  • DOWNLOAD THIS BOOKS INTO AVAILABLE FORMAT (2019 Update) ......................................................................................................................... ......................................................................................................................... Download Full PDF EBOOK here { https://soo.gd/irt2 } ......................................................................................................................... Download Full EPUB Ebook here { https://soo.gd/irt2 } ......................................................................................................................... Download Full doc Ebook here { https://soo.gd/irt2 } ......................................................................................................................... Download PDF EBOOK here { https://soo.gd/irt2 } ......................................................................................................................... Download EPUB Ebook here { https://soo.gd/irt2 } ......................................................................................................................... Download doc Ebook here { https://soo.gd/irt2 } ......................................................................................................................... ......................................................................................................................... ................................................................................................................................... eBook is an electronic version of a traditional print book THIS can be read by using a personal computer or by using an eBook reader. (An eBook reader can be a software application for use on a computer such as Microsoft's free Reader application, or a book-sized computer THIS is used solely as a reading device such as Nuvomedia's Rocket eBook.) Users can purchase an eBook on diskette or CD, but the most popular method of getting an eBook is to purchase a downloadable file of the eBook (or other reading material) from a Web site (such as Barnes and Noble) to be read from the user's computer or reading device. Generally, an eBook can be downloaded in five minutes or less ......................................................................................................................... .............. Browse by Genre Available eBooks .............................................................................................................................. Art, Biography, Business, Chick Lit, Children's, Christian, Classics, Comics, Contemporary, Cookbooks, Manga, Memoir, Music, Mystery, Non Fiction, Paranormal, Philosophy, Poetry, Psychology, Religion, Romance, Science, Science Fiction, Self Help, Suspense, Spirituality, Sports, Thriller, Travel, Young Adult, Crime, Ebooks, Fantasy, Fiction, Graphic Novels, Historical Fiction, History, Horror, Humor And Comedy, ......................................................................................................................... ......................................................................................................................... .....BEST SELLER FOR EBOOK RECOMMEND............................................................. ......................................................................................................................... Blowout: Corrupted Democracy, Rogue State Russia, and the Richest, Most Destructive Industry on Earth,-- The Ride of a Lifetime: Lessons Learned from 15 Years as CEO of the Walt Disney Company,-- Call Sign Chaos: Learning to Lead,-- StrengthsFinder 2.0,-- Stillness Is the Key,-- She Said: Breaking the Sexual Harassment Story THIS Helped Ignite a Movement,-- Atomic Habits: An Easy & Proven Way to Build Good Habits & Break Bad Ones,-- Everything Is Figureoutable,-- What It Takes: Lessons in the Pursuit of Excellence,-- Rich Dad Poor Dad: What the Rich Teach Their Kids About Money THIS the Poor and Middle Class Do Not!,-- The Total Money Makeover: Classic Edition: A Proven Plan for Financial Fitness,-- Shut Up and Listen!: Hard Business Truths THIS Will Help You Succeed, ......................................................................................................................... .........................................................................................................................
       Reply 
    Are you sure you want to  Yes  No
    Your message goes here

