Step One: Disassembly
© 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017
3
After Each Patient, Remove the Mouthpiece
from th...
© 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017
4 Remove the Vacuum Pipe from the Control Head
Pres...
© 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017
5 Remove the Control Head from the Vacuum Hose
Care...
© 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017
6 Cap the End of the Vacuum Hose with the Suction L...
© 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017
7
Titanium Control Base
Vacuum Levers
Remove the Va...
© 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017
8 Your System Is Now Properly Disassembled
Vacuum P...
© 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017
9
Step Two: Surface Disinfection
Between Each Patie...
© 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017
10
Between Each Patient, Perform a Surface
Disinfec...
© 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017
11
Between each patient, perform a surface disinfec...
© 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017
12
Step Three: Cleaning
© 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017
13
Six Vacuum Pipes Are Provided So One
Is Always R...
© 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017
14 Clean the Vacuum Pipe
Using the provided Double ...
© 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017
15 Clean the Titanium Control Base
Using the provid...
© 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017
16
According to Frequency of Use, Regularly Clean t...
© 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017
17
Step Three: Sterilization
© 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017
18
Follow the sterilization
recommendations provide...
© 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017
19 Before Use on a Patient, Sterilize the Vacuum Pi...
© 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017
20
Follow the sterilization
recommendations provide...
© 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017
21
Weekly, or as Needed per Standard Office Protoco...
© 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017
22
Step Four: Lubrication and Reassembly
© 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017
23 Inspect the Vacuum Lever O-rings for Damage
If t...
© 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017
24 If Necessary, Remove the O-ring from the Vacuum ...
© 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017
25 Replace the O-ring on the Vacuum Levers
Roll the...
© 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017
26 Lubricate the Vacuum Lever O-rings
Coat the O-ri...
© 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017
27 Inspect the Vacuum Hose O-ring for Damage
If the...
© 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017
28 If Necessary, Remove the O-ring from the Vacuum ...
© 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017
29 Replace the O-ring on the Vacuum Hose
Pinch the ...
© 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017
30 Lubricate the Vacuum Hose O-ring
Coat the O-ring...
© 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017
31
Reconnect the Control Head to the Vacuum Hose
an...
© 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017
32
THANK YOU.
PLEASE CONTACT US IF YOU HAVE ANY QUE...
Upcoming SlideShare
Loading in …5
×

Isodry Cleaning & Maintenance

5,487 views

Published on

How to Clean and Maintain Your Isodry Dental Isolation System

Published in: Health & Medicine
  • Manifest Absolutely Anything. Discover the Universe's "7 Sacred Signs" that guide the way to unlocking your heart�s greatest desires. Access your free report now! ➤➤ https://tinyurl.com/y6pnne55
       Reply 
    Are you sure you want to  Yes  No
    Your message goes here
  • Be the first to like this

