  1. 1. MICROPRECISION CALIBRATION INC. BLOCK 2 LOT 4 CALAMBA PREMIERE INTERNATIONAL PARK BATINO, CALAMBA CITY, LAGUNA, PHILS 4027 +63495023015 Certificate of CalibrationDate: 5/16/2008 Lab # 935.05 Certificate #: 315634 Customer: SAMPLE 85/18 BUMDUREN KRASANG Purchase Order: 12345 BURIRAM, , 31160 Work Order: N/A MPC Control #: US37290511 Serial Number: US37290511 Asset ID: 8957 Department: N/A Gage Type: POWER SENSOR Performed By: DANNY BOY B. BUTIAL Manufacturer: HEWLETT PACKARD Received Condition: IN TOLERANCE Model Number: 8481A Returned Condition: IN TOLERANCE Size: 10MHZ~18GHZ Cal Date: October 31, 2007 Temp./RH: 23 ° C / 49 % Cal. Interval: 12 MONTHS Cal. Due Date: October 31, 2008 *Calibration Notes: Site loc: MPC-Lab. See attached data report. Standards Used To Calibrate Equipment I.D. Description Model Serial Manufacturer Cal. Due Date Traceability # AD3312 SIGNAL GENERATOR 83640B 3844A00758 HEWLETT PACKARD 1/16/2009 227210 AI2393 DIRECTIONAL COUPLER 11692D 1212A02566 HEWLETT PACKARD 8/29/2008 139290 AN1816 POWER METER E4419B GB40208110 AGILENT 3/5/2009 260329 AN1869 POWER SPLITTER 11667A 04040 HEWLETT PACKARD 5/9/2009 310631 AU6515 POWER SENSOR 8481A 1726A03110 AGILENT 1/25/2009 1-1201452485-1 R1640 DIRECTIONAL COUPLER 778D 15050 HEWLETT PACKARD 8/20/2008 133363 Procedures Used In This Event: Procedure Name Description 848X SERIES, HP POWER SENSOR, COAXIAL (HP)Calibrating Technician: QC Approval: DANNY BOY B. BUTIAL Jarrod TrammellUncertainty Estimated at 2 to 1Uncertainties have been estimated at a 95 percent confidence level (k=2). All calibrations performed conform to ISO 17025:2005, ISO 9001:2000, ANSI Z540-1, MPC Quality Manual, Calibration System Rev5, andcustomer requirementsAll standards used are either traceable to the National Institute of Standards Technology or have intrinsic accuracy. All services performed have used proper manufacturer and industrial service techniques. Thisreport may not be reproduced in part without written permission of Micro Precisions Quality Assurance Manager. Page 1 of 1 (CERT, Rev 0)
  2. 2. Hewlett Packard 8481A Power Sensor Procedure Reference: 848X SERIES, HP Work Order: n/a ID Number: US37290511 Serial Number: US37290511 Asset Number: VM#8957 / PW#003850 Calibration Date: Oct. 31, 2007 Temperature: 23.0°C Humidity: 49.0%Section 1 - RHO and Calibration Factors Reference Cal Factor: 100.0 % Frequency ( Ghz ) Cal Factor CF Uncertainty Measured RHO RHO Upper Limit .01 100.2 % 1.3 % .048 < .166 .03 100.1 % 1.3 % .048 < .166 .1 100.1 % 1.3 % .023 < .048 .3 100.2 % 1.3 % .020 < .048 1 98.8 % 1.3 % .020 < .048 2 98.8 % 1.4 % .020 < .083 3 97.7 % 1.6 % .021 < .083 4 96.7 % 1.6 % .021 < .083 5 97.2 % 1.7 % .018 < .083 6 96.3 % 2.0 % .020 < .083 7 97.8 % 2.0 % .029 < .083 8 95.1 % 2.1 % .030 < .083 9 95.2 % 2.3 % .030 < .083 10 92.9 % 2.5 % .042 < .083 11 96.0 % 2.6 % .043 < .083 12 93.1 % 2.5 % .027 < .083 12.4 93.0 % 2.5 % .022 < .083 13 94.3 % 2.6 % .016 < .123 14 91.6 % 2.7 % .016 < .123 15 92.2 % 2.7 % .030 < .123 16 93.7 % 2.6 % .058 < .123 17 91.9 % 2.8 % .060 < .123 18 89.5 % 2.9 % .054 < .123 Page 1 of 2
  3. 3. ID Number: US37290511 Calibration Factor Chart 100.0 % 98.0 % 96.0 % 94.0 %Cal Factor % 92.0 % 90.0 % 88.0 % 86.0 % 84.0 % .01 .03 .1 .3 1 2 3 4 5 6 7 8 9 10 11 12 12.4 13 14 15 16 17 18 Frequency ( Ghz ) RHO Chart .200 .180 .160 .140 .120 RHO .100 Measured RHO .080 Upper Limit .060 .040 .020 .000 .01 .03 .1 .3 1 2 3 4 5 6 7 8 9 10 11 1212.413 14 15 16 17 18 Frequency ( Ghz ) Page 2 of 2

×