Non-recurring:
(a)Basic portable X-ray machine
(b)X-ray protection devices (e.g. lead aprons,
Gloves, Goggles)
(c)X-ray mo...
Fixed type: Ceiling suspended or floor mounted Mobile type
Portable type Ultra portable (Dental)
Fluoroscopy CT
Digital radiographyPlain and computed radiography
X-ray cassettesX-ray cassettes X-ray gridsX-ray grids
X-ray films in light-proof boxesX-ray films in light-proof boxes
Lea...
X-ray automatic processorX-ray automatic processor
X-ray processing tanks
Dry imager and printer
X-rays may be
produced
when a beam
of electrons of
sufficient
energy from
cathode
interact with
anode matter
inside the X-...
The heat of cathode filament is directly proportional to the magnitude of the current
beam (mA) passed through it.
Thus ...
 X-rays are directed for a brief period of time in a defined angle towards the
body part to be examined radiographically....
 The X-rays are differentially absorbed in body
tissues through an interaction known as
photoelectric absorption.
 Thus ...
 The X-ray film gets selectively modified by these X-rays*
depending upon the number of X-rays reaching to its
different ...
 Such modified X-ray film can be chemically processed
to convert this latent image in to a visible one.
 The parts of Ag...
 Thus area of X-ray film receiving greater number of X-rays
(or light rays) turn blacker than other; the area of film not...
 Minimum magnification of the image of structures being examined (FFD
and subject film distance)
 Least possible distort...
Formulation of radiographic technique chartFormulation of radiographic technique chart
Thickness of part (X)
in cms
Sugges...
kVp change required with change in mAskVp change required with change in mAs
kVp range kVp change required when mAs is dou...
Exposure latitudeExposure latitude
kVp
range
Exposure latitude
46-55 ±2 kVp
56-65 ±4 kVp
66-75 ±6 kVp
76-85 ±8 kVp
86-95 ±...
Processing solution tanks
A: Developing solution tank
B: Rinsing solution tank
C: Fixing solution tank
D: Washing (running...
12 mAs, 16 mAs, 20 mAs, 24 mAs kVp constant at 7512 mAs, 16 mAs, 20 mAs, 24 mAs kVp constant at 75
Nov 2013 radiographic exposure and processing technique sp tyagi
Nov 2013 radiographic exposure and processing technique sp tyagi
Nov 2013 radiographic exposure and processing technique sp tyagi
Nov 2013 radiographic exposure and processing technique sp tyagi
Nov 2013 radiographic exposure and processing technique sp tyagi
Nov 2013 radiographic exposure and processing technique sp tyagi
Nov 2013 radiographic exposure and processing technique sp tyagi
Nov 2013 radiographic exposure and processing technique sp tyagi
Upcoming SlideShare
Loading in …5
×

