SlideShare a Scribd company logo
1 of 76
Download to read offline
Look us up on www.additel.com or call today (1)714-998-6899 
3UHFLVLRQ XQGHU SUHVVXUH
Pressure Calibration Equipment
Catalog 2017
Additel Corporation is one of the leading worldwide providers of pressure calibration tools.
We are dedicated to designing, manufacturing, and delivering the highest quality test tools
and portable calibrators for the process and calibration industries. For many years Additel
has successfully developed pressure controlers, automated pressure calibrators, digital
pressure test gauges, digital pressure calibrators, pressure test and calibration pumps, and
multifunction process calibrators. Additel products are currently used in over 70 countries
worldwide.
Product quality and customer service along with innovative engineering have been our top
priorities and will continue to be our guiding principles going forward. We are committed to
customer satisfaction through quality products, competitive pricing, unmatched services/
technical support and continued introduction of new and innovative products.
$'',7(/ 25325$7,21
RUSRUDWH +HDGTXDUWHUV
Additel Corporation
2900 Saturn Street #B
Brea, CA 92821, USA
6DOW /DNH LW 2IILFH
Additel Corporation
1314 West State Rd Suite 101
Pleasant Grove, UT 84062, USA
3KRQH  )D[  (PDLO VDOHV#DGGLWHOFRP
(XURSHDQ 2IILFH
Additel Corporation
Holkebjergvej 79
5250 Odense, Denmark
$VLD 2IILFH
Additel Corporation
Bldg 5, No. 3 Fengxiu Middle Road,
Haidian, Beijing, 100094, China
$XWRPDWLF 3UHVVXUH DOLEUDWRU
Automatic Handheld Pressure Calibrator Selection Guide.......................01
Additel 760 Automatic Handheld Pressure Calibrator..............................02
Automated Pressure Calibrators Selection Guide....................................07
Additel 761 Automated Pressure Calibrators...........................................08
3UHVVXUH RQWUROOHUV
Additel 780 Series Pressure Controller....................................................13
'LJLWDO 3UHVVXUH DOLEUDWRUV  *DXJHV
Additel Digital Pressure Gauges Selection Guide....................................20
Additel 672 Digital Pressure Calibrators.................................................. 22
Application notes......................................................................................25
Additel 681 Digital Pressure Gauges........................................................27
Additel 680 Digital Pressure Gauges........................................................30
Application notes......................................................................................32
0XOWLIXQFWLRQ 3URFHVV DOLEUDWRUV
Multifunction Process Calibrators Selection Guide...................................34
Additel 209 210 Series Loop Calibrator....................................................35
Additel 220 Multifunction Loop Calibrator.................................................37
Additel 221A 222A 223A /WNVKſEVKQP QEWOGPVKPI 2TQEGUU %CNKDTCVQTU
'LJLWDO 3UHVVXUH 0RGXOHV
Additel 160A Series Intelligent Digital Pressure Modules........................42
3UHVVXUH7HVWDOLEUDWLRQ 3XPSV
Pressure Test/Calibration Pumps Selection Guide..................................45
Additel 901A Low Pressure Test Pump...................................................46
Additel 912 Low Pressure Test Pump......................................................47
Additel 914 Handheld Pneumatic Pressure Test Pump...........................48
7$%/( 2) 217(176
Additel 916 Pneumatic Pressure Test Pump........................................... 49
Additel 917 Pneumatic Pressure Test Pump........................................... 50
Additel 918 Pneumatic Pressure Test Pump........................................... 51
Additel 919A Pneumatic High Pressure Test Pump................................ 52
Application notes..................................................................................... 53
Additel 920 Pneumatic High Pressure Test Pump.................................. 54
Additel 925 Handheld Hydraulic Pressure Test Pump............................ 55
Application notes..................................................................................... 56
Additel 927 Hydraulic Pressure Test Pump............................................. 57
Additel 928 Hydraulic Pressure Test Pump............................................. 58
Additel 936 Hydraulic High Pressure Calibration Pump...........................59
Additel 937 Hydraulic High Pressure Calibration Pump...........................60
Additel 938 Hydraulic High Pressure Calibration Pump...........................61
Additel 946 Hydraulic High Pressure Calibration Pump...........................62
Additel 949 Hydraulic Ultra-high Pressure Test Pump.............................63
Additel 959 Hydraulic Ultra-high Pressure Test Pump.............................64
3UHVVXUH 0DQLIROGV....................................................................................65
)LOWHUV.........................................................................................................66
+RVHV $GDSWHUV DQG )LWWLQJV....................................................................67
6RIWZDUH.....................................................................................................70
7$%/( 2) 217(176
Look us up on www.additel.com or call today (1)714-998-6899 
0RGHO
)HDWXUHV
//3 //3'/ ' ''/ 0$ 0$'/
3UHVVXUH
5DQJH@
±30 in H2O
(±75 mbar)
±30 in H2O
(±75 mbar)
-12.5 to 35 psi
(-0.86 to 2.5 bar)
-12.5 to 35 psi
(-0.86 to 2.5 bar)
-12.5 to 300 psig
(-0.86 to 20 bar)
-12.5 to 300 psig
(-0.86 to 20 bar)
$FFXUDF  )6
0.05[1]
0.05[1]
0.02[1]
0.02[1]
0.02[1][3]
0.02[1][3]
6WDELOLW  )6
0.005[2]
0.005[2]
0.005[2]
0.005[2]
0.005[2]
0.005[2]
*DXJH 3UHVVXUH
Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ
'LIIHUHQWLDO 3UHVVXUH
Ŗ Ŗ Ŗ Ŗ
$EVROXWH 3UHVVXUH
Ŗ Ŗ
%DURPHWULF 3UHVVXUH
Ŗ Ŗ
5HPRYDEOH ,QWHUQDO
0RGXOH Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ
([WHUQDO 3UHVVXUH
0RGXOH Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ
7HPSHUDWXUH
RPSHQVDWHG Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ
%XLOWLQ (OHFWULDO 3XPS
Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ
%XLOWLQ )LOWHU
Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ
%XLOWLQ /LTXLG7UDS
Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ
6RXUFH6LPXODWH  P$
Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ
0HDVXUH P$
RU 9 Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ
9 /RRS SRZHU
Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ
3UHVVXUH VZLWFK WHVW
Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ
+$57 RPPXQLFDWLRQ
Ŗ Ŗ Ŗ
7DVN 'RFXPHQWDWLRQ
Ŗ Ŗ Ŗ
'DWD /RJJLQJ
Ŗ Ŗ Ŗ
KDQQHOV 4 4 4 4 4 4
86%:L)L DQG
%OXHWRRWK Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ
$XWRPDWLF +DQGKHOG 3UHVVXUH DOLEUDWRU
6HOHFWLRQ *XLGH
$GGLWHO 
@ )6 VSHFL¿FDWLRQ DSSOLHV WR WKH VSDQ RI WKH PRGXOH UDQJH
@ 6WDELOLW EDVHG RQ )6 RI WKH LQWHUQDO SUHVVXUH PRGXOH ,QWHUQDO PRGXOH LV VZLWFKDEOH
@ 6SHFL¿FDWLRQ EDVHG RQ JDXJH PHDVXUHPHQW $Q DGGLWLRQDO  SD XQFHUWDLQW ZLOO QHHG WR EH LQFOXGHG ZKHQ PHDVXULQJ LQ DEVROXWH PRGH
Pressure/ProcessCalibrationEquipmentAdditelCatalog2016
Look us up on www.additel.com or call today (1)714-998-6899
$XWRPDWLF +DQGKHOG 3UHVVXUH DOLEUDWRU
$GGLWHO 
)XOO DXWRPDWLF FDOLEUDWRU ZLWK EXLOWLQ SXPS DQG FRQWUROOHU
6ZLWFKDEOH LQWHUQDO SUHVVXUH PRGXOHV IRU H[SDQGDEOH UDQJHV
$FFXUDF  HDU
RI  )6
([WHUQDO SUHVVXUH PRGXOHV DYDLODEOH PHDVXUH RQO
/HVV WKDQ  OEV  NJ
IRU KDQGKHOG RSHUDWLRQ
6RXUFH SUHVVXUH PHDVXUH SUHVVXUH DQG HOHFWULFDO
 FKDQQHOV
2SWLRQDO +$57 FRPPXQLFDWLRQV
2SWLRQDO GDWD ORJJLQJ DQG WDVN GRFXPHQWLQJ
86%:L)L DQG %OXHWRRWK FRPPXQLFDWLRQV
29(59,(:
A portable automated pressure calibrator in the palm of your hand—this could be
our most exciting product yet! The Additel 760 series Automatic Handheld Pressure
Calibrator takes portable pressure calibration to new levels. Weighing less than 4 lbs
(1.8 kg), the ADT760’s innovative design contains a built-in pump, precision pressure
sensor, internal controller and a large touch-screen color display. To generate
pressure, simply key in the desired pressure and the Additel 760 will do the rest.
Each unit has four channels: one internal pressure channel for source and measure
pressures, two external pressure measurement channels, and one electronic measure
and source channel. This series of calibrator has three standard models with the option
of adding HART communications, documentation and data logging.
Look us up on www.additel.com or call today (1)714-998-6899 
)($785(6
ADT760-LLP
The 760-LLP is designed for low pressure calibration and comes with a built-in pressure module to ±30 inH2O (±75 mbar)
and provides an accuracy to 0.05%FS. If you need to go lower in pressure, you can purchase additional internal pressure
modules (ADT155 series), which for the 760-LLP provide a variety of ranges down to ±0.25 inH2O (±0.62 mbar). The
accuracy of 0.05%FS and control stability 0.005%FS is based on the internal module’s span. Measurements can be
made in gauge or differential mode.
ADT760-D
The 760-D gives you differential and gauge pressure but at a higher pressure range than the ADT760-LLP. Covering the
range of -12.5 to 35 psi (-0.86 to 2.5 bar) and with an accuracy of 0.02%FS, the ADT760-D is an ideal solution to cover
very common gauge and differential pressure measurements. The Additel 760-D comes with an internal module covering
its full range, but lower pressure modules down to ±10 inH2O differential can be purchased to improve accuracy at lower
pressures.
ADT760-MA
The 760-MA generates and controls pressure from vacuum pressures up to 300 psig (20 bar.g)
with an accuracy of 0.02%FS. Equipped with a built-in barometric reference, each unit can switch
between gauge and absolute pressure types. A variety of internal sensors are available which
offer lower pressure ranges for improved performance.
Documenting Process Functionality
Each model of the Additel 760 series has an option incorporating documentation and
communication functions turning your 760 into a multifunction documenting process calibrator.
This feature provides HART communication, task documentation and data logging.
6SHFLILFDWLRQ //3 ' 0$
0D[ 3UHVVXUH 5DQJH
±30 inH2O
(75 mbar)
-12.5 to 35 psi
(-0.86 to 2.5 bar)
-12.5 to 300 psig
(-0.86 to 20 bar.g)
$FFXUDF 0.05%FS[1]
0.02%FS[1]
0.02%FS[1][3]
6WDELOLW 0.005%FS[2]
0.005%FS[2]
0.005%FS[2]
3UHVVXUH7SH Differential, Gauge Differential, Gauge Gauge, Absolute
2YHU 5DQJH ,QGLFDWLRQ 120%
5HVROXWLRQ 6 digits
0HDVXUHPHQW 8QLWV
Pa, hPa, kPa, mPa, bar, mbar, psi, mmHg@0°C, cmHg@0°C, mHg@0°C, inHg@0°C, inH2O@4°C, mmH2O@4°C,
cmH2O@4°C, mH2O@4°C, mmH2O@20°C, cmH2O@20°C, mH2O@20°C, inH2O@20°C, kg/m2, mtorr, torr, lb/ft2, tsi, custom
%DURPHWULF $FFXUDF N/A N/A 60 Pa[4]
RQQHFWLRQ $CTD ſVVKPI
*QUG  HV 
 O YKVJ DWKNVKP ſNVGT VQ
1/4BSPF, 1/4NPTF, and M20F adapters
*QUG  HV 
 O YKVJ DWKNVKP ſNVGT VQ
1/4BSPF, 1/4NPTF, and M20F adapters
3UHVVXUH 2XWSXW 5DWH 30 Seconds (30 inH2O/100 ml) 10 Seconds (35 psi/5 ml) 90 Seconds (300 psi/5 ml)
3UHVVXUH 6SHFLILFDWLRQV
@ )6 VSHFL¿FDWLRQ DSSOLHV WR WKH VSDQ RI WKH PRGXOH UDQJH
@ 6WDELOLW EDVHG RQ )6 RI WKH LQWHUQDO SUHVVXUH PRGXOH 6WDELOLW LV  )6 RU  SD ZKLFKHYHU LV JUHDWHU ,QWHUQDO PRGXOH LV VZLWFKDEOH
@ 6SHFL¿FDWLRQ EDVHG RQ JDXJH PHDVXUHPHQW $Q DGGLWLRQDO  SD XQFHUWDLQW ZLOO QHHG WR EH LQFOXGHG ZKHQ PHDVXULQJ LQ DEVROXWH PRGH
@  3D XQFHUWDLQW N
LQFOXGHV FDOLEUDWLRQ XQFHUWDLQW OLQHDULW DQG ORQJWHUP VWDELOLW  3D SHU HDU
%DURPHWHU UDQJH RI  WR  N3D
6SHFLILFDWLRQ 5DQJH 5HVROXWLRQ $FFXUDF 1RWH
P$ 0HDVXUH ±30 mA 0.0001 mA 0.01%RD+0.005%FS +ORGFCPEG š
9 0HDVXUH ±30 V 0.0001 V 0.01%RD+0.005%FS +ORGFCPEG /š
P$ 6RXUFH 24 mA 0.001 mA 0.01%RD+0.005%FS 20 mA @ 1K
/RRS 3RZHU 6RXUFH 25 V N/A ±1 V 100 mA (Max Loading)
3UHVVXUH 6ZLWFK Open, close. Support for mechanical switches and NPN/PNP digital switches.
7HPSHUDWXUH RPSHQVDWLRQ 41°F to 95°F (5°C to 35°C)
7HPSHUDWXUH RHIILFLHQW  ± ( 0.001%RD + 0.001%FS ) / °C outside of 5°C to 35°C
(OHFWULFDO 6SHFLILFDWLRQV
Pressure/ProcessCalibrationEquipmentAdditelCatalog2016
Look us up on www.additel.com or call today (1)714-998-6899
0RGXOH
0RGXOH 5DQJH
0HGLD
$FFXUDF
 )6
@
%XUVW
3UHVVXUH
//3 ' 0$
LQ+2 PEDU
'3[4]
±0.25 ±0.62 G 0.2[2]
100x
Ŗ
'3[4]
±0.5 ±1.25 G 0.1[2]
100x
Ŗ
'3[4]
±1 ±2.5 G 0.05[2]
100x
Ŗ
'3[4]
±2 ±5 G 0.05[2]
100x
Ŗ
'3 ±5 ±10 G 0.05[2]
50x
Ŗ
'3 ±10 ±25 G 0.05[2]
20x
Ŗ Ŗ
'3 ±20 ±50 G 0.05 20x
Ŗ Ŗ
'3 ±30 ±75 G 0.05 20x
Ŗ Ŗ
'3 ±50 ±160 G 0.05 3x
Ŗ
'3 ±150 ±350 G 0.02 3x
Ŗ
'3 ±300 ±700 G 0.02 3x
Ŗ
'3
-350 to 400
(-12.5 to 15 psi)
-860 to 1K
(-0.86 to 1 bar)
G 0.02 3x
Ŗ
'3
-350 to 800
(-12.5 to 30 psi)
-860 to 2K
(-0.86 to 2 bar)
G 0.02 3x
Ŗ
'3.
-350 to 1K
(-12.5 to 35 psi)
-860 to 2.5K
(-0.86 to 2.5
bar)
G 0.02 3x
Ŗ
*DXJH
3UHVVXUH
psi bar
3 ±10 ±0.7 G 0.02[3]
3x
Ŗ Ŗ
3 ±15 ±1 G 0.02[3]
3x
Ŗ Ŗ
3 -12.5 to 30 -0.86 to 2 G 0.02[3]
3x
Ŗ Ŗ
3 -12.5 to 35 -0.86 to 2.5 G 0.02[3]
3x
Ŗ Ŗ
3 -12.5 to 50 -0.86 to 3.5 G 0.02[3]
3x
Ŗ
3 -12.5 to 100 -0.86 to 7 G 0.02[3]
3x
Ŗ
3 -12.5 to 150 -0.86 to 10 G 0.02[3]
3x
Ŗ
3 -12.5 to 200 -0.86 to 14 G 0.02[3]
3x
Ŗ
3 -12.5 to 300 -0.86 to 20 G 0.02[3]
3x
Ŗ
,QWHUQDO 0RGXOHV 6SHFLILFDWLRQV DQG RPSDWLELOLW
=? (5 URGEKſECVKQP CRRNKGU VQ VJG URCP QH VJG OQFWNG TCPIG #EEWTCE[ KPENWFGU QPG[GCT UVCDKNKV[ GZEGRV HQT 2 VQ 2 OQFWNGU
[2] 0.05%FS accuracy (incl 6-month stability). One-year accuracy is 0.05%FS calibration accuracy combined with 0.05%FS one-year stability.
=? 5RGEKſECVKQP DCUGF QP ICWIG OGCUWTGOGPV #P CFFKVKQPCN  RC WPEGTVCKPV[ YKNN PGGF VQ DG KPENWFGF YJGP OGCUWTKPI KP CDUQNWVG OQFG #RRNKECDNG QPN[
for use with the ADT760-MA
[4] 0.05 Pa pressure control stability
Look us up on www.additel.com or call today (1)714-998-6899 
*HQHUDO 6SHFLILFDWLRQV
6SHFLILFDWLRQ 'HVFULSWLRQ
KDQQHOV
Four total: one electrical, one internal
pressure, two external pressure (measure
only)
(QFORVXUH ,3 5DWLQJ IP54 water and dust proof
%DWWHU
Rechargeable Li-Ion battery, typically 10
hours of operation, recharges in less than 4
hours
'LVSOD Color 800 x 480 TFT 5-inch touch screen
RPPXQLFDWLRQV USB, WiFi, Bluetooth
:HLJKW 4 lbs (1.8 kg)
6L]H 9.3 x 4.3 x 2.8 in (235 x 110 x 70 mm)
HUWLILFDWLRQ 0+56VTCEGCDNG EGTVKſECVG YKVJ FCVC KPENWFGF
+$57 RPPXQLFDWLRQV Optional (ADT760-X-DL model)
'DWD /RJJLQJ
Optional (ADT760-X-DL model), up to
1,000,000 readings (date and time stamped)
7DVN 'RFXPHQWDWLRQ
Optional (ADT760-X-DL model) up to 250
tasks
$XWRPDWLRQ )XQFWLRQV Switch test, auto step, leak test
0LVXVH 3URWHFWLRQ Up to 30 V on any two sockets
0XOWL /LQJXDO ,QWHUIDFH
English, German, French, Italian, Spanish,
Portuguese, Chinese, Japanese, and
Russian
3XPS /LIH 500,000 cycles
3RZHU
Rechargeable Li-Ion battery, external power:
110/220 V power adapter 10 V
(QYLURQPHQW 6SHFLILFDWLRQV
Operation: 32°F to 122°F (0°C to 50°C),
0-90% RH, less than 3,000 meters
Compensated temperature: 32°F to 122°F
(0°C to 50°C)
Storage temperature: -4°F to 158°F (-20°C to
70°C)
9LEUDWLRQ DQG 6KRFN
Vibration: 4g (20 to 2,000 Hz)
Shock: 8g, 1 meter drop test
RPSOLDQFH CE
:DUUDQW 1 year
2SWLRQDO ,QWHUQDO 3UHVVXUH 0RGXOHV
0RGXOH 1XPEHU 'HVFULSWLRQ
$'7'3
Pressure module for ADT760-LLP, ±0.25
inH2O, ±0.2%FS
$'7'3
Pressure module for ADT760-LLP, ±0.5
inH2O, ±0.1%FS
$'7'3
Pressure module for ADT760-LLP, ±1 inH2O,
±0.05%FS
$'7'3
Pressure module for ADT760-LLP, ±2 inH2O,
±0.05%FS
$'7'3
Pressure module for ADT760-LLP, ±5 inH2O,
±0.05%FS
$'7'3
Pressure module for ADT760-LLP  -D ±10
inH2O, ±0.05%FS
$'7'3
Pressure module for ADT760-LLP  -D, ±20
inH2O, ±0.05%FS
$'7'3
Pressure module for ADT760-LLP  -D, ±30
inH2O, ±0.05%FS
$'7'3
Pressure module for ADT760-D, ±50 inH2O,
±0.05%FS
$'7'3
Pressure module for ADT760-D, ±150 inH2O,
±0.02%FS
$'7'3
Pressure module for ADT760-D, ±300 inH2O,
±0.02%FS
$'7'3
Pressure module for ADT760-D, -350 to 400
inH2O (-12.5 to 15 psi), ±0.02%FS
$'7'3
Pressure module for ADT760-D, -350 to 800
inH2O (-12.5 to 30 psi), ±0.02%FS
$'7'3.
Pressure module for ADT760-D, -350 to 1K
inH2O (-12.5 to 35 psi), ±0.02%FS
$'73
Pressure module for ADT760-D  -MA, ±10
psi, ±0.02%FS
$'73
Pressure module for ADT760-D  -MA, -12.5
to 15 psi, ±0.02%FS
$'73
Pressure module for ADT760-D  -MA, -12.5
to 30 psi, ±0.02%FS
$'73
Pressure module for ADT760-D  -MA, -12.5
to 35 psi, ±0.02%FS
$'73
Pressure module for ADT760-D  -MA, -12.5
to 50 psi, ±0.02%FS
$'73
Pressure module for ADT760-MA, -12.5 to
100 psi, ±0.02%FS
$'73
Pressure module for ADT760-MA, -12.5 to
150 psi, ±0.02%FS
$'73
Pressure module for ADT760-MA, -12.5 to
200 psi, ±0.02%FS
$'73
Pressure module for ADT760-MA, -12.5 to
300 psi, ±0.02%FS
Pressure/ProcessCalibrationEquipmentAdditelCatalog2016
Look us up on www.additel.com or call today (1)714-998-6899
$FFHVVRULHV ,QFOXGHG
0RGHO 4XDQWLW 3LFWXUH
ADT100-760-KIT
adapter set
(excl ADT760-LLP)
1 set (to 1/4BSPF, to
1/4NPTF, to M20F)
ADT100-760 Hose
(excl ADT760-LLP)
1 pc
9816-X 110V/220V
external power adapter
1 pc
9724 chargeable Li-ion
battery
1 pc
9025 test leads for
calibrator
1 sets (3 pcs)
CD Manual 1 pc
NIST-traceable
EGTVKſECVG
1 pc
* Additel/Land software available for free download at www.additel.com
2SWLRQDO $FFHVVRULHV
0RGHO 'HVFULSWLRQ 3LFWXUH
ADT160A
Pressure modules (reference
160A data sheet; module
connection cable sold
separately)
9060
Pressure module connection
cable
9510
Additel/Cal task management
software
ADT100-760-N
Special Male Connector to
1/4NPT quick connector
ADT100-760-B
Special Male Connector to
1/4BSP quick connector
ADT100-760-M
Special Male Connector to
M20x1.5 quick connector
ADT100-760-N2
Special Male Connector to
1/2NPT quick connector
ADT100-760-B2
Special Male Connector
to1/2BSP quick connector
ADT128-B
Pneumatic Pressure Manifold
for ADT760, -15 to 3000 psi,
single port, 1/4 BSP F
ADT128-B2
Pneumatic Pressure Manifold
for ADT760, -15 to 3000 psi,
single port, 1/2 BSP F
ADT128-N
Pneumatic Pressure Manifold
for ADT760, -15 to 3000 psi,
single port, 1/4 NPT F
ADT128-N2
Pneumatic Pressure Manifold
for ADT760, -15 to 3000 psi,
single port, 1/2 NPT F
ADT128-M
Pneumatic Pressure Manifold
for ADT760, -15 to 3000 psi,
single port, M20x1.5
9240 DP gauge holder, rubber
1220211087 Filter, set of 2 pcs
9913-760-SC
Soft carrying case for
ADT760, test leads, and
many accessories
9914-760
Carry case for ADT760 and
various accessories
ADT100-760-CNT
Special male connector to
special male connector (to
be used to adapt from the
ADT760 to ADT100-760-KIT
adapters)
2UGHULQJ ,QIRUPDWLRQ
0RGHO7SH
0RGHO 1XPEHU 'HVFULSWLRQ
$'7//3
Automatic Handheld Pressure Calibrator, ±30
inH2O (DP30 sensor included)
$'7//3'/
Automatic Handheld Pressure Calibrator, ±30
inH2O (DP30 sensor included) w/HART and
data logging
$'7'
Automatic Handheld Pressure Calibrator,
-12.5 to 35 psi (CP35 sensor included)
$'7''/
Automatic Handheld Pressure Calibrator,
-12.5 to 35 psi (CP35 sensor included) w/
HART and data logging
$'70$
Automatic Handheld Pressure Calibrator,
-12.5 to 300 psi (CP300 sensor included)
$'70$'/
Automatic Handheld Pressure Calibrator,
-12.5 to 300 psi (CP300 sensor included) w/
HART and data logging
3UHVVXUH UDQJH
See Model Type table.
$'7 //3
0RGHO 1XPEHU
Look us up on www.additel.com or call today (1)714-998-6899 
0RGHO
)HDWXUHV
//3 ' / /$ 0 0$ + +$ %3
3UHVVXUH
5DQJH
-25 to 25 mbar
(-10 to 10 inH2O)
-0.95 to 1 bar
(-13.5 to 15 psi)
-0.95 to 7 bar
(-13.5 to 100
psi)
0.05 to 8 bar.a
(1.2 to 115 psi.a)
-0.90 to 25 bar
(-13 to 375 psi)
0.1 to 26 bar.a
(1.7 to 390 psi.a)
-0.90 to 40 bar
(-13 to 600 psi)
0.1 to 41 bar.a
(1.7 to 615 psi.a)
100 to 1200 hPa
3UHVVXUH
0RGXOH 
-2.5 to 2.5 mbar
(-1 to 1 inH2O)
-25 to 25 mbar
(-10 to 10 inH2O)
-0.95 to 2.5 bar
(-13.5 to 35 psi)
0.05 to 3.5 bar.a
(1.2 to 50 psi.a)
-0.9 to 2.5 bar
(-13 to 35 psi)
0.1 to 3.5 bar.a
(1.7 to 50 psi.a)
-0.9 to 2.5 bar
(-13 to 35 psi)
0.1 to 3.5 bar.a
(1.7 to 50 psi.a)
100 to 1200 hPa
3UHVVXUH
0RGXOH 
-25 to 25 mbar
(-10 to 10 inH2O)
-0.95 to 1 bar
(-13.5 to 15 psi)
0 to 7 bar
(0 to 100 psi)
0.05 to 8 bar.a
(1.2 to 115 psi.a)
0 to 25 bar
(0 to 375 psi)
0.1 to 26 bar.a
(1.7 to 390 psi.a)
0 to 40 bar
(0 to 600 psi)
0.1 to 41 bar.a
(1.7 to 615 psi.a)
NA
'LIIHUHQWLDO
3UHVVXUH Ŗ Ŗ
*DXJH 3UHVVXUH
Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ
$EVROXWH
3UHVVXUH Ŗ Ŗ Ŗ Ŗ
%DURPHWULF
3UHVVXUH Ŗ Ŗ Ŗ Ŗ
7HPSHUDWXUH
RPSHQVDWHG Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ
%XLOWLQ (OHFWULDO
3XPS Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ
%XLOWLQ )LOWHU
Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ
%XLOWLQ /LTXLG
7UDS Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ
6RXUFH  WR
P$ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ
0HDVXUH P$
RU 9 Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ
0HDVXUH
H[WHUQDO
SUHVVXUH
PRGXOH
Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ
9 /RRS SRZHU
Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ
3UHVVXUH VZLWFK
WHVW Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ
+$57
RPPXQLFDWLRQ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ
'RFXPHQWLQJ
Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ
$XWRPDWHG 3UHVVXUH DOLEUDWRUV
6HOHFWLRQ *XLGH
$GGLWHO
Pressure/ProcessCalibrationEquipmentAdditelCatalog2016
Look us up on www.additel.com or call today (1)714-998-6899
$XWRPDWHG 3UHVVXUH DOLEUDWRUV
$GGLWHO 
)XOO DXWRPDWHG SUHVVXUH FDOLEUDWRU ZLWK EXLOWLQ SUHVVXUH JHQHUDWRU  FRQWUROOHU
WR DV KLJK DV  SVL  EDU
RU DV ORZ DV  3D  LQ+2
)6 DFFXUDF
'XDO SUHVVXUH PRGXOHV
%XLOWLQ ILOWHU DQG OLTXLG WUDS ZLWK YHQWLQJ VVWHP SUHYHQWV FRQWDPLQDWLRQ RI WKH
FDOLEUDWRU
3RUWDEOH  OE
29(59,(:
With a built-in high performance electronic pump and precision pressure controller,
the 761 series portable automated pressure calibrators provide a turnkey solution for
ECNKDTCVKQP QH ICWIGU VTCPUOKVVGTU CPF UYKVEJGU DQVJ KP VJG ſGNF CPF KP VJG NCDQTCVQT[
In a portable package, this calibrator can automatically generate pressures from 90%
vacuum to 600 psi (40 bar) with 0.005% FS pressure control stability and 0.02% FS
accuracy. To improve the calibrator accuracy, two pressure modules with differing
ranges are built-in and integrated with the internal pump and controller.
The 761-LLP is specially designed for low pressure calibration, and pressures can be
set to as low as 0.01 Pa (0.00004 inH2O) with a control stability better than 0.05 Pa
(0.0002 inH2O).
The 761-BP is designed for barometric gauge calibration. With a quartz pressure
sensor, the accuracy of 761-BP can achieve 0.01%FS accuracy.
With optional external pressure modules (160A series), the 761 can measure
pressures up to 10,000 psi (700 bar) with 0.02%FS accuracy. In addition to the
pressure generation, control, and measurement capabilities, the 761 also features
HART communication capability, supplies 24V loop power, and reads the current or
voltage produced by the pressure transducers. The 761 is a state of the art automated
RTGUUWTG ECNKDTCVQT YJKEJ DTKPIU CWVQOCVGF RTGUUWTG ECNKDTCVKQP VQ VJG ſGNF
Look us up on www.additel.com or call today (1)714-998-6899 
)($785(6
3UHVVXUH 5DQJH
Pressure generated by built-in pump and controlled
automatically from 90% vacuum to 600 psi (40 bar)
Pressure measured with external pressure modules
(160A series) to 10,000 psi (700 bar)
$FFXUDF )XOO WHPSHUDWXUH FRPSHQVDWHG
0.02%FS
0.05%FS (ADT761-LLP)
0.01%FS (ADT761-BP)
RQWURO 6WDELOLW
0.005%FS
ADT761-LLP:  0.05 Pa (0.0002 inH2O)
RQWDPLQDWLRQ SUHYHQWLRQ VVWHP
Built-in filter and liquid trap with venting system
prevent contamination of the calibrator
5QNKF RCTVKENGU ECP DG DNQEMGF D[ DWKNVKP ſNVGT
A small amount of liquid can be collected by the built-
in liquid trap
Both solid particles and liquid can be blown out
through the venting system
$XWRPDWLRQ DQG7HVW 3URJUDP
Fully Automated pressure calibration
Task preset and auto run
Auto step
(DV WR 8VH
User friendly interface
Intuitive icon driven menu structure
6RXUFLQJ DQG 0HDVXULQJ HOHFWULFDO VLJQDOV
Measure mA with 0.01% RD + 1.5μA accuracy
Measure V with 0.01% RD + 1.5mV accuracy
Source/sink 0 to 22mA current
24 VDC loop power supplied to pressure transmitter
during test
Automated switch test
0LVXVH 3URWHFWLRQ
Up to 30V misuse protection on any two sockets
'LVSOD
7 LCD color screen (800X480)
'RFXPHQWLQJ DQG RQGHPDQG ORJJLQJ
Task management capability
Internal memory stores 200 tasks
Download tasks and upload results
900 snapshots
+$57 RPPXQLFDWLRQ FDSDELOLW
Support HART instrumentation
3UHVVXUH RXWSXW UDWH
2.4 liters/minute
5HFKDUJHDEOH EDWWHU
Rechargeable battery with up to 8 hours operation
between charges
/HDN WHVWLQJ
Apply leak test to an external pressure system to
determine the magnitude of pressure variations due to
leaks
3LSHOLQH 3URWHFWLRQ Special design the pipeline layout to
avoid possible blockage and leakage
3RUWDEOH Only 12.4 lb
0XOWL /LQJXDO ,QWHUIDFH
English, German, French, Italian, Spanish, Portuguese,
5KORNKſGF %JKPGUG
(Traditional Chinese, Japanese and Russian are
available per request)
1,677UDFHDEOH NIST traceable calibration with data
(included)
Pressure/ProcessCalibrationEquipmentAdditelCatalog2016
Look us up on www.additel.com or call today (1)714-998-6899
7DVN 0DQXDO
%XLOWLQ 8VHU
V 0DQXDO
3UHVVXUH JDXJH  WUDQVPLWWHU  VZLWFK FDOLEUDWLRQ
'RFXPHQWLQJ
RQWURO 6WDELOLW
0.005%FS[1]
;
ADT761-LLP:  0.05 Pa (0.0002 inH2O)
3UHVVXUH RQQHFWLRQV
Pressure connection: ADT761-L/M/H/LA/MA/HA
One ADT100-761 Hose Test Kit with quick connector
1/8 BSP Female on calibrator
Two 1/4NPT(or 1/2NPT, 1/4BSP, 1/2BSP, M20×1.5)
female hand-tight connectors on the external pressure
manifold
Pressure connection: ADT761-LLP/BP
5VCKPNGUU UVGGN JQUG DCTD ſVVKPI
([WHUQDO 3UHVVXUH 0RGXOH RQQHFWLRQ
Accepts Intelligent Pressure Module
(OHFWULFDO RQQHFWLRQ
0.156 inch (Ø4mm) banana sockets
'RFXPHQWLQJ DQG RQGHPDQG ORJJLQJ
Task management capability
Internal memory stores 200 tasks
Download tasks and upload results
'LVSOD
7 LCD color screen (800×480)
Display rate: 5 readings per second
Resolution 4, 5 or 6 digit, user selectable
3UHVVXUH 8QLWV
Pa, kPa, MPa, psi, bar, mbar, inH2O@4°C,
mmH2O@4°C, inH2O@20°C, mmH2O@20°C,
inHg@0°C, mmHg@0°C, kgf/cm2
, and hPa (ADT761-
BP)
3XPS OLIH 100,000 cycles
3RZHU
Battery: Rechargeable Li-ion Polymer battery (included)
Li-Battery working time: ~8 hours
Recharge time: ~3 hours
External Power: 110/220V Power adapter (DC 27 V)
(Included)
(QYLURQPHQWDO 6SHFLILFDWLRQ
Operating temperature: 32°F to 122°F (0°C to 50°C)
Compensated Temperature: 32°F to 122°F (0°C to
50°C)
Storage temperature: -20°C to 60°C (-4°F to 120°F)
Humidity: 90%, non-condensing
(XURSHDQ RPSOLDQFH CE Marked
RPPXQLFDWLRQ
RS232 (DB9/F, environmentally sealed)
Baud rate: 2400, 4800, 9600, 19200, 38400, 57600,
115200
Data length: 7 Bits, 8 Bits
Stop bit: 1 Bit, 2 Bits
Address: from 1 to 121
'LPHQVLRQ
11.65 X 7.32 X 7.09 in (296 X186 X180 mm)
:HLJKW 12.4 to 16.3 lb (5.6 to 7.4 kg)
:DUUDQW 18 months
=? (5 URGEKſECVKQP CRRNKGU VQ VJG URCP QH VJG TCPIG
63(,),$7,216
Look us up on www.additel.com or call today (1)714-998-6899 
$XWRPDWHG 3UHVVXUH DOLEUDWRUV
$GGLWHO 
3UHVVXUH 5DQJHV
(OHFWULFDO 0HDVXUHPHQW DQG 6RXUFH $FFXUDF
63(,),$7,216
=? 1PG [GCT CEEWTCE[ 
KPENWFKPI  [GCT UVCDKNKV[ (5 URGEKſECVKQP CRRNKGU VQ VJG URCP QH VJG OQFWNG TCPIG
[2] 0.05% FS accuracy (including 6 months stability). One year accuracy is 0.05% FS calibration accuracy combined with 0.05% FS one year stability.
* Additel 761 calibrators support 160A series intelligent digital pressure modules that are available for gauge, vacuum and absolute pressure
from -15 psi to 10,000 psi (-1 bar to 700 bar). Accuracy from 0.02%FS includes operation over 14°F to 122°F (-10°C to 50°C), one year
UVCDKNKV[ CPF ECNKDTCVKQP WPEGTVCKPV[ (QT FGVCKNGF URGEKſECVKQP TGHGT VQ # UGTKGU RTGUUWTG OQFWNGU FCVCUJGGV
7SH 0RGHO 3UHVVXUH 5DQJH
3UHVVXUH 0RGXOH  3UHVVXUH 0RGXOH 
5DQJH $FFXUDF@
5DQJH $FFXUDF@
Low/Differential
Pressure
ADT761-LLP
-25 to 25 mbar
(-10 to 10 inH2O)
-2.5 to 2.5 mbar
(-1 to 1 inH2O)
0.05%FS[2] -25 to 25 mbar
(-10 to 10 inH2O)
0.05%FS[2]
ADT761-D
-0.95 to 1 bar
(-13.5 to 15 psi)
-25 to 25 mbar
(-10 to 10 inH2O)
0.05%FS[2] -0.95 to 1 bar
(-13.5 to 15 psi)
0.02%FS
Gauge Pressure
ADT761-L
-0.95 to 7 bar
(-13.5 to 100 psi)
-0.95 to 2.5 bar
(-13.5 to 35 psi)
0.02%FS
0 to 7 bar
(0 to 100 psi)
0.02%FS
ADT761-M
-0.90 to 25 bar
(-13 to 375 psi)
-0.9 to 2.5 bar
(-13 to 35 psi)
0.02%FS
0 to 25 bar
(0 to 375 psi)
0.02%FS
ADT761-H
-0.90 to 40 bar
(-13 to 600 psi)
-0.9 to 2.5 bar
(-13 to 35 psi)
0.02%FS
0 to 40 bar
(0 to 600 psi)
0.02%FS
Gauge/Absolute
Pressure
ADT761-LA
-0.95 to 7 bar
(-13.5 to 100 psi)
-0.95 to 2.5 bar
(-13.5 to 35 psi)
0.02%FS
-0.95 to 7 bar
(-13.5 to 100 psi)
0.02%FS
0.05 to 8 bar.a
(1.2 to 115 psi.a)
0.05 to 3.5 bar.a
(1.2 to 50 psi.a)
0.03%FS
0.05 to 8 bar.a
(1.2 to 115 psi.a)
0.025%FS
ADT761-MA
-0.90 to 25 bar
(-13 to 375 psi)
-0.9 to 2.5 bar
(-13 to 35 psi)
0.02%FS
-0.9 to 25 bar
(-13 to 375 psi)
0.02%FS
0.1 to 26 bar.a
(1.7 to 390 psi.a)
0.1 to 3.5 bar.a
(1.7 to 50 psi.a)
0.03%FS
0.1 to 26 bar.a
(1.7 to 390 psi.a)
0.025%FS
ADT761-HA
-0.90 to 40 bar
(-13 to 600 psi)
-0.9 to 2.5 bar
(-13 to 35 psi)
0.02%FS
-0.9 to 40 bar
(-13 to 600 psi)
0.02%FS
0.1 to 41 bar.a
(1.7 to 615 psi.a)
0.1 to 3.5 bar.a
(1.7 to 50 psi.a)
0.03%FS
0.1 to 41 bar.a
(1.7 to 615 psi.a)
0.025%FS
Barometric pressure ADT761-BP 100 to 1,200 hPa 100 to 1,200 hPa 0.01%FS NA NA
5DQJH 5HVROXWLRQ $FFXUDF
9ROWDJH 0HDVXUHPHQW ±30.0000V 0.1mV ±(0.01%RD + 1.5mV)
XUUHQW 0HDVXUHPHQW ±30.0000mA 0.1μA ±(0.01%RD + 1.5μA)
XUUHQW 6RXUFH 0.000 to 22.000mA 1μA ±(0.02%RD + 2.2μA)
6ZLWFK7HVW If the switch has detective Voltage, the range is from 3V to 24V
' 9 RXWSXW 24V ± 0.5V, max: 50mA
Pressure/ProcessCalibrationEquipmentAdditelCatalog2016
Look us up on www.additel.com or call today (1)714-998-6899
2SWLRQDO $FFHVVRULHV
0RGHO 1XPEHU 'HVFULSWLRQ 3LFWXUH
ADT100-761-N4
Hose Test Kit, 5 feet
ƀGZKDNG JQUG 
(Ø4mm) tube to 1/4NPT
hand-tight quick
connector
ADT100-761-B4
Hose Test Kit, 5 feet
ƀGZKDNG JQUG 
(Ø4mm) tube to 1/4BSP
hand-tight quick
connector
ADT106-N
External Pressure
Manifold (with two
1/4NPT hand-tight quick
connectors)
ADT106-B
External Pressure
Manifold (with two
1/4BSP hand-tight quick
connectors)
ADT106-M
External Pressure
Manifold (with two
M20X1.5 hand-tight
quick connectors)
ADT160A
Pressure modules
(please check Additel
160A datasheet)
9050
USB to RS232 (DB9/M)
Adapter
9060
Pressure module
connection cable
9510
Additel/Cal task
management software
ADT100-761Hose ADT761 hose, 5 feet
$FFHVVRULHV LQFOXGHG
External Pressure
Manifold (ADT106)
([FHSW //3
1 pc

