Experiences on Electropolishing
set up at Ettore Zanon SpA
New EP facility developed for the treatment
of EXFEL 1.3GHz cavities.
TTC Meeting DESY – Hamburg
24 - 27 March 2014
The new electropolishing facility at Ettore Zanon SpA
• Horizontal EP facility for 1.3 GHz – 9 cell
cavities
• Developed by the technical department of
Ettore Zanon SpA
▫ specifications and hints given by experts from
DESY.
• More than 100 cavities treated from July 2013.
• Now working at a rate of 5 cavities/week.
• Treatment data:
▫ 140 µm bulk EP as first main polishing
 Final surface treatment is done with 10 µm BCP
▫ Constant 17 V applied on cavity for 6 hours
▫ Mean current value: 270 A
▫ Mean temperature value: 31°C.
2
Experiences on Electropolishing set up at Ettore Zanon SpA
TTC Meeting DESY – Hamburg
24 - 27 March 2014
Before EP
After EP
Steps for qualification
3
Experiences on Electropolishing set up at Ettore Zanon SpA
TTC Meeting DESY – Hamburg
24 - 27 March 2014
• Project with 3D drawing of the bench
• Mechanical and hydraulic installation by Zanon’s staff
• First tests with dummy cavities (DCVs) for parameters setting
• Qualification procedure:
▫ First EP test on reference cavity (RCVn°1):
 40 µm EP vertical test failed
▫ Modification of cathode shielding made with teflon tape
▫ Second EP test on reference cavity (RCVn°2)
 40 µm EP  vertical test OK
▫ Test with two series cavities
 140 µm EP, CAV599 vertical test OK
 140 µm EP, CAV600 vertical test OK
Go on with treatment of series cavities
• Aluminium cathode 99.5% purity:
▫ 5 mm holes for acid distribution,
▫ Teflon tape to shield cathode at irises.
• Fluorinated polymers PTFE, PVD, PFA
▫ For parts in contact with acid
• Ultrapure water 18MΩ*cm
• 99.999% pure nitrogen
• Acid mixture: H2SO4+HF (9:1 ratio)
• Usage up to 10 g/l Niobium dissolved
EP facility: some details
4
Experiences on Electropolishing set up at Ettore Zanon SpA
TTC Meeting DESY – Hamburg
24 - 27 March 2014
Cavity ready for
installation on EP bench
Cavity installed
on EP facility
3D drawing
EP facility: automation of the process
5
Experiences on Electropolishing set up at Ettore Zanon SpA
TTC Meeting DESY – Hamburg
24 - 27 March 2014
Process sequence handled by a PLC
• Sensors for: temperature, pressure,
flowrate, acid level
• H2 and HF sensors for personnel and
explosion safety.
• Movements and rotation controlled by
position sensors.
Automatic control of
current and temperature:
ON/OFF controller on
cooling line
Reaction rate control
Thank you for your attention!
6
Experiences on Electropolishing set up at Ettore Zanon SpA
TTC Meeting DESY – Hamburg
24 - 27 March 2014
1. First configuration:
▫ 40 µm EP test  removal by weight: 42.8 µm
▫ vertical test failed!
▫ Hypothesis:
 polishing at equators not sufficient?
 Surface not perfectly smooth?
2. Second configuration:
▫ 40 µm EP test  removal by weight: 50 µm
▫ vertical test OK!
Tests with different cathode shielding made with teflon tape
7
Experiences on Electropolishing set up at Ettore Zanon SpA
TTC Meeting DESY – Hamburg
24 - 27 March 2014
Constant V = 17V
Mean I = 230A
Mean T = 31°C
Constant V = 17V
Mean I = 267A
Mean T = 30°C
▫ 140 µm bulk EP as first main polishing
 Usually more than 140 µm are removed to be on the safe side
▫ Constant 17 V applied on cavity for ~6 hours
▫ Mean current value: 270 A
▫ Mean temperature value: 31°C.
▫ Mean removal rate: 0.42 µm/min
▫ Average acid flowrate: 10 l/min
▫ Holes diameter: 5 mm
▫ Acid velocity at single hole: 0.94 m/s
▫ Average nitrogen overlay flowrate: 50 l/min
▫ Cavity rotation: 1 rpm/min
EP parameters
8
Experiences on Electropolishing set up at Ettore Zanon SpA
TTC Meeting DESY – Hamburg
24 - 27 March 2014
Ettore Zanon SpA follows the BCP flash production scheme:
• 140 µm EP as first cleaning,
• 10 µm BCP after helium tank integration.
MAIN STEPS:
1. US cleaning in clean room of ISO7 standard
2. Installation of cavity in EP bench CAVITY is HORIZONTAL
3. Pressure test with ultrapure water CAVITY is VERTICAL
4. Draining with nitrogen CAVITY is VERTICAL
5. Filling with acid CAVITY is HORIZONTAL
6. 17 V applied on cavity for approx. 6 hours CAVITY is HORIZONTAL
7. Draining with nitrogen CAVITY is VERTICAL
8. Rinsing with ultrapure water to safe pH CAVITY is VERTICAL
9. Removal from EP bench CAVITY is HORIZONTAL
10.Transportation to clean room ISO7 and rinsing to 12 MΩ*cm
11.30 minutes 100bar high pressure rinsing (HPR) in ISO4 cleanroom
Main EP sequence
9
Experiences on Electropolishing set up at Ettore Zanon SpA
TTC Meeting DESY – Hamburg
24 - 27 March 2014
140 µm removal: how is it calculated?
10
Two different ways to estimate mean thickness removal
1. by cavity weight before and after EP
2. by the integral of current over time
No infos about actual removal at irises and equators
 For this topic, see Ambra Gresele’s presentation
Statistics on EP treatment of EXFEL cavities in series production
TTC Meeting DESY – Hamburg
24 - 27 March 2014
µm =
𝑊 𝑏𝑒𝑓𝑜𝑟𝑒 −𝑊 𝑎𝑓𝑡𝑒𝑟
7.6
µm = (1.484 ∙ 10−3 ∙ 𝐼 ∙ ∆𝑡)
𝑡=𝑡 𝐸𝑃
𝑡=0
140 µm removal: how is it achieved?
Two main parameters: Time and Current
11
TIME
• With an error of ~10µm, duration of EP is the best way to control total
removal.
+ 10 minutes  + 5 micron
Statistics on EP treatment of EXFEL cavities in series production
TTC Meeting DESY – Hamburg
24 - 27 March 2014
Data used are from
weight measurements
140 µm removal: how is it achieved?
Two main parameters: Time and Current
12
CURRENT
Surface removal control by current is more difficult:
• Current is not constant during EP
• Not a free parameter: related to temperature
Statistics on EP treatment of EXFEL cavities in series production
TTC Meeting DESY – Hamburg
24 - 27 March 2014
Data used are from
weight measurements
Comparing surface removal and frequency variation
ΔF as a different method to estimate material removal
13
1. All methods presented can only estimate the MEAN REMOVAL
2. ΔF determines the cavity lenght after tuning
▫ Is it possible to predict correctly ΔF  cavity final lenght from EP variables?
3. Correlation between ΔF and µm removed by weight
▫ Data are scattered  no clear correlation
Statistics on EP treatment of EXFEL cavities in series production
TTC Meeting DESY – Hamburg
24 - 27 March 2014
Data used are from
weight measurements
About removal at irises and equators
14
• Surface thickness measured by means of a ultrasonic thickness gauge.
▫ Krautkramer Branson CL3 Ultrasonic Thickness Gauge with sensor Alpha
20/125 016dry.
▫ Data taken from four cavities.
• Removal at irises is higher than at the equators.
▫ 100 µm guaranteed also at equators.
Statistics on EP treatment of EXFEL cavities in series production
TTC Meeting DESY – Hamburg
24 - 27 March 2014
High measuring error
It makes sense only with long
EP treatment (es. 140 µm)

