Ventilador Trilogy 100 catálogo

2,356 views

Published on

  • Be the first to comment

  • Be the first to like this

Ventilador Trilogy 100 catálogo

  1. 1. Proximal pressure line Exhalation valve line Quick start guide 1 2 3 4 Overview Trilogy100 Circuit set-up Set-up screen Prescription set-up
  2. 2. Overview 1 Removable air pathway Remote alarm connector Ethernet connector Serial connector – to connect to a computer or Alice polysomnographs Air inlet and grey foam filter External battery connector Detachable battery Oxygen valve port (flow source up to 15 LPM) Exit cm H2O Primary: S/T AVAPS 0 5 10 15 20 25 30 24/04/2007 04:53 PM Navigate Select 7 1 2 3 Cancel action or exit menu Navigate menus or change settings Select/modify/confirm parameters Audio pause button Start/stop button AC power inlet 1 2 3 4 5 6 7 SD card to download ventilation data 4 Prescriber environment, full access mode SD card inserted in the ventilator Audio pause active Ramp active Status of detachable battery Status of internal battery Status of external battery Battery is charging Symbols and icons 5 6
  3. 3. 2 Circuit set-up Before operating Trilogy100, you need to configure the circuit type. Set-up with exhalation valve Set-up with exhalation port Tee connector Proximal pressure line Mask without integrated leak Exhalation valve line Flexible tubing Tracheostomy tubing Proximal pressure port Exhalation valve Bacterial/viral filter (optional) Flexible tubing Whisper Swivel II (exhalation port) Whisper Swivel II (exhalation port) Mask with integrated leak Mask without integrated leak Tracheostomy tubing Bacterial/viral filter (optional) To change the porting block Proximal pressure line Exhalation valve line Please check that circuit type is properly configured.
  4. 4. 3 Set-up screen Setting the Active or Passive circuit Make sure Trilogy100 is turned off before taking the following steps Menu 1/10 Settings And Alarms Options Alarm log Event log Information Exit Navigate Select Select SettingsAndAlarms Menu Settings and Alarms 2/23 Dual Prescription OFF Circuit Type Active PAP Mode S/T Cancel Edit OK Use to choose Active PAP (exhalation valve set-up) or Passive (exhalation port set-up) and then press twice to confirm B D Menu Settings and Alarms 2/23 Dual Prescription OFF Circuit Type Passive Mode S/T Exit Navigate Modify A C Press and hold and for a few seconds to access Set-up page Navigate to CircuitType and press
  5. 5. A Settings And Alarms Options Alarm log Event log Information Menu Options 1/18 Menu Access Limited Detailed View OFF Language English Navigate to Options and press Press to enter MenuAccess C E Menu Options 2/18 Menu Access Full Detailed View OFF Language English Press to begin therapy or continue navigating through the menu to set-up the ventilation parameters To enable full access to the settings B D Menu Options 1/18 Menu Access Full Detailed View OFF Language English Use to choose Full or LimitedAccess and press to confirm Press and to return to Initial Screen Exit Navigate Select Finish Navigate Modify Cancel Edit OK Finish Navigate Modify Setting menu access
  6. 6. Settings And Alarms Options Alarm log Exit Navigate Select Select SettingsAndAlarms 4 Prescription set-up B Menu A From main screen, press Menu AVAPS ON Tidal Volume 400 ml IPAP Max Pressure 25.0 cmH20 Dual Prescription Menu Settings And Alarms 1/20 Dual Prescription ON Mode S/T AVAPS ON Tidal Volume 400 ml IPAP Max Pressure 25.0 cmH20 Finish Navigate Modify If Dual Prescription is enabled, both prescriptions need to be adjusted as previously explained AVAPS ON Tidal Volume 400 ml IPAP Max Pressure 25.0 cmH20 RR 20BPM PIP 5.9cm H2O Vte 435ml Leak 21l/min Menu 3/8 Switch to Secondary Settings Safely Remove SD Card Primary Settings and Alarms Secondary Settings and Alarms Options Exit Navigate Select F Two prescriptions to set Peak Flow 67.9l/min MinVent 8.7l/min Cancel Edit OK Press to validate new setting Cancel Edit OK E Use to change the value or status of setting
  7. 7. AVAPS OFF IPAP 30.0 cmH20 EPAP 8.0 cmH20 Finish Mode S/T AVAPS OFF IPAP 30.0 cmH20 EPAP 8.0 cmH20 Navigate Modify Finish Navigate Modify Press to change the parameter highlighted in blue C D Scroll up and down using to navigate through the settings and alarms Mode S/T AVAPS ON Tidal Volume 400 ml IPAP Max Pressure 25.0 cmH20 Finish Navigate Modify G Press and to return to the main screen
  8. 8. Philips Healthcare is part of Royal Philips Electronics How to reach us www.philips.com/healthcare healthcare@philips.com fax: +31 40 27 64 887 © 2009 Koninklijke Philips Electronics N.V. All rights are reserved. Philips Healthcare reserves the right to make changes in specifications and/or to discontinue any product at any time without notice or obligation and will not be liable for any consequences resulting from the use of this publication. GeyerWMB 1/19/09 MCI 4102101 PN 1059349 Asia +852 2821 5888 Europe,Middle East,Africa +49 7031 463 2254 Latin America +55 11 2125 0744 North America +1 425 487 7000 800 285 5585 (toll free,US only) Philips Respironics 1010 Murry Ridge Lane Murrysville, PA 15668 Customer Service 724 387 4000 800 345 6443 (toll free) Respironics Europe,Africa,Middle East +33 1 47 52 30 00 www.philips.com/respironics Respironics,Trilogy,Whisper Swivel, and Alice are trademarks of Respironics, Inc. and its affiliates.

×