Bright Field Microscope
Troubleshooting Challenge
Instructions:
• Work in groups.
• 2 minutes per challenge.
• Diagnose the problem, explain the cause, propose a solution.
Scoring: Diagnosis (2), Explanation (2), Solution (2).
Challenge 1
Field of view is completely dark. Identify at least 3 causes and explain your troubleshooting sequence.
Answer 1
Diagnosis: Light is not reaching the objective.
Possible causes:
• Illuminator switched off.
• Brightness too low.
• Iris diaphragm fully closed.
• Condenser lowered.
• Objective not clicked into position.
• Slide not over the light path.
Troubleshooting:
1. Turn on light and increase intensity.
2. Check objective clicks into place.
3. Centre slide.
4. Raise condenser.
5. Partially open iris diaphragm.
6. Refocus at 10×.
Challenge 2
The specimen is blurry at every magnification. What is causing the problem?
Answer 2
Possible causes:
• Dirty objective, eyepiece or slide.
• Incorrect focusing.
• Condenser misaligned.
• Damaged coverslip.
Troubleshooting:
Clean optics using lens paper only; refocus from 10×; adjust condenser and iris.
Principle: Image sharpness depends on clean optics, correct focal plane and proper illumination.
Challenge 3
The specimen disappears after changing from 10× to 40×.
Answer 3
Causes:
• Specimen not centred at low power.
• Slide shifted.
• Improper refocusing.
Solution:
Return to 10×, relocate and centre the specimen, then switch to 40× and use only the fine-focus knob.
Principle: Higher magnification reduces the field of view.
Challenge 4
The image is very bright but details are difficult to see.
Answer 4
Cause: Excess illumination reduces contrast.
Solution:
• Partially close iris diaphragm.
• Adjust condenser height.
• Reduce lamp intensity.
Principle: Contrast improves when illumination is optimized, not maximized.
Challenge 5
Black dots appear in every field.
Answer 5
Identify source:
• Rotate eyepiece—dots move: eyepiece dirty.
• Rotate objective—dots move: objective dirty.
• Move slide—dots move: slide dirty.
Clean with lens paper.
Principle: Dust blocks light and creates artefacts.
Challenge 6
The objective lens touches the slide while focusing.
Answer 6
Cause: Coarse adjustment used at high magnification or stage raised excessively.
Solution:
• Start with 10×.
• Use coarse focus only at low power.
• At 40×/100× use only fine focus.
Principle: High-power objectives have a very small working distance.