ZMPCRM073000.12.02 Quick Reference Guide
1. CURRENT
TM
SOLUTIONS
A Roscoe Medical Company
QUICK REFERENCE GUIDE
PAGE 1
The purpose of this guide is to assist you in correctly operating your Intensity™ 5000 Hybrid unit.
This device should only be used under the continued supervision of a licensed physician. This
guide is not intended to replace or substitute the Intensity™ 5000 Hybrid Instruction Manual.
PLEASE READ THE INSTRUCTION MANUAL COMPLETELY BEFORE USING THIS DEVICE. Read,
understand and practice the warnings, cautions and operating instructions. Always follow the
operating instructions prescribed by your healthcare provider.
Top
15
1
2
16
Front
3
6
Side
5
4
Back
17
7
8
9
10
11
12
13
19
14
18
INTENSITY™
5000 HYBRID
UNIT
(2 SETS)
12 Pulse width button
3 Channel 1 intensity increase button
13 Pulse rate button
4 Channel 1 intensity decrease button
14 On/Off button
5 Channel 2 intensity increase button
15 Channel 1 output socket
16 Channel 2 output socket
17 Belt clip
8 Therapy mode LED indicator
LEAD WIRES
2 Channel 2 Intensity output indicator
7 Key-press cover
(2 SETS)
11 Therapy time LED indicator
6 Channel 2 intensity decrease button
ELECTRODES
1 Channel 1 Intensity output indicator
18 Battery compartment cover
19 Adapter receptacle
9 Therapy mode button
10 Therapy time button
For more information about your Intensity™ 5000 Hybrid unit, visit us at currentsolutionsnow.com or call 1-800-871-7858.
2. CURRENT
TM
SOLUTIONS
A Roscoe Medical Company
QUICK REFERENCE GUIDE
PAGE 2
Steps 1 – 4 should be performed with the device turned OFF.
IMPORTANT! Once the unit has been turned ON, do not remove electrodes
without first turning OFF the device.
1
INSERT BATTERY
Press arrow and slide battery compartment
cover open. Insert 9V battery, matching the
+ and – ends to those in unit.
The included AC adapter may
also be used as a power source.
4
PLACE ELECTRODES
ON SKIN
Remove electrodes from
protective backing. Keep
backing to properly store
electrodes after use.
2
CONNECT ELECTRODES
TO LEAD WIRES
Turn the PW knob clockwise to select pulse width.
The pulse width is adjustable from 30 uS to 260 uS.
5
TURN DEVICE ON
Press the ON/OFF button located out the
bottom of the control panel. The mode LED indicator
and timer LED indicator will light up.
Turn the PR knob clockwise to select pulse rate.
The pulse rate is adjustable from 2 Hz to 150 Hz.
SELECT
THERAPY MODE
There are 5 therapeutic modes available:
B (burst), N (normal), M1 (pulse width
modulation), M2 (pulse rate modulation) and
SD (strength duration).
MODE
8
9
SELECT
THERAPY TIME
There are 4 therapy times available: 15 minutes, 30
minutes, 60 minutes and C for continuous.
N
15
M1
30
M2
60
N
15
M1
30
M2
60
SD
C
TIMER
Press the MODE button and continue to press to
select desired mode.
ON/OFF
MODE
P.R. (Hz)
6
B
B
P.W. (uS)
This unit operates
with one or two sets
of electrodes. When
using one set, plug
lead wire into Channel
1 on top of the unit.
Use Channel 2 to
operate a second set
if directed to do so
by your healthcare
provider.
Make sure the bare metal of the pins is not exposed.
See Electrode Placement Guide for more information.
ADJUST PULSE WIDTH
AND PULSE RATE
CONNECT LEAD WIRES
TO TOP OF DEVICE
Firmly insert
the lead wire
connectors into
the electrode
connectors.
Apply electrodes to the
exact site indicated by
your doctor or therapist.
Make sure the skin at the
application site is clean
and dry.
7
3
SD
C
TIMER
Press the TIMER button and the appropriate time LED
indicator will illuminate. Continue to press to select
desired time.
ADJUST CHANNEL
INTENSITY
Press the intensity control button
s and t to control the intensity
output. You will hear a “beep” sound.
Slowly press the intensity button
control until you reach the setting
recommended by your physician or
therapist. Repeat for other channel if
both channels are being used.
The Safety Lock Feature automatically locks the
settings if no changes have been made in the panel for
20 seconds. This feature prevents accidental changes to
your settings. Press the t button to unlock the device.
PROBLEMS WITH YOUR DEVICE? Please see pages 24 and 25 of your Intensity™ 5000 Hybrid Instruction Manual for troubleshooting tips.