More Related Content

What's hot

How to measure solids flow? Mass flow of Bulk Solids?
How to measure solids flow? Mass flow of Bulk Solids?How to measure solids flow? Mass flow of Bulk Solids?
How to measure solids flow? Mass flow of Bulk Solids?Tiago Morais
 
Quality Assurance in the Metals Lab
Quality Assurance in the Metals LabQuality Assurance in the Metals Lab
Quality Assurance in the Metals LabRyan Brennan
 
Bulk flow detection solids FLOWJAM
Bulk flow detection solids FLOWJAMBulk flow detection solids FLOWJAM
Bulk flow detection solids FLOWJAMTiago Morais
 
Horiba pg250 datasheet
Horiba pg250 datasheetHoriba pg250 datasheet
Horiba pg250 datasheeta1-cbiss
 
Experimentalstudy measurement of sonic speed of drilling muds under shear stress
Experimentalstudy measurement of sonic speed of drilling muds under shear stressExperimentalstudy measurement of sonic speed of drilling muds under shear stress
Experimentalstudy measurement of sonic speed of drilling muds under shear stressFrank-Michael Jäger
 
Outside Plant Presentation
Outside Plant PresentationOutside Plant Presentation
Outside Plant Presentationpols66
 
Total (Organic) Carbon, Nitrogen and Sulfur Elemental Analysis of Soils and E...
Total (Organic) Carbon, Nitrogen and Sulfur Elemental Analysis of Soils and E...Total (Organic) Carbon, Nitrogen and Sulfur Elemental Analysis of Soils and E...
Total (Organic) Carbon, Nitrogen and Sulfur Elemental Analysis of Soils and E...Elementar Analysensysteme GmbH
 
Analytical Microprocessor Based Conductivity Meter by ACMAS Technologies Pvt ...
Analytical Microprocessor Based Conductivity Meter by ACMAS Technologies Pvt ...Analytical Microprocessor Based Conductivity Meter by ACMAS Technologies Pvt ...
Analytical Microprocessor Based Conductivity Meter by ACMAS Technologies Pvt ...Acmas Technologies Pvt. Ltd.
 
Irga portable photosynthetic system
Irga portable photosynthetic systemIrga portable photosynthetic system
Irga portable photosynthetic systemZuby Gohar Ansari
 
L17_Comparing measurement principles DLG1 AP
L17_Comparing measurement principles DLG1 APL17_Comparing measurement principles DLG1 AP
L17_Comparing measurement principles DLG1 APDaniel Gore
 
Lyssy l80 5000 Water Vapour Permeability Tester
Lyssy l80 5000 Water Vapour Permeability TesterLyssy l80 5000 Water Vapour Permeability Tester
Lyssy l80 5000 Water Vapour Permeability TesterSystechIllinois
 
Total Nitrogen & Phosphorus Estimation Reactor by ACMAS Technologies Pvt Ltd.
Total Nitrogen & Phosphorus Estimation Reactor by ACMAS Technologies Pvt Ltd.Total Nitrogen & Phosphorus Estimation Reactor by ACMAS Technologies Pvt Ltd.
Total Nitrogen & Phosphorus Estimation Reactor by ACMAS Technologies Pvt Ltd.Acmas Technologies Pvt. Ltd.
 
Steam water analysis system
Steam water analysis systemSteam water analysis system
Steam water analysis systempriyank.modi
 
2013 PITTCON presentation
2013 PITTCON presentation2013 PITTCON presentation
2013 PITTCON presentationITS (Sensafe)
 
Monitoring Ozone In Ambient Air
Monitoring Ozone In Ambient AirMonitoring Ozone In Ambient Air
Monitoring Ozone In Ambient AirECRD IN
 

What's hot (20)

How to measure solids flow? Mass flow of Bulk Solids?
How to measure solids flow? Mass flow of Bulk Solids?How to measure solids flow? Mass flow of Bulk Solids?
How to measure solids flow? Mass flow of Bulk Solids?
 
Quality Assurance in the Metals Lab
Quality Assurance in the Metals LabQuality Assurance in the Metals Lab
Quality Assurance in the Metals Lab
 
Bulk flow detection solids FLOWJAM
Bulk flow detection solids FLOWJAMBulk flow detection solids FLOWJAM
Bulk flow detection solids FLOWJAM
 
Horiba pg250 datasheet
Horiba pg250 datasheetHoriba pg250 datasheet
Horiba pg250 datasheet
 
Screening parameters in water analysis
Screening parameters in water analysisScreening parameters in water analysis
Screening parameters in water analysis
 
Experimentalstudy measurement of sonic speed of drilling muds under shear stress
Experimentalstudy measurement of sonic speed of drilling muds under shear stressExperimentalstudy measurement of sonic speed of drilling muds under shear stress
Experimentalstudy measurement of sonic speed of drilling muds under shear stress
 
Outside Plant Presentation
Outside Plant PresentationOutside Plant Presentation
Outside Plant Presentation
 
Application training on the vario MAX N/CN cube
Application training on the vario MAX N/CN cubeApplication training on the vario MAX N/CN cube
Application training on the vario MAX N/CN cube
 
Instruments chns overview_2011
Instruments chns overview_2011Instruments chns overview_2011
Instruments chns overview_2011
 
Total (Organic) Carbon, Nitrogen and Sulfur Elemental Analysis of Soils and E...
Total (Organic) Carbon, Nitrogen and Sulfur Elemental Analysis of Soils and E...Total (Organic) Carbon, Nitrogen and Sulfur Elemental Analysis of Soils and E...
Total (Organic) Carbon, Nitrogen and Sulfur Elemental Analysis of Soils and E...
 
Analytical Microprocessor Based Conductivity Meter by ACMAS Technologies Pvt ...
Analytical Microprocessor Based Conductivity Meter by ACMAS Technologies Pvt ...Analytical Microprocessor Based Conductivity Meter by ACMAS Technologies Pvt ...
Analytical Microprocessor Based Conductivity Meter by ACMAS Technologies Pvt ...
 
