▣ Laboratory quality control is designed to detect,
reduce , and correct deficiencies in a laboratory’s
internal analytical process prior to the release of
patient results, in order to improve-the quality of
results reported by the laboratory .
▣ Quality control is a measure of precision or how
well the measurement system reproduce the same
result over time and under varying operating
conditions .
2
• Quality control involves implementation of both Internal Quality
Control (IQC) and External Quality Control (EQC) which is known as
Proficiency Testing (PT).
• The use of internal quality control (IQC) have helped the
laboratories to continually improve the quality control of laboratory.
• In order to achieve the precision, accuracy and reliability of
laboratory test results, accreditation and external quality
assessment (EQA) are very important procedures
When QC is performed
Quality control is usually run -
~ At the beginning of each shift
~After an instrument is serviced
~When reagent lots are changed
~After equipment calibration
~whenever patients results seem inappropriate.
4
The steps for implementing a QC
programme are:
▣ establish policies and procedures
▣ assign responsibility for monitoring and reviewing
▣ train all staff in how to properly follow policies and procedures
▣ select good QC material
▣ establish control ranges for the selected material
▣ develop graphs to plot control values—these are called Levey–Jennings
charts
▣ establish a system for monitoring control values
▣ take immediate corrective action if needed
▣ maintain records of QC results and any corrective actions taken.
5
▣ Quality control material should approximate the
same matrix as patient specimens, taking into
account properties such as viscosity, turbidity,
composition, and color.
▣ It should be simple to use, with minimal vial-to-vial
variability, because variability could be
misinterpreted as systematic error in the method
or instrument.
▣ It should be stable for long periods of time, and
available in large enough quantities for a single
batch to last at least one year.
6
Properties
of QC
material
▣ Liquid controls are more convenient than lyophilized (freeze-
dried) controls because they do not have to be reconstituted,
minimising pipetting error.
▣ Dried Tube Specimen (DTS) is slightly cumbersome as a QC
material but it is very low-cost, stable over long periods and
efficient, especially useful for resource-restricted settings in
under-developed and developing countries.
7
Establishing the value range
for the control material
8
What is laboratory error?
▣ Laboratory error: Any defect that occurs during the entire testing
process,from ordering test to reporting results,that in any way influences
the quality of laboratory services
9
Total Testing process
▣ Begins and ends with the patient
▣ Three phases of testing
▣ 1)Pre-analytical -prior to testing
▣ 2)Analytical-All processes involved in testing
▣ 3)Post analytical -After test analysis
▣ Errors can occur at any phase of testing
10
▣Pre-Analytical
errors
11
Pre-Analytical errors
▣ Error that occur in the pre testing phase
▣ Occur both inside and outside of the laboratory and are made by both
laboratory personnel and non laboratory personnel
▣ Includes test requests, patient and specimen identification, specimen
collection, transport, processing
12
Common Pre-analytical Errors
▣ Patient identification
▣ Patient preparation
▣ Selecting the site
▣ Site preparation
▣ Tourniquet application and time
▣ Proper venipuncture technique
▣ Order of draw
• Proper tube mixing Correct
specimen volume
• Proper tube handling and
specimen processing
• Serum samples
• Plasma samples
• Centrifugation
• Special handling of blood
samples
• Stability for whole blood ,
serum and plasma
13
▣Analytical
Errors
14
• Errors that occur during the testing phase
• Includes specimen testing, equipment issues, reagent issues, work environment,
inadequate staffing
• Not always a failure to follow policy, not always a lack of knowledge
• Occurs in the laboratory
15
The analytical phase includes those processes that occur during sample analysis
or testing:
▣ verification of instrument linearity,
▣ verification of instrument precision
▣ verification of instrument accuracy,
▣ verification of reagent lots, through the use of quality control (QC) samples,
procedures, and QC rules.