Nov 2013 radiographic exposure and processing technique sp tyagi

874 views

Published on

Nov 2013 radiographic exposure and processing technique sp tyagi

Published in: Education
  • Be the first to comment

Nov 2013 radiographic exposure and processing technique sp tyagi

  1. 1. Non-recurring: (a)Basic portable X-ray machine (b)X-ray protection devices (e.g. lead aprons, Gloves, Goggles) (c)X-ray monitoring badges (d)X-ray film storage box (e)Different sized X-ray film cassettes (e.g. 8X10”, 10X12”, 12X15” etc) (f)Different types Grid (e.g. 5:1, 8:1, 12:1 as per grid ratio) (g)Different sized X-ray film hangers (At least 4 per size) (h)Different sized lead blockers (i)Lead markers (e.g. Alphabets and numbers) (j)Animal positioning devices (k)Safe lights (Usually low wattage red lights) (l)Film processing tanks (m)Film drier (n)X-ray film viewer/illumination box (o)X-ray record keeping files Recurring: (a)Different sized X-ray films (b)X-ray film developer, replenisher and fixing agents. (c)Intensifying screen cleaning solution (d)Contrast media (e)Tranquilizers/sedatives (f)Anaesthetic agents (g)Emergency drugs/items stock (h)Animal restraining basic devices
  2. 2. Fixed type: Ceiling suspended or floor mounted Mobile type Portable type Ultra portable (Dental)
  3. 3. Fluoroscopy CT Digital radiographyPlain and computed radiography
  4. 4. X-ray cassettesX-ray cassettes X-ray gridsX-ray grids X-ray films in light-proof boxesX-ray films in light-proof boxes Lead markersLead markers
  5. 5. X-ray automatic processorX-ray automatic processor X-ray processing tanks Dry imager and printer
  6. 6. X-rays may be produced when a beam of electrons of sufficient energy from cathode interact with anode matter inside the X- ray tube.
  7. 7. The heat of cathode filament is directly proportional to the magnitude of the current beam (mA) passed through it. Thus more is the ma, the more will be the heat and more will be the number of electrons released from filament atoms. The more is the number of available electrons to interact with the target, more will be number of X-rays produced Another way to control the number of electrons is to increase the time for which machine is operated. This is controlled by another knob on the control dial which select the time (S). The kinetic energy of electrons travelling from cathode to anode in the X-ray tube is dependent on the applied KVP. The higher is the KVP, the higher will be energy of electrons and The higher is the energy of incoming electrons, the higher will be the energy of resulting X-ray photons.
  8. 8.  X-rays are directed for a brief period of time in a defined angle towards the body part to be examined radiographically.  An X-ray sensitive film in a light proof cassette is kept on the other side of part to catch those X-rays that are able to penetrate the part.
  9. 9.  The X-rays are differentially absorbed in body tissues through an interaction known as photoelectric absorption.  Thus its primary beam gets attenuated and reach the X-ray film carrying useful information about the composition of the body part exposed. X ray film in light-proof cassette AirAir Bon e Bon e Soft tissue • Highest absorption • Intermediate absorption • Lowest absorption
  10. 10.  The X-ray film gets selectively modified by these X-rays* depending upon the number of X-rays reaching to its different parts thus forming a ‘LATENT IMAGE’ of the part being exposed.  *X-ray films are in fact mostly exposed by the light rays emitted by the intensifying screen fitted inside the cassette. (The intensifying screen convert X-rays in to light.)
  11. 11.  Such modified X-ray film can be chemically processed to convert this latent image in to a visible one.  The parts of AgBr grains where metallic silver atoms have been deposited is more vulnerable to the action of reducing solution.  Here the Ag ions readily acquire electrons from the developing solution and converts in to black metallic silver.  In the fixing and clearing solution, the part of emulsion containing metallic silver is fixed and unaffected AgBr grains are solubilized and removed.
  12. 12.  Thus area of X-ray film receiving greater number of X-rays (or light rays) turn blacker than other; the area of film not receiving the X-rays turns transparent.  When viewed against white light, the transparent area appears white.  Accordingly a grayish image forms over radiograph depicting the details of the section of the part exposed. Thus on a radiograph bone looks whitish, lungs look blackish and the rest of the soft tissue appears in shades of gray.
  13. 13.  Minimum magnification of the image of structures being examined (FFD and subject film distance)  Least possible distortion of the same (Patient positioning and X ray beam angles)  Sharp delineation of the structures (kVp, mAs, patient restraining, use of grid, processing of film)  Adequate contrast (kVp, mAs, processing of film)  Sufficient radiographic density (kVp, mAs, processing of film)  Necessary numbers of radiographic views  Additionally required contrast techniques  Absence of radiographic artifacts and processing faults (Quality of X ray accessories, processing and storage of films)
  14. 14. Formulation of radiographic technique chartFormulation of radiographic technique chart Thickness of part (X) in cms Suggestive kVp Up to 80 kVp 80-100 kVp Above 100 kVp X-3 A-6 A-9 A-12 X-2 A-4 A-6 A-8 X-1 A-2 A-3 A-4 X A A A X+1 A+2 A+3 A+4 X+2 A+4 A+6 A+8 X+3 A+6 A+9 A+12 • *A is the kVp that provided best radiographic quality on trial exposures. • ** Mas is to kept constant as per the best combination on trial exposures • *** This technique chart will be applicable for similar kind of tissues only.
  15. 15. kVp change required with change in mAskVp change required with change in mAs kVp range kVp change required when mAs is doubled or halved Doubled Halved 41-50 -4 +4 51-60 -6 +6 61-70 -8 +8 71-80 -10 +10 81-90 -12 +12 91-100 -14 +14 101-110 -16 +16 mAs change required with use of grid Grid ratio Increase required in mAs by a factor of 5:1 2 8:1 3 12:1 4
  16. 16. Exposure latitudeExposure latitude kVp range Exposure latitude 46-55 ±2 kVp 56-65 ±4 kVp 66-75 ±6 kVp 76-85 ±8 kVp 86-95 ±10 kVp Degree of variation from the correct exposure factors that still produces a diagnostic radiograph. Degree of variation from the correct exposure factors that still produces a diagnostic radiograph. Thumb rules •Use higher kVp •Use highest possible mA •Use shortest possible exposure time (s) •Use constant FFD (90-100 cms.)
  17. 17. Processing solution tanks A: Developing solution tank B: Rinsing solution tank C: Fixing solution tank D: Washing (running water) tank E: All enclosing warm water (~200 C) tank Processing solution tanks A: Developing solution tank B: Rinsing solution tank C: Fixing solution tank D: Washing (running water) tank E: All enclosing warm water (~200 C) tank Off to drierOff to drier BA C DE 5 min5 min 1/2 min1/2 min 10 min10 min 20 min20 min Timings are suggestive only
  18. 18. 12 mAs, 16 mAs, 20 mAs, 24 mAs kVp constant at 7512 mAs, 16 mAs, 20 mAs, 24 mAs kVp constant at 75

×