110V/220V external
Power adapter
1 pc

Chargeable Li-ion
battery
1 pc

Test Leads for calibrator
2 sets
(4 pcs)

Short circuit cable
1 set
(2 pcs)

Carrying case for 761
calibrator and accessory
1 pc

Pressure module
connection cable
1 pc

DP gauge holder, rubber
2QO IRU $'7//3
1 pc
$'7
([FHSW //3
1 pc
O-rings([FHSW //3
20 pcs
Manual 1 pc
Allen wrench 1 set
ADT100-761Hose
(Except 761-LLP: 2 pcs)
1 pcs
NIST traceable
ECNKDTCVKQP EGTVKſECVG
1 pc
$XWRPDWHG 3UHVVXUH DOLEUDWRUV
$GGLWHO 
25'(5,1* ,1)250$7,21
3UHVVXUH UDQJH
See pressure range
table.
3UHVVXUH SRUW WSH RQ ([WHUQDO
3UHVVXUH 0DQLIROG
N - 1/4NPT female
N2 - 1/2NPT female
B - 1/4BSP female
B2 - 1/2BSP female
M - M20X1.5 female
* Additel/Land software could be downloaded for free at www.additel.com
$'7 / 1
0RGHO 1XPEHU
Look us up on www.additel.com or call today (1)714-998-6899 
$GGLWHO 
6HULHV 3UHVVXUH RQWUROOHU
3UHVVXUH UDQJHV IURP YDFXXP WR  SVL  EDU
5HPRYDEOH LQWHUFKDQJHDEOH LQWHOOLJHQW VHQVRUV
3UHFLVLRQ DFFXUDF RI  RI UHDGLQJ SOXV  RI IXOO VFDOH
6WDQGDORQH VROXWLRQ WR  SVL  EDU
QR JDV ERWWOH UHTXLUHG
ZKHQ XVHG ZLWK WKH (OHFWULF 3XPS DQG . *DV %RRVWHU
6WDQGDORQH VROXWLRQ WR  SVL  EDU
QR JDV ERWWOH UHTXLUHG
ZKHQ XVHG ZLWK WKH (OHFWULF 3XPS
6WDQGDUG PRGHO DFFXUDF RI  RI IXOO VFDOH
([WHUQDO SUHVVXUH PRGXOHV WR  SVL  EDU
:L)L HQDEOHG FRPPXQLFDWLRQV
)XOO WHPSHUDWXUH FRPSHQVDWHG DFFXUDF RYHU °C WR °C
+$57 RPPXQLFDWLRQ DQG 3URILEXV 3$
/DUJH ´ FRORU WRXFK VFUHHQ GLVSOD
RQWURO VWDELOLW RI  )6
%XLOWLQ EDURPHWHU
(DVWRXVH LFRQ EDVHG XVHU LQWHUIDFH
29(59,(:
For years, we’ve provided the most durable, accurate, quality pressure calibration
products for field applications. The Additel 780 series controller incorporates the same
durability, accuracy, and quality into a new bench top controller packed with features
and functionality that is remarkably easy to use. The Additel 780 series offers two base
ranges: to 1,000 psi (70 bar) and to 3,000 psi (200 bar). The base range establishes the
OCZKOWO EQPVTQNNKPI TCPIG QH VJG EQPVTQNNGT 'CEJ EQPſIWTCVKQP KPENWFGU C EQPVTQN UGPUQT
which is preselected to the sensor range best suited for your application. External and
internal sensors can be used which allows for expanded range and accuracy capability in
the future.
There are also two controller types that can be selected: the Additel 780S is the standard
controller option without any measurement capability. The Additel 780 has expanded
HWPEVKQPCNKV[ KPENWFKPI GNGEVTKECN OGCUWTGOGPV CPF *#46 CPF 2TQſDWU 2# EQOOWPKECVKQP
Pressure/ProcessCalibrationEquipmentAdditelCatalog2016
Look us up on www.additel.com or call today (1)714-998-6899
02'8/$5 '(6,*1
Each unit comes with one Intelligent Pressure Module configured to the many range offerings provided. Standard
accuracy sensors (ADT160A-CPXXX) are silicone pressure sensors with a 1 year accuracy of 0.02% FS. The
RTGEKUKQP SWCTV DCUGF UGPUQTU KORTQXG VJG  [GCT CEEWTCE[ URGEKſECVKQP VQ  QH TGCFKPI 
  (5 'CEJ
sensor has been specially aged, tested and screened before assembly. After assembly each sensor is temperature
compensated over the range of 0°C to 50°C. The Additel 780 series allows for one
internal pressure sensor and one external pressure sensor. The modular design of this
unit provides for interchangeability of both the internal and external sensors with other
Intelligent Pressure Modules.
In addition to the Intelligent Pressure Modules, the Additel 780 series has a built in
barometric sensor. This allows for switching between gauge pressures to absolute
pressures.
67$1'$/21( 62/87,21
Typical pressure controllers will require a nitrogen bottle for the gas supply which make it
FKHſEWNV VQ OQXG VJG EQPVTQNNGT CTQWPF YKVJQWV JCXKPI VQ OQXG QT EQPPGEV VQ CPQVJGT DQVVNG
The Additel 780 series is unique in that with the optional electric pump, you can generate
pressures to 1,000 psi (70 bar) without the need of a gas bottle. Add the booster to the
EQPſIWTCVKQP CPF  RUK 
 DCT KU CEJKGXCDNG YKVJ VJG #FFKVGN  EQPVTQNNGT UGTKGU
The 780 Series Pressure Controller is considered to handle a very wide range of applications
which may normally require a pressure calibrator. Built-in capability, includes current and
voltage measurement capability, 24 volt loop power, HART®
CPF 2TQſDWU 2# EQOOWPKECVKQP
switch measurement capability, and much more.
6SHFLILFDWLRQ $'7. $'7.
3UHVVXUH UDQJH -14.5 to 1,000 psi (-0.95 to 70 bar) -14.5 to 3,000 psi (-0.95 to 200 bar)[1]
RQWURO VWDELOLW 0.003% FS (stability based on % FS of control sensor range)
3UHFLVLRQ
LQFOXGHV  HDU VWDELOLW
See pressure range table
HDU VWDELOLW
0.01% FS standard accuracy sensors
0.005% FS precision accuracy sensors
0HGLD Clean gas
2YHUUDQJH LQGLFDWLRQ 103% to 120% (based on sensor)
5HVROXWLRQ 4, 5, 6, or 7 digits (user selectable)[2]
3UHVVXUH WSH Gauge, Absolute
:DUP XS WLPH 15 minutes
7SLFDO SUHVVXUH VHWWOLQJ WLPH
ZLWKLQ  )6
20 seconds (10%FS pressure change assuming 50 mL test volume)
0HDVXUHPHQW XQLWV
Pa, hPa, kPa, MPa, bar, mbar, psi, mmHg@0°C,cmHg@0°C, mHg@0°C, inHg@0°C,
inH2O@4°C, mmH2O@4°C, cmH2O@4°C, mH2O@4°C, mmH2O@20°C, cmH2O@20°C,
mH2O@20°C, kg/m2
, kg/cm2
, mtorr, torr, atm, lb/ft2
, tsi, user selectable
0LQLPXP FRQWURO SUHVVXUH@
0.0001 psi 0.001 psi
[1] Gas booster or HP gas supply required to reach 3,000 psi (200 bar). Booster ordered separately.
[2] 7 digit resolution for precision model only.
[3] Dependent on pressure module.
0($685(0(17 63(,),$7,216
352(66 )817,21$/,7 H[FO $'76 YHUVLRQV
Look us up on www.additel.com or call today (1)714-998-6899 
6SHFLILFDWLRQ 5DQJH 5HVROXWLRQ $FFXUDF
9ROWV '
-30 to 30 V 0.1 mV ±0.01% rdg + 1.5 mV
-300 to 300 mV 1 μV ±0.01% rdg + 15 μV
XUUHQW ' -30 to 30 mA 0.1 μA ±0.01% rdg + 1.5 μA
6ZLWFK WHVW If the switch has detected voltage, the range is from 3 – 24 V
' 9 RXWSXW 24V ± 0.24V, max 30 mA
[1] Not available in ADT780S versions
(/(75,$/ 0($685( 63(,),$7,216@
6SHFLILFDWLRQ ADT780-1K  ADT780-3K
3RZHU 100 to 240V, 50/60 Hz
3UHVVXUH SRUWV G1/8 F
6WRUDJH WHPSHUDWXUH -20°C to 70°C
2SHUDWLQJ HQYLURQPHQW 0-90% RH non-condensing
'LVSOD 7 inch (17.8 cm) color, touch screen display
:HLJKW 33 lbs (15 kg)
'LPHQVLRQV ':+
16.5 x 17.3 x 5.2 inch (419 x 440 x 132 mm)
0RXQWLQJ Standard desktop, optional rack mount kit
6KRFN 4G
9LEUDWLRQ 1G 10Hz~500Hz
3+6,$/ 63(,),$7,216
:L)L VSHFLILFDWLRQV 802.11 b, g, and n
9HQW Front panel vent and safety release button (only for 780S)
RPPXQLFDWLRQV RS232,USB,LAN,WIFI
6WDELOLW LQGLFDWRU User selectable
7SLFDO VOHZ UDWH Typically 20 seconds
2SHUDWLQJ PRGHV Control, measure, and vent
'LVSOD PRGHV
Controller – show pressure indication and control
Calibrator – shows pressure indication, control, and electrical measurement (excl ADT780S)
/RFDOL]DWLRQ 'PINKUJ %JKPGUG 
UKORNKſGF )GTOCP 5RCPKUJ (TGPEJ +VCNKCP 2QTVWIWGUG 4WUUKCP ,CRCPGUG
RQIRUPLW CE
DOLEUDWLRQ FHUWLILFDWLRQ 0+56VTCEGCDNG EGTVKſECVG QH ECNKDTCVKQP YKVJ FCVC KPENWFGF
:DUUDQW 1 year
27+(5 63(,),$7,216
6HQVRU0RGH *DXJH $EVROXWH
ADT160-02-CPXX N/A 40 Pa[1]
ADT160-01-APXXQ 3 Pa[2]
N/A
Barometer range (60~110)kPa, the accuracy is 40 Pa (11 Pa is optional)
[1] 40 Pa uncertainty (k=2) includes calibration uncertainty, linearity, and long term stability (30 Pa per year). Barometer range of 60 to 120 kPa.
[2] Combined linearity, hysteresis, and repeatability. Add 3 Pa when used in gauge mode. When using the ADT160-01-APXXQ sensors with the
ADT780 controller in gauge mode, regular zeroing will realign the barometric reading to provide the most accurate result.
%$520(75, 0($685(0(17 63(,),$7,21
Pressure/ProcessCalibrationEquipmentAdditelCatalog2016
Look us up on www.additel.com or call today (1)714-998-6899
$'7; ([WHUQDO
0DQLIROG
1 pc
0DQLIROG FRQQHFWLRQ
KRVH
1 pc
3RZHU DGDSWHU 1 pc
7HVW OHDGV
H[FHSW 6
3 sets (6 pc)
 6KRUW FLUFXLW
FDEOH
H[FHSW 6
1 set (2 pc)
 3UHVVXUH PRGXOH
FRQQHFWLRQ FDEOH
1 pc
$GDSWHU VHW DGDSWHUV
WR ILW WKH $'7; SRUW
WR PDOH ILWWLQJV
1 set
(2 pcs G1/8M
to G1/4M, 2 pcs
G1/8M to Festo,
release valve)
' 0DQXDO 1 pc
1,67 WUDFHDEOH
FHUWLILFDWH
1 pc
2ULQJV 2 types
9DFXXP9HQW WXELQJ 2 pc
67$1'$5' $(6625,(6
$'7$
Pressure modules (see module
information)
$'7;
External pneumatic pressure
manifold, 3,000 psi (200 bar),
four hand-tight quick connectors
$'7.%67 3,000 psi (200 bar) gas booster
$'7.(3803
1,000 psi (70 bar) electrical
pump
 RS232 to USB adapter
(;7
RS 232 (DB9/M) extension
cable, 9 feet
$'7)/7 Liquid trap