▣ testing of Samples
16
17
Common causes
▣ Quality control errors, calibration errors
▣ Manual pipetting errors
▣ Reagent issues
▣ Specimen interference
▣ Dilution and/or calculation errors
▣ Inadequate staffing
18
▣Post-
Analytical
Errors
19
▣ • Errors that occur in the post-testing (post-analytical) phase
The second major source of errors in the total testing process (18-30%)
▣ • Includes clerical errors, reporting test results, interpretation of results, follow-up, specimen
storage, retesting
▣ • Failure to adhere to policies and procedures
▣ • Occur both inside and outside of the laboratory and are made by both laboratory personnel &
non-laboratory personnel
▣ • Includes Physician receiving, interpreting, and acting on laboratory results 20
Procedures of the post-analytical phase
1. Evaluation of test results
2. Decision to release test results
3. Preparation of the laboratory test report
4. Release of the laboratory test report
5. Reporting of test results
6. Sample storage and disposal
7.Archiving of laboratory documentation
8. Post-analytical quality indicators 21
Post-analytical quality indicators
▣ Minimum recommended quality indicators for the post-analytical phase are: turnaround time (TAT), percentage of
incorrect laboratory test reports, and notification of critical results.
• Turnaround time - Turnaround time (TAT) is the time interval from the time when the laboratory receives the
sample until the test results for that sample are validated and released.
▣ Monitoring of turnaround time can be expressed in terms of percentage of tests not performed within a given
time
• Errors during transcription of results/ Incorrect laboratory reports - The percentage of laboratory test reports
that are incorrect is another essential Quality indicator in the post-analytical phase
• Notification of critical results - All procedures related to the reporting of critical results have to be recorded and
the data periodically analyzed (number or proportion of critical results reported within a defined period of time).
22
Identifying Errors
▣ Identifying & monitoring errors allows for opportunities to improve lab quality
▣ How can we identify errors?
• Incident reports
• Patient complaints
• Complaints from physician and other healthcare staff
• Laboratory reports (such as TAT, recollects, repeated tests, etc.)
23
Common post-analytical errors
• Critical results not reported or delayed.
• Results not reviewed or verified before being reported.
• Transcriptional error
• Sample Retention
• Abnormal results not recognized by Physician.
• Failure of Physician to order appropriate follow-up testing.
• Failure of Physician to communicate results to patient. 24
ACCURACY & PRICISION
• Accuracy - Accuracy refers to the degree of agreement between a
measured value and its ‘true’ value. It is generally measured by direct
comparison to a reference value or more commonly by using quality
control serum, with an accurate value assigned to it by the manufacturer.
• Precision - Precision refers to the reproducibility or the agreement
between replicate measurements.
• Precision is quantitatively expressed as Standard Deviation (SD) or more
precisely as Coefficient of Variation (CV) of the results in a set of replicate
measurements.
SOME OTHER TERMS USED IN
LABROTORY
• Specificity - Specificity describes to the ability of a method to measure solely
the component of interest.
• Sensitivity - Sensitivity is the ability to detect small quantities of a measured
component. It will subsequently affect both precision and accuracy, when
attempting to measure levels at the bottom end of the clinical range.
• Repeatability - It is the degree of agreement between successive
measurements which have been done on the same sample under similar
conditions (e.g. same analyzer, same user, same laboratory, same methods,
and same reagent lot) within in a very short time
Statistical Tools Used in Laboratory for
Quality Control
Normal or Gaussian distribution
This distribution chart is a biaxial bell shaped
diagram (X/Y) in which X-axis shows the values of
a variable’s observation and Y-axis shows the
frequency of each value.
The distribution curve or Gaussian curve
indicate that about 68% of all values would fall
within 1SD from the mean or average value, 95%
would be expected to fall within 2SD and 99.7%
would be expected to fall within 3SD value.
Measures Of Central Tendency
• (i) Mean
Mean (x
̄ ) is defined as the arithmetic average of a group of values and is determined
by summing the values and dividing by the number of values. x
̄ =(Σ xi) / n 6
• (ii) Median
Median value Is the value which divides the variable observations into two equal
parts and represents the center of the distribution. It is often used with skewed data.