Irga portable photosynthetic system
Irga portable photosynthetic systemIrga portable photosynthetic system
Irga portable photosynthetic system
 
L17_Comparing measurement principles DLG1 AP
L17_Comparing measurement principles DLG1 APL17_Comparing measurement principles DLG1 AP
L17_Comparing measurement principles DLG1 AP
 
Lyssy l80 5000 Water Vapour Permeability Tester
Lyssy l80 5000 Water Vapour Permeability TesterLyssy l80 5000 Water Vapour Permeability Tester
Lyssy l80 5000 Water Vapour Permeability Tester
 
Total Nitrogen & Phosphorus Estimation Reactor by ACMAS Technologies Pvt Ltd.
Total Nitrogen & Phosphorus Estimation Reactor by ACMAS Technologies Pvt Ltd.Total Nitrogen & Phosphorus Estimation Reactor by ACMAS Technologies Pvt Ltd.
Total Nitrogen & Phosphorus Estimation Reactor by ACMAS Technologies Pvt Ltd.
 
Steam water analysis system
Steam water analysis systemSteam water analysis system
Steam water analysis system
 
2013 PITTCON presentation
2013 PITTCON presentation2013 PITTCON presentation
2013 PITTCON presentation
 
Sulfur dioxide sensor 4 electrode - so2-a4
Sulfur dioxide sensor 4 electrode - so2-a4Sulfur dioxide sensor 4 electrode - so2-a4
Sulfur dioxide sensor 4 electrode - so2-a4
 
Ozone
OzoneOzone
Ozone
 
Monitoring Ozone In Ambient Air
Monitoring Ozone In Ambient AirMonitoring Ozone In Ambient Air
Monitoring Ozone In Ambient Air
 

Viewers also liked

Solar Elyse And Jeong
Solar   Elyse And JeongSolar   Elyse And Jeong
Solar Elyse And Jeongguest71b2ef
 
As rochas e os minerais Catarina Santos
As rochas e os minerais   Catarina SantosAs rochas e os minerais   Catarina Santos
As rochas e os minerais Catarina SantosIsabela Afra
 
Organizaciones tradicionales y actuales
Organizaciones tradicionales y actualesOrganizaciones tradicionales y actuales
Organizaciones tradicionales y actualeselizabethguarinc
 
Lmcp1532 pembangunan bandar mapan
Lmcp1532 pembangunan bandar mapanLmcp1532 pembangunan bandar mapan
Lmcp1532 pembangunan bandar mapanSoo Voo Key
 
Guia para estudo dos pronomes pessoais 4º ano
Guia para estudo dos pronomes pessoais 4º anoGuia para estudo dos pronomes pessoais 4º ano
Guia para estudo dos pronomes pessoais 4º anoPaulo Alves de Araujo
 
Israel-Gaza Exercise
Israel-Gaza ExerciseIsrael-Gaza Exercise
Israel-Gaza ExerciseLisa M. Beck
 
Dynamic Eagle Company Profile 2
Dynamic Eagle Company Profile 2Dynamic Eagle Company Profile 2
Dynamic Eagle Company Profile 2Nkululeko Ngubane
 
Legislation changes to improve bc ev charging infrastructure
Legislation changes to improve bc ev charging infrastructureLegislation changes to improve bc ev charging infrastructure
Legislation changes to improve bc ev charging infrastructureGordon James
 
1.01 Typography Usability and Readability
1.01 Typography Usability and Readability1.01 Typography Usability and Readability
1.01 Typography Usability and ReadabilityMsGilbert
 
To Have or Have Not.PDF
To Have or Have Not.PDFTo Have or Have Not.PDF
To Have or Have Not.PDFLisa M. Beck
 
BookazineBits Thursday June 25 2015
BookazineBits Thursday June 25 2015BookazineBits Thursday June 25 2015
BookazineBits Thursday June 25 2015Rick Gallagher
 

Viewers also liked (16)

Nódulo tiroideo
Nódulo tiroideoNódulo tiroideo
Nódulo tiroideo
 
Atividade Avaliativa História
Atividade Avaliativa HistóriaAtividade Avaliativa História
Atividade Avaliativa História
 
Solar Elyse And Jeong
Solar   Elyse And JeongSolar   Elyse And Jeong
Solar Elyse And Jeong
 
Agile coaching: Why You Should Care - program handouts
Agile coaching: Why You Should Care  - program handoutsAgile coaching: Why You Should Care  - program handouts
Agile coaching: Why You Should Care - program handouts
 