Shipment case for the ADT780
and ADT780S Controller
 $GGLWHO3DO Manual calibration software
 $GGLWHO$DO
Automated calibration software,
Task and asset management
 Rack mount kit
$'7.LW
Adapter set, total 9 pcs.
G1/8 M to 1/4NPT M (3 pcs)
G1/8 M to G1/4 M (3 pcs)
G1/8 M to M20x1.5 M (3 pcs)
237,21$/ $(6625,(6
$'7 25'(5,1* ,1)250$7,21
0RGHO 1XPEHU
0RGHO
ADT780
ADT780S
+LJK $FFXUDF %DURPHWHU
0DQLIROG SRUW WSH
N-1/4NPT male
N2-1/2NPT male
B-1/4BSP male
B2-1/2BSP male
M-M20x1.5 male
$'7 3 +$%31N
3UHVVXUH UDQJH
See ADT160A range
table on page 17
0D[PXP 5DQJH
3K-3,000 psi (200 bar)
1K-1,000 psi (70 bar)
RQWUROOHU ZLWKRXW . SVL ERRVWHU
Look us up on www.additel.com or call today (1)714-998-6899 
31
3UHVVXUH 5DQJH
0HGLD $FFXUDF
3UHVVXUH 5DWLQJ 6XJJHVWHG
RQWUROOHU
RPSDWLELOLW
,QWHUQDO
([WHUQDO
8VHSVLD EDUD %XUVW 2YHU 3UHVVXUH
3UHFLVLRQ 4XDUW] 6HQVRUV@
$34 0 to 45 3.0 G 0.005% rdg + 0.005% FS 3x 1.2x ADT780-1K only Both
$34 0 to 100 7.0 G 0.005% rdg + 0.005% FS 3x 1.2x ADT780-1K only Both
$34 0 to 300 20 G 0.005% rdg + 0.005% FS 3x 1.2x Both Both
$34 0 to 400 28 G 0.005% rdg + 0.005% FS 3x 1.2x Both Both
$34 0 to 500 35 G 0.005% rdg + 0.005% FS 3x 1.03x Both Both
$3.4 0 to 1,000 70 G 0.005% rdg + 0.005% FS 3x 1.1x Both Both
$3.4 0 to 2,000 140 G 0.005% rdg + 0.005% FS 3x 1.1x ADT780-3K only Both
$3.4 0 to 3,000 200 G 0.005% rdg + 0.005% FS 3x 1.1x ADT780-3K only Both
6WDQGDUG 3UHVVXUH 6HQVRUV
31
3UHVVXUH 5DQJH
0HGLD $FFXUDF
3UHVVXUH 5DWLQJ 6XJJHVWHG
RQWUROOHU
RPSDWLELOLW
,QWHUQDO
([WHUQDO
8VHSVLJ EDUJ %XUVW 2YHU 3UHVVXUH
3 ±10 ±0.7 G 0.02% FS 3x 1.2x ADT780-1K only Both
3 ±15 ±1.0 G 0.02% FS 3x 1.2x ADT780-1K only Both
3 -15 to 30 -1 to 2.0 G 0.02% FS 3x 1.2x ADT780-1K only Both
3 -15 to 50 -1 to 3.5 G 0.02% FS 3x 1.2x ADT780-1K only Both
3 -15 to 100 -1 to 7.0 G,L 0.02% FS 3x 1.2x ADT780-1K only Both
3 -15 to 300 -1 to 20 G,L 0.02% FS 3x 1.2x Both Both
3 -15 to 500 -1 to 35 G,L 0.02% FS 3x 1.2x Both Both
3 -15 to 600 -1 to 40 G,L 0.02% FS 3x 1.2x Both Both
3. -15 to 1,000 -1 to 70 G,L 0.02% FS 3x 1.2x Both Both
3. -15 to 2000 -1 to 140 G,L 0.02% FS 3x 1.2x ADT780-3K only Both
3. -15 to 3,000 -1 to 200 G,L 0.02% FS 3x 1.2x ADT780-3K only Both
3. -15 to 5000 -1 to 350 G,L 0.02% FS 3x 1.2x N/A External
3. -15 to 10,000 -1 to 700 G,L 0.02% FS 2x 1.2x N/A External
[1] Contact Additel for other range options.
**Low pressure sensors (ADT160A-XX-DPX) available for low pressure and differential pressure measurement. Also available for low
pressure control (gauge mode only). Ranges from ±1 inH20 (2.5 mbar) to 300 inH2O (700 mbar).
$'7$ ,QWHOOLJHQW 'LJLWDO 3UHVVXUH 0RGXOHV
*DXJH SUHVVXUH 'LIIHUHQWLDO SUHVVXUH
63(,),$7,216
Pressure/ProcessCalibrationEquipmentAdditelCatalog2016
Look us up on www.additel.com or call today (1)714-998-6899
6WDQGDUG $FFXUDF 3UHFLVLRQ $FFXUDF
3;;; $3.4 $3.4 $34
7HPSHUDWXUH FRPSHQVDWLRQ -10°C to 50°C (14°F to 122°F) 0°C to 50°C (32°F to 122°F)
2SHUDWLQJ WHPSHUDWXUH -10°C to 50°C (14°F to 122°F) 0°C to 50°C (32°F to 122°F)
6WRUDJH WHPSHUDWXUH -20°C to 70°C (-4°F to 158°F) -20°C to 70°C (-4°F to 158°F)
5HODWLYH KXPLGLW 95% RH 95% RH
3UHVVXUH FRQQHFWLRQV
IRU H[WHUQDO XVH RQO
1/4NPT, 1/4BSP, 1/2BSP, M20x15 1/4NPT, 1/4BSP, 1/2BSP, M20x15
(QFORVXUH
IRU H[WHUQDO XVH RQO
SS enclosure SS enclosure
'LPHQVLRQV 'LD [ +
33 mm x 123 mm (1.3” x 4.84”) 44 mm x 220mm (1.7” x 8.7”) 60 mm x 190 mm (2.4” x7.5”)
:HLJKW 0.4 kg (0.99 lb) 0.8 kg (1.8 lb) 1.2 kg (2.6 lb)
[1] Internal sensors used to go inside the ADT780 need to have a M20 connection and connection kit.
25'(5,1* ,1)250$7,21
0RGHO 1XPEHU
0RGHO 5DQJH WSH
PSI - range by psi
BAR - range by bar
([WHUQDO SUHVVXUH SRUW WSH
N - 1/4NPT male
N2 - 1/2NPT male
B - 1/4BSP male
B2 - 1/2BSP male
M - M20x1.5 male
$'7$ 3 36, 1
3UHVVXUH UDQJH
See range table
$FFXUDF
01 – 0.005% rdg + 0.005% FS
02 – 0.02% FS
6(1625 63(,),$7,216
Look us up on www.additel.com or call today (1)714-998-6899 
(/(75, 3803
0LQLPXP LQOHW SUHVVXUH 300 psi (20 bar)
2XW SUHVVXUH UDQJH 300 to 3,600 psi (250 bar)
,QOHW IORZ UDWH 500ml/min
6L]H :+'
17.8 X 7.3 X 23.9 inch (451 X 185 X 607 mm)
9ROXPH 375 mL
:HLJKW 74 lbs (33.6 kg)
3RZHU 100 to 240V
6XSSO PHGLD Nitrogen or clean gas
 SVL  EDU
*DV %RRVWHU 6SHFLILFDWLRQV
3UHVVXUH UDQJH -13.5 to 1,080 psi (-0.95 to 75 bar)
:HLJKW 69.6 lbs (31.6 kg)
6L]H:+'
17.3 X 9.3 X 21.7 inch (440 X 235X 550 mm)
3RZHU 100 to 240V
0HGLD Air
9ROXPH 150 mL
2XWOHW SRUW G1/8F
6WRUDJH WHPSHUDWXUH -20°C to 70°C
2SHUDWLQJ HQYLURQPHQW 0-90% RH non-condensing
0RXQWLQJ Standard desktop, optional rack mount
RQWURO RSHUDWLRQ 2 LED displays with pressure limit settings
7SLFDO PD[ SUHVVXUH WLPH Approx 2 min with ADT780 Controller
 SVL  EDU
(OHFWULF 3XPS 6SHFLILFDWLRQV
$GGLWHO .(3803
%2267(5$GGLWHO .%67
*HQHUDWHV 9DFXXP WR  SVL  EDU
%XLOWLQ ILOWHU DQG OLTXLG WUDS
(OHFWULF 3XPS DQG %RRVWHU
Pressure/ProcessCalibrationEquipmentAdditelCatalog2016
Look us up on www.additel.com or call today (1)714-998-6899
$GGLWHO 3UHVVXUH *DXJH
6HOHFWLRQ *XLGH
6HULHV
)HDWXUH
$'7 6HULHV
'LJLWDO 3UHVVXUH DOLEUDWRU
$'7 6HULHV
'LJLWDO 3UHVVXUH *DXJH
$'7 6HULHV
'LJLWDO 3UHVVXUH *DXJH
*DXJH 3UHVVXUH Ŗ Ŗ Ŗ
RPSRXQG 3UHVVXUH Ŗ Ŗ Ŗ
$EVROXWH 3UHVVXUH Ŗ Ŗ
'LIIHUHQWLDO 3UHVVXUH Ŗ Ŗ
$FFXUDF ODVVHV
0.02%  0.05%FS
15K  20K psi: 0.05%FS
 20K psi: 0.1%FS
0.02%, 0.05%, 0.1%,  0.2%FS  0.1%RD
15K  20K psi: 0.05%, 0.1%, 0.2%FS 
0.1%RD
20K psi: 0.1%  0.2%FS
0.05%, 0.1%,  0.25%FS
 20K psi: 0.1%  0.25%FS
'LJLWDO 'LVSOD Ŗ Ŗ Ŗ
$QDORJ 'LVSOD )DQ6KDSHG
,QGLFDWLRQ
Ŗ
)XOO7HPSHUDWXUH RPSHQVDWLRQ
IURP Ĉ WR Ĉ
Ŗ Ŗ Ŗ
5HVROXWLRQ
'LJLW 5HVROXWLRQ Ŗ
'LJLW 5HVROXWLRQ Ŗ Ŗ Ŗ
6HOHFWDEOH 3UHVVXUH 8QLWV 11 11 19
%DFNOLJKW Ŗ Ŗ Ŗ
2YHU 3UHVVXUH ,QGLFDWLRQ Ŗ Ŗ Ŗ
,6 HUWLILFDWLRQ RSWLRQDO ă QRW
DYDLODEOH IRU SDQHO PRXQW
#6+5 QPN[ 
Ů)2
,3 HUWLILFDWLRQ Ŗ
3DQHO 0RXQW RSWLRQDO
Ŗ
:LUHOHVV 680W only
'DWD /RJJLQJ Ŗ Optional 680W only
0LQ0D[ Ŗ Ŗ Ŗ
%XLOWLQ /HDN7HVW Ŗ Ŗ
+$57 RPPXQLFDWLRQ Ŗ
0HDVXUH P$ DQG 9 Ŗ
9 /RRS 3RZHU Ŗ
6ZLWFK7HVW Ŗ
1,677UDFHDEOH HUWLILFDWH RI
DOLEUDWLRQ
Ŗ Ŗ Ŗ
3RZHU Rechargeable battery 9V battery (120/220V adapter is optional) 2AA batteries
3UHVVXUH
3UHVVXUH 5DQJH
0HGLD
$'7 6HULHV
'LJLWDO 3UHVVXUH DOLEUDWRU
$'7 6HULHV
'LJLWDO 3UHVVXUH *DXJH
$'7 6HULHV
'LJLWDO 3UHVVXUH *DXJH
SVL EDU
*DXJH
9 -15 to 0 -1 to 0 G Ŗ Ŗ Ŗ
*3 0 to 5 0 to 0.35 G Ŗ Ŗ
*3 0 to 10 0 to 0.7 G Ŗ Ŗ
*3 0 to 15 0 to 1 G, L Ŗ Ŗ Ŗ
*3 0 to 30 0 to 2 G, L Ŗ Ŗ Ŗ
*3 0 to 50 0 to 3.5 G, L Ŗ Ŗ
*3 0 to 100 0 to 7 G, L Ŗ Ŗ Ŗ
*3 0 to 300 0 to 10 G, L Ŗ Ŗ Ŗ
*3 0 to 500 0 to 35 G, L Ŗ Ŗ Ŗ
6HULHV
Look us up on www.additel.com or call today (1)714-998-6899 
3UHVVXUH
3UHVVXUH 5DQJH
0HGLD
$'7 6HULHV
'LJLWDO 3UHVVXUH DOLEUDWRU
$'7 6HULHV
'LJLWDO 3UHVVXUH *DXJH
$'7 6HULHV
'LJLWDO 3UHVVXUH *DXJH
SVL EDU
*3 0 to 600 0 to 40 G, L Ŗ Ŗ
*3. 0 to 1K 0 to 70 G, L Ŗ Ŗ Ŗ
*3. 0 to 2K 0 to 140 G, L Ŗ Ŗ
*3. 0 to 3K 0 to 200 G, L Ŗ Ŗ Ŗ
*3. 0 to 5K 0 to 350 G, L Ŗ Ŗ Ŗ
*3. 0 to 10K 0 to 700 G, L Ŗ Ŗ Ŗ
*3. 0 to 15K 0 to 1K G, L Ŗ Ŗ Ŗ
*3. 0 to 20K 0 to 1.4K G, L Ŗ Ŗ Ŗ
*3. 0 to 25K 0 to 1.6K G, L Ŗ Ŗ Ŗ
*3. 0 to 30K 0 to 2K G, L Ŗ Ŗ Ŗ
*3. 0 to 36K 0 to 2.5K G, L Ŗ Ŗ Ŗ
*3. 0 to 40K 0 to 2.8K G, L Ŗ Ŗ Ŗ
RPSRXQG
3 ±2 ±0.16 G Ŗ Ŗ
3 ±5 ±0.35 G Ŗ Ŗ
3 ±10 ±0.7 G Ŗ Ŗ
3 ±15 ±1 G Ŗ Ŗ Ŗ
3 -15 to 30 -1 to 2 G Ŗ Ŗ Ŗ
3 -15 to 100 -1 to 7 G, L Ŗ Ŗ
3 -15 to 300 -1 to 20 G, L Ŗ Ŗ
$EVROXWH
$3 5 0.35 G Ŗ Ŗ
$3 10 0.7 G Ŗ Ŗ
$3 15 1 G Ŗ Ŗ
$3 30 2 G Ŗ Ŗ
$3 50 3.5 G Ŗ Ŗ
$3 100 7 G, L Ŗ Ŗ
$3 300 20 G, L Ŗ Ŗ
$3 500 35 G, L Ŗ Ŗ
$3. 1K 70 G, L Ŗ Ŗ
$3. 3K 200 G, L Ŗ Ŗ
$3. 5K 350 G, L Ŗ Ŗ
'LIIHUHQWLDO inH2O mbar
'3 ±1 ±2.5 G Ŗ Ŗ
'3 ±2 ±5 G Ŗ Ŗ
'3 ±5 ±10 G Ŗ Ŗ
'3 ±10 ±25 G Ŗ Ŗ
'3 ±20 ±50 G Ŗ Ŗ
'3 ±30 ±75 G Ŗ Ŗ
'3 ±50 ±160 G Ŗ Ŗ
'3 ±150 ±350 G Ŗ Ŗ
'3 ±300 ±700 G Ŗ Ŗ
6HULHV
Pressure/ProcessCalibrationEquipmentAdditelCatalog2016
Look us up on www.additel.com or call today (1)714-998-6899
$FFXUDF
ADT672-02: 0.02% of full scale
ADT672-05: 0.05% of full scale
20,000 psi (1,400 bar): 0.1% FS
*DXJH 7SHV
Gauge pressure
Compound Pressure
Absolute pressure
Differential pressure
'LVSOD
Description: Dual-line 6 full digit FSTN LCD with LED
backlight
Display rate: 3.5 readings per second (Default setting)
Numeral display height: 16.5mm (0.65)
3UHVVXUH 8QLWV
Pa, kPa, MPa, psi, bar, mbar, inH2O@4°C,
mmH2O@4°C, inHg@0°C, mmHg@0°C, kgf/cm2
(QYLURQPHQWDO
Compensated Temperature: 14°F to 122°F (-10°C to
50°C)
Operating Temperature: 14°F to 122°F (-10°C to
50°C)
Storage Temperature: -4°F to 158°F (-20°C to 70°C)
Humidity: 95%
29(59,(:
*DXJH SUHVVXUH 'LIIHUHQWLDO SUHVVXUH
#V ſTUV INCPEG VJG  UGTKGU RTGEKUKQP RTGUUWTG ECNKDTCVQTU NQQM NKMG CP QTFKPCT[ RTGUUWTG ICWIG $WV VJKU UGTKGU KU
OWEJ OQTG VJCP QTFKPCT[ CPF FGſPKVGN[ OQTG VJCP LWUV C RTGUUWTG ICWIGōKVŏU C RTGUUWTG ECNKDTCVQT 9KVJ CFXCPEGF
microprocessor technology and state-of-the-art silicon pressure sensors, the 672 series precision pressure calibrators
provide a pressure calibration solution for gauges, transmitters, and switches over a wide pressure range. The 672
is the size of a pressure gauge but with the functionality of a calibrator: It measures pressure precisely with a built-
in pressure sensor, as well as reads the current or mV produced by a transducer. It can even supply an excitation
voltage to power sensors or transmitters during calibration. In order to reach 0.02%FS accuracy up to 10,000 psi
(700 bar) and 0.1%FS accuracy up to 40,000 psi (2,800 bar), every silicon pressure sensor has been specially aged,
tested, and screened before assembly. The 672 series precision pressure calibrators are unmatched in performance
and reliability.
)($785(6
Pressure ranges to 40,000 psi (2800 bar)
Measure mA with 0.01% RD + 1.5 μA accuracy
Measure V with 0.01% RD + 1.5 mV accuracy
Power transmitters during test using 24V loop supply
Pressure switch test
HART Communication capability
Advanced temperature compensation
Dual readout
Min/Max/Hold to capture changing measurements
Data logging
Large, easy to read display with 6-digit resolution
Backlit display
Rechargeable battery or AC adapter
NIST traceable calibration with data (Included)
63(,),$7,216
'LJLWDO 3UHVVXUH DOLEUDWRUV
$GGLWHO 
3UHVVXUH UDQJHV WR  SVL  EDU
+$57 RPPXQLFDWLRQ FDSDELOLW
0HDVXUH P$ RU 9 DQG ZLWK 9 ORRS SRZHU
(DVWRXVH LQH[SHQVLYH SUHVVXUH FDOLEUDWRU ZLWK
XQFHUWDLQW EHWWHU WKDQ  )6
Look us up on www.additel.com or call today (1)714-998-6899 
*DXJH 3UHVVXUH [1]
P/N
Pressure Range
Media[2]
Accuracy(%FS)
Burst
Pressure(psi) (bar)
V15 -15 -1.0 G 0.02 (0.05) 3×
GP2 2 0.16 G 0.05 3×
GP5 5 0.35 G 0.05 3×
GP10 10 0.7 G 0.02 (0.05) 3×
GP15 15 1.0 G, L[3]
0.02 (0.05) 3×
GP30 30 2.0 G, L[3]
0.02 (0.05) 3×
GP50 50 3.5 G, L 0.02 (0.05) 3×
GP100 100 7.0 G, L 0.02 (0.05) 3×
GP300 300 20 G, L 0.02 (0.05) 3×
GP500 500 35 G, L 0.02 (0.05) 3×
GP600 600 40 G, L 0.02 (0.05) 3×
GP1K 1,000 70 G, L 0.02 (0.05) 3×
GP2K 2,000 140 G, L 0.02 (0.05) 3×
GP3K 3,000 200 G, L 0.02 (0.05) 3×
GP5K 5,000 350 G, L 0.02 (0.05) 3×
GP10K 10,000 700 G, L 0.02 (0.05) 3×
GP15K 15,000 1,000 G, L 0.05 (0.1) 2×
GP20K 20,000 1,400 G, L 0.05 (0.1) 1.5×
GP25K 25,000 1,600 G, L 0.1 1.5×
GP30K 30,000 2,000 G, L 0.1 1.5×
GP36K 36,000 2,500 G, L 0.1 1.5×
GP40K 40,000 2,800 G, L 0.1 1.1×
$EVROXWH 3UHVVXUH
P/N
Pressure Range
Media Accuracy(%FS)
Burst
Pressure(psi) (bar)
AP5 5 0.35 G 0.1 3×
AP10 10 0.7 G 0.1 3×
AP15 15 1.0 G 0.1 3×
AP30 30 2.0 G 0.1 3×
AP50 50 3.5 G 0.1 3×
AP100 100 7.0 G , L 0.05 (0.1) 3×
AP300 300 20 G , L 0.05 (0.1) 3×
AP500 500 35 G , L 0.05 (0.1) 3×
AP1K 1,000 70 G , L 0.05 (0.1) 3×
AP3K 3,000 200 G , L 0.05 (0.1) 3×
AP5K 5,000 350 G , L 0.05 (0.1) 3×
'LIIHUHQWLDO 3UHVVXUH
P/N
Pressure Range
Media
Accuracy
(%FS)[1]
Burst
Pressure
Static
Pressure
Range(inH2O) (mbar)
DP1 ±1 ±2.5 G 0.05[2]
100× ±10 psi
DP2 ±2 ±5.0 G 0.05[2]
100× ±10 psi
DP5 ±5 ±10 G 0.05[2]
50× ±10 psi
DP10 ±10 ±25 G 0.05[2]
20× ±10 psi
DP20 ±20 ±50 G 0.05 20× ±10 psi
DP30 ±30 ±75 G 0.05 20× ±10 psi
DP50 ±50 ±160 G 0.05 3× ±10 psi
DP150 ±150 ±350 G 0.02 3× 50 psi
DP300 ±300 ±700 G 0.02 3× 50 psi
(OHFWULFDO 0HDVXUHPHQW $FFXUDF
Range Accuracy
Voltage DC ±30.0000V ± (0.01%RD + 1.5 mV)
Current DC ±30.0000mA ± (0.01%RD + 1.5 μA)
DC 24V 24V±0.5V max:50mA Protect at: 120mA
Switch[1]
Status OPEN/CLOSED
RPSRXQG 3UHVVXUH
P/N
Pressure Range
Media Accuracy(%FS)[1] Burst
Pressure(psi) (bar)
CP2 ±2 ±0.16 G 0.05 3×
CP5 ±5 ±0.35 G 0.02 (0.05) 3×
CP10 ±10 ±0.7 G 0.02 (0.05) 3×
CP15 ±15 ±1 G 0.02 (0.05) 3×
CP30 -15 to 30 -1 to 2 G 0.02 (0.05) 3×
CP100 -15 to 100 -1 to 7 G, L 0.02 (0.05) 3×
CP300 -15 to 300 -1 to 20 G, L 0.02 (0.05) 3×
3UHVVXUH 5DQJHV
[1] Sealed gauge pressure for above 1000 psi
[2] G=Gas, L=Liquid
[3] 0.02% FS for gas media only
[1] 1V~12V if switch has detective voltage
63(,),$7,216
=? (5 URGEKſECVKQP CRRNKGU VQ VJG URCP QH VJG TCPIG #EEWTCE[ KPENWFGU
one year stability.
[2] 0.05%FS accuracy (incl 6 months stability). One year accuracy is
0.05%FS calibration accuracy combined with 0.05%FS one year stability.
=? (5 URGEKſECVKQP CRRNKGU VQ VJG URCP QH VJG TCPIG
Pressure/ProcessCalibrationEquipmentAdditelCatalog2016
Look us up on www.additel.com or call today (1)714-998-6899
63(,),$7,216
25'(5,1* ,1)250$7,21
3UHVVXUH RQQHFWLRQV
1/4NPT male, 1/2NPT male, 1/4BSP male, 1/2BSP
OCNG /: OCNG 
ŭ RUK
1/4HP female or 1/4HP male (15,000 psi)
1/4HP female: Autoclave F-250-C, 9/16 - 18 UNF-2B
1/4HP male: Autoclave M-250-C, 9/16 - 18 UNF-2A
0.236 inch (Ø6mm) test hose (for differential pressure)
Other connections available per request
(OHFWULFDO RQQHFWLRQ
0.156 inch (Ø4mm) sockets
2YHU 3UHVVXUH :DUQLQJ 120%
'DWD /RJJLQJ
5VQTCIG ECRCEKV[  ſNGU  TGEQTFU RGT ſNG
Mode: manual and automatic
Hourly-record: record the data every hour
Interval-record: set by user
0RGHO 1XPEHU
0RGHO
$FFXUDF
02-0.02% of full scale
05-0.05% of full scale
10-0.1% of full scale
5DQJH WSH
PSI-range by psi
BAR-range by bar
H2O-range by inH2O
3UHVVXUH SRUW WSH
N-1/4NPT male
N2-1/2NPT male
B-1/4BSP male
B2-1/2BSP male
M-M20X1.5 male
AF-Autoclave F-250-C female
AM-Autoclave M-250-C male
3UHVVXUH UDQJH 31
See pressure range table
3RZHU
Battery: Rechargeable Li-ion polymer battery
Li-Battery working time: 40 hours
Recharge time: 4 hours
External power: 110V/220V power adapter (DC10V)
(QFORVXUH
Case material: Aluminum alloy
Wetted parts: 316L SS
Dimension: Ø120mm X 46mm depth X 184mm height
Weight: 0.7kg
(XURSHDQ RPSOLDQFH CE marked
RPPXQLFDWLRQ
RS232 (DB9/F, environmentally sealed)
Baud rate: 1200,2400,4800,9600
Data length: 8 bits
Stop bit: 2 bits
Address: from 1 to 112
:DUUDQW1 year
$FFHVVRULHV LQFOXGHG
110V/220V external power adapter (DC 10V)
2 pieces 1.5-meter test leads
2 pieces 0.236 inch (Ø6 mm) test hose
(for differential pressure gauge only)
Additel/Land software
(free download at www.additel.com)
Manual
0+56 VTCEGCDNG ECNKDTCVKQP EGTVKſECVG
2SWLRQDO $FFHVVRULHV
0RGHO
QXPEHU
'HVFULSWLRQ
9702 Spare rechargeable Li-ion polymer battery for 672
9816
Spare 110V/220V external power adapter (DC 10V)
for ADT22X and ADT672 calibrator
9502
Additel/Log II real time data logging and graphical
software for 681 and 672
9506 Additel/PCal pressure calibration software
9050 USB to RS232 (DB9/M) Adapter
9050-EXT RS 232 (DB9/M) extension cable, 9 feet
9900-672 Carrying Case for one 672 digital pressure gauge
9022 Spare 2 pieces test leads (1.5-meter) for 672
$'7 *3 36, 1
'LJLWDO 3UHVVXUH DOLEUDWRUV
$GGLWHO
Look us up on www.additel.com or call today (1)714-998-6899 
3HUFHQWDJH RI 5HDGLQJ $FFXUDF
$SSOLFDWLRQ 1RWH
8QGHUVWDQGLQJ $FFXUDF 6SHFLILFDWLRQV IRU 'LJLWDO 3UHVVXUH 6HQVRUV ă 3HUFHQWDJH RI )XOO 6FDOH 9HUVXV
3HUFHQWDJH RI 5HDGLQJ
5RGEKſECVKQPU HQT FKIKVCN RTGUUWTG ICWIGU ECP UQOGVKOGU UGGO EQPHWUKPI QT QXGTYJGNOKPI GURGEKCNN[ KH [QW CTG
unfamiliar with the terminology. Some pressure sensors will specify accuracy as a percent of full scale (FS) while
QVJGTU RTQXKFG VJG URGEKſECVKQP CU C RGTEGPV QH TGCFKPI 5Q YJ[ CTG VJGTG FKHHGTGPV YC[U QH URGEKH[KPI VJG CEEWTCE[ QH
pressure sensors and is percent of reading more accurate than percent of full scale or vise versa? This brief technical
note will discuss the two differences and answer these questions.
Accuracy as a percentage of reading is accomplished by multiplying
the accuracy percentage by the pressure reading. Thus, the lower
the pressure measurement, the better the accuracy. Instruments that
have a percent reading specification are accompanied with a floor
URGEKſECVKQP 6JG ƀQQT URGEKſECVKQP VCMGU KPVQ CEEQWPV WPEGTVCKPVKGU
such as resolution and measurement noise which may be negligible
at higher pressures but are of much more significance at lower
pressures.
For example, an accuracy specification may read 0.1% of reading
for 20 to 100% of range and 0.02% of full scale below 20% of the
TCPIG 6JG  QH HWNN UECNG URGEKſECVKQP KU EQPUKFGTGF VJG ƀQQT
URGEKſECVKQP 6Q WPFGTUVCPF VJG CEEWTCE[ QH VJG UGPUQT VJG WUGT KU
VJGP TGSWKTGF VQ MPQY YJGTG VJG ƀQQT URGE KU CRRNKECDNG CPF VJG HWNN
scale of the sensor.
6JKU OGVJQF QH URGEKſECVKQP KU QHVGP WUGF DGECWUG KV CNKIPU YGNN YKVJ
the typical performance of pressure gauges. Typically, the closer you
measure to barometric pressure the better the performance of the
ICWIG (KIWTGU  CPF VJG ITCRJ DGNQY UJQY CP GZCORNG URGEKſECVKQP
for a 100 psi gauge and its accuracy in psi.
)LJXUH   3HUFHQW UHDGLQJ DFFXUDF H[DPSOH
)XOO VFDOH 0 to 100 psi
$FFXUDF 20 to 100% FS: 0.1% of reading
0 to 20% FS: 0.02% of FS
SVL
$FFXUDF
SVL
0 0.02
0.02%FS10 0.02
20 0.02
0.1% of
Reading
30 0.03
40 0.04
50 0.05
60 0.06
70 0.07
80 0.08
90 0.09
100 0.10
Pressure/ProcessCalibrationEquipmentAdditelCatalog2016
Look us up on www.additel.com or call today (1)714-998-6899
3HUFHQWDJH RI )XOO 6FDOH $FFXUDF
$ RPSDULVRQ RI 3HUFHQW RI )XOO 6FDOH DQG 3HUFHQW RI 5HDGLQJ $FFXUDFLHV
Accuracy as a percentage of full scale is calculated by multiplying the
accuracy percentage by the full scale pressure of the gauge. This is
QDXKQWUN[ C OQTG UKORNG OGVJQF QH URGEKſECVKQP CPF KU OQUV EQOOQPN[
used in industry because it is easy to calculate and interpret. Denoting the
accuracy as percent full scale is a more conservative way of specifying
the pressure sensor because typically the sensor doesn't perform the
same over its full range. It usually will perform more accurately as you
CRRTQCEJ DCTQOGVTKE RTGUUWTG 6JKU V[RG QH URGEKſECVKQP KU OQUV EQOOQP
for industrial gauges which make it easier to compare one gauge versus
CPQVJGT (KIWTG  KU CP GZCORNG URGEKſECVKQP HQT C  RUK ICWIG CPF KVU
accuracy in psi.
SVL
$FFXUDF
SVL
0 0.05
0.05%FS
10 0.05
20 0.05
30 0.05
40 0.05
50 0.05
60 0.05
70 0.05
80 0.05
90 0.05
100 0.05
SVL
$FFXUDF
SVL
RI
5HDGLQJ
 RI )6
0 0.02 0.05
10 0.02 0.05
20 0.02 0.05
30 0.03 0.05
40 0.04 0.05
50 0.05 0.05
60 0.06 0.05
70 0.07 0.05
80 0.08 0.05
90 0.09 0.05
100 0.10 0.05
So you may ask, Which is more accurate? The answer is that it depends on the
RTGUUWTG DGKPI OGCUWTGF +P VJG VYQ GZCORNGU IKXGP VJG ICWIG URGEKſGF CV 
of reading is more accurate as you measure lower pressures in its range. However,
CU [QW OQXG CDQXG  QH VJG TCPIG VJG ICWIG URGEKſGF CV  QH HWNN UECNG
becomes more accurate than the 0.1% of reading gauge. This can be seen clearly
in the chart (left) and graph (below) where the two gauges are compared in terms
of psi accuracy. To properly compare these, two gauges you should convert the
accuracy to pressure units, such as psi or bar. Then they can be properly matched
one against another in like units of measure.
+P EQPENWUKQP QPG OGVJQF QH URGEKſECVKQP KU PQV DGVVGT VJCP CPQVJGT KV KU LWUV
different. Given this difference it becomes important to know how to interpret the
FKHHGTGPV URGEKſECVKQPU V[RGU CPF DG CDNG VQ EQORCTG QPG XGTUWU CPQVJGT
Look us up on www.additel.com or call today (1)714-998-6899 
29(59,(:
*DXJH SUHVVXUH 'LIIHUHQWLDO SUHVVXUH
With advanced microprocessor technology and state-of-the-art silicon pressure sensors, the 681 series digital
pressure gauges provide an accurate, reliable, and economic solution for a wide range of pressure applications.
They are loaded with functionality and remarkably easy to use. To reach the best performance, every silicon
pressure sensor in our gauges is specially aged, tested and screened before assembly. At Additel, fully temperature
compensated accuracy means every sensor is pressure tested at several environmental temperatures from
u% VQ u% 9KVJ VJKU VGUV FCVC KPFKXKFWCN EQGHſEKGPVU CTG IGPGTCVGF CPF UVQTGF KP VJG ICWIG EJCTCEVGTKKPI KVU
performance over the full temperature compensated range. And now the ADT681IS comes with an IP67 rating
meaning it is dust resistant and water proof, submersible in 1 meter of water.
'HVLJQHG WR ILW RXU QHHG
Additel pressure gauges give you the widest variety of sensor choices on the market. Whether you require low
inches of water measurement or very high pressure measurement, we have a gauge that will meet your need. We
offer sensors which are ±1 inH2O (±2.5 mbar) to 40K psi (2,800 bar) and everything in between.
Do your applications require you to measure both positive and vacuum pressure? Our compound gauges do not
EQORTQOKUG CEEWTCE[ CPF RTQXKFGU [QW YKVJ VJG UCOG JKIJ CEEWTCE[ URGEKſECVKQP QP DQVJ RQUKVKXG CPF XCEWWO
pressures. We offer a wide variety up to 300 psi (20 bar). If you need a higher range, just contact us and we can
likely customize one to meet your need. We also offer absolute pressure sensors to 5K psi (350 bar) and a full
range of differential pressure sensors from ±1 inH2O (±2.5 mbar) to ±300 inH2O (±700 mbar). Are you looking for a
RTGUUWTG ICWIG VQ WUG KP JCCTFQWU CTGCU! 1WT #6':EGTVKſGF KPVTKPUKECNN[ UCHG OQFGNU 
+5 CTG FGUKIPGF HQT
pressure measurement in hazardous areas.
If you need to panel mount our sensors, we offer the option (see ordering information) for a back-mounted pressure
RQTV CPF ICWIG JQWUKPI FGUKIPGF VQ ſV KP C RCPGN #PF OQUV TGEGPVN[ YG XG CFFGF VJG QRVKQP VQ FQ UVCPFCNQPG FCVC
logging with the 681. Now you can record more than 21,000 records internal to the 681 series. Each record includes
date, time, pressure and temperature readings. Download the logged data with our free Additel/Land software or
you can purchase our Additel/Log II for real-time logging and data analysis. The 681 series digital pressure gauges
are unmatched in performance and reliability. Best of all, they are very affordable.
3UHVVXUH UDQJHV WR  SVL  EDU

More Related Content

Viewers also liked

INTRODUCTION of CENTRAL COALFIELD LIMITED
INTRODUCTION of CENTRAL COALFIELD LIMITEDINTRODUCTION of CENTRAL COALFIELD LIMITED
INTRODUCTION of CENTRAL COALFIELD LIMITEDsneh samrani
 
Ans 15203468 source_ppt
Ans 15203468 source_pptAns 15203468 source_ppt
Ans 15203468 source_pptkamancc
 
Mba Tomtom merger TeleAtlas
Mba Tomtom merger TeleAtlasMba Tomtom merger TeleAtlas
Mba Tomtom merger TeleAtlasRoel_Kock
 
Trinity Daily Feb 16, 2017
Trinity Daily Feb 16, 2017Trinity Daily Feb 16, 2017
Trinity Daily Feb 16, 2017Arun Surendran
 
Переход на новый порядок использования ККТ: основные вопросы
Переход на новый порядок использования ККТ: основные вопросы Переход на новый порядок использования ККТ: основные вопросы
Переход на новый порядок использования ККТ: основные вопросы MoySklad
 
CV. S. Jones Latest
CV. S. Jones LatestCV. S. Jones Latest
CV. S. Jones LatestSimon Jones
 
POOJA BHEDA REUME 1
POOJA BHEDA REUME 1POOJA BHEDA REUME 1
POOJA BHEDA REUME 1POOJA BHEDA
 

Viewers also liked (8)

INTRODUCTION of CENTRAL COALFIELD LIMITED
INTRODUCTION of CENTRAL COALFIELD LIMITEDINTRODUCTION of CENTRAL COALFIELD LIMITED
INTRODUCTION of CENTRAL COALFIELD LIMITED
 
Ans 15203468 source_ppt
Ans 15203468 source_pptAns 15203468 source_ppt
Ans 15203468 source_ppt
 
Mba Tomtom merger TeleAtlas
Mba Tomtom merger TeleAtlasMba Tomtom merger TeleAtlas
Mba Tomtom merger TeleAtlas
 
Trinity Daily Feb 16, 2017
Trinity Daily Feb 16, 2017Trinity Daily Feb 16, 2017
Trinity Daily Feb 16, 2017
 
Переход на новый порядок использования ККТ: основные вопросы
Переход на новый порядок использования ККТ: основные вопросы Переход на новый порядок использования ККТ: основные вопросы
Переход на новый порядок использования ККТ: основные вопросы
 
CV. S. Jones Latest
CV. S. Jones LatestCV. S. Jones Latest
CV. S. Jones Latest
 
POOJA BHEDA REUME 1
POOJA BHEDA REUME 1POOJA BHEDA REUME 1
POOJA BHEDA REUME 1
 
SS CMMS Training
SS CMMS TrainingSS CMMS Training
SS CMMS Training
 

Similar to Additel Catalog 2017 | Additel Katalog 2017

Katalog Additel - Distributor Additel di Indonesia - Siwali 021 45850618
Katalog Additel - Distributor Additel di Indonesia - Siwali 021 45850618Katalog Additel - Distributor Additel di Indonesia - Siwali 021 45850618
Katalog Additel - Distributor Additel di Indonesia - Siwali 021 45850618PT. Siwali Swantika
 
Fluke Catalog | Pressure Calibration Application and Solutions
Fluke Catalog | Pressure Calibration Application and SolutionsFluke Catalog | Pressure Calibration Application and Solutions
Fluke Catalog | Pressure Calibration Application and SolutionsPT. Siwali Swantika
 
Cashco - Reguladores Industriais
Cashco - Reguladores IndustriaisCashco - Reguladores Industriais
Cashco - Reguladores IndustriaisRenan Sehn Santos
 
Data Supplement for Kunkle Safety and Relief Products
Data Supplement for Kunkle Safety and Relief ProductsData Supplement for Kunkle Safety and Relief Products
Data Supplement for Kunkle Safety and Relief ProductsMead O'Brien, Inc.
 