• (iii) Mode
Mode value is the value with the highest frequency and is used to describe the data
with two centers (bimodal).
Measure of Spread of the Data
• Standard Deviation
Standard deviation (SD or s) is the measure of dispersion of a group of values
around the mean. It is derived from the curve of normal distribution and is used
to assess precision.
• Coefficient of Variation
The coefficient of variation (CV) is defined as the ratio of standard deviation to the
mean and is expressed as percentage. It is another measure of percentage of
imprecision.
• Standard Deviation Index
It is the difference between indiviual value subtracted from the group mean
divided by the SD of the group also known as Z- statistic. It is used for peer-group
comparison.
Setting up ‘mean’ of control –
• The mean value of a control material provides an estimate of the central
tendency of the distribution that is expected if method performance
remains stable.
• Mean= ∑ Xi
n
• Any change in accuracy such as a systematic shift or drift, would be
reflected in a change in the mean value of control.
• Which would be shown a shift or drift of the distribution of control results.
How many control measurements should be collected before making calculations
• The rule of thumb is to collect at least 20 measurements over two
weeks or 10 working days .
Procedure -
Include control materials as part of your daily work for an
appropriate period to observe the variation expected in your
laboratory.
Longer periods help to include more operators and more method
changes , such as pre and post maintenance performance, changes
in reagent lots, sample probes and pipettes.
Standard deviation
• Related to the spread or distribution of control results about the expected mean.
• The bigger the standard deviation, the wider the distribution, the greater the
random error, and the poorer the precision of the method.
• The smaller the standard deviation, narrower and sharper the distribution the
smaller the random error, and better the method precision.
• In our control charts we have 1 SD, 2 SD and 3 SD limits as points for
evaluation of QC datas.
Co-efficient of variation ( CV% )
• CV describes the standard deviation as a percentage of
the mean,
• CV % = SD x 100
MEAN
It is measure of percentage of imprecision.
How is CV useful –
• It provides a better estimate of method performance over a range of concentrations.
• It helps in deriving imprecision of a measurement procedure
• Helps in monitoring long term performance of measuring procedure.
• It gives information required to find out the ‘uncertainty of
measurement’/Measurement Uncertainty(MU) of a measuring procedure.
Sources of uncertainty
• sampling,
• sample preparation,
• sample portion selection,
• calibrators,
• reference materials,
• input quantities,
• equipment used,
• environmental conditions,
• condition of the sample
• changes of operator.
Measurement Uncertainty(MU)
• Measurement uncertainty (MU)(Uncertainty of measurement) is
an inherent property of any quantitative measurement result which
express the lack of knowledge of the true value and quantify the
uncertainty of a result, incorporating the factors known to influence
it.
• “a parameter associated with the result of a measurement, that
characterises the dispersion of the values that could reasonably be
attributed to the measurand”(VIM)
Measurement Uncertainty(MU)
• The uncertainty of measurement information most relevant to interpreting a
lab’s test results is the imprecision of the test results across as many routine
operating conditions as possible(multiple calibrator and reagent batches,
multiple operators, equipment maintenance) .
• This imprecision is most easily derived from long-term internal quality
control (QC) data, calculated as standard deviation (SD) or coefficient of
variation (CV%).
• From CV% the MU can be calculated as the 95% conifidence interval i.e ±
1.96 x CV%/SD.
• As per the reportable range of an analyte MU should be calculated at
more then one level of quality control.
Internal Quality Control ( IQC)
• (IQC) is performed daily in the laboratory and involves the use of
- calibrated glassware,
- reagents and equipment.
• The laboratory staff should be qualified professionals.
• Recommendation is to use at-least two control levels for each analyte.
• Samples are internally evaluated in the laboratory.
