistulasFistulas
Fistulas
 
Nebulizers and spacers for aerosol delivery through adult nasal cannula at lo...
Nebulizers and spacers for aerosol delivery through adult nasal cannula at lo...Nebulizers and spacers for aerosol delivery through adult nasal cannula at lo...
Nebulizers and spacers for aerosol delivery through adult nasal cannula at lo...
 
Anaesthesia ventilators
Anaesthesia ventilatorsAnaesthesia ventilators
Anaesthesia ventilators
 
Respiration Pathway_Dräger Ventilation Story (1).pptx
Respiration Pathway_Dräger Ventilation Story (1).pptxRespiration Pathway_Dräger Ventilation Story (1).pptx
Respiration Pathway_Dräger Ventilation Story (1).pptx
 
Cirugia bariatrica reclutamiento y peep
Cirugia bariatrica reclutamiento y peepCirugia bariatrica reclutamiento y peep
Cirugia bariatrica reclutamiento y peep
 
3. breathing system bsc
3. breathing system bsc3. breathing system bsc
3. breathing system bsc
 
weaning 986.full.pdf
weaning 986.full.pdfweaning 986.full.pdf
weaning 986.full.pdf
 
Chapter 9 room ventilation systems (1)
Chapter 9 room ventilation systems (1)Chapter 9 room ventilation systems (1)
Chapter 9 room ventilation systems (1)
 
Tracheostomy overview
Tracheostomy overviewTracheostomy overview
Tracheostomy overview
 

Recently uploaded

Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original PhotosCall Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photosnarwatsonia7
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalorenarwatsonia7
 
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service MumbaiVIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbaisonalikaur4
 
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...narwatsonia7
 
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...narwatsonia7
 
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...Miss joya
 
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiCall Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiNehru place Escorts
 
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowSonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowRiya Pathan
 
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...Miss joya
 
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls ServiceCall Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Servicesonalikaur4
 
VIP Call Girls Lucknow Nandini 7001305949 Independent Escort Service Lucknow
VIP Call Girls Lucknow Nandini 7001305949 Independent Escort Service LucknowVIP Call Girls Lucknow Nandini 7001305949 Independent Escort Service Lucknow
VIP Call Girls Lucknow Nandini 7001305949 Independent Escort Service Lucknownarwatsonia7
 
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort ServiceCall Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Serviceparulsinha
 
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safenarwatsonia7
 
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceCollege Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceNehru place Escorts
 
Call Girl Surat Madhuri 7001305949 Independent Escort Service Surat
Call Girl Surat Madhuri 7001305949 Independent Escort Service SuratCall Girl Surat Madhuri 7001305949 Independent Escort Service Surat
Call Girl Surat Madhuri 7001305949 Independent Escort Service Suratnarwatsonia7
 
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Call Girls Budhwar Peth 7001305949 All Area Service COD available Any Time
Call Girls Budhwar Peth 7001305949 All Area Service COD available Any TimeCall Girls Budhwar Peth 7001305949 All Area Service COD available Any Time
Call Girls Budhwar Peth 7001305949 All Area Service COD available Any Timevijaych2041
 
Aspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliAspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliRewAs ALI
 

Recently uploaded (20)

Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original PhotosCall Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
 
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service MumbaiVIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
 
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
 
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
 
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
 
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiCall Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
 
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowSonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
 
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
 
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
VIP Call Girls Pune Vrinda 9907093804 Short 1500 Night 6000 Best call girls S...
 
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls ServiceCall Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Service
 
VIP Call Girls Lucknow Nandini 7001305949 Independent Escort Service Lucknow
VIP Call Girls Lucknow Nandini 7001305949 Independent Escort Service LucknowVIP Call Girls Lucknow Nandini 7001305949 Independent Escort Service Lucknow
VIP Call Girls Lucknow Nandini 7001305949 Independent Escort Service Lucknow
 
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort ServiceCall Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
 
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
 
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceCollege Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
 
Call Girl Surat Madhuri 7001305949 Independent Escort Service Surat
Call Girl Surat Madhuri 7001305949 Independent Escort Service SuratCall Girl Surat Madhuri 7001305949 Independent Escort Service Surat
Call Girl Surat Madhuri 7001305949 Independent Escort Service Surat
 
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
 
Call Girls Budhwar Peth 7001305949 All Area Service COD available Any Time
Call Girls Budhwar Peth 7001305949 All Area Service COD available Any TimeCall Girls Budhwar Peth 7001305949 All Area Service COD available Any Time
Call Girls Budhwar Peth 7001305949 All Area Service COD available Any Time
 