Case cx220 c hydraulic excavator service repair manual
Case cx220 c hydraulic excavator service repair manualCase cx220 c hydraulic excavator service repair manual
Case cx220 c hydraulic excavator service repair manualfjjskekksemmm
 
TL9000 and the Process Approach
TL9000 and the Process ApproachTL9000 and the Process Approach
TL9000 and the Process ApproachJeff Bostow
 
Mounting and maintenance
Mounting and maintenanceMounting and maintenance
Mounting and maintenancemkpq pasha
 
Mounting and maintenance
Mounting and maintenanceMounting and maintenance
Mounting and maintenancemkpq pasha
 
Catálogo Williams _ V series
Catálogo Williams _ V series Catálogo Williams _ V series
Catálogo Williams _ V series Vibropac
 
Catálogo Williams _ X series
Catálogo Williams _ X series Catálogo Williams _ X series
Catálogo Williams _ X series Vibropac
 
Auto Tools Fyler – CarQuest – PetesPaint.ca – Q3 2018
Auto Tools Fyler – CarQuest – PetesPaint.ca – Q3 2018Auto Tools Fyler – CarQuest – PetesPaint.ca – Q3 2018
Auto Tools Fyler – CarQuest – PetesPaint.ca – Q3 2018Petes Paint & Auto Parts
 
Auto Tools Fyler – CarQuest – DTAuto.ca – Q2 2018
Auto Tools Fyler – CarQuest – DTAuto.ca – Q2 2018Auto Tools Fyler – CarQuest – DTAuto.ca – Q2 2018
Auto Tools Fyler – CarQuest – DTAuto.ca – Q2 2018DT AUTO PARTS Leamington
 
Meeting Q1 2023 PKU.pptx
Meeting Q1 2023 PKU.pptxMeeting Q1 2023 PKU.pptx
Meeting Q1 2023 PKU.pptxPandauCctv
 
Auto Tools Fyler – CarQuest – DTAuto.ca – Q2 2018
Auto Tools Fyler – CarQuest – DTAuto.ca – Q2 2018Auto Tools Fyler – CarQuest – DTAuto.ca – Q2 2018
Auto Tools Fyler – CarQuest – DTAuto.ca – Q2 2018DT AUTO PARTS Leamington
 

Similar to Additel Catalog 2017 | Additel Katalog 2017 (20)

Katalog Additel - Distributor Additel di Indonesia - Siwali 021 45850618
Katalog Additel - Distributor Additel di Indonesia - Siwali 021 45850618Katalog Additel - Distributor Additel di Indonesia - Siwali 021 45850618
Katalog Additel - Distributor Additel di Indonesia - Siwali 021 45850618
 
Fluke Catalog | Pressure Calibration Application and Solutions
Fluke Catalog | Pressure Calibration Application and SolutionsFluke Catalog | Pressure Calibration Application and Solutions
Fluke Catalog | Pressure Calibration Application and Solutions
 
Katalog Additel 2018
Katalog Additel 2018Katalog Additel 2018
Katalog Additel 2018
 
Cashco - Reguladores Industriais
Cashco - Reguladores IndustriaisCashco - Reguladores Industriais
Cashco - Reguladores Industriais
 
Calibracion de presion
Calibracion de presionCalibracion de presion
Calibracion de presion
 
Data Supplement for Kunkle Safety and Relief Products
Data Supplement for Kunkle Safety and Relief ProductsData Supplement for Kunkle Safety and Relief Products
Data Supplement for Kunkle Safety and Relief Products
 
하이플럭스 영문 카달로그 2014 (English)
하이플럭스 영문 카달로그 2014 (English)하이플럭스 영문 카달로그 2014 (English)
하이플럭스 영문 카달로그 2014 (English)
 
41TE.pdf
41TE.pdf41TE.pdf
41TE.pdf
 
Case cx220 c hydraulic excavator service repair manual
Case cx220 c hydraulic excavator service repair manualCase cx220 c hydraulic excavator service repair manual
Case cx220 c hydraulic excavator service repair manual
 
Jet cleaning machine-PressureJet
Jet cleaning machine-PressureJetJet cleaning machine-PressureJet
Jet cleaning machine-PressureJet
 
TL9000 and the Process Approach
TL9000 and the Process ApproachTL9000 and the Process Approach
TL9000 and the Process Approach
 
Mounting and maintenance
Mounting and maintenanceMounting and maintenance
Mounting and maintenance
 
Mounting and maintenance
Mounting and maintenanceMounting and maintenance
Mounting and maintenance
 
Catálogo Williams _ V series
Catálogo Williams _ V series Catálogo Williams _ V series
Catálogo Williams _ V series
 
Catálogo Williams _ X series
Catálogo Williams _ X series Catálogo Williams _ X series
Catálogo Williams _ X series
 
Auto Tools Fyler – CarQuest – PetesPaint.ca – Q3 2018
Auto Tools Fyler – CarQuest – PetesPaint.ca – Q3 2018Auto Tools Fyler – CarQuest – PetesPaint.ca – Q3 2018
Auto Tools Fyler – CarQuest – PetesPaint.ca – Q3 2018
 
Auto Tools Fyler – CarQuest – AutoBarn.ca
Auto Tools Fyler – CarQuest – AutoBarn.ca Auto Tools Fyler – CarQuest – AutoBarn.ca
Auto Tools Fyler – CarQuest – AutoBarn.ca
 
Auto Tools Fyler – CarQuest – DTAuto.ca – Q2 2018
Auto Tools Fyler – CarQuest – DTAuto.ca – Q2 2018Auto Tools Fyler – CarQuest – DTAuto.ca – Q2 2018
Auto Tools Fyler – CarQuest – DTAuto.ca – Q2 2018
 
Meeting Q1 2023 PKU.pptx
Meeting Q1 2023 PKU.pptxMeeting Q1 2023 PKU.pptx
Meeting Q1 2023 PKU.pptx
 
Auto Tools Fyler – CarQuest – DTAuto.ca – Q2 2018
Auto Tools Fyler – CarQuest – DTAuto.ca – Q2 2018Auto Tools Fyler – CarQuest – DTAuto.ca – Q2 2018
Auto Tools Fyler – CarQuest – DTAuto.ca – Q2 2018
 

More from PT. Siwali Swantika

Fluke 418x Infrared Calibrator.pdf
Fluke 418x Infrared Calibrator.pdfFluke 418x Infrared Calibrator.pdf
Fluke 418x Infrared Calibrator.pdfPT. Siwali Swantika
 
Kalibrasi dengan metode komparasi
Kalibrasi dengan metode komparasiKalibrasi dengan metode komparasi
Kalibrasi dengan metode komparasiPT. Siwali Swantika
 
10 Langkah Kalibrasi Termometer.pdf
10 Langkah Kalibrasi Termometer.pdf10 Langkah Kalibrasi Termometer.pdf
10 Langkah Kalibrasi Termometer.pdfPT. Siwali Swantika
 
Pengujian Sistem Photovoltaic.pdf
Pengujian Sistem Photovoltaic.pdfPengujian Sistem Photovoltaic.pdf
Pengujian Sistem Photovoltaic.pdfPT. Siwali Swantika
 
Fluke Process Calibration Tools Catalog
Fluke Process Calibration Tools CatalogFluke Process Calibration Tools Catalog
Fluke Process Calibration Tools CatalogPT. Siwali Swantika
 
Fluke Temperature Calibration Catalog
Fluke Temperature Calibration CatalogFluke Temperature Calibration Catalog
Fluke Temperature Calibration CatalogPT. Siwali Swantika
 

More from PT. Siwali Swantika (20)

NetAlly LinkRunner 10G
NetAlly LinkRunner 10GNetAlly LinkRunner 10G
NetAlly LinkRunner 10G
 
NetAlly Etherscope nXG
NetAlly Etherscope nXGNetAlly Etherscope nXG
NetAlly Etherscope nXG
 
Fluke 418x Infrared Calibrator.pdf
Fluke 418x Infrared Calibrator.pdfFluke 418x Infrared Calibrator.pdf
Fluke 418x Infrared Calibrator.pdf
 
Kalibrasi dengan metode komparasi
Kalibrasi dengan metode komparasiKalibrasi dengan metode komparasi
Kalibrasi dengan metode komparasi
 
Kalibrasi Transmitter
Kalibrasi TransmitterKalibrasi Transmitter
Kalibrasi Transmitter
 
Kategori Pengukuran Kelistrikan
Kategori Pengukuran KelistrikanKategori Pengukuran Kelistrikan
Kategori Pengukuran Kelistrikan
 
5 Langkah Kalibrasi RTD
5 Langkah Kalibrasi RTD5 Langkah Kalibrasi RTD
5 Langkah Kalibrasi RTD
 
Keuntungan AnyWare Cloud.pdf
Keuntungan AnyWare Cloud.pdfKeuntungan AnyWare Cloud.pdf
Keuntungan AnyWare Cloud.pdf
 
3 Metode Kalibrasi Sensor Suhu
3 Metode Kalibrasi Sensor Suhu3 Metode Kalibrasi Sensor Suhu
3 Metode Kalibrasi Sensor Suhu
 
10 Langkah Kalibrasi Termometer.pdf
10 Langkah Kalibrasi Termometer.pdf10 Langkah Kalibrasi Termometer.pdf
10 Langkah Kalibrasi Termometer.pdf
 
Pengujian Sistem Photovoltaic.pdf
Pengujian Sistem Photovoltaic.pdfPengujian Sistem Photovoltaic.pdf
Pengujian Sistem Photovoltaic.pdf
 
Fluke Process Calibration Tools Catalog
Fluke Process Calibration Tools CatalogFluke Process Calibration Tools Catalog
Fluke Process Calibration Tools Catalog
 
Fluke Temperature Calibration Catalog
Fluke Temperature Calibration CatalogFluke Temperature Calibration Catalog
Fluke Temperature Calibration Catalog
 
Vitrek Catalog
Vitrek CatalogVitrek Catalog
Vitrek Catalog
 
Rotronic Water Activity Series
Rotronic Water Activity SeriesRotronic Water Activity Series
Rotronic Water Activity Series
 
Rotronic Catalog 2022
Rotronic Catalog 2022Rotronic Catalog 2022
Rotronic Catalog 2022
 
Rotronic RMS Catalog
Rotronic RMS CatalogRotronic RMS Catalog
Rotronic RMS Catalog
 
Fluke Catalog 2022
Fluke Catalog 2022Fluke Catalog 2022
Fluke Catalog 2022
 
Metrel Catalog 2022
Metrel Catalog 2022Metrel Catalog 2022
Metrel Catalog 2022
 
NetAlly Product Family Brochure
NetAlly Product Family BrochureNetAlly Product Family Brochure
NetAlly Product Family Brochure
 

Recently uploaded

IVE Industry Focused Event - Defence Sector 2024
IVE Industry Focused Event - Defence Sector 2024IVE Industry Focused Event - Defence Sector 2024
IVE Industry Focused Event - Defence Sector 2024Mark Billinghurst
 
Software and Systems Engineering Standards: Verification and Validation of Sy...
Software and Systems Engineering Standards: Verification and Validation of Sy...Software and Systems Engineering Standards: Verification and Validation of Sy...
Software and Systems Engineering Standards: Verification and Validation of Sy...VICTOR MAESTRE RAMIREZ
 
Class 1 | NFPA 72 | Overview Fire Alarm System
Class 1 | NFPA 72 | Overview Fire Alarm SystemClass 1 | NFPA 72 | Overview Fire Alarm System
Class 1 | NFPA 72 | Overview Fire Alarm Systemirfanmechengr
 
Gurgaon ✡️9711147426✨Call In girls Gurgaon Sector 51 escort service
Gurgaon ✡️9711147426✨Call In girls Gurgaon Sector 51 escort serviceGurgaon ✡️9711147426✨Call In girls Gurgaon Sector 51 escort service
Gurgaon ✡️9711147426✨Call In girls Gurgaon Sector 51 escort servicejennyeacort
 
Past, Present and Future of Generative AI
Past, Present and Future of Generative AIPast, Present and Future of Generative AI
Past, Present and Future of Generative AIabhishek36461
 
Work Experience-Dalton Park.pptxfvvvvvvv
Work Experience-Dalton Park.pptxfvvvvvvvWork Experience-Dalton Park.pptxfvvvvvvv
Work Experience-Dalton Park.pptxfvvvvvvvLewisJB
 
Arduino_CSE ece ppt for working and principal of arduino.ppt
Arduino_CSE ece ppt for working and principal of arduino.pptArduino_CSE ece ppt for working and principal of arduino.ppt
Arduino_CSE ece ppt for working and principal of arduino.pptSAURABHKUMAR892774
 
TechTAC® CFD Report Summary: A Comparison of Two Types of Tubing Anchor Catchers
TechTAC® CFD Report Summary: A Comparison of Two Types of Tubing Anchor CatchersTechTAC® CFD Report Summary: A Comparison of Two Types of Tubing Anchor Catchers
TechTAC® CFD Report Summary: A Comparison of Two Types of Tubing Anchor Catcherssdickerson1
 
CCS355 Neural Networks & Deep Learning Unit 1 PDF notes with Question bank .pdf
CCS355 Neural Networks & Deep Learning Unit 1 PDF notes with Question bank .pdfCCS355 Neural Networks & Deep Learning Unit 1 PDF notes with Question bank .pdf
CCS355 Neural Networks & Deep Learning Unit 1 PDF notes with Question bank .pdfAsst.prof M.Gokilavani
 
Concrete Mix Design - IS 10262-2019 - .pptx
Concrete Mix Design - IS 10262-2019 - .pptxConcrete Mix Design - IS 10262-2019 - .pptx
Concrete Mix Design - IS 10262-2019 - .pptxKartikeyaDwivedi3
 
Risk Assessment For Installation of Drainage Pipes.pdf
Risk Assessment For Installation of Drainage Pipes.pdfRisk Assessment For Installation of Drainage Pipes.pdf
Risk Assessment For Installation of Drainage Pipes.pdfROCENODodongVILLACER
 
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube Exchanger
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube ExchangerStudy on Air-Water & Water-Water Heat Exchange in a Finned Tube Exchanger
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube ExchangerAnamika Sarkar
 
Unit7-DC_Motors nkkjnsdkfnfcdfknfdgfggfg
Unit7-DC_Motors nkkjnsdkfnfcdfknfdgfggfgUnit7-DC_Motors nkkjnsdkfnfcdfknfdgfggfg
Unit7-DC_Motors nkkjnsdkfnfcdfknfdgfggfgsaravananr517913
 
US Department of Education FAFSA Week of Action
US Department of Education FAFSA Week of ActionUS Department of Education FAFSA Week of Action
US Department of Education FAFSA Week of ActionMebane Rash
 
computer application and construction management
computer application and construction managementcomputer application and construction management
computer application and construction managementMariconPadriquez1
 
Architect Hassan Khalil Portfolio for 2024
Architect Hassan Khalil Portfolio for 2024Architect Hassan Khalil Portfolio for 2024
Architect Hassan Khalil Portfolio for 2024hassan khalil
 
Call Us ≽ 8377877756 ≼ Call Girls In Shastri Nagar (Delhi)
Call Us ≽ 8377877756 ≼ Call Girls In Shastri Nagar (Delhi)Call Us ≽ 8377877756 ≼ Call Girls In Shastri Nagar (Delhi)
Call Us ≽ 8377877756 ≼ Call Girls In Shastri Nagar (Delhi)dollysharma2066
 

Recently uploaded (20)

IVE Industry Focused Event - Defence Sector 2024
IVE Industry Focused Event - Defence Sector 2024IVE Industry Focused Event - Defence Sector 2024
IVE Industry Focused Event - Defence Sector 2024
 
Software and Systems Engineering Standards: Verification and Validation of Sy...
Software and Systems Engineering Standards: Verification and Validation of Sy...Software and Systems Engineering Standards: Verification and Validation of Sy...
Software and Systems Engineering Standards: Verification and Validation of Sy...
 
Class 1 | NFPA 72 | Overview Fire Alarm System
Class 1 | NFPA 72 | Overview Fire Alarm SystemClass 1 | NFPA 72 | Overview Fire Alarm System
Class 1 | NFPA 72 | Overview Fire Alarm System
 
Gurgaon ✡️9711147426✨Call In girls Gurgaon Sector 51 escort service
Gurgaon ✡️9711147426✨Call In girls Gurgaon Sector 51 escort serviceGurgaon ✡️9711147426✨Call In girls Gurgaon Sector 51 escort service
Gurgaon ✡️9711147426✨Call In girls Gurgaon Sector 51 escort service
 
Past, Present and Future of Generative AI
Past, Present and Future of Generative AIPast, Present and Future of Generative AI
Past, Present and Future of Generative AI
 
Work Experience-Dalton Park.pptxfvvvvvvv
Work Experience-Dalton Park.pptxfvvvvvvvWork Experience-Dalton Park.pptxfvvvvvvv
Work Experience-Dalton Park.pptxfvvvvvvv
 
Arduino_CSE ece ppt for working and principal of arduino.ppt
Arduino_CSE ece ppt for working and principal of arduino.pptArduino_CSE ece ppt for working and principal of arduino.ppt
Arduino_CSE ece ppt for working and principal of arduino.ppt
 
TechTAC® CFD Report Summary: A Comparison of Two Types of Tubing Anchor Catchers
TechTAC® CFD Report Summary: A Comparison of Two Types of Tubing Anchor CatchersTechTAC® CFD Report Summary: A Comparison of Two Types of Tubing Anchor Catchers
TechTAC® CFD Report Summary: A Comparison of Two Types of Tubing Anchor Catchers
 
CCS355 Neural Networks & Deep Learning Unit 1 PDF notes with Question bank .pdf
CCS355 Neural Networks & Deep Learning Unit 1 PDF notes with Question bank .pdfCCS355 Neural Networks & Deep Learning Unit 1 PDF notes with Question bank .pdf
CCS355 Neural Networks & Deep Learning Unit 1 PDF notes with Question bank .pdf
 
Concrete Mix Design - IS 10262-2019 - .pptx
Concrete Mix Design - IS 10262-2019 - .pptxConcrete Mix Design - IS 10262-2019 - .pptx
Concrete Mix Design - IS 10262-2019 - .pptx
 
Exploring_Network_Security_with_JA3_by_Rakesh Seal.pptx
Exploring_Network_Security_with_JA3_by_Rakesh Seal.pptxExploring_Network_Security_with_JA3_by_Rakesh Seal.pptx
Exploring_Network_Security_with_JA3_by_Rakesh Seal.pptx
 
young call girls in Rajiv Chowk🔝 9953056974 🔝 Delhi escort Service
young call girls in Rajiv Chowk🔝 9953056974 🔝 Delhi escort Serviceyoung call girls in Rajiv Chowk🔝 9953056974 🔝 Delhi escort Service
young call girls in Rajiv Chowk🔝 9953056974 🔝 Delhi escort Service
 
Risk Assessment For Installation of Drainage Pipes.pdf
Risk Assessment For Installation of Drainage Pipes.pdfRisk Assessment For Installation of Drainage Pipes.pdf
Risk Assessment For Installation of Drainage Pipes.pdf
 
Design and analysis of solar grass cutter.pdf
Design and analysis of solar grass cutter.pdfDesign and analysis of solar grass cutter.pdf
Design and analysis of solar grass cutter.pdf
 
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube Exchanger
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube ExchangerStudy on Air-Water & Water-Water Heat Exchange in a Finned Tube Exchanger
Study on Air-Water & Water-Water Heat Exchange in a Finned Tube Exchanger
 
Unit7-DC_Motors nkkjnsdkfnfcdfknfdgfggfg
Unit7-DC_Motors nkkjnsdkfnfcdfknfdgfggfgUnit7-DC_Motors nkkjnsdkfnfcdfknfdgfggfg
Unit7-DC_Motors nkkjnsdkfnfcdfknfdgfggfg
 
US Department of Education FAFSA Week of Action
US Department of Education FAFSA Week of ActionUS Department of Education FAFSA Week of Action
US Department of Education FAFSA Week of Action
 
computer application and construction management
computer application and construction managementcomputer application and construction management
computer application and construction management
 
Architect Hassan Khalil Portfolio for 2024
Architect Hassan Khalil Portfolio for 2024Architect Hassan Khalil Portfolio for 2024
Architect Hassan Khalil Portfolio for 2024
 
Call Us ≽ 8377877756 ≼ Call Girls In Shastri Nagar (Delhi)
Call Us ≽ 8377877756 ≼ Call Girls In Shastri Nagar (Delhi)Call Us ≽ 8377877756 ≼ Call Girls In Shastri Nagar (Delhi)
Call Us ≽ 8377877756 ≼ Call Girls In Shastri Nagar (Delhi)
 