• The main purpose of IQC is the precision (repeatability or reproducibility) of the method
Interpretation
▣ Interpretation of quality control data involves both graphical and statistical methods.
▣ Quality control data is most easily visualised using a Levey-Jennings Chart.
▣ The dates of analyses are plotted along the x-axis and control values are plotted along
the y-axis.
▣ The pattern of plotted points provides a simple way to detect increased random error
and shifts or trends in calibration.
39
The Control charts
▣ Control charts are a statistical approach to the study of manufacturing
process variation for the purpose of improving the economic effectiveness of
the process.
▣ These methods are based on continuous monitoring of process variation.
▣ The control chart, also known as the Shewhart chart or Process behaviour
chart, is a statistical tool intended to assess the nature of variation in a
process and to facilitate forecasting and management.
40
LJ chart & Westgard rule
• This test is used to check either the quality control value that we run on daily basis in our
laboratories is running well or not.
• Either the LJ graph is normal or abnormal. To rule out This, Westgard rule is very important.
• Every control that we run is plotted on LJ graph.
• LJ graph has x and y axis. On X axis there will be no of days/time while on y axis there will be value
of mean and standard deviation (difference between mean and actual value.
41
Levey Jennings (L-J) Chart
• Levey Jennings chart (L-J chart) is one of the
most commonly used charts to monitor
quality control results.
• It is a graphical method for displaying the
values of controls.
• The control values are plotted versus the
days of the month which is indicated on the
x-axis and value of controls as mean±1SD,
mean±2SD and mean±3SD are indicated on
the y-axis.
• So we daily run our control on LJ graph and our values are plotted on this
graph and graph is drawn. Now in automatic, Westgard rule can be activated
directly. There following 8 major rules and 3 other multi rules for LJ graph in
Westgard Rules.
• Nowadays the automations find the SD automatically It may be performed by
drawing the graph and the SD will be known to us by kit.
43
Rule -12S
▣ Criteria
• One measurement exceeds 2 SD either above or below the mean of
the reference range
▣ Suspected
• Inaccuracy and/or Imprecision
44
Rule 13S
▣ Criteria
• One measurement exceeds 3 SD either above or below the mean
of the reference range.
▣ Suspected
• Inaccuracy and/or Imprecision
45
Rule 22S
▣ Criteria
• 2 consecutive measurement exceeds 2 SD of the reference range,
and on the same side of the mean.
▣ Suspected
• Inaccuracy and/or Imprecision
46
Rule 4S
▣ Criteria
• 2 measurements in the same run have a 4 SD difference (such as
one exceeding 2 SD above the mean, and another exceeding 2 SD
below the mean).
▣ Suspected
• Imprecision.
47
Rule 41S
▣ Criteria
• 4 consecutive measurement exceeds 1 SD on the same side of the
mean.
▣ Suspected
• Inaccuracy.
48
Rule 10X
▣ Criteria
• 10 consecutive measurement are on the same side of the mean.
▣ Suspected
• Inaccuracy.
49
Corrective Action and Preventive
Action (CAPA)
• Corrective action is that action which should be used to stop the
occurrence of non-conformities.
• Corrective action starts with the root cause analysis which forms the most
important part of corrective action.
• The results of the corrective action taken need to be recorded and
monitoring should be done to verify the completion of actions taken and
also to see its effectiveness.
• Preventive action is that which should give the opportunity to prevent
potential nonconformities.
STEPS OF CAPA
a. Reviewing and defining the problem or non-conformity.
b. Finding the cause of problem.
c. Develop an action plan to correct the problem and prevent the recurrence.
d. Implementing the plan.
e. Evaluating the effectiveness of the correction in preventing the problem.
Corrective Actions
1. Repeat assay with another control aliquotes .
2. Repeat assay with another newly reconstituted control aliquotes.
3. Look for clot , Reagent levels , Mechanical fault .
4. Recalibrate .
5. Change Reagent bottle and recalibrate.
6. Check and Do periodic maintenance.
7. Use alternative controls of similar strength – to rule out
manufacturing issues of this lot of control
8. Call service engineer .
9. Use commercial accuracy- based materials / Perform on second
instrument
External Quality Control
• External quality control (EQC) or proficiency testing (PT) is performed as a
test of competency.