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Servicesauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
 
Aspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliAspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas Ali
 

Jaem

  • 1. A Single Ventilator for Multiple Simulated Patients to Meet Disaster Surge Greg Neyman, MD, Charlene Babcock Irvin, MD Abstract Objectives: To determine if a ventilator available in an emergency department could quickly be modified to provide ventilation for four adults simultaneously. Methods: Using lung simulators, readily available plastic tubing, and ventilators (840 Series Ventilator; Puritan-Bennett), human lung simulators were added in parallel until the ventilator was ventilating the equivalent of four adults. Data collected included peak pressure, positive end-expiratory pressure, total tidal volume, and total minute ventilation. Any obvious asymmetry in the delivery of gas to the lung simu- lators was also documented. The ventilator was run for almost 12 consecutive hours (5.5 hours of pressure control and more than six hours of volume control). Results: Using readily available plastic tubing set up to minimize dead space volume, the four lung simu- lators were easily ventilated for 12 hours using one ventilator. In pressure control (set at 25 mm H2O), the mean tidal volume was 1,884 mL (approximately 471 mL/lung simulator) with an average minute ven- tilation of 30.2 L/min (or 7.5 L/min/lung simulator). In volume control (set at 2 L), the mean peak pressure was 28 cm H2O and the minute ventilation was 32.5 L/min total (8.1 L/min/lung simulator). Conclusions: A single ventilator may be quickly modified to ventilate four simulated adults for a limited time. The volumes delivered in this simulation should be able to sustain four 70-kg individuals. While further study is necessary, this pilot study suggests significant potential for the expanded use of a single ventilator during cases of disaster surge involving multiple casualties with respiratory failure. ACADEMIC EMERGENCY MEDICINE 2006; 13:1246–1249 ª 2006 by the Society for Academic Emergency Medicine Keywords: disaster, ventilator, respiratory failure, surge capacity A fter the events of September 11, 2001, and the recent hurricanes in the Gulf Coast, there has been a focus on anticipating the need for medi- cal care for large numbers of victims.1 Addressing surge capacity requires a multitiered approach involving local and federal agencies as well as resource management (i.e., personnel, patient space, supplies, and special equipment). Recent experiences have shown that hospi- tals may be rapidly extended to operate at 120%–130% capacity.2 Depending on the nature of the disaster, many other- wise plentiful hospital supplies, such as ventilators, may suddenly become insufficient to support the demand.3 In the event of a large influx of patients in respiratory distress (e.g., a large outbreak of botulism), the number of ventilators available may not be enough to support all of the patients. While government resources would eventually be available, there may be a time when hospitals will need to provide ventilatory support to a greater number of pa- tients than the available number of ventilators. Manual ventilation (‘‘bagging’’) is possible, but it is possible that the additional personnel required would not be available. Some institutions have begun to stockpile disposable automatic ventilators for use in the event of a disaster.4 One commercially available device is Vortran’s Auto- matic Resuscitator, a single-patient, single-use, pressure- powered ventilator. It could be kept in stock and deployed when necessary. It runs on wall oxygen (50 psi) and is pressure cycled off the wall oxygen source, and seven can be run simultaneously off a single oxygen supply line by using the Vortran E-Vent Case multioutlet manifold device. One advantage is that this type of automatic venti- lator is gas driven (from the oxygen source) and requires no electricity (this may be an advantage in certain types of disasters), is disposable, and is modifiable to provide ventilation for up to seven individuals off one oxygen From the Department of Emergency Medicine, St. John’s Hospi- tal and Medical Center (GN, CBI), Detroit, MI. Received March 3, 2006; revisions received May 4, 2006, and May 7, 2006; accepted May 8, 2006. Lung simulators were donated by Puritan-Bennett. Address for correspondence and reprints: Greg Neyman, MD, Department of Emergency Medicine, St. John Hospital and Medical Center, 22101 Moross Road, Detroit, MI 48236. E-mail: greg.neyman@stjohn.org. ISSN 1069-6563 ª 2006 by the Society for Academic Emergency Medicine PII ISSN 1069-6563583 doi: 10.1197/j.aem.2006.05.0091246