As rochas e os minerais Catarina Santos
As rochas e os minerais   Catarina SantosAs rochas e os minerais   Catarina Santos
As rochas e os minerais Catarina Santos
 
Organizaciones tradicionales y actuales
Organizaciones tradicionales y actualesOrganizaciones tradicionales y actuales
Organizaciones tradicionales y actuales
 
Lmcp1532 pembangunan bandar mapan
Lmcp1532 pembangunan bandar mapanLmcp1532 pembangunan bandar mapan
Lmcp1532 pembangunan bandar mapan
 
Guia para estudo dos pronomes pessoais 4º ano
Guia para estudo dos pronomes pessoais 4º anoGuia para estudo dos pronomes pessoais 4º ano
Guia para estudo dos pronomes pessoais 4º ano
 
Israel-Gaza Exercise
Israel-Gaza ExerciseIsrael-Gaza Exercise
Israel-Gaza Exercise
 
Dynamic Eagle Company Profile 2
Dynamic Eagle Company Profile 2Dynamic Eagle Company Profile 2
Dynamic Eagle Company Profile 2
 
Legislation changes to improve bc ev charging infrastructure
Legislation changes to improve bc ev charging infrastructureLegislation changes to improve bc ev charging infrastructure
Legislation changes to improve bc ev charging infrastructure
 
Php curl
Php curlPhp curl
Php curl
 
Evolution
EvolutionEvolution
Evolution
 
1.01 Typography Usability and Readability
1.01 Typography Usability and Readability1.01 Typography Usability and Readability
1.01 Typography Usability and Readability
 
To Have or Have Not.PDF
To Have or Have Not.PDFTo Have or Have Not.PDF
To Have or Have Not.PDF
 
BookazineBits Thursday June 25 2015
BookazineBits Thursday June 25 2015BookazineBits Thursday June 25 2015
BookazineBits Thursday June 25 2015
 

Similar to Experiences on Electropolishing Facility at Ettore Zanon SpA

Process Solutions from Wet Chemistry to Near-Infrared Spectroscopy
Process Solutions from Wet Chemistry to Near-Infrared SpectroscopyProcess Solutions from Wet Chemistry to Near-Infrared Spectroscopy
Process Solutions from Wet Chemistry to Near-Infrared SpectroscopyMetrohm USA
 
Amesa mercury sorbent trap system datasheet
Amesa mercury sorbent trap system datasheetAmesa mercury sorbent trap system datasheet
Amesa mercury sorbent trap system datasheeta1-cbiss
 
Varion plus 700 iq nh4 no3 sensor-wtw
Varion plus 700 iq nh4 no3 sensor-wtwVarion plus 700 iq nh4 no3 sensor-wtw
Varion plus 700 iq nh4 no3 sensor-wtwENVIMART
 
Endress+Hauser : Overview & Innovation in W&WWTP Instrumentation
Endress+Hauser : Overview & Innovation in W&WWTP InstrumentationEndress+Hauser : Overview & Innovation in W&WWTP Instrumentation
Endress+Hauser : Overview & Innovation in W&WWTP Instrumentationyann lafforgue
 
Turbimax cus31 endress+hauser datasheet-turbidity sensor
Turbimax cus31 endress+hauser datasheet-turbidity sensorTurbimax cus31 endress+hauser datasheet-turbidity sensor
Turbimax cus31 endress+hauser datasheet-turbidity sensorENVIMART
 
PhD seminars: Rainwater runoff on porous building materials
PhD seminars: Rainwater runoff on porous building materialsPhD seminars: Rainwater runoff on porous building materials
PhD seminars: Rainwater runoff on porous building materialsThijs Van den Brande
 
Workshop - Voorkomen van faling door vibraties - mechanical validation
Workshop - Voorkomen van faling door vibraties - mechanical validationWorkshop - Voorkomen van faling door vibraties - mechanical validation
Workshop - Voorkomen van faling door vibraties - mechanical validationSirris
 
TESTING OF COST-EFFECTIVE LEAK LOCALIZATION METHOD
TESTING OF COST-EFFECTIVE LEAK LOCALIZATION METHODTESTING OF COST-EFFECTIVE LEAK LOCALIZATION METHOD
TESTING OF COST-EFFECTIVE LEAK LOCALIZATION METHODiQHub
 
Oxymax cos41 endress+hauser datasheet-measurement of the dissolved oxygen
Oxymax cos41 endress+hauser datasheet-measurement of the dissolved oxygenOxymax cos41 endress+hauser datasheet-measurement of the dissolved oxygen
Oxymax cos41 endress+hauser datasheet-measurement of the dissolved oxygenENVIMART
 