• It includes the participation of the laboratory in an external quality
assessment scheme which provides samples for analysis every month.
• They have to be analyzed by the laboratory professionals using the same
procedures as used for testing of quality control samples and patient
samples.
• The results obtained from analysis of EQC samples are reported to the
outside agency running external quality assessment scheme (EQAS).
• They then provide a report for the participating laboratory based on mean,
coefficient of variation and standard deviation index( Z-Score) of the all the
participating laboratories.
z – score ( SDI)
• The z score is the number of standard deviations a control
result is from the expected mean.
• Formula = OBSERVED RESULT – GROUP MEAN
STANDARD DEVIATION
Steps for minimizing
pre analytical errors
55
Patient identification
▣ It is important to identify a patient properly so that blood is being collected from the
correct person. Drawing blood from wrong person or labelling the correct patient's
samples with a different patient's label can certainly contribute to laboratory error.
▣ Blood should not be drawn from a patient before confirming that he is the right
person. A nurse ,physician, relative or guardian should identify patients that are
unable to speak or identify themselves.
56
Patient preparation
▣ Prior to collecting specimen for chemistry certain patient variables need to be
considered. For certain chemistry analytes such as glucose and cholesterol, patients
need to be fasting for at least 12hours prior to venipuncture.
▣ Other analytes such as cortisol and adrenocorticotropin, have diurnal variations,
where the analytes is at its highest level in the morning and the levels gradually
decrease during the course of the day
57
Selecting the site
▣ Selecting the appropriate site for venipuncture can contribute to a better quality sample. The
preferred site is the median cubital vein. This vein is usually the easiest to access.
▣ Generally, there is less need to probe to find the vein, which in turn should cause fewer traumas
during the venipuncture. This will usually be the most comfortable for the patient. If the median
cubital vein cannot be used, the next choice would be the cephalic vein.
▣ The last vein to consider for venipuncture is the basilic vein. This vein is in close proximity to the
median nerve and brachial artery, and extreme caution must be used so that only the basilic vein is
being punctured.
58
Tourniquet Application and Time
▣ The tourniquet should be applied approximately three to four inches above the
venipuncture site. The tourniquet should be on the arm no longer than one minute.
▣ A good rule of thumb to determine the one-minute tourniquet time is to remove the
tourniquet when blood starts to flow into the first tube of blood being drawn.
▣ Prolonged tourniquet time can lead to an increase in various chemistry analytes,
including serum protein, potassium and lactic acid due to haem concentration of blood
at the puncture site.
59
Proper Venipuncture Technique
▣ During phlebotomy, avoid probing to find the vein and achieve blood flow. Excessive
probing and/or “fishing” to find a vein can result in a poor quality sample, including
hemolysis. As mentioned previously, hemolysis can affect several chemistry analytes.
60
61
Centrifugation
▣ The next step in sample processing is the centrifugation of the blood collection tubes In
a swinging bucket centrifuge (preferred type of spin for gel separation tubes), the tubes
should be spun for ten minutes at a speed of 1100 to 1300 relative centrifugal force
(RCF) A fifteen-minute spin at the same speed is required for spinning tubes in a fixed-
angle centrifuge Serum and plasma tubes without gel can be spun at a speed of 1000
RCF for ten minutes.
62
• Develop clear, written standard operating procedure.
• Monitoring Quality indicators.
63
Steps to minimize
analytical errors
64
▣ Use of automated systems
▣ Develop written procedures & policies
▣ Implement a quality assurance program to assess & evaluate procedures
▣ Rigorous training programs
▣ Staff the lab adequately
65
Types of errors
Systematic
Error
Random Error
Cause of systematic error
▣ Sudden failure or change in light source
▣ Change in reagent formulation
▣ Inaccurate calibration
▣ Gradual deterioration of control materials
▣ Room temperature
67
Causes of Random errors
Improperly mixed /dissolved reagents
Air bubbles in reagents and reagents lines, sampling or reagents syringes.