Isodry Cleaning & Maintenance

  1. 1. © 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017 HOW TO CLEAN AND MAINTAIN YOUR DENTAL ISOLATION SYSTEM
  2. 2. © 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017 2 Step One: Disassembly
  3. 3. © 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017 3 After Each Patient, Remove the Mouthpiece from the Control Head and Safely Dispose of It. Re-use of the mouthpiece can result in the transmission of infectious pathogens SINGLE-USE MEANS SAFE. Mouthpieces are single-use only to eliminate the risk of patient cross contamination.
  4. 4. © 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017 4 Remove the Vacuum Pipe from the Control Head Press and hold the rectangular release button on the underside of the Control Head to release and disconnect the Vacuum Pipe.
  5. 5. © 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017 5 Remove the Control Head from the Vacuum Hose Carefully remove the Control Head by grasping the end of the Vacuum Hose and pulling it straight off. DO NOT TWIST
  6. 6. © 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017 6 Cap the End of the Vacuum Hose with the Suction Line Plug Suction Line Plug
  7. 7. © 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017 7 Titanium Control Base Vacuum Levers Remove the Vacuum Levers from the Titanium Control Base Remove the Vacuum Levers by rotating the levers to the off position. Keep pushing gently until each lever pops out.
  8. 8. © 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017 8 Your System Is Now Properly Disassembled Vacuum Pipe Titanium Control Base Vacuum Levers
  9. 9. © 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017 9 Step Two: Surface Disinfection Between Each Patient
  10. 10. © 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017 10 Between Each Patient, Perform a Surface Disinfection of the Titanium Control Base Wipe down the surface of the Titanium Control Base with an intermediate level disinfectant wipe with a claim to inactivate Mycobacterium tuberculosis. Titanium Control Base Vacuum Levers
  11. 11. © 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017 11 Between each patient, perform a surface disinfection of the Vacuum Hose Wipe down the surface of the Vacuum Hose with an intermediate level disinfectant wipe with a claim to inactivate Mycobacterium tuberculosis.
  12. 12. © 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017 12 Step Three: Cleaning
  13. 13. © 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017 13 Six Vacuum Pipes Are Provided So One Is Always Ready When You Need It
  14. 14. © 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017 14 Clean the Vacuum Pipe Using the provided Double End Tube Brush and a neutral pH (6.5-7.5) enzymatic cleaning solution, remove biomatter and debris from the suction channels. After cleaning, rinse with tap water and dry. DO NOT HOLD PARTS IN CLEANING SOLUTION.
  15. 15. © 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017 15 Clean the Titanium Control Base Using the provided Double End Tube Brush and a neutral pH (6.5-7.5) enzymatic cleaning solution, remove biomatter and debris from the suction channels. After cleaning, rinse with tap water and dry. DO NOT HOLD PARTS IN CLEANING SOLUTION OR USE ULTRASONIC CLEANERS
  16. 16. © 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017 16 According to Frequency of Use, Regularly Clean the inside of the Vacuum Hose to Prevent Accumulation of Biofilm Start by vacuuming water through the Vacuum Hose. Then, vacuum a dental hose cleaner with a pH between 6 and 8, through the Vacuum Hose per manufacturer’s instructions.
  17. 17. © 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017 17 Step Three: Sterilization
  18. 18. © 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017 18 Follow the sterilization recommendations provided in the Instructions for Use Before Use on a Patient, Sterilize the Vacuum Pipes
  19. 19. © 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017 19 Before Use on a Patient, Sterilize the Vacuum Pipes Place the Vacuum Pipes into an autoclave pouch and sterilize in a steam autoclave using these validated sterilization cycles. Cycle Temperature Min. Exposure Time Min. Drying Time Standard 121 °C / 250 °F 30 min 30 min Standard 132 °C / 270 °F 15 min 15 min Prevacuum 132 °C / 270 °F 4 min 20 min Prevacuum 134 °C / 273 °F 3 min 20 min
  20. 20. © 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017 20 Follow the sterilization recommendations provided in the Instructions for Use Titanium Control Base Vacuum Levers Weekly, or as Needed per Standard Office Protocol, Sterilize the Titanium Control Base
  21. 21. © 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017 21 Weekly, or as Needed per Standard Office Protocol, Sterilize the Titanium Control Base Place the Titanium Control Base and Vacuum Levers into an autoclave pouch and sterilize in a steam autoclave using these validated sterilization cycles. Cycle Temperature Min. Exposure Time Min. Drying Time Standard 121 °C / 250 °F 30 min 30 min Standard 132 °C / 270 °F 15 min 15 min Prevacuum 132 °C / 270 °F 4 min 20 min Prevacuum 134 °C / 273 °F 3 min 20 min
  22. 22. © 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017 22 Step Four: Lubrication and Reassembly
  23. 23. © 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017 23 Inspect the Vacuum Lever O-rings for Damage If the O-rings shows signs of cracking, breakage, or crusty build-up that can not be completely cleaned with a wet cotton swab, the O-ring will need to be removed and replaced. If properly maintained and lubricated, the O-ring can last for many months or years.
  24. 24. © 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017 24 If Necessary, Remove the O-ring from the Vacuum Levers Using a probe with a point, pinch and slide the O-ring to one side to create a small loop. Slide the O-ring off. Insert the probe point under the O-ring.
  25. 25. © 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017 25 Replace the O-ring on the Vacuum Levers Roll the O-ring onto the end of the Vacuum Lever valve barrel and into the groove. O-ring in place.
  26. 26. © 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017 26 Lubricate the Vacuum Lever O-rings Coat the O-ring by applying a small amount of the Dow 111 O-ring Lubricant provided in the O-ring Maintenance Kit. THE VACUUM LEVER O-RINGS REQUIRE REGULAR LUBRICATION TO OPERATE AT PEAK EFFICIENCY.
  27. 27. © 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017 27 Inspect the Vacuum Hose O-ring for Damage If the O-ring shows signs of cracking, breakage, or crusty build-up that can not be completely cleaned with a wet cotton swab, the O-ring will need to be removed and replaced. The O-ring is located less than ½ inch inside the grey plastic fitting at the head of the Vacuum Hose.
  28. 28. © 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017 28 If Necessary, Remove the O-ring from the Vacuum Hose Hold the head of the Vacuum Hose and insert a probe with a point under the O-ring. Gently guide the probe around the groove. The O-ring should come out. Pull out the O-ring.
  29. 29. © 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017 29 Replace the O-ring on the Vacuum Hose Pinch the O-ring and insert one end into the groove. Using the probe, massage the O-ring into the groove.
  30. 30. © 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017 30 Lubricate the Vacuum Hose O-ring Coat the O-ring by applying a small amount of the Dow 111 O-ring Lubricant provided in the O-ring Maintenance Kit. THE Vacuum Hose O-RING REQUIRES REGULAR LUBRICATION TO OPERATE AT PEAK EFFICIENCY
  31. 31. © 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017 31 Reconnect the Control Head to the Vacuum Hose and Reattach the Vacuum Pipe to the Control Head
  32. 32. © 2017 Isolite® Systems. All rights reserved. Last Revised: 03/01/2017 32 THANK YOU. PLEASE CONTACT US IF YOU HAVE ANY QUESTIONS OR NEED ANY ADDITIONAL SUPPORT.

×