  • 2. source. However, this device requires anticipatory pur- chase, and it lacks the computerized monitoring that stan- dard ventilators have, necessitating more intensive staff support.5 Another option for providing increased ventilation ca- pacity includes a proprietary device (patent pending) that is essentially a control system for splitting one ventilator to provide ventilation for two patients.6 This does not have the advantages of the previously described dispos- able ventilator, because it would rely on an electrically driven ventilator. Additionally, it would require the pur- chase of additional equipment (the proprietary control system). Although larger urban hospitals may be able to justify the resources to stockpile disposable ventilators, smaller hospitals may not. In the event of a need for more venti- lators than are currently available in a hospital, it may be valuable to explore methods to maximally utilize the ven- tilators that are already available. Hospitals do have gen- erators and would likely be able to support electrical service for a short time after the disaster. Given this po- tential, it may be appropriate to consider a simple modi- fication of the currently available hospital ventilators to provide more patients with ventilator support. Our hy- pothesis was that using simple, rapidly deployable mod- ifications, a single ventilator could be used to ventilate multiple casualties when the number of victims exceeds the number of ventilators. METHODS Study Design This was a simulator-based pilot study. This study was considered exempt after review by the institutional review board. Study Protocol Four sets of standard ventilator tubing (Hudson) were connected to a single ventilator (Puritan-Bennett, 840 se- ries) via two flow splitters (one on the patient inflow limb of the circuit, and one on the patient exhaust limb). Each flow splitter was constructed of three Briggs T-tubes with included connection adapters (Hudson) (Figure 1), with the valves removed. The Briggs T-tube is utilized clinically (and generally available) for flow-by oxygen or humidity for a patient with an endotracheal or tracheos- tomy tube, or for in-line aerosol treatments of ventilated patients. The T-tubes were arranged so that the two side ports of a central T-tube were attached to the bottom ports of the two side T-tubes via adapters that come with the T-tube. The final configuration of the three T-tubes is seen in Figure 2 (with a trimmed section of standard ven- tilation tubing at the hub for connection to the ventila- tor); it allowed for air flowing from the ventilator to be split evenly to four simulated patients and for the air returning from the four patients to flow back into the one exhaust port on the ventilator. The ventilator tubing was run from the inflow splitter to the outflow splitter, with four test lungs (Puritan-Ben- nett) in the center. The test lungs were used to simulate one patient each on the modified ventilator circuit. The final configuration was a simulation of four patients on a single ventilator in parallel operation (Figure 3). To test this circuit, a time frame was arbitrarily chosen as approximately six hours. There were two reasons for this. First, this is a simple feasibility study, and we would expect someone to inspect the system at least once in six hours if ever used in a real disaster. Second, we realize that beyond this feasibility study, animal studies are needed, and this could allow for observation of the func- tion of this circuit for a longer period. Finally, in many potential disaster situations, by six hours additional sup- port may be available. Pressure control operation was randomly selected (via coin toss) to precede volume control. To approximate physiologic parameters, the ventilator settings were di- aled to a peak pressure of 25 cm H2O, 0 cm of positive end-expiratory pressure, and a respiratory rate of 16 breaths/min. The ventilator software chose an inspira- tory/expiratory ratio of 1:2 automatically. After cumula- tive random interval inspections, total pressure control Figure 1. Briggs T-tube with included connection adapter. Figure 2. Flow splitter configuration of three T-tubes with connection adapters (shown here with trimmed section of ventilator tubing). ACAD EMERG MED November 2006, Vol. 13, No. 11 www.aemj.org 1247
  • 3. operation was 5 hours 33 minutes. Volume control settings of 2,000 mL tidal volume (500 mL per test lung) and a respi- ratory rate of 16 breaths/min were chosen to approximate physiologic parameters. The ventilator software chose an inspiratory/expiratory ratio of 1:1 automatically. After cumulative random interval inspections, total volume control operation was 6 hours 11 minutes. Measurements During operation, the circuit was inspected at random in- tervals in between examinations of patients in a busy metropolitan emergency department to simulate the in- tervals that the circuit would be inspected in a mass casu- alty event. The inspections occurred approximately every 23 (Æ18) minutes, and the ventilator display readouts were recorded. Simultaneously, the lungs were subjec- tively inspected for symmetry of excursion and evidence of respiratory stacking. Specifically, the examiner moni- tored for asymmetric inflation of individual test lungs