Pipette tips is not fitting properly.
Unstable temperature and incubation.
Unstable power supply.
Poor operator technique
Improper mixing of processed samples.
▣ Incorrect reconstitution of the control material.
68
Prevention of analytical errors
1. Quality of distilled water.
2. Calibration of measuring and testing instruments
3. Regular servicing and maintenance of equipment.
4. Use of calibrator which is traceable to national/international
reference material.
5. Include quality control specimens in each procedure on a daily basis.
6. Laboratory personnel should be aware that the stability of all reagents and controls
69
Steps for minimizing
post-analytical
errors
70
• Use appropriate technology, such as barcode readers.
• Implement automatic verification of normal results.
• Develop clear, written procedure for reporting results.
• In order to avoid transcriptional errors in the results of the test, the reporting /
signatory personal shall verify the results before the results are reported or dispatched.
• Person receiving the reports will sign in the Record Entry Book after collecting reports.
• Education on utilizing resources available for results interpretation. 71
Sample Retention Duration for
Biochemistry Lab
• Whole blood, Serum & Plasma – 48hrs after report has been
finalized.
• Body fluids – 48hrs after report has been finalized.
• Urine Routine – 24hrs after report has been finalized.
• 24hr Urines samples discarded 48hrs after report has been
finalized.
72
Why is this so important?
▣ Best practice for patient care
▣ Physician depend on accurate lab results to properly diagnose & treat patients
(-70% of clinical decisions are based on lab results)
▣ Errors lead to increased healthcare costs & decreased patient satisfaction
▣ Increase in operating costs for the laboratory & facility
▣ Can weaken confidence in healthcare services
73
74
Validation of results in
clinical laboratory
▣ 1. Method validation
▣ 2. Instrument validation
▣ 3. Preanalytical validation procedures
▣ 4. Analytical validation procedures
▣ 5. Postanalytical validation procedures
75
Method Validation
▣ A method is defined as a reagent kit in combination with one or more calibrators and
measurement equipment
▣ For kits used as recommended, an implementation validation is sufficient. This
validation is well described in the clinical laboratory Improvement Amendments (CLIA
procedure. ’88) requirements: calibration, calibration verification, accuracy
verification, reportable range, precision verification, and reference range verification
76
INSTRUMENT VALIDATION
▣ Instrument validation not only includes the measurement equipment, but also control
of speed of centrifuges, small utensils such as pipettes, dispensors, dilutors,
incubators, waterbath, washing devices, and forth. In some cases instrument
validation is included in the system validation; but it is necessary to know the
elements included in system validation.
▣ One of the most difficult validation procedures is software, hardware, and barcode
readers. CLIA explicitly requires that laboratories follow and document the necessary
maintenance stated by the laboratory information system (LIS) manufacturer or
established by the laboratory for the LIS computer and its components.
77
▣ All input/output components such as monitors, printers, and modem transmission
equipment must be maintained to assure accurate, clear, and interference-free
transmission. If the LIS performs any calculation functions to determine a laboratory
result, the function should be verified immediately after the LIS is programmed and
prior to calculation of patient results.
▣ A new instrument cannot be used without a preliminary evaluation. The laboratory
must establish a standard procedure for the evaluation of new instruments.
78
PREANALYTICAL
VALIDATION
PROCEDURES
79
▣ Most errors in clinical laboratories are made during the preanalytical phase.
▣ Preanalytical validation starts with correct sampling procedures.
▣ Correct sampling also includes patient preparation, and correct labeling
and handling of samples before transport and transfer to the laboratory.
▣ There should also be clear instructions for preliminary sample treatment
and storage prior to the analysis.