Tube Wall Temperature Measurement On Steam Reformers - Best Practices
Tube Wall Temperature Measurement On Steam Reformers - Best PracticesTube Wall Temperature Measurement On Steam Reformers - Best Practices
Tube Wall Temperature Measurement On Steam Reformers - Best PracticesGerard B. Hawkins
 
Micronics Portaflow PF440IP - Portable Ultrasonic Clamp On Flow Meter
Micronics Portaflow PF440IP - Portable Ultrasonic Clamp On Flow Meter Micronics Portaflow PF440IP - Portable Ultrasonic Clamp On Flow Meter
Micronics Portaflow PF440IP - Portable Ultrasonic Clamp On Flow Meter Thorne & Derrick UK
 
Ha Ngoc Hieu - ultrasound pdm pres
Ha Ngoc Hieu - ultrasound pdm presHa Ngoc Hieu - ultrasound pdm pres
Ha Ngoc Hieu - ultrasound pdm preshieuhangoc
 
PPT - Electronics Industry Wastewater
PPT - Electronics Industry WastewaterPPT - Electronics Industry Wastewater
PPT - Electronics Industry WastewaterNitin Yadav
 

Similar to Experiences on Electropolishing Facility at Ettore Zanon SpA (20)

Process Solutions from Wet Chemistry to Near-Infrared Spectroscopy
Process Solutions from Wet Chemistry to Near-Infrared SpectroscopyProcess Solutions from Wet Chemistry to Near-Infrared Spectroscopy
Process Solutions from Wet Chemistry to Near-Infrared Spectroscopy
 
Amesa mercury sorbent trap system datasheet
Amesa mercury sorbent trap system datasheetAmesa mercury sorbent trap system datasheet
Amesa mercury sorbent trap system datasheet
 
Qualification of high performance liquid chromatography
Qualification of high performance liquid chromatographyQualification of high performance liquid chromatography
Qualification of high performance liquid chromatography
 
Combination of pH and ORP electrode
Combination of pH and ORP electrodeCombination of pH and ORP electrode
Combination of pH and ORP electrode
 
Varion plus 700 iq nh4 no3 sensor-wtw
Varion plus 700 iq nh4 no3 sensor-wtwVarion plus 700 iq nh4 no3 sensor-wtw
Varion plus 700 iq nh4 no3 sensor-wtw
 
Endress+Hauser : Overview & Innovation in W&WWTP Instrumentation
Endress+Hauser : Overview & Innovation in W&WWTP InstrumentationEndress+Hauser : Overview & Innovation in W&WWTP Instrumentation
Endress+Hauser : Overview & Innovation in W&WWTP Instrumentation
 
pH electrodes-pH sensors CPS41
pH electrodes-pH sensors CPS41pH electrodes-pH sensors CPS41
pH electrodes-pH sensors CPS41
 
Turbidity sensor-Turbimax CUS31
Turbidity sensor-Turbimax CUS31Turbidity sensor-Turbimax CUS31
Turbidity sensor-Turbimax CUS31
 
ICOS ATC Labeling Process
ICOS ATC Labeling ProcessICOS ATC Labeling Process
ICOS ATC Labeling Process
 
Turbimax cus31 endress+hauser datasheet-turbidity sensor
Turbimax cus31 endress+hauser datasheet-turbidity sensorTurbimax cus31 endress+hauser datasheet-turbidity sensor
Turbimax cus31 endress+hauser datasheet-turbidity sensor
 
PhD seminars: Rainwater runoff on porous building materials
PhD seminars: Rainwater runoff on porous building materialsPhD seminars: Rainwater runoff on porous building materials
PhD seminars: Rainwater runoff on porous building materials
 
Workshop - Voorkomen van faling door vibraties - mechanical validation
Workshop - Voorkomen van faling door vibraties - mechanical validationWorkshop - Voorkomen van faling door vibraties - mechanical validation
Workshop - Voorkomen van faling door vibraties - mechanical validation
 
TESTING OF COST-EFFECTIVE LEAK LOCALIZATION METHOD
TESTING OF COST-EFFECTIVE LEAK LOCALIZATION METHODTESTING OF COST-EFFECTIVE LEAK LOCALIZATION METHOD
TESTING OF COST-EFFECTIVE LEAK LOCALIZATION METHOD
 
Photometric sensor for SAC or nitrate measurement
Photometric sensor for SAC or nitrate measurementPhotometric sensor for SAC or nitrate measurement
Photometric sensor for SAC or nitrate measurement
 