4 computed tomography Dr. Muhammad Bin Zulfiqar

  1. 1. 4Computed Tomography DR MUHAMMAD BIN ZULFIQAR PGR III FCPS Services institute of Medical Sciences/ Services Hospital Lahore GRAINGER & ALLISON’S DIAGNOSTIC RADIOLOGY
  2. 2. • FIGURE 4-1 ■ CT performance. Performance doubled every two years. This trend has stopped around 2010.
  3. 3. • FIGURE 4-2 ■ (A) Multidetector (MDCT) Principle. Modern multidetector CT machines use a third-generation setup with a rotating tube–detector combination. The detector is split into multiple thin parallel detector rows. (B) Dual-source CT. Two tube– detector units are combined in one machine. This requires only a quarter rotation to complete a 180° half-scan. (C) z-Flying focal spot (from Flohr et al.16). A z-FFS rapidly toggles the position of the focal spot in the z-direction. This produces twice the number of projections per rotation but does not increase detector width, image acquisition speed or number of available detector rows. In addition, the tissue covered by the two beams is separated towards the tube and overlaps towards the detector.
  4. 4. • FIGURE 4-3 ■ Two pairs of images demonstrating the effect of using different reconstruction kernels. Images through the thorax on lung windows at the level of the aortic arch using (A) soft tissue and (B) lung algorithms show the latter to give a much sharper image of the focal lesion within the left upper lobe and of the background emphysema. Images through the upper lumbar spine on bone windows using (C) soft tissue and (D) bone algorithms show the latter to give a much sharper image of the focal metastatic deposit within the vertebral body.
  5. 5. • FIGURE 4-3 ■ Two pairs of images demonstrating the effect of using different reconstruction kernels. Images through the thorax on lung windows at the level of the aortic arch using (A) soft tissue and (B) lung algorithms show the latter to give a much sharper image of the focal lesion within the left upper lobe and of the background emphysema. Images through the upper lumbar spine on bone windows using (C) soft tissue and (D) bone algorithms show the latter to give a much sharper image of the focal metastatic deposit within the vertebral body.
  6. 6. • FIGURE 4-4 ■ Comparison of filtered back projection (FBP). (A) and (C) and iterative reconstructions (IR; (B) and (D)). If the filter settings of IR are set to maximum noise reduction, a plastic look with isolated noise pixels may arise (B). For appropriate settings, image quality can be substantially improved, especially in areas of high absorption, such as the shoulders, or high contrast, such as the lungs (D).
  7. 7. • FIGURE 4-4 ■ Comparison of filtered back projection (FBP). (A) and (C) and iterative reconstructions (IR; (B) and (D)). If the filter settings of IR are set to maximum noise reduction, a plastic look with isolated noise pixels may arise (B). For appropriate settings, image quality can be substantially improved, especially in areas of high absorption, such as the shoulders, or high contrast, such as the lungs (D).
  8. 8. • FIGURE 4-5 ■ Effect of imaging and reconstruction thickness. Sagittal images through the lumbar spine using (A) 1-mm imaging reconstructed in 1-mm-thick slices, (B) 1- mm-thick slices reconstructed in 5-mm-thick slices and (C) 5-mm-thick images reconstructed in 5-mm thick slices. Note the progressive deterioration in image detail, development of a step artefact and blurring of the images.
  9. 9. • FIGURE 4-5 ■ Effect of imaging and reconstruction thickness. Sagittal images through the lumbar spine using (A) 1-mm imaging reconstructed in 1-mm- thick slices, (B) 1-mm-thick slices reconstructed in 5- mm-thick slices and (C) 5- mm-thick images reconstructed in 5-mm thick slices. Note the progressive deterioration in image detail, development of a step artefact and blurring of the images.
  10. 10. • FIGURE 4-6 ■ CT-guided drainage procedure. The diagnostic study (A) demonstrates an abscess within the sigmoid mesentery following an episode of diverticulitis. Under CT guidance (B) a drainage catheter is inserted from the left side of the anterior abdominal wall and the abscess completely drained.
  11. 11. • FIGURE 4-7 ■ ECG-synchronised CT acquisitions. Spiral CT with retrospective gating (A) keeps the tube current constant throughout the entire data acquisition process while only a small portion (8–20%) of the data (grey) is used for image reconstruction. ECG dose modulation (B) reduces the tube current during systole and has a higher dose efficiency of 16–40%. Prospective triggering (C) with a stop-and-start technique uses 50–100% of the data, depending on the amount of padding. ‘Flash scanning’ (D) uses a very fast spiral data acquisition process with prospective triggering, and has a dose efficiency of nearly 100%.
  12. 12. • FIGURE 4-8 ■ Effect of cardiac gating. Conventional axial CT image (A) through the left atrium in a patient with a history of ischaemic stroke shows extensive motion artefacts. However, retrospective gating of the images (B) clearly demonstrates thrombus within the left atrial appendage.
  13. 13. • FIGURE 4-9 ■ (A) The Hounsfield units of different structures vary with the energy of the X-ray photons and, more importantly, so does the rate of change. Dual-energy imaging exploits this to identify individual chemical components within a structure. Imaging of a thoracic aortic dissection with (B) conventional unenhanced and (C) arterial phase imaging. A ‘virtual unenhanced’ image (D), created by subtracting the iodine from the arterial phase study, enhances the conspicuity of the dissection flap and removes the need for the unenhanced phase.
  14. 14. • FIGURE 4-9 ■ (A) The Hounsfield units of different structures vary with the energy of the X-ray photons and, more importantly, so does the rate of change. Dual-energy imaging exploits this to identify individual chemical components within a structure. Imaging of a thoracic aortic dissection with (B) conventional unenhanced and (C) arterial phase imaging. A ‘virtual unenhanced’ image (D), created by subtracting the iodine from the arterial phase study, enhances the conspicuity of the dissection flap and removes the need for the unenhanced phase.
  15. 15. • FIGURE 4-10 ■ Perfusion CT in left middle cerebral artery territory infarct. Mean transit time (MTT) map (A) shows an area of delayed MTT in the posterior part of the left middle cerebral artery territory. Cerebral blood flow (CBF) map (B) shows a larger area of reduced CBF indicating ischaemic and infarcted tissue. Cerebral blood volume CBV map (C) shows a small area of reduced CBV in the left parietal convexity corresponding to core infarct. The area of mismatch between the regions of reduced CBF and CBV is potentially salvageable ischaemic penumbra.
  16. 16. • FIGURE 4-10 ■ Perfusion CT in left middle cerebral artery territory infarct. Mean transit time (MTT) map (A) shows an area of delayed MTT in the posterior part of the left middle cerebral artery territory. Cerebral blood flow (CBF) map (B) shows a larger area of reduced CBF indicating ischaemic and infarcted tissue. Cerebral blood volume CBV map (C) shows a small area of reduced CBV in the left parietal convexity corresponding to core infarct. The area of mismatch between the regions of reduced CBF and CBV is potentially salvageable ischaemic penumbra.

×