Additel Catalog 2017 | Additel Katalog 2017

  • 1. Look us up on www.additel.com or call today (1)714-998-6899 3UHFLVLRQ XQGHU SUHVVXUH Pressure Calibration Equipment Catalog 2017
  • 2. Additel Corporation is one of the leading worldwide providers of pressure calibration tools. We are dedicated to designing, manufacturing, and delivering the highest quality test tools and portable calibrators for the process and calibration industries. For many years Additel has successfully developed pressure controlers, automated pressure calibrators, digital pressure test gauges, digital pressure calibrators, pressure test and calibration pumps, and multifunction process calibrators. Additel products are currently used in over 70 countries worldwide. Product quality and customer service along with innovative engineering have been our top priorities and will continue to be our guiding principles going forward. We are committed to customer satisfaction through quality products, competitive pricing, unmatched services/ technical support and continued introduction of new and innovative products. $'',7(/ 25325$7,21 RUSRUDWH +HDGTXDUWHUV Additel Corporation 2900 Saturn Street #B Brea, CA 92821, USA 6DOW /DNH LW 2IILFH Additel Corporation 1314 West State Rd Suite 101 Pleasant Grove, UT 84062, USA 3KRQH )D[ (PDLO VDOHV#DGGLWHOFRP (XURSHDQ 2IILFH Additel Corporation Holkebjergvej 79 5250 Odense, Denmark $VLD 2IILFH Additel Corporation Bldg 5, No. 3 Fengxiu Middle Road, Haidian, Beijing, 100094, China
  • 3. $XWRPDWLF 3UHVVXUH DOLEUDWRU Automatic Handheld Pressure Calibrator Selection Guide.......................01 Additel 760 Automatic Handheld Pressure Calibrator..............................02 Automated Pressure Calibrators Selection Guide....................................07 Additel 761 Automated Pressure Calibrators...........................................08 3UHVVXUH RQWUROOHUV Additel 780 Series Pressure Controller....................................................13 'LJLWDO 3UHVVXUH DOLEUDWRUV *DXJHV Additel Digital Pressure Gauges Selection Guide....................................20 Additel 672 Digital Pressure Calibrators.................................................. 22 Application notes......................................................................................25 Additel 681 Digital Pressure Gauges........................................................27 Additel 680 Digital Pressure Gauges........................................................30 Application notes......................................................................................32 0XOWLIXQFWLRQ 3URFHVV DOLEUDWRUV Multifunction Process Calibrators Selection Guide...................................34 Additel 209 210 Series Loop Calibrator....................................................35 Additel 220 Multifunction Loop Calibrator.................................................37 Additel 221A 222A 223A /WNVKſEVKQP QEWOGPVKPI 2TQEGUU %CNKDTCVQTU 'LJLWDO 3UHVVXUH 0RGXOHV Additel 160A Series Intelligent Digital Pressure Modules........................42 3UHVVXUH7HVWDOLEUDWLRQ 3XPSV Pressure Test/Calibration Pumps Selection Guide..................................45 Additel 901A Low Pressure Test Pump...................................................46 Additel 912 Low Pressure Test Pump......................................................47 Additel 914 Handheld Pneumatic Pressure Test Pump...........................48 7$%/( 2) 217(176
  • 4. Additel 916 Pneumatic Pressure Test Pump........................................... 49 Additel 917 Pneumatic Pressure Test Pump........................................... 50 Additel 918 Pneumatic Pressure Test Pump........................................... 51 Additel 919A Pneumatic High Pressure Test Pump................................ 52 Application notes..................................................................................... 53 Additel 920 Pneumatic High Pressure Test Pump.................................. 54 Additel 925 Handheld Hydraulic Pressure Test Pump............................ 55 Application notes..................................................................................... 56 Additel 927 Hydraulic Pressure Test Pump............................................. 57 Additel 928 Hydraulic Pressure Test Pump............................................. 58 Additel 936 Hydraulic High Pressure Calibration Pump...........................59 Additel 937 Hydraulic High Pressure Calibration Pump...........................60 Additel 938 Hydraulic High Pressure Calibration Pump...........................61 Additel 946 Hydraulic High Pressure Calibration Pump...........................62 Additel 949 Hydraulic Ultra-high Pressure Test Pump.............................63 Additel 959 Hydraulic Ultra-high Pressure Test Pump.............................64 3UHVVXUH 0DQLIROGV....................................................................................65 )LOWHUV.........................................................................................................66 +RVHV $GDSWHUV DQG )LWWLQJV....................................................................67 6RIWZDUH.....................................................................................................70 7$%/( 2) 217(176
  • 5. Look us up on www.additel.com or call today (1)714-998-6899 0RGHO )HDWXUHV //3 //3'/ ' ''/ 0$ 0$'/ 3UHVVXUH 5DQJH@ ±30 in H2O (±75 mbar) ±30 in H2O (±75 mbar) -12.5 to 35 psi (-0.86 to 2.5 bar) -12.5 to 35 psi (-0.86 to 2.5 bar) -12.5 to 300 psig (-0.86 to 20 bar) -12.5 to 300 psig (-0.86 to 20 bar) $FFXUDF )6
  • 7. 0.005[2] 0.005[2] 0.005[2] 0.005[2] 0.005[2] 0.005[2] *DXJH 3UHVVXUH Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ 'LIIHUHQWLDO 3UHVVXUH Ŗ Ŗ Ŗ Ŗ $EVROXWH 3UHVVXUH Ŗ Ŗ %DURPHWULF 3UHVVXUH Ŗ Ŗ 5HPRYDEOH ,QWHUQDO 0RGXOH Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ ([WHUQDO 3UHVVXUH 0RGXOH Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ 7HPSHUDWXUH RPSHQVDWHG Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ %XLOWLQ (OHFWULDO 3XPS Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ %XLOWLQ )LOWHU Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ %XLOWLQ /LTXLG7UDS Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ 6RXUFH6LPXODWH P$ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ 0HDVXUH P$ RU 9 Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ 9 /RRS SRZHU Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ 3UHVVXUH VZLWFK WHVW Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ +$57 RPPXQLFDWLRQ Ŗ Ŗ Ŗ 7DVN 'RFXPHQWDWLRQ Ŗ Ŗ Ŗ 'DWD /RJJLQJ Ŗ Ŗ Ŗ KDQQHOV 4 4 4 4 4 4 86%:L)L DQG %OXHWRRWK Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ $XWRPDWLF +DQGKHOG 3UHVVXUH DOLEUDWRU 6HOHFWLRQ *XLGH $GGLWHO @ )6 VSHFL¿FDWLRQ DSSOLHV WR WKH VSDQ RI WKH PRGXOH UDQJH @ 6WDELOLW EDVHG RQ )6 RI WKH LQWHUQDO SUHVVXUH PRGXOH ,QWHUQDO PRGXOH LV VZLWFKDEOH @ 6SHFL¿FDWLRQ EDVHG RQ JDXJH PHDVXUHPHQW $Q DGGLWLRQDO SD XQFHUWDLQW ZLOO QHHG WR EH LQFOXGHG ZKHQ PHDVXULQJ LQ DEVROXWH PRGH
  • 8. Pressure/ProcessCalibrationEquipmentAdditelCatalog2016 Look us up on www.additel.com or call today (1)714-998-6899 $XWRPDWLF +DQGKHOG 3UHVVXUH DOLEUDWRU $GGLWHO )XOO DXWRPDWLF FDOLEUDWRU ZLWK EXLOWLQ SXPS DQG FRQWUROOHU 6ZLWFKDEOH LQWHUQDO SUHVVXUH PRGXOHV IRU H[SDQGDEOH UDQJHV $FFXUDF HDU
  • 9. RI )6 ([WHUQDO SUHVVXUH PRGXOHV DYDLODEOH PHDVXUH RQO
  • 10. /HVV WKDQ OEV NJ
  • 11. IRU KDQGKHOG RSHUDWLRQ 6RXUFH SUHVVXUH PHDVXUH SUHVVXUH DQG HOHFWULFDO FKDQQHOV 2SWLRQDO +$57 FRPPXQLFDWLRQV 2SWLRQDO GDWD ORJJLQJ DQG WDVN GRFXPHQWLQJ 86%:L)L DQG %OXHWRRWK FRPPXQLFDWLRQV 29(59,(: A portable automated pressure calibrator in the palm of your hand—this could be our most exciting product yet! The Additel 760 series Automatic Handheld Pressure Calibrator takes portable pressure calibration to new levels. Weighing less than 4 lbs (1.8 kg), the ADT760’s innovative design contains a built-in pump, precision pressure sensor, internal controller and a large touch-screen color display. To generate pressure, simply key in the desired pressure and the Additel 760 will do the rest. Each unit has four channels: one internal pressure channel for source and measure pressures, two external pressure measurement channels, and one electronic measure and source channel. This series of calibrator has three standard models with the option of adding HART communications, documentation and data logging.
  • 12. Look us up on www.additel.com or call today (1)714-998-6899 )($785(6 ADT760-LLP The 760-LLP is designed for low pressure calibration and comes with a built-in pressure module to ±30 inH2O (±75 mbar) and provides an accuracy to 0.05%FS. If you need to go lower in pressure, you can purchase additional internal pressure modules (ADT155 series), which for the 760-LLP provide a variety of ranges down to ±0.25 inH2O (±0.62 mbar). The accuracy of 0.05%FS and control stability 0.005%FS is based on the internal module’s span. Measurements can be made in gauge or differential mode. ADT760-D The 760-D gives you differential and gauge pressure but at a higher pressure range than the ADT760-LLP. Covering the range of -12.5 to 35 psi (-0.86 to 2.5 bar) and with an accuracy of 0.02%FS, the ADT760-D is an ideal solution to cover very common gauge and differential pressure measurements. The Additel 760-D comes with an internal module covering its full range, but lower pressure modules down to ±10 inH2O differential can be purchased to improve accuracy at lower pressures. ADT760-MA The 760-MA generates and controls pressure from vacuum pressures up to 300 psig (20 bar.g) with an accuracy of 0.02%FS. Equipped with a built-in barometric reference, each unit can switch between gauge and absolute pressure types. A variety of internal sensors are available which offer lower pressure ranges for improved performance. Documenting Process Functionality Each model of the Additel 760 series has an option incorporating documentation and communication functions turning your 760 into a multifunction documenting process calibrator. This feature provides HART communication, task documentation and data logging. 6SHFLILFDWLRQ //3 ' 0$ 0D[ 3UHVVXUH 5DQJH ±30 inH2O (75 mbar) -12.5 to 35 psi (-0.86 to 2.5 bar) -12.5 to 300 psig (-0.86 to 20 bar.g) $FFXUDF 0.05%FS[1] 0.02%FS[1] 0.02%FS[1][3] 6WDELOLW 0.005%FS[2] 0.005%FS[2] 0.005%FS[2] 3UHVVXUH7SH Differential, Gauge Differential, Gauge Gauge, Absolute 2YHU 5DQJH ,QGLFDWLRQ 120% 5HVROXWLRQ 6 digits 0HDVXUHPHQW 8QLWV Pa, hPa, kPa, mPa, bar, mbar, psi, mmHg@0°C, cmHg@0°C, mHg@0°C, inHg@0°C, inH2O@4°C, mmH2O@4°C, cmH2O@4°C, mH2O@4°C, mmH2O@20°C, cmH2O@20°C, mH2O@20°C, inH2O@20°C, kg/m2, mtorr, torr, lb/ft2, tsi, custom %DURPHWULF $FFXUDF N/A N/A 60 Pa[4] RQQHFWLRQ $CTD ſVVKPI *QUG HV O YKVJ DWKNVKP ſNVGT VQ 1/4BSPF, 1/4NPTF, and M20F adapters *QUG HV O YKVJ DWKNVKP ſNVGT VQ 1/4BSPF, 1/4NPTF, and M20F adapters 3UHVVXUH 2XWSXW 5DWH 30 Seconds (30 inH2O/100 ml) 10 Seconds (35 psi/5 ml) 90 Seconds (300 psi/5 ml) 3UHVVXUH 6SHFLILFDWLRQV @ )6 VSHFL¿FDWLRQ DSSOLHV WR WKH VSDQ RI WKH PRGXOH UDQJH @ 6WDELOLW EDVHG RQ )6 RI WKH LQWHUQDO SUHVVXUH PRGXOH 6WDELOLW LV )6 RU SD ZKLFKHYHU LV JUHDWHU ,QWHUQDO PRGXOH LV VZLWFKDEOH @ 6SHFL¿FDWLRQ EDVHG RQ JDXJH PHDVXUHPHQW $Q DGGLWLRQDO SD XQFHUWDLQW ZLOO QHHG WR EH LQFOXGHG ZKHQ PHDVXULQJ LQ DEVROXWH PRGH @ 3D XQFHUWDLQW N
  • 13. LQFOXGHV FDOLEUDWLRQ XQFHUWDLQW OLQHDULW DQG ORQJWHUP VWDELOLW 3D SHU HDU
  • 14. %DURPHWHU UDQJH RI WR N3D 6SHFLILFDWLRQ 5DQJH 5HVROXWLRQ $FFXUDF 1RWH P$ 0HDVXUH ±30 mA 0.0001 mA 0.01%RD+0.005%FS +ORGFCPEG š 9 0HDVXUH ±30 V 0.0001 V 0.01%RD+0.005%FS +ORGFCPEG /š P$ 6RXUFH 24 mA 0.001 mA 0.01%RD+0.005%FS 20 mA @ 1K /RRS 3RZHU 6RXUFH 25 V N/A ±1 V 100 mA (Max Loading) 3UHVVXUH 6ZLWFK Open, close. Support for mechanical switches and NPN/PNP digital switches. 7HPSHUDWXUH RPSHQVDWLRQ 41°F to 95°F (5°C to 35°C) 7HPSHUDWXUH RHIILFLHQW ± ( 0.001%RD + 0.001%FS ) / °C outside of 5°C to 35°C (OHFWULFDO 6SHFLILFDWLRQV
  • 15. Pressure/ProcessCalibrationEquipmentAdditelCatalog2016 Look us up on www.additel.com or call today (1)714-998-6899 0RGXOH 0RGXOH 5DQJH 0HGLD $FFXUDF )6
  • 16. @ %XUVW 3UHVVXUH //3 ' 0$ LQ+2 PEDU '3[4] ±0.25 ±0.62 G 0.2[2] 100x Ŗ '3[4] ±0.5 ±1.25 G 0.1[2] 100x Ŗ '3[4] ±1 ±2.5 G 0.05[2] 100x Ŗ '3[4] ±2 ±5 G 0.05[2] 100x Ŗ '3 ±5 ±10 G 0.05[2] 50x Ŗ '3 ±10 ±25 G 0.05[2] 20x Ŗ Ŗ '3 ±20 ±50 G 0.05 20x Ŗ Ŗ '3 ±30 ±75 G 0.05 20x Ŗ Ŗ '3 ±50 ±160 G 0.05 3x Ŗ '3 ±150 ±350 G 0.02 3x Ŗ '3 ±300 ±700 G 0.02 3x Ŗ '3 -350 to 400 (-12.5 to 15 psi) -860 to 1K (-0.86 to 1 bar) G 0.02 3x Ŗ '3 -350 to 800 (-12.5 to 30 psi) -860 to 2K (-0.86 to 2 bar) G 0.02 3x Ŗ '3. -350 to 1K (-12.5 to 35 psi) -860 to 2.5K (-0.86 to 2.5 bar) G 0.02 3x Ŗ *DXJH 3UHVVXUH psi bar 3 ±10 ±0.7 G 0.02[3] 3x Ŗ Ŗ 3 ±15 ±1 G 0.02[3] 3x Ŗ Ŗ 3 -12.5 to 30 -0.86 to 2 G 0.02[3] 3x Ŗ Ŗ 3 -12.5 to 35 -0.86 to 2.5 G 0.02[3] 3x Ŗ Ŗ 3 -12.5 to 50 -0.86 to 3.5 G 0.02[3] 3x Ŗ 3 -12.5 to 100 -0.86 to 7 G 0.02[3] 3x Ŗ 3 -12.5 to 150 -0.86 to 10 G 0.02[3] 3x Ŗ 3 -12.5 to 200 -0.86 to 14 G 0.02[3] 3x Ŗ 3 -12.5 to 300 -0.86 to 20 G 0.02[3] 3x Ŗ ,QWHUQDO 0RGXOHV 6SHFLILFDWLRQV DQG RPSDWLELOLW =? (5 URGEKſECVKQP CRRNKGU VQ VJG URCP QH VJG OQFWNG TCPIG #EEWTCE[ KPENWFGU QPG[GCT UVCDKNKV[ GZEGRV HQT 2 VQ 2 OQFWNGU [2] 0.05%FS accuracy (incl 6-month stability). One-year accuracy is 0.05%FS calibration accuracy combined with 0.05%FS one-year stability. =? 5RGEKſECVKQP DCUGF QP ICWIG OGCUWTGOGPV #P CFFKVKQPCN RC WPEGTVCKPV[ YKNN PGGF VQ DG KPENWFGF YJGP OGCUWTKPI KP CDUQNWVG OQFG #RRNKECDNG QPN[ for use with the ADT760-MA [4] 0.05 Pa pressure control stability
  • 17. Look us up on www.additel.com or call today (1)714-998-6899 *HQHUDO 6SHFLILFDWLRQV 6SHFLILFDWLRQ 'HVFULSWLRQ KDQQHOV Four total: one electrical, one internal pressure, two external pressure (measure only) (QFORVXUH ,3 5DWLQJ IP54 water and dust proof %DWWHU Rechargeable Li-Ion battery, typically 10 hours of operation, recharges in less than 4 hours 'LVSOD Color 800 x 480 TFT 5-inch touch screen RPPXQLFDWLRQV USB, WiFi, Bluetooth :HLJKW 4 lbs (1.8 kg) 6L]H 9.3 x 4.3 x 2.8 in (235 x 110 x 70 mm) HUWLILFDWLRQ 0+56VTCEGCDNG EGTVKſECVG YKVJ FCVC KPENWFGF +$57 RPPXQLFDWLRQV Optional (ADT760-X-DL model) 'DWD /RJJLQJ Optional (ADT760-X-DL model), up to 1,000,000 readings (date and time stamped) 7DVN 'RFXPHQWDWLRQ Optional (ADT760-X-DL model) up to 250 tasks $XWRPDWLRQ )XQFWLRQV Switch test, auto step, leak test 0LVXVH 3URWHFWLRQ Up to 30 V on any two sockets 0XOWL /LQJXDO ,QWHUIDFH English, German, French, Italian, Spanish, Portuguese, Chinese, Japanese, and Russian 3XPS /LIH 500,000 cycles 3RZHU Rechargeable Li-Ion battery, external power: 110/220 V power adapter 10 V (QYLURQPHQW 6SHFLILFDWLRQV Operation: 32°F to 122°F (0°C to 50°C), 0-90% RH, less than 3,000 meters Compensated temperature: 32°F to 122°F (0°C to 50°C) Storage temperature: -4°F to 158°F (-20°C to 70°C) 9LEUDWLRQ DQG 6KRFN Vibration: 4g (20 to 2,000 Hz) Shock: 8g, 1 meter drop test RPSOLDQFH CE :DUUDQW 1 year 2SWLRQDO ,QWHUQDO 3UHVVXUH 0RGXOHV 0RGXOH 1XPEHU 'HVFULSWLRQ $'7'3 Pressure module for ADT760-LLP, ±0.25 inH2O, ±0.2%FS $'7'3 Pressure module for ADT760-LLP, ±0.5 inH2O, ±0.1%FS $'7'3 Pressure module for ADT760-LLP, ±1 inH2O, ±0.05%FS $'7'3 Pressure module for ADT760-LLP, ±2 inH2O, ±0.05%FS $'7'3 Pressure module for ADT760-LLP, ±5 inH2O, ±0.05%FS $'7'3 Pressure module for ADT760-LLP -D ±10 inH2O, ±0.05%FS $'7'3 Pressure module for ADT760-LLP -D, ±20 inH2O, ±0.05%FS $'7'3 Pressure module for ADT760-LLP -D, ±30 inH2O, ±0.05%FS $'7'3 Pressure module for ADT760-D, ±50 inH2O, ±0.05%FS $'7'3 Pressure module for ADT760-D, ±150 inH2O, ±0.02%FS $'7'3 Pressure module for ADT760-D, ±300 inH2O, ±0.02%FS $'7'3 Pressure module for ADT760-D, -350 to 400 inH2O (-12.5 to 15 psi), ±0.02%FS $'7'3 Pressure module for ADT760-D, -350 to 800 inH2O (-12.5 to 30 psi), ±0.02%FS $'7'3. Pressure module for ADT760-D, -350 to 1K inH2O (-12.5 to 35 psi), ±0.02%FS $'73 Pressure module for ADT760-D -MA, ±10 psi, ±0.02%FS $'73 Pressure module for ADT760-D -MA, -12.5 to 15 psi, ±0.02%FS $'73 Pressure module for ADT760-D -MA, -12.5 to 30 psi, ±0.02%FS $'73 Pressure module for ADT760-D -MA, -12.5 to 35 psi, ±0.02%FS $'73 Pressure module for ADT760-D -MA, -12.5 to 50 psi, ±0.02%FS $'73 Pressure module for ADT760-MA, -12.5 to 100 psi, ±0.02%FS $'73 Pressure module for ADT760-MA, -12.5 to 150 psi, ±0.02%FS $'73 Pressure module for ADT760-MA, -12.5 to 200 psi, ±0.02%FS $'73 Pressure module for ADT760-MA, -12.5 to 300 psi, ±0.02%FS
  • 18. Pressure/ProcessCalibrationEquipmentAdditelCatalog2016 Look us up on www.additel.com or call today (1)714-998-6899 $FFHVVRULHV ,QFOXGHG
  • 19. 0RGHO 4XDQWLW 3LFWXUH ADT100-760-KIT adapter set (excl ADT760-LLP) 1 set (to 1/4BSPF, to 1/4NPTF, to M20F) ADT100-760 Hose (excl ADT760-LLP) 1 pc 9816-X 110V/220V external power adapter 1 pc 9724 chargeable Li-ion battery 1 pc 9025 test leads for calibrator 1 sets (3 pcs) CD Manual 1 pc NIST-traceable EGTVKſECVG 1 pc * Additel/Land software available for free download at www.additel.com 2SWLRQDO $FFHVVRULHV 0RGHO 'HVFULSWLRQ 3LFWXUH ADT160A Pressure modules (reference 160A data sheet; module connection cable sold separately) 9060 Pressure module connection cable 9510 Additel/Cal task management software ADT100-760-N Special Male Connector to 1/4NPT quick connector ADT100-760-B Special Male Connector to 1/4BSP quick connector ADT100-760-M Special Male Connector to M20x1.5 quick connector ADT100-760-N2 Special Male Connector to 1/2NPT quick connector ADT100-760-B2 Special Male Connector to1/2BSP quick connector ADT128-B Pneumatic Pressure Manifold for ADT760, -15 to 3000 psi, single port, 1/4 BSP F ADT128-B2 Pneumatic Pressure Manifold for ADT760, -15 to 3000 psi, single port, 1/2 BSP F ADT128-N Pneumatic Pressure Manifold for ADT760, -15 to 3000 psi, single port, 1/4 NPT F ADT128-N2 Pneumatic Pressure Manifold for ADT760, -15 to 3000 psi, single port, 1/2 NPT F ADT128-M Pneumatic Pressure Manifold for ADT760, -15 to 3000 psi, single port, M20x1.5 9240 DP gauge holder, rubber 1220211087 Filter, set of 2 pcs 9913-760-SC Soft carrying case for ADT760, test leads, and many accessories 9914-760 Carry case for ADT760 and various accessories ADT100-760-CNT Special male connector to special male connector (to be used to adapt from the ADT760 to ADT100-760-KIT adapters) 2UGHULQJ ,QIRUPDWLRQ 0RGHO7SH 0RGHO 1XPEHU 'HVFULSWLRQ $'7//3 Automatic Handheld Pressure Calibrator, ±30 inH2O (DP30 sensor included) $'7//3'/ Automatic Handheld Pressure Calibrator, ±30 inH2O (DP30 sensor included) w/HART and data logging $'7' Automatic Handheld Pressure Calibrator, -12.5 to 35 psi (CP35 sensor included) $'7''/ Automatic Handheld Pressure Calibrator, -12.5 to 35 psi (CP35 sensor included) w/ HART and data logging $'70$ Automatic Handheld Pressure Calibrator, -12.5 to 300 psi (CP300 sensor included) $'70$'/ Automatic Handheld Pressure Calibrator, -12.5 to 300 psi (CP300 sensor included) w/ HART and data logging 3UHVVXUH UDQJH See Model Type table. $'7 //3 0RGHO 1XPEHU
  • 20. Look us up on www.additel.com or call today (1)714-998-6899 0RGHO )HDWXUHV //3 ' / /$ 0 0$ + +$ %3 3UHVVXUH 5DQJH -25 to 25 mbar (-10 to 10 inH2O) -0.95 to 1 bar (-13.5 to 15 psi) -0.95 to 7 bar (-13.5 to 100 psi) 0.05 to 8 bar.a (1.2 to 115 psi.a) -0.90 to 25 bar (-13 to 375 psi) 0.1 to 26 bar.a (1.7 to 390 psi.a) -0.90 to 40 bar (-13 to 600 psi) 0.1 to 41 bar.a (1.7 to 615 psi.a) 100 to 1200 hPa 3UHVVXUH 0RGXOH -2.5 to 2.5 mbar (-1 to 1 inH2O) -25 to 25 mbar (-10 to 10 inH2O) -0.95 to 2.5 bar (-13.5 to 35 psi) 0.05 to 3.5 bar.a (1.2 to 50 psi.a) -0.9 to 2.5 bar (-13 to 35 psi) 0.1 to 3.5 bar.a (1.7 to 50 psi.a) -0.9 to 2.5 bar (-13 to 35 psi) 0.1 to 3.5 bar.a (1.7 to 50 psi.a) 100 to 1200 hPa 3UHVVXUH 0RGXOH -25 to 25 mbar (-10 to 10 inH2O) -0.95 to 1 bar (-13.5 to 15 psi) 0 to 7 bar (0 to 100 psi) 0.05 to 8 bar.a (1.2 to 115 psi.a) 0 to 25 bar (0 to 375 psi) 0.1 to 26 bar.a (1.7 to 390 psi.a) 0 to 40 bar (0 to 600 psi) 0.1 to 41 bar.a (1.7 to 615 psi.a) NA 'LIIHUHQWLDO 3UHVVXUH Ŗ Ŗ *DXJH 3UHVVXUH Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ $EVROXWH 3UHVVXUH Ŗ Ŗ Ŗ Ŗ %DURPHWULF 3UHVVXUH Ŗ Ŗ Ŗ Ŗ 7HPSHUDWXUH RPSHQVDWHG Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ %XLOWLQ (OHFWULDO 3XPS Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ %XLOWLQ )LOWHU Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ %XLOWLQ /LTXLG 7UDS Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ 6RXUFH WR P$ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ 0HDVXUH P$ RU 9 Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ 0HDVXUH H[WHUQDO SUHVVXUH PRGXOH Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ 9 /RRS SRZHU Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ 3UHVVXUH VZLWFK WHVW Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ +$57 RPPXQLFDWLRQ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ 'RFXPHQWLQJ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ $XWRPDWHG 3UHVVXUH DOLEUDWRUV 6HOHFWLRQ *XLGH $GGLWHO
  • 21. Pressure/ProcessCalibrationEquipmentAdditelCatalog2016 Look us up on www.additel.com or call today (1)714-998-6899 $XWRPDWHG 3UHVVXUH DOLEUDWRUV $GGLWHO )XOO DXWRPDWHG SUHVVXUH FDOLEUDWRU ZLWK EXLOWLQ SUHVVXUH JHQHUDWRU FRQWUROOHU WR DV KLJK DV SVL EDU
  • 22. RU DV ORZ DV 3D LQ+2
  • 23. )6 DFFXUDF 'XDO SUHVVXUH PRGXOHV %XLOWLQ ILOWHU DQG OLTXLG WUDS ZLWK YHQWLQJ VVWHP SUHYHQWV FRQWDPLQDWLRQ RI WKH FDOLEUDWRU 3RUWDEOH OE
  • 24. 29(59,(: With a built-in high performance electronic pump and precision pressure controller, the 761 series portable automated pressure calibrators provide a turnkey solution for ECNKDTCVKQP QH ICWIGU VTCPUOKVVGTU CPF UYKVEJGU DQVJ KP VJG ſGNF CPF KP VJG NCDQTCVQT[ In a portable package, this calibrator can automatically generate pressures from 90% vacuum to 600 psi (40 bar) with 0.005% FS pressure control stability and 0.02% FS accuracy. To improve the calibrator accuracy, two pressure modules with differing ranges are built-in and integrated with the internal pump and controller. The 761-LLP is specially designed for low pressure calibration, and pressures can be set to as low as 0.01 Pa (0.00004 inH2O) with a control stability better than 0.05 Pa (0.0002 inH2O). The 761-BP is designed for barometric gauge calibration. With a quartz pressure sensor, the accuracy of 761-BP can achieve 0.01%FS accuracy. With optional external pressure modules (160A series), the 761 can measure pressures up to 10,000 psi (700 bar) with 0.02%FS accuracy. In addition to the pressure generation, control, and measurement capabilities, the 761 also features HART communication capability, supplies 24V loop power, and reads the current or voltage produced by the pressure transducers. The 761 is a state of the art automated RTGUUWTG ECNKDTCVQT YJKEJ DTKPIU CWVQOCVGF RTGUUWTG ECNKDTCVKQP VQ VJG ſGNF
  • 25. Look us up on www.additel.com or call today (1)714-998-6899 )($785(6 3UHVVXUH 5DQJH Pressure generated by built-in pump and controlled automatically from 90% vacuum to 600 psi (40 bar) Pressure measured with external pressure modules (160A series) to 10,000 psi (700 bar) $FFXUDF )XOO WHPSHUDWXUH FRPSHQVDWHG
