Tcpo2 calibration using PeriFlux 6000

  1. 1. Calibration PeriFlux 6000 | tcpO2 made intelligent
  2. 2. Disclaimer The information contained in this document is intended to provide general information only. It is not intended to be, nor does it constitute, medical advice. Under no circumstances is the information contained in this document to be interpreted as a recommendation for a particular treatment for specific individuals. In all cases it is recommended that clinicians perform their own interpretations of data in conjunction with the clinical assessment of their patient. Due to Perimed’s commitment to continuous improvement of our products, all specifications are subject to change without notice. All information and content in this document is protected by copyright. All rights are reserved. Users are prohibited from modifying, copying, distributing, transmitting, displaying, publishing, selling, licensing, creating derivative works, or using any information available in or through the document for commercial or public purposes. All responsibility for any liability, loss or risk, personal or otherwise, which is incurred as a consequence, directly or indirectly, of the use and application of any of the material in this document is specifically disclaimed.
  3. 3. • Calibration • tcpO2 Site Selection • Electrode Maintenance • Interpreting Curves and Results
  4. 4. Prepare for Calibration Make sure all electrodes are positioned in the electrode holders.
  5. 5. Prepare for Calibration Uncheck channels if necessary.
  6. 6. Start Calibration Tap CAL to start calibration
  7. 7. Calibrating
  8. 8. Calibration OK
  9. 9. Check Calibration Temperature Check temperature. Typically 45°C.
  10. 10. Comments Always perform a calibration: • Before each monitoring period • When changing patients or changing sites • Every time an electrode is re-membraned If the calibration takes more than 5 minutes, replace the membranes.
  11. 11. Changing Temperature Setting MENU →Settings →Calibration Edit temperature.
  12. 12. Thank You! PeriFlux 6000 | tcpO2 made intelligent www.perimed-instruments.com

×