and incomplete deflation before subsequent inflation. RESULTS After the configuration was sealed, the ventilator system did not alarm. Visual inspection showed roughly equiva- lent excursion of all lung models. No respiratory stacking was seen. Averages of ventilator display readout sam- plings over thecourse of thestudy are presentedin Table 1. DISCUSSION A four-patient configuration operated successfully on a single ventilator for almost 12 hours. Pressures did not exceed 35 cm H2O. Airway pressures beyond 35 mm H2O are associated with ventilator-induced lung injury.7 Individual tidal volumes reached 471–507 mL, which ap- proximates 7 mL/kg for a 70-kg individual. Studies have shown that ventilation with 6–8 mL/kg is associated with improved outcome in injured lungs.8 No evidence of respiratory stacking or preferential filling of individual lung simulators was observed. LIMITATIONS The chief limitation of this study is that it is a simula- tor study. Therefore, only successful physical ventilation could be demonstrated. Adequate oxygenation and the potential for ventilator-associated lung injury could not be addressed. The presumption of equal ventilation to all four lung simulators presumed equal lung physiology. A patient with asthma with greater resistance to ventilations may not receive equal ventilation with this system. Further animal studies are necessary to address this concern. The inability to directly measure volumes delivered to the individual test lungs may bias the results and thus change the actual method in which this ventilator configuration would be deployed. Potential infectious complications from sharing one ventilator were not investigated. Again, further study in this area would be beneficial. Because this was a pilot study, further research is in- dicated to test the efficacy and safety of the modified circuit. Replication of the study in an animal model is in- dicated. Ventilator software may allow for ventilation of more patients than was explored in this study. This may be an important consideration in usefulness of ventila- tors in the disaster situation. Finally, development of quantitative measurement techniques of individual tidal volumes transferred would enhance further research efforts as well as clinical delivery. CONCLUSIONS This pilot study suggests that the physics of a ventilator/ patient circuit could accommodate more than one Figure 3. Final configuration of multiple patient ventilation (*outflow limbs darkly stained for purposes of graphic clar- ity only). Table 1 Pressure and Flow Characteristics of Lung Simulators in Multiple Patient Ventilation Configuration as an Average of Random Collections Pressure Control Volume Control Total time (hr/min) 5:33 6:11 Peak pressure Æ SD (cm H2O) 25 Æ 0 28 Æ 2 Mean pressure Æ SD (cm H2O) 8.7 Æ 0.1 10.4 Æ 0.5 Peak expiratory end pressure Æ SD (cm H2O) 1.3 Æ 0.1 1.8 Æ 0.1 Tidal volume Æ SD (mL) 1,884 Æ 90 2029 Æ 3 Estimated tidal volume per simulator Æ SD (mL) 471 Æ 22 507 Æ 1 Minute ventilation Æ SD (L) 30.2 Æ 1.4 32.5 Æ 0.1 Estimated minute ventilation per simulator Æ SD (L) 7.5 Æ 0.4 8.1 Æ 0.0 Respiratory rate (breaths/min) 16 16 Inspiratory/expiratory ratio 1:2 1:1 1248 Neyman and Irvin SINGLE VENTILATOR FOR MULTIPLE SIMULATED PATIENTS
  • 4. patient. In a catastrophic situation, when there are more patients who require ventilators than there are ventila- tors available, simple modification of the ventilator cir- cuit could help absorb the extra burden. The authors thank the entire Respiratory Therapy Department at St. John’s Medical Center (Detroit, MI) and Dr. Peter Hoffmann (Physics Department, Wayne State University, Detroit, MI) for consultation regarding fundamental physics. References 1. American College of Emergency Physicians. Health care system surge capacity recognition, prepared- ness, and response. Ann Emerg Med. 2005; 45: 239. 2. Hick JL, Hanfling D, Burstein JL, et al. Health care fa- cility and community strategies for patient care surge capacity. Ann Emerg Med. 2004; 44:253–61. 3. Hick JL, O’Laughlin DT. Concept of operations for triage of mechanical ventilation in an epidemic. Acad Emerg Med. 2006; 13:223–9. 4. Yale New Haven Health. Emergency Preparedness Briefing. Available at: http://yalenewhavenhealth.org/ emergency/commu/briefings/EPBriefingV3I49.pdf. Accessed Apr 25, 2006. 5. VORTRAN Medical Technology 1, Inc. VORTRAN Automatic Resuscitator (VAR) User’s Guide. Available at: http://www.vortran.com/VAR_User_Guide_2005A. pdf. p 3. Accessed May 8, 2006. 6. Darowski M, Englisz M. Artificial ventilation of the lungs for emergencies. Front Med Biol Eng. 2000; 10:177–83. 7. MacIntyre NR. Principles of mechanical ventilation. In: Mason RJ, Murray JF, Broaddus VC, Nadel JA, eds. Murray Nadel’s Textbook of Respiratory Medicine. 4th ed. Philadelphia, PA: Harcourt Health Sciences Group, 2005, p. 2342. 8. Michaels AJ. Management of post traumatic respira- tory failure. Crit Care Clin. 2004; 20:83–99. ACAD EMERG MED November 2006, Vol. 13, No. 11 www.aemj.org 1249