  • 26. 0.02%FS 0.05%FS (ADT761-LLP) 0.01%FS (ADT761-BP) RQWURO 6WDELOLW 0.005%FS ADT761-LLP: 0.05 Pa (0.0002 inH2O) RQWDPLQDWLRQ SUHYHQWLRQ VVWHP Built-in filter and liquid trap with venting system prevent contamination of the calibrator 5QNKF RCTVKENGU ECP DG DNQEMGF D[ DWKNVKP ſNVGT A small amount of liquid can be collected by the built- in liquid trap Both solid particles and liquid can be blown out through the venting system $XWRPDWLRQ DQG7HVW 3URJUDP Fully Automated pressure calibration Task preset and auto run Auto step (DV WR 8VH User friendly interface Intuitive icon driven menu structure 6RXUFLQJ DQG 0HDVXULQJ HOHFWULFDO VLJQDOV Measure mA with 0.01% RD + 1.5μA accuracy Measure V with 0.01% RD + 1.5mV accuracy Source/sink 0 to 22mA current 24 VDC loop power supplied to pressure transmitter during test Automated switch test 0LVXVH 3URWHFWLRQ Up to 30V misuse protection on any two sockets 'LVSOD 7 LCD color screen (800X480) 'RFXPHQWLQJ DQG RQGHPDQG ORJJLQJ Task management capability Internal memory stores 200 tasks Download tasks and upload results 900 snapshots +$57 RPPXQLFDWLRQ FDSDELOLW Support HART instrumentation 3UHVVXUH RXWSXW UDWH 2.4 liters/minute 5HFKDUJHDEOH EDWWHU Rechargeable battery with up to 8 hours operation between charges /HDN WHVWLQJ Apply leak test to an external pressure system to determine the magnitude of pressure variations due to leaks 3LSHOLQH 3URWHFWLRQ Special design the pipeline layout to avoid possible blockage and leakage 3RUWDEOH Only 12.4 lb 0XOWL /LQJXDO ,QWHUIDFH English, German, French, Italian, Spanish, Portuguese, 5KORNKſGF %JKPGUG (Traditional Chinese, Japanese and Russian are available per request) 1,677UDFHDEOH NIST traceable calibration with data (included)
  • 27. Pressure/ProcessCalibrationEquipmentAdditelCatalog2016 Look us up on www.additel.com or call today (1)714-998-6899 7DVN 0DQXDO %XLOWLQ 8VHU V 0DQXDO 3UHVVXUH JDXJH WUDQVPLWWHU VZLWFK FDOLEUDWLRQ 'RFXPHQWLQJ RQWURO 6WDELOLW 0.005%FS[1] ; ADT761-LLP: 0.05 Pa (0.0002 inH2O) 3UHVVXUH RQQHFWLRQV Pressure connection: ADT761-L/M/H/LA/MA/HA One ADT100-761 Hose Test Kit with quick connector 1/8 BSP Female on calibrator Two 1/4NPT(or 1/2NPT, 1/4BSP, 1/2BSP, M20×1.5) female hand-tight connectors on the external pressure manifold Pressure connection: ADT761-LLP/BP 5VCKPNGUU UVGGN JQUG DCTD ſVVKPI ([WHUQDO 3UHVVXUH 0RGXOH RQQHFWLRQ Accepts Intelligent Pressure Module (OHFWULFDO RQQHFWLRQ 0.156 inch (Ø4mm) banana sockets 'RFXPHQWLQJ DQG RQGHPDQG ORJJLQJ Task management capability Internal memory stores 200 tasks Download tasks and upload results 'LVSOD 7 LCD color screen (800×480) Display rate: 5 readings per second Resolution 4, 5 or 6 digit, user selectable 3UHVVXUH 8QLWV Pa, kPa, MPa, psi, bar, mbar, inH2O@4°C, mmH2O@4°C, inH2O@20°C, mmH2O@20°C, inHg@0°C, mmHg@0°C, kgf/cm2 , and hPa (ADT761- BP) 3XPS OLIH 100,000 cycles 3RZHU Battery: Rechargeable Li-ion Polymer battery (included) Li-Battery working time: ~8 hours Recharge time: ~3 hours External Power: 110/220V Power adapter (DC 27 V) (Included) (QYLURQPHQWDO 6SHFLILFDWLRQ Operating temperature: 32°F to 122°F (0°C to 50°C) Compensated Temperature: 32°F to 122°F (0°C to 50°C) Storage temperature: -20°C to 60°C (-4°F to 120°F) Humidity: 90%, non-condensing (XURSHDQ RPSOLDQFH CE Marked RPPXQLFDWLRQ RS232 (DB9/F, environmentally sealed) Baud rate: 2400, 4800, 9600, 19200, 38400, 57600, 115200 Data length: 7 Bits, 8 Bits Stop bit: 1 Bit, 2 Bits Address: from 1 to 121 'LPHQVLRQ 11.65 X 7.32 X 7.09 in (296 X186 X180 mm) :HLJKW 12.4 to 16.3 lb (5.6 to 7.4 kg) :DUUDQW 18 months =? (5 URGEKſECVKQP CRRNKGU VQ VJG URCP QH VJG TCPIG 63(,),$7,216
  • 28. Look us up on www.additel.com or call today (1)714-998-6899 $XWRPDWHG 3UHVVXUH DOLEUDWRUV $GGLWHO 3UHVVXUH 5DQJHV (OHFWULFDO 0HDVXUHPHQW DQG 6RXUFH $FFXUDF 63(,),$7,216 =? 1PG [GCT CEEWTCE[ KPENWFKPI [GCT UVCDKNKV[ (5 URGEKſECVKQP CRRNKGU VQ VJG URCP QH VJG OQFWNG TCPIG [2] 0.05% FS accuracy (including 6 months stability). One year accuracy is 0.05% FS calibration accuracy combined with 0.05% FS one year stability. * Additel 761 calibrators support 160A series intelligent digital pressure modules that are available for gauge, vacuum and absolute pressure from -15 psi to 10,000 psi (-1 bar to 700 bar). Accuracy from 0.02%FS includes operation over 14°F to 122°F (-10°C to 50°C), one year UVCDKNKV[ CPF ECNKDTCVKQP WPEGTVCKPV[ (QT FGVCKNGF URGEKſECVKQP TGHGT VQ # UGTKGU RTGUUWTG OQFWNGU FCVCUJGGV 7SH 0RGHO 3UHVVXUH 5DQJH 3UHVVXUH 0RGXOH 3UHVVXUH 0RGXOH 5DQJH $FFXUDF@ 5DQJH $FFXUDF@ Low/Differential Pressure ADT761-LLP -25 to 25 mbar (-10 to 10 inH2O) -2.5 to 2.5 mbar (-1 to 1 inH2O) 0.05%FS[2] -25 to 25 mbar (-10 to 10 inH2O) 0.05%FS[2] ADT761-D -0.95 to 1 bar (-13.5 to 15 psi) -25 to 25 mbar (-10 to 10 inH2O) 0.05%FS[2] -0.95 to 1 bar (-13.5 to 15 psi) 0.02%FS Gauge Pressure ADT761-L -0.95 to 7 bar (-13.5 to 100 psi) -0.95 to 2.5 bar (-13.5 to 35 psi) 0.02%FS 0 to 7 bar (0 to 100 psi) 0.02%FS ADT761-M -0.90 to 25 bar (-13 to 375 psi) -0.9 to 2.5 bar (-13 to 35 psi) 0.02%FS 0 to 25 bar (0 to 375 psi) 0.02%FS ADT761-H -0.90 to 40 bar (-13 to 600 psi) -0.9 to 2.5 bar (-13 to 35 psi) 0.02%FS 0 to 40 bar (0 to 600 psi) 0.02%FS Gauge/Absolute Pressure ADT761-LA -0.95 to 7 bar (-13.5 to 100 psi) -0.95 to 2.5 bar (-13.5 to 35 psi) 0.02%FS -0.95 to 7 bar (-13.5 to 100 psi) 0.02%FS 0.05 to 8 bar.a (1.2 to 115 psi.a) 0.05 to 3.5 bar.a (1.2 to 50 psi.a) 0.03%FS 0.05 to 8 bar.a (1.2 to 115 psi.a) 0.025%FS ADT761-MA -0.90 to 25 bar (-13 to 375 psi) -0.9 to 2.5 bar (-13 to 35 psi) 0.02%FS -0.9 to 25 bar (-13 to 375 psi) 0.02%FS 0.1 to 26 bar.a (1.7 to 390 psi.a) 0.1 to 3.5 bar.a (1.7 to 50 psi.a) 0.03%FS 0.1 to 26 bar.a (1.7 to 390 psi.a) 0.025%FS ADT761-HA -0.90 to 40 bar (-13 to 600 psi) -0.9 to 2.5 bar (-13 to 35 psi) 0.02%FS -0.9 to 40 bar (-13 to 600 psi) 0.02%FS 0.1 to 41 bar.a (1.7 to 615 psi.a) 0.1 to 3.5 bar.a (1.7 to 50 psi.a) 0.03%FS 0.1 to 41 bar.a (1.7 to 615 psi.a) 0.025%FS Barometric pressure ADT761-BP 100 to 1,200 hPa 100 to 1,200 hPa 0.01%FS NA NA 5DQJH 5HVROXWLRQ $FFXUDF 9ROWDJH 0HDVXUHPHQW ±30.0000V 0.1mV ±(0.01%RD + 1.5mV) XUUHQW 0HDVXUHPHQW ±30.0000mA 0.1μA ±(0.01%RD + 1.5μA) XUUHQW 6RXUFH 0.000 to 22.000mA 1μA ±(0.02%RD + 2.2μA) 6ZLWFK7HVW If the switch has detective Voltage, the range is from 3V to 24V ' 9 RXWSXW 24V ± 0.5V, max: 50mA
  • 29. Pressure/ProcessCalibrationEquipmentAdditelCatalog2016 Look us up on www.additel.com or call today (1)714-998-6899 2SWLRQDO $FFHVVRULHV 0RGHO 1XPEHU 'HVFULSWLRQ 3LFWXUH ADT100-761-N4 Hose Test Kit, 5 feet ƀGZKDNG JQUG (Ø4mm) tube to 1/4NPT hand-tight quick connector ADT100-761-B4 Hose Test Kit, 5 feet ƀGZKDNG JQUG (Ø4mm) tube to 1/4BSP hand-tight quick connector ADT106-N External Pressure Manifold (with two 1/4NPT hand-tight quick connectors) ADT106-B External Pressure Manifold (with two 1/4BSP hand-tight quick connectors) ADT106-M External Pressure Manifold (with two M20X1.5 hand-tight quick connectors) ADT160A Pressure modules (please check Additel 160A datasheet) 9050 USB to RS232 (DB9/M) Adapter 9060 Pressure module connection cable 9510 Additel/Cal task management software ADT100-761Hose ADT761 hose, 5 feet $FFHVVRULHV LQFOXGHG
  • 31. 1 pc 110V/220V external Power adapter 1 pc Chargeable Li-ion battery 1 pc Test Leads for calibrator 2 sets (4 pcs) Short circuit cable 1 set (2 pcs) Carrying case for 761 calibrator and accessory 1 pc Pressure module connection cable 1 pc DP gauge holder, rubber 2QO IRU $'7//3
  • 34. 20 pcs Manual 1 pc Allen wrench 1 set ADT100-761Hose (Except 761-LLP: 2 pcs) 1 pcs NIST traceable ECNKDTCVKQP EGTVKſECVG 1 pc $XWRPDWHG 3UHVVXUH DOLEUDWRUV $GGLWHO 25'(5,1* ,1)250$7,21 3UHVVXUH UDQJH See pressure range table. 3UHVVXUH SRUW WSH RQ ([WHUQDO 3UHVVXUH 0DQLIROG N - 1/4NPT female N2 - 1/2NPT female B - 1/4BSP female B2 - 1/2BSP female M - M20X1.5 female * Additel/Land software could be downloaded for free at www.additel.com $'7 / 1 0RGHO 1XPEHU
  • 35. Look us up on www.additel.com or call today (1)714-998-6899 $GGLWHO 6HULHV 3UHVVXUH RQWUROOHU 3UHVVXUH UDQJHV IURP YDFXXP WR SVL EDU
  • 36. 5HPRYDEOH LQWHUFKDQJHDEOH LQWHOOLJHQW VHQVRUV 3UHFLVLRQ DFFXUDF RI RI UHDGLQJ SOXV RI IXOO VFDOH 6WDQGDORQH VROXWLRQ WR SVL EDU
  • 37. QR JDV ERWWOH UHTXLUHG ZKHQ XVHG ZLWK WKH (OHFWULF 3XPS DQG . *DV %RRVWHU 6WDQGDORQH VROXWLRQ WR SVL EDU
  • 38. QR JDV ERWWOH UHTXLUHG ZKHQ XVHG ZLWK WKH (OHFWULF 3XPS 6WDQGDUG PRGHO DFFXUDF RI RI IXOO VFDOH ([WHUQDO SUHVVXUH PRGXOHV WR SVL EDU
  • 39. :L)L HQDEOHG FRPPXQLFDWLRQV )XOO WHPSHUDWXUH FRPSHQVDWHG DFFXUDF RYHU °C WR °C +$57 RPPXQLFDWLRQ DQG 3URILEXV 3$ /DUJH ´ FRORU WRXFK VFUHHQ GLVSOD RQWURO VWDELOLW RI )6 %XLOWLQ EDURPHWHU (DVWRXVH LFRQ EDVHG XVHU LQWHUIDFH 29(59,(: For years, we’ve provided the most durable, accurate, quality pressure calibration products for field applications. The Additel 780 series controller incorporates the same durability, accuracy, and quality into a new bench top controller packed with features and functionality that is remarkably easy to use. The Additel 780 series offers two base ranges: to 1,000 psi (70 bar) and to 3,000 psi (200 bar). The base range establishes the OCZKOWO EQPVTQNNKPI TCPIG QH VJG EQPVTQNNGT 'CEJ EQPſIWTCVKQP KPENWFGU C EQPVTQN UGPUQT which is preselected to the sensor range best suited for your application. External and internal sensors can be used which allows for expanded range and accuracy capability in the future. There are also two controller types that can be selected: the Additel 780S is the standard controller option without any measurement capability. The Additel 780 has expanded HWPEVKQPCNKV[ KPENWFKPI GNGEVTKECN OGCUWTGOGPV CPF *#46 CPF 2TQſDWU 2# EQOOWPKECVKQP
  • 40. Pressure/ProcessCalibrationEquipmentAdditelCatalog2016 Look us up on www.additel.com or call today (1)714-998-6899 02'8/$5 '(6,*1 Each unit comes with one Intelligent Pressure Module configured to the many range offerings provided. Standard accuracy sensors (ADT160A-CPXXX) are silicone pressure sensors with a 1 year accuracy of 0.02% FS. The RTGEKUKQP SWCTV DCUGF UGPUQTU KORTQXG VJG [GCT CEEWTCE[ URGEKſECVKQP VQ QH TGCFKPI (5 'CEJ sensor has been specially aged, tested and screened before assembly. After assembly each sensor is temperature compensated over the range of 0°C to 50°C. The Additel 780 series allows for one internal pressure sensor and one external pressure sensor. The modular design of this unit provides for interchangeability of both the internal and external sensors with other Intelligent Pressure Modules. In addition to the Intelligent Pressure Modules, the Additel 780 series has a built in barometric sensor. This allows for switching between gauge pressures to absolute pressures. 67$1'$/21( 62/87,21 Typical pressure controllers will require a nitrogen bottle for the gas supply which make it FKHſEWNV VQ OQXG VJG EQPVTQNNGT CTQWPF YKVJQWV JCXKPI VQ OQXG QT EQPPGEV VQ CPQVJGT DQVVNG The Additel 780 series is unique in that with the optional electric pump, you can generate pressures to 1,000 psi (70 bar) without the need of a gas bottle. Add the booster to the EQPſIWTCVKQP CPF RUK DCT KU CEJKGXCDNG YKVJ VJG #FFKVGN EQPVTQNNGT UGTKGU The 780 Series Pressure Controller is considered to handle a very wide range of applications which may normally require a pressure calibrator. Built-in capability, includes current and voltage measurement capability, 24 volt loop power, HART® CPF 2TQſDWU 2# EQOOWPKECVKQP switch measurement capability, and much more. 6SHFLILFDWLRQ $'7. $'7. 3UHVVXUH UDQJH -14.5 to 1,000 psi (-0.95 to 70 bar) -14.5 to 3,000 psi (-0.95 to 200 bar)[1] RQWURO VWDELOLW 0.003% FS (stability based on % FS of control sensor range) 3UHFLVLRQ LQFOXGHV HDU VWDELOLW
  • 41. See pressure range table HDU VWDELOLW 0.01% FS standard accuracy sensors 0.005% FS precision accuracy sensors 0HGLD Clean gas 2YHUUDQJH LQGLFDWLRQ 103% to 120% (based on sensor) 5HVROXWLRQ 4, 5, 6, or 7 digits (user selectable)[2] 3UHVVXUH WSH Gauge, Absolute :DUP XS WLPH 15 minutes 7SLFDO SUHVVXUH VHWWOLQJ WLPH ZLWKLQ )6
  • 42. 20 seconds (10%FS pressure change assuming 50 mL test volume) 0HDVXUHPHQW XQLWV Pa, hPa, kPa, MPa, bar, mbar, psi, mmHg@0°C,cmHg@0°C, mHg@0°C, inHg@0°C, inH2O@4°C, mmH2O@4°C, cmH2O@4°C, mH2O@4°C, mmH2O@20°C, cmH2O@20°C, mH2O@20°C, kg/m2 , kg/cm2 , mtorr, torr, atm, lb/ft2 , tsi, user selectable 0LQLPXP FRQWURO SUHVVXUH@ 0.0001 psi 0.001 psi [1] Gas booster or HP gas supply required to reach 3,000 psi (200 bar). Booster ordered separately. [2] 7 digit resolution for precision model only. [3] Dependent on pressure module. 0($685(0(17 63(,),$7,216 352(66 )817,21$/,7 H[FO $'76 YHUVLRQV
  • 43.
  • 44. Look us up on www.additel.com or call today (1)714-998-6899 6SHFLILFDWLRQ 5DQJH 5HVROXWLRQ $FFXUDF 9ROWV ' -30 to 30 V 0.1 mV ±0.01% rdg + 1.5 mV -300 to 300 mV 1 μV ±0.01% rdg + 15 μV XUUHQW ' -30 to 30 mA 0.1 μA ±0.01% rdg + 1.5 μA 6ZLWFK WHVW If the switch has detected voltage, the range is from 3 – 24 V ' 9 RXWSXW 24V ± 0.24V, max 30 mA [1] Not available in ADT780S versions (/(75,$/ 0($685( 63(,),$7,216@ 6SHFLILFDWLRQ ADT780-1K ADT780-3K 3RZHU 100 to 240V, 50/60 Hz 3UHVVXUH SRUWV G1/8 F 6WRUDJH WHPSHUDWXUH -20°C to 70°C 2SHUDWLQJ HQYLURQPHQW 0-90% RH non-condensing 'LVSOD 7 inch (17.8 cm) color, touch screen display :HLJKW 33 lbs (15 kg) 'LPHQVLRQV ':+
  • 45. 16.5 x 17.3 x 5.2 inch (419 x 440 x 132 mm) 0RXQWLQJ Standard desktop, optional rack mount kit 6KRFN 4G 9LEUDWLRQ 1G 10Hz~500Hz 3+6,$/ 63(,),$7,216 :L)L VSHFLILFDWLRQV 802.11 b, g, and n 9HQW Front panel vent and safety release button (only for 780S) RPPXQLFDWLRQV RS232,USB,LAN,WIFI 6WDELOLW LQGLFDWRU User selectable 7SLFDO VOHZ UDWH Typically 20 seconds 2SHUDWLQJ PRGHV Control, measure, and vent 'LVSOD PRGHV Controller – show pressure indication and control Calibrator – shows pressure indication, control, and electrical measurement (excl ADT780S) /RFDOL]DWLRQ 'PINKUJ %JKPGUG UKORNKſGF )GTOCP 5RCPKUJ (TGPEJ +VCNKCP 2QTVWIWGUG 4WUUKCP ,CRCPGUG RQIRUPLW CE DOLEUDWLRQ FHUWLILFDWLRQ 0+56VTCEGCDNG EGTVKſECVG QH ECNKDTCVKQP YKVJ FCVC KPENWFGF :DUUDQW 1 year 27+(5 63(,),$7,216 6HQVRU0RGH *DXJH $EVROXWH ADT160-02-CPXX N/A 40 Pa[1] ADT160-01-APXXQ 3 Pa[2] N/A Barometer range (60~110)kPa, the accuracy is 40 Pa (11 Pa is optional) [1] 40 Pa uncertainty (k=2) includes calibration uncertainty, linearity, and long term stability (30 Pa per year). Barometer range of 60 to 120 kPa. [2] Combined linearity, hysteresis, and repeatability. Add 3 Pa when used in gauge mode. When using the ADT160-01-APXXQ sensors with the ADT780 controller in gauge mode, regular zeroing will realign the barometric reading to provide the most accurate result. %$520(75, 0($685(0(17 63(,),$7,21
  • 46. Pressure/ProcessCalibrationEquipmentAdditelCatalog2016 Look us up on www.additel.com or call today (1)714-998-6899 $'7; ([WHUQDO 0DQLIROG 1 pc 0DQLIROG FRQQHFWLRQ KRVH 1 pc 3RZHU DGDSWHU 1 pc 7HVW OHDGV H[FHSW 6
  • 47. 3 sets (6 pc) 6KRUW FLUFXLW FDEOH H[FHSW 6
  • 48. 1 set (2 pc) 3UHVVXUH PRGXOH FRQQHFWLRQ FDEOH 1 pc $GDSWHU VHW DGDSWHUV WR ILW WKH $'7; SRUW WR PDOH ILWWLQJV
  • 49. 1 set (2 pcs G1/8M to G1/4M, 2 pcs G1/8M to Festo, release valve) ' 0DQXDO 1 pc 1,67 WUDFHDEOH FHUWLILFDWH 1 pc 2ULQJV 2 types 9DFXXP9HQW WXELQJ 2 pc 67$1'$5' $(6625,(6 $'7$ Pressure modules (see module information) $'7; External pneumatic pressure manifold, 3,000 psi (200 bar), four hand-tight quick connectors $'7.%67 3,000 psi (200 bar) gas booster $'7.(3803 1,000 psi (70 bar) electrical pump RS232 to USB adapter (;7 RS 232 (DB9/M) extension cable, 9 feet $'7)/7 Liquid trap Shipment case for the ADT780 and ADT780S Controller $GGLWHO3DO Manual calibration software $GGLWHO$DO Automated calibration software, Task and asset management Rack mount kit $'7.LW Adapter set, total 9 pcs. G1/8 M to 1/4NPT M (3 pcs) G1/8 M to G1/4 M (3 pcs) G1/8 M to M20x1.5 M (3 pcs) 237,21$/ $(6625,(6 $'7 25'(5,1* ,1)250$7,21 0RGHO 1XPEHU 0RGHO ADT780 ADT780S +LJK $FFXUDF %DURPHWHU 0DQLIROG SRUW WSH N-1/4NPT male N2-1/2NPT male B-1/4BSP male B2-1/2BSP male M-M20x1.5 male $'7 3 +$%31N 3UHVVXUH UDQJH See ADT160A range table on page 17 0D[PXP 5DQJH 3K-3,000 psi (200 bar) 1K-1,000 psi (70 bar) RQWUROOHU ZLWKRXW . SVL ERRVWHU
  • 50.
  • 51. Look us up on www.additel.com or call today (1)714-998-6899 31 3UHVVXUH 5DQJH 0HGLD $FFXUDF 3UHVVXUH 5DWLQJ 6XJJHVWHG RQWUROOHU RPSDWLELOLW ,QWHUQDO ([WHUQDO 8VHSVLD EDUD %XUVW 2YHU 3UHVVXUH 3UHFLVLRQ 4XDUW] 6HQVRUV@ $34 0 to 45 3.0 G 0.005% rdg + 0.005% FS 3x 1.2x ADT780-1K only Both $34 0 to 100 7.0 G 0.005% rdg + 0.005% FS 3x 1.2x ADT780-1K only Both $34 0 to 300 20 G 0.005% rdg + 0.005% FS 3x 1.2x Both Both $34 0 to 400 28 G 0.005% rdg + 0.005% FS 3x 1.2x Both Both $34 0 to 500 35 G 0.005% rdg + 0.005% FS 3x 1.03x Both Both $3.4 0 to 1,000 70 G 0.005% rdg + 0.005% FS 3x 1.1x Both Both $3.4 0 to 2,000 140 G 0.005% rdg + 0.005% FS 3x 1.1x ADT780-3K only Both $3.4 0 to 3,000 200 G 0.005% rdg + 0.005% FS 3x 1.1x ADT780-3K only Both 6WDQGDUG 3UHVVXUH 6HQVRUV 31 3UHVVXUH 5DQJH 0HGLD $FFXUDF 3UHVVXUH 5DWLQJ 6XJJHVWHG RQWUROOHU RPSDWLELOLW ,QWHUQDO ([WHUQDO 8VHSVLJ EDUJ %XUVW 2YHU 3UHVVXUH 3 ±10 ±0.7 G 0.02% FS 3x 1.2x ADT780-1K only Both 3 ±15 ±1.0 G 0.02% FS 3x 1.2x ADT780-1K only Both 3 -15 to 30 -1 to 2.0 G 0.02% FS 3x 1.2x ADT780-1K only Both 3 -15 to 50 -1 to 3.5 G 0.02% FS 3x 1.2x ADT780-1K only Both 3 -15 to 100 -1 to 7.0 G,L 0.02% FS 3x 1.2x ADT780-1K only Both 3 -15 to 300 -1 to 20 G,L 0.02% FS 3x 1.2x Both Both 3 -15 to 500 -1 to 35 G,L 0.02% FS 3x 1.2x Both Both 3 -15 to 600 -1 to 40 G,L 0.02% FS 3x 1.2x Both Both 3. -15 to 1,000 -1 to 70 G,L 0.02% FS 3x 1.2x Both Both 3. -15 to 2000 -1 to 140 G,L 0.02% FS 3x 1.2x ADT780-3K only Both 3. -15 to 3,000 -1 to 200 G,L 0.02% FS 3x 1.2x ADT780-3K only Both 3. -15 to 5000 -1 to 350 G,L 0.02% FS 3x 1.2x N/A External 3. -15 to 10,000 -1 to 700 G,L 0.02% FS 2x 1.2x N/A External [1] Contact Additel for other range options. **Low pressure sensors (ADT160A-XX-DPX) available for low pressure and differential pressure measurement. Also available for low pressure control (gauge mode only). Ranges from ±1 inH20 (2.5 mbar) to 300 inH2O (700 mbar). $'7$ ,QWHOOLJHQW 'LJLWDO 3UHVVXUH 0RGXOHV *DXJH SUHVVXUH 'LIIHUHQWLDO SUHVVXUH 63(,),$7,216
  • 52. Pressure/ProcessCalibrationEquipmentAdditelCatalog2016 Look us up on www.additel.com or call today (1)714-998-6899 6WDQGDUG $FFXUDF 3UHFLVLRQ $FFXUDF 3;;; $3.4 $3.4 $34 7HPSHUDWXUH FRPSHQVDWLRQ -10°C to 50°C (14°F to 122°F) 0°C to 50°C (32°F to 122°F) 2SHUDWLQJ WHPSHUDWXUH -10°C to 50°C (14°F to 122°F) 0°C to 50°C (32°F to 122°F) 6WRUDJH WHPSHUDWXUH -20°C to 70°C (-4°F to 158°F) -20°C to 70°C (-4°F to 158°F) 5HODWLYH KXPLGLW 95% RH 95% RH 3UHVVXUH FRQQHFWLRQV IRU H[WHUQDO XVH RQO
  • 53. 1/4NPT, 1/4BSP, 1/2BSP, M20x15 1/4NPT, 1/4BSP, 1/2BSP, M20x15 (QFORVXUH IRU H[WHUQDO XVH RQO
  • 54. SS enclosure SS enclosure 'LPHQVLRQV 'LD [ +
  • 55. 33 mm x 123 mm (1.3” x 4.84”) 44 mm x 220mm (1.7” x 8.7”) 60 mm x 190 mm (2.4” x7.5”) :HLJKW 0.4 kg (0.99 lb) 0.8 kg (1.8 lb) 1.2 kg (2.6 lb) [1] Internal sensors used to go inside the ADT780 need to have a M20 connection and connection kit. 25'(5,1* ,1)250$7,21 0RGHO 1XPEHU 0RGHO 5DQJH WSH PSI - range by psi BAR - range by bar ([WHUQDO SUHVVXUH SRUW WSH N - 1/4NPT male N2 - 1/2NPT male B - 1/4BSP male B2 - 1/2BSP male M - M20x1.5 male $'7$ 3 36, 1 3UHVVXUH UDQJH See range table $FFXUDF 01 – 0.005% rdg + 0.005% FS 02 – 0.02% FS 6(1625 63(,),$7,216
  • 56. Look us up on www.additel.com or call today (1)714-998-6899 (/(75, 3803 0LQLPXP LQOHW SUHVVXUH 300 psi (20 bar) 2XW SUHVVXUH UDQJH 300 to 3,600 psi (250 bar) ,QOHW IORZ UDWH 500ml/min 6L]H :+'
  • 57. 17.8 X 7.3 X 23.9 inch (451 X 185 X 607 mm) 9ROXPH 375 mL :HLJKW 74 lbs (33.6 kg) 3RZHU 100 to 240V 6XSSO PHGLD Nitrogen or clean gas SVL EDU
  • 58. *DV %RRVWHU 6SHFLILFDWLRQV 3UHVVXUH UDQJH -13.5 to 1,080 psi (-0.95 to 75 bar) :HLJKW 69.6 lbs (31.6 kg) 6L]H:+'
  • 59. 17.3 X 9.3 X 21.7 inch (440 X 235X 550 mm) 3RZHU 100 to 240V 0HGLD Air 9ROXPH 150 mL 2XWOHW SRUW G1/8F 6WRUDJH WHPSHUDWXUH -20°C to 70°C 2SHUDWLQJ HQYLURQPHQW 0-90% RH non-condensing 0RXQWLQJ Standard desktop, optional rack mount RQWURO RSHUDWLRQ 2 LED displays with pressure limit settings 7SLFDO PD[ SUHVVXUH WLPH Approx 2 min with ADT780 Controller SVL EDU
  • 60. (OHFWULF 3XPS 6SHFLILFDWLRQV $GGLWHO .(3803 %2267(5$GGLWHO .%67 *HQHUDWHV 9DFXXP WR SVL EDU
  • 61. %XLOWLQ ILOWHU DQG OLTXLG WUDS (OHFWULF 3XPS DQG %RRVWHU
  • 62. Pressure/ProcessCalibrationEquipmentAdditelCatalog2016 Look us up on www.additel.com or call today (1)714-998-6899 $GGLWHO 3UHVVXUH *DXJH 6HOHFWLRQ *XLGH 6HULHV )HDWXUH $'7 6HULHV 'LJLWDO 3UHVVXUH DOLEUDWRU $'7 6HULHV 'LJLWDO 3UHVVXUH *DXJH $'7 6HULHV 'LJLWDO 3UHVVXUH *DXJH *DXJH 3UHVVXUH Ŗ Ŗ Ŗ RPSRXQG 3UHVVXUH Ŗ Ŗ Ŗ $EVROXWH 3UHVVXUH Ŗ Ŗ 'LIIHUHQWLDO 3UHVVXUH Ŗ Ŗ $FFXUDF ODVVHV 0.02% 0.05%FS 15K 20K psi: 0.05%FS 20K psi: 0.1%FS 0.02%, 0.05%, 0.1%, 0.2%FS 0.1%RD 15K 20K psi: 0.05%, 0.1%, 0.2%FS 0.1%RD 20K psi: 0.1% 0.2%FS 0.05%, 0.1%, 0.25%FS 20K psi: 0.1% 0.25%FS 'LJLWDO 'LVSOD Ŗ Ŗ Ŗ $QDORJ 'LVSOD )DQ6KDSHG ,QGLFDWLRQ
  • 63. Ŗ )XOO7HPSHUDWXUH RPSHQVDWLRQ IURP Ĉ WR Ĉ Ŗ Ŗ Ŗ 5HVROXWLRQ 'LJLW 5HVROXWLRQ Ŗ 'LJLW 5HVROXWLRQ Ŗ Ŗ Ŗ 6HOHFWDEOH 3UHVVXUH 8QLWV 11 11 19 %DFNOLJKW Ŗ Ŗ Ŗ 2YHU 3UHVVXUH ,QGLFDWLRQ Ŗ Ŗ Ŗ ,6 HUWLILFDWLRQ RSWLRQDO ă QRW DYDLODEOH IRU SDQHO PRXQW
  • 64. #6+5 QPN[ Ů)2 ,3 HUWLILFDWLRQ Ŗ 3DQHO 0RXQW RSWLRQDO
  • 65. Ŗ :LUHOHVV 680W only 'DWD /RJJLQJ Ŗ Optional 680W only 0LQ0D[ Ŗ Ŗ Ŗ %XLOWLQ /HDN7HVW Ŗ Ŗ +$57 RPPXQLFDWLRQ Ŗ 0HDVXUH P$ DQG 9 Ŗ 9 /RRS 3RZHU Ŗ 6ZLWFK7HVW Ŗ 1,677UDFHDEOH HUWLILFDWH RI DOLEUDWLRQ Ŗ Ŗ Ŗ 3RZHU Rechargeable battery 9V battery (120/220V adapter is optional) 2AA batteries 3UHVVXUH 3UHVVXUH 5DQJH 0HGLD $'7 6HULHV 'LJLWDO 3UHVVXUH DOLEUDWRU $'7 6HULHV 'LJLWDO 3UHVVXUH *DXJH $'7 6HULHV 'LJLWDO 3UHVVXUH *DXJH SVL EDU *DXJH 9 -15 to 0 -1 to 0 G Ŗ Ŗ Ŗ *3 0 to 5 0 to 0.35 G Ŗ Ŗ *3 0 to 10 0 to 0.7 G Ŗ Ŗ *3 0 to 15 0 to 1 G, L Ŗ Ŗ Ŗ *3 0 to 30 0 to 2 G, L Ŗ Ŗ Ŗ *3 0 to 50 0 to 3.5 G, L Ŗ Ŗ *3 0 to 100 0 to 7 G, L Ŗ Ŗ Ŗ *3 0 to 300 0 to 10 G, L Ŗ Ŗ Ŗ *3 0 to 500 0 to 35 G, L Ŗ Ŗ Ŗ 6HULHV
  • 66. Look us up on www.additel.com or call today (1)714-998-6899 3UHVVXUH 3UHVVXUH 5DQJH 0HGLD $'7 6HULHV 'LJLWDO 3UHVVXUH DOLEUDWRU $'7 6HULHV 'LJLWDO 3UHVVXUH *DXJH $'7 6HULHV 'LJLWDO 3UHVVXUH *DXJH SVL EDU *3 0 to 600 0 to 40 G, L Ŗ Ŗ *3. 0 to 1K 0 to 70 G, L Ŗ Ŗ Ŗ *3. 0 to 2K 0 to 140 G, L Ŗ Ŗ *3. 0 to 3K 0 to 200 G, L Ŗ Ŗ Ŗ *3. 0 to 5K 0 to 350 G, L Ŗ Ŗ Ŗ *3. 0 to 10K 0 to 700 G, L Ŗ Ŗ Ŗ *3. 0 to 15K 0 to 1K G, L Ŗ Ŗ Ŗ *3. 0 to 20K 0 to 1.4K G, L Ŗ Ŗ Ŗ *3. 0 to 25K 0 to 1.6K G, L Ŗ Ŗ Ŗ *3. 0 to 30K 0 to 2K G, L Ŗ Ŗ Ŗ *3. 0 to 36K 0 to 2.5K G, L Ŗ Ŗ Ŗ *3. 0 to 40K 0 to 2.8K G, L Ŗ Ŗ Ŗ RPSRXQG 3 ±2 ±0.16 G Ŗ Ŗ 3 ±5 ±0.35 G Ŗ Ŗ 3 ±10 ±0.7 G Ŗ Ŗ 3 ±15 ±1 G Ŗ Ŗ Ŗ 3 -15 to 30 -1 to 2 G Ŗ Ŗ Ŗ 3 -15 to 100 -1 to 7 G, L Ŗ Ŗ 3 -15 to 300 -1 to 20 G, L Ŗ Ŗ $EVROXWH $3 5 0.35 G Ŗ Ŗ $3 10 0.7 G Ŗ Ŗ $3 15 1 G Ŗ Ŗ $3 30 2 G Ŗ Ŗ $3 50 3.5 G Ŗ Ŗ $3 100 7 G, L Ŗ Ŗ $3 300 20 G, L Ŗ Ŗ $3 500 35 G, L Ŗ Ŗ $3. 1K 70 G, L Ŗ Ŗ $3. 3K 200 G, L Ŗ Ŗ $3. 5K 350 G, L Ŗ Ŗ 'LIIHUHQWLDO inH2O mbar '3 ±1 ±2.5 G Ŗ Ŗ '3 ±2 ±5 G Ŗ Ŗ '3 ±5 ±10 G Ŗ Ŗ '3 ±10 ±25 G Ŗ Ŗ '3 ±20 ±50 G Ŗ Ŗ '3 ±30 ±75 G Ŗ Ŗ '3 ±50 ±160 G Ŗ Ŗ '3 ±150 ±350 G Ŗ Ŗ '3 ±300 ±700 G Ŗ Ŗ 6HULHV