80
▣ Careful inspection of samples upon arrival in the laboratory can disclose
systematic sampling error.
▣ Transport must be organized in such a way that the quality of samples is not
affected.
▣ Once samples arrive in the laboratory, a check must be performed on forms
and samples: correctly filled-in request forms, verification of sample
identity, the right vial, correct volume, and validity of sampling vials.
▣ After serum/plasma separation, the sample view must be verified:
hemolyzed, lipemic, and icteric samples must be identified.
81
ANALYTICAL
VALIDATION
82
▣ Analytical validation includes internal quality control (IQC) and external quality
assessment (EQA).
▣ Internal quality control must disclose deviations from stable performance, while
external quality assessment must disclose errors in the individual laboratory’s stable
performance.
▣ Beside these control aspects, analytical validation also pinpoints reagent validation
and system validation procedures.
83
▣ The laboratory must establish and follow written quality control procedures
for monitoring and evaluating the quality of the analytical testing process of
each method to assure the accuracy and reliability of patient test results
and reports.
▣ Depending on the type of analysis and the type of equipment, the number
of control samples and the frequency of controls will be different. CLIA ’88
asks for at least two levels of control materials each day of testing.
▣ QC target values and acceptable performance limits are established to
optimize the probability to detect a measurement defect that is large
enough to have an impact on clinical care.
84
External Quality Assessment
EQA/PT is used to evaluate measurement procedure performance by
comparing a laboratory’s results with those of other laboratory’s for
the same set of samples.
Two types of samples:
• Commutable
• Non-commutable
85
▣ Commutable samples:
▣ Prepared by using an individual donor’s specimen or by pooling clinical
patient samples with minimal processing or additives to avoid alteration of
the sample matrix.
▣ Commutable samples are preferred whenever available.
▣ Results are comparable among all measurement procedures.
86
▣ Non-commutable:
▣ Samples are altered in the process such that matrix is modified.
▣ Do not have same measurement characteristics as unaltered clinical patient samples.
▣ Some EQA/PT samples (e.g., urine, csf, blood gas) are prepared as synthetic materials
not derived from patient fluids.
▣ Results are comparable only among a “peer group” using similar measurement
procedure.
87
Potential Problems Investigating Proficiency
testing Results:
1. Clerical errors: Incorrect reporting, mislabeling, Incorrect units, etc.
2. Measurement procedure problems: Incorrect manufacturing of reagents, inadequate
specificity/sensitivity,inadequate QC.
3. Equipment Problems: Obstruction by clot, misaligned probes, inappropriate maintenance.
4. Technical problems cause by personnel errors: incorrect operation, incorrect handling of reagents,
Misinterpretation.
5. Problem with EQA/PT material: Incorrect handling, Incorrect vials.
88
POSTANALYTICAL
VALIDATION
89
▣ Most clinical chemistry and hematology analyzers in clinical laboratories are coupled
to the laboratory informatic system so that transcription errors between raw data
and data on the reports are avoided.
▣ All manual data input must be controlled in order to avoid mistakes in keying.
▣ Postanalytical validation also includes biological and nosological validation.
90
▣ In screening tests, biological validation includes the comparison of an individual
result with the result of a reference group for which reference values are determined.
In a monitoring situation, results are compared with previous results from the same
person.
▣ It is the responsibility of the clinician to interpret the results. This is defined as clinical
or nosological validation . During this process, single results are related to other
results.
91
References
▣ TIETZ TEXTBOOK OF LABORATORY MEDICINE, 7TH EDITION
▣ TEXTBOOK OF BIOCHEMISTRY FOR MEDICAL STUDENTS, Revised
Reprint 10/e
▣ Wilson And Walkers Principles And Techniques Of Biochemistry
And Molecular Biology 8Ed
▣ Harper's Illustrated Biochemistry, Thirty-Second Edition
▣ WHO Laboratory Quality Management System Handbook
▣ Internet resources.
92
93
Thanks!