Capnography
Capnography  Capnography
Capnography
 
Oxymax cos41 endress+hauser datasheet-measurement of the dissolved oxygen
Oxymax cos41 endress+hauser datasheet-measurement of the dissolved oxygenOxymax cos41 endress+hauser datasheet-measurement of the dissolved oxygen
Oxymax cos41 endress+hauser datasheet-measurement of the dissolved oxygen
 
Tube Wall Temperature Measurement On Steam Reformers - Best Practices
Tube Wall Temperature Measurement On Steam Reformers - Best PracticesTube Wall Temperature Measurement On Steam Reformers - Best Practices
Tube Wall Temperature Measurement On Steam Reformers - Best Practices
 
Micronics Portaflow PF440IP - Portable Ultrasonic Clamp On Flow Meter
Micronics Portaflow PF440IP - Portable Ultrasonic Clamp On Flow Meter Micronics Portaflow PF440IP - Portable Ultrasonic Clamp On Flow Meter
Micronics Portaflow PF440IP - Portable Ultrasonic Clamp On Flow Meter
 
Ha Ngoc Hieu - ultrasound pdm pres
Ha Ngoc Hieu - ultrasound pdm presHa Ngoc Hieu - ultrasound pdm pres
Ha Ngoc Hieu - ultrasound pdm pres
 
PPT - Electronics Industry Wastewater
PPT - Electronics Industry WastewaterPPT - Electronics Industry Wastewater
PPT - Electronics Industry Wastewater
 

Experiences on Electropolishing Facility at Ettore Zanon SpA

  • 1. Experiences on Electropolishing set up at Ettore Zanon SpA New EP facility developed for the treatment of EXFEL 1.3GHz cavities. TTC Meeting DESY – Hamburg 24 - 27 March 2014
  • 2. The new electropolishing facility at Ettore Zanon SpA • Horizontal EP facility for 1.3 GHz – 9 cell cavities • Developed by the technical department of Ettore Zanon SpA ▫ specifications and hints given by experts from DESY. • More than 100 cavities treated from July 2013. • Now working at a rate of 5 cavities/week. • Treatment data: ▫ 140 µm bulk EP as first main polishing  Final surface treatment is done with 10 µm BCP ▫ Constant 17 V applied on cavity for 6 hours ▫ Mean current value: 270 A ▫ Mean temperature value: 31°C. 2 Experiences on Electropolishing set up at Ettore Zanon SpA TTC Meeting DESY – Hamburg 24 - 27 March 2014 Before EP After EP
  • 3. Steps for qualification 3 Experiences on Electropolishing set up at Ettore Zanon SpA TTC Meeting DESY – Hamburg 24 - 27 March 2014 • Project with 3D drawing of the bench • Mechanical and hydraulic installation by Zanon’s staff • First tests with dummy cavities (DCVs) for parameters setting • Qualification procedure: ▫ First EP test on reference cavity (RCVn°1):  40 µm EP vertical test failed ▫ Modification of cathode shielding made with teflon tape ▫ Second EP test on reference cavity (RCVn°2)  40 µm EP  vertical test OK ▫ Test with two series cavities  140 µm EP, CAV599 vertical test OK  140 µm EP, CAV600 vertical test OK Go on with treatment of series cavities
  • 4. • Aluminium cathode 99.5% purity: ▫ 5 mm holes for acid distribution, ▫ Teflon tape to shield cathode at irises. • Fluorinated polymers PTFE, PVD, PFA ▫ For parts in contact with acid • Ultrapure water 18MΩ*cm • 99.999% pure nitrogen • Acid mixture: H2SO4+HF (9:1 ratio) • Usage up to 10 g/l Niobium dissolved EP facility: some details 4 Experiences on Electropolishing set up at Ettore Zanon SpA TTC Meeting DESY – Hamburg 24 - 27 March 2014 Cavity ready for installation on EP bench Cavity installed on EP facility 3D drawing
  • 5. EP facility: automation of the process 5 Experiences on Electropolishing set up at Ettore Zanon SpA TTC Meeting DESY – Hamburg 24 - 27 March 2014 Process sequence handled by a PLC • Sensors for: temperature, pressure, flowrate, acid level • H2 and HF sensors for personnel and explosion safety. • Movements and rotation controlled by position sensors. Automatic control of current and temperature: ON/OFF controller on cooling line Reaction rate control
  • 6. Thank you for your attention! 6 Experiences on Electropolishing set up at Ettore Zanon SpA TTC Meeting DESY – Hamburg 24 - 27 March 2014
  • 7. 1. First configuration: ▫ 40 µm EP test  removal by weight: 42.8 µm ▫ vertical test failed! ▫ Hypothesis:  polishing at equators not sufficient?  Surface not perfectly smooth? 2. Second configuration: ▫ 40 µm EP test  removal by weight: 50 µm ▫ vertical test OK! Tests with different cathode shielding made with teflon tape 7 Experiences on Electropolishing set up at Ettore Zanon SpA TTC Meeting DESY – Hamburg 24 - 27 March 2014 Constant V = 17V Mean I = 230A Mean T = 31°C Constant V = 17V Mean I = 267A Mean T = 30°C
  • 8. ▫ 140 µm bulk EP as first main polishing  Usually more than 140 µm are removed to be on the safe side ▫ Constant 17 V applied on cavity for ~6 hours ▫ Mean current value: 270 A ▫ Mean temperature value: 31°C. ▫ Mean removal rate: 0.42 µm/min ▫ Average acid flowrate: 10 l/min ▫ Holes diameter: 5 mm ▫ Acid velocity at single hole: 0.94 m/s ▫ Average nitrogen overlay flowrate: 50 l/min ▫ Cavity rotation: 1 rpm/min EP parameters 8 Experiences on Electropolishing set up at Ettore Zanon SpA TTC Meeting DESY – Hamburg 24 - 27 March 2014
  • 9. Ettore Zanon SpA follows the BCP flash production scheme: • 140 µm EP as first cleaning, • 10 µm BCP after helium tank integration. MAIN STEPS: 1. US cleaning in clean room of ISO7 standard 2. Installation of cavity in EP bench CAVITY is HORIZONTAL 3. Pressure test with ultrapure water CAVITY is VERTICAL 4. Draining with nitrogen CAVITY is VERTICAL 5. Filling with acid CAVITY is HORIZONTAL 6. 17 V applied on cavity for approx. 6 hours CAVITY is HORIZONTAL 7. Draining with nitrogen CAVITY is VERTICAL 8. Rinsing with ultrapure water to safe pH CAVITY is VERTICAL 9. Removal from EP bench CAVITY is HORIZONTAL 10.Transportation to clean room ISO7 and rinsing to 12 MΩ*cm 11.30 minutes 100bar high pressure rinsing (HPR) in ISO4 cleanroom Main EP sequence 9 Experiences on Electropolishing set up at Ettore Zanon SpA TTC Meeting DESY – Hamburg 24 - 27 March 2014
  • 10. 140 µm removal: how is it calculated? 10 Two different ways to estimate mean thickness removal 1. by cavity weight before and after EP 2. by the integral of current over time No infos about actual removal at irises and equators  For this topic, see Ambra Gresele’s presentation Statistics on EP treatment of EXFEL cavities in series production TTC Meeting DESY – Hamburg 24 - 27 March 2014 µm = 𝑊 𝑏𝑒𝑓𝑜𝑟𝑒 −𝑊 𝑎𝑓𝑡𝑒𝑟 7.6 µm = (1.484 ∙ 10−3 ∙ 𝐼 ∙ ∆𝑡) 𝑡=𝑡 𝐸𝑃 𝑡=0
  • 11. 140 µm removal: how is it achieved? Two main parameters: Time and Current 11 TIME • With an error of ~10µm, duration of EP is the best way to control total removal. + 10 minutes  + 5 micron Statistics on EP treatment of EXFEL cavities in series production TTC Meeting DESY – Hamburg 24 - 27 March 2014 Data used are from weight measurements
  • 12. 140 µm removal: how is it achieved? Two main parameters: Time and Current 12 CURRENT Surface removal control by current is more difficult: • Current is not constant during EP • Not a free parameter: related to temperature Statistics on EP treatment of EXFEL cavities in series production TTC Meeting DESY – Hamburg 24 - 27 March 2014 Data used are from weight measurements
  • 13. Comparing surface removal and frequency variation ΔF as a different method to estimate material removal 13 1. All methods presented can only estimate the MEAN REMOVAL 2. ΔF determines the cavity lenght after tuning ▫ Is it possible to predict correctly ΔF  cavity final lenght from EP variables? 3. Correlation between ΔF and µm removed by weight ▫ Data are scattered  no clear correlation Statistics on EP treatment of EXFEL cavities in series production TTC Meeting DESY – Hamburg 24 - 27 March 2014 Data used are from weight measurements
  • 14. About removal at irises and equators 14 • Surface thickness measured by means of a ultrasonic thickness gauge. ▫ Krautkramer Branson CL3 Ultrasonic Thickness Gauge with sensor Alpha 20/125 016dry. ▫ Data taken from four cavities. • Removal at irises is higher than at the equators. ▫ 100 µm guaranteed also at equators. Statistics on EP treatment of EXFEL cavities in series production TTC Meeting DESY – Hamburg 24 - 27 March 2014 High measuring error It makes sense only with long EP treatment (es. 140 µm)