  • 67. Pressure/ProcessCalibrationEquipmentAdditelCatalog2016 Look us up on www.additel.com or call today (1)714-998-6899 $FFXUDF ADT672-02: 0.02% of full scale ADT672-05: 0.05% of full scale 20,000 psi (1,400 bar): 0.1% FS *DXJH 7SHV Gauge pressure Compound Pressure Absolute pressure Differential pressure 'LVSOD Description: Dual-line 6 full digit FSTN LCD with LED backlight Display rate: 3.5 readings per second (Default setting) Numeral display height: 16.5mm (0.65) 3UHVVXUH 8QLWV Pa, kPa, MPa, psi, bar, mbar, inH2O@4°C, mmH2O@4°C, inHg@0°C, mmHg@0°C, kgf/cm2 (QYLURQPHQWDO Compensated Temperature: 14°F to 122°F (-10°C to 50°C) Operating Temperature: 14°F to 122°F (-10°C to 50°C) Storage Temperature: -4°F to 158°F (-20°C to 70°C) Humidity: 95% 29(59,(: *DXJH SUHVVXUH 'LIIHUHQWLDO SUHVVXUH #V ſTUV INCPEG VJG UGTKGU RTGEKUKQP RTGUUWTG ECNKDTCVQTU NQQM NKMG CP QTFKPCT[ RTGUUWTG ICWIG $WV VJKU UGTKGU KU OWEJ OQTG VJCP QTFKPCT[ CPF FGſPKVGN[ OQTG VJCP LWUV C RTGUUWTG ICWIGōKVŏU C RTGUUWTG ECNKDTCVQT 9KVJ CFXCPEGF microprocessor technology and state-of-the-art silicon pressure sensors, the 672 series precision pressure calibrators provide a pressure calibration solution for gauges, transmitters, and switches over a wide pressure range. The 672 is the size of a pressure gauge but with the functionality of a calibrator: It measures pressure precisely with a built- in pressure sensor, as well as reads the current or mV produced by a transducer. It can even supply an excitation voltage to power sensors or transmitters during calibration. In order to reach 0.02%FS accuracy up to 10,000 psi (700 bar) and 0.1%FS accuracy up to 40,000 psi (2,800 bar), every silicon pressure sensor has been specially aged, tested, and screened before assembly. The 672 series precision pressure calibrators are unmatched in performance and reliability. )($785(6 Pressure ranges to 40,000 psi (2800 bar) Measure mA with 0.01% RD + 1.5 μA accuracy Measure V with 0.01% RD + 1.5 mV accuracy Power transmitters during test using 24V loop supply Pressure switch test HART Communication capability Advanced temperature compensation Dual readout Min/Max/Hold to capture changing measurements Data logging Large, easy to read display with 6-digit resolution Backlit display Rechargeable battery or AC adapter NIST traceable calibration with data (Included) 63(,),$7,216 'LJLWDO 3UHVVXUH DOLEUDWRUV $GGLWHO 3UHVVXUH UDQJHV WR SVL EDU
  • 68. +$57 RPPXQLFDWLRQ FDSDELOLW 0HDVXUH P$ RU 9 DQG ZLWK 9 ORRS SRZHU (DVWRXVH LQH[SHQVLYH SUHVVXUH FDOLEUDWRU ZLWK XQFHUWDLQW EHWWHU WKDQ )6
  • 69. Look us up on www.additel.com or call today (1)714-998-6899 *DXJH 3UHVVXUH [1] P/N Pressure Range Media[2] Accuracy(%FS) Burst Pressure(psi) (bar) V15 -15 -1.0 G 0.02 (0.05) 3× GP2 2 0.16 G 0.05 3× GP5 5 0.35 G 0.05 3× GP10 10 0.7 G 0.02 (0.05) 3× GP15 15 1.0 G, L[3] 0.02 (0.05) 3× GP30 30 2.0 G, L[3] 0.02 (0.05) 3× GP50 50 3.5 G, L 0.02 (0.05) 3× GP100 100 7.0 G, L 0.02 (0.05) 3× GP300 300 20 G, L 0.02 (0.05) 3× GP500 500 35 G, L 0.02 (0.05) 3× GP600 600 40 G, L 0.02 (0.05) 3× GP1K 1,000 70 G, L 0.02 (0.05) 3× GP2K 2,000 140 G, L 0.02 (0.05) 3× GP3K 3,000 200 G, L 0.02 (0.05) 3× GP5K 5,000 350 G, L 0.02 (0.05) 3× GP10K 10,000 700 G, L 0.02 (0.05) 3× GP15K 15,000 1,000 G, L 0.05 (0.1) 2× GP20K 20,000 1,400 G, L 0.05 (0.1) 1.5× GP25K 25,000 1,600 G, L 0.1 1.5× GP30K 30,000 2,000 G, L 0.1 1.5× GP36K 36,000 2,500 G, L 0.1 1.5× GP40K 40,000 2,800 G, L 0.1 1.1× $EVROXWH 3UHVVXUH P/N Pressure Range Media Accuracy(%FS) Burst Pressure(psi) (bar) AP5 5 0.35 G 0.1 3× AP10 10 0.7 G 0.1 3× AP15 15 1.0 G 0.1 3× AP30 30 2.0 G 0.1 3× AP50 50 3.5 G 0.1 3× AP100 100 7.0 G , L 0.05 (0.1) 3× AP300 300 20 G , L 0.05 (0.1) 3× AP500 500 35 G , L 0.05 (0.1) 3× AP1K 1,000 70 G , L 0.05 (0.1) 3× AP3K 3,000 200 G , L 0.05 (0.1) 3× AP5K 5,000 350 G , L 0.05 (0.1) 3× 'LIIHUHQWLDO 3UHVVXUH P/N Pressure Range Media Accuracy (%FS)[1] Burst Pressure Static Pressure Range(inH2O) (mbar) DP1 ±1 ±2.5 G 0.05[2] 100× ±10 psi DP2 ±2 ±5.0 G 0.05[2] 100× ±10 psi DP5 ±5 ±10 G 0.05[2] 50× ±10 psi DP10 ±10 ±25 G 0.05[2] 20× ±10 psi DP20 ±20 ±50 G 0.05 20× ±10 psi DP30 ±30 ±75 G 0.05 20× ±10 psi DP50 ±50 ±160 G 0.05 3× ±10 psi DP150 ±150 ±350 G 0.02 3× 50 psi DP300 ±300 ±700 G 0.02 3× 50 psi (OHFWULFDO 0HDVXUHPHQW $FFXUDF Range Accuracy Voltage DC ±30.0000V ± (0.01%RD + 1.5 mV) Current DC ±30.0000mA ± (0.01%RD + 1.5 μA) DC 24V 24V±0.5V max:50mA Protect at: 120mA Switch[1] Status OPEN/CLOSED RPSRXQG 3UHVVXUH P/N Pressure Range Media Accuracy(%FS)[1] Burst Pressure(psi) (bar) CP2 ±2 ±0.16 G 0.05 3× CP5 ±5 ±0.35 G 0.02 (0.05) 3× CP10 ±10 ±0.7 G 0.02 (0.05) 3× CP15 ±15 ±1 G 0.02 (0.05) 3× CP30 -15 to 30 -1 to 2 G 0.02 (0.05) 3× CP100 -15 to 100 -1 to 7 G, L 0.02 (0.05) 3× CP300 -15 to 300 -1 to 20 G, L 0.02 (0.05) 3× 3UHVVXUH 5DQJHV [1] Sealed gauge pressure for above 1000 psi [2] G=Gas, L=Liquid [3] 0.02% FS for gas media only [1] 1V~12V if switch has detective voltage 63(,),$7,216 =? (5 URGEKſECVKQP CRRNKGU VQ VJG URCP QH VJG TCPIG #EEWTCE[ KPENWFGU one year stability. [2] 0.05%FS accuracy (incl 6 months stability). One year accuracy is 0.05%FS calibration accuracy combined with 0.05%FS one year stability. =? (5 URGEKſECVKQP CRRNKGU VQ VJG URCP QH VJG TCPIG
  • 70. Pressure/ProcessCalibrationEquipmentAdditelCatalog2016 Look us up on www.additel.com or call today (1)714-998-6899 63(,),$7,216 25'(5,1* ,1)250$7,21 3UHVVXUH RQQHFWLRQV 1/4NPT male, 1/2NPT male, 1/4BSP male, 1/2BSP OCNG /: OCNG ŭ RUK 1/4HP female or 1/4HP male (15,000 psi) 1/4HP female: Autoclave F-250-C, 9/16 - 18 UNF-2B 1/4HP male: Autoclave M-250-C, 9/16 - 18 UNF-2A 0.236 inch (Ø6mm) test hose (for differential pressure) Other connections available per request (OHFWULFDO RQQHFWLRQ 0.156 inch (Ø4mm) sockets 2YHU 3UHVVXUH :DUQLQJ 120% 'DWD /RJJLQJ 5VQTCIG ECRCEKV[ ſNGU TGEQTFU RGT ſNG Mode: manual and automatic Hourly-record: record the data every hour Interval-record: set by user 0RGHO 1XPEHU 0RGHO $FFXUDF 02-0.02% of full scale 05-0.05% of full scale 10-0.1% of full scale 5DQJH WSH PSI-range by psi BAR-range by bar H2O-range by inH2O 3UHVVXUH SRUW WSH N-1/4NPT male N2-1/2NPT male B-1/4BSP male B2-1/2BSP male M-M20X1.5 male AF-Autoclave F-250-C female AM-Autoclave M-250-C male 3UHVVXUH UDQJH 31 See pressure range table 3RZHU Battery: Rechargeable Li-ion polymer battery Li-Battery working time: 40 hours Recharge time: 4 hours External power: 110V/220V power adapter (DC10V) (QFORVXUH Case material: Aluminum alloy Wetted parts: 316L SS Dimension: Ø120mm X 46mm depth X 184mm height Weight: 0.7kg (XURSHDQ RPSOLDQFH CE marked RPPXQLFDWLRQ RS232 (DB9/F, environmentally sealed) Baud rate: 1200,2400,4800,9600 Data length: 8 bits Stop bit: 2 bits Address: from 1 to 112 :DUUDQW1 year $FFHVVRULHV LQFOXGHG 110V/220V external power adapter (DC 10V) 2 pieces 1.5-meter test leads 2 pieces 0.236 inch (Ø6 mm) test hose (for differential pressure gauge only) Additel/Land software (free download at www.additel.com) Manual 0+56 VTCEGCDNG ECNKDTCVKQP EGTVKſECVG 2SWLRQDO $FFHVVRULHV 0RGHO QXPEHU 'HVFULSWLRQ 9702 Spare rechargeable Li-ion polymer battery for 672 9816 Spare 110V/220V external power adapter (DC 10V) for ADT22X and ADT672 calibrator 9502 Additel/Log II real time data logging and graphical software for 681 and 672 9506 Additel/PCal pressure calibration software 9050 USB to RS232 (DB9/M) Adapter 9050-EXT RS 232 (DB9/M) extension cable, 9 feet 9900-672 Carrying Case for one 672 digital pressure gauge 9022 Spare 2 pieces test leads (1.5-meter) for 672 $'7 *3 36, 1 'LJLWDO 3UHVVXUH DOLEUDWRUV $GGLWHO
  • 71. Look us up on www.additel.com or call today (1)714-998-6899 3HUFHQWDJH RI 5HDGLQJ $FFXUDF $SSOLFDWLRQ 1RWH 8QGHUVWDQGLQJ $FFXUDF 6SHFLILFDWLRQV IRU 'LJLWDO 3UHVVXUH 6HQVRUV ă 3HUFHQWDJH RI )XOO 6FDOH 9HUVXV 3HUFHQWDJH RI 5HDGLQJ 5RGEKſECVKQPU HQT FKIKVCN RTGUUWTG ICWIGU ECP UQOGVKOGU UGGO EQPHWUKPI QT QXGTYJGNOKPI GURGEKCNN[ KH [QW CTG unfamiliar with the terminology. Some pressure sensors will specify accuracy as a percent of full scale (FS) while QVJGTU RTQXKFG VJG URGEKſECVKQP CU C RGTEGPV QH TGCFKPI 5Q YJ[ CTG VJGTG FKHHGTGPV YC[U QH URGEKH[KPI VJG CEEWTCE[ QH pressure sensors and is percent of reading more accurate than percent of full scale or vise versa? This brief technical note will discuss the two differences and answer these questions. Accuracy as a percentage of reading is accomplished by multiplying the accuracy percentage by the pressure reading. Thus, the lower the pressure measurement, the better the accuracy. Instruments that have a percent reading specification are accompanied with a floor URGEKſECVKQP 6JG ƀQQT URGEKſECVKQP VCMGU KPVQ CEEQWPV WPEGTVCKPVKGU such as resolution and measurement noise which may be negligible at higher pressures but are of much more significance at lower pressures. For example, an accuracy specification may read 0.1% of reading for 20 to 100% of range and 0.02% of full scale below 20% of the TCPIG 6JG QH HWNN UECNG URGEKſECVKQP KU EQPUKFGTGF VJG ƀQQT URGEKſECVKQP 6Q WPFGTUVCPF VJG CEEWTCE[ QH VJG UGPUQT VJG WUGT KU VJGP TGSWKTGF VQ MPQY YJGTG VJG ƀQQT URGE KU CRRNKECDNG CPF VJG HWNN scale of the sensor. 6JKU OGVJQF QH URGEKſECVKQP KU QHVGP WUGF DGECWUG KV CNKIPU YGNN YKVJ the typical performance of pressure gauges. Typically, the closer you measure to barometric pressure the better the performance of the ICWIG (KIWTGU CPF VJG ITCRJ DGNQY UJQY CP GZCORNG URGEKſECVKQP for a 100 psi gauge and its accuracy in psi. )LJXUH 3HUFHQW UHDGLQJ DFFXUDF H[DPSOH )XOO VFDOH 0 to 100 psi $FFXUDF 20 to 100% FS: 0.1% of reading 0 to 20% FS: 0.02% of FS SVL $FFXUDF SVL
  • 72. 0 0.02 0.02%FS10 0.02 20 0.02 0.1% of Reading 30 0.03 40 0.04 50 0.05 60 0.06 70 0.07 80 0.08 90 0.09 100 0.10
  • 73. Pressure/ProcessCalibrationEquipmentAdditelCatalog2016 Look us up on www.additel.com or call today (1)714-998-6899 3HUFHQWDJH RI )XOO 6FDOH $FFXUDF $ RPSDULVRQ RI 3HUFHQW RI )XOO 6FDOH DQG 3HUFHQW RI 5HDGLQJ $FFXUDFLHV Accuracy as a percentage of full scale is calculated by multiplying the accuracy percentage by the full scale pressure of the gauge. This is QDXKQWUN[ C OQTG UKORNG OGVJQF QH URGEKſECVKQP CPF KU OQUV EQOOQPN[ used in industry because it is easy to calculate and interpret. Denoting the accuracy as percent full scale is a more conservative way of specifying the pressure sensor because typically the sensor doesn't perform the same over its full range. It usually will perform more accurately as you CRRTQCEJ DCTQOGVTKE RTGUUWTG 6JKU V[RG QH URGEKſECVKQP KU OQUV EQOOQP for industrial gauges which make it easier to compare one gauge versus CPQVJGT (KIWTG KU CP GZCORNG URGEKſECVKQP HQT C RUK ICWIG CPF KVU accuracy in psi. SVL $FFXUDF SVL
  • 74. 0 0.05 0.05%FS 10 0.05 20 0.05 30 0.05 40 0.05 50 0.05 60 0.05 70 0.05 80 0.05 90 0.05 100 0.05 SVL $FFXUDF SVL
  • 75. RI 5HDGLQJ RI )6 0 0.02 0.05 10 0.02 0.05 20 0.02 0.05 30 0.03 0.05 40 0.04 0.05 50 0.05 0.05 60 0.06 0.05 70 0.07 0.05 80 0.08 0.05 90 0.09 0.05 100 0.10 0.05 So you may ask, Which is more accurate? The answer is that it depends on the RTGUUWTG DGKPI OGCUWTGF +P VJG VYQ GZCORNGU IKXGP VJG ICWIG URGEKſGF CV of reading is more accurate as you measure lower pressures in its range. However, CU [QW OQXG CDQXG QH VJG TCPIG VJG ICWIG URGEKſGF CV QH HWNN UECNG becomes more accurate than the 0.1% of reading gauge. This can be seen clearly in the chart (left) and graph (below) where the two gauges are compared in terms of psi accuracy. To properly compare these, two gauges you should convert the accuracy to pressure units, such as psi or bar. Then they can be properly matched one against another in like units of measure. +P EQPENWUKQP QPG OGVJQF QH URGEKſECVKQP KU PQV DGVVGT VJCP CPQVJGT KV KU LWUV different. Given this difference it becomes important to know how to interpret the FKHHGTGPV URGEKſECVKQPU V[RGU CPF DG CDNG VQ EQORCTG QPG XGTUWU CPQVJGT
  • 76. Look us up on www.additel.com or call today (1)714-998-6899 29(59,(: *DXJH SUHVVXUH 'LIIHUHQWLDO SUHVVXUH With advanced microprocessor technology and state-of-the-art silicon pressure sensors, the 681 series digital pressure gauges provide an accurate, reliable, and economic solution for a wide range of pressure applications. They are loaded with functionality and remarkably easy to use. To reach the best performance, every silicon pressure sensor in our gauges is specially aged, tested and screened before assembly. At Additel, fully temperature compensated accuracy means every sensor is pressure tested at several environmental temperatures from u% VQ u% 9KVJ VJKU VGUV FCVC KPFKXKFWCN EQGHſEKGPVU CTG IGPGTCVGF CPF UVQTGF KP VJG ICWIG EJCTCEVGTKKPI KVU performance over the full temperature compensated range. And now the ADT681IS comes with an IP67 rating meaning it is dust resistant and water proof, submersible in 1 meter of water. 'HVLJQHG WR ILW RXU QHHG Additel pressure gauges give you the widest variety of sensor choices on the market. Whether you require low inches of water measurement or very high pressure measurement, we have a gauge that will meet your need. We offer sensors which are ±1 inH2O (±2.5 mbar) to 40K psi (2,800 bar) and everything in between. Do your applications require you to measure both positive and vacuum pressure? Our compound gauges do not EQORTQOKUG CEEWTCE[ CPF RTQXKFGU [QW YKVJ VJG UCOG JKIJ CEEWTCE[ URGEKſECVKQP QP DQVJ RQUKVKXG CPF XCEWWO pressures. We offer a wide variety up to 300 psi (20 bar). If you need a higher range, just contact us and we can likely customize one to meet your need. We also offer absolute pressure sensors to 5K psi (350 bar) and a full range of differential pressure sensors from ±1 inH2O (±2.5 mbar) to ±300 inH2O (±700 mbar). Are you looking for a RTGUUWTG ICWIG VQ WUG KP JCCTFQWU CTGCU! 1WT #6':EGTVKſGF KPVTKPUKECNN[ UCHG OQFGNU +5 CTG FGUKIPGF HQT pressure measurement in hazardous areas. If you need to panel mount our sensors, we offer the option (see ordering information) for a back-mounted pressure RQTV CPF ICWIG JQWUKPI FGUKIPGF VQ ſV KP C RCPGN #PF OQUV TGEGPVN[ YG XG CFFGF VJG QRVKQP VQ FQ UVCPFCNQPG FCVC logging with the 681. Now you can record more than 21,000 records internal to the 681 series. Each record includes date, time, pressure and temperature readings. Download the logged data with our free Additel/Land software or you can purchase our Additel/Log II for real-time logging and data analysis. The 681 series digital pressure gauges are unmatched in performance and reliability. Best of all, they are very affordable. 3UHVVXUH UDQJHV WR SVL EDU
  • 77. )6 RU 5' DFFXUDF SUHVVXUH LQGLFDWLRQ ZLWK IDQVKDSHG JUDSK VFDOH IRU YLVXDO UHIHUHQFH )XOO WHPSHUDWXUH FRPSHQVDWHG DFFXUDF 3DQHO PRXQW JDXJHV DUH DYDLODEOH ,QWULQVLFDOO VDIH YHUVLRQ ,6
  • 79. 'LJLWDO 3UHVVXUH *DXJHV $GGLWHO )($785(6 Pressure ranges to 40,000 psi (2,800 bar) 0.02% full scale accuracy (681-02) 0.05% full scale accuracy (681-05) 0.1% full scale accuracy (681-10) 0.2% full scale accuracy (681-20) 0.1% reading scale accuracy (681-RD) IP67 rated: Submersible in 1 meter of water(681IS) Fully temperature compensated accuracy from 14°F to 122°F (-10°C to 50°C) Up to eleven selectable pressure units Large, easy to read display with 5-digit resolution Backlit display % pressure indication with fan-shaped graph scale for visual reference KURNC[ ƀCUJ YCTPKPI YJGP RTGUUWTG QXGT QH (5 Bottom mount or panel mount #6': EGTVKſGF KPVTKPUKECNN[ UCHG /QFGN +5 NIST traceable calibration with data(included) 9V battery power or AC adapter (optional) Data logging to 21,843 records (includes date, time, pressure and temperature) Now IP67 Rated!
  • 80. Pressure/ProcessCalibrationEquipmentAdditelCatalog2016 Look us up on www.additel.com or call today (1)714-998-6899 *DXJH 3UHVVXUH @ P/N Pressure Range Media [2] Accuracy Burst Pressure(psi) (bar) %FS %RD V15 -15 -1.0 G 0.02 (0.05, 0.1, 0.2) N/A 3× GP2 2 0.16 G 0.05 (0.1, 0.2) N/A 3× GP5 5 0.35 G 0.05 (0.1, 0.2) 0.1 3× GP10 10 0.7 G 0.02 (0.05, 0.1, 0.2) 0.1 3× GP15 15 1.0 G, L[3] 0.02 (0.05, 0.1, 0.2) 0.1 3× GP30 30 2.0 G, L[3] 0.02 (0.05, 0.1, 0.2) 0.1 3× GP50 50 3.5 G, L 0.02 (0.05, 0.1, 0.2) 0.1 3× GP100 100 7.0 G, L 0.02 (0.05, 0.1, 0.2) 0.1 3× GP300 300 20 G, L 0.02 (0.05, 0.1, 0.2) 0.1 3× GP500 500 35 G, L 0.02 (0.05, 0.1, 0.2) 0.1 3× GP600 600 40 G, L 0.02 (0.05, 0.1, 0.2) 0.1 3× GP1K 1,000 70 G, L 0.02 (0.05, 0.1, 0.2) 0.1 3× GP2K 2,000 140 G, L 0.02 (0.05, 0.1, 0.2) 0.1 3× GP3K 3,000 200 G, L 0.02 (0.05, 0.1, 0.2) 0.1 3× GP5K 5,000 350 G, L 0.02 (0.05, 0.1, 0.2) 0.1 3× GP10K 10,000 700 G, L 0.02 (0.05, 0.1, 0.2) 0.1 3× GP15K 15,000 1,000 G, L 0.05 (0.1,0.2) 0.1 2× GP20K 20,000 1,400 G, L 0.05 (0.1,0.2) N/A 1.5× GP25K 25,000 1,600 G, L 0.1 (0.2) N/A 1.5× GP30K 30,000 2,000 G, L 0.1 (0.2) N/A 1.5× GP36K 36,000 2,500 G, L 0.1 (0.2) N/A 1.5× GP40K 40,000 2,800 G, L 0.1 (0.2) N/A 1.1× 3UHVVXUH 5DQJHV [1] Sealed gauge pressure for above 1,000 psi [2] G=Gas, L=Liquid [3] 0.02% FS for gas media only 63(,),$7,216 0RGHO ADT681: Regular gauge #6+5 #6': EGTVKſGF KPVTKPUKECNN[ UCHG $FFXUDF 681(IS)-02 : 0.02% of full scale 681(IS)-05 : 0.05% of full scale 681(IS)-10 : 0.1% of full scale 681(IS)-20 : 0.2% of full scale 681(IS)-RD: 0% to 20% of Range: 0.02% of full scale 20% to 110% of Range: 0.1% of reading Vacuum: 0.25% of full scale[1][2] [1] FS = -14.5 psi [2] Applicable ADT681-RD-CPX (For detailed accuracy, please see pressure range table) *DXJH7SHV Gauge pressure Compound pressure Absolute pressure Differential pressure )DQVKDSHG *UDSK 6FDOH Similar to analog dials, including pressure swing, % indication with fan-shaped graph scale for visual reference, low/high alarm. 'LVSOD Description: 5 full digit FSTN LCD Display rate: 3 readings per second (Default setting). Adjustable from 10 readings per second to 1 reading every ten seconds Numeral display height: 16.5mm (0.65) 3UHVVXUH 8QLWV Pa, kPa, MPa, psi, bar, mbar, kgf/cm2 , inH2O@4°C mmH2O@4°C, inHg@0°C, mmHg@0°C (QYLURQPHQWDO Compensated Temperature: 14°F to 122°F (-10°C to 50°C) Operating Temperature: 14°F to 122°F (-10°C to 50°C) Storage Temperature: -4°F to 158°F (-20°C to 70°C) Humidity: 95% 3UHVVXUH 3RUW 1/4NPT male, 1/2NPT male, 1/4BSP male, 1/2BSP OCNG /ª OCNG ŭ RUK 1/4HP female or 1/4HP male( 15,000 PSI) *1/4HP female: Autoclave F-250-C, 9/16 - 18 UNF-2B *1/4HP male: Autoclave M-250-C, 9/16 - 18 UNF-2A 0.236 inch (Ø6 mm) test hose (for differential pressure) Other connections available per request 3RZHU Battery: One 9V alkaline battery (included) Battery life: 300 hours (10 readings/s), 600 hours (3 reading/s), or 6000 hours (1 reading/10s) Power auto-off: 60 minutes power auto-off. Auto-off may be disable External power: 110/220V external power adapter (optional) (QFORVXUH Case material: Aluminum alloy Wetted parts: 316L SS Dimension: Ø110mm X 35mm depth X 176mm height (panel mount gauge: Ø140mm X 86mm depth) Weight: 0.6kg RPSOLDQFH Protection Level: IP67(available for 681IS GP15-40K) Vibration: 5g (20-2000 Hz) Shock Resistance: 100g/11ms 'DWD /RJJLQJ Data storage: 21,843 records (each record includes date, time, pressure and temperature) Rate: user-selectable from 1 to 99,999 second intervals ,QWULQVLF 6DIHW DQG (XURSHDQ RPSOLDQFH CE marked #6': EGTVKſGF KPVTKPUKECNN[ UCHG #6+5 II 1G EX ia IIC T4 Ga RPPXQLFDWLRQ RS232 *(Do not use the RS-232 connector in a hazardous atmosphere) :DUUDQW 1 year
  • 81. Look us up on www.additel.com or call today (1)714-998-6899 $EVROXWH 3UHVVXUH P/N Pressure Range Media [1] Accuracy(%FS) Burst Pressure(psi) (bar) AP5 5 0.35 G 0.1 (0.2) 3× AP10 10 0.7 G 0.1 (0.2) 3× AP15 15 1.0 G 0.1 (0.2) 3× AP30 30 2.0 G 0.1 (0.2) 3× AP50 50 3.5 G 0.1 (0.2) 3× AP100 100 7.0 G, L 0.05 (0.1, 0.2) 3× AP300 300 20 G, L 0.05 (0.1, 0.2) 3× AP500 500 35 G, L 0.05 (0.1, 0.2) 3× AP1K 1,000 70 G, L 0.05 (0.1, 0.2) 3× AP3K 3,000 200 G, L 0.05 (0.1, 0.2) 3× AP5K 5,000 350 G, L 0.05 (0.1, 0.2) 3× 'LIIHUHQWLDO 3UHVVXUH P/N Pressure Range Media Accuracy (%FS)[1] Burst Pressure Static Pressure Range(inH2O) (mbar) DP1 ±1 ±2.5 G 0.05[2] 100× ±10 psi DP2 ±2 ±5.0 G 0.05[2] 100× ±10 psi DP5 ±5 ±10 G 0.05[2] 50× ±10 psi DP10 ±10 ±25 G 0.05[2] 20× ±10 psi DP20 ±20 ±50 G 0.05 20× ±10 psi DP30 ±30 ±75 G 0.05 20× ±10 psi DP50 ±50 ±160 G 0.05 3× ±10 psi DP150 ±150 ±350 G 0.02 (0.05) 3× 50 psi DP300 ±300 ±700 G 0.02 (0.05) 3× 50 psi RPSRXQG 3UHVVXUH@ P/N Pressure Range Media [2] Accuracy Burst Pressure (psi) (bar) %FS[3] %RD CP2 ±2 ±0.16 G 0.05 (0.1,0.2) N/A 3× CP5 ±5 ±0.35 G 0.02 (0.05,0.1,0.2) 0.1 3× CP10 ±10 ±0.7 G 0.02 (0.05,0.1,0.2) 0.1 3× CP15 ±15 ±1 G 0.02 (0.05,0.1,0.2) 0.1 3× CP30 -15 to 30 -1 to 2 G 0.02 (0.05,0.1,0.2) 0.1 3× CP100 -15 to 100 -1 to 7 G, L 0.02 (0.05,0.1,0.2) 0.1 3× CP300 -15 to 300 -1 to 20 G, L 0.02 (0.05,0.1,0.2) 0.1 3× [1] Sealed gauge pressure for above 1,000 psi. [2] G=Gas, L=Liquid =? (5 URGEKſECVKQP CRRNKGU VQ VJG URCP QH VJG TCPIG Note: For oil-free applications contact Additel. $FFHVVRULHV LQFOXGHG Rubber boot (Except panel mount) 9V alkaline battery (1 pc) Manual 0+56 VTCEGCDNG ECNKDTCVKQP EGTVKſECVG 2SWLRQDO $FFHVVRULHV 0RGHO QXPEHU 'HVFULSWLRQ 9812 110V/220V external power adapter (DC 9V) for 681 digital pressure gauge. 9502 Additel/Log II real time data logging and graphical software for 681 and 672. 9506 Additel/PCal pressure calibration software 9050 USB to RS232 (DB9/M) Adapter 9050-EXT RS 232 (DB9/M) extension cable, 9 feet 9900-681 Carrying Case for one 681 digital pressure gauge 9251 Rubber boot for ADT681 25'(5,1* ,1)250$7,21 0RGHO 1XPEHU 0RGHO ADT681 ADT681IS $FFXUDF 02-0.02% of full scale 05-0.05% of full scale 10-0.1% of full scale 20-0.2% of full scale RD-0.1% of reading 5DQJH WSH PSI-range by psi BAR-range by bar H2O-range by inH2O 3UHVVXUH SRUW WSH N- 1/4NPT male N2- 1/2NPT male B-1/4BSP male B2-1/2BSP male M-M20X1.5 male AF-Autoclave F-250-C female AM-Autoclave M-250-C male '/ Data logging 3UHVVXUH UDQJH 31 See pressure range table 3% Panel mount with back pressure port $'7 *3 36, 3% '/ 1 $GGLWHO 'LJLWDO 3UHVVXUH *DXJHV =? (5 URGEKſECVKQP CRRNKGU VQ VJG URCP QH VJG TCPIG #EEWTCE[ KPENWFGU one year stability. [2] 0.05%FS accuracy (incl 6 months stability). One year accuracy is 0.05%FS calibration accuracy combined with 0.05%FS one year stability. [1] G=Gas, L=Liquid
  • 82. Pressure/ProcessCalibrationEquipmentAdditelCatalog2016 Look us up on www.additel.com or call today (1)714-998-6899 3HUIRUPDQFH Accuracy: 0.05%, 0.1% or 0.25%FS Update Rate: 10 times/Sec ,3 times /Sec (default), 1 time /Sec ,1 time/15 Seconds Operating Temperature: 14°F to 122°F (-10°C to 50°C) Compensated Temperature: 14°F to 122°F (-10°C to 50°C), accuracy guaranteed Storage Temperature: -4°F to 158°F (-20°C to 70°C) Overload Pressure: 1.2X 3RZHU 6XSSO Battery: 2 x AA alkaline (LR6) batteries *Recommend: Duracell MX1500, EnergizerE91, Panasonic LR6, or Duracell 15AU Battery Life: 10 rdg/sec: 1500 hrs; 3 rdg/sec: 3000 hrs 1 rdg/sec: 6000 hrs; 1 rdg/15 sec: 12000 hrs. Rated Power: 3.0VDC Imax=30mA Pmax=90mW 0DWHULDO Wetted Surface: All wetted surfaces are 316L Stainless 5VGGN YKVJ YGNFGF FGUKIP HQT ICWIGU ŭ RUK PQ O-ring, thread tape, epoxy, or sealant on any part of the sensor assembly), and copper O-ring design for gauges 10,000 psi. Case: PC+ABS Protective Cover: Silicone Rubber 29(59,(: We designed the 680 series digital pressure gauges with two main objectives in mind. First, to provide an affordable FKIKVCN ICWIG VQ TGRNCEG OGEJCPKECN ICWIGU +H [QWŏTG NQQMKPI VQ OQXG HTQO FKCN ICWIGU VQ C FKIKVCN ICWIG [QWŏNN ſPF the 680 standard version gauge to be of high quality and suited for your need in terms of price and performance. With advanced microprocessor technology and state-of-the-art silicon pressure sensors, the 680 series digital pressure gauges provide an accurate, reliable, and economic solution for a wide range of pressure applications. They are loaded with functionality and remarkably easy to use. To reach the best performance, every silicon pressure sensor is specially aged, tested and screened before assembly. The second objective was to provide a high-precision pressure gauge capable of wireless communication and data logging. Our 680W series provides just this along with several accuracy and pressure range options to meet your need. This wireless unit is compatible with the Additel/Land Wireless software, which is available for a free download from our website. Data can be recorded standalone with the 680W and then downloaded wirelessly to Additel/Land Wireless. For more advanced logging and data analysis, Additel/Log II Wireless is specially designed to communicate with the 680W. Each unit can store up to 140,000 readings which consist of date, time, pressure, and internal temperature. The 680 series digital pressure gauges are unmatched in performance and reliability. Best of all they are very affordable. )($785(6 Pressure ranges to 40,000 psi (2,800 bar) 0.05%, 0.1% or 0.25% full scale accuracy Fully temperature compensated accuracy from 14°F to 122°F (-10°C to 50°C) Up to 13 selectable pressure units, 6 customized FGſPKVKQP GPIKPGGTKPI WPKVU Large, easy to read display with 5-digit resolution Backlit display Icon-based menu KURNC[ ƀCUJ YCTPKPI YJGP RTGUUWTG GZEGGFU 120% of FS Stainless wetted surface construction IP67 (submersible in 1 meter of water) Drop-tested from 1 meter 2 AA alkaline batteries %' 4 66' (%% + +% + %GTVKſECVGU NIST traceable calibration with data(included) 63(,),$7,216 'LJLWDO 3UHVVXUH *DXJHV $GGLWHO 6HULHV 3UHVVXUH UDQJHV WR SVL EDU
  • 83. RU )6 DFFXUDF )XOO WHPSHUDWXUH FRPSHQVDWHG DFFXUDF ,3 'DWD ORJJLQJ DQG ZLUHOHVV :
  • 84. ZLWK GDWD ORJJLQJ DQG ZLUHOHVV RSWLRQDO
  • 85. 0RGHOV ADT680: Regular version ADT680W: Data logging and wireless version Pressure Type: Gauge and Compound
  • 86. Look us up on www.additel.com or call today (1)714-998-6899 *DXJH 3UHVVXUH P/N Pressure Range Accuracy(FS%) Media[3] Burst Pressure(psi)[1] (bar) V15 -15 -1.0 0.05 (0.1, 0.25) G 3× GP15 15 1.0 0.05 (0.1, 0.25) G,L 3× GP30 30 2.0 0.05 (0.1, 0.25) G,L 3× GP100 100 7.0 0.05 (0.1, 0.25) G,L 3× GP300 300 20 0.05 (0.1, 0.25) G,L 3× GP500 500 35 0.05 (0.1, 0.25) G,L 3× GP1K 1,000 70 0.05 (0.1, 0.25) G,L 3× GP3K 3,000 200 0.05 (0.1, 0.25) G,L 3× GP5K 5,000 350 0.05 (0.1, 0.25) G,L 3× GP10K 10,000 700 0.05 (0.1, 0.25) G,L 3× GP15K 15,000 1,000 0.05 (0.1, 0.25) G,L 2× GP20K 20,000 1,400 0.05 (0.1, 0.25) G,L 1.5× GP25K 25,000 1,600 0.1 (0.25) G,L 1.5× GP30K 30,000 2,000 0.1 (0.25) G,L 1.5× GP36K 36,000 2,500 0.1 (0.25) G,L 1.5× GP40K 40,000 2,800 0.1 (0.25) G,L 1.1× [1] Sealed gauge pressure for above 1,000 psi. =? (5 URGEKſECVKQP CRRNKGU VQ VJG URCP QH VJG TCPIG [3] G=Gas, L=Liquid 'LVSOD .% 5RGEKſECVKQP (560.%8KUWCN UEQRG ZOO Full 5 digits,15.2mm High; 7 segment analog bar graph scaled from 0-100% of FS Backlight: White :LUHOHVV RPPXQLFDWLRQ $'7: RQO
  • 87. Wireless Frequency: 2.4G ISM Bands, 20 meter range Number of wireless Channels : Chanel 1-15 Software: Wireless network demo software included Read upto 20 gauges. 'DWD /RJJLQJ $'7: RQO
  • 88. Storage Capacity: 140,000 readings(time, pressure, and temperature) Storage Interval: Adjustable from 1-9999 Sec Single-button-press data logging enabled RPSOLDQFHV %GTVKſECVGU%' 4 66' (%% + +% + Protection Level: IP67 Vibration: 5g(20-2000Hz) Shock resistance: 100g/11ms 2WKHUV Dimensions: 100mm x 40mm, total height:157mm Weight: 500g Fitting: 1/4NPT male, 1/2NPT male, 1/4BSP male, $52 OCNG /ª OCNG ŭ 25+ 1/4HP female or 1/4HP male( 15,000 PSI) *1/4HP female: Autoclave F-250-C, 9/16 - 18 UNF-2B *1/4HP male: Autoclave M-250-C, 9/16 - 18 UNF-2A Other connections available per request 63(,),$7,216 0RGHO 1XPEHU 0RGHO ADT680 ADT680W $FFXUDF 05-0.05% of full scale 10-0.1% of full scale 25-0.25% of full scale 5DQJH WSH PSI-range by psi BAR-range by bar 3UHVVXUH SRUW WSH N-1/4NPT male N2-1/2NPT male B-1/4BSP male B2-1/2BSP male M-M20X1.5 male AF-Autoclave F-250-C female AM-Autoclave M-250-C male 3UHVVXUH UDQJH 31 See pressure range table $'7 *3 36, 1 $XWR RII: Disabled,15, 30, 45, 60, 90, or 120 Minutes. )LOWHULQJ #XGTCIKPI VQ UCORNGU QT NQYRCUU ſTUV QTFGT ſNVGT (QJLQHHULQJ XQLWV: Pa, kPa, MPa, bar, mba, psi, kgf/ cm2 , mmH2O, mmHg, inH2O, inHg, ozf/in, %, °C, °F. extend units: inH2O(20°C), inH2O(60°F), mmH2O(20°C), mmH2O(15°C), ftH2O(60°F), or ftH2O(4°C) %DFNOLJKW 'XUDWLRQ: Not auto off,15,30,45,60 seconds optional. 0D[0LQ GDWD FDSWXUH: Saves Max and Min data during pressure measurement. .H /RFNRXW: When the gauge is in auto-storing mode, the front panel buttons will be automatically locked. 2YHUSUHVVXUH $ODUP KURNC[ YKNN ƀCUJ QXGT (5 %DWWHU YROWDJH ,QGLFDWRU: Displays the battery life remaining. When the battery voltage is too low, the gauge will power-off automatically. 2YHUSUHVVXUH 5HFRUG: Gauge will record max pressure data when the gauge is over pressured by 120% of FS. /HDNDJH WHVW: In leak test mode, the gauge will record beginning RTGUUWTG GPFKPI RTGUUWTG CPF UJQY VJG FKHHGTGPEG ţ2 )DFWRU 5HVHW: Resets all settings back to factory default, except the calibration parameters. :DUUDQW: 1 year 25'(5,1* ,1)250$7,21 2SWLRQDO $FFHVVRULHV $FFHVVRULHV LQFOXGHG ## DCVVGT[ REU /CPWCN 0+56 VTCEGCDNG ECNKDTCVKQP EGTVKſECVG Rubber boot for Additel 680 gauge; Additel/Land Wireless software for 680W (free download at www. additel.com) 0RGHO QXPEHU 'HVFULSWLRQ 9503 Additel/Log II Wireless real time data logging and graphical software for 680W 9030 Spare wireless master device (USB dongle) for ADT680W gauge. RPSRXQG 3UHVVXUH@ P/N Pressure Range Accuracy(FS%)[2] Media[3] Burst Pressure(psi) (bar) CP15 ±15 ±1 0.05(0.1, 0.25) G,L 3× CP30 -15 to 30 -1 to 2 0.05(0.1, 0.25) G,L 3× Note: For oil-free applications contact Additel.
  • 89. Pressure/ProcessCalibrationEquipmentAdditelCatalog2016 Look us up on www.additel.com or call today (1)714-998-6899 :K7HPSHUDWXUH RPSHQVDWLRQ 5HDOO 0DWWHUV IRU 3UHVVXUH 0HDVXUHPHQW Have you ever wondered how much impact environmental temperature has on your pressure sensors? Nearly every pressure sensor has some sort of environmental temperature specification on its data sheet. This technical note explains the environmental temperature effects on pressure sensors, quantifying the impact, and ways to minimize the impact. Much like anything else in the physical measurement world, pressure sensors are subject to changes in environmental conditions. Temperature effects tend to have the largest impact on pressure measurement accuracy. Temperature GHHGEVU FKTGEVN[ KPƀWGPEG VJG RTGUUWTG UGPUQT CPF VJG EKTEWKVT[ WUGF VQ OGCUWTG VJG UGPUQT KIKVCN RTGUUWTG UGPUQTU use electronic circuits which provide an analog output proportional to the inlet pressure. There are three factors of a sensor's circuitry that are affected by environmental temperature changes: zero pressure output voltage, pressure sensitivity span and bridge resistance. Temperature-compensated sensors employee some techniques to correct for and minimize the impact of temperature changes on these factors. 6Q WPFGTUVCPF VJG GPXKTQPOGPVCN VGORGTCVWTG GHHGEV QP [QWT UGPUQT KV KU JGNRHWN VQ ſTUV WPFGTUVCPF UQOG EQOOQP VGTOU [QW OC[ UGG QP C RTGUUWTG UGPUQT URGEKſECVKQP UJGGV 2SHUDWLQJ 7HPSHUDWXUH 5DQJH This is the temperature range over which the sensor can be used without causing damage. 7HPSHUDWXUH RPSHQVDWHG $FFXUDF 5DQJH This refers to the environmental temperature range over which the accuracy of the sensor is applicable. 7HPSHUDWXUH RHIILFLHQW An additional error that needs to be considered when used outside of the temperature compensated accuracy range. Many sensors are only tested and calibrated at laboratory temperatures. In this case, VJG VGORGTCVWTG EQGHſEKGPV YKNN PGGF VQ DG EQPUKFGTGF KP VJG OGCUWTGOGPV CEEWTCE[ YJGP WUKPI VJG UGPUQT QWVUKFG QH laboratory temperatures. So how much will the ambient temperature impact your measurement accuracy? Well, this will depend on the temperature compensated accuracy range and the temperature coefficient. To demonstrate this, let's consider VJTGG FKHHGTGPV ICWIGU #U [QW ECP UGG HTQO VJG URGEKſECVKQPU DGNQY ſIWTG VJG[ CNN JCXG VJG UCOG CEEWTCE[ specification of 0.05% FS. However, as you consider the temperature compensated accuracy range and the VGORGTCVWTG EQGHſEKGPV [QW NN UGG C HCKTN[ NCTIG XCTKCVKQP DGVYGGP VJG VJTGG ICWIGU The graph below shows the total specified accuracy when considering the temperature effects on the pressure gauges. As you can see in one case here, the lack of temperature compensation and inclusion of the temperature EQGHſEKGPV URGEKſECVKQP OQTG VJCP VTKRNG VJG (5 CEEWTCE[ URGEKſECVKQP :K SUHVVXUH VHQVRUV DUH LPSDFWHG E HQYLURQPHQWDO WHPSHUDWXUH FKDQJHV 4XDQWLILQJ WKH HQYLURQPHQWDO WHPSHUDWXUH HIIHFW )LJXUH 0DQXIDFWXUHU 0DQXIDFWXUHU 0DQXIDFWXUHU $FFXUDF 0.05% FS 0.05% FS 0.05% FS 7HPSHUDWXUH RPSHQVDWHG $FFXUDF 5DQJH N/A 15°C to 35°C -10°C to 50°C 7HPSHUDWXUH RHIILFLHQW Add 0.003% FS/°C from 23°C Add 0.003% FS/°C: -10°C to 15°C, 35°C to 50°C N/A $SSOLFDWLRQ 1RWH
  • 90. Look us up on www.additel.com or call today (1)714-998-6899 To further show temperature compensation has real effect, we placed a non-temperature compensated pressure gauge in a temperature chamber and pressure tested it from 0 to 580 psi (0 to 40 bar) and over the environmental temperature range of 0°C to 40°C. We then performed the same test on a temperature compensated gauge. As you may expect—the higher the pressure, the larger the impact from the environmental temperature. Below is a chart comparing the non-temperature compensated gauge with the temperature compensated gauge. The temperature effect on a pressure sensor will be negligible when used at the same laboratory temperature in which it was calibrated. This, however, is often not practical for your measurements. With sensor technology advances, we have found a variety of ways to minimize the temperature effect on pressure sensors and with confidence define a large temperature compensated accuracy range. First, regularly zero your digital pressure gauges. By zeroing the pressure gauge, you are aligning the zero pressure output voltage to the current environmental conditions. You should only zero the pressure gauge when you do not have any inlet pressure on the gauge. Because each sensor is unique and performs differently due to environmental temperature changes, at Additel, we pressure test every sensor in a thermal chamber at different temperatures so we understand its pressure performance relative to environmental changes. Each sensor contains a temperature-compensated circuit which we NQCF EQGHſEKGPVU TGRTGUGPVKPI VJG VGORGTCVWTG VGUVKPI QH VJG ICWIG 6JKU CNNQYU HQT [QW VQ EQPſFGPVN[ WUG QWT UGPUQTU QXGT VJG TCPIG u% VQ u% YKVJQWV JCXKPI VQ CFF C VGORGTCVWTG EQGHſEKGPV GTTQT VQ VJG CEEWTCE[ 7HPSHUDWXUH FRPSHQVDWLRQ WHVW UHVXOWV 0LQLPL]LQJ HQYLURQPHQWDO WHPSHUDWXUH HUURU
  • 91. Pressure/ProcessCalibrationEquipmentAdditelCatalog2016 Look us up on www.additel.com or call today (1)714-998-6899 0RGHOV $GGLWHO /RRS DOLEUDWRU $GGLWHO /RRS DOLEUDWRU $GGLWHO 0XOWLIXQFWLRQ /RRS DOLEUDWRU $GGLWHO $ 0XOWLIXQFWLRQ 7HPSHUDWXUH DOLEUDWRU $GGLWHO $ 0XOWLIXQFWLRQ 3URFHVV DOLEUDWRU $GGLWHO $ 'RFXPHQWLQJ 3URFHVV DOLEUDWRU 0HDVXUH 9ROWDJHP9
  • 92. ±300 mV ±75 mV ±75 mV ±75 mV 9ROWDJH9
  • 93. 0 to 30 V 0 to 30 V ±60 V ±30 V ±30 V ±30 V XUUHQW P$
  • 94. 0 to 24 mA 0 to 24 mA ±30 mA ±30 mA ±30 mA ±30 mA 5HVLVWDQFHRKP
  • 95. 0 to 2,000 š 0 to 4,000 š 0 to 4,000 š 0 to 4,000 š )UHTXHQF+]
  • 96. 1 to 50,000 Hz 1 to 50,000 Hz 1 to 50,000 Hz 3XOVH 0 to 999,999 0 to 999,999 0 to 999,999 /LPLW 6ZLWFK 3 to 24V 3 to 24V 3 to 24V 3 to 24V 3UHVVXUH Ŗ[1] Ŗ[1] 6RXUFH 6LPXODWH 9ROWDJHP9
  • 97. -10 to 200 mV -10 to 75 mV -10 to 75 mV -10 to 75 mV 9ROWDJH9
  • 98. 0 to 12 V 0 to 12 V 0 to 12 V 0 to 12 V XUUHQW P$
  • 99. 0 to 24 mA 0 to 24 mA 0 to 22 mA 0 to 22 mA 0 to 22 mA 0 to 22 mA 5HVLVWDQFHRKP
  • 100. 1 to 4,000 š 1 to 4,000 š 1 to 4,000 š )UHTXHQF+]
  • 101. 0 to 50,000 Hz 0 to 50,000 Hz 0 to 50,000 Hz 3XOVH 0 to 999,999 0 to 999,999 0 to 999,999 ' 9 Ŗ Ŗ Ŗ Ŗ 5HFRUG 6FDOLQJ Ŗ Ŗ Ŗ Ŗ 0LQ0D[$YJ7DUH Ŗ +ROG Ŗ $V IRXQG $V OHIW Ŗ Ŗ Ŗ Ŗ 2QGHPDQG ORJJLQJ Ŗ Ŗ Ŗ Ŗ 8SORDG GDWD WR 3 Ŗ Ŗ Ŗ )HDWXUHV 9 ORRS VXSSO Ŗ Ŗ Ŗ Ŗ 5DPS 6WHS Ŗ Ŗ Ŗ Ŗ 6LPXODWH7UDQVPLWWHU Ŗ Ŗ Ŗ 57' 11 types[2] 11 types[2] 11 types[2] 7KHUPRFRXSOH 13 types[3] 13 types[3] 13 types[3] +$57 FRPPXQLFDWLRQ Ŗ[4] Ŗ 'RFXPHQWLQJ Ŗ Ŗ Ŗ :DUUDQW 3 years 3 years 3 years 3 years 3 years 3 years 1,67 WUDFHDEOH FDOLEUDWLRQ Ŗ Ŗ Ŗ Ŗ Ŗ Ŗ 6HH 3DJH P35 P35 P37 P39 P39 P39 6HOHFWLRQ *XLGH 0XOWLIXQFWLRQ 3URFHVV DOLEUDWRUV [1] Additel 160A External Pressure Module required [2] RTD Includes Pt10(385), Pt100(385), Pt100(3916), Pt100(3926), Pt500(385), Pt1000(385), Cu10(427), Cu50(428), Cu100(428), Ni100(617), Ni120(672),Ni1000 and customized [3] Thermocouple Includes S, R, B, K, N, E, J, T, C, D, G, L, U =? $WKNVKP š TGUKUVQT HQT *#46 EQOOWPKECVKQPU
  • 102. Look us up on www.additel.com or call today (1)714-998-6899 $GGLWHO DQG 6HULHV /RRS DOLEUDWRU 29(59,(: The new Additel 209 and 210 loop calibrator series combine ease of use and functionality, making them the ideal tools to troubleshoot your process loop. The ADT209 has an accuracy of 0.03% of reading whereas the ADT210 holds an accuracy of 0.01% of reading. If you want to source, simulate or simply measure, the Additel Loop calibrator series will fit your need. The ADT209 and ADT210 allow for measurement of current, voltage and a switch. You can also simulate or source mA or a process transmitter. With a push of a button, you can switch to zero and span values, auto ramp, and auto step throughout the range. Each loop calibrator has a large, easy to read screen which simultaneously displays the measurement with the % of span. (/(75,$/ 0($685( 63(,),$7,216 $FFXUDF WR RI UHDGLQJ 6PDOO DQG UXJJHG KDQGKHOG GHVLJQ 0HDVXUH VRXUFH RU VLPXODWH ORRS FXUUHQW 0HDVXUH ' YROWV 6LPXOWDQHRXVO P$ DQG VSDQ GLVSOD 6ZLWFK IXQFWLRQDOLW 6HOHFWDEOH UDPS DQG VWHS IXQFWLRQV (DV WR UHDG GLVSOD DQG XVHU LQWHUIDFH +$57 š UHVLVWRU LQ VHULHV ZLWK 9 ORRS 5DQJH 5HVROXWLRQ $'7 $FFXUDF $'7 $FFXUDF 9ROWDJH ' 0 to 30 V 1 mV 0.03%RD + 2mV 0.01%RD + 2mV XUUHQW ' 0 to 24 mA 1 μA 0.03%RD + 2μA 0.01%RD + 2μA 6ZLWFK WHVW +PRWV TGUKUVCPEG OQTG VJCP /š Trigger level: low level 0.3V; high level: 2V =? /š KPRWV TGUKUVCPEG =? .QQR VTCPUOKVVGT EWTTGPV OGCUWTG š OCZKOWO (/(75,$/ 6285( 63(,),$7,216 5DQJH 5HVROXWLRQ $'7 $FFXUDF $'7 $FFXUDF XUUHQW ' 0 to 24 mA 1 μA 0.03%RD + 2μA 0.01%RD + 2μA 6RXUFH PRGH š O# OCZKOWO 6LQN PRGH External loop voltage nominal 24V, maximum 30V, minimum 12V
  • 103. Pressure/ProcessCalibrationEquipmentAdditelCatalog2016 Look us up on www.additel.com or call today (1)714-998-6899 25'(5,1* ,1)250$7,21 $'7 $'7 0RGHO 1XPEHU 2SWLRQDO $FFHVVRULHV $FFHVVRULHV LQFOXGHG 9812 110V/220V external power adapter (DC 9V) 9024 Test lead set 1 set Alligator clips 2 pcs User manual 1 pc 9V Alkaline battery 1 pc *(1(5$/ 63(,),$7,216 9ROWDJH OLPLW 30V between terminals or between terminals and ground 0HDVXUHPHQW )XQFWLRQV Auto step, auto ramp, span step 'LVSOD VA LCD display. 2.04 x 2.04 in (52 x 52 mm) /RRS SRZHU 24V 2YHUYROWDJH SURWHFWLRQ 30V DC (240V AC) 2YHUORDG FXUUHQW SURWHFWLRQ 33 mA DC 6WRUDJH WHPSHUDWXUH -20°C to 70°C :RUNLQJ (QYLURQPHQW -10 to 50°C, 95%RH :RUNLQJ $OWLWXGH 3,000 m 9LEUDWLRQVKRFN Random 2G 5 to 500Hz 1 meter drop 3RZHU One 9V alkaline battery (ANSI/NEDA 1604A or IEC) DC9V optional adapter available %DWWHU OLIH WSLFDO
  • 104. 1WVRWV OQFG JQWTU O#š Measure mode: 50 hours 6L]H /[:[+
  • 105. 163 x 83 x 41 mm :HLJKW 350 g DOLEUDWLRQ HUWLILFDWLRQ 0+56VTCEGCDNG EGTVKſECVG QH ECNKDTCVKQP YKVJ FCVC RPSOLDQFH HUWLILFDWLRQ ADT209: CSA, CE, DNV ADT210: CSA, CE, DNV :DUUDQW 3 years
  • 106. Look us up on www.additel.com or call today (1)714-998-6899 0XOWLIXQFWLRQ /RRS DOLEUDWRU $GGLWHO 0HDVXUH DQG VRXUFH ORRS FXUUHQW P9 DQG 9 0HDVXUH DQG VRXUFH VLPXOWDQHRXVO /RRS RQWLQXLW7HVW FDSDELOLW 9 ORRS VXSSO ZLWK VLPXOWDQHRXV FXUUHQW PHDVXUHPHQW HDU ZDUUDQW 29(59,(: The 220 is a highly integrated loop calibrator featuring several patented technologies. The calibrator is an ultra-compact, rugged, and best of all, easy to use hand-held device that will source, simulate and measure loop current, mV/V, Loop Continuity Test, and also perform switch testing. Its smart phone-like menu and interface makes it simple and easy to WUG 6JG KU KFGCN HQT ECNKDTCVKPI OCKPVCKPKPI CPF VTQWDNGUJQQVKPI XCTKQWU NQQR FGXKEGU KP VJG ſGNF 0HDVXULQJ DQG VRXUFLQJ ORRS FXUUHQW P9 DQG 9 Measures loop current up to 30mA with 0.1μA resolution Measures voltage signals up to 60 V, and mV signals up to 300 mV Sources/sink loop current, mV and V 24V loop power supply with simultaneous current measurement Simultaneous dual reading capability Capture switch 'DWD VWDWLVWLFV DQG DQDOVLV Display max, min, peak-to-peak, average, and percentage simultaneously Convert data to a real physical value through re-scaling function 2QGHPDQG ORJJLQJ High-capacity storage with up to twenty thousand records On-line view, index, analyze, auto-curve, and list the data 9DULRXV VRXUFHV DQG FRQYHUWLQJ PRGH 5WRRQTV NKPGCT ƀQY XCNXG CPF UVGRRKPI UQWTEG Auto-source of ramping and stepping Gauge span examination $V DO RQVLWH FDOLEUDWLRQ WRROV Four on-site calibration tools preset: loop indicator, sensor/transmitter, signal isolator/converter, and limit alarm/switch 6XSSRUW : WUDQVPLWWHU FRQQHFWLRQ 0XOWL OLQJXDO LQWHUIDFH 'PINKUJ5KORNKſGF %JKPGUG RQYHQLHQW WRROV Calculator, thermocouple converter, HOLD screen lock 'LVSOD UDWH 4 readings/second (DV WR XVH Smartphone-like menu and interface make the operation simpler and easier Ultra-compact, size 3.9 x 7.6 x 2.0 (100mm x 192mm x 52mm), and weight 1.6 lb (0.7 kg) One hand operation 6QDSVKRW Save and manage up to 100 snapshots 'LVSOD 3.5 inch TFT color screen 4, 5, or 6 digits adjustable display 5XJJHG Rugged design for harsh environments. Passed a 1-meter drop test Three years warranty for the 220, and one year for the battery pack 0LVXVH SURWHFWLRQ DQG (OHFWULFDO LVRODWLRQ Up to 30V voltage on any two sockets and up to 1A current on current sockets will not damage the calibrator. The calibrator will return to normal condition as soon as the voltage or current is removed. Measuring, sourcing and loop power circuits are electrically isolated each other. 5HFKDUJHDEOH EDWWHU Rechargeable Li-ion battery for 15 hours uninterrupted use. Battery life will be reduced when 24V is applied. The rechargeable battery is replaceable. 1,677UDFHDEOH DOLEUDWLRQ ZLWK GDWD )LUPZDUH XSJUDGH 5WRRQTV ſTOYCTG WRITCFG YJGP CXCKNCDNG :DUUDQW 3 years )($785(6
  • 107. Pressure/ProcessCalibrationEquipmentAdditelCatalog2016 Look us up on www.additel.com or call today (1)714-998-6899 0XOWLIXQFWLRQ /RRS DOLEUDWRU $GGLWHO $33/,$7,216 63(,),$7,216 25'(5,1* ,1)250$7,21 The 220 loop calibrator is a process tool for measuring, sourcing and simulating mA, mV and V, captures switch values and provides 24V loop power. It is a high performance solution for calibration, repair and maintenance of various loops devices, Loop Continuity Test, and switch capture. ADT220 * Additel/Land software could be downloaded for free at www.additel.com 0HDVXUHPHQW $FFXUDF 5DQJH 5HVROXWLRQ $FFXUDF 9ROWDJH ' -300.000 to 300.000 mV 1μV 0.01%RD+15μV -60.0000 to 60.0000 V 0.1mV 0.01%RD+3mV XUUHQW ' -30.0000 to 30.0000 mA 0.1μA 0.01%RD+1.5μA /RRS RQWLQXLW 0 to 2000.0 ohm š 4 š 6ZLWFK7HVW For the contact with potential, the voltage within the range 3V to 24V. 6RXUFH $FFXUDF 9ROWDJH ' -10.00 to 200.00 mV 0.01mV 0.02%RD+10.5μV 0 to 12.000 V 1mV 0.02%RD+0.6mV XUUHQW ' 0 to 22.000 mA 1μA 0.02%RD+1.1μA '9 N/A N/A 0.5V (QYLURQPHQWDO 6SHFLILFDWLRQV 2SHUDWLQJ 7HPSHUDWXUH -10°C to 50°C 6WRUDJH7HPSHUDWXUH -20°C to 60°C +XPLGLW 90%, non-condensing 6DIHW 6SHFLILFDWLRQV (XURSHDQ RPSOLDQFH CE Mark 0HFKDQLFDO 6SHFLILFDWLRQV 'LVSOD 3.5 inch TFT color screen (OHFWULFDO RQQHFWLRQ «OO UQEMGVU CPF ƀCV OKPKLCEM VJGTOQEQWRNG socket 56 ,QWHUIDFH Standard RS232-DB9 socket 6L]H 3.9 x 7.6 x 2.0 (100mm x 192mm x 52mm) :HLJKW 1.6 lb (0.7 kg) 3RZHU 6XSSO Polymer Li-ion rechargeable battery,or 10V DC adaptor %DWWHU Rechargeable Li-ion battery (included) %DWWHU /LIH 15 hours uninterrupted use Battery life will be reduced when 24V is applied %DWWHU KDUJH 110V/220V external power adapter (included) (OHFWULFDO 6SHFLILFDWLRQV *HQHUDO 6SHFLILFDWLRQV 0RGHO 1XPEHU $FFHVVRULHV LQFOXGHG
  • 108. 9816-X 110V/220V external power adapter 1 pc 9712 Chargeable Li-ion battery 1 pc 9022 Test leads 2 sets(4 pcs) 9020 Short circuit cable 1 sets(2 pcs) User Manual 1 pc 0+56 VTCEGCDNG ECNKDTCVKQP EGTVKſECVG 1 pc 2SWLRQDO $FFHVVRULHV 0RGHO QXPEHU 'HVFULSWLRQ 9050 USB to RS232 (DB-9 Male) Adapter 9050-EXT RS 232 (DB9/M) extension cable, 9 feet 9712 Spare chargeable Li-ion battery for multifunction calibrator 9816 110V/220V external power adapter 9906 Carrying case for multifunction calibrator
  • 109. Look us up on www.additel.com or call today (1)714-998-6899 0XOWLIXQFWLRQ 'RFXPHQWLQJ 3URFHVV DOLEUDWRUV $GGLWHO $ $ $ 6RXUFLQJ VLPXODWLQJ DQG PHDVXULQJ SUHVVXUH WHPSHUDWXUH DQG HOHFWULFDO VLJQDOV 6PDUWSKRQHOLNH PHQX DQG LQWHUIDFH PDNH WKH RSHUDWLRQ VLPSOH +$57 RPPXQLFDWLRQ FDSDELOLW $
  • 110. 7KH LQWHUQDO FROG MXQFWLRQ FRPSHQVDWLRQ VHQVRU FDQ EH UHFDOLEUDWHG DW LFH SRLQW E XVHU 8OWUDFRPSDFW [ [ DQG OE NJ
  • 111. 29(59,(: This series of highly integrated multifunction calibrators feature several patented technologies. These are an ultra- compact, rugged, and easy to use hand-held device for sourcing, simulating and measuring pressure, temperature, and electrical signals. Their smartphone-like menu and interface make the operation simple. Automation and documentation capabilities make the these calibrators a turnkey solution. $GGLWHO $ The Additel 221A is very unique to the multifunction temperature calibrator market. Not only does it provide you with the ability to source, simulate and measure temperature and electrical functions but it also incorporates full documenting capability and many other solutions other products do not provide. As a standalone device, you can create tasks, run tests and store the results. With the use of Additel/Land Software or Additel/Cal software, all saved tasks and data can be downloaded and managed. With it's unique internal could junction compensation sensor, cold junction compensation is very simple and easy. $GGLWHO $ The Additel 222A Multifunction Process Calibrator takes all the functionality of the 221A and adds the ability to measure and source pressure using the ADT160A Intelligent Pressure Modules. With calibration characteristics programmed directly into the ADT160A pressure modules you can simply connect it to the 222A and it will automatically recognize and display the module as the source or measure pressure. For ultimate convenience, one 222A allows for modules to be hot swapped with a simple connection at the top of the module. $GGLWHO $ The 223A has all the capability of the 221A and 222A and HART Communication in one small, easy-to-use package. The 223A has a full HART library that allows for the reading of HART-smart devices and also the capability to write to devices. Combined with full task automation and documentation, the 223A is an ideal tool to accomplish many of your important tasks. 6RXUFLQJ VLPXODWLQJ DQG PHDVXULQJ WHPSHUDWXUH DQG HOHFWULFDO VLJQDOV Sources and measures mV, mA, ohms, RTDs, thermocouples, frequency, and pulses Simulates and measures 13 thermocouples and 11 RTDs to calibrate transmitters Measures and sources pressure using Additel 160A series Intelligent Digital Pressure Modules from -15 psi to 10,000 psi (-1 bar to 700 bar) 24V loop power supply Simultaneous dual reading capability Automatic switch test Supports square root transmitter 2WNUG HTGSWGPE[ QWVRWV HQT VJG ECNKDTCVKQP QH ƀQY VQVCNKGT (DV WR XVH Smartphone-like menu and interface make the operation simpler and easier Ultra-compact, size 3.9 x 7.6 x 2.0 (100mm x 192mm x 52mm), and weight 1.6 lb (0.7 kg) One hand operation DOLEUDWHG FROG MXQFWLRQ FRPSHQVDWLRQ 3DWHQWHG
  • 112. Cold junction equivalent block in the calibrator A calibrated PRT element with flexible leads is installed in the equivalent block for thermocouple cold junction compensation This PRT element can be pulled out from the calibrator and re-calibrated and corrected at ice point by users 'RFXPHQWLQJ DQG DXWRPDWHG SURFHGXUH FDSDELOLW Manage the information of the device under test. Set up automated calibration procedures, and 223A performs the test, calculates the errors, displays and/ or stores the results in the memory, and highlights the out-of-tolerance points. As-found and As-left functions allow recording and documenting results for quality control. Download tasks and upload the results. Snapshots allow you to capture and save work. 1,677UDFHEOH DEOH ZLWK